Performing ultrasound probe quality assurance assessments: A How-to Guide

Similar documents
Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Breast Ultrasound QA Phantom Recommended by Japan Association of Breast and Thyroid Sonology

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Photomultiplier Tube

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

Design Your Performance

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

12/26/2017. Alberto Ardon M.D.

Real-time B-mode ultrasound quality control test procedures a Report of AAPM Ultrasound Task Group No. 1

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

UGEO H60. Performance in Style. Features

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Ultrasound physical principles in today s technology

The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line

CRIMP VS. SOLDER: PROS & CONS

ACOUSTIC MICRO IMAGING ANALYSIS METHODS FOR 3D PACKAGES

4 Working With Scan Modes

Vivid S5. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

A COMPACT SYSTEM WITH ADVANCED PERFORMANCE

Ultrasound & Artifacts

Medical Imaging (EL582/BE620/GA4426)

Basic functions of the universal flaw detector GEKKO

3. Ultrasound Imaging(2)

PULSE ECHO ULTRASOUND IMAGING SYSTEMS: PERFORMANCE TESTS AND CRITERIA

Quick Reference Guide

Why partial discharge testing makes good sense

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

DACON INSPECTION SERVICES. Phased Array Ultrasonic Testing

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

MIL-STD-883H METHOD ULTRASONIC INSPECTION OF DIE ATTACH

Enhanced Resonant Inspection Using Component Weight Compensation. Richard W. Bono and Gail R. Stultz The Modal Shop, Inc. Cincinnati, OH 45241

Technician License Course Chapter 4. Lesson Plan Module 9 Antenna Fundamentals, Feed Lines & SWR

KAERI Feeder Tube Inspection Using EMAT Generated Circumferential Guided Waves

Invisible sophistication. Visible simplicity. CS Welcome to the simplicity of compact panoramic imaging

Detection of Internal OR External Pits from Inside OR Outside a tube with New Technology (EMIT)

InterBOARD TM 12 Channel Transmitter and Receiver Evaluation Board User Guide

The Script of ZST + Presentation. MIS Upstream Marketing Team [ 日期 ]

MIRA Purpose MIRA Tomographer MIRA MIRA Principle MIRA MIRA shear waves MIRA

SMART LASER SENSORS SIMPLIFY TIRE AND RUBBER INSPECTION

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Technician License. Course

Sonic Distance Sensors

Kit for building your own THz Time-Domain Spectrometer

Retrospective Transmit Beamformation. Whitepaper. ACUSON SC2000 Volume Imaging Ultrasound System. Answers for life.

A NOVEL HIGH SPEED, HIGH RESOLUTION, ULTRASOUND IMAGING SYSTEM

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Certified Precision Sheet Metal Operator Level I

Laser Telemetric System (Metrology)

Ultrasound Imaging Ultr Michael Dadd 2007

The importance of partial discharge testing throughout the development and operation of power transformers

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

INTRODUCTION. Have applications for imaging, detection and navigation.

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa. Joanna X.Qiao 1, Matthias Jobst 2

Ultrasonic Phased Array Crack Detection Update

Model OB-ITF Infrared Through-Beam Fiber Optic Optical Barrier Operator s Manual. Version

Understanding Infrared Camera Thermal Image Quality

Pulse Arrested Spark Discharge (PASD) Wiring Diagnostic**

GE Healthcare LOGIQ P3. Staying ahead of the curve

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Ludlum Medical Physics

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

GE Healthcare. Vivid S5. Cardiovascular ultrasound system

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

Surface Deployed / Ground Sensors

How to Avoid Thermal Sensor Damage & Out of Tolerance Conditions

QC Testing for Computed Tomography (CT) Scanner

The History and Future of Measurement Technology in Sumitomo Electric

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.

SAMPLE. Determining the health of your power transformer begins with Transformer Clinic s SAMPLE testing programs.

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

Thermography. White Paper: Understanding Infrared Camera Thermal Image Quality

Crimp Type Winding Connections in Hermetic Motors

CMUT and PMUT: New Technology Platform for Medical Ultrasound Rob van Schaijk

Diagnostic Ultrasound System. Operation Note

YAZAKI SPECIFICATION YNA-IS-001

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

Capabilities of Flip Chip Defects Inspection Method by Using Laser Techniques

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Installation of vibration sensors

Hiding In Plain Sight. How Ultrasonics Can Help You Find the Smallest Bonded Wafer and Device Defects. A Sonix White Paper

BLADE AND SHAFT CRACK DETECTION USING TORSIONAL VIBRATION MEASUREMENTS PART 1: FEASIBILITY STUDIES

SOLDER SLEEVES AIRWORTHINESS BULLETIN. Overview. Repair and Maintenance Techniques for Wiring Systems Date : 29 July 2005.

DC-6. Diagnostic Ultrasound System

NDT-PRO Services expands service offering

Nuclear Associates , , CT Head and Body Dose Phantom

ULTRASONIC IMAGING of COPPER MATERIAL USING HARMONIC COMPONENTS

QL55 Series Luminescence Sensor

UNPACK & ASSEMBLY. Done! CAUTION! THE MILL WILL BE VERY HEAVY - GET ASSISTANCE Pepe Tools.

SUPERSEDED. NOT the LATEST REVISION

CEELOK FAS-T* High-Speed Circular Connector System

Please refer to the figure on the following page which shows the relationship between sound fields.

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

Transcription:

Performing ultrasound probe quality assurance assessments: A How-to Guide

A comprehensive quality assurance program has the potential to directly contribute to better patient outcomes. Regular testing provides a mechanism to monitor probe performance and correct as needed. The end result is a continuous quality improvement program that ensures optimal diagnostic outcomes. As vital as it is to perform regular testing, there is no one standardized process or criteria for effectively field-testing ultrasound probes in the industry. Innovatus Imaging has designed a guide to assist technicians with understanding industry standard testing criteria for assessing probe performance. The guide presents background information on probe design, detailed testing methods, root cause analysis and troubleshooting techniques. WHAT YOU LL LEARN Key terminology Overview of the components and technologies used in modern probes Common quality assurance tests from the most popular accreditation boards including: - American College of Radiology (ACR) - American Institute of Ultrasound in Medicine (AIUM) Best practices and ground rules for performing image quality testing Common testing methods, pass/fail criteria and techniques for troubleshooting image quality problems How to assess probe performance in the field How to acquire and assemble the quality control data required for accreditation

Contents Industry Standards...2 Key Terminology...3 Best Practices...5 Common Testing Methods Image Uniformity...7 Maximum Depth of Penetration and Functional Resolution..10 Geometric Accuracy...13 Spatial Resolution...15 Contrast Resolution...17 Cable Noise Test...19 Electro-Mechanical Functionality...21 Documentation Template...22

Industry Standards With more than 300 diagnostic ultrasound probe models on the market, determining which criteria and methods are best for testing image quality can be confusing. The physical and technical design of one probe can greatly vary from one model to the next. Without a firm knowledge base, consistent methodology and welldefined pass/fail criteria, test results can be highly variable and inconsistent. Standards established by several accreditation boards offer a framework for testing and solid criteria for assessing the performance of devices used within diagnostic ultrasound. THE TWO PRIMARY ACCREDITATION BOARDS American College of Radiology (ACR) American Institute of Ultrasound in Medicine (AIUM) COMMON REQUIREMENTS FOR ACCREDITATION Case study submissions Personnel education and continued training Document storage and record-keeping Policies and procedures to safeguard patients, ultrasound personnel and equipment COMMON FUNCTIONAL TESTS FOR ACCREDITATION Image uniformity Maximum depth of penetration Geometric accuracy Spatial accuracy Contrast resolution Tests for cable noise and electro-mechanical functionality are not addressed by these boards, but offer a more comprehensive performance assessment. To help close the industry s standardization gap, Innovatus Imaging has established a set of guidelines that will help you perform consistent ultrasound probe quality assurance assessments. The goals are to build technical confidence and to minimize variability so that clinicians can trust the tools that they rely upon the most to provide the most accurate diagnosis possible. 2

Key Terminology Each component within the probe plays a vital role in the overall performance of the device. It s critical to understand the purposes and the functions associated with basic probe construction prior to presenting best practices for assessing performance. A failure in one area of the probe may affect the performance of or the results provided by another. ACOUSTIC STACK ELEMENTS Lens Mechanical focus of ultrasound beam Model-specific materials and dimensioning Electrical insulation Fluid barrier Single or multilayer materials, must be ISO 10993 compliant Matching layer(s) Maximizes transmission of energy from acoustic array to tissue by reducing reflection and increases spectrum of frequencies (bandwidth) emitted by probe Commonly consists of one to seven different layers based on probe model Generally ¼ wavelength of the array s center frequency in thickness Acoustic Array Converts electrical energy to mechanical energy (pressure wave) and vice versa Consists of several to thousands of individual elements Backing material Dampens crystal vibrations to reduce pulse duration, which increases axial resolution 3

OTHER PROBE ELEMENTS Flex circuit Flexible circuit board that connects the interconnect board to the individual elements Interconnect/Scanhead electronics Bridge between individual coaxial cables and flex circuit May include multiplexing circuitry, as well as beamforming circuitry Shielding Reduces the effects of electromagnetic interference Typically surrounds the array, the backing material and scanhead electronics Miniature coax Bridge between the connector electronics and the scanhead electronics Internal signal wire enclosed within a braided shield Typically less than the diameter of a hair Model-specific length, impedance and capacitance Main cable Contains several to hundreds of individual coaxial wires Model specific shielding to minimize the effects of electromagnetic interference Strain relief Reduces stress on main and coaxial cables Cable jacket Protects main probe cable 4

Best Practices Quality assurance testing on probes is a key contributor to improving the accuracy of patient diagnoses, treatment plans and ultimately patient outcomes. Routine and consistent testing is vital to ensuring that devices are performing at a level very similar to that when they were purchased. By following these best practices, you ll be able to feel confident in equipment quality. The following can be considered a baseline for beginning the QA process. System and environment Ensure a quality equipment ground. Clean the scanner s probe ports. Disconnect all probes except the one being tested. Disconnect the network cable from the rear of the scanner. Adjust room lighting to typical scanning room intensity. Display brightness and contrast: - Use the system grayscale bar to adjust brightness so the darkest shade of gray is barely visible. - Assure that the grayscale bars are graduated or stepped. - Contrast is adjusted for white whites but not blooming. - Confirm any recommendations or adjustments with the sonographer. Testing devices Specialized high-tech testing devices are not needed to assess probe performance in the clinical setting. A tissue-mimicking phantom MUST be used to properly assess image quality. - The same model and type of phantom should be used between each assessment. - Suppliers include Gammex and ATS. - Measurements should remain consistent between like systems and a single probe or a single system and like probe models over the lifespan of the devices. - Do not attempt to assess image quality viewing a patient scan. - Echogenicity or the ability to bounce an echo varies greatly from person to person. 5

Consistency between testing is key Tests should be conducted always using identical system settings and presets. Example: Every Philips C5-1 should be tested using the same preset. Suggested presets Linear probes Vascular preset Curved probes Abdominal preset Endo-cavity OB present Sector Cardiac preset The number of focal points should be minimal, only one to two. Turn off all options such as harmonic imaging, spatial compounding and image smoothing (these options may mask probe and scanner deficiencies). OEMS have brand specific names for these options, some of which are noted below: Brand specifications Philips Siemens GE THI, SonoCT and xres THI, MultiHz and SieClear Octave, Crossbeam and CHI Periodic inspections should occur at original equipment manufacturer recommended intervals, if not more often. (Accreditation boards recommend at least annually.) Acceptance testing should be performed when: Equipment is new and is being installed Equipment has been in storage and is being returned to service Image-related assemblies, including probes, have been repaired or replaced Visual inspections should be conducted by sonographers daily By addressing minor physical deficiencies, prior to them progressing to a point where they impact image quality, long-term service costs can be minimized. Inspect components for the following concerns: Lens Housing Holes, cuts, missing sealant, bulges, air bubbles, worn corners, separation in materials Cracks, separation, missing sealant, sharp edges Strain reliefs Separation from housing, cuts, holes, excessive stiffness Cable Connector Pin-bank Cuts, holes, exposed wiring, roll-over damage, excessive stiffness Deformations, cracks, missing/malfunctioning hardware Bent pins, corrosion, burn marks, excessive dust/debris 6

How to Perform Field-based Probe Testing The following are instructions on how to conduct common functional tests required for accreditation. Image Uniformity Common names Element testing, channel testing Purpose To perform element-to-element or channel-to-channel comparison Region of interest The entire width of image Scanner settings Frequency Highest possible Depth 3-6 cm of depth Focus Single focal point located in the very near field Gain Adjust overall gain and TGC so that mid-range gray level exists over the entire image Recommended testing method 1 Adjust the depth to 3-6 cm (this is probe dependent) 2 Adjust the number of focal points to one (or the focal zone to the smallest possible size) 3 Adjust the focal zone to the top-most position of the image 4 Adjust the TGCs so that the background is as uniform as possible and is shaded in the middle-range of the grayscale 5 Slowly move the probe back and forth along the phantom s surface and monitor the image on the display for any vertical non-uniformity (weak elements minor dropout will be easier to visualize when the probe is in motion) 7

Visual criteria Use the following information to determine next steps for service: Fine shadow on a single channel: Minor flaw Multiple fine shadows or wide shadow on multiple channels or elements: Major flaw Ranking and action levels Ranking Visual Criteria Potential Impact Action 1 No flaws are present Operating as expected No action required 2 One or two minor flaws are present Considered operational and can be used for scanning Inspect occasionally for possible additional deterioration over time 3 Three or more minor flaws are present Borderline based upon location Replace as soon as convenient 4 Major flaws are present Unacceptable for clinical use Remove from service immediately Minor Major Major 8

Potential root causes for the artifact Probe Acoustic lens Acoustic array Cable/wiring harness Pins/connector Lens delamination: wide shadowing A single or multiple missing or weak elements may show small artifacts Test for intermittencies Inspect regularly: clean dust from pinned connectors, clean pin-less connector interfaces System Connector board Front-end board Inspect regularly: Examine for damaged pins. Clean dust from scanner ports. Test the probe on multiple ports. Test the probe on another scanner to rule-out a probe/ connector issue. 9

Maximum Depth of Penetration and Functional Resolution Common names Maximum visualization, relative penetration or functional resolution Purpose To provide an indication of overall sensitivity of a scanner/probe to detect weak signals Regions of interest Fiber targets Maximum depth Anechoic targets Functional resolution Scanner settings You must use same preset for each individual probe model and same model phantom. Frequency Typical for the probe model Depth Probe dependent Focus Adjusted to meet needs Maximum Depth of Penetration The maximum distance between the top of the image and the deepest vertical target that can be visualized Recommended testing method 1 Position the probe over the vertical group of line targets until a clear image is obtained 2 Adjust the depth to a point no deeper than needed to display as many vertical targets as possible This will be different for each make and model of probe Adjust the focal zone to match the depth of the deepest visible target 3 Utilize cursors to measure the distance between the top of the phantom and the deepest target that can be clearly imaged This distance should remain consistent over the life of the probe 10

Examples 17 cm penetration 8 cm penetration Functional Resolution The maximum distance between the top of the image and the deepest and smallest anechoic target that can be visualized. Recommended testing method 1 Position the probe over the anechoic target structures until a clear image is obtained The actual depth will depend on the specific make and model of the probe 2 Adjust the focal zone to match the depth of the smallest anechoic target that is clearly visible 3 Examine the image and determine the depth of the smallest anechoic target that is clearly visible This distance should remain consistent over the life of the probe Examples 2 mm target at 5 cm 11

Ranking and action levels A 5% decrease in maximum depth of penetration (from the first measurement) is cause of concern A 10% decrease is cause for corrective action to the probe or scanner Potential root causes for change Probe Root cause Lens thickness: improper repair Acoustic lens Fluid/gel infiltration Damage to matching layers Acoustic array Arrays can degrade over time (>8-10 years) System System Inconsistent preset Display settings Environmental Environmental Ambient lighting 12

Geometric Accuracy Common names Measurement accuracy Purpose To verify both horizontal and vertical accuracy Region of interest Respective target groups Scanner settings Frequency Depth Focus Typical for the probe model Probe dependent Single focal point located in the center of the area to be measured Vertical measurement 1 Position the probe above the column of reflector targets in the center of the image 2 Choose two targets separated by a distance that is consistent with the type of probe 3 Place cursors in the middle of each target 4 Measurements should remain consistent between like systems and a single probe or a single system and like probe models over the life span of the devices 13

Horizontal measurement 1 Choose two targets that are separated by a distance of at least half of the image width 2 Place cursors in the middle of each target 3 Verify that the target distance is within 2% or 2mm (whichever is greater) Potential root causes for inaccuracies Probe Acoustic lens Lens thickness, lens material: improper repair Compare measurements between multiple probes of the same model System System Gross inaccuracies may indicate a major system issue Compare measurements between multiple systems of the same model 14

Spatial Resolution Common names Axial/lateral resolution Purpose To verify the minimum distance at which two targets can be individually visualized Region of interest Resolution targets Scanner settings Frequency Depth Focus Typical for the probe model Phantom dependent Single focal point located in the center of the area to be measured Lateral resolution (X) Ability to distinguish between two objects that lie on a line perpendicular to ultrasound beam Varies with depth and focal point(s) Axial resolution (Y) Ability to distinguish between two objects that lie on a line parallel to ultrasound beam Does not vary with depth Elevational resolution (Z) Ability to distinguish between two objects perpendicular to scan plane (slice thickness) Varies with depth 15

Recommended testing method 1 Position the probe over the axial-lateral resolution group 2 Adjust the focal zone to the center (or just above the center of the targets) 3 Examine the image and determine the distance at which the targets are clearly displayed as separate target points The distance should remain consistent over the life of the probe Potential root causes for changes Probe Lens delamination Acoustic lens Lens thickness, lens material: improper repair Compare measurements between multiple probes of the same model System Improper focal point location System Gross inaccuracies may indicate a major system issue Compare measurements between multiple systems of the same model 16

Contrast Resolution Common names Gray scale, gray scale resolution Purpose To provide an indication of the system s and the probe s ability to distinguish between objects of similar and varying densities Region of interest Gray scale targets Scanner settings Frequency Depth Focus Gain Typical for the probe model Dependent upon phantom model Positioned at or right-above targets Adjust overall gain and TGC so that mid-range gray level exists over the entire image Recommended testing method 1 Position the probe over the gray scale target group until a clear image is obtained 2 Adjust the focal zone to the center or just above the center of the target 3 Examine the image 4 All shades of gray should be clearly visualized (background and targets). The targets should appear circular in shape and vary in the degree of brightness ranging from low to high levels of contrast 17

Probe Lens thickness: improper repair Acoustic lens Fluid/gel infiltration Damage to matching layers Compare results between multiple probes of the same model Acoustic array Arrays can degrade over time (>8-10 years) System System Inconsistent preset Display settings Environmental Environmental Ambient lighting 18

Cable Noise Test Purpose To identify intermittencies and breakdowns in signal wiring in the cable harness Region of interest Entire width of image Scanner settings Frequency Depth Cardiac probes Non-cardiac probes Typical for the probe model 6-8 cm CW Doppler mode Color Doppler mode Recommended testing method: cardiac probes 1 Assure that the lens is clean and dry 2 Enable CW Mode or Color Doppler Mode 3 Adjust the volume and/or Doppler Gain to an acceptable level 4 Use the trackball to move the cursor line slightly left or right of center Avoid a straight vertical line 5 Flex each strain relief looking and listening for unacceptable noise 6 Move the cursor line to several other locations and evaluate the noise level 7 If cable has roll-over damage, flex/stress the damaged region Recommended testing method: non-cardiac 1 Assure that the lens is clean and dry 2 Enable color Doppler mode 3 Size and position the color box to span the entire width of the image 4 Depth of color box should span less than 5 cm 5 Adjust the color gain so that there is only very minor speckling in the image 6 Flex each strain relief looking for streaks of flashing color 7 If cable has roll-over damage, flex/stress the damaged region 19

Potential root causes for artifact Probe Wiring harness Intermittent wiring: Heavy static sounds in CW mode, streaks of flashing color in color mode Major wiring damage: intermittent dropout in the 2D imaging 20

Electro-mechanical Functionality Purpose To identify electro-mechanical failures in 3D/4D volumetric probes Region of interest Probe only Internal electro-mechanics Motor Sensor Scanner settings Mode 3D/4D Recommended testing method 1 Visually inspect the dome for abrasions and signs of impact 2 Enable continuous 3D/4D acquisition 3 Apply slight pressure to dome 4 An error message should not be displayed Stress strain reliefs and any area on the cable that shows evidence of roll-over damage or pinching. Potential root cause of failure Probe Impact to dome: internal electro-mechanical damage Probe Motor failure Sensor failure 21

Documentation Template Currently, there is no standardized form within the industry for assessing system or probe performance. Below is an example of a form which can be created using Microsoft Excel. The Innovatus Imaging team can help to customize a form to meet your needs. 22

Performing ultrasound probe quality assurance assessments: A How-to Guide INNOVATUS IMAGING 575 Epsilon Drive Pittsburgh, PA 15238 www.innovatusimaging.com CUSTOMER CARE/ORDERS 1-844-687-5100 customercare@innovatusimaging.com