A NEW BROADBAND PULSED HIGH VOLTAGE MONITOR *

Similar documents
FLASH X-RAY (FXR) ACCELERATOR OPTIMIZATION BEAM-INDUCED VOLTAGE SIMULATION AND TDR MEASUREMENTS *

ANALYSIS OF A PULSED CORONA CIRCUIT

DEVELOPMENT OF AN ULTRA-COMPACT EXPLOSIVELY DRIVEN MAGNETIC FLUX COMPRESSION GENERATOR SYSTEM

An experimental system was constructed in which

PULSED BREAKDOWN CHARACTERISTICS OF HELIUM IN PARTIAL VACUUM IN KHZ RANGE

FLASH X-RAY (FXR) ACCELERATOR OPTIMIZATION INJECTOR VOLTAGE-VARIATION COMPENSATION VIA BEAM-INDUCED GAP VOLTAGE *

ANALYSIS OF SWITCH PERFORMANCE ON THE MERCURY PULSED- POWER GENERATOR *

INVESTIGATION OF A HIGH VOLTAGE, HIGH FREQUENCY POWER CONDITIONING SYSTEM FOR USE WITH FLUX COMPRESSION GENERATORS

TRANSMISSION LINE AND ELECTROMAGNETIC MODELS OF THE MYKONOS-2 ACCELERATOR*

PHASING CAPABILITY. Abstract ARRAY. level. up to. to 12 GW. device s outpu antenna array. Electric Mode. same physical dimensions.

PERFORMANCE OF A 10 KV, 625 KA, 85 KJ ENERGY DISCHARGE MODULE UTILIZING A SOLID DIELECTRIC SWITCH.*

Strategic Technical Baselines for UK Nuclear Clean-up Programmes. Presented by Brian Ensor Strategy and Engineering Manager NDA

PULSED POWER SWITCHING OF 4H-SIC VERTICAL D-MOSFET AND DEVICE CHARACTERIZATION

David L. Lockwood. Ralph I. McNall Jr., Richard F. Whitbeck Thermal Technology Laboratory, Inc., Buffalo, N.Y.

OPTICAL EMISSION CHARACTERISTICS OF HELIUM BREAKDOWN AT PARTIAL VACUUM FOR POINT TO PLANE GEOMETRY

RAVEN, A 5 kj, 1.5 MV REPETITIVE PULSER* G. J. Rohwein Sandia National Laboratories Albuquerque, New Mexico 87185

Experimental Observation of RF Radiation Generated by an Explosively Driven Voltage Generator

MEASUREMENTS OF THE RADIATED FIELDS AND CONDUCTED CURRENT LEAKAGE FROM THE PULSED POWER SYSTEMS IN THE NATIONAL IGNITION FACILITY AT LLNL

Development of a charged-particle accumulator using an RF confinement method FA

Investigation of a Forward Looking Conformal Broadband Antenna for Airborne Wide Area Surveillance

SILICON CARBIDE FOR NEXT GENERATION VEHICULAR POWER CONVERTERS. John Kajs SAIC August UNCLASSIFIED: Dist A. Approved for public release

8.2. Washington, D. C delivered 65 kj into a matched load with 63 ns FWHM. Peak power was about 1 TW.

"OPTIMAL SIMULATION TECHNIQUES FOR DISTRIBUTED ENERGY STORE RAILGUNS WITH SOLID STATE SWITCHES"

IREAP. MURI 2001 Review. John Rodgers, T. M. Firestone,V. L. Granatstein, M. Walter

9.4 A HIGH CURRENT PULSER FOR EXPERIMENT 11225, "NEUTRINO ELECTRON ELASTIC SCATTERING" C. Dalton, G. Krausse, and J. Sarjeant

Report Documentation Page

COMPACT FLASH X-RAY UNITS. Abstract

ULTRA FAST, HIGH REP RATE, HIGH VOLTAGE SPARK GAP PULSER

0.18 μm CMOS Fully Differential CTIA for a 32x16 ROIC for 3D Ladar Imaging Systems

A COMPACT, 1-MV, 6-kA RADIOGRAPHY SOURCE WITH A ONE- METER EXTENSION AND RIGHT-ANGLE BEND

Fuse and Load Testing With Mid-Sized, High Energy Density Flux Compression Generators

W. L. Bird University of Texas, Austin, Center for Electro-Mechanics, Taylor Hall 167, Austin, TX 78712

Signal Processing Architectures for Ultra-Wideband Wide-Angle Synthetic Aperture Radar Applications

Adaptation of ASTERIX to Positive Polarity for 2 to 4-MV Rod-Pinch Diode Experiments and Diode Electrical Analysis *

Robotics and Artificial Intelligence. Rodney Brooks Director, MIT Computer Science and Artificial Intelligence Laboratory CTO, irobot Corp

IB2-1 HIGH AVERAGE POWER TESTS OF A CROSSED-FIELD CLOSING SWITCH>:< Robin J. Harvey and Robert W. Holly

Modeling Antennas on Automobiles in the VHF and UHF Frequency Bands, Comparisons of Predictions and Measurements

CHARACTERIZATION OF PASCHEN CURVE ANOMOLIES AT HIGH P*D VALUES

Active Denial Array. Directed Energy. Technology, Modeling, and Assessment

Technology Maturation Planning for the Autonomous Approach and Landing Capability (AALC) Program

EFFECT OF TRANSFORMER LEAKAGE INDUCTANCE ON THE THREE PHASE CAPACITIVE INPUT RECTIFIER

A RENEWED SPIRIT OF DISCOVERY

L. B. Gordon Space Power Institute 231 Leach Center Auburn University, Alabama 36849

UPGRADES TO THE DARHT SECOND AXIS INDUCTION CELLS

[Research Title]: Electro-spun fine fibers of shape memory polymer used as an engineering part. Contractor (PI): Hirohisa Tamagawa

Underwater Intelligent Sensor Protection System

ACCELERATOR FAST KICKER R&D WITH ULTRA COMPACT 50MVA NANO-SECOND FID PULSE GENERATOR

COM DEV AIS Initiative. TEXAS II Meeting September 03, 2008 Ian D Souza

FAA Research and Development Efforts in SHM

Non-Data Aided Doppler Shift Estimation for Underwater Acoustic Communication

D. M. Barrett University of California Lawrence Livermore National Laboratory Livermore, CA (b) PFL Voltage at Point A. (c) Voltage Across Load

MINIATURIZED ANTENNAS FOR COMPACT SOLDIER COMBAT SYSTEMS

DEVELOPMENT OF STITCH SUPER-GTOS FOR PULSED POWER

0.9Vo II. SYNTHESIZER APPROACH

Lattice Spacing Effect on Scan Loss for Bat-Wing Phased Array Antennas

Solar Radar Experiments

VHF/UHF Imagery of Targets, Decoys, and Trees

Pulse Power Performance of the Cygnus 1 and 2 Radiographic Sources

Advances in SiC Power Technology

Innovative 3D Visualization of Electro-optic Data for MCM

Modeling of Ionospheric Refraction of UHF Radar Signals at High Latitudes

Remote Sediment Property From Chirp Data Collected During ASIAEX

Loop-Dipole Antenna Modeling using the FEKO code

Frequency Stabilization Using Matched Fabry-Perots as References

Mathematics, Information, and Life Sciences

Durable Aircraft. February 7, 2011

Student Independent Research Project : Evaluation of Thermal Voltage Converters Low-Frequency Errors

Key Issues in Modulating Retroreflector Technology

14. Model Based Systems Engineering: Issues of application to Soft Systems

REGULATED CAPACITOR CHARGING CIRCUIT USING A HIGH REACTANCE TRANSFORMER 1

HIGH TEMPERATURE (250 C) SIC POWER MODULE FOR MILITARY HYBRID ELECTRICAL VEHICLE APPLICATIONS

Thermal Simulation of a Silicon Carbide (SiC) Insulated-Gate Bipolar Transistor (IGBT) in Continuous Switching Mode

Smart Sensor Network for Aircraft Corrosion Monitoring

REPORT DOCUMENTATION PAGE

Wavelength Division Multiplexing (WDM) Technology for Naval Air Applications

NEURAL NETWORKS IN ANTENNA ENGINEERING BEYOND BLACK-BOX MODELING

Rump Session: Advanced Silicon Technology Foundry Access Options for DoD Research. Prof. Ken Shepard. Columbia University

EVALUATION OF RESISTORS FOR TRANSIENT HIGH-VOLTAGE APPLICATIONS

DESIGN OPTIONS FOR A PULSED-POWER UPGRADE OF THE Z ACCELERATOR *

DIELECTRIC ROTMAN LENS ALTERNATIVES FOR BROADBAND MULTIPLE BEAM ANTENNAS IN MULTI-FUNCTION RF APPLICATIONS. O. Kilic U.S. Army Research Laboratory

A R Miller Maxwell Laboratories, Inc Balboa Ave., San Diego, CA

EFFECTS OF ELECTROMAGNETIC PULSES ON A MULTILAYERED SYSTEM

A Comparison of Two Computational Technologies for Digital Pulse Compression

UNCLASSIFIED UNCLASSIFIED 1

Range-Depth Tracking of Sounds from a Single-Point Deployment by Exploiting the Deep-Water Sound Speed Minimum

-6 MV VACUUM VOLTMETER DEVELOPMENT

DoDTechipedia. Technology Awareness. Technology and the Modern World

Department of Energy Technology Readiness Assessments Process Guide and Training Plan

JOCOTAS. Strategic Alliances: Government & Industry. Amy Soo Lagoon. JOCOTAS Chairman, Shelter Technology. Laura Biszko. Engineer

Design, Construction, and Testing of an Inductive Pulsed-Power Supply for a Small Railgun

Research on High Power Railguns at the Naval Research Laboratory

ANTENNA DEVELOPMENT FOR MULTIFUNCTIONAL ARMOR APPLICATIONS USING EMBEDDED SPIN-TORQUE NANO-OSCILLATOR (STNO) AS A MICROWAVE DETECTOR

REPORT DOCUMENTATION PAGE. Thermal transport and measurement of specific heat in artificially sculpted nanostructures. Dr. Mandar Madhokar Deshmukh

SA Joint USN/USMC Spectrum Conference. Gerry Fitzgerald. Organization: G036 Project: 0710V250-A1

Coherent distributed radar for highresolution

Design of Synchronization Sequences in a MIMO Demonstration System 1

The Energy Spectrum of Accelerated Electrons from Waveplasma Interactions in the Ionosphere

Modeling an HF NVIS Towel-Bar Antenna on a Coast Guard Patrol Boat A Comparison of WIPL-D and the Numerical Electromagnetics Code (NEC)

Basic Studies in Microwave Sciences FA

CALIBRATION OF THE BEV GPS RECEIVER BY USING TWSTFT

Transcription:

A NEW BROADBAND PULSED HIGH VOLTAGE MONITOR * W. R. Cravey, Bob Anderson, Paul Wheeler, Dave Kraybill, Nicole Molau, and Deborah Wojtowicz University of California, Lawrence Livermore National Laboratory Livermore, California 94550 Abstract Experimental results of a new GHz-bandwidth diagnostic for measuring pulsed high voltage are presented. The probe is a small helical coil that self-magnetically insulates to prevent turn-to-tum flashover. A small external B-dot sensor measures!-dot, the time rate of change of current in the probe. The voltage to be measured is equal to L Idot. This paper describes the experimental apparatus used to measure a 1 MV, 60 ns pulse with a rise time of 1.5 ns generated by LLNL's 2MV accelerator 1. Excellent agreement is obtained when comparing the output of the new probe with a conventional resistive divider technique. The major difference in the temporal shape of the waveforms was determined to be caused by the bandwidth limitation of the resistive divider probe. Introduction The measurement of a high voltage fast rise time pulse across a vacuum gap is a technical challenge. Capacitive dividers are sensitive to stray charges and are difficult to calibrate. Resistive dividers have limited bandwidth since they must be long and fabricated very carefully to avoid surface flashover. This paper describes two experiments using a new, high bandwidth, high voltage probe. The first experiment was a high impedance test and was designed to use the probe 2 to measure a high voltage pulse at the end of a vacuum insulated transmission line (VITL). Principles of operation, failure modes, and frequency response of the probe where a few of the questions to be answered from the high impedance experiment. A low impedance experiment was also performed in which the probe was used to measure the voltage across a vacuum diode at the end of a three meter magnetically insulated pulse sharpening line (MITL). The MITL and the VITL were driven by the LLNL 2 MV accelerator. Principles of Operation Except for small terms due to inductance of leads and the leakage inductance between turns, the inductance of a wire-wound helical coil depends only on its radius, length, and number of turns, not on the wire diameter. Therefore, even if the wire should expand from heating, or if it is magnetically pinched, the inductance of the helix remains relatively constant. Since the inductance is constant, the output voltage is L I-dot. Turn-to-tum flashover is prevented by having sufficient spacing between turns and by limiting the impedance across the gap (including the coil inductance). Then, when the electric field at the wire surface reaches field-emission threshold, there will be sufficient magnetic flux between adjacent turns to magnetically insulate against turn-to-turn flashover. Experimental Setup High Impedance Teat The first experiment was conducted on a vacuum insulated transmission line (VITL), see Figure 1. The VITL was driven by a 1 MV, 60 ns, 1.5 ns rise time pulse from the LLNL 2 MV accelerator 1. The compact voltage probe (CVP) was connected across the vacuum end of the line. Diagnostics used on the line included resistive voltage probes, B-dot sensors, and a framing camera. Two voltage probes were used on the experiment, a proven resistive divider probe and the new probe. High speed optical diagnostics were used to measure the coil's mechanical movement as a function of time and the onset of flashover phenomena. The resistive voltage probe was used as a voltage monitor for comparison with the CVP output. B-dot probes were placed on the VITL to measure the current through the line. Failure Modes Two failure tests were conducted on the CVP to check its robustness against flashover. First, the self-magnetic field was reduced by increasing the inductance of the CVP by increasing the turns radius of the coil. The CVP was found to self insulate as predicted 2. * This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under Contract No. W-7405-ENG-48 919

Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated to average 1 hour per response, including the time for reviewing instructions, searching existing data sources, gathering and maintaining the data needed, and completing and reviewing the collection of information. Send comments regarding this burden estimate or any other aspect of this collection of information, including suggestions for reducing this burden, to Washington Headquarters Services, Directorate for Information Operations and Reports, 1215 Jefferson Davis Highway, Suite 1204, Arlington VA 22202-4302. Respondents should be aware that notwithstanding any other provision of law, no person shall be subject to a penalty for failing to comply with a collection of information if it does not display a currently valid OMB control number. 1. REPORT DATE JUN 1989 2. REPORT TYPE N/A 3. DATES COVERED - 4. TITLE AND SUBTITLE A New Broadband Pulsed High Voltage Monitor 5a. CONTRACT NUMBER 5b. GRANT NUMBER 5c. PROGRAM ELEMENT NUMBER 6. AUTHOR(S) 5d. PROJECT NUMBER 5e. TASK NUMBER 5f. WORK UNIT NUMBER 7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) University of California, Lawrence Livermore National Laboratory Livermore, California 94550 8. PERFORMING ORGANIZATION REPORT NUMBER 9. SPONSORING/MONITORING AGENCY NAME(S) AND ADDRESS(ES) 10. SPONSOR/MONITOR S ACRONYM(S) 12. DISTRIBUTION/AVAILABILITY STATEMENT Approved for public release, distribution unlimited 11. SPONSOR/MONITOR S REPORT NUMBER(S) 13. SUPPLEMENTARY NOTES See also ADM002371. 2013 IEEE Pulsed Power Conference, Digest of Technical Papers 1976-2013, and Abstracts of the 2013 IEEE International Conference on Plasma Science. Held in San Francisco, CA on 16-21 June 2013. U.S. Government or Federal Purpose Rights License. 14. ABSTRACT Experimental results of a new GHz-bandwidth diagnostic for measuring pulsed high voltage are presented. The probe is a small helical coil that self-magnetically insulates to prevent turn-to-tum flashover. A small external B-dot sensor measures!-dot, the time rate of change of current in the probe. The voltage to be measured is equal to Lâ Idot. This paper describes the experimental apparatus used to measure a 1 MV, 60 ns pulse with a rise time of 1.5 ns generated by LLNL s 2MV accelerator1. Excellent agreement is obtained when comparing the output of the new probe with a conventional resistive divider technique. The major difference in the temporal shape of the waveforms was determined to be caused by the bandwidth limitation of the resistive divider probe. 15. SUBJECT TERMS 16. SECURITY CLASSIFICATION OF: 17. LIMITATION OF ABSTRACT SAR a. REPORT b. ABSTRACT c. THIS PAGE 18. NUMBER OF PAGES 4 19a. NAME OF RESPONSIBLE PERSON Standard Form 298 (Rev. 8-98) Prescribed by ANSI Std Z39-18

Next, the electric field breakdown of the CVP was tested by reducing the gap between turns on the coil. The gap was reduced by using a ribbon conductor instead of a round wire, Figure 2. Both tests were successful in showing that the coil remained in magnetic insulation, and there was no observed electrical field breakdown. Frequency Response One of the most distinguishing characteristics of the probe is its bandwidth. The frequency response of the probe was measured using the cold test setup shown in Figure 3. The CVP was setup in the same manner as in the MITL experiment, except the line was driven with a 35 volt pulse with a rise time of 100 ps. The input of the line signal was measured with a 7854 oscilloscope. The output was taken from a B-dot probe near the CVP. The B-dot probes were calibrated in an air line. The measured bandwidth was approximately 4 GHz. Knowing that the output signal is proportional to!-dot, VOUT was determined. The input was deconvolved from VOUT to obtain the impulse response. Analysis of this data indicates the bandwidth of the new probe to be 2.5 GHz, see Figure 4. Lifetime Tests A framing camera was used to measure the lifetime of the coil by observing its mechanical change. Figure 5 shows the results of one test for a time of 1.3 microseconds after the CVP was excited. The CVP was found to stay intact for several microseconds which was adequate for the intended use. also measured without the CVP on the line. Illustrated in Figure 7 is the output of the Faraday cup with and without the CVP on the line. The current flow past the probe was inhibited by approximately a factor of 2. Temporal Characteristics Temporal characteristics were compared by integrating the B-dot output,which is proportional to the voltage, and comparing this output with the voltage at the probe. The output of the B-dot was integrated in two ways: digitally from the oscilloscope signal, and with an RC integrator placed in series with the line. Output wave forms of all three signals are shown in Figure 8. Excellent agreement was found between the temporal characteristics of all three signals. The high frequency structure appearing on the pulse is due to the probe's higher bandwidth. Linearity The linearity of the probe was checked by varying the output voltage of the accelerator and comparing this change in output with the output of the CVP. This test showed that the compact voltage probe was linear over a broad range of line voltages which is of great importance in high voltage measurements. The results of this test are illustrated in Figure 9. Conclusion Experimental Layout Low Impedance Teat A 6 Q, 3 meter, magnetically insulated transmission line (MITL) was connected to the the output of the LLNL 2 MV accelerator for the low impedance test of the compact voltage probe. The CVP was connected to the MITL via a low inductance connection as shown in Figure 6. Diagnostics for the low impedance tests consisted of B-dot probes, x-ray diodes, a RC integrator and a Faraday Cup. Disruption of MITL Power Flow The CVP-VITL experiment was successful in showing that the probe does magnetically insulate and is robust to designed electrical fields, against wire mechanical changes, and to higher impedances(lower insulating fields). The results of comparing the temporal waveforms from the two voltage probes was excellent. The new probe shows more high frequency structure on the pulse, which is attributed to the probe's higher bandwidth of 2.5 GHz. The probe was found to stay intact for a long period of time (i.e. microseconds). The CVP also performed well in the low impedance tests. The measured signal gain was linear over greater than a factor of 2. Overall the probe preformed as expected and the result were excellent. It was of interest to know how the power flow of the MITL would be interrupted by the CVP. The power flow test was done by recording the current of the MITL using a Faraday Cup at the end of the line past the CVP as shown in Figure 6. The current was 920

Reference 1. "Megavolt Accelerator", Freytag, E. K. and Di Capua, M., Thrust Area Report, Lawrence Livermore National Laboratory, UCRL-53700, 1985. 2. Inciuctiye Voltage Diyider, Wheeler, Paul C., Lawrence Livermore National Laboratory, UCID 20521, 1985. Figure 1. High impedance vacuum insulated transmission line. Magnetic insulation failure test: Reduce self magnetic field by increasing inductance Increasing turns radius TF ~ ~ - - - r r r r=l.2 r 4 =3.0 r=s.o Electric field breakdown test: Increase electric field by reducing turn-to-tum spacing From Cold Tests We get estimates of the probe bandwidth eo(probo) = u(l)@ Sys(l) @ probe(l) We idemify cavity modes that caused the ringing signal during the experiment Figure 2. Failure mode tests. Figure 3. Schematic of cold test. C'IP TIIH RESPONSE frow A VOLTAGE ST[P INPUT CVP rr 0UOIC' RESPONSE froi.i A VOllo\Gf Sll\' lt1plll f llt1 1l p 1 nee b~ l!!rt bdlrcda fihouency (Gill) Figure 4. Response from cold test data. 921

.. I 111111 -------+-IJft~ Figure 5. Framing camera photograph taken 1.3 ~s after voltage pulse with a 10 nanosecond exposure. Figure 6. Low impedance magnetically insulated transmission line. MITL Output Current Figure 7. Power Flow disruption in MITL. CVP SHOTf44 B-dotf1 [solid] B-dotf7 [AC INT) [dot] B-dotf7 (DIG INT) [dash].b 900 eso 800 750 700 Linearity of Compact Voltage Probe CVP Voltage V. Line Input Voltage.6. 4 Vollage (kv) 650 600 550 500 450.2 50 Figure B. 60 val tage probe. -. 70 BO 90 100 110 120 130 140 150 TIME lnsecl Temporal characteristics of compact 400 350 300 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 Figure 9. Linearity of compact voltage probe. VTL (kv) 922