Technical Basis Document (TBD) and User Guides

Similar documents
The ACT External HEPA Push-Through Filter Assembly. A. A. Frigo, S. G. Wiedmeyer, D. E. Preuss, E. F. Bielick, and R. F. Malecha

IMU integration into Sensor suite for Inspection of H-Canyon

Hardware-in-the-Loop Testing of Wireless Systems in Realistic Environments

High-]FrequencyElectric Field Measurement Using a Toroidal Antenna

Detector And Front-End Electronics Of A Fissile Mass Flow Monitoring System

National Accelerator Laboratory

Y f OAK RIDGE Y4 2 PLANT. Lionel Levinson General Electric Company. November 24, Approved for Public Release; distribution is unlimited.

Report on Ghosting in LL94 RAR Data

J.C. Courtney Nuclear Science Center Louisiana State University Baton Rouge, LA

High Explosive Radio Telemetry System. Federal Manufacturing & Technologies. R. Johnson, FM&T; B. Mclaughlin, FM&T;

STP-NU ROADMAP TO DEVELOP ASME CODE RULES FOR THE CONSTRUCTION OF HIGH TEMPERATURE GAS COOLED REACTORS (HTGRS)

U.S. DEPARTMENT OF ENERGY. YlAMT-485 Y-I 2. Project Accomplishment Summary for Project Number 92-Y12P-013-B2 HYDROFORMING DESIGN AND PROCESS ADVISOR

Field Testing of a Portable Radiation System

Radio Frequency Current Drive for Small Aspect Ratio Tori

PHOTOMULTIPLIER TUBE SATURATION INDICATOR. Jeffery F. Ruch P.O. Box 79 West Mifflin, PA. David J. Urban. West Mifflin, PA DISCLAIMER

Utilizing OSIsoft Visualizing PI Data System for Tank Level Data

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

AN IN-LINE POWER MONITOR FOR HE11 LOW LOSS TRANSMISSION LINES

Comparisons Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems

Measurements of edge density profile modifications during IBW on TFTR

Cullet Manufacture Using the Cylindrical Induction Melter

Argonne National Laboratory P.O. Box 2528 Idaho Falls, ID

Model: M /800 MHz Mobile Radio

22/ HASL-109 THE ATTENUATION OF NATURAL ENVIRONMENTAL RADIATION BY AN AUTOMOBILE

Controlling Changes Lessons Learned from Waste Management Facilities 8

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays

Introduction to Radar Systems. The Radar Equation. MIT Lincoln Laboratory _P_1Y.ppt ODonnell

UCRL-ID Broad-Band Characterization of the Complex Permittivity and Permeability of Materials. Carlos A. Avalle

DISCLAIMER. Portions of this document may be illegible in electronic image products. Images are produced from the best available original document.

The Development of an Enhanced Strain Measurement Device to Support Testing of Radioactive Material Packages*

Assisting DOE EM 4.12, Office of Groundwater and Subsurface Closure

AIGaAs/InGaAIP Tunnel Junctions for Multifunction Solar Cells. Sharps, N. Y. Li, J. S. Hills, and H. Hou EMCORE Photovoltaics

THE MEASURED PERFORMANCE OF A 170 GHz REMOTE STEERING LAUNCHER

STP-PT-054 CONCENTRATED SOLAR POWER (CSP) CODES AND STANDARDS GAP ANALYSIS

Recent advances in ALAMO

Armlication For United States Patent For HOT CELL SHIELD PLUG EXTRACTION APPARATUS. Philip A. Knapp Moore, ID. and. Larry K. Manhart Pingree, ID

National Accelerator LaboratoryFERMILAB-TM-1966

STP-PT-032 BUCKLING OF CYLINDRICAL, THIN WALL, TRAILER TRUCK TANKS AND ASME SECTION XII

MAPPING INDUCED POLARIZATION WITH NATURAL ELECTROMAGNETIC FIELDS FOR EXPLORATION AND RESOURCES CHARACTERIZATION BY THE MINING INDUSTRY

GA A22776 THE DESIGN AND PERFORMANCE OF WAVEGUIDE TRANSMISSION LINE COMPONENTS FOR PLASMA ELECTRON CYCLOTRON HEATING (ECH) SYSTEMS

MONITORING POWER PLANT EFFICIENCY USING THE MICROWAVE-EXCITED PHOTOACOUSTIC EFFECT TO MEASURE UNBURNED CARBON. Quarterly Technical Progress Report

Introduction to Radar Systems. Radar Antennas. MIT Lincoln Laboratory. Radar Antennas - 1 PRH 6/18/02

Integration of MGDS Design into the Licensing Process' This paper presents an overview of how the Mined Geologic Disposal System (MGDS) design

D. F. Spencer R. Aryaeinejad E. L. Reber. October 2001

GA A22583 FAST WAVE ANTENNA ARRAY FEED CIRCUITS TOLERANT OF TIME-VARYING LOADING FOR DIII D

D&D Knowledge Management through Contributions in Wikipedia

Downloaded from 1. THE FOLLOWING PAGES OF MIL-STD-462D HAVE BEEN REVISED AND SUPERSEDE THE PAGES LISTED:

Order Number : GETEC-C FCC Part 1 Test Report Number : GETEC-E Page 2 / 15 CONTENTS

FD: l-a3-97 f /WE#Tt5- u$-af79f733

Rotational Absolute Magnetic Kit Encoder Version 33 LP and HP Displacement Sensor

A Multilevel Voltage-Source Converter System with Balanced DC Voltages' Abstract

MASTER --3. Gtl.- DISTRIBUTION. THiS DOCUMENT IS UNLIMITED PNL-SA Shaw Whiteman Anderson Alzheimer G. A. March 1995

Total Ionizing Dose Test Report. Z-Series DC-DC Converter

Specification of APS Corrector Magnet Power Supplies from Closed Orbit Feedback Considerations.

MAPPING INDUCED POLARIZATION WITH NATURAL ELECTROMAGNETIC FIELDS FOR EXPLORATION AND RESOURCES CHARACTERIZATION BY THE MINING INDUSTRY

A REGULATED POWER SUPPLY FOR THE FILAMENTS OF A HIGH POWER GYROTRON

Agilent 5DX System. X-ray Safety Test Procedure. for use with Software Release 8.20 and later

GA A22897 QUASI-OPTIC COMPONENTS IN OVERSIZED CORRUGATED WAVEGUIDE FOR MILLIMETER-WAVE TRANSMISSION SYSTEMS

Inovision Model 451P Pressurized µr Ion Chamber Survey Meter

Simulation of a Mobile Robotic Platform in Gazebo and RViz using ROS

Emerging NDE Technology for Aging Aircraft

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

FEATURES ± [0.64 ± 0.05] dia. (no. 22 AWG) tinned nickel leads solderable and weldable

National Accelerator Laboratory

LUDLUM MODEL BGO Scintillator. June 2018

Performance of Keck Adaptive Optics with Sodium Laser Guide Stars

Detection of Targets in Noise and Pulse Compression Techniques

Thin Film Bar MOS Capacitors

Mechanical Pyroshoek Shrmlations for Payload Systems*

INTERMEDIATE HEAT EXCHANGER (IHX) STP-NU-038

Evaluation of Oil Shale Bitumen as a Pavement Asphalt Additive to Reduce Moisture Damage Susceptibility

THE 110 GHz MICROWAVE HEATING SYSTEM ON THE DIII D TOKAMAK

MG MW S-Band Magnetron

GA A23983 AN ADVANCED COLLABORATIVE ENVIRONMENT TO ENHANCE MAGNETIC FUSION RESEARCH

RF HIGH VOLTAGE PERFORMANCE OF RF TRANSMISSION LINE COMPONENTS ON THE DIII-D FAST WAVE CURRENT DRIVE (FWCD) SYSTEM

Anibal Bonino, Nuclear Regulatory Authority (ARN) Av del Libertador 8250 Buenos Aires, Argentina 541/ FAX: 541/

Five-beam Fabry-Perot velocimeter

Laser Surface Profiler

GYROTRON-BASED MILLIMETER-WAVE: BEAMS FOR MATERIAL PROCESSING. Thomas Hardek Wayne Cooke. William P e r r y D a n i e l Rees

Lawrence Berkeley National Laboratory Recent Work

GA A26816 DESIGNS OF NEW COMPONENTS FOR ITER ECH&CD TRANSMISSION LINES

Los A LA-UR Los Alamos National Laboratory Los Alamos, New Mexico 87545

GA A22574 ADVANTAGES OF TRAVELING WAVE RESONANT ANTENNAS FOR FAST WAVE HEATING SYSTEMS

The Technology Development Office

GA A24030 ECE RADIOMETER UPGRADE ON THE DIII D TOKAMAK

MASTER. Self-Stressing Structures for Wafer-Level Oxide Breakdown to 200 MHz. n. SELF-STRESSING OXIDE STRUCIURE

Evaluation of Roof Bolting Requirements Based on In-Mine Roof Bolter Drilling

Thick Film Resistor Networks, Military, MIL-PRF Qualified, Type RZ010 and RZ020 Dual-In-Line, Molded DIP

cycle to cycle, so errors can be used to update the reference waveforms for future cycles. At A P S, updates are

TABLE OF CONTENTS. References

GA A FABRICATION OF A 35 GHz WAVEGUIDE TWT CIRCUIT USING RAPID PROTOTYPE TECHNIQUES by J.P. ANDERSON, R. OUEDRAOGO, and D.

Performance of Image Intensifiers in Radiographic Systems

Health Issues. Introduction. Ionizing vs. Non-Ionizing Radiation. Health Issues 18.1

Sandia National Laboratories MS 1153, PO 5800, Albuquerque, NM Phone: , Fax: ,

Technical Description and Operator s Manual. Survey Meter SM 5 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau

v-~ -8 m w Abstract Framework for Sandia Technolow Transfer Process Introduction

GA A22577 AN ELM-RESILIENT RF ARC DETECTION SYSTEM FOR DIII D BASED ON ELECTROMAGNETIC AND SOUND EMISSIONS FROM THE ARC

TEST REPORT FROM RFI GLOBAL SERVICES LTD

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON

Nuclear Science and Security Consortium: Advancing Nonproliferation Policy Education

Transcription:

ORNUM-6604 Technical Basis Document (TBD) and User Guides P. J. Chiaro, Jr.nwE\VED

This report has been reproduced directly from the best available copy. Available to DOE and DOE contractors from the Office of Scientific and Technical Information. P.O. Box 62. Oak Ridge, TN 37831; prices available from (61 5) 576-8401. Available to the public from the National Technical Information Service, U.S. Department of Commerce, 5285 Port Royal Rd., Springfield. VA 22161. I 1 This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States nor any agency thereof, nor any of their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof..

This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the Unittd States Government nor any agency thereof, nor any of their empioyees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or ~ M c by e trade name, trademark, manufacturer, or otherwise does not nectssarily constitute or imply its endorsement, recornmend;ltion, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

DISCLAIMER Portions of this document may be illegible in electronic image products. Images are produced from the best available original document.

ORNL/M-6604 Instrumentation and Controls Division TECHNICAL BASIS DOCUMENT (TBD) AND USER GUIDES P. J. Chiaro, Jr. Date Published: September 1998 Prepared by OAK RIDGE NATIONAL LABORATORY Oak Ridge, Tennessee 3783 1-6285 Managed by LOCKHEED MARTIN ENERGY RESEARCH CORP. for the U.S. DEPARTMENT OF ENERGY under contract DE-AC05-960R22464

4

CONTENTS BACKGROUND... V 1. TECHNICAL BASIS DOCUMENT... 1.1 BACKGROUND AND SCOPE... 1.2 APPLICATION... 1.3 ESTABLISHING OPERATIONAL LIMITATIONS... 1.3.1 Radiation Types... 1.3.2 Temperature and Humidity... 1.3.3 Radio Frequency and Magnetic Fields... 1.4 REQUIREMENT REFERENCE... 1 1 1 1 1 2 2 2 2. USER GUIDE... 2.1 SCOPE... 2.2 SPECIFICS... 2.3 EXAMPLES... 3 3 3 4 3. INSTRUMENT USER GUIDE. EBERLINE MODEL RO-20... 3.1 DESCRIPTION... 3.2 TYPICAL USES... 3.3 OPERATIONAL INFORMATION... 3.4 TEST INFORMATION... 3.5 USER LIST... 3.6 USER COMMENTS... 5 5 5 5 7 7 7 4. INSTRUMENT USER GUIDE. BICRON SURVEYOR M WITH PANCAKE GM PROBE 8 8 4.1 DESCRIPTION... 8 4.2 TYPICAL USES... 8 4.3 OPERATIONAL INFORMATION... 9 4.4 TEST INFORMATION... 9 4.5 USER LIST... 4.6 USER COMMENTS... 10 5. INSTRUMENT USER GUIDE. LUDLUM MODEL 12 WITH A MODEL 43-65 PROBE.. 1 1 5.1 DESCRIPTION... 11 5.2 TYPICAL USES... 11 5.3 OPERATIONAL INFORMATION... 11 5.4 TEST INFORMATION... 13 5.5 USER LIST... 13 5.6 USER COMMENTS... 13... 111

4

BACKGROUND To ensure that radiation protection instruments are appropriate for (1) the type(s), levels, and energies of the radiation(s) encountered, and (2) for existing environmental conditions, [ 10 CFR 835.401(~)(2)& (3)], some U. S. Department of Energy (DOE) facilities have developed Technical Basis Documents (TBDs) and Instrument User Guides. Most are paper-based and not easily updated or shared. Due to the need to share information, the development of a standard TBD and a web-based User Guide format was needed. The combination of both documents should help those involved in the purchase, approval, and use of radiation instruments with ensuring that a specific instrument model is appropriate for use at their facility. a V

. 1. TECHNICAL BASIS DOCUMENT 1.1 BACKGROUND AND SCOPE A Technical Basis Document (TBD) should provide the background information for establishment of an instrument s operational requirements. Due to the amount and location of DOE facilities, no one set of requirements is possible. Operational requirements will vary based on the local environments and missions at each facility. Environmental conditions that can affect an instrument s operations are ambient temperature, humidity, and radio frequency, and to a lesser extent, magnetic fields, and interfering ionizing radiations. Consideration should also be made regarding how an instrument is to be used. If an instrument will be transported around the facility, vibration and shock can cause problems if they are not addressed in the TBD. This document provides guidance for the development of a TBD. 1.2 APPLICATION This document applies to radiation instruments used for personnel and equipment contamination monitoring, dose rate monitoring, and air monitoring. 1.3 ESTABLISHING OPERATIONAL LIMITATIONS Tests performed by instrument manufacturers and independent facilities have found that certain environmental conditions can cause detrimental effects on a radiation instrument s ability to operate correctly. Observable effects usually include reduced response, increased response, erratic response, or no response. Without prior knowledge of these operational limitations, instruments could be placed into service that are not appropriate for the expected environmental conditions. The TBD should contain specific information about the expected environmental conditions and information about the radiation types and levels expected to be encountered at the facility. The following sections provide guidance in a format that can be used to establish a facility-specific TBD. 1.3.1 Radiation Types This section should contain a statement regarding the types of radiation encountered at the facility, i.e., alpha, beta, gamma, and/or neutron. Information about expected energy levels or isotopes is also suggested. Example Expected radiation types: Beta, Gamma. o, 9 0 ~ Common Isotopes: I cs, 6 0 ~ 90Sr/ Interfering Ionizing Radiations-An instrument s response to radiations other than those it was calibrated for should be known. Typically, radiation instrumentation users will be the personnel who are most aware of the various radiations at the facility. 1

Example Instruments shall be capable of monitoring alpha radiation in the presence of beta-gamma and neutron radiation; beta-gamma radiation in the presence of alpha and neutron radiation; and neutron radiation in the presence of alpha and beta-gamma radiation. 1.3.2 Temperature and Humidity Atmospheric conditions will obviously vary across the complex. Conditions within weather protected and non-weather protected locations should be used as a basis for establishing operating requirements. These can usually be obtained from local weather records or through an analysis of individual operations within each location. Climatic information is also available in existing Military Standards, such as MIL-STD 810. Besides minimum and maximum conditions, exposure times should be considered. Instruments used for brief periods at elevated temperature and high humidity should not be required to meet the same requirements as instruments used for extended periods of times in those conditions. Example Outdoor Conditions Expected temperature range: +I 0 OF to +I I O O F. Typical Relative Humidity Levels: 20 to 95% RH. Barometric Pressure: 28 to 31 in Hg. Indoor Conditions Possible temperature range: +40 O F to +I20 O F. Expected temperature range: +65 O F to +80 O F. Possible relative humidity range: 20 to 95% RH. Expected relative humidity range: <75% RH. Note: Most facilities are environmentally controlled. These conditions may be external to the facilities while the operating environment is stable and controlled inside most facilities. 1.3.3 Radio Frequency and Magnetic Fields Radio frequencies (non-ionizing radiation) and magnetic fields must be considered when evaluating radiation instrumentation. The prevalence of radios and cellular communication, as well as the use of other non-communications equipment that generate RF fields have increased the amount of interfering non-ionizing radiation in the environment. Documented susceptibilities include response effects (increased and decreased), and other operational changes such as alarm activation, display effects, etc. Information about local RF emitting devices such as their output frequency is necessary to ensure that the instrument is properly evaluated prior to use. The local communications group could probably provide the necessary information for the establishment of operating criteria. The use of a scan evaluation over a frequency range of 20 or 80 MHz to 1000 MHz at 20 Volts/meter will ensure that an instrument s susceptibility is known. 1.4 REQUIREMENT REFERENCE Title 10, Code of Federal Regulations, Part 835;Occupational Radiation Protection 2

2. USERGUIDE 2.1 SCOPE A User Guide is a consolidated source of information for an instrument model s operating limitations and susceptibilities. The information is based on controlled evaluations performed across the DOE complex, from field-use experience and manufacturer-supplied data. It can be used with a TBD to determine if a particular model meets the operational requirements of a specific facility, or as a means to share information about a particular model with other users. The user guide is set up for use on the World Wide Web to enable easier updating and viewing. Access to the web site is uncontrolled, but registration is required when providing operational comments. A description of each user guide section follows. 2.2 SPECIFICS A typical user guide is divided into the following sections: 1. Description-Describes the instrument model, probe/detector type, operating range(s), basic operating description, etc. 2. Typical Uses-Describes 3. Operational (test-based)-provides a breakdown of the limitations on the instrument s use as determined by controlled testing performed either by the manufacturer or independent laboratory. The subsections are as follows: how the instrument is commonly used, i.e., dose rate measurement, contamination, etc. 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 Temperature, Humidity, Radio Frequency, Magnetic Field, Vibration and Shqck, Energy Response, Interfering Ionizing Radiations, and Other. 4. Test Information-Provides World Wide Web. 5. User List-Provides a listing of the instrument s current users. The list is updated as needed and used to notify interested people of new information regarding a specific instrument model automatically, as needed. 6. User Comments-Contains a list of field-use observations and experiences provided by other users. All comments are shown by date and user name. Each comment is reviewed before being added to the users guide. links to appropriate testing information available on the

2.3 EXAMPLES Three example user guides are attached as Appendices 1 through 3. Appendix 1, Eberline RO-20 Appendix 2, Bicron Surveyor with Pancake GM Probe Appendix 3, Ludlum Model 12 with a 43-65 Probe 4

3. INSTRUMENT USER GUIDE EBEIUINE MODEL RO-20 3.1 DESCRIPTION The Eberline Model RO-20 is a portable, battery powered air ionization chamber used to detect beta, gamma, and x-ray radiation. It has five linear ranges of 0-5, O - 50, and 0-500 mr/hr, and 0-5, and 0-50 R/hr. The ion chamber is vented to the atmosphere so that changes in atmospheric pressure will affect the response of the instrument. 3.2 TYPICAL USES The Eberline RO-20 is typically used to perform dose rate surveys. It is commonly used to establish stay times or monitor areas to establish protection boundaries. 3.3 OPERATIONAL INFORMATION TemDerature Manufacturer: -40 F to +140"F(-40 C to +60"C) Available Testing: -10 C to 50 C (+14"Fto +122"F) RO-20s can be susceptible to rapid changes in temperature especially when operated using the most sensitive range. The susceptibility is greater when exposed to changes that are lower than ambient (22 C to -10 C). When it is necessary to operate an RO-20 after a temperature change of >lo"c, it is recommended that the instrument reach thermal equilibrium prior to use. Humidity Manufacturer: Operable from 0 to 95%, non-condensing Available Testing: Exposed to 95% (non-condensing) for eight hours, and 40% for 4 hours at 22 f 2 C. None for non-condensing atmospheres. None.. Radio Freauency Manufacturer: None provided. Available Testing: Tested at 20 Volts/meter over a frequency band from 100 khz to 1000 MHz and at 2450 MHz. 5 '

Some susceptibility may exist from 730 to 800 MHz when operated using the most sensitive range. The RO-20 should not be operated using the most sensitive range in RF fields from 730 to 800 MHz. Mametic Field Manufacturer: None provided. Available Testing: 10 Gauss (10 Oersted) DC, and 60 Hz (1.26 Gauss) AC in two orientations. None for field intensities up to those performed during testing. If field conditions are unknown, it is recommended that a response check be performed prior to use at the location of use. Some effect may be seen when operated in high magnetic fields. Vibration and Shock Manufacturer: None provided. Available Testing: Exposure to 15 Hz and 28 Hz at 2 G in three orientations. No shock information available. Typical electronic equipment handling procedures. It is recommended that a response check be performed prior to use if it is suspected that the unit had been exposed to an extreme vibration condition or shock event. Energy ResDonse Note: The following information is provided by the manufacturer. Information is normalized to a I3 Cs calibration, slide closed and facing the, source. Photon +: 30% from 8 kev to 1.3 MeV with open slide facing the source. +: 15%from 33 kev to 1.3 MeV with the slide closed. +: 15% from 55 kev to 1.3 MeV through the side of the case. Beta Uranium slab - 30% of the true mrad/hr field behind 7 mg/cm2 window with RO-20 resting on the slab, slide open. 90 S r - 9 - approximately 93% of true mradhr field at 30 cm with the slide open. Fast Neutron Response (PuBe) 8% in mr/hr of true neutron field in mrem/hr. Note - Available test results indicated a rate of 6%. 6

Interfering Ionizing Radiations Manufacturer: N/A. Available Testing: N/A. N/A. This instrument is designed to respond to x-ray, gamma, and beta. It also responds to neutrons. Information regarding its neutron response is available in the energy response section. Other The RO-20 has been revised since its inception. Revisions were made to address fundamental susceptibilities with changes in temperature. Questions regarding currently available units should be addressed to the manufacturer. Questions not addressed directly in this document should be addressed to other users or the manufacturer. 3.4 TEST INFORMATION PNNL - http://www.pnl.gov/health/health-prot/instrume.html ORNL - http://www.ic.ornl.gov/centers/ite/hpinfo.htm LANL - http://esh-4.lanl.gov/ric1/result.htm a 3.5 USERLIST This section provides a listing of the instrument s current users. The list is updated and used to notify people of new information regarding a specific instrument model, as needed. 3.6 USER COMMENTS This section contains a list of field-use observations and experiences that have been provided by other users. All comments are shown by date and user name. Each comment is reviewed before being added to the TBD. User Name Date Comment Key Word (Main Issue of Comment) 7

4. INSTRUMENT USER GUIDE BICRON SURVEYOR M WITH PANCAKE GM PROBE j 4.1 DESCRIPTION The Surveyor M is a portable, battery operated count rate meter. The pancake GM tube is a Bicron Model PGM. The ranges are X 1 (0-500 cpm), X 10 (0-5000 cpm), X 100 (0-50000 cpm), and X 1000 (0-500000 cpm). The X 1000 range is not recommended for use with this instrumentldetector combination. 4.2 TYPICAL USES The Bicron Surveyor M/PGM is typically used for betdgamma contamination surveys. It is commonly used to check for contamination on equipment or personnel. 4.3 OPERATIONAL INFORMATION TemDerature Manufacturer: -20 C to +50 C (-29 C to +122"F) Available Testing: -10 C to 50 C (+14"C to +122"F) None when operated within the stated temperature range. If operation is required outside the stated temperature range, a response check should be performed after the instrument reaches thermal equilibrium with the ambient conditions prior to use. Humidity Manufacturer: 5% change in reading from 10 to 95% relative humidity, Available Testing: Exposed to 95% (non-condensing) for eight hours, and 40% for 4 hours at 22 2 C. None for non-condensing atmospheres. None. * Radio Freauencv Manufacturer: None provided. Available Testing: Tested using a scan from 0.3 MHz to 35 MHz, and at 140 MHz at 50 voltdmeter. Also tested at 915 MHz and 2.45 GHz at 0.4 watts/meter*and 2.0 wattdmeter' respectively. None. None. If field conditions are unknown, it is recommended that a response check be performed prior to use at the location of use. 8 '

Magnetic Field Manufacturer: None provided. Available Testing: 10 Gauss DC. This instrument/probe combination shows a susceptibility to magnetic fields when operated using the X 1 range. Other, less sensitive ranges do not seem susceptible. The instrument/probe combination should not be operated using the X1 range in areas where magnetic fields may exist. If field conditions are unknown, it is recommended that a response check be performed prior to use at the location of use. Vibration and Shock Manufacturer: Up to loog per lightweight machine of MIL-STD 202C, method 202B (shock). Up to 5 g in three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz. Available Testing: Exposure to 15 Hz and 28 Hz at 2 G in three orientations. No shock information available. Typical electronic equipment handling procedures. It is recommended that a response check be performed prior to use if it is suspected that the unit had been exposed to an extreme vibration condition or shock event. Energy Resoonse No information available. Interferinp Ionizing Radiations This instrument/probe combination responds to most all ionizing radiations. Other None. 4.4 TEST INFORMATION ORNL - http://www.ic.ornl.gov/centers/ite/hpinfo.htm 4.5 USERLIST This section provides a listing of the instrument s current users. The list is updated and used to notify people of new information regarding a specific instrument model, as needed. 9

4.6 User Comments This section contains a list of field-use observations and experiences provided by other users. All comments are shown by date and user name. Each comment is reviewed before being added to the TBD. User Name Date Comment Key Word (Main Issue of Comment) 10

5. INSTRUMENT USER GUIDE LUDLUM MODEL 12 WITH A MODEL 43-65 PROBE. 5.1 DESCRIPTION The Ludlum Model 12 is a portable, battery operated count ratemeter. The Ludlum Model 43-65 detects alpha radiation. The Ludlum Model 12 has a scale of 0-500 counts per minute with four range multipliers; X1, X10, X100, and X1000. 5.2 TYPICAL USES The Model 12/43-65 instrument combination is typically used to do alpha surface Contamination surveys. It is commonly used to check for contamination on equipment or personnel. 5.3 OPERATIONAL INFORMATION This section provides a breakdown of the limitations on the instrument's use as determined by controlled testing, or as stated by the manufacturer. I Temperature Manufacturer: -40 C to 50 C (-40 to 122 F) Available Testing: -10 C to 50 C (+14 to +122"F) The instrument is susceptible to rapid changes in temperature especially when operated using the X1 range. The X1 is also affected when ambient temperatures are less than +lo"c (+50"F)or greater than +40"C (+104"F). If operation is required outside of the stated temperature range, a response check should be performed after the instrument reaches thermal equilibrium with the ambient conditions. This instrument/probe combination should not be used on the X1 range when the ambient temperature is less than +1O"C (+50"F)or greater than +40"C (+104"F). When exposing this instrument/probe combination to a temperature change of greater than 10 -C is necessary, it is recommended that the instrument reach thermal equilibrium and then be successfully response checked before use. Humiditv Manufacturer: No information available. Available Testing: Exposed to 95% (non-condensing) for eight hours, and 40% for 4 hours at 22 & 2 C. 11

None for non-condensing atmospheres. None. Radio Freauencv Manufacturer: None provided. Available Testing: Tested using a scan from 0.3 MHz to 35 MHz, and at 140 MHz at 50 voltdmeter. Also tested at 915 MHz and 2.45 GHz at 0.4 watts/meter2 and 2.0 watts/meter2 respectively. The instrument/probe combination is susceptible to RF fields. Susceptibilities were indicated at frequencies below 35 MHz, and at 140 and 915 MHz, primarily on the X1 range. The Ludlum Model 12 with a Ludlum Model 43-65 should not be operated in radio frequency fields on the X1 range. If field conditions are unknown, or when operated using other ranges, it is recommended that a response check be performed before use at the location of use. Magnetic Field Manufacturer: None provided. Available Testing: 10 Gauss (10 Oersted) DC, and 60 Hz (1.26 Gauss) AC in two orientations. The instrument/probe combination is susceptible to magnetic fields. The Ludlum Model 12 with a Ludlum Model 43-65 should not be used in areas where magnetic fields exist. Vibration and Shock Manufacturer: None provided. Available Testing: No test information available. Typical electronic equipment handling procedures. Energv ResDonse NIA It is recommended that the instrument be response checked before use if it is suspected that the unit had been exposed to an extreme vibration condition or shock event. Interfering Ionizing Radiations Manufacturer: None. Available Testing: Gamma - Exposed to 11.7 Whr 137Cs gamma field. Neutron - Exposed to a 1 rem/hr unmoderated "'Cf neutron field. 12

The response of the Ludlum Model 12 with a Ludlum Model 43-65 was 2000 to 4000 counts per minute when exposed to the neutron field. No response was observed when exposed to the gamma field. It is recommended that this instrument/probe combination not be used when neutrons are present. Other Questions not addressed directly in this document should be addressed to other users or the manufacturer. 5.4 TEST INFORMATION This section provides links to national laboratory testing information available on the world wide web. ORNL - http://www.ic.orni.gov/centers/ite/hpinfo.htm 5.5 USERLIST This section provides a listing of the instrument s current users. The list is updated and used to notify people of new information regarding a specific instrument model, as needed. 5.6 USER COMMENTS This section contains a list of field-use observations and experiences provided by other users. All comments are shown by date and user nanie. Each comment is reviewed before being added to the TBD. User Name Date Comment Key Word (Main Issue of Comment)

ORNL/M-6604 INTERNAL DISTRIBUTION 1-2. P. J. Chiaro, Jr. 3. M. A. Phipps 4. R. T. Rosebeny 5. E.R.Tapp 6. J. D. White 7. 8. 9. 10-11. M. C. Wright Central Research Library ORNL Laboratory Records-RC ORNL Laboratory Records-OSTI EXTERNAL DISTRIBUTION 12. Murari Sharma, U.S. Department of Energy, EH-52, 19901 Germantown Road, Room 5057, Germantown, MD 20874