Model: PR-55. High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL

Similar documents
For model: HVP-10R / HVP-18HF / HVP-28HF / HVP-39pro

Keysight N2771B 30 kv High Voltage Probe

Operator s Manual. PP022 Passive Probe

Operator s Manual. PP017 and PP018 Passive Probes

Operator s Manual. PP016 Passive Probe

400 MHz Passive High-Voltage Probe R&S RT-ZH

TA MHz oscilloscope probe TA MHz oscilloscope probe

PKT 512A-RO High Impedance Passive Cable Divider

PHV RO High Voltage Passive Probe. Instruction Manual

TA MHz ±30 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT I, Pollution Degree 2.

TETRIS User's Guide. High Impedance Active Probe DO177-1

User s Manual. Miniature Passive Probe. IM EN 3rd Edition

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

TETRIS 1000 High Impedance Active Probe. Instruction Manual

TT-SI MHz Active Differential Probe

TA MHz x10/x100 Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

TA MHz ±700 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

User s Manual Current Probe IM E. 8th Edition

TT-SI MHz DIFFERENTIAL PROBE

TT-SI 9001 / TT-SI MHz Active Differential Probes

User s Manual Current Probe. IM E 1st Edition. Yokogawa Electric Corporation

PHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

TT-SI MHz Active Differential Probe

User s Manual Current Probe. IM E 2nd Edition IM E

Supplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

Digital Clamp Meter Model: &

User s Manual Current Probe. IM E 2nd Edition. Yokogawa Electric Corporation

AC/DC CLAMP METER USER S MANUAL

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

MHz FUNCTION GENERATOR INSTRUCTION MANUAL

DDS Function Generator

PT-2000series AC/DCcurrentprobe PT-2710 PT-2720 PT-2740 PT-2770 MANUAL INSTRUCTION

supplies and frequency converters. BumbleBee is also very effective in fast transient measurements

PHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design

MD 200A High voltage differential probe

KMD-S04 Multímetro de bolsillo

User s Manual Current Probe IM E. 6th Edition

Tablet Oscilloscope Safety Information

PMT 211A-RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

Electrical Multimeter

BumbleBee. High Voltage Differential Probe Ord.-No: Features:

PMT 221A. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design. Interchangeable Spring Contact Tip

AC/DC Current Probe GCP-100 QUICK START GUIDE ISO-9001 CERTIFIED MANUFACTURER

80i-110s AC/DC CURRENT PROBE

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

Model 4007DDS. 7 MHz Sweep Function Generator

USER'S MANUAL DMR-6700

Electrical Multimeter

CONTENTS MS2033A. 1.Introduction Safety Information...01

Instruction Manual. P6015A 1000X High Voltage Probe

NOTE: Fully read and understand this manual before using this Digital Multimeter.

N2790A Differential Voltage Probe

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

Keysight N2790A Differential Voltage Probe. User Guide

TK 100 Probe Calibrator. Instruction Manual

UT232. Operating Manual. Digital Power Clamp Meter

Agilent N2780A, N2781A, N2782A, and N2783A Current Probes

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

PML 711A-RO High impedance passive probe Features:

374/375/376 Clamp Meter. Users Manual

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

INSTRUCTION MANUAL CLAMP ON PROBE

Model /701913/ PBA1000/2500/1500 Active Probe

80i-110s. AC/DC Current Probe Instructions. Introduction

Model : OPERATING MANUAL Table of Contents (1)

netzerotools.com 374/375/376 Users Manual Clamp Meter

1147A 50 MHz Current Probe

15B+/17B+/18B+ Calibration Manual. Digital Multimeter

323/324/325 Clamp Meter

374/375/376 Clamp Meter

SA1500X. Operating Basics. Cautions DEVICE SPECIFICATIONS. High Impedance Active Probe

UT207A/208A/209A Operating Manual. Table of Contents

374 FC/375 FC/376 FC Clamp Meter

AC/DC Clamp Meter. Owner's Manual. Model No Safety Operation Maintenance Español

Model UT10A: OPERATING MANUAL Table of Contents (1)

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

DM-45 Digital Multimeter

T5-600/T Electrical Tester

Analog Technologies. Multimeter 15B and17b

Dawson DDM190. Digital Multimeter User s Manual

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

Handheld Digital Multimeter PRO-50A

P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Model MHz Arbitrary Waveform / Function Generator

RP1003C/RP1004C/RP1005C

Digital Multimeter, Pen Type

DS1000E, DS1000D Series Digital Oscilloscope

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95

Model 4210-MMPC-L. Multi-measurement Prober Cable Kit. Overview. Quick start guide topics. Related documents

RIGOL. User s Guide. RP1003C/RP1004C/RP1005C Current Probe. Mar RIGOL Technologies, Inc

DDM350 Pen-Type Digital Multimeter User s Manual

BENNING. CM7 Electrical Tester. Service Information

374/375/376 Clamp Meter

True RMS Digital Multimeter Model:

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

AC/DC True RMS Power Clamp Meter ELECTRONIC TEST INSTRUMENTS MODEL 325

Transcription:

Model: PR-55 High Voltage Oscilloscope Probe 10 kv, 40 MHz USER MANUAL 1

2017 All rights reserved. B&K Precision products are covered by US and foreign patents, issued and pending. Information is this publication supersedes all earlier versions. Specifications subject to change without notice. For product information, sales, service, and technical support: Call 1-714-921-9095 or visit: bkprecision.com Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems). This product is subject to Directive 2012/19/EU of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements. 2

Safety Summary WARNING CAUTION To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Only qualified personnel should use this probe. This high voltage probe is designed to be used by personnel who are trained, experienced, or otherwise qualified to recognize hazardous situations and who are trained in the safety precautions necessary to avoid possible injury when using such a device. This instrument is intended for usage within Measurement Category I (CAT I) only. Do not exceed 60 seconds of contact with high voltage. When making contact between this probe and any circuit with voltages 8 kv DC, 5 kvrms AC, or 15 kvpk-pk AC, wait for 5 minutes between such measurements. Do not work alone when working with high voltages. Do not apply to the input any potential that exceeds the maximum rating of the probe. For your own safety, inspect the probes for cracks and frayed or broken leads before each use. If defects are noted, DO NOT USE the probe. Hands, shoes, floor and work bench must be dry. 3

Avoid making measurements under humidity, dampness or other environmental conditions that might affect safety. Turn the high voltage source off before connecting or disconnecting the probe whenever possible. The probe body should be kept clean and free of any conductive contamination. Do not remove the probe casing. Removal of the probe s casing may expose you to electric shock. Keep fingers behind the probe s finger guard during use. Always properly ground the probe with the lead before contacting high voltage circuits with the probe tip. Always disconnect the probe tip from high voltage circuits before disconnecting the ground lead. Do not connect the probe ground lead to any point which is at a potential other than earth ground. Use only on test instruments where the chassis or return lead is properly grounded. Use only in office-type indoor setting The instrument is designed to be used in office-type indoor environments. Do not operate the instrument: In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates. In environments where there is a danger of any liquid being spilled on the instrument. In air temperatures exceeding the specified operating temperatures. 4

In atmospheric pressures outside the specified altitude limits or where the surrounding gas is not air. Use in an indoor Pollution Degree 2 environment. Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields. Not for critical applications. This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system. Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists. Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained. A WARNING statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in injury or death to personnel. A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in damage to or destruction of parts or the entire product. 5

1 General Information 1.1 Product Overview The B&K Precision PR-55 is a 10 kv High Voltage divider probe for use with both digital and analog oscilloscopes. 1.2 Features Measure up to: 10 kv (DC + AC pk ) or 7 kv rms AC 40 MHz bandwidth Voltage dividing of 1000:1 Frequency compensation 2 Using the Probe 1. Connect the probe to the BNC input of the oscilloscope. 2. Select the desired volts/division range on your oscilloscope. 3. If possible, always turn the high voltage source off before connecting or disconnecting the probe. 4. Connect the divider probe ground lead (alligator clip) to a good earth ground or reliable chassis ground. 5. Before turning on the high voltage source, be certain the operator is not touching any part of the device under test. 6. Turn on the high voltage source. 6

7. You will now be able to analyze the voltage waveform on your oscilloscope. 8. Turn off the high voltage source. 9. Disconnect the High Voltage Probe from the high voltage source BEFORE disconnecting the ground clip lead. WARNING Do not exceed 60 seconds of contact between this probe and any circuit with voltages 8 kv DC, 5 kv rms AC, or 15 kv pk-pk AC. Wait for 5 minutes between such measurements. 3 Frequency Compensation Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope s input capacitance. Compensation should be adjusted whenever the probe is connected to, or transferred between, oscilloscopes. The procedure is the same as for a x10 passive probe. Connect the probe to the oscilloscope. Connect probe tip to square-wave generator. Adjust the square wave generator for 10 V amplitude. 7

Adjust the oscilloscope for a time base of 500 µs/div. Turn trimmer (Figure 1) to adjust the compensation on the probe making a flat pulse top. Figure 1: Trimmer on BNC Figure 2 shows how over- and undercompensated pulse responses will look. Figure 2: Over/Under Compensation 8

Figure 3 shows a perfectly compensated probe. Figure 3: Good Compensation 9

4 Specifications All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature range of 25 C ± 5 C. PR-55 10 kv High Voltage Probe Operating Parameters Maximum Input Voltage (CAT I) 10 kv (DC + AC pk) 7 kv rms AC Division Ratio 1000:1 Bandwidth 40 MHz (-3 db) Compensation Range 5 pf to 50 pf Temperature Coefficient 200 ppm/ºc Input Resistance 100 MΩ ± 5% Rise Time 8.8 ns Accuracy DC Volts ±3% (below 10 kv) ±3% at 1 khz AC Volts ±3 db 0 to 40 MHz General Cable Length 6.6 ft (2 m) Designed for Use in Pollution Degree 2 Operating Environment 32ºF to 122ºF (0ºC to 50ºC) Storage Temperature -4ºF to 158ºF (-20ºC to 70ºC) 80% relative humidity at Humidity 104ºF (40ºC) Dimensions 13 in (34 cm) Weight 0.55 lbs (250 g) Specifications are subject to change without notice. To ensure the most current version of this manual, please download the current version from our website: bkprecision.com. 10

5 Voltage Derating Curve WARNING When measuring higher frequency signals, be sure to comply with the Voltage vs Frequency Derating Curve. Figure 4: Voltage Derating Curve 6 Cleaning Clean only the exterior probe body and cables. Use a soft cotton cloth lightly moistened with a mild solution of detergent and water. Do not allow any portion of the probe to be submerged at any time. WARNING Dry the probe thoroughly before attempting to make voltage measurements. CAUTION Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables. 11

B&K Precision 22820 Savi Ranch Parkway Yorba Linda, CA 92887 800-462-9832 or 714-921-9095 bkprecision.com 2017 B&K Precision 12