NUCLEAR POWER CORPORATION OF INDIA LIMITED (A Govt. of India Enterprise)

Similar documents
S. GURESH 4 JAN 2017 S. JOHNSON 4 JAN 2017

Standard Practice for Ultrasonic Examination of Turbine and Generator Steel Rotor Forgings 1

This British Standard, having been prepared under the direction of the Iron and Steel Standards Policy Committee, was published under the authority of

PROPOSED CHANGES TO: APPENDIX IV - PHASED ARRAY E-SCAN AND S-SCAN MANUAL RASTER EXAMINATION TECHNIQUES

American Petroleum Institute Task Group Line Pipe

Optimized Semi-Flexible Matrix Array Probes for Large Rotor Shafts and DGS Sizing Diagram Simulation Tool

DESIGN & VALIDATION OF A SEMI-FLEXIBLE PAUT PROBE FOR THE MANUFACTURING INSPECTIONS OF LARGE FORGED ROTORS

Engineering Policy & Procedure

BASICS ULTRASONIC TESTING METHOD. -Dr.Oruganti Prabhakar Proprietor-OP-TECH

TB/T Translated English of Chinese Standard: TB/T RAILWAY INDUSTRY STANDARD

RELIABILITY OF GUIDED WAVE ULTRASONIC TESTING. Dr. Mark EVANS and Dr. Thomas VOGT Guided Ultrasonics Ltd. Nottingham, UK

EMAT Application on Incoloy furnace Tubing Ramamohan Reddy M (ASNT Level III UT, PCN Level III UT,PAUT&TOFD)

Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems 1

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Ultrasonic testing Characterization of indications in welds

Annexure - V. Specifications, Typical Drawing & Data Sheet for Split TEE (Hot Tapping Material)

GB/T Translated English of Chinese Standard: GB//T NATIONAL STANDARD OF THE

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic thickness measurement

GUIDELINES FOR THE APPLICATION OF TIME-OF-FLIGHT DIFFRACTION (TOFD) AND PHASED ARRAY ULTRASONIC TESTING (PAUT) TECHNIQUES

RADIOGRAPHIC EXAMINATION PROCEDURE

NB/T Translated English of Chinese Standard: NB/T

NDI Techniques Supporting Steel Pipe Products

Structural UT: Variables Affecting Attenuation and Review of the 2 db per Inch Model

2010 ULTRASONIC BENCHMARKS

BINDT Telford. Guided Wave Testing and Monitoring Over Long and Short Ranges

Ripple and Uniformity Measurement of a Phased-Array Testing-Machine for round-bar Testing

Standard Practice for Measuring Thickness by Manual Ultrasonic Pulse-Echo Contact Method 1

Phased Array UT Application For Boiler Tube Inspection in Manufacturing And In-Service Anandamurugan S 1, Siva Sankar Y 2

Fluid Sealing Association

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

ASNT Level III Study Guide: Ultrasonic Testing Method, second edition Errata 1st Printing 09/13

Spectral Distance Amplitude Control for Ultrasonic Inspection of Composite Components

Pipeline Technology Conference 2010

Modelling Probe Wedge and Pipe Geometry as Critical Parameters in Pipe Girth Weld Ultrasonic Inspections Using Civa Simulation Software

A PHASED ARRAY ULTRASONIC TESTING OF A MANUAL THICK AUSTENITIC WELD FEEDBACK

In-Line EMAT Ultrasonic Weld Inspection for ERW Tube Mill Using Guided Ultrasonic Waves

UNIVERSAL ULTRASONIC FLAW DETECTOR UD4-76. TOFD version.

CRACK DETECTION AND DEFECT CLASSIFICATION USING THE LLT - TECHNIQUE. Wolfgang Gebhardt and Friedhelm Walte

PAUT as Tool for Corrosion Damage Monitoring

English version. Steel and steel products Location and preparation of samples and test pieces for mechanical testing (ISO 377:1997)

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

Long Range Ultrasonic Testing - Case Studies

Simulation of Ultrasonic Testing of Rail Wheel Face using Phased Array and DDF technique

Ultrasonic Phased Array Crack Detection Update

Phased Array&TOFD Probes

0.20. Record Page 1 of 19

Two aging problems were discovered in the mid 90`s: thinning and cracking.

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

Specifications for EA31002 U Tube Bundle for CDU 3

Pipeline & Specialty Services (P&SS)

ULTRASONIC QUALIFICATION CARD GUIDANCE INSTRUCTIONS

Nondestructive Testing (CALIBRATION)

Kerkau Manufacturing. B16.5 Flange Book

Improved Inspection of CRA-Clad Pipeline Girth Welds with the Use of Accessible Advanced Ultrasonic Phased-Array Technology

18th World Conference on Non-destructive Testing, April 2012, Durban, South Africa

PIPING NOMENCLATURE. Header Branch connection Valve Flange Expansion joint Expansion loop Pipe support Reducer Elbow

Ultrasonic Examination Of Tube To Tube Sheet Weld Joints In High Pressure Heat Exchanger For The Urea Process

Fig.2: Scanner VistaScan for image plates

A COMPARISON BETWEEN ASTM E588 AND SEP 1927 RELATING RESOLUTION LIMITS AT DETERMINATION OF THE PURITY GRADE

Part 1: General principles

TESTING OF HEXAGONAL AND OTHER EDGE PROFILED MATERIALS BY USING ULTRASONIC PHASED ARRAY PROBES ARRANGED ANNULARLY

MULTI-CHANNEL ULTRASONIC FLAW DETECTOR ОКО-22М-UT THE BEST INDUSTRIAL OEM SOLUTION FOR IN-LINE AND IN-SERVICE SYSTEMS

FLOW SWITCH 600 Series Velocity Flow Sensor. Instruction Manual

DETECTION OF CORROSION IN BOTTOM PLATES OF GAS AND OIL TANKS USING GUIDED ULTRASONIC WAVES AND ELECTROMAGNETIC ULTRASONIC (EMAT) TRANSDUCERS

Curved arrays for improved horizontal sizing in small pipe welds

ULTRASONIC FLAW DETECTORS A1212 MASTER, A1214 EXPERT OPERATION MANUAL

Reference Standards Calibration Blocks

Model R7900. Instruction Manual. Ultrasonic Thickness Gauge. reedinstruments. www. com

INTRODUCTION. Strong Performance: High resolution and penetration, achieving precise flaw detection

1112. Dimensional evaluation of metal discontinuities by geometrical parameters of their patterns on imaging flaw detector monitor

Considerations on Linear Phased Array transducers with Circular Crystals ECNDT Prague 2014

Cone type Phased Array Design for High Speed Hollow Axle Inspection

PHASED ARRAYS FOR SMALL DIAMETER, THIN-WALLED PIPING INSPECTIONS

METAL FABRICATION SECTION 8: METAL FABRICATION SCOPE MATERIAL.

QUALITY ASSURANCE SPECIFICATION

Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars

KAERI Feeder Tube Inspection Using EMAT Generated Circumferential Guided Waves

Multi-Parameter POD for Industrial Applications The Influence of the Material Attenuation as an Example

Probability of Rejection - In conformance with DNV OS F101

MILITARY SPECIFICATION LIGHTING, INSTRUMENT, INTEGRAL, WHITE GENERAL SPECIFICATION FOR

Application of Ultrasonic Guided Waves for Characterization of Defects in Pipeline of Nuclear Power Plants. Younho Cho

ISO INTERNATIONAL STANDARD. Non-destructive testing Acoustic emission inspection Secondary calibration of acoustic emission sensors

Basic functions of the universal flaw detector GEKKO

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

T 519 om-96 SUGGESTED METHOD 1970 OFFICIAL STANDARD 1978 OFFICIAL TEST METHOD 1980 REVISED 1986 REVISED 1991 REVISED TAPPI

American Institute of Timber Construction 7012 South Revere Parkway Suite 140 Centennial, CO Phone: 303/ Fax: 303/

Sonotron NDT 4, Pekeris str., Rabin Science Park, Rehovot, 76702, Israel Phone:++972-(0) Fax:++972-(0)

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION

CNC Cooltool - Milling Machine

New Multi-Technology In-Line Inspection Tool For The Quantitative Wall Thickness Measurement Of Gas Pipelines

Principles and Applications of Air-Coupled Ultrasonics. Joe Buckley, Sonatest Plc

Testing of friction properties of fasteners

High Frequency Ultrasonic Systems with Frequency Ranges of 35 to 200 MHz

ULTRASONIC FLAW DETECTOR +TOFD VERSION. CE MARKING EN Compliant

OPERATING PAVEMENT PROFILOGRAPH AND EVALUATING PROFILES

Standard Practice for Ultrasonic Examinations Using Electromagnetic Acoustic Transducer (EMAT) Techniques 1

Encoding and Code Wheel Proposal for TCUT1800X01

Rev 2.0 September 2010 Copyright Lecoeur Electronique corporation all rights reserved -

ULTRASONIC GUIDED WAVE FOCUSING BEYOND WELDS IN A PIPELINE

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic inspection Evaluating electronic characteristics of ultrasonic test instruments

Welding Use of time-of-flight diffraction technique (TOFD) for examination of welds

Transcription:

NUCLEAR POWER CORPORATION OF INDIA LIMITED (A Govt. of India Enterprise) DIRECTORATE OF QUALITY ASSURANCE NO.: PP-P-2014 REV. NO. 2 TITLE FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE REVISION NO. 0 1 2 DATE OF ISSUE (Month/Year) TOTAL NO. OF PAGES (Including cover sheet) March 2001 May 2007 15 18 18 ORIGINAL PREPARED BY NAME SIGNATURE DATE S.S. Som SSE(QA), KAPS Sd/- March 2001 REVIEWED BY APPROVED BY M. Mohan Babu, ACE (ISI) C. Awasthi, ED (QA) March 2001 March 2001

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 1 of 17 REVISION CONTROL SHEET NO.: PP-P-2014 REV. NO. 2 FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Rev. No. & date Description of Revision Revised By Reviewed By Approved By 1 May 2007 2 Feb. 2013 Editorial changes and changes as per latest codes/ standards Editorial changes and changes as per latest codes/ standards P.V. Khambatkone CE (QA) A.K. Palit, CE(QA) A. K. Palit, CE (QA) M. Mohan Babu CE (ISIS), QA A. K. Palit CE (QA) M. Mohan Babu CE (ISIS), QA K. P. Dwivedi ED (QA)

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 2 of 17 1.0 SCOPE This procedure describes the requirements for ultrasonic examination of steel forgings by pulse-echo, manual contact method and is applicable to forged flanges, bars, and pipe fittings such as Elbows, Tees and Couplings. Forgings and forged or rolled bars which are to be bored to form tubular products shall be examined after boring. 2.0 REFERENCES This procedure is based on the following: a) ASME Section-V, Article 5, (2010 Edition). b) ASME Section - III, Subsection, NB, NC (2010 Edition). c) SA-388, Recommended practice for ultrasonic examination of steel forgings, (2010 Edition). d) PP-P-1981 Rev. 2, General requirements and Formats for Ultrasonic Examination procedure. 3.0 SCAN DIRECTIONS Following scans are applicable for the forgings of following types a) Solid cylindrical sections: Radial scan (Clause 3.1) and Axial scan (Clause 3.2.1 & 3.2.2 as applicable). b) Hollow cylindrical sections: Circumferential scan (clause 3.3) and axial scan (clause 3.2.2). c) Disk forging: Radial scan from the circumference (clause 3.1) and axial scan from flat face (clause 3.2.1). d) Rectangular forgings: Normal beam examination in accordance to the technique mentioned in clause No. 3.1. 3.1 Radial Scan A radial normal beam scan using longitudinal waves shall be conducted. Scan surfaces shall include all solid cylindrical surfaces of the object. This scan shall be carried out by the Distance Amplitude Correction Curve (DAC) method using reference flat bottom hole standard. 3.2 Axial scan 3.2.1 Normal beam examination An axial normal beam scan using longitudinal waves shall be carried out, if possible, from both ends of the forging using the flat bottom hole method. 3.2.2 Angle beam examination An axial angle beam scan using shear wave shall be carried out in place of axial normal beam scan for cylindrical sections of long length due to attenuation and also for ring and hollow forgings having an outside diameter to inside diameter ratio of less than 2 to 1 which can not be examined axially using a straight beam in axial direction.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 3 of 17 3.3 Circumferential scan A circumferential angle beam scan using shear wave from the OD surface shall be carried out for ring & hollow forgings that have a ratio of outside diameter to inside diameter of less than 2 to 1. Use of 35 probe is desirable when the OD / ID ratio is 1.75 to 1 minimum. 4.0 INSTRUMENTATION 4.1 Ultrasonic instrument Pulse-echo type instruments shall be used. The amplitude linearity, amplitude control linearity and time base linearity shall meet the requirements defined in Annexure 4. These parameters shall be checked at the time of commencement of the examination unless valid records less than six months old exist proving conformance and no repair of instrument has been necessitated. Witnessing personnel may call for rechecks of these parameters at the start of examination or at any time during examination, if there is reason to doubt instrument performance in these regards. Following table is suggestive of make & model of some UT machines: Sr. Models Models 1 M/s. Roop Ultrasonix, India. 4400, 440A, 4400M 2 M/s. Electronic and Engineering Co., India ESM-2, 2M, EX-10,100 3 M/s Modsonic Einstein 4 M/s. Krautkramer, Germany, USA USIP11, USD10, USL 48 5 M/s. Karl Deutsch, Germany 1023, 1024, 1025, 1030 6 M/s. Panametrics, USA. Epoch-2000, Epoch-II 7 M/s. Sonic, USA. FTS MK-I, FTS MK-IV However, Head QA / Supdt. QA of applicable unit shall have the authority to permit use of any other equipment which meets the requirements of Annexure 4. 4.2 Search Units / Probes The probes recommended for particular application, are as follows: 4.2.1 Normal Beam probe a) Size: A maximum active are of 650 mm 2 with 13 mm minimum to 30 mm maximum dimension or 19 mm diameter minimum dimension shall be used. b) Frequency: Nominally probes of 2MHz to 5 MHz shall be used. Whenever necessary, to compensate for attenuation, 0.5 MHz probes may be used.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 4 of 17 4.2.2 Angle beam probe a) Size: Probes equipped from 12 by 25mm to 25 by 25 mm transducers shall be used. b) Beam Angle: 45 o angle beam search unit shall be used wherever applicable. c) Frequency: The nominal test frequency shall be 2 to 5 MHz. The exit point of the sound beam and the actual refracted beam angle of angle beam search unit shall be determined on an IIW reference block. The permissible error on either side of the nominal angle shall be 2 o and the accuracy of measurement shall be 1 o 5.0 PERSONNEL QUALIFICATIONS Scanning shall be carried out by personnel qualified by the ISNT or the ASNT to UT Level I or higher. Verification of calibration, scanning, analysis of indications and reporting shall be by personnel qualified by the ISNT or the ASNT to UT Level II or higher. The Head of the QA Unit may conduct evaluation tests to check the theoretical and practical capability of the personnel performing UT. 6.0 GENERAL REQUIREMENTS 6.1 Surface Conditions The surface of the forging to be examined shall be free of loose scale, paint, dirt etc. The surface roughness of scanning surfaces shall not exceed 6.35 µm. 6.2 Couplant Oil, grease, glycerine, water or cellulose paste shall be used as couplant. Select a suitable couplant based on the surface conditions and orientation; roughness and nonhorizontal conditions require a more viscous couplant. Any requirements for protection of the material under test, such as for sulphur and halogen content, shall be met by the couplant. 6.3 Scanning Speed The maximum rate of probe (normal beam search unit) movement shall be established by moving the probe on the calibration block for which the amplitude from FBH reflector shall be at least 50% of the PLR.The maximum rate of probe (angle beam search unit) movement shall be established by moving the probe on the calibration block for which the amplitude from reflector notch shall be at least 50% of the reference amplitude when picked up at a distance of full vee path for outer notch and be not less than 60% of the reference amplitude when at a distance of ½ vee path for inside notch. However the scanning speed shall not exceed 150 mm / sec. 6.4 Probe Overlap To assure complete coverage of the forging volume, the overlap shall be at least 10% of the probe diameter. 6.5 Calibration Identify and Validity

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 5 of 17 Any change of instrument or control settings (other than temporary changes of controls which operate on fixed steps), probe, couplant or probe cables shall require re-calibration. Operating variables including the test instrument, its control settings, its power source, probe, probe cable and couplant shall be unchanged once calibrated throughout scanning and evaluation, except for readjustments of gain between evaluation and scan as required during examination. After initial calibration, rechecks the proper functioning of the equipment shall be checked by the use of the reference specimens, as a minimum - 6.6 Reject a) At the beginning of each production run of a given size and thickness of the given material. b) After each 4 hours or less during production run. c) At the end of the production run. This control is also called suppression in some instruments and shall be maintained in the minimum position (or OFF position, if available) throughout the examination. 6.7 Continuously Variable Controls Continuously variable controls such as damping and continuously variable gain shall be maintained at either of the extremes of their positions to the extent possible to avoid inadvertent changes. 7.0 RADIAL SCAN (For Solid Cylindrical Forgings and for Disc Forgings) This scan shall be carried out by normal probe at all cylindrical surfaces of the test object. 7.1 Calibration Blocks Calibration blocks shall be of the same material specification and heat treatment as applicable to the test object. However, where the actual forging material belongs to P1 (Refer ASME Sec. IX), use of calibration block made of mild steel is permitted, provided the block is made from forged material. The surface finish at the entry surface of the calibration block should be similar to that of the test object. Prior to machining the block, the raw material shall be ultrasonic examination checked and shall be adequately defect free. The hole diameter shall be 3% of the thickness and minimum 3 mm diameter. The holes shall be machined truly flat bottomed, preferably by end milling. The diameter of the calibration blocks and block length shall be as shown in the Table of Annexure 2. 7.2 Time Base Calibration Basis Mechanically measure the test object diameter or thickness as applicable within 1 mm accuracy. Select a round figure for the required range such as 0 to 50, 0 to 100 or 0 to

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 6 of 17 200 mm etc. The range shall be such that the first back wall echo is between 60% and 90% of the time base. Calibration on reference block such as V1 block is accepted for the cases where job configuration does not give back wall echo. 7.3 Time Base Calibration Procedure Calculate the time base position for the first back wall echo. Set the time base controls such that the second back wall echo is set to this value, with the first back wall echo being set to zero. Now, lock all the time base controls other than delay. From this setting, use the delay control to shift the first back wall echo to the calculated position. The time base is now calibrated, lock delay also. Confirm calibration from the position of echoes of other smaller diameter / thickness area wherever possible and correct any errors if necessary. 7.4 Sensitivity Calibration 7.4.1 Setting the primary reference level At a suitable gain setting find out which hole provides the highest response. Set the amplitude of this echo to between 70% FSH and 80% FSH using the gain control. This is the reference gain and should not be changed. 7.4.2 Drawing the DAC curve Use the instrument screen as a graph. At the metal path corresponding to the first hole (which gives highest response), mark the maximized amplitude of the echo. Repeat this for all the other holes. Draw a smooth line through the points so obtained. This is the distance amplitude correction (DAC) curve. Use a suitable marking pen which should not damage the screen protection sheet. 7.5 Scan, Recording and Evaluation Scan at a gain value higher than the above reference gain by 6 db. Recording & Evaluation shall be done at reference gain for - a) Any indication whose amplitude equals or exceeds 50% of the established DAC curve,. b) Travelling, clustered or continuous indications that have amplitudes equal to or greater than 25% of the DAC curve. c) any reduction in back reflection (associated with a discontinuity indication) exceeding 20% of the calibration back reflection. Evaluate the amplitude of each indication by comparison with the height of the DAC curve at the same metal path as the indication and report it as a percentage of DAC. 7.6 Acceptance Level Indications of following nature shall cause for rejection - a) Any indication whose amplitude exceeds 100% of the established DAC curve, b) Travelling, clustered or continuous indications that have amplitudes equal to or greater than 50% of the DAC curve,

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 7 of 17 c) Reduction in black reflection exceeding 50% not associated with geometric configuration are unacceptable. 8.0 CIRCUMFERENTIAL SCAN (For Ring Forgings and Hollow Forgings) This scan shall be carried out by an angle probe on the OD surface of the test object. 8.1 Calibration Block Calibration blocks shall be of the same material specification and heat treatment as applicable to the test object. However, where the actual forging material belongs to P1 (Refer ASME Sec. IX), use of calibration block made of mild steel is permitted, provided the block is made from forged material. The surface finish at the contact surface of the calibration block should be similar to that of the test object. Prior to machining the block, the raw material shall be ultrasonic examination checked and shall be adequately defect-free. The block shall contain rectangular notches on the ID and OD surface in the axial direction and parallel to the axis of forging. The notch depth shall be 3% maximum of the thickness or 6 mm, whichever is smaller and its length shall be 25 mm. The width of the notch shall not be greater than twice its depth. A sketch of the notch is provided as Annexure 3. 8.2 Time Base Calibration Basis The screen range chosen shall be the shortest metal path distance to include at least 1½ skip distance for the material to be examined. Select a round figure for the required range such as 0 to 20, 0 to 25, 0 to 50 etc. Calibration on reference block such as the V1 block is accepted for cases where the job configuration does not produce a back wall echo. 8.3 Time Base Calibration Procedure Calculate the time base position for the ID notch and OD notch. Using delay control and time-base control, set the echoes from ID and OD notch to their respective position on the screen. 8.4 Sensitivity Calibration At the metal path corresponding to ID notch, obtain an indication amplitude of approximately 75% FSH at a suitable gain setting and mark it on the screen. This is the reference gain and should not be changed with above gain setting. Mark the maximized echo amplitude from OD notch corresponding to its metal path on the screen. Draw a smooth line through the points so obtained. This is the distance amplitude correction (DAC) curve. 8.5 Scan, Recording & Evaluation Scan at a gain value higher than the above reference gain by 6 db. Perform the examination by scanning over the entire surface area circumferentially in both clockwise and counter clock wise directions from the OD surface.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 8 of 17 Recording & Evaluation shall be done at reference gain for any indication that has amplitude equal to or greater than 50% of the DAC curve. 8.6 Acceptance Level Any indication above 100 % DAC shall cause for rejection. 9.0 AXIAL SCANNING (for Solid Cylindrical, Disk and Hollow Forgings) This scanning shall be carried out by a normal probe by flat bottom hole technique or by an angle probe as per applicability mentioned in the clause No. 3.0 9.1 Angle Beam Scanning (for Hollow Forgings) 9.1.1 Calibration block Calibration blocks shall be of the same material specification and heat treatment as applicable to the test object. However, where the actual forging material belongs to P1 (Refer ASME Sec. IX), use of calibration block made of mild steel is permitted, provided the block is made from forged material. The surface finish at the entry surface of the calibration block should be similar to that of the test object. Prior to machining the block, the raw material shall be ultrasonic examination checked and shall be adequately defect free. The block shall contain rectangular notch on the ID and OD surface in the direction perpendicular to the axis of forging. The notch depth shall be 3% maximum of the thickness or 6 mm, whichever is smaller and its length 25 mm. The width shall not be greater than twice its depth. 9.1.2 Time base calibration basis The screen range chosen shall be the shortest metal path distance to include at least 1½ skip distance for the material to be examined. Select a round figure for the required range such as 0 to 20, 0 to 25, 0 to 50 etc. Calibration on reference block such as V1 block is accepted for the cases where job configuration does not give back wall echo. 9.1.3 Time base calibration procedure Calculate the time base position for the ID notch and OD notch. Using delay control and time-base control, set the echoes from ID and OD notch to their respective position on the screen. 9.1.4 Sensitivity calibration At the metal path corresponding to ID notch, obtain an indication amplitude of approximately 75% FSH at a suitable gain setting and mark it on the screen. This is the reference gain and should not be changed with above gain setting. Mark the maximized echo amplitude from OD notch corresponding to its metal path on the screen. Draw a smooth line through the points so obtained. This is the distance amplitude correction (DAC) curve. 9.1.5 Scanning, Recording & Evaluation

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 9 of 17 Scanning shall be done axially from both end of the test object. Scan at a gain value higher than the above reference gain by 6 db. Perform the examination by scanning over the entire surface area circumferentially in both clockwise and counter clock wise directions from the OD surface. Evaluation shall be done at reference gain for any indication that has amplitude equal to or greater than 50% of the DAC curve. 9.1.6 Acceptance level Any indication above 100 % DAC shall be the cause for rejection. 9.2 Normal Beam Scanning (for Cylindrical Forgings and Disk Forgings) 9.2.1 Calibration Block Calibration blocks shall be of the same material specification and heat treatment as applicable to the test object. However, where the actual forging material belongs to P1 (Refer ASME Sec. IX), use of calibration block made of mild steel is permitted, provided the block is made from forged material. The surface finish at the entry surface of the calibration block should be similar to that of the test object. Prior to machining the block, the raw material shall be ultrasonic examination checked and shall be adequately defect free. Prior to machining the block, the raw material shall be ultrasonic examination checked and shall be adequately defect free. The hole diameter shall be 3% of the thickness and minimum 3 mm diameter. The holes shall be machined truly flat bottomed, preferably by end milling. The diameter of the calibration blocks and block length shall be as shown in the Annexure 2. 9.2.2 Time base calibration basis Mechanically measure the test object diameter or thickness as applicable within 1 mm accuracy. Select a round figure for the required range such as 0 to 50, 0 to 100 or 0 to 200 mm etc. the range shall be such that the first back wall echo is between 60% and 90% of the time base. Calibration on reference block such as V1 block is accepted for the cases where job configuration does not give back wall echo. 9.2.3 Time base calibration procedure Calculate the time base position for the first back wall echo. Set the time base controls such that the second back wall echo is set to this value, with the first back wall echo being set to zero. Now, lock all the time base controls other than delay. From this setting, use the delay control to shift the first back wall echo to the calculated position. The time base is now calibrated, lock delay also confirm calibration from the position of echoes of other smaller diameter / thickness area wherever possible and correct any errors if necessary. 9.2.4 Sensitivity 9.2.4.1 Setting the primary reference level

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 10 of 17 At a suitable gain setting find out which hole provides the highest response. Set the amplitude of this echo to between 70% FSH and 80% FSH using the gain control. This is the reference gain and should not be changed. 9.2.4.2 Drawing the DAC curve Use the instrument screen as a graph. At the metal path corresponding to the first hole (which gives highest response), mark the maximized amplitude of the echo. Repeat this for all the other holes. Draw a smooth line through the points so obtained. This is the distance amplitude correction (DAC) curve. Use a suitable marking pen which should not damage the screen protection sheet. 9.2.5 Scanning, Recording & Evaluation Scan at a gain value higher than the above reference gain by 6 db. Evaluation shall be done at reference gain for - Any indication whose amplitude equals or exceeds 50% of the established DAC curve. 9.2.6 Acceptance level Indications of following nature shall cause for rejection- Any indication whose amplitude exceeds 100% of the established DAC curve. 10.0 REPORTING All relevant data of the product, Ref document details, Calibration data and examination results shall be entered in the Proforma enclosed in Annexure 1. 11.0 POST EXAMINATION CLEANING After the examination the test surface shall be thoroughly cleaned free from grease, oil etc using dry cloth, acetone etc.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 11 of 17 ULTRASONIC EXAMINATATION REPORT ANNEXURE 1 REPORT NO.:. DATE OF TESTING : REFERENCE NO.: REFERENCE CODE NO.: MATERIAL DESCRIPTION WORK ORDER NO. : MATERIAL : IDENTIFICATION: INSTRUMENT DESCRIPTION DIMENSIONS: SURFACE : MACHINED/ROLLED/WELDED CONDITION INSTRUMENT USED : MACHINE No.: MAKE: SEARCH UNIT : COUPLANT : OIL / WATER CALIBRATION RANGE CALIBRATION: CALIBRATION : STANDARD METHOD & TECHNIQUE: SENSITIVITY: EXAMINATION SCANNING: EXAMINATION STARTED AT : HRS. CLOSED AT : HRS CALIBRATION 1. Hrs 2. Hrs 3. Hrs RECHECKED AT OBSERVATION OF RECALIBRATION CHECK 1. 2. 3. OBSERVATION : EVALUATION :

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 12 of 17 ANNEXURE 2 B C D STANDARD REFERENCE BLOCK (For Axial Scanning) Note: For Radial Scanning, the hole should be drilled radially on the circumference of the Reference Block.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 13 of 17 DIMENSIONS OF REFERENCE BLOCKS Hole diameter (mm) Metal distance B (mm) Overall length C (mm) Diameter D (mm) 3 25 45 50 3 50 70 50 3 75 95 50 3 150 172 75 3 250 273 100 3 B B+ 19 125 NOTE B FOR ADDITIONAL SUPPLEMENTAL BLOCKS FOR THICKNESS > 250mm.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 14 of 17 Notch Details ANNEXURE 3 Depth Width Length

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 15 of 17 ANNEXURE 4 (Appendix 1 to PP-P-1981, Performance Checks of UT Instrument) A. SCOPE This Appendix describes the procedures to be followed and provides the requirements to be met to check and confirm that an ultrasonic test instrument is performing adequately, if such a check is required by a Specification or Procedure. In using the following directions, indications may need to be maintained on the screen over a length of time, with assurance that there is no change in received echo energy. An immersion fixture with an immersion probe can be very convenient for maintaining this constancy over this period. However, use of such a fixture is not mandatory, though then, care has to be taken that changes do not take place and a recheck shall particularly be done with the old control settings. If the performance of the instrument does not meet the requirements, it shall be serviced and all performance parameters re-evaluated. Such servicing is not within the purview of UT personnel, but must be conducted by the manufacturer, their authorized agents or other electronic servicing personnel of adequate electronic capability. B. AMPLITUDE LINEARITY This parameter is also called Screen Height Linearity. Use any convenient angle beam probe. Set the Suppression or Reject control to minimum (or OFF, if available) which value shall be maintained undisturbed further. Pick up two simultaneous indications from ½ and 3/4T side drilled holes in any convenient test piece or block which can adjusted such as to provide a ratio of 1:2 between indication heights. That is, one echo must be twice the height of the other. It is immaterial which echo appears first on the time-base. In order to obtain this condition, one method is to obtain a stepped block (i.e. one with two distinct thicknesses) and scan it from the un-stepped flat side. On scanning the zone of thickness change, two echoes, backwall echoes from the two thicknesses, will be obtained. By selecting an appropriate probe position, the two echoes can be set at any amplitude ratio to each other, in this case, 1:2 or 2:1. Using Gain, Damping and Filter or any combination of these controls, set the amplitude of the bigger of the two echoes to 80 % full-screen-height. Without permitting probe movement, re-adjust any of the above controls to change the amplitude of the larger of the two echoes successively from 100 % to 20 % in steps of 10 % full-screen-height. Record the amplitude of the smaller echo in each case. This value must be within 5 % full-screen-height of half the bigger echo, rounded off to the nearest 1 %. For example, for the case of the bigger echo being at 60 %, the smaller echo must be within the range of 30 % full-screen-height, ± 5 %. Values outside these call for instrument servicing. C. AMPLITUDE CONTROL LINEARITY Use any convenient probe. Set the Suppression or Reject control to minimum (or OFF, if available) which value shall be maintained undisturbed further. Pick up

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 16 of 17 indication from from ½T side drilled in any convenient test piece or block which can readily be maintained stable. Using Gain, Damping and Filter or any combination of these controls, set the amplitude of this echo to 80 % full-screen-height. Only the Gain control or controls shall be adjusted further. Without permitting probe movement, change the Gain by ± 6 db and ± 12 db and record the amplitudes. The table below gives the requirements: Initial amplitude set at % of full screen Gain Control Change Acceptance Limits of Amplitude % of full Screen 80 % - 6 db 32 to 48 % 80 % - 12 db 16 to 24 % 40 % + 6 db 64 to 96 % 20 % + 12 db 64 to 96 % The settings and readings shall be estimated to the nearest 1% of full screen This check shall be done starting with many representative values of Gain, such that the performance of the Gain control system is evaluated over the entire range of gain control. Where more than one calibrated Gain control is available, such as two separate knobs, each of them shall be checked over their range of control. In order to check the gain controls at various ranges, other controls such as Damping and Filter may be set at different combinations at the start of each test sequence of varying the Gain by ± 6 and ± 12 db. The values of 6 and 12 db may be obtained by combining two of the controls. When not using the DGS method, depending on the application, it may be adequate to limit the check to the general range of values in use during actual scans. D. INSTRUMENT TIME -BASE LINEARITY This parameter shall be checked at or near range values corresponding to those normally in use. In the following descriptions, the percentage values refer to the time-base, 0 % and 100 % being the left and right hand extreme graduations. Select a block of uniform thickness of approximately one-fifth of the range to be checked. Use any convenient normal beam probe, and suitable settings for control settings other than those related to the time-base. Calibrate the instrument time-base to five times this thickness, by positioning the second back-wall echo to 40 % and third back-wall echo to 60 % of the screen. This would require adjustments to the Range, Material Calibration and Delay controls or their equivalent. Lock all of these controls other than Delay.

FOR ULTRASONIC EXAMINATION OF STEEL FORGINGS FOR GENERAL PURPOSE Page 17 of 17 Using Delay, shift the second back-wall echo to a selected screen division and record the position of the next echo, i.e. the third back-wall echo. Selected screen divisions, in turn, shall be zero, 20 %, 40 % and 80 %. The expected reading shall, in each case be obviously 20 % larger than the selected screen division. In each case the permissible error from expected reading shall be within 2 % of the time-base. In instruments with a total of 50 divisions for the time-base, this will correspond to one such small division. In many instruments, the display does not extend beyond the final time-base mark. In such cases, the check of the last point shall be done in reverse, by shifting the third back-wall echo to the last mark and reading the position of the second back-wall echo. Specifically in the case of the initial or last portions of the time-base twice the error of 2 %, i.e. 4 % may be permitted. This however, is conditional on checks of the 5 to 25 % or the 75 to 95 % or both respectively, additionally being done and the results being within 2 %, (in instruments with 2% as the small division size, these values may be changed to 6 to 26 % and 76 to 96 % respectively).