AC-DC TCC with Built-in Tee Connector for Accurate Calibrations

Similar documents
The Effect of Changing the Applied Sequence Using the TVC on the Accuracy of the AC Signal Calibration

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

New generation of cage-type current shunts at CMI

Validation Of The New Automatic System For AC Voltage Comparisons. Umberto Pogliano, Gian Carlo Bosco and Marco Lanzillotti

Study of the Long Term Performance on the Calibration Data of the Coaxial Thermistor Mounts up to 18 GHz

NRC Publications Archive Archives des publications du CNRC

New generation of cage type current shunts developed using model analysis

An improvement for dual channel sampling wattmeter

Verifying the Wideband Input of an AC Measurement Standard

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

Ac-dc transfer standard shunts for frequencies up to 1 MHz. U. Pogliano, C.G. Bosco, M. Lanzillotti, D Serazio

Application Note. Calibration of 7000 Series Precision LCR Meters

PXIe Contents. Required Software CALIBRATION PROCEDURE

AC-DC TRANSFER DIFFERENCE STANDARDS AND CALIBRATIONS AT NRC CANADA

5790A Automated AC Measurement Standard

IMPROVEMENTS IN THE NIST CALIBRATION SERVICE FOR THERMAL TRANSFER STANDARDS

Assoc. Prof. Dr. Mohammed Helmy Abd El-Raouf

Discover. Blue Box. the. Difference. High Resistance Metrology Products Guide

Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge. Ivan Leniček 1, Roman Malarić 2, Alan Šala 3

Wideband resistive voltage divider for a standard wattmeter

THE ELECTROMETRIC AC-DC TRANSFER STANDARD AS PRIMARY STANDARD AT IEN FOR AC VOLTAGES FROM 300 V TO 1000 V

Extension and Improvement of Capacitance and Induction Measurement Capabilities at NMISA. Name: Michael Khoza Date:

Model AccuBridge 6242D Resistance Bridge

CALIBRATION PROCESS QUANTITY REDUCTION OF THE THERMAL VOLTAGE CONVERTER STANDARD USING A THREE-STAGE BUILD-UP AND BUILD-DOWN METHOD

Model 6600A Dual Source High Resistance Bridge

AC Resistance Thermometry Bridges and their Advantages By Peter Andrews

COAXIAL CURRENT SHUNTS FROM 1 ma TO 100 A

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

The Metrology Behind Wideband/RF Improvements to the Fluke Calibration 5790B AC Measurement Standard

MODEL INFORMATION. AccuBridge Model 6010D. Automated Primary Resistance/ Thermometry Bridge

Standard Resistors Precision DC Shunts Digital Thermometers Non-inductive AC Shunts Intelligent Teraohmmeter Precision Decade Standards Precision Air

Mitigation of Electromagnetic Interference Generated by GPIB Control-Network in AC-DC Transfer Measurement System

1 Ω 10 kω High Precision Resistance Setup to calibrate Multifunction Electrical instruments

Certificate of Calibration

Phase Comparison of High Current Shunts up to 100 khz

Wideband resistive voltage divider for a standard wattmeter

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V

Comparison of the Josephson Voltage Standards of the INTI and the BIPM

2-3 Calibration of Standard Voltage and Current Generator

EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

Electronics Test and Development Centre, STQC Directorate, Agriculture College Campus, Shivajinagar, Pune, Maharashtra

Performance of the Single Junction Thermal Voltage Converter at 1 MHz via Equivalent Circuit Simulation

FINAL REPORT. EUROMET project No Bilateral comparison of DC and AC voltages BEV - NCM. W. Waldmann (BEV, pilot laboratory) P.

arxiv: v1 [physics.ins-det] 4 Oct 2011

Peter Dack, Vice President Sales & Marketing Radian Research, Inc.

CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

Calsytech, # 38 North Mada Street Nandambakkam, Chennai, Tamil Nadu. Discipline Electro-Technical Calibration Issue Date

Realization and traceability of AC power standard at frequency of 50 Hz

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL

1433 SERIES. High-Accuracy Decade Resistor User and Service Manual

Application of Digital Sampling Method for Voltage Transformer Test Set Calibrations. Hüseyin Çaycı

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

American Association for Laboratory Accreditation

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

5790A/AF AC Measurement Standard

Modeling and Practical Suggestions to Improve ESD Immunity Test Repeatability

MODEL INFORMATION MODEL 6010D. Automated Primary Resistance/ Thermometry Bridge

13 th Asian Physics Olympiad India Experimental Competition Wednesday, 2 nd May 2012

Contents CALIBRATION PROCEDURE NI PXI-5422

About the power standards calibration in the National Institute of Metrology, Romania

ULTRA-PRECISE AIR RESISTANCE STANDARDS

PRECISION TEMPERATURE SYSTEMS. Compact Modular and Upgradeable Thermometry Measurement Systems

Solutions In Calibration TRANSMILLE PRODUCT GUIDE 2015/16 PROVIDING SOLUTIONS IN CALIBRATION WORLDWIDE. www. transmillecalibration.

Calibration Technique for SFP10X family of measurement ICs

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TIME AND FREQUENCY METROLOGY

Fallstricke präziser DC- Messungen

NIS Activities in the Field of Electricity and Magnetism

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM

6622A Series. Direct Current Comparator Resistance Bridges 6622A SERIES FEATURES

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

AC Voltage Standards With Quantum Traceability

50 OHM MULTIJUNCTION THERMAL CONVERTERS FOR AC VOLTAGE MEASUREMENTS UP TO 100 MHZ

NI 4070/4072 6½-Digit FlexDMM

High Voltage Component Production Testing with Two Model 2410 SourceMeter Units. Application Note Series. Introduction. Test System Configuration

SCXI HIGH-VOLTAGE ATTENUATOR TERMINAL BLOCK

Report on the Activities in Electricity and Magnetism within National Institute of Metrology (NIM), China CCEM 2017

5790B/5/AF AC Measurement Standard

Maximizing your reference multimeter, minimizing measurement uncertainties

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

Calibration of high-quality reference standards

AC/DC Resistance Standard DRR-112

NI PXI ½ Digit FlexDMM Calibration Procedure

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

SIM Regional Comparison of AC-DC VOLTAGE TRANSFER DIFFERENCE. SIM.EM-K6a, SIM.EM-K9, SIM.EM-K11 and SIM.EM-Supplementary 120 V / 53 Hz FINAL REPORT

Power Quality Measurements the Importance of Traceable Calibration

Using Reference Multimeters for Precision Measurements

CCEM Meeting, March 2017

Model 6625T Series TEMPERATURE MEASUREMENT SYSTEMS

in the genrad tradition

EE2210 Laboratory Project 1 Fall 2013 Function Generator and Oscilloscope

Physikalisch-Technische Bundesanstalt EUROMET supplementary comparison EUROMET.EM-S38 Comparison of small current sources

EMRP ENG05 Metrology for Solid State Lighting. D2.1.1 Comparison Report on Measurement of Electrical Quantities of Selected SSL Lamps

RF-POWER STANDARD FROM AC-DC THERMAL CONVERTER. L. Brunetti, L. Oberto, M. Sellone

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

riešenia na presné meranie Model 1281/1271 Selfcal Digital Multimeters

CDAX 605 High Precision Capacitance & Dissipation Factor Test Set

Test Laboratory No Accredited by CAI for Electromagnetic Compatibility, Electrical Safety and Electrical Cable Tests

Transcription:

AC-DC TCC with Built-in Tee Connector for Accurate Calibrations Rasha S. M. Ali Electrical Quantities Metrology Laboratory, National Institute of Standards (NIS), Egypt E-mail: rasha_sama79@hotmail.com Abstract Accuracy of thermal current converters (TCCs) calibration to determine their ac-dc differences is important for ac currents measurement traceability. They can be calibrated either by connecting the unknown TCC in series with the standard TCC or connecting them in parallel by using tee connector. In this paper, TCCs are established for this study. One is developed for 5 ma range, and the other for range 2.5 A. The 5 ma TCC consists of only single-junction thermal element (SJTE) with rating 5 ma, 90. The other TCC constructed from a SJTE connected in parallel with a shunt metal film resistor to withstand 2.5 A. They are calibrated against standard TCCs by connecting them in series, and in parallel. Due to some drawbacks and to overcome them, a built-in tee connector is also fabricated and introduced in this paper. These built-in tee connectors are connected to the fabricated TCCs to construct built-in TCCs. They are calibrated against the same standard TCCs. Their ac-dc differences are compared to each other. The uncertainty budget for the TCCs calibration and the ac currents measurements are estimated. The performances of the TCCs which are calibrated with the different connections are also evaluated by using them in measuring ac currents. Keywords: Ac current measurement, Ac-dc thermal converter, Thermal current converter, Ac-dc current transfer difference,. 1. Introduction Ac voltage and current are considered one of the very important quantities in the electrical field of metrology [1]. Ac current is accurately measured in terms of standard and reference dc current by means of ac-dc TCC to provide traceability to the dc standard. TCCs are mainly made of thermal converters particularly for the milliampere ranges [2]. For higher current ranges, shunt resistors are used to be connected in parallel with the thermal converters. These TCCs are calibrated by using other standard TCCs which are traceable to national metrology institute (NMI) primary standard of ac current. This calibration is done for the accurate determination of the ac-dc current transfer difference which is the metrological factor of TCCs [3]. There are two different connections for calibrating these TCCs where each NMI has its own method. One of them is made by connecting the unknown TCC and the standard TCC in series. The other is connecting them in parallel by using a tee connector. For the series connection, the two standards have the same current but there is a problem in grounding [4]. Due to using tee connector in the parallel connection, its admittance can lead to errors in the calibration [5] which increase by increasing the current range and the frequency. 1

So in this paper to study the effect of the different connections on the TCCs calibration precision and accuracy and to improve them, standard TCCs are developed at the National Institute of Standards (NIS), Egypt. These standards are automatically calibrated by the two connections against standards TCCs using a LabVIEW program. The introduced TCCs are then connected with built-in tee connectors. The construction of these TCCs is introduced at this paper. The TCCs with the built-in tee connectors are calibrated against the same standards TCCs to verify its behavior. The ranges of the fabricated TCCs are 5 ma, and 2.5 A. The calibrated TCCs are then used to measure ac currents at different frequencies. This is made to evaluate their performance and to study the effect of the different calibration systems of the TCCs set-up on the measurement accuracy of the ac currents. 2. Calibration Methods of TCCs TCCs are established at NIS for this study. One of them consists of SJTE only with rating 5 ma, 90 which used as TCC for 5 ma range. The other TCC constructed from a SJTE connected in parallel with a shunt metal film resistor with value 0.2 and power rating 3 W for 2.5 A current range. The used resistor has low inductance value and good stability. These two TCCs can be used from 50% to 120% of their current ratings. They are automatically calibrated by comparison against other standard TCCs which are traceable to PTB, Germany to obtain their ac-dc current transfer difference ( c ). The unit under test (UUT) TCC and the standard (St.) TCC are compared by connecting them in series, and in parallel by using tee connector. The 5 ma TCC is calibrated at 5 ma, and the 2.5 A TCC is calibrated at 2 A. Figure 1 shows the system set-up for the two connections of the unknown and standard TCCs which are done for calibration. The unknown TCCs are calibrated at frequencies 55 Hz, 300 Hz, 1 khz, and 10 khz. The set-up of the calibration systems consists mainly from a calibrator and two digital multimeters (DMM) to measure the output emfs from the two TCCs. Shielded cables are used for the output emf of the TCCs and connected to the guards of the DMMs to insure the same potential for all the used instruments [4]. (a) Series connection (b) Parallel connection Figure 1 Set-up of calibration systems for series and parallel connections The ac-dc difference is obtained from [6]. The results; c and Type A at the different ranges and frequencies by the two connections are compared to each other. Table 1 2

shows the obtained results for the 5 ma, and 2.5 A TCCs by the series and parallel connections. Table 1 c and Type A for the fabricated TCCs by the two connections TCC Connection Obtained Results, Frequency (Hz) Range Method µa/a 55 300 1 k 10 k 5 ma Series Parallel Ac-dc difference, c 16.2 8.7 1.9 9.8 Type A 2.6 3.2 4.1 3.2 Ac-dc difference, c 10.4 5.3 3.9 11.6 Type A 7.0 4.3 5.4 4.1 2.5 A Series Parallel Ac-dc difference, c -108-190 152 91.4 Type A 5.0 3.7 4.5 5.3 Ac-dc difference, c -149-247 165 247 Type A 8.0 7.4 15 12 It is found that the parallel connection appears higher ac-dc differences and Type A than the series connection at the various frequencies especially at the high current. That is due to the small impedance of the TCC with respect to the tee connector impedance. So, built-in tee connector is developed and studied at this paper. 3. TCC with Built-in Tee Connector The fabricated TCCs are connected with built-in tee connector to overcome the drawbacks of the tee connector. The introduced connector is mounted on a printed board as shown in Figure 2. One of the input and output currents terminals are connected to N-type male connectors. Figure 2 Built-in tee connector The other output terminal of the introduced tee connector is connected directly to a lead of the fabricated TCCs to be used as standards for the ac current. The other lead is connected to the low terminal of the input connector to have the same potential. One of the developed built-in tee connector is for the 5 ma, and the other for 2.5 A. The 2.5 A built-in tee connector is constructed to withstand this applied current. They have very small internal impedance with respect to the internal impedance of the commercial tee connector. These built-in TCCs are calibrated against other standard TCCs as shown in Figure 3. 3

Figure 3 Calibration system by using the built-in TCCs Figure 4 shows the ac-dc differences of the 5 ma, and 2.5 A TCCs by the different connections; series, parallel, and built-in tee at frequencies 55 Hz, 300 Hz, 1 khz, and 10 khz. Figure 4 Ac-dc differences of the TCCs obtained by the different connections The figures illustrate that the built-in TCCs introduces lower ac-dc differences at the different ranges and frequencies particularly in higher current ranges. The 2.5 A builtin TCC shows better frequency dependence than the TCCs which are calibrated by other connections. 4. Behavior Evaluation of the Introduced TCCs Calibration Methods The effect of the different calibration connections on the measurement accuracy and the uncertainty of the ac current has been evaluated. Ac currents with values 5 ma and 1.9 A have been measured by TCCs that are calibrated by the three different connections. The actual values of the ac currents are obtained from calibrating them at CMI, Czech. The value of the ac current by using the TCCs is calculated by [2,7]: 4

I ac = I dc (1 + ) (1) Where, I dc is the traceable dc current. = m + c c is the ac dc current transfer difference of the used TCC. Eac Edc m is equal to m = ne dc (2) The expanded uncertainty is estimated for the TCCs calibration by the different connections and also for the ac currents measurement according to the GUM; Guide to the Expression of in Measurement [8]. Table 2 and Table 3 demonstrate the uncertainty budget for the calibration of the built-in tee connector TCC 5 ma, 300 Hz and the measurement of ac current at 1.9 A, 300 Hz which obtained by using the built-in tee connector TCC respectively as an example. Sources Repeatability (n = 12) Calibration of Standard TCC Calibration of DC Current Tee Connector Effect Table 2 budget for the 5 ma TCC at frequency 300 Hz Standard Degree Probability Divider of distribution C ( A/A) i Freedom contribution ( A/A) 0.8 Normal 1 1 11 0.8 3 Normal 2 1 1.5 10 Normal 2 1 5 0.87 Rectangular 3 1 0.5 Combined standard uncertainty: Effective degrees of freedom: Expanded at confidence level 95%, (k = 2): ±5.6 A/A ±11.2 A/A Table 3 budget for the 1.9 A ac current measurement at frequency 300 Hz Probability Degree of Divisor contribution Sources distribution Freedom ( A/A) Repeatability (n = 12) Normal 1 11 1.6 Standard TCC Calibration Normal 2 11 DC Current Calibration Normal 2 8.7 Calibration of DMM Normal 2 1.7 Resolution Rectangular 3 1.7 10-8 Thermal emf Rectangular 3 6 Combined standard uncertainty: Effective degrees of freedom: Expanded at confidence level 95%, (k = 2): ±15.4 A/A ±30.8 A/A 5

Figure 5 shows the different results of the 5 ma and 1.9 A ac current at frequency 300 Hz accompanied with their expanded uncertainties. The results are represented with four error bars. Figure 5 Comparison of ac currents measurements For the 5 ma TCC, the error bars illustrate that the measured values using the three connections are acceptable and near to the 5 ma actual value especially the built-in TCC and also it has a smaller expanded uncertainty. However, the value of the ac current obtained by using the 2.5 A built-in TCC is more accurate than other connections. The obtained value using the series connection is slightly near to the actual value so, it can be accepted. While the value obtained by the parallel connection is so far consequently it is not accepted. It illustrates that at high currents, the built-in TCC is more accurate ac-dc current transfer standard and precise than other TCCs which are calibrated by the other connections. 5. Conclusion To study the TCCs calibration connections effect on the accuracy, TCCs are fabricated and introduced in this paper. Their ranges are 5 ma and 2.5 A. They are calibrated against other standard TCCs by two different connections; series and parallel using tee connector. The ac-dc differences of the TCCs which are calibrated by using the parallel connection introduce higher values. So, a built-in tee connecter TCC is developed and studied in this paper. The introduced tee connector is connected directly to the fabricated TCCs to form built-in TCCs. They are used as standards for the 5 ma and 2.5 A ac currents. The built-in TCCs are also calibrated against the same standard TCCs. They demonstrate lower ac-dc transfer differences, and the 2.5 A built-in TCC shows good frequency dependence. The performance of the built-in TCCs are verified by using them in measuring 5 ma, and 1.9 A at 300 Hz. 6

The obtained values are compared with their actual values and also with the results obtained by using the same TCCs which are calibrated by the series and parallel connections. The error bars for this comparison illustrates that the 5 ma TCC which has been calibrated by the different connections has acceptable values specifically the built-in TCC which is near to the 5 ma actual value. While the 2.5 A built-in TCC is more accurate and accepted than others. It verifies that the built-in TCC improves the ac current measurement accuracy with reduced expanded uncertainty. Furthermore, the calibration method by the parallel connection is not recommended for the high currents TCCs due to having high impedance compared to the TCCs impedances. References [1] A. K Govil, Saood Ahmad, Bijendra Pal, and P C Kothari, "Development of an automated AC-DC Transfer Measurement System for Voltage and Current al Low Frequencies", Conference on Precision Electromagnetic Measurements (CPEM), June 2008. [2] Kåre Lind, Tore Sørsdal, and Harald Slinde, "Design, Modeling, and Verification of High- Performance AC DC Current Shunts From Inexpensive Components", IEEE Transaction on instrumentation and measurement, Vol. 57, No. 1, pp. 176-181, January 2008. [3] Bruno Trinchera, Luca Roncaglione Tet, and Marco Lanzillotti, "AC-DC Current Transfer Difference Estimation of Thin-Film Multijunction TCs up to 1 MHz", Conference on Precision Electromagnetic Measurements (CPEM), 2016. [4] Piotr S. Filipski and Michael Boecker, AC-DC Current Transfer Standards and Calibration at NRC, Simposio de Metrologia, 25-27 Oct. 2006. [5] Karl-Erik Rydler, "High Precision Automated Measuring System for ac-dc Current Transfer Standards", IEEE Transaction on instrumentation and measurement, Vol. 42, No. 2, pp. 608-611, April 1993. [6] B. Pal, S. Ahmad, and A. K. Govil, "Automation and Evaluation of Two Different Techniques to Calibrate Precision Calibrators for Low Frequency Voltage using Thermal Devices", MAPAN-Journal of Metrology Society of India, DOI 10.1007/s12647-012-0038-5, Published online 11 January 2013. [7] Rasha S.M. Ali, AC Current automatic calibration using two different TCC designs, Journal of Measurement Science and Instrumentation, Vol. 4, No. 3, PP. 205-209, September 2013. [8] Guide to the expression of uncertainty in measurement, JCGM 100:Sept.2008. http://www.bipm.org/utils/common/documents/jcgm/jcgm_100_2008_e.pdf 7