Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Type B

Similar documents
Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I

ROTANODE TM E7252X E7252FX E7252GX Rotating Anode X-Ray Tube Housing Assembly. General Data. IEC Classification... Class I Electrical :

Rotating Anode X-Ray Tube Housing Assembly. General Data

Product Information PI

Product Information PI

Rotating Anode X-Ray Tube Housing Assembly. General Data

Product Information PI

Stationary Anode X-Ray Tube. General Data

Product Information PI

Product Information PI

Product Information PI

Rotating Anode X-ray Tube Assembly

Rotating Anode X-ray Tube

CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

Industrial X-ray Tube KL

MG MW S-Band Magnetron

Atomscope PORTABLE VETERINARY X-RAY EQUIPMENT INSTRUCTION MANUAL. Version 1.0

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

PET1606J2F. Pilani Electron Tubes & Devices Pvt. Ltd. Water Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1606J2F

Abridged Data. General Data. MG7095 Tunable S-Band Magnetron for Switched Energy Applications. Cooling. Electrical. Accessories.

M5028 Precision Tuned Magnetron

MG7095 Tunable S-Band Magnetron

X-band Magnetron. Cooling (note 5) Water Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

ARU-01CF Portable X-Ray NSN:

MG5223F S-Band Magnetron

X band Magnetron. Water: Anode cavity Forced-air: Input ceramics and terminals Output coupling (note 6) UG51/U Magnet (note 7) Integral, Permanent

PET1610F. Pilani Electron Tubes & Devices Pvt. Ltd. Forced-Air Cooled Triode. For Industrial RF Heating. Drop in equivalent of BW1610F

MG6090 Tunable S-Band Magnetron

Phase-sequence Phase-loss Relay

Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL

RadView Manual Collimator

MG5193 Tunable S-Band Magnetron

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

MEDIUM-MU AIR-COOLED POWER TRIODE 3CX15,000H3

MILADY. Product Data. Page 1 of 8

Corsix R 32 Accu PRODUCT DATA

Technical Data TD. X-Ray FLAT PANEL IMAGER FDX3543RP FEATURING: Active Area: 35(H) 43(V)cm (14 17 ) Excellent Sensitivity & Resolution

Digital Panoramic X- Ray unit. Product Data

CX1140 Hydrogen Thyratron

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

TLP3543 TLP Applications. 2. General. 3. Features. 4. Packaging and Pin Assignment Rev.3.0. Start of commercial production

Recommended Land Pattern: [mm]

Type 297, High-Voltage Mica Capacitors Corona-free Mica Coupling Capacitors for Medium-Voltage PDA s

BCA 9SK. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

E2V Technologies MG5223F S-Band Magnetron

TLP3558A,TLP3558AF TLP3558A,TLP3558AF. 1. Applications. 2. General. 3. Features Rev.1.0. Start of commercial production

KP 376 KP/MMKP 376. AC and pulse metallized polypropylene film capacitors. Fig.1 Simplified outlines.

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

E2V Technologies MG5222 X-Band Magnetron

USER MANUAL 600A AC Clamp Meter + NCV Model MA610

TD62786AP,TD62786AF,TD62787AP,TD62787AF

SPECIFICATION, CONTROLS AND ACCESSORIES

TD62308AP,TD62308AF TD62308AP/AF. 4ch Low Input Active High-Current Darlington Sink Driver. Features. Pin Assignment (top view)

VGM Series. 100 & 60 W, Efficient, CV Class 2 LED Drivers for Signage Applications ORDERING INFORMATION

ASMB-KTF0-0A306-DS100

TOSHIBA BiCD Digital Integrated Circuit Silicon Monolithic TB62756FUG

E2V Technologies MG6028 Fast Tuned Magnetron

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS

400Amp True RMS AC Clamp Meter + NCV

ACS600, ACS800. Installation Guide

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

32mm Glass Transponder. Read Only, Read/Write. Reference Guide

Wireless Room Thermostat Installation and Operation Y6630D1007

MS2109A AC/DC Clamp Meter. User Manual. Contents

X-ray Tube and Generator Basic principles and construction

MODEL 3810/2 Line Impedance Stabilization Network

ASMB-TTF0-0A20B-DS101

SIG Series Spark Gaps

TLP3924 TELECOMMUNICATION PROGRAMMABLE CONTROLLERS MOSFET GATE DRIVER. Features. Pin Configuration (top view)

Ultra low quiescent current, Fast Load Transient 300 ma CMOS Low Drop-Out Regulator in ultra small package

Computer-14d - xx - 144a

AMM-1022 Digital Multimeter USER`S MANUAL

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

SSM3J356R SSM3J356R. 1. Applications. 2. Features. 3. Packaging and Pin Assignment Rev.3.0. Silicon P-Channel MOS (U-MOS )

TOSHIBA Fast Recovery Diode Silicon Diffused Type CMF (50 Hz) (Note 2)

Digital Clamp Meter. User Manual

Operation Manual. SWF Series Switch Mode Power Supply. SANKEN ELECTRIC CO., LTD No.PAN E-01.

TOSHIBA Photocoupler GaAs IRED & Photo-Triac TLP161J. Marking of Classification

EMC Feedthrough Capacitors

TD62318APG,TD62318AFG

LED Driver Board 150PW02F

High switching capacity 1a/1c 30A power relays

SSM3K341R SSM3K341R. 1. Applications. 2. Features. 3. Packaging and Pin Assignment Rev.5.0. Silicon N-channel MOS (U-MOS -H)

HITEK POWER XR KW X-RAY POWER SUPPLY MODULE

TBD62308AFAG TBD62308AFAG. TOSHIBA BiCD Integrated Circuit Silicon Monolithic. 4channel Low active high current sink type DMOS transistor array

Pressure transmitter AKS 32 and AKS 33

MODEL TS-113 VOLT-OHM-MILLIAMMETER. Operator s Manual WARNING READ AND UNDERSTAND THIS MANUAL BEFORE USING THE INSTRUMENT

Keysight 86205B RF Bridge

MFAC. Grid Solutions. High Impedance Differential Relay. MFAC Types. Key Benefits. Application. Description. Imagination at work

E2V Technologies M5187F X-Band Magnetron

Power Quality Solutions

U5700 Submersible Liquid Level Pressure Transducer

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

High Voltage Surge Resistor

This section is specifically about safety matters

TLP3902 TLP3902 SOLID STATE RELAY PROGRAMMABLE CONTROLLERS MOSFET GATE DRIVER. Features. Pin Configuration (top view)

MXRP-160C. Product Description. Product Specifications

OKI Electronics Components OM5653C-30B 1. DESCRIPTION 2. FEATURES

TOSHIBA Fast Recovery Diode Silicon Diffused Type CMF01

TD62783AP,TD62783AF,TD62784AP,TD62784AF

Transcription:

Technical Data TD ROTANODE E7239X E7239FX E7239GX 0197 Rotating Anode X-Ray Tube Housing Assembly Rotating anode X-ray tube housing assembly for the purpose of general diagnostic X-ray procedures. Specially processed Rhenium-tungsten faced molybdenum target of 74 mm diameter. These tubes have foci 2.0 mm and 1.0 mm, and are available for a maximum tube voltage 125 kv with Three-phase generator. Accommodated with IEC 60526 type high-voltage cable receptacles. General Data IEC Classification... Class I Type B Electrical: Circuit (Center-grounded)... Three-phase full-wave rectified Nominal Tube Voltage: Radiographic... 40 ~ 125 kv Max. Nominal Focal Spot Value: Large Focus... 2.0 mm Small Focus... 1.0 mm Nominal Anode Input Power (at 0.1s): 60 Hz Large Focus... 47 kw Small Focus... 22.5 kw The information contained herein is presented only as a guide for the applications of our products. No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others. The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of equipment incorporating this product. For further particulars apply to TETD. No. TE-E7239X/FX/GX 2004-11-01

Motor Ratings: Duty Starting Running Power source (Hz) 50/60 50/60 Input power (W) 1050 270 43 Voltage 1) (V) 200 100 40 Current (A) 6.0 3.0 1.2 Min. Speed up 2) (s) 0.8 1.5 - Capacitor (µf) 24 24 24 Note: 1) The every applied voltage must be never exceeded 110% of the above specification. 2) The speed-up time is allowed up to 110% of the above specification. Anode Speed: 50 Hz... 2700 min -1 Min. 60 Hz... 3200 min -1 Min. Stator impedance: Common-Main Winding... 27.5 Ω Common-Auxiliary Winding... 58.0 Ω Resistance between Housing and Low Voltage Terminals... 2 MΩ Min. Normal operating range of the housing temperature... 16 ~ 75 C Mechanical: Dimensions:... See dimensional outline Overall Length... 479 mm Maximum Diameter... 152.4 mm Target: Angle... 16 degrees Construction... Rhenium-Tungsten faced Molybdenum Permanent Filtration... 0.9 mm Al / 75 kv IEC 60522 / 2003 Radiation Protection (To meet the requirements of IEC 60601-1-3): Leakage Technique Factor... 125 kv 4 ma X-ray Coverage... 354 354 mm at SID 750 mm Weight (Approx.)... 16 kg High Tension Terminals... To meet the requirements of IEC 60526 Cooling Method... Natural or forced air Tube Housing Model Number: E7239X... XH-121 E7239FX... XH-126 E7239GX... XH-150-2 -

Absolute Maximum and Minimum Ratings (At any time, these values must not be exceeded.) Maximum Tube Voltage: Radiographic... 125 kv Maximum Voltage to Ground... 65 kv Minimum Tube Voltage... 40 kv Maximum Tube Current: Large Focus... 570 ma Small Focus... 340 ma Maximum Filament Current: Large Focus... 5.1 A Small Focus... 5.1 A Filament Voltage: Large Focus (At max. filament current 5.2 A)... 7.7 ~ 10.4 V Small Focus (At max. filament current 5.2 A)... 5.8 ~ 7.8 V Filament Frequency Limits... 0 ~ 25 khz Average Input Power... 60 W (85 HU/s) (Repeated radiographic exposure) Thermal Characteristics: Anode Heat Storage Capacity... 100 kj (140 khu) Maximum Anode Heat Dissipation Rate... 475 W (667 HU/s) Housing Heat Storage Capacity... 900 kj (1250 khu) Maximum Housing Heat Dissipation Rate: Without Air-circulator... 180 W (15 khu/min) - 3 -

Environmental Limits Operating Limits: Temperature... 10 ~ 40 C Humidity... 30 ~ 85 % (No condensation) Atmospheric Pressure... 70 ~ 106 kpa Shipping and Storage Limits: Temperature... -20 ~ 70 C Humidity... 20 ~ 90 % (No condensation) Atmospheric Pressure... 50 ~ 106 kpa - 4 -

Warning Warning to Interface with X-ray Generator 1. Housing Rupture Never input over-rated power to x-ray tube assembly. If the input power is extremely higher than specification, it may cause the over temperature of anode, insert tube glass shatter and ultimately the following serious problems due to generating over-pressure by oil vaporization inside housing assembly. In such a critical condition, the safety thermal switch can not protect x-ray tube even if it works. * Housing sealing parts (cathode side) rupture * Human injury including burns due to hot oil escape * Fire accident due to flaming anode target We strongly request that the x-ray generator should have a protective function which manages input power to x-ray tube assembly. - 5 -

Cautions Caution to Interface with X-ray Generator 1. Over Rating X-ray tube assembly can be broken with applying just one over rated shot. Please read the technical data sheets carefully and follow the instructions. 2. Permanent Filtration The total filtration and the distance between x-ray focal spot and human body are regulated legally. They should be complied with the regulation. 3. Safety Thermal Switch X-ray tube assembly has safety thermal switch to prohibit further input power when the tube housing reaches to the temperature of switch-open. The switch should be hooked up with the x-ray generator which control output power to x-ray tube assembly. Even if the switch works, never turn the system power off and the cooling unit should be activated. 4. Unexpected Malfunction X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or failure. If the serious problems caused by the above risk is expected, we recommend to have a contingency plan to avoid such a case. 5. New Application If you use the product with new application not to be mentioned in this specification or with different type of x-ray generator, please contact to us for confirming its availability. - 6 -

Caution for Installation, Adjustment and Maintenance 1. Qualified Persons Only qualified persons who have technical training and professional knowledge can handle x-ray tube assembly. 2. Fragile Glass X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or pulsed shock over 19.6m/s 2 (2G). Careful handling is required to treat or transport. 3. Ground Terminal X-ray tube assembly has ground terminal. Ground cable should be connected. 4. High Voltage All x-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the high voltage delivered plugs or terminals. When direct access to such parts is required, the primary circuit should be disabled and high voltage capacitors/cables discharged. 5. High Voltage Plug High voltage plug should be cleaned up and free from any physical damages. Silicon compound application is required for high voltage stability. 6. Operation Atmosphere X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas. 7. Protective Cover X-ray tube assembly is not allowed to use without the protective cover attached. 8. Handling Appropriate jig or tools are required for tube installation to avoid physical damages. 9.Returning Tube X-ray tube assembly should be repackaged with the original material when it is returned back for quality examination in our factory. Be careful to put the tube upside cathode. If the packaging is not proper, the tube may not be correctly examined. - 7 -

Caution in Operation 1. X-Ray Radiation X-ray tube assembly should have the beam limiting equipment mounted on the x-ray port to protect unnecessary radiation. 2. Dielectric Oil X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for human health, if it is exposed to the non-restricted area, it should be disposed as following to the local regulation. 3. Operation Atmosphere X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas. 4. Lead Disposition X-ray tube housing is lined with lead to protect unnecessary radiation. As the lead powder or vapor is harmful for human health, it should be disposed as following to the local regulation or returned back to us with your cost of transportation. We dispose it in our facility with free of charge. 5. X-ray tube housing temperature Do not touch on X-ray tube housing surface just after operation due to high temperature. Stay X-ray tube to be cooled. 6. Any Malfunction Please contact to your system service person immediately, if any malfunction is noticed. - 8 -

Caution Label (a) This label is a caution label to notify the user of the following point. "Housing end cap is used to protect the electric shock and x-ray leakage." Attachment position: X-ray tube assembly housing end cap - 9 -

Maximum Rating Charts (Absolute Maximum Rating Charts) Conditions : Tube Voltage Three-Phase Stator Power Frequency 60Hz Focal Spot : 2.0 mm Focal Spot : 1.0 mm TUBE CURRENT [ma] 570 100kV 90kV 550 110kV 500 125kV 450 400 350 300 250 200 150 80kV 70kV60kV50kV 40kV TUBE CURRENT [ma] 340 300 250 200 150 100 80kV 90kV 100kV 110kV 125kV 70kV 60kV 50kV 40kV 100 50 50 0 0.01 0.03 0.07 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10 0.3 0.7 EXPOSURE TIME [s] 0 0.01 0.03 0.07 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10 0.3 0.7 EXPOSURE TIME [s] Conditions : Tube Voltage Three-Phase Stator Power Frequency 50Hz Focal Spot : 2.0 mm Focal Spot : 1.0 mm TUBE CURRENT [ma] 570 550 500 450 400 350 300 250 200 150 90kV 80kV 70kV 60kV 50kV 100kV 110kV 125kV 40kV TUBE CURRENT [ma] 340 300 250 200 150 100 70kV 60kV 50kV 80kV 90kV 100kV 110kV 125kV 40kV 100 50 50 0 0.01 0.03 0.07 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10 0.3 0.7 EXPOSURE TIME [s] 0 0.01 0.03 0.07 0.02 0.05 0.1 0.2 0.5 1 2 3 5 7 10 0.3 0.7 EXPOSURE TIME [s] - 10 -

Emission & Filament Characteristics Three-Phase Focal Spot : 2.0 mm 600 18 TUBE CURRENT [ma] 500 400 300 200 125kV 100kV Ef 80kV 40kV 50kV 60kV 16 14 12 10 FILAMENT VOLTAGE [V] 100 8 0 6 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1 FILAMENT CURRENT [A] Focal Spot : 1.0 mm 350 300 100kV 80kV 16 14 TUBE CURRENT [ma] 250 200 150 100 125kV 40kV 50kV 60kV 12 10 8 6 FILAMENT VOLTAGE [V] 50 Ef 4 0 2 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1 FILAMENT CURRENT [A] - 11 -

Thermal Characteristics Housing Thermal Characteristics 900 800 180W HEAT STORAGE [kj] 700 600 500 400 300 COOLING 200 100 HEATING 0 0 20 40 60 80 100 120 140 160 180 200 TIME [min] Cooling Without Air Circulator. 100 90 Anode Thermal Characteristics HEAT STRAGE [kj] 80 70 60 50 40 30 20 COOLING 60W 10 HEATING 0 0 1 2 3 4 5 6 7 8 TIME [min] The heating curves are showing example of average input power to anode in operation. - 12 -

Dimensional Outline of E7239X Unit mm 140 4-M6 8 DEEP 41.27 4-M5 10 DEEP 92 45 152.4 90 11.1 11.1 125.4 170 156 479 CENTRAL RAY AND REFERENCE AXIS L S ANODE 16 CATHODE C TERMINAL CONNECTIONS 81 53 11.1 121 2-M63.5, P=1.27 9 8 7 6 5 NCNCNC 1 2 3 4 E M A TEMPERATURE RELAY (NORMALLY CLOSED) NC ET 70 Note) Do not connect terminal No.1 and No.5 or 6 in series circuit. FOCAL SPOT A CENTRAL OF FOCAL SPOT B A:±1.5mm,B:±1.5mm C : COMMON L : LARGE FOCUS S : SMALL FOCUS M : MAIN WINDING OF THE STATOR A : AUX. WINDING OF THE STATOR NC: NON-CONNECTION ET: EARTH TERMINAL : CENTRAL X-RAY ANODE & CATHODE TERMINAL : IEC 60526 TYPE - 13 -

Dimensional Outline of E7239FX Unit mm 140 ANODE 16 CATHODE L S CENTRAL RAY AND REFERENCE AXIS 479 170 156 125.4 11.1 11.1 90 45 92 152.4 4-M6 8 DEEP 41.27 4-M5 10 DEEP C TERMINAL CONNECTIONS 9 8 7 6 5 1 2 3 NCNCNC M A 4 E NC ET 2-M63.5, P=1.27 121 11.1 53 TEMPERATURE RELAY (NORMALLY CLOSED) 81 Note) Do not connect terminal No.1 and No.5 or 6 in series circuit. 70 A CENTRAL OF FOCAL SPOT B FOCAL SPOT A:±1.5mm,B:±1.5mm C : COMMON L : LARGE FOCUS S : SMALL FOCUS M : MAIN WINDING OF THE STATOR A : AUX. WINDING OF THE STATOR NC: NON-CONNECTION ET: EARTH TERMINAL : CENTRAL X-RAY ANODE & CATHODE TERMINAL : IEC 60526 TYPE - 14 -

Dimensional Outline of E7239GX Unit mm 4-M6 8 DEEP 41.27 92 45 90 CENTRAL RAY AND REFERENCE AXIS 140 16 11.1 125.4 170 479 ANODE 4-M5 10 DEEP 152.4 11.1 156 L S CATHODE C 81 121 A CENTRAL OF FOCAL SPOT 2-M63.5, P=1.27 TERMINAL CONNECTIONS B A:±1.5mm,B:±1.5mm 70 9 8 7 6 NC NC NC 5 1 M 2 3 4 A NC E TEMPERATURE RELAY (NORMALLY CLOSED) ET 53 FOCAL SPOT Note) Do not connect terminal No.1 and No.5 or 6 in series circuit. C : COMMON L : LARGE FOCUS S : SMALL FOCUS M : MAIN WINDING OF THE STATOR A : AUX. WINDING OF THE STATOR NC : NON-CONNECTION ET : EARTH TERMINAL : CENTRAL X-RAY ANODE & CATHODE TERMINAL : IEC 60526 TYPE - 15 -

OVERSEAS SUBSIDIARIES AND AFFILIATES EU REPRESENTATIVE TOSHIBA ELECTRONICS EUROPE RIVERSIDE WAY, CAMBERLEY, SURREY. GU15 3YA U.K. PHONE (0) 1276 694600 FAX (0) 1276 694800 For Sales & Technical Services, please contact the following representative: TOSHIBA ELECTRONICS EUROPE RIVERSIDE WAY, CAMBERLEY, SURREY. GU15 3YA U.K. PHONE (0) 1276 694600 FAX (0) 1276 694800 TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC. ONE PARKWAY NORTH, SUITE 500, DEERFIELD, IL 60015-2547, USA PHONE (847) 945-1500 FAX (847) 945-1044 TOSHIBA ELECTRON DEVICES & MATERIALS TRADING (SHANGHAI) CO., LTD. (TEMS) No.689, GUANG DONG RD, SHANGHAI, 200001,CHAINA RM607, HAITONG SECURITIES TOWER PHONE (21) 6341-0055 FAX (21) 6341-0990 Sales & Marketing Department 1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN PHONE : +81-287-26-6668 FAX : +81-287-26-6061 http://www.toshiba-tetd.co.jp/ Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO 14001 Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO 9001, ISO 13485