Contents. 2 qutag Manual

Similar documents
PULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL

CIRCUIT-TEST ELECTRONICS

Model 9302 Amplifier-Discriminator Operating and Service Manual

ADQ108. Datasheet. Features. Introduction. Applications. Software support. ADQ Development Kit. Ordering information

Power Meter. Measurement Guide. for Anritsu RF and Microwave Handheld Instruments BTS Master Site Master Spectrum Master Cell Master

User s Manual Current Probe. IM E 2nd Edition IM E

Random & Sinusoidal Jitter Injector. Main Unit Operation Manual

Model 533 Dual Sum and Invert Amplifier Operating and Service Manual

PA8HF power amplifier Operating guide

COM-POWER OPERATION MANUAL ACS W

Model 863 Quad Timing Filter Amplifier Operating and Service Manual

InGaAs SPAD BIOMEDICAL APPLICATION INDUSTRIAL APPLICATION ASTRONOMY APPLICATION QUANTUM APPLICATION

UR200SI / UR200WE ENGLISH

SPM-50 RF Spectrum Power Meter PC Software User Manual

Documentation. PerfectPulse Fast Edge Signal Generator

RIGOL. Quick Guide. DG2000 Series Function/Arbitrary Waveform Generator. Sept RIGOL Technologies, Inc.

HZJF-9007 Fully Functional Partial Discharge Inspector USER MANUAL. Huazheng Electric Manufacturing(Baoding) Co.,Ltd

Pulse Amplifier DIV 20 Instruction Manual 03/2016

User s Manual Current Probe. IM E 2nd Edition. Yokogawa Electric Corporation

Model 9305 Fast Preamplifier Operating and Service Manual

Model 416A Gate and Delay Generator Operating and Service Manual

EPA104/254. User Manual.

TA MHz ±700 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

312, 316, 318. Clamp Meter. Users Manual

TA MHz ±30 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT I, Pollution Degree 2.

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

TT-SI 9001 / TT-SI MHz Active Differential Probes

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

TETRIS 1000 High Impedance Active Probe. Instruction Manual

PI-150 Broadband Power Indicator

SUBNANOSECOND PULSE GENERATOR PPG-2/1000 USER MANUAL

3B SCIENTIFIC PHYSICS

DM-45 Digital Multimeter

plifier D-501 otion Am Tactile M

EPA152/252/502. User Manual.

Instruction Manual ICM 3091N Digital AC Clampmeter EN FR IT DE ES

DEUTSCH ITALANIO ENGLISH NORSK FRANÇAIS DANSK NEDERLANDS Hersteller UR8200

SDI SPECTRADYNAMICS, INC. HIGH PERFORMANCE DISTRIBUTION AMPLIFIER OPERATING MANUAL

Dual 500ns ADC User Manual

PI-10 Broadband Power Indicator

Visit us at

REVAMP4120T Instruction manual

Wilcoxon Research PA8HF power amplifier Operating guide

TT-SI MHz Active Differential Probe

9200 Series, 300 MHz Programmable Pulse Generator

TT-SI MHz DIFFERENTIAL PROBE

InGaAs SPAD freerunning

DS1000E, DS1000D Series Digital Oscilloscope

STEREO POWER AMPLIFIER OWNER MANUAL PR-150 DESIGNED IN U.K. PDF created with FinePrint pdffactory trial version

Opus 21 s80 Integrated Amplifier Owner's Manual

2015 RIGOL TECHNOLOGIES, INC.

TT-SI MHz Active Differential Probe

Model and

Radio Control Installation and Operating Instructions System 4

Model # PCYFC-10kW-250A PCYFC-20kW-250A PowerCycle Battery Conditioner Operation Manual

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

M302RM OPERATING MANUAL

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER

HTA125A/250A. Power Amplifiers. Installation & Use Manual

NANOSECOND PULSE GENERATOR NPG-15/2000(N) USER MANUAL

TA MHz x10/x100 Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

200Amp AC Clamp Meter + NCV Model MA250

GFT bit High Speed Digitizer

DAA AES/EBU Digital Audio Distribution Amplifier. User Manual. I.R.T. Communications Pty Ltd

Antenna Splitter ASA 1. Instruction manual

TMP40. User Manual.

Analog-to-Digital-Converter User Manual

DS1000B Series Digital Oscilloscope

VFSC9 ELECTRONIC SPEED CONTROLLER. Mounting and operating instructions

400Amp True RMS AC Clamp Meter + NCV

ExpoM - ELF User Manual

Mixer Amplifier series

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Model 5100F. Advanced Test Equipment Rentals ATEC (2832) OWNER S MANUAL RF POWER AMPLIFIER

DPA-1200 ORDERCODE D4180 DPA-2400 ORDERCODE D4181 DPA-3400 ORDERCODE D4182

SAGA PRO SERIES STEREO POWER AMPLIFIER OPERATION MANUAL

USER MANUAL V Valencia, Spain -

R-Series R235LS 2-Channel Power Amplifier with Local Source Switching

USB Multifunction Arbitrary Waveform Generator AWG2300. User Guide

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

LED PowerBlinder 4 ORDERCODE 41320

Part No. Z , IB Operation Manual. Pseudo Voice Generator KSG3600

Model Hz to 10MHz Precision Phasemeter. Operating Manual

INSTRUCTION MANUAL ENGLISH

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

CONTENTS. SAFETY PRECAUTIONS: Before use, read the following safety precautions

Series 70 Servo NXT - Modulating Controller Installation, Operation and Maintenance Manual

Dual Channel Function/Arbitrary Waveform Generators 4050B Series

MD 200A High voltage differential probe

PROFESSIONAL SMPS STEREO AMPLIFIERS USER GUIDE

DPA74/154 AUDAC PROFESSIONAL AUDIO EQUIPMENT. DPA74/154 Quad Channel Class-D Amplifier. User Manual & Installation Guide

INSTRUCTION MANUAL FUNCTION GENERATOR FG-274 B

Rubix RGB 3D Panel. User Manual. Order code: LEDJ476

User s Guide. Digital AC/DC Clamp Meter Model 38394

Model MHz Arbitrary Waveform / Function Generator

N2792A and N2793A Differential Probes User s Guide

AWG801 8 GSPS 11-bit Arbitrary Waveform Generator

REVAMP4100 Instruction manual

USB Line Camera 8M. Coptonix GmbH

INFRARED WIRELESS DISTRIBUTOR

Transcription:

qutag Manual V1.0.0

Contents 1. Introduction... 3 2. Safety and Maintenance... 3 2.1. Legend... 3 2.2. General Instructions... 3 2.3. Environmental Conditions... 4 2.4. Electrical Installation... 4 2.5. Preventive Maintenance... 4 2.5.1. Safety Testing... 5 2.5.2. Cleaning... 5 3. Technical Information... 6 3.1. Technical Specifications... 7 3.2. Additional Features... 8 4. First Time Setup... 10 5. Firmware Update... 12 6. Using the qutag: Displays on front panel... 11 7. Daisy... 13 7.1. Input Parameter Settings... 14 7.2. (Coincidence) Counting... 16 7.3. Histograms and Timestamping... 17 7.4. Correlation Measurement: HBT... 19 8. Synchronization... 21 9. Command Line Interface... 22 10. Using the qutag with custom software (DLL)... 22 11. Revision History... 23 2 qutag Manual www.qutools.com

1. Introduction The qutag is a time-to-digital converter that measures electric signals and mark them with a time tag. This stream of time tags can be used in various and versatile applications from measurements in the picosecond range up to days. The universal time tagging method allows correlation measurements (cross-correlation, auto-correlation), lifetime measurements (start-(multi)stop) and many more possibilities within one measurement. The saved stream of time tags contains all information needed to reconstruct every measurement and analysis performed. 2. Safety and Maintenance 2.1. Legend Caution General risk of danger. Warning An instruction which draws attention to the risk of injury or death. Warning Risk of electric shock. High voltages present. 2.2. General Instructions The equipment, as described herein, is designed for use by personnel properly trained in the use and handling of mains powered electrical equipment. Only personnel trained in the servicing and maintenance of this equipment should remove its covers or attempt any repairs or adjustments. If malfunction is suspected, immediately return the part to the vendor for repair or replacement. There are no user-serviceable parts inside the electronics. Modified or opened electronics cannot be covered by the warranty anymore. Take special care if connecting products from other manufacturers. Follow the General Accident Prevention Rules. If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Do not operate the instrument outside its rated supply voltages or environmental range. In particular, excessive moisture may impair safety. www.qutools.com qutag Manual 3

For laboratory use only. This unit is intended for operation from a normal, single phase supply, in the temperature range 5 to 40 C, 20% to 80% RH. 2.3. Environmental Conditions Warning. Operation outside the following environmental limits may adversely affect operator safety: Indoor use only Maximum altitude 2000 m Temperature range 5 C to 40 C Maximum humidity less than 80% RH (non-condensing) at 31 C To ensure reliable operation the unit should not be exposed to corrosive agents or excessive moisture, heat or dust. If the unit has been stored at a low temperature or in an environment of high humidity, it must be allowed to reach ambient conditions before being powered up. 2.4. Electrical Installation The unit must be connected only to an earthed fused supply of 100 to 240V (Japan, USA and Europe). Use only power supply cables supplied by the manufacturer, other cables may not be rated to the same current. The unit is shipped with appropriate power cables for use in Europe. When shipped to other territories the appropriate power plug must be fitted by the user. Never connect any cabling to the electronics when contacts are exposed! Avoid short-cuts. Prevent electrical shock from electronic. To prevent electrical shock do not remove the cover of the control unit. Unplug power cord and all other electrical connections and consult qualified service personnel when servicing or cleaning. Operate only under dry conditions and at room temperature conditions. 2.5. Preventive Maintenance The equipment contains no user serviceable parts. There is a risk of severe electrical shock if the equipment is operated with the covers removed. Only personnel authorized by the vendor and trained in the maintenance of this equipment should remove its covers or attempt any repairs or adjustments. Maintenance is limited to safety testing and cleaning as described in the following sections. 4 qutag Manual www.qutools.com

2.5.1. Safety Testing Safety testing in accordance with local regulations, should be performed on a regular basis, (typically once per year for an instrument in daily use). Caution. The instrument contains a power supply filter. Insulation testing of the power supply connector should be performed using a DC voltage. 2.5.2. Cleaning Disconnect the power supply before cleaning the unit. Never attempt to clean the qutag by immersion into any liquid. Never allow water to get inside the case. The qutag parts are sensitive to any kind of liquid. Do not use any type of abrasive pad, scouring powder, or solvent, e.g. alcohol or benzene. Please note that all parts of qutag are cleaned in our production facility. www.qutools.com qutag Manual 5

3. Technical Information The key features of the qutag device are: High timing resolution (bin size 1 ps) High event rates USB3.0 interface Easy-to-use The device can be used for a variety of Applications: Time correlated single photon counting (TCSPC) Fluorescence lifetime imaging Quantum information experiments LIDAR High energy/accelerator physics High precision time measurements Channels 4 Bin width (ps) 1 Input pulse high level (V) Min -3.3 Max 3.3 Input pulse width (ps) Min 300 Input pulse separation (ns) Min 40 Input Impedance (Ohm) 50 Table 1: Input Specifications 6 qutag Manual www.qutools.com

3.1. Technical Specifications Time to digital converters Digital resolution Timing jitter, RMS Timing jitter, FWHM Max. event rate per channel Sustained throughput rate Delay range Delay resolution Min. pulse to pulse separation Differential non-linearity <1% Max. acquisition time without overflow 13 days 1 ps < 10 ps < 25 ps 25 Mcps 200 MHz periodic (with divider) 100 Mcps (USB3.0) -100 +100 ns 1 ps 40 ns Input Channels Number of channels 1 start, 4 stop (model dependent) Input connector SMA Signal levels (threshold comparator) -3 +3V e.g. LVTTL, NIM Threshold level resolution 1.46 mv Max. input level ± 3.3V Edge Rising, falling Min. input pulse width 300 ps Termination 50 Ohms Divider On start input (model dependent) Clock Input (model dependent) Frequency 10 MHz (model dependent) Signal levels (threshold) -5 +5V Signal form Sinusoidal, square wave Termination 50 Ohms Input connector SMA Synchronization (model dependent) Number of synchronizable qutags 4 Number of synchronized stop channels 16 Operation Interface Operating system Power consumption USB3.0 Windows, Linux < 50W at 100 to 230 VAC, 50-60 Hz www.qutools.com qutag Manual 7

3.2. Additional Features Note: The additional features are model-dependent. Contact us for details. Marker Inputs In addition to the 4+1 channel input, the device features marker inputs that insert marker timestamps in your timeline. You can connect these inputs e.g. to your pixel clock or line clock. This helps you to sort the timestamps to the correct pixel in your FLIM setup. Markers 4 Input pulse width (ns) Min 10 Input pulse level (V) Low -0.5 0.8 High 2.0 5.5 Input Impedance (Ohm) 470 Frequency (MHz) Max 10 Table 1: Input Specifications Figure 1: D-Sub connector at the qutag back panel with marker inputs 1-4 at pin 5-8 and ground at pin 17-26 Markers are saved within the data stream of Daisy and Tarek as timestamps with channel numbers 100-103. A marker with channel number 104 is generated every ms. Within synchroniced devices Channel numbers 105-109 correspond in the same way to the second device and so on. Start-Channel as Input The Start channel can be converted to another stop channel, allowing the device to have 5 completely equal input channels. 8 qutag Manual www.qutools.com

User-defined Clock frequency Allow to use any frequency between 1 100 MHz as a clock input for long-term accuracy. Divider for stop channels This option allows you to enable the divider on all stop channels. This allows higher frequency periodic signals to be recorded. www.qutools.com qutag Manual 9

4. First Time Setup Figure 2: Rear side of the qutag. 1: USB3.0 connection, 2: GPIO connection, 3: synchronization input, 4: synchronization output, 5: AC power input and power switch. Make sure the cooling fans are not obstructed. Otherwise the qutag may be damaged due to overheating. Supply power to the qutag via the AC power input and turn it on using the power switch. Connect the qutag to a PC using USB 3.0 connector. An USB 2.0 cable can be used, but limits the transfer rate. Note: The Ser, Ethernet and USB 2.0 connectors are reserved for future use and will not work. Software and Drivers (Installer or zip archive) o Windows: Start the Daisy qutag Installer and follow the instructions, or extract the zip archive to a directory of your choice and install the device driver in the usbdriver directory. o Linux: Start daisy.exe. Extract the tgz archive Follow the instructions in install/readme.linux.txt. 10 qutag Manual www.qutools.com

5. Displays on front panel Single tap on the left display to toggle on and off the illumination of all displays within one device. Channel mode Start / Stop and channel Number. Displaying the counts per second. Shows coincidences with other channels within the device. Threshold level of channel input with signal type or threshold voltage. Auto calibrate the threshold by long pressing the display of the desired channel. www.qutools.com qutag Manual 11

6. Firmware Update Attention: Naturally, firmware updates are a delicate process and can potentially harm the device if done incorrectly. Please read and follow the instructions carefully. Figure 3: The firmware update tool nhflash. Start nhflash.exe. If the tool displays Firmware: outdated as in Figure 2, please stick closely to the following steps, when an update is required: 1. If the device is running, switch it off. 2. If any qutag software is still running, close it. 3. Turn on the device. 4. Start nhflash.exe. After some time the LED turns green. Proceed only if the tools displays Firmware: outdated. 5. Press Flash. Make absolutely sure the device is not shut down or disconnected during this step! The flash process takes several minutes and may seem to hang without progress for up to 3 minutes. After the update is completed successfully, the LED stops flashing and the text next to it shows Update complete. 6. Close the firmware update utility. 7. Turn the qutag device off and on again. 12 qutag Manual www.qutools.com

7. Daisy The Daisy GUI Data Analysis and Imaging System - is started by daisy.exe. It provides several functions which will be explained in the following chapters. The opening screen is shown in Figure 4, the different features are accessed by clicking on the different tabs. Double-clicking a tab decouples it from the main window, making it possible to view multiple tabs at once. Figure 4: The opening screen of the Daisy GUI. Different tasks can be accessed by the different tabs at the top. Please note that this window will display additional options for upgraded qutag devices, so the appearance varies depending on the used model. www.qutools.com qutag Manual 13

7.1. Input Parameter Settings Input parameter settings like the trigger level and rising/falling edge detection can be configured in the Detector Parameters" tab shown in Figure 5. After a restart of the device all channels are by default configured to detect the rising edge of input signals at a trigger level of 1 V, which is suitable for LVTTL signals. Please note that these settings are saved in the device and therefore persistent between software restart. Only a restart of the device will reset all settings to their default values. Figure 5: Input parameter settings in Daisy. 14 qutag Manual www.qutools.com

The checkboxes next to Enable Channels can disable individual inputs. The setting in Dead Time enforces dead time after each generated timestamp. This can be used to enforce the specified dead time of 40 ns. After an event has been detected on a specific channel, subsequent events on the same channel are ignored during the Dead Time. The setting can be used to replace the natural dead time of the detectors of about 40ns by an exactly defined time span. Values below the natural dead time have no effect. The trigger level and rising/falling edge detection of each input can be configured with the inputs under Signal Conditioning. Presets for the common LVTTL and NIM signal standards can be chosen under Signal Type. The Signal Delay setting can be used to compensate runtime differences between -100 ns and 100 ns for each input. The Divider setting configures the input divider on the start input. Setting of n > 1 cause only every n-th input pulse to be measured, allowing event rates above 25 MHz. The Reconstruct setting selects whether events lost to the divider should be reconstructed assuming a periodic input signal. www.qutools.com qutag Manual 15

7.2. (Coincidence) Counting All single (each channel individually) and coincidence rates (channel 1 to 5) are displayed in the Counters Tab shown in Figure 6. Figure 6: Coincidence Counters Tab of the Daisy. The integration (or exposure ) time can be set to any ms value between 1 ms and 2 16-1 ms 65.5 s. The coincidence time window can be set to any value between 0 ps and 2 32-1 ps 4.3 ms. Two (or more) detection events will be counted as a coincident event if the difference of their time stamps is less or equal than the specified time window. The trend of a selected count rates can be displayed in the line graphs. 16 qutag Manual www.qutools.com

7.3. Histograms and Timestamping Timestamps histograms and the raw timestamps are accessible in the Time diffs tab shown in Figure 7. Figure 7: Histogram and timestamping tab in Daisy. Raw timestamps can be saved to a file using the buttons next to Save Time Tags. To configure the histogram measurement, the bin width, bin count and the value (offset) of the first bin can be configured under Histogram Window. Which channels contribute to the measurement can be chosen in the Histogram drop-down menu. If Global is selected all Timestamps from all channels used are used as start and stop events. www.qutools.com qutag Manual 17

The integration time of each measurement is set at Exp. Time. If Sync is selected the finished histogram is displayed after every measurement, otherwise a live view of the running measurement is shown. The Refresh button can be used to display the running measurement is Sync is selected. If Diffs is selected, the differences of consecutive Timestamps are added to the histogram. Use this setting if the time differences between events on different stop inputs are of interest. If Diffs is not selected the absolute value of individual timestamps are added to the histogram. Use this setting if the time between events on the start input and a stop input is of interest. The data of the currently displayed histogram or of every finished measurement can be saved using the buttons next to Save histogram. 18 qutag Manual www.qutools.com

7.4. Correlation Measurement: HBT The HBT measurement and fit feature is accessible in the HBT tab shown in Figure 8. Figure 8: HBT measurement and fit tab in Daisy. The measurement is configured in the Histogram field. Here the histogram width, a bin count as well as the two input channels are selected. www.qutools.com qutag Manual 19

The Results shows several parameters of the current measurement including the event rates, measurement time and the g (2) (0) value. In the Fitting field a model function for the fit can be chosen. The result of the fit is displayed in the bottom field. The fit is only valid, if the number next to Fit Success is positive. If no fit can be found, please adjust the starting values using the Edit Start Values button. The available models for the correlation function are shown below. With these, thermal photons, antibunching and three-level-systems can be described. Thermal g (2) th (τ) = A e (τ τ 0 )2 2c 2 + B Antibunching g (2) ab (τ) = 1 e τ τ 0 τa Three-Level-System g (2) tl (τ) = 1 + p 2 f (c e τ τ 0 τ b (1 + c) e τ τ 0 τa ) Additionally models which take detector offsets and jitter into account are available. If they are used, the appropriate values for your detectors have to be entered in the Detector Parameters" field. 20 qutag Manual www.qutools.com

8. Synchronization If more than four inputs channels are required for a measurement, up to four qutags can be synchronized. To synchronize devices, turn off their power and connect them using the provided cables as shown in Figure 9. Figure 9: Diagram showing cable configuration for synchronizing qutags with bottom device as master. We recommend putting the devices on top of each other. For each pair of neighboring devices the qutag OUT connector of the bottom device is connected to the IN connector of the top device using the provided cables. After the devices are turned on, you can obverse the changed channel numbers on the front displays. For alignment and initial measurements with synchronized devices Daisy can be used. After startup a dialog will appear which allows you to choose which device will be connected. Multiple instances of Daisy can be run to work with different synchronized devices. For measurements using Timestamps from all synchronized devices the Tarec tool must be used to record data from all devices and merge and sort them into a single file. The tool is started with tarec.exe. Please close all instances of Daisy before starting Tarec. The Tarec GUI is shown in Figure 10. www.qutools.com qutag Manual 21

Figure 10: tarec tool recording data from two synchronized qutags. The serial number and an approximate number of recorded events are shown on the bottom for each connected device. The Config buttons on each line can be used to configure the inputs of the devices similar to the Detector Parameters tab in daisy introduced in chapter 7.1. Whether a text or a binary output file is created and their filenames can be chosen next to Text Output and Binary Output. Recording data is controlled using the Start and Stop buttons. After the recording is stopped a progress bar is displayed while the data is sorted and saved in the output file. 9. Command Line Interface The command line interface can be used for some simple tasks as a one-time readout of the count rates, writing time stamps to a file or saving histogram data. To use it, open a Windows command prompt (Press Start, type cmd and press Enter), go to the software directory and call tdccli.exe. Use the parameter -h for additional information. The source code is available in \userlib\cli. 10. Using the qutag with custom software (DLL) The qutag can be accessed from custom software using the DLL provided in \userlib. Examples in C are provided in \userlib\src. The complete documentation of the API can be found in \userlib\doc\index.html. 22 qutag Manual www.qutools.com

11. Revision History V1.0.0 (2018-02-09): initial release qutools GmbH : Kistlerhofstraße 70 (Geb. 88) : 81379 München : Germany e-mail: info@qutools.com Phone: +49 89 321 649 59-0 : Fax: +49 89 321 649 59-9 Web: http://www.qutools.com Copyright 2018 by qutools GmbH. The information, specifications and technical data contained in this manual are subject to change at any time by qutools GmbH without prior notice. qutools GmbH assumes no responsibility or liability for any errors or inaccuracies that may appear in this document and any consequences thereof. www.qutools.com qutag Manual 23