perioperativecpd.com The basics of diathermy

Similar documents
Electrosurgery Units (ESU s) and their variants

Electrosurgery Units (ESU s) and their variants. D. J. McMahon rev cewood

RAISING THE LEVEL OF PROTECTION IN YOUR ORS

What tests should be done?

Andy Falconer. Principles of surgical diathermy

ELECTROSURGICAL GENERATOR USER S GUIDE

MICROWAVE DIATHERMY AND SURGICAL DIATHERMY DIATHERMICS

J-470 ELECTROSURGE. Radio Frequency Electrosurgery Unit. Operation Instructions

User s Guide Aaron 950

Application Notes. Electrosurgical Analyzer Primer

USER S GUIDE. User s Guide Bovie IDS-400

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence

Meg1 (with Ligasmart)

An Alternative Technique to Perform Surgeriesin Hospital by Surgical Diathermy

User s Guide Aaron 940

BOVIE SURGI-CENTER PRO. Electrosurgical Generator. User's Guide

ERBE ERBOTOM ICC 300. Instruction manual 08.00

ELEKTROTOM 610. Operating Instructions. Valid from version 1130

Instruction Manual. (English-English)

Surgitron Dual Frequency 4.0 MHz Radiowave Technology

Innovating Together. An introduction to Electrosurgery

Review Article Electrical Safety in Operating Room. Abstract

White Paper: Technology and Clinical Overview of the Genii TouchSoft Coagulator monopolar probe

Low / Medium Power Electrosurgery Review 2002

DSL# EC 2009 Ethicon Endo-Surgery. All rights reserved PRODUCT CATALOG

INSTRUCTIONS ESG-400

radiosurg One device for all surgical treatments!

EXCELL NHP/T. A new step AheAd to the future ELECTROSURGICAL UNITS FOR MONOPOLAR AND BIPOLAR USE ISO ALSA IT ISO

ERBE. ERBOTOM ICC 80 / 50 Instruction manual 08.00

OPERATING MANUAL ELECTROSURGICAL UNIT

DERM 941 DERM 942 High-Frequency Desiccators

Introduction to Medical Electronics Industry Test Analysis and Solution

Operator s Manual E L E C T R O S U R G I C A L U N I T

Electrosurgery. KLS Martin ME MB 2 SAFETY BY EASY USE

OPERATING MANUAL ELECTROSURGICAL UNIT

surtron.com SURTRON ELECTROSURGICAL units

ALSATOM SU MPC SU 50 MPC SU 100 MPC SU 140 MPC SU 140/D MPC SU 140/BD MPC

Electrical noise in the OR

ESU-2000 Series Product Overview

PULSAR Generator Operator s Manual

Electrosurgery. KLS Martin ME 102. The compact Electrosurgical Unit with high Capacity

Potential Risks of MRI in Device Patients

LEEP System 1000 Table of Contents


maxium electrosurgery unit with maxium Beamer

HF- GENERATORS.

RF leakage current in electrosurgical units: Influence of the layout in taking measurements

ARC 200 HF SURGICAL GENERATOR SIMPLY SAFE

Rolyan Splint Pan OPERATION MANUAL. Item # Small Item # Large

*Notebook is excluded

Current-Programmed Mode Control Strategies for Electrosurgical Generators

Aquamantys 3 Pump Generator User Guide. Revision F

ULP Wireless Technology for Biosensors and Energy Harvesting

Safety Warnings Features Specifications Instrument Layout Operation Preparation AC Current Measurement How to Use Peak Hold Function How to Use The

The use of sutures and ligatures is an integral part of surgical procedures. Being proficient in suturing and ligating will not only make your

Crossfire Footswitch. User Guide

Electrosurgical Generator and Smoke Evacuation System Service Manual. Utah Medical Products, Inc. Catalog Numbers ESU-110 & ESU-220

Biomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer:

Safety Warnings Features Specifications Instrument Layout Operation Preparation AC Current Measurement How to Use Peak Hold Function How to Use The

HAZARDS OF NON-IONIZING RADIOFREQUENCY (RF) RADIATION

79/26 Series III Multimeter

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

Radiofrequency: Principles & Devices

EDL Group #3 Final Report - Surface Electromyograph System

Physical Therapy. Methods in Physical Therapy. Non-electric heat therapy Therapeutique application of electric current

BME 599a Applied Electrophysiology Midterm (Thursday 10/12/00 09:30)

Application Note # 5438

DIATHERMY UNITS HIGH FREQUENCY HEAT THERAPY:

Radio Frequency Electrosurgical System Specialist

: 0089 GTIN

V SALAI SELVAM, AP & HOD, ECE, Sriram Engg. College, Perumalpattu 1 MEDICAL ELECTRONICS UNIT V. Solar spectrum

Introduction EMC. Filter parameters. Definition of EMC / EMI. X-Capacitor. Sources of EMI. Coupling mechanism. Y-Capacitor.

M30 SPOT WELDER INSTRUCTION MANUAL 230V 1PH.

Sonicator Plus 940. Instruction Manual

(12) Patent Application Publication (10) Pub. No.: US 2007/ A1

USER'S MANUAL DMR-6700

Institute of Surgical Research A Module - Surgical Techniques

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

DIATERMO MB 122 AND MB 160 MONO-BIPOLAR W W

Central Blood Pressure Meter Model cbp301. Operating Manual

National Voluntary Laboratory Accreditation Program

5570M Series Intercom Installation and Operation Guide

RF Amplifiers and RF Heating for Medical Applications FTF-IND-F0690

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

Investigation of skin effect on coaxial cables

OPERATING MANUAL MINIDOP ES-100VX POCKET DOPPLER

GET MORE. WITH LESS. Less sticking. 1, Less cleaning. 2, Less eschar buildup. 2, More benefits. LigaSure Maryland jaw device with nano-coated jaws

National Voluntary Laboratory Accreditation Program

Leksell Neuro Generator

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان

Marx Generator 2.0 Kit

SCANLAN LEGACY. High Precision Titanium Instruments

Device Under Test: ALTEA VS- 24-I VS-24-I. 0 24/09/12 First issue A. Peretto L. Peretto 1 24/06/16 All text review E. Scala L. Peretto J. L.

Electrocardiogram (ECG)

INTRINSIC SAFETY BASIC PRINCIPLES

ET 40 - Electrician Theory Examination Marking Schedule

BM VET 80, 120, 140, 180 W

Unit 3 Magnetism...21 Introduction The Natural Magnet Magnetic Polarities Magnetic Compass...21

TRANSMITTER ZM-920PA/930PA

Transcription:

perioperativecpd.com continuing professional development The basics of diathermy By the perioperativecpd team Introduction Electrosurgery or diathermy is the passage of a high-frequency alternating current through the body to produce a desirable surgical effect. It has allowed faster operating, reduced blood loss and new surgical techniques. It does have its dangers and its safe use in operating theatres requires a grasp of its mechanisms of action and potential pitfalls. perioperativecpd@gmail.com 1

Electrocautery or Electrosurgery? The terms electrocautery and electrosurgery have often been confused. Electrocautery is an old system for haemostasis which is no longer used. It uses direct current (DC) to generate heat to the tip of a metal instrument, thereby causing blood to coagulate. During electrocautery, current does not enter the patient s body. Only the heated wire comes in contact with tissue. Electrosurgery uses modern instruments that harness alternating current (AC) and the patient is included in the circuit. Diathermy circuits Historically, diathermy used mains voltage referenced to earth and the patient return electrode or pad was earthed. If this pad became loose, was incorrectly applied, or the current found an alternative path to earth (e.g. via a limb in contact with the table) burns would result. Since the 1980 s modern monopolar diathermy generators use an isolated circuit with output voltages referenced to the generator itself (figure 1). This requires current to flow back to the generator to complete a circuit. If this circuit is broken (e.g. if the patient return electrode is not connected), current will not flow. Therefore in modern theatres calling the patient return pad/electrode an earthing or grounding pad it is not correct. Only the metal casing of the diathermy machine is earthed in case of insulation failure. Modern monopolar circuit bipolar circuit Figure 1: The electricity circuit from a monopolar and a bipolar circuit Frequency Standard electrical current alternates at a frequency of 50/60* cycles per second (Hz). Diathermy systems could function at this frequency, but as the current would be transmitted through body tissue at 60 cycles, excessive neuromuscular stimulation and perhaps electrocution would result. Nerve and muscle stimulation cease at 100,000 cycles per second (100 khz), therefore diathermy can be performed safely at frequencies above 100 khz (Figure 2). These are sometimes called the radio frequencies. perioperativecpd@gmail.com 2

An electrosurgical generator takes the 50/60 cycle current and increases the frequency to over 200,000 cycles per second (200 khz). At this frequency electrosurgical energy can pass through the patient with minimal neuromuscular stimulation and no risk of electrocution. The electrical energy is converted to heat and tissue temperatures at the active electrode tip can be as high 1000 0 C. *some countries use 50Hz (U.K.) and some 60Hz (U.S.A) Household Ceasing of muscle Appliances stimulation AM Radio Television 60Hz 100Hz 550-1550kHz 54-880MHz Figure 2: Frequencies of electrosurgery Electrosurgery 200 khz 3.3 MHz Monopolar diathermy Monopolar is the most commonly used electrosurgical modality, due to its versatility and clinical effectiveness. In monopolar diathermy, the active electrode is in the surgical site. The dispersive or patient return electrode is somewhere else on the patient s body (figure 1). Monopolar diathermy should also never be used on endarterial organs which include fingers, ears, the nose and the penis. If the main supplying artery is thrombosed, it can result in necrosis and self-amputation. Bipolar In bipolar electrosurgery, both the active electrode and return electrode functions are performed by a pair of forceps at the site of surgery (Figure 1). Only the tissue grasped is included in the electrical circuit. It has limited ability to cut and coagulate large bleeding areas, and it is more ideally used for those procedures where tissues can be easily grabbed on both sides by the forceps electrodes. Because the return function is performed by one tine of the forceps, no patient return electrode is needed. Modern devices are capable of sealing vessels up to 7 mm in diameter by a combination of mechanical pressure and diathermy. Microprocessors use tissue response generators to adjust current and voltage based on the sensed tissue impedance, so vessels can be reliably sealed to withstand three times normal systolic pressure. perioperativecpd@gmail.com 3

Patient return pad safety Historically, patient return electrode burns have accounted for 70 percent of the injuries reported during the use of electrosurgery. Patient return electrodes are not inactive or passive. The only difference between the active electrode and the patient return electrode is their size and relative conductivity. The active electrode (i.e. diathermy tip) has a small surface area for contact with the target tissues; all of the current is forced to flow through the tiny area where the active electrode makes contact with the tissue. At this point, the current flow is concentrated intensely. The heat at the site is great enough to achieve cutting and coagulation (figure 3). Current leaves the body via a dispersive electrode or patient return pad. The pad has a large surface area thus the current density is low. It is typically applied to the thigh, buttock or back. As long as the surface area is large, (i.e over 100cm 2 ) and the electrode makes good contact with the skin it will offer a passage of least resistance for safe exit of the current from the patient. An incorrectly applied pad can cause burns due to the reduced surface area increasing the current density and heat. Things to consider in placing the return pad include avoiding areas of vascular insufficiency, irregular body contours and bony prominences. The plate must also be placed away from any metal implants to avoid the current passing into this and causing heat damage. More modern machines actively monitor the amount of impedance at the patient/pad interface because there is a direct relationship between this impedance and the pad contact area. The systems are designed to deactivate the generator before an injury can occur if it detects a dangerously high level of impedance caused by a pad becoming partially detached(split pads). Figure 3: Current density diagram perioperativecpd@gmail.com 4

Modes Diathermy generators are able to produce a variety of electrical waveforms. As waveforms change, so will the corresponding tissue effects. Cut (Yellow button or pedal) The cutting action of monopolar is achieved by a continuous low voltage electric current waveform action which vaporises the tissues on contact (Figure 4). This waveform produces heat very rapidly and allows the cutting of tissues without coagulation and the effect is similar to being cut with a scalpel. The electrode does not need to be in contact with the tissue. Although it can be used for the initial skin incision a scalpel is still preferred. On the skin a scalpel reduces tissue damage, produces neater wound edges and reduces infection rates. The Cut setting sometimes has extra settings called Blend depending on the levels of energy involved. A blended current is not a mixture of both cutting and coagulation current but rather a cut mode with a modification of the duty cycle. A lower duty cycle produces less heat. Coagulation or Coag (Blue button or pedal) This uses a higher voltage AC current but the waveform is only on 6% of the time (see Figure 4). This interrupted waveform will produce less heat and instead of tissue vaporization, a coagulation effect is produced. Coag can be used directly through the active electrode or through a conducting device such as insulated forceps to direct the coagulation more accurately. As with Cut there are two modes: desiccation (also called forced coag ) and fulguration ( spray coag ). Figure 4: Differences in electricity currents between cutting and coagulation mode of diathermy Desiccation ( Forced Coag ) This allows more precise coagulation and the electrode needs to be in touch with the target tissue. The voltages are somewhat lower in desiccation mode compared to Fulguration but with slightly higher currents (0.5W vs 0.1W). perioperativecpd@gmail.com 5

Fulguration ( Spray Coag ) This is good for haemostasis and sprays a shower of sparks a few millimetre away from the targeted tissue. It is achieved by using very high voltages (around 6000V) with lower currents. It should be avoided on delicate organs like bowel and near large vessels as the effect is less controlled than desiccation and can cause thermal injuries. Other precautions Fire Most flammable anaesthetic gases have been eliminated from the operating theatre but alcohol preps and other flammable vapours can ignite if present when an active electrode is used in the same area. Care should be used if using alcohol preps and if they need to be used then avoid pooling. The inadvertent activation of an active electrode positioned on sponges, drapes, or in an oxygen-enriched atmosphere can also result in fires. Capacitive coupling A capacitor is simply two conductors (one of which may be the active electrode) separated by an insulator such as air. High frequency currents flow through a capacitor since an electrostatic field is created between the two conductors, and when the insulating capacity is exceeded, a current is induced in the other conductor. If this current is high enough it can cause a burn at sites of patient contact. This is a hazard during laparoscopic surgery between the active electrode and a metal trochar or scope. (Figure 5) Figure 5: two types of capacitive coupling perioperativecpd@gmail.com 6

Electromagnetic interference (EMI) Diathermy can interfere with both EEG and ECG monitoring electrodes. Of greater consequence is the interaction between EMI and cardiac pacemakers or implantable defibrillators. Bipolar is safer than monopolar, but can still cause interference. The effects are unpredictable and include inappropriate pacing, damage to the device, inappropriate defibrillation, and myocardial heat damage. Modern pacemakers have a titanium shell and interference monitor to protect them from EMI. The use of a magnet to reset pacemakers to asynchronous continuous pacing can be unpredictable. Current advice suggests limiting the use of diathermy to short low-power bursts and avoiding monopolar where possible. Placement of the dispersive electrode away from the device increases safety. Where appropriate the device should be checked and reprogrammed to monitoring mode prior to surgery. Diathermy surgical smoke Diathermy smoke consists of 95% steam and 5% cellular debris, with a mean diameter of 0.07 m, containing a variety of toxic mutagenic chemicals including hydrogen cyanide and benzene. To prevent viruses and viable cancer cells being transmitted in surgical smoke and diathermy machines should now contain a surgical smoke evacuator with a 0.1-m filter, attached to the diathermy pencil less than 2 cm from the site of smoke production. Similar devices Ultrasonic surgical tools are not diathermy devices. They operate by containing elements that vibrate at 55.5 khz. No electrical energy flows. The cutting and coagulation effect occurs at lower temperatures than diathermy, by a combination of mechanical and heat energy that denatures proteins. References: Ahmad NZ, Ahmed A. Meta-analysis of the effectiveness of surgical scalpel or diathermy in making abdominal skin incisions. 2011. Ann Surg. 253: 8 13. Allen M. Pacemakers and implantable caridoverter defibrillators. Anaesthesia 2006; 61: 883e90. Bovie Medical. Understanding Electrosurgery. http://www.boviemedical.com/downloads/understandingelectrosurgerylit_r3.pdf. Accesssed 15/11/2016 Boyd, Daniel E. et al. Surgical diathermy.naesthesia & Intensive Care Medicine, Volume 14, Issue 10, 431 433 Carter PL, The life and legacy of William T. Bovie. 2013. The American Journal of Surgery, 205(5): 488-491. Edward J, Fitzgerald F, Malik M, Ahmed I. A single-blind controlled study of electrocautery and ultrasonic scalpel smoke plumes in laparoscopic surgery. 2012. Surgical Endoscopy. 26(2): 337-342 Ethicon, Inc. or Ethicon Endo-Surgery, Inc., Data on file. Harmonic Ace, Enseal G2. http://www.ethicon.com/healthcareprofessionals/products/advanced-energy - Accessed 18/04/2016. Namal RS, Shrestha A, Ling Au-Yeung KC. Basic Principles of Electrosurgery and Energized Dissection. 2014. WJMER 6(1) Mehta SP, Bhananker SM, Posner KL, Domino KB. Operating room fires: a closed claims analysis. 2013. Anesthesiology. 118 (5):1133-9. Principles of Electrosurgery. Department of Education, Covidien Inc. 2008http://www.valleylab.com/education/poes/ - Accessed 01/10/2016 The SAGES Manual on the Fundamental Use of Surgical Energy (FUSE). Springer New York 2012-01-01P. Ulmer BC. The hazards of surgical smoke. AORN J 2008; 87: 721e34. perioperativecpd@gmail.com 7