Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic Density

Similar documents
Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Nuclear Associates

Safelight Fog does what to contrast and density on film?

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Beam-Restricting Devices

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

Effect of Backscattered Radiation on X-Ray Image Contrast

DIGITAL RADIOGRAPHY ARTIFACTS

- KiloVoltage. Technique 101: Getting Back to Basics

Visibility of Detail

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

1-1. GENERAL 1-2. DISCOVERY OF X-RAYS

LECTURE 1 The Radiographic Image

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

Teaching Digital Radiography and Fluoroscopic Radiation Protection

RADIOGRAPHIC EXPOSURE

Nuclear Associates

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

Physical characteristics of four roentgenographic chains proposed for chest radiography

Film Replacement in Radiographic Weld Inspection The New ISO Standard

Engineering Policy & Procedure

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

American University of Beirut Faculty of Health Sciences Medical Imaging Sciences MIMG 203. (Medical Imaging Equipment I) Fall Semester

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

X-ray Tube and Generator Basic principles and construction

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

X-RAY. Lecture No.4. Image Characteristics:

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

P UBLIC Law passed by the Congress in 1969 and often referred to

Objective Evaluation of Radiographic Contrast- Enhancement Masks

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Exposure System Selection

Overview. Professor Roentgen was a Physicist!!! The Physics of Radiation Oncology X-ray Imaging

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Nuclear Associates

Radiographic testing: Increased detection sensitivity using optimum source to object distance

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Acquisition, Processing and Display

Fig.2: Scanner VistaScan for image plates

X-RAYS - NO UNAUTHORISED ENTRY

NDE SOLUTIONS RADIOGRAPHY COURSE OUTLINE

ZANCO Journal of Pure and Applied Sciences. Image Quality Assessment of Diagnostic X-ray machines of Different Hospitals in Erbil

CR Basics and FAQ. Overview. Historical Perspective

Amorphous Selenium Direct Radiography for Industrial Imaging

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Determination ofradiological Quality Parameters Using Optical Densitometer andsimple Fabricated Equipment

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

Ayaz Jhanorwala, Rishikesh Kumar, Satish Tilva. L & T MHPS Boilers Pvt Ltd., Hazira, Surat , India

RADIOGRAPHIC EXAMINATION PROCEDURE

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

QC Testing for Computed Tomography (CT) Scanner

Attikon, Rimini 1, , Athens, Greece , Athens, Greece , Athens, Greece

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

PD233: Design of Biomedical Devices and Systems

of sufficient quality and quantity

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Do you have any other questions? Please call us at (Toll Free) or , or

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System

JEFFERSON COLLEGE. Radiographic Exposures

U.S. ARMY MEDICAL DEPARTMENT CENTER AND SCHOOL FORT SAM HOUSTON, TEXAS PRINCIPLES OF RADIOGRAPHIC EXPOSURE SUBCOURSE MD0952 EDITION 200

ProX Intraoral X-ray. PLANMECA is proud to introduce a new intraoral X-ray unit to its comprehensive collection of imaging products- the ProX.

2017 West Coast Educators Conference Orlando. Projection Geometry. 1. Review hierarchy of image qualities (amplified version):

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

Principle of X-Ray Systems

Nuclear Associates

(CR) PORTABLE CHEST EXAM Baseline Definition Worksheet

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Your worldwide source of grids. What a difference a grid makes

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

Digital Imaging Considerations Computed Radiography

Nuclear Associates , , , , , ,

High-Resolution Radiographs of the Hand

STUDIES ON THE STATUS OF LIGHT BEAM DIAPHRAGMS IN CALABAR: EFFECTS AND IMPLICATIONS ON RADIATION PROTECTION

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

MOBILEART eco C504-E009G

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)

ILFORD FP4 PROCESS CONTROL STRIPS & FILM PROCESS CONTROL

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors

10/3/2012. Study Harder

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

TILMAN EHRENSTEIN DIAGNOSTIC IMAGING

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Nuclear Associates

INTRODUCTION: D = Log I / I (1)

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

JMSCR Vol 06 Issue 12 Page December 2018

10/26/2015. Study Harder

Transcription:

Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic I. Musdalifah 1,*, S. Suryani 2, N. Rauf 2 1 Academy of Radiodiagnostic Engineering and Radiotherapy of Muhammadiyah, Makassar Indonesia 2 Department of Physics, Hasanuddin University, Makassar 90245, Indonesia E-mail: * indahmusdalifah86@yahoo.com ABSTRACT Research has been conducted to determine the radiographic density value resulting from tube voltage variation (kv) and current tim e strength (mas) with 15% rule. The method used in this research is descriptive method with direct experiment. The study used tubular (kv) and strong current time times (mas) with a 15% rule of 60 kv 10 mas, 51 kv 20 mas, 69 kv 5 mas, 80 kv 6 mas, 68 kv 12 mas and 92 kv 3 mas. Instead of the object is a step wedge. The results of this study can be concluded that the density value with the use of 15% rule on radiography is no difference. Keywords: Tube Voltage, Strong Flow and Exposure Time,, Rule 15% Article history: received January 10 th 2018, received in revised May 05 th, 2018 1. INTRODUCTION In line with the development of science and technology in the field of health, technology in the field of radiology also experienced a drastic change. Beginning with the discovery of X-rays by German physicist Wilhelm Conrad Roentgen in 1895 [1]. The first object to be Roentgen's experiment was his wife's hand that was wearing a metal ring. But the first radiographic image produced by Roentgen certainly does not have the same qualities compared to the quality of the radiographs produced in today's sophisticated times. The quality of a radiograph is determined by the exposure factor setting. The exposure factor is a factor that determines the quality and quantity of X-ray radiation radiation required in the manufacture of radiography. The exposure factor is the factor that influences and determines the quality and quantity of X-ray radiation radiation required in the manufacture of radiography [7]. Exposure factors are divided into two parts, namely the main factor and factors that affect the system characteristics of radiography. According to Bushong ( 245-250) [4], the main factors affecting the quality and quantity of X-rays are kvp (kilovolt peak) or kv, ma (milli Ampere), s (second), and distance (FFD) while influencing factors on radiographic characteristic system ie focal spot size, filter use, high level generator. In addition 5

there are several other factors that also affect the formation of radiography, the collimation, the use of grids, the use of a combination of screen film and film prosessing [6]. Giving exposure factor to get certain density value is used some variation choice such as tube voltage variation (kv), milliampere times second (mas), FFD (focus film distance), filter usage, radiation field (collimation) and grid usage. Another literature is also found a rule of exposure value variation known as the 15% rule, which states that if kv is increased by 15% to produce the same density on the radiograph then mas is halved whereas if kv is decreased by 15% to produce the same density on radiography then mas duplicated [5]. The use of this 15% rule is to maintain the same density value on a radio. Kilovolt (kv) not only affects the density but also other aspects of the radiograph. So kilovolt (kv) is not a major factor in manipulating radiography in density changes. However, it is sometimes necessary to manipulate kilovolt (kv) to maintain or adjust the density [10]. Maintaining or adjusting radiographic density can be achieved using the 15% rule. The 15% rule is a rule that states that if kv is raised by 15% to produce the same density on the radiograph then mas is halved whereas if kv is decreased by 15% to produce the same density on radiography then mas is doubled. The change in tube voltage is the preferred method to change the desired density of the image, as a result by using the 15% rule will always change the contrast of the image. Then formulated as follows by increasing kv by 15% then mas decreased by half. If the kv decreases by 15% then the mas is doubled. The effects of changes in kv are not uniform across the kv range. When low or high kv is used, the amount of kv change necessary to maintain the density may be more or less than 15% [5]. To meet the high quality of the radiograph image, a radiograph must meet several aspects to be assessed on a radiograph that is density, contrast, sharpness and detai [8]. A. is the degree of blackness in the film. The result of the exposure of the film after being processed produces a blackening effect because it corresponds to the emulsion nature of the film which will darken when exposed. This degree of blackness depends on the level of exposure received in both kv and mas. This density can be measured so that the density itself will have a value. The highest density that can be generated is 4 and the lowest density value starts from 0.2. While the value of density that can be seen with ordinary eye ranged between 0.25-2 [8]. on radiography is defined as follows: description: 6

d density Io rays leading to the film It rays passed to the film B. Contrast Contrast is the difference in density between two points. a radiograph has a good contrast if it can be distinguished between parts of one another. according to nova rahman [8] film contrast is the contrast produced due to the nature of the film. every movie produced by a company has its own character. There are films that have characters with high film response to exposure by either X-rays or visible light. The film response to this exposure will affect the resulting density value. Films that are highly responsive to exposure will produce higher densities when compared to films that are less responsive to exposure. The radiographic contrast is formulated as follows: Description: C contrast D2 at Region D1 at Region 1 C. Sharpness Sharpness notices how the density changes on the boundary between adjacent areas. The boundary between the two areas that appear can be very sharp, this is because there is a drastic change in the value of density at that limit. The higher the contrast value the sharper the resulting image [9]. The irregularities of the radiograph are influenced by several factors, namely geometric factors, movement factors, and photographic effects consisting of Intensifying screen and parallax. D. Details Detail is the ability to show the sharpness and the smallest structures on a radiograph. In a radiographic examination, there is a portion of the picture that has a very small but very important structure in diagnosis. Factors that affect the details of the size of focal spot size, FFD (Focus Film Distance) and OFD (Object Film Distance) [3-2]. In hospitals the application of this 15% rule is very rare. Most use of exposure factors is limited only to the provisions contained in the aircraft control panel or solely forecast by Radiogafer. From the research conducted on the implementation of Terri L. Fauber's theory on tube voltage variation (k V) and strong time-current (mas) to radiographic density, it was found that density value with the use of 15% rule on radiography was no difference. 2. MATERIALS AND METHODS The research equipment used in this research is x-ray planes, step wedge, Radiographic Cassettes, Automatic Processing, densitometer. In doing this research, the researcher used Fuji film as much as 3 sheets measuring 24x30 cm with step wedge as the object exposed with tube voltage (kv) and different current time time (mas) that is 60 kv 10 mas, 51 kv 20 mas, 69 kv 5 mas, 80 kv 6 7

mas, 68 kv 12 mas and 92 kv 3 mas with the same FFD usage at each exposure of 100 cm. After the film is exposed then it is processed through processing automatic. 3 pieces of film consisting of 6 samples that have been processed then measured the value of each density by using a densitometer tool from step 1 to step 11 on the film. Data processing is done by making table and comparison graph from result of calculation of difference of density value which result from different exposure factor. To support the discussion then used the application of SPSS or Statistical Product and Service Solutions and the method of sensitometry with analysis of the characteristic curve using Microsoft Office Excel program. This research is done in accordance with the scheme of the research flow below. 3. RESULT From result of measurement done by researcher then obtained result of measurement from population 1 for density value from sample 1 by using kv and mas base that is 60 kv and 10 mas as in table 1. Table 1. Results of measurement of density values from sample 1. Fog 1 2 3 Average Level 1 0,67 0,65 0,70 0,67 0,41 2 0,87 0,86 0,92 0,88 0,41 3 1,23 1,22 1,26 1,23 0,41 4 1,65 1,65 1,68 1,66 0,41 5 2,05 2,04 2,02 2,03 0,41 6 2,32 2,33 2,29 2,31 0,41 7 2,45 2,46 2,44 2,45 0,41 8 2,51 2,51 2,51 2,51 0,41 9 2,54 2,54 2,54 2,54 0,41 10 2,54 2,55 2,55 2,54 0,41 11 2,54 2,55 2,54 2,54 0,41 Mean 1 1,94 Basic deviation 0,716 The contrast value in each sample can be obtained by using the formula : C Dmax Dmin Thus, the contrast value for sample 1 with the use of 60 kv and 10 mas was 1.87 For sample 2 in population 1 with a decrease of kv 15% and mas doubled ie 51 kv and 20 mas obtained by measurement results according to table 2. 8

Table 2. Results of measurement of density values from sample 2. Fog Level 1 2 3 Average 1 0,47 0,46 0,47 0,47 0,40 2 0,63 0,63 0,64 0,63 0,40 3 0,98 0,98 0,99 0,98 0,40 4 1,52 1,54 1,54 1,53 0,40 5 2,09 2,11 2,09 2,09 0,40 6 2,27 2,26 2,27 2,27 0,40 7 2,36 2,36 2,37 2,36 0,40 8 2,41 2,42 2,41 2,41 0,40 9 2,51 2,52 2,52 2,51 0,40 10 2,54 2,55 2,54 2,54 0,40 11 2,55 2,55 2,55 2,55 0,40 Mean 1 1,84 Basic deviation 0,803 From the above measurement results obtained the contrast of the sample 2 of 2.08 For sample 3 with a 15% increase in kv and a reduction in mas ½ of 69 kv and 5 mas obtained by measurement results according to Table 3. Table 3. Results of measurement of density values of sample 3. 1 2 3 Average Fog Level 1 1,39 1,39 1,36 1,38 0,35 2 1,73 1,73 1,70 1,72 0,35 3 2,01 2,03 2,02 2,02 0,35 4 2,24 2,25 2,24 2,24 0,35 5 2,38 2,39 2,39 2,38 0,35 6 2,46 2,48 2,48 2,47 0,35 7 2,52 2,53 2,53 2,52 0,35 8 2,55 2,55 2,55 2,55 0,35 9 2,56 2,56 2,56 2,56 0,35 10 2,57 2,57 2,56 2,57 0,35 11 2,57 2,57 2,57 2,57 0,35 Mean 1 2,27 Basic deviation 0,401 9

From table 3 we get contrast value equal to 1,19. Here is a graph of the average density measurement results of samples 1, 2 and 3 ie Fig 1. Graph of Radiographic of Population 1 For the result of measurement from population 2 obtained table as follows. Table 4. The result of measurement of density value from sample 4 by using kv and base mas is 80 kv and 6 mas. Fog 1 2 3 Average Level 1 2,49 2,49 2,48 2,48 0,50 2 2,56 2,57 2,56 2,56 0,50 3 2,60 2,60 2,60 2,60 0,50 4 2,63 2,63 2,63 2,63 0,50 5 2,64 2,64 2,64 2,64 0,50 6 2,64 2,64 2,65 2,64 0,50 7 2,65 2,65 2,65 2,65 0,50 8 2,65 2,65 2,65 2,65 0,50 9 2,66 2,66 2,66 2,66 0,50 10 2,66 2,66 2,66 2,66 0,50 11 2,66 2,67 2,67 2,67 0,50 Mean 1 2,62 Basic deviation 0,056 Table 5. Results of measurement of density values from sample 5 with decreased kv 15% and mas duplicated ie 68 kv and 12 mas Fog 1 2 3 Average Level 1 2,30 2,32 2,31 2,31 0,35 2 2,45 2,46 2,45 2,45 0,35 3 2,54 2,55 2,55 2,54 0,35 4 2,60 2,60 2,60 2,60 0,35 5 2,62 2,62 2,62 2,62 0,35 6 2,64 2,63 2,64 2,63 0,35 7 2,65 2,64 2,65 2,64 0,35 8 2,65 2,64 2,65 2,64 0,35 9 2,65 2,65 2,65 2,65 0,35 10 2,66 2,65 2,65 2,65 0,35 11 2,65 2,66 2,66 2,65 0,35 Mean 1 2,58 Basic deviation 0,108 From table 4 in population 2 obtained a contrast value of 0.19. 10

Table 6. The result of measuring the density value of sample 6 with a 15% increase in KV and a reduction of ½ mas of 92 kv and 3 mas Averag Fog Level 1 2 3 e 1 2,58 2,59 2,59 2,58 0,84 2 2,62 2,63 2,62 2,62 0,84 3 2,65 2,64 2,65 2,64 0,84 4 2,65 2,65 2,65 2,65 0,84 5 2,66 2,65 2,65 2,65 0,84 6 2,66 2,66 2,66 2,66 0,84 7 2,66 2,66 2,66 2,66 0,84 8 2,67 2,67 2,66 2,67 0,84 9 2,67 2,67 2,67 2,67 0,84 10 2,67 2,67 2,67 2,67 0,84 11 2,67 2,67 2,67 2,67 0,84 Mean 1 2,64 Basic deviation 0,027 Fig. 2. Graph of Radiographic of Population 2 11

The contrasting value of sample 5 with the use of 68 kv and 12 mas was 0.34. The contrast value of sample 6 with the use of 92 kv and 3 mas was 0.09. Here is a graph of the average density measurements of each sample of 4.5, and 6 ie. An overview of the tables and curves of samples 1,2 and 3 in population 1 shows a slight difference in the resulting density values. So to ensure there is no significant difference then the researchers conducted a statistical test on the density value of 3 samples in the population 1. First of all researchers test the data normality that aims to determine the type of statistical tests to be used next. Normality test using Shapiro-wilk test because sample less than (<) 50. After normality test is done significant value <0,05 so that data is not normally distributed. then the researchers do the next statistical test Kruskall-Wallis H test to see the difference in the density value of the 3 samples. After testing the significant value is 0.240. Because the significance value> 0,05 then it is concluded that the three data there is no difference. In the table and curve of samples 4, 5 and 6 for population 2, there was no significant difference in density value, no statistical tests were required and it was concluded that there were no differences between the films exposed to both basic kv and mas using the 15% rule. So the relationship of high use of tube voltage (kv) with contrast value is concluded that the higher the use of kv, the higher the density value produced while the lower the contrast. 4. CONCLUSION Based on the observations made by the testers it can be concluded that there is no difference in the average density generated, either by the film exposed by using tube voltage (kv) and the strong base time current (mas) or using the tube voltage (kv) and the current strength time (mas) 15% rule. This means that the resulting density is the same. High kv usage will result in higher density value while the contrast value is lower. REFERENCES [1] Akhadi, Mukhlis. 2001. Pengantar Dasar-dasar Proteksi Radiasi. Jakarta:Rineka Cipta. [2] Sunyoto, Danang., (2012). Statistik Kesehatan, (1). Yogjakarta:Nuha Medika. [3] Bushong, S. Carlyle., (2008). Radiologic Science for Technologist, (9). Canada:Mosby Elsevier. [4] Meredith, W. J. dan Massey, J. B., (1977). Fundamental Physics of Radiology, Bristol:John Wright and So ns Ltd. [5] Fauber, Terri L., (2000). Radiographic Imaging and Exposure. Toronto:Mosby Elsevier. [6] Sartinah, dkk. (2008). Jurnal Berkala Fisika Volume 11, No.2, Hal 45-52, tersedia pada http://eprints.undip.ac.id/2293 diakses 21 April 2018 pukul 10.30. [7] Rahman, Nova, (2009). Radiofotografi. Padang:Universitas Baiturrahmah [8] Fridawanty, Astuty, (2012). Jurnal Variasi Pemilihan Faktor Expose terhadap Kontras pada Teknik Radiografi 12

Jaringan Lunak, tersedia pada http://repository.unhas.ac.id/ diakses 25 April 2018 Pukul 14.10 [9] Bushong, S. Carlyle., (2001). Radiologic Science for Technologist, (7). Toronto:Mosby Company [10] Ballinger, P. W dan Frank, E. D., (2003). Merril s Atlas Of Radiographic Positions and Radiologic Procedures, Volume I : 10. USA:Mosby. 13

14