Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof

Similar documents
Pilot Study EURAMET.AUV.V-P1: Bilateral comparison on magnitude of the complex charge sensitivity of accelerometers from 10 Hz to 10 khz

Continuous development of the national standard for vibration

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz

Vibration Transducer Calibration System

UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES

Calibration of piezoelectric accelerometers at INTI

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

CALIBRATION OF ACCELEROMETERS USING PARAMETER IDENTIFICATION TARGETING A VERSATILE NEW STANDARD

State of progress of dynamic calibration of force, torque and pressure sensors including conditioners

Principles of Vibration Measurement and Analysis. Dr. Colin Novak, P.Eng July 29, 2015

Dynamic measurement activities at PTB Hans-Jürgen von Martens Physikalisch-Technische Bundesanstalt (PTB), Germany

REPORT ON KEY COMPARISON APMP.AUV.A K3

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens

PRODUCT DATA. Piezoelectric Accelerometer Miniature Triaxial DeltaTron Accelerometers Types 4524, 4524 B

A study of Savitzky-Golay filters for derivatives in primary shock calibration

INTERNATIONAL OIML R 103 RECOMMENDATION

Primary vibration calibration by laser interferometry requirements, problems and first experience with a new calibration system

VALIDATION OF A LOW COST SYSTEM FOR VIBRATION MONITORING

Vibration Calibration Technique and basics of Vibration Measurement

EUROMET PROJECT 401 Harmonisation of Audiometry Measurements within Europe

TECHNICAL PROTOCOL FOR AFRIMETS RMO COMPARISON AFRIMETS.AUV.A-S1. Version 4

Physikalisch Technische Bundesanstalt

PRODUCT DATA. Piezoelectric Charge Accelerometer Types 4326-A and 4326-A-001. Uses and Features

Table of Contents. Compendium SPEKTRA. Calibration Systems CS18. Vibration and Shock Exciters. Vibration Control Systems. Services

Model CAACS Commercial Automated Accelerometer Calibration System

Model CAACS Commercial Automated Accelerometer Calibration System

PRODUCT DATA 4397A/ 8A/ 9A. Description. Miniature DeltaTron Accelerometer Type 4397A DeltaTron Accelerometers Types 4398A, 4399A FEATURES USES

Dr. Takashi Usuda (NMIJ), CCAUV President Dr. Gianna Panfilo (BIPM), Secretary

Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition

Introduction to Measurement Systems

Bureau International des Poids et Mesures. International Recognition of NMI Calibration and Measurement Capabilities: The CIPM MRA

TECHNICAL PROTOCOL FOR KEY COMPARISON CCAUV.A-K4

Sensors for Vibration, Acceleration, and Shock Measurement. Product Catalog

COPY FOR INFORMATION ONLY

Enhancing the capability of primary calibration system for shock acceleration in NML

CIPM and CCPR What are these organizations and how do they affect my testing results. Maria Nadal Photometry, Surface Color and Appearance NIST

This document is a preview generated by EVS

Report on key comparison CCAUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

The Importance of Global Metrology for Standards, Industry and Trade: Metrology in a Dynamic World

DEC Country Report. National Institute of Metrology (Thailand) By Mr.Pairoj Rattanangkul

ILAC input to CIPM MRA Review Workshop October 2015

Calibration of High-Voltage Test Equipment

PRODUCT DATA. Piezoelectric Accelerometer Miniature Triaxial Charge Accelerometers Types 4326-A and 4326-A-001

Joint ILAC CIPM Communication regarding the. Accreditation of Calibration and Measurement Services. of National Metrology Institutes.

Accelerometer Mounting

AN5E Application Note

Validation of primary hydrophone calibrations by inter-laboratory comparisons and by independent calibration methods

Introduction to LIVM Accelerometers

Technical Documentation

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005

26 Endevco Isotron accelerometers

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

APPLICATION NOTE How to Deal with Electrical Noise and Interference in the Measuring Chain using a NEXUS

Physikalisch-Technische Bundesanstalt EUROMET supplementary comparison EUROMET.EM-S38 Comparison of small current sources

SMD ENS Metrology - National Standards. Short overview. Hugo Pirée. World Metrology Day 2018 Brussels.

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION

High Sensitivity Triaxial CCLD Accelerometer Type 4506-B-003

Acceleration Sensor AS - 022

Calibration of high-quality reference standards

Assessment of the Metrological Performance of Seismic Tables for a QMS Recognition

Part 1: Fundamentals

Internationally accepted framework for metrology

Development of Shock Acceleration Calibration Machine in NMIJ

Final Report Key Comparison COOMET.EM.BIPM-K10.b. Comparison of the 10 V Josephson Voltage Standards COOMET 542/RU/11. A.S. Katkov 1, P.A.

LASER VIBROMETER CALIBRATION AT HIGH FREQUENCIES USING CONVENTIONAL CALIBRATION EQUIPMENT

Artificial Mastoid Calibration System at NIM, China. ZHONG Bo

Terms and expressions for specifying torque transducers

MIL-STD-202G SHOCK (SPECIFIED PULSE)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005, ANSI/NCSL Z

Reciprocity calibration of Laboratory. Standard Microphones. There is simply no substitute for knowing what you re doing

Free-field calibration of an underwater sound receiver in a laboratory water tank at very low frequencies

Very High Frequency Calibration of Laser Vibrometer up to 350 khz

Ears Project Newsletter No 3. Welcome. C o n t e n t s. Welcome Preface of the coordinator. News & facts Latest news

VARIABLE INDUCTANCE TRANSDUCER

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

INTERNATIONAL STANDARD

Report of the TC Time and Frequency. Ramiz Hamid TC-TF Chair, TÜBİTAK UME, Turkey

Maximizing LPM Accuracy AN 25

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 including Sound Intensity Microphone Pair Type 4197 and Dual Preamplifier Type 2683

MEASUREMENT AND STANDARDS

METROLOGICAL EVALUATION OF AN OPTICAL FIBER ACCELEROMETER FOR POWER TRANSMISSION LINES MONITORING

Chapter 7. Optical Measurement and Interferometry

Publishable JRP Summary for Project T4 J03 JOSY. Next generation of quantum voltage systems for wide range applications

COOMET.AUV.W-S1 supplementary comparison of free-field hydrophone calibrations in the frequency range 250 Hz to 8 khz

Calibration Certificate

New Long Stroke Vibration Shaker Design using Linear Motor Technology

Report on Research Activities of KRISS CCAUV

FINAL REPORT. EUROMET project No Bilateral comparison of DC and AC voltages BEV - NCM. W. Waldmann (BEV, pilot laboratory) P.

SIM.EM S9.b, 1 Ω and 10 kω

Model 626B01. Low Frequency Industrial ICP Accelerometer. Installation and Operating Manual

CCAUV.A-K5, COOMET.AUV.A-K5 and AFRIMETS.AUV.A-K5. Key comparison CCAUV.A-K5

Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge. Ivan Leniček 1, Roman Malarić 2, Alan Šala 3

Voltage Sensors URV5-Z

Organisation Internationale de Métrologie Légale

Reverberation time and structure loss factor

ASC IMU 7.X.Y. Inertial Measurement Unit (IMU) Description.

INTERNATIONAL COMPARISON OF ACCELERATION STANDARDS - PRELIMINARY RESULTS OF SIM PROJECT AV-1

Sensors for Small Forces

SmartSenseCom Introduces Next Generation Seismic Sensor Systems

Transcription:

Final report 2011 04 20 Bilateral Comparison EURAMET.AUV.V-K1.2 (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof Abstract This report describes the results obtained in a bilateral comparison of primary vibration calibration facilities. Two piezoelectric standard accelerometers (one single-ended as well one back-to-back type) were calibrated first at METAS, then at the PTB and again at METAS after their return. The METAS and the PTB calibrations were found to be in excellent agreement, with a mean difference smaller than the expanded uncertainty (k = 2) of the comparison. Metrologia 2011 48 Tech. Suppl.09003

1. Introduction This report summarizes the results obtained in the bilateral comparison EURAMET.AUV.V-K1.2 carried out from October 2008 to April 2009. 2. Participants The following two laboratories participated in the bilateral EURAMET comparison: METAS / Switzerland Hof Christian pilot laboratory Phone: +41 31 32 34 750, Fax: +41 31 32 33210 E-mail: christian.hof@metas.ch Lindenweg 50 3003 Bern-Wabern PTB / Germany Bruns Thomas Phone: +49531 592 1220 Fax: +49531 592 1241 E-mail: thomas.bruns@ptb.de Bundesallee 100 38116 Braunschweig 3. Aim and task of the comparison This bilateral EURAMET comparison of accelerometer calibration was intended as a follow-up to the EUROMET.AUV.V-K1 comparison to underpin the improved measurement capabilities of METAS in the vibration and acceleration field. The EUROMET.AUV.V-K1 comparison was organized to disseminate the key comparison reference values, which were established earlier in the CIPM key comparison CCAUV.V-K1 within the RMO EURAMET. METAS did participate in EUROMET.AUV.V-K1 with calibration facilities on the secondary level. However, after this comparison, METAS put into service a new set-up for primary calibration. Its performance is validated with the comparison reported here. The principal task of the comparison is to measure the charge sensitivity of two accelerometer standards (one of single-ended design and one of back-to-back design) at different frequencies and acceleration amplitudes specified in clause 4. The charge sensitivity is calculated as the ratio of the amplitude of the output Metrologia 2011 48 Tech. Suppl. 09003 2/2

charge of the accelerometer to the amplitude of the acceleration at its reference surface. The reference surface is the base or mounting surface of the accelerometer of single-ended design, and the top surface of the accelerometer of back-to-back design. The charge sensitivity is given in pico coulombs per metres per second squared: pc/(m/s 2 ). To calibrate the two accelerometers, primary vibration calibration by laser interferometry in accordance with ISO 16063-11:1999 was used. A calibrated charge amplifier was used to measure the output charge of the accelerometer standards. The participating laboratories reported the measurement results of the charge sensitivity and the associated uncertainties individually as they were calculated for any specified measurement condition (in particular, for a given frequency), without applying any curve fitting procedure (as frequently used to suppress deviations from a "flat" frequency response). 4. Measurement Conditions frequencies: 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 khz and 5 khz (160 Hz is the reference frequency); optionally the laboratories can measure at other frequencies (such as frequencies included in the third-octave frequency series); amplitudes: preferred value 100 m/s 2. A range of 10 m/s 2 to 200 m/s 2 were complied with; tolerated ambient and accelerometer temperature during the calibration: 23 C ± 3 C (actual values to be stated with an uncertainty of 0.5 C); relative humidity: max. 75%; mounting torque of the accelerometer: (2 ± 0.1) N m. 5. Transfer standards Two types of piezoelectric standard accelerometers were used as transfer standards: one single-ended Brüel & Kjær type 8305 WH (Accelerometer A), and one back-to-back Brüel & Kjær type 8305 (Accelerometer B). Specifications of Accelerometer A: single ended transfer standard accelerometer; Brüel & Kjær 8305 WH SN 2495771; weight: 26 grams; length: 22 mm; width over flats of hexagonal faces: 16 mm; mounting thread: 10-32 UNF-2B; electrical connector: coaxial 10-32 UNF-2A thread; accelerometer Metrologia 2011 48 Tech. Suppl. 09003 3/3

capacitance: 75 pf; sensitivity: 0.13 pc/(m/s 2 ); max. transverse sensitivity at 30 Hz: 1%. Specifications of Accelerometer B: back-to-back reference standard accelerometer; Brüel & Kjær 8305 SN 2456549 ); weight: 40 grams; length: 29 mm; width over flats of hexagonal faces: 16 mm; mounting thread: 10 32 UNF-2B; electrical connector: coaxial 10-32 UNF-2A thread; accelerometer capacitance: 75 pf; sensitivity: 0.13 pc/(m/s 2 ); max. transverse sensitivity at 30 Hz: 1%. 6. Measurement information The measurand was the magnitude of the complex charge sensitivity. Calibration of Accelerometer A by laser interferometry: the reference surface for acceleration measurement was by definition the base or mounting surface of the accelerometer. As this surface is covered during the calibration, the motion was sensed on the moving part close to the accelerometer. (Alternatively, the motion could have been sensed at the mounting surface of the accelerometer via longitudinal holes in the moving part of the vibration exciter. However, neither of the laboratories did this). Calibration of Accelerometer B by laser interferometry: the motion was sensed at the top surface without any dummy mass; no reflector (e.g. corner cube) was attached to the top surface. The charge amplifier used in the laboratories was calibrated using a standard capacitor and standard voltmeter, both traceable to national standards. The calibration of the charge amplifier was carried out shortly before the calibration, using values of the electrical quantities similar to those occurring in the accelerometer calibration. In order to suppress the effect of any non-rectilinear motion, the displacement was measured at a minimum of three different points. These points were equally spaced on the top surface of the back-to-back accelerometer or on the mounting surface of the single-ended accelerometer. The mounting surfaces of the accelerometers and the moving part of the vibration exciter were slightly lubricated before mounting. The calibration of the accelerometers was carried out in accordance with the usual procedure of the corresponding laboratory. Metrologia 2011 48 Tech. Suppl. 09003 4/4

7. Communication of the results METAS performed an initial calibration of the standards and then submitted the calibration results to the CCAUV Executive Secretary (24 October 2008) before sending the artifacts to the PTB. PTB performed the calibration of the artifacts (November 2008) and submitted its calibration report to the CCAUV Executive Secretary and to the pilot laboratory (March 2009) The calibration reports contained detailed descriptions of: the calibration equipment; the calibration methods used; the ambient conditions; the mounting technique; the calibration results including the relative expanded uncertainty; For reporting the calibration results, clause 10 of ISO 16063-11:1999 was taken into account. 8. Circulation type From METAS to PTB and back. 9. Transportation The transfer standards were transported in a closed box by an international transportation agency. 10. Results For the purpose of linking to the earlier determined CCAUV.V-K1 reference values, the participants of the EUROMET.AUV.V-K1 comparison were requested to provide measurement results for the following frequencies: 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 khz and 5 khz. The results communicated by the two participants are summarized in Table 1 and Table 2. Metrologia 2011 48 Tech. Suppl. 09003 5/5

Table 1. Summary of the results obtained while calibrating the single ended reference transducer (SN 2495771). frequency (Hz) results obtained by the METAS sensitivity (amplitude) expanded uncertainty (k=2) (%) results obtained by the PTB sensitivity (amplitude) expanded uncertainty (k=2) (%) 40 0.12906 0.4 0.12904 0.1 80 0.12906 0.4 0.12902 0.1 160 0.12904 0.4 0.12908 0.1 800 0.12912 0.4 0.12912 0.1 2000 0.12960 0.4 0.12953 0.1 5000 0.13261 0.4 0.13260 0.1 Table 2. Summary of the results obtained while calibrating the back to back reference transducer (SN 2456549). frequency (Hz) results obtained by the METAS sensitivity (amplitude) expanded uncertainty (k=2) (%) results obtained by the PTB sensitivity (amplitude) expanded uncertainty (k=2) (%) 40 0.12619 0.4 0.12610 0.1 80 0.12612 0.4 0.12609 0.1 160 0.12611 0.4 0.12610 0.1 800 0.12622 0.4 0.12621 0.1 2000 0.12660 0.4 0.12662 0.1 5000 0.12861 0.4 0.12863 0.1 The results of these measurements appear to be in excellent agreement. The information of Tables 1 and 2 is represented graphically in Figures 1 to 4. Metrologia 2011 48 Tech. Suppl. 09003 6/6

Figure 1. Sensitivity of the back-to-back reference transducer as measured by METAS in comparison with the determination by PTB - the results are virtually indistinguishable. Figure 2. Relative deviation from the mean of the sensitivity of the back-to-back reference transducer as measured by METAS in comparison with the determination by PTB - the results are in excellent agreement. Metrologia 2011 48 Tech. Suppl. 09003 7/7

Figure 3. Sensitivity of the single-ended reference transducer as measured by METAS in comparison with the determination by PTB - the results are virtually indistinguishable. Figure 4. Relative deviation of the sensitivity of the single-ended reference transducer from the mean as measured by METAS in comparison with the determination by PTB - the results are in excellent agreement. Metrologia 2011 48 Tech. Suppl. 09003 8/8

A more thorough mathematical analysis of these data allows to confirm in quantitative terms the actual consistency of the obtained results. Table 3 and Table 4 summarize the mean of the measurand at different frequencies for both transducers, the chi-squared values (CHI_sq), the deviations of the METASresults from the mean (d i ), the deviations of the PTB-results from the mean (d j ), the corresponding uncertainties and the degrees of equivalence. Table 3. Analysis of the results obtained during the calibration of the back-toback reference transducer (SN 2456549) by METAS (i) and PTB (j). frequency (Hz) weighted mean SD (mean) CHI_sq D i*10 4 U(D i)*10 4 D j*10 4 U(D j)*10 4 D ij*10 4 U(D ij)*10 4 40 0.12611 0.00006 0.11970 0.84713 4.89710-0.05287 0.30563 0.90000 5.20273 80 0.12609 0.00006 0.01331 0.28236 4.89424-0.01764 0.30574 0.30000 5.19999 160 0.12610 0.00006 0.00148 0.09412 4.89381-0.00588 0.30581 0.10000 5.19962 800 0.12621 0.00006 0.00148 0.09412 4.89808-0.00588 0.30608 0.10000 5.20416 2000 0.12662 0.00006 0.00587-0.18823 4.91276 0.01177 0.30714-0.20000 5.21990 5000 0.12863 0.00006 0.00569-0.18823 4.99076 0.01177 0.31202-0.20000 5.30277 Table 4. Analysis of the results obtained during the calibration of the singleended reference transducer (SN 2495771) by METAS (i) and PTB (j). frequency (Hz) weighted mean SD (mean) CHI_sq D i*10 4 U(D i)*10 4 D j*10 4 U(D j)*10 4 D ij*10 4 U(D ij)*10 4 40 0.12904 0.00006 0.00696 0.20891 5.00831-0.01305 0.31291 0.22196 5.32123 80 0.12902 0.00006 0.02300 0.37979 5.00836-0.02372 0.31283 0.40351 5.32118 160 0.12907 0.00006 0.01783-0.33431 5.00741 0.02091 0.31314-0.35522 5.32054 800 0.12912 0.00006 0.00001 0.00904 5.01059-0.00056 0.31316 0.00960 5.32375 2000 0.12953 0.00006 0.07045 0.66745 5.02938-0.04167 0.31399 0.70912 5.34337 5000 0.13260 0.00006 0.00117 0.08810 5.14605-0.00551 0.32158 0.09360 5.46763 An excellent agreement between the measurement results was found for the calibration applying the primary method over the requested frequency range according to CCAUV.V-K1. Metrologia 2011 48 Tech. Suppl. 09003 9/9

11. Conclusion The reported comparison demonstrates a mutual equivalence of the calibration results obtained by the participating institutes within the declared uncertainty and over the considered frequency range. It further allowed to investigate the uncertainty contributions in an enhanced frequency range, which and to improve the estimated uncertainty budget. Metrologia 2011 48 Tech. Suppl. 09003 10/10