Enhancing the capability of primary calibration system for shock acceleration in NML

Similar documents
Development of Shock Acceleration Calibration Machine in NMIJ

A study of Savitzky-Golay filters for derivatives in primary shock calibration

UPGRADE OF THE MEDIUM AND HIGH FREQUENCY VIBRATION CALIBRATION REFERENCE EQUIPMENT AND EXTENSION TO LOW FREQUENCIES

CALIBRATION OF LASER VIBROMETER STANDARDS ACCORDING TO ISO

Continuous development of the national standard for vibration

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

Principles of Vibration Measurement and Analysis. Dr. Colin Novak, P.Eng July 29, 2015

Dynamic Structure Evaluation of Isolation Seismic Block for Primary Vibration Calibration System

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Microscopic Laser Doppler Vibrometer

LASER VIBROMETER CALIBRATION AT HIGH FREQUENCIES USING CONVENTIONAL CALIBRATION EQUIPMENT

Vibration Transducer Calibration System

System Inputs, Physical Modeling, and Time & Frequency Domains

Model CAACS Commercial Automated Accelerometer Calibration System

Model CAACS Commercial Automated Accelerometer Calibration System

Pilot Study EURAMET.AUV.V-P1: Bilateral comparison on magnitude of the complex charge sensitivity of accelerometers from 10 Hz to 10 khz

New Long Stroke Vibration Shaker Design using Linear Motor Technology

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

APMP.AUV.V-K1 Regional Key Comparison of Standard Accelerometer. Final Report. Project Number: APMP-IC Shing Chen and Hans-Jürgen von Martens

Very High Frequency Calibration of Laser Vibrometer up to 350 khz

D.C. Emmony, M.W. Godfrey and R.G. White

INTERFEROMETRIC VIBRATION DISPLACEMENT MEASUREMENT

Optical Encoder Applications for Vibration Analysis

430. The Research System for Vibration Analysis in Domestic Installation Pipes

2 Study of an embarked vibro-impact system: experimental analysis

Study on a Single-Axis Fabry-Perot Fiber-Optic Accelerometer and its Signal Demodulation Method

School of Instrument Science and Opto-electronics Engineering, Hefei University of Technology, Hefei, China 2

The Virgo detector. L. Rolland LAPP-Annecy GraSPA summer school L. Rolland GraSPA2013 Annecy le Vieux

1272. Phase-controlled vibrational laser percussion drilling

Primary vibration calibration by laser interferometry requirements, problems and first experience with a new calibration system

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

A New Solution for Shock and Vibration Calibration of Accelerometers

Fabry Perot Resonator (CA-1140)

Department of Electrical Engineering and Computer Science

1433. A wavelet-based algorithm for numerical integration on vibration acceleration measurement data

CALIBRATION OF ACCELEROMETERS USING PARAMETER IDENTIFICATION TARGETING A VERSATILE NEW STANDARD

Laboratory 14. Lab 14. Vibration Measurement With an Accelerometer

VIBRATION CONTROL APPLIED TO AN ELECTRODYNAMIC EXCITER TO IMPROVE ACCELEROMETER CALIBRATIONS

point at zero displacement string 80 scale / cm Fig. 4.1

T40FH. Torque flange. Special features. Data sheet

Technical data. Your advantages

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

Angle Encoder Modules

Properties and Applications

Bilateral Comparison EURAMET.AUV.V-K1.2. (vibration acceleration) Swiss Federal Office of Metrology METAS Christian Hof

Dynamic measurement activities at PTB Hans-Jürgen von Martens Physikalisch-Technische Bundesanstalt (PTB), Germany

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Stability of a Fiber-Fed Heterodyne Interferometer

The units of vibration depend on the vibrational parameter, as follows:

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

This document is a preview generated by EVS

MAGNETIC LEVITATION SUSPENSION CONTROL SYSTEM FOR REACTION WHEEL

Implementation of an Accelerometer Transverse Sensitivity Measurement System. By: Ian Veldman 3 to 5 September 2012

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

PRODUCT DATA. Piezoelectric Accelerometer Miniature Triaxial DeltaTron Accelerometers Types 4524, 4524 B

Theory and Applications of Frequency Domain Laser Ultrasonics

INTERNATIONAL STANDARD

LONGITUDINAL TRACKING OF DIRECT DRIVE INERTIAL FUSION TARGETS. 2 General Atomics, P.O. Box 85608, San Diego CA

Calibration of piezoelectric accelerometers at INTI

The ENDEVCO high-g shock triaxial accelerometer: A smaller, more cost-effective solution to making triaxial measurements. Endevco technical paper 334

ni.com Sensor Measurement Fundamentals Series

Instrumentation (ch. 4 in Lecture notes)

Model DML 100 Accuracy class 0.1 No. Nominal torque M nom (bidirectional) Nm 100 Speed range rpm 0 10,000 Output signal Unit Value

Directly Chirped Laser Source for Chirped Pulse Amplification

HAMEG Programmable Measuring Instruments Series 8100

DRTS 66 The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

16.3 Standing Waves on a String.notebook February 16, 2018

Torque Sensor DML 500 S080S10X

Agilent AEDA-3300 Series Ultra Miniature, High Resolution Incremental Kit Encoders Data Sheet

5. Transducers Definition and General Concept of Transducer Classification of Transducers

Development of innovative fringe locking strategies for vibration-resistant white light vertical scanning interferometry (VSI)

The Study on the Method of Eliminating Errors of PSD

Research on the Transient Response and Measure Method of Engineering Vibration Sensors

Assessment of the Metrological Performance of Seismic Tables for a QMS Recognition

AGN 008 Vibration DESCRIPTION. Cummins Generator Technologies manufacture ac generators (alternators) to ensure compliance with BS 5000, Part 3.

Annex to the Accreditation Certificate D-K according to DIN EN ISO/IEC 17025:2005

Realization of 16-channel digital PGC demodulator for fiber laser sensor array

Do all accelerometers behave the same? Meggitt-Endevco, Anthony Chu

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers

DEVELOPMENT OF STABILIZED AND HIGH SENSITIVE OPTICAL FI- BER ACOUSTIC EMISSION SYSTEM AND ITS APPLICATION

The PIA-devices are used for high g-accelerating an attached mass or to produce a compression impact into extended mass-loaded structures.

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz

ULTRASONIC TRANSDUCER PEAK-TO-PEAK OPTICAL MEASUREMENT

VCSO Mechanical Shock Compensation

T10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Application of Fiber Optic Sensors for Stator End Winding Vibration Monitoring. M. Sasic, R. Sadanandan, G. Stone Iris Power Qualitrol

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 4 Fall Term

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION

Dynamic Vibration Absorber

Part VI: Requirements for Integrated Services Digital Network Terminal Equipment

CCAUV/ CCAUV. Activities in KRISS. Wan-Cho Cho

A Fast Phase meter for Interferometric Applications with an Accuracy in the Picometer Regime

Sensing. Autonomous systems. Properties. Classification. Key requirement of autonomous systems. An AS should be connected to the outside world.

Resonance Tube Lab 9

Passive Wireless Sensors

Development of a Package for a Triaxial High-G Accelerometer Optimized for High Signal Fidelity

VCSO Mechanical Shock Compensation

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.

Characterization of Silicon-based Ultrasonic Nozzles

Transcription:

Enhancing the capability of primary calibration system for shock acceleration in NML Jiun-Kai CHEN 1 ; Yen-Jong HUANG 1 1 Center for Measurement Standards, Industrial Technology Research Institute, R.O.C. ABSTRACT As the guardian of the national measurement standards, the Vibration laboratory of National Measurement Laboratory (NML) in Taiwan in compliance with the standard ISO 16063-13 had established primary shock calibration system in 2009, which was based on the rigid body collision method, with peak acceleration ranging from 200 m/s 2 to 5 000 m/s 2 and shock pulse duration less than 2.5 milliseconds. Hereafter, upgrading the capacity for measuring shock acceleration by increasing the maximum acceleration from 5 000 m/s 2 to 10 000 m/s 2 in 2011.The system had participated regional pilot comparison of the APMP Pilot Comparison (APMP.AUV.V-P1) in 2014. During the comparison, a weak point of capability was shown that the shock pulse was shorter than other NMIs. This research discusses the improvement of increasing the duration time of shock calibration system. Upgrade the electromagnetic shock source and reform the relative position of airborne hammer and anvil. Controlling different input DC voltage value and refining different hardness rubber will produce different acceleration and duration time. After enhancing the capability of system, the maximum shock pulse duration time is increasing from 2.5 milliseconds to 4 milliseconds. Finally, the system performance data is provided together with evaluation of measurement uncertainty. Keywords: Shock calibration, Primary, I-INCE Classification of Subjects Number(s):72.2.2 (See. http://www.inceusa.org/links/subj%20class%20-%20formatted.pdf.) 1. INTRODUCTION The electronic and acoustical products, vehicle electronics, information goods and so on, must be executed with impact test during R&D process. In the meantime, it is also required that the shock accelerometer be applied to the vehicle hit-test to monitor acceleration. Therefore, shock calibration technology and the associated industrial development are obvious to be closely linked. As the guardian of the national measurement standards and echoing to industrial requirements, the Vibration laboratory of National Measurement Laboratory (NML) in Taiwan in compliance with the standard ISO 16063-13 (1) had established shock calibration system in 2010, which is based on the rigid body collision method, with peak acceleration ranging from 200 m/s 2 to 5 000 m/s 2 and shock pulse duration less than 2.5 milliseconds.(2) The system is different from other NMIs in the use of electromagnetic exciter, which shows the superiority of good reproducibility in the generation of peak shock acceleration.(3) Hereafter we had kept upgrading the capacity for measuring shock acceleration by increasing the maximum acceleration from 5 000 m/s 2 to 10 000 m/s 2 in 2011(4). ISO 16063 mainly specifies two mechanical vibration modes to calibrate accelerometer. They are sinusoidal vibration and shock vibration, respectively. For the main difference between the two typical vibration modes, the sinusoidal vibration mode is for calibration at each discrete frequency point with small vibration amplitude, common acceleration amplitude does not over 1 000 m/s 2 ;while shock vibration mode is a transient shock pulse with short duration and large amplitude (acceleration amplitude up to 100 000 m/s 2 ). Mechanical vibration measurement methods are divided into primary method and secondary method. Primary method adopts laser interferometry to determine the input mechanical vibration acceleration accelerometer, so it has very high calibration precision. 1 jkchen@itri.org.tw 670

2. SYSTEM DESCRIPTION The primary calibration system for shock acceleration is established by referring to ISO 16063-13 rigid body motion method. The system mainly includes Shock level activating module, Laser interferometer module and Signal access device. The system connection diagram is showed in Figure-1. Shock level activating module refers to 1 in Figure-1, and Laser interferometer module is showed as 11 in Figure-1. The symbols of element labeling are as followed: 1. Shock level activating module 2. EM activating controller 3. Programmable DC power supply 4. High speed EM hammer 5. Shock hammer 6.Trigger device 7. Rubber mat 8. Shock anvil 9. Shock accelerometer 10. Charge amplifier or power supply 11. Laser interferometer module 12. Single frequency stable frequency laser tube 13. Light detection module 14. Signal picks up device (Digital oscilloscope card) 15. Optics table Figure-1 Diagram of system connection The present primary calibration system for shock acceleration makes use of an accelerometer, ENDEVCO/2270, fixed on an airborne anvil after shock, and then through charge amplifier, ENDEVCO/133, to output voltage signal. Time-sequence signals are recorded by a digital waveform recorder, filtering and calculating maximum shock voltage signal V max. When an accelerometer is shocked, dual-channel interfering signals will be synchronously recorded via laser interferometer, and then time-sequence signals will be recorded via another digital waveform recorder. After dual-channel interfering signals are processed through low-pass filter and twice differential, maximum shock acceleration A max can be calculated when an accelerometer is shocked. Through the ration between V max and A max, voltage sensitivity of an accelerometer, S sh =V max /A max, can be determined. 2.1 Upgrade shock level activating module The shock source of shock level activating module in Figure-2 is that one voltage input by electromagnetic excitation controller drives a high speed electromagnetic hammer through programmable DC supply to produce external shock force, which hits an airborne hammer supported with air bushing. After hit by a high speed electromagnetic anvil, an airborne hammer will continue to hit another airborne anvil supported by air bushing at an approximate uniform motion. In this study, we upgrade an electromagnetic excitation power supply and increase the rod s radius of electromagnetic hammer from 25 mm to 30 mm. The front end of airborne anvil is fixed by PU rubber with different Shore hardness. The back end fixes a shock accelerometer. When an airborne anvil is hit by an airborne hammer, it will conduct 671

shock force by rigid motion to an accelerometer. At the meantime, an accelerometer will output one voltage with approximate half-sine shock waveform under external shock force. The maximum voltage will be V max after low-pass filtering. The radius of airbone hammer and airbone anvil increase from 25 mm to 30 mm. An airborne hammer and anvil in this module are supported by two radial porous air bushings and the corresponding air bushings also upgrade to right size. The radial gap between axis and air bushing is about 4 μm. The central line of both airborne hammer and anvil is aligned as ±0.2 mm. The axial linear and rotational motion can freely perform under low friction. In the actual occurrence of shock, the lateral motion of non-axis motion will take place as well. In order to evaluate this characteristic, we make use of a triaxial high shock accelerometer to install on the back end of an airborne anvil and measure the lateral motion ratio. The ratio of maximum lateral motion for that system is less than 4 %.On the other hand, controlling different input DC voltage value and refining different hardness rubber will produce different acceleration and duration time. Radius of electromagnetic hammer: upgrade from 25 mm to 30 mm, and bigger power supply Radius of hamer and anvil: upgrade from 25 mm to 30 mm Air bushings: upgrade from 25 mm to 30 mm Figure 2 Upgrade the shock level activating module Shock accelerometers used for the primary calibration system for shock acceleration are accelerometer ENDEVCO/2270 and charge amplifier ENDEVCO/133, which that accelerometer provides the excellent vibration characteristics: such as maximum transverse sensitivity 0.3%, annual maximum change of charge stability ±0.2%, maximum shock acceleration 150,000 m/s 2 and a linear amplitude increment of 10,000 m/s 2 about 0.1%. In order to simulate mass for an accelerometer under calibration, the back end of a shock accelerometer installs a dummy mass. This mass can be designed as 20 50 g. The dummy mass is made of stainless steel. Uniformity on the mirror end is less than 1/4 wavelength. Attention must be paid to parallelism for the both ends on-mirror and mirror. After dummy mass is installed, the relative motion between the vibration plane of an accelerometer and measurement plane of laser is less than 0.5%. 2.2 Interferometer This laser interferometer is modified Michelson interferometer as prototype, which modifies the current structure. This structure is a combination of modified Michelson interferometer and Mach Zehnder interferometer.(5) Interferometer and shock machine are put on optical table together, where frequency stabilized He Ne laser is applied with a wavelength of 632.99 nm. In order to record high-speed shock data, bandwidth of two photo detectors output from interfering signals needs to meet f max = V max 3.16 106. Therefore, the measuring bandwidth of photo detectors in this system is 150 MHz. Figure-3 shows an illustration of optical interferometer. 672

Figure-3 Illustration of optical interferometer The interferometer mainly makes use of interfering signals u 1 (t) and u 2 (t) of dual-channel phase difference 90 to analyze the phase, where u 1 (t) and u 2 (t) are interfering signals detected by photo detectors. In order to reduce the uncertainty of accelerometer sensitivity, the relative amplitude variation of u 1 (t) and u 2 (t) should be adjusted and controller less than ±5 %. The phase difference should be 85 95, u 1 (t) and u 2 (t) are as shown in Figure-4. Figure-4 Interfering signal u 1 (t) and u 2 (t) of phase difference 90. 2.3 Digital waveform recorder Through Lab View programming, NI 5922 Digital Waveform Recorder will be synchronously triggered by proximity switch to record voltage output from an accelerometer. NI 5124 Digital Waveform Recorder is used to record dual-channel interfering signals, where sampling frequency of interfering signal is 200 MS/s with 12-bit resolution and that of accelerometer voltage is 15 MS/s with 24-bit resolution. 2.4 Digital signal process In order analyze the phase of u 1 (t) and u 2 (t), the phase is defined as: ϕ(t) = tan 1 u 1(t) u 2 (t) This phase is time-sequence data. Upon the phase is determined, the shock displacement is defined as: D(t) = λ 4π φ(t) (2) As interfering signals are influenced by noise, this displacement signal will be filtered by low-pass Butterworth filter and differentiated once to obtain speed signal. After filtered by low-pass filter and differentiated once since speed signal is determined, shock acceleration signal can be obtained. The maximum value in this shock acceleration waveform is A max. This system is digitally filtered with 5 khz cut-off frequency of four-order Butterworth low-pass filter. After the value is filtered by Butterworth low-pass filter, the value string will take place phase (1) 673

delay. Therefore, the front and rear value string will be overturned and filtered by two-order backward Butterworth low-pass filter after filtered by two-order forward Butterworth low-pass filter. This delay will be compensated. 3. SYSTEM EVALUATION This system mainly evaluated the shock accelerometer set Endevco 2270/133 voltage sensitivity uncertainty. The voltage sensitivity of the accelerometer is calculated base on the following formula: S sh = V max A max (3) Where S sh is the accelerometer set voltage sensitivity (mv/(m/s 2 )); V max the voltage of maximum filtered accelerometer output (mv) and A max is the maximum calculated acceleration from interferometer signal (m/s 2 ). According to ISO 16063-13, the mathematical function relationship of uncertainty for accelerometer can be written as y = f(x, e 1, e n ) = x + e 1 + e 2 + + e n. Where the y is the accelerometer sensitivity, x means repeat measurements for accelerometer sensitivity and e 1, e 2, e n each factor means source of uncertainty respectively. From the result of measurement system validation procedure for shock accelerometer-phase operational method, Table 1 shows the example of uncertainty budget for shock level 200 m/s 2. The relative expanded uncertainty is less than 0.8 % within coverage factor k = 2 in the acceleration 200 m/s 2 and the duration time is bigger than 4 milliseconds. Source of uncertainty Table 1 Uncertainty budget for shock level 200 m/s 2 Factor Type A /B Relative standard uncertainty component Repeating evaluation of system 1 A 0.0122 % 9 digital oscilloscope card (NI 5922) resolution 3 B 0.0176 % 200 digital oscilloscope card (NI 5922) accuracy 2 B 0.25 % 200 digital oscilloscope card (NI 5922) 3 B 0.0011 % 200 digital filtering 3 B 0.0483 % 200 hum and noise 1 A 0.0494 % 9 Influence of sensitivity measurement owing to charge stability for accelerometer 3 B 0.1155 % 200 transverse swing 2 B 0.0106 % 200 Influence of sensitivity measurement owing to accelerometer amplitude linearity 3 B 0.0289 % 200 accelerometer amplitude and frequency variation 3 B 0.1403 % 200 charge amplifier linearity 3 B 0.0577 % 200 charge amplifier amplitude and frequency variation 3 B 0.0073 % 200 owing to charge amplifier gain variation 3 B 0.0069 % 200 owing to dual-channel interference phase signal to 3 B 0.0231 % 200 noise DOF ν(x i ) 674

owing to digital filtering 3 owing to relative motion 3 owing to laser wavelength 3 owing to digital oscilloscope card (NI 5124) timing 3 B 0.0483 % 200 B 0.231 % 200 B 0.00053 % 200 B 0.0075 % 200 4. CONCLUSIONS This system utilizes electromagnetism excitation equipment to generate steady impulse force, and then guide the hammer and anvil with rigid body motion collision in an alignment by air bearing support. This research discusses the improvement of increasing the duration time of shock calibration system. Upgrade the electromagnetic shock source and reform the relative position of airborne hammer and anvil. Controlling different input DC voltage value and refining different hardness rubber produce different acceleration and duration time. After enhancing the capability of system, the maximum shock pulse duration time is increasing from 2.5 milliseconds to 4 milliseconds. We use the modified Michelson and Mach Zehnder interferometer to collect the laser interferometer signal and then acceleration of the accelerometer is derived through two numerical differentiations and two low-pass digital filtering from the signal. At present the acceleration capability of primary shock calibration system is from 200 m/s 2 to 10 000 m/s 2, the shock pulse duration time is less than 4 milliseconds. The final evaluated relative expanded uncertainty is less than 0.8 % within coverage factor k = 2. REFERENCES 1. ISO 16063-13, Methods for the calibration of vibration and shock transducers Part 13: Primary shock calibration using laser interferometry, 2001. 2. Yu-Chung Huang, Jiun-Kai Chen, Hsin-Chia Ho, Chung-Sheng Tu, Chao-Jung Chen, The set up of primary calibration system for shock acceleration in NML, Measurement, Volume 45, Issue 10, December 2012, p. 2382-2387. 3. Jiun-Kai Chen, Yu-Chung Huang, Hong-bo Hu, Feasibility study on comparing laser interferometry calibration system with different shock exciters, XX IMEKO World Congress, September 9-14, 2012. 4. Jiun-Kai Chen, Yu-Chung Huang, Increase Shock Energy for Primary Low Level Shock Calibration System, 8 th Chinese Metrology Society Congress, October 18-19, 2010. 5. Hsin-Chia Ho, Jiun-Kai Chen, Chao-Jung Chen, Yu-Chung Huang, The study about laser interferometer applying on shock calibration, The Modern precision theory and application conference between cross-strait, 56 (2009), pp. 8.25 8.26 675