VATTSpec Instructions Rev. 10/23/2015

Similar documents
Goodman Cookbook. Goodman Spectrograph. Adapted by D. Sanmartim from L. Fraga's Guide. Sep SOAR Telescope

DBSP Observing Manual

SARG: The Graphical User Interface Manual

Brief description of GIRAFFE

Cerro Tololo Inter-American Observatory. CHIRON manual. A. Tokovinin Version 2. May 25, 2011 (manual.pdf)

m Copernico Telescope, Ekar User Manual

FLAMINGOS at the KPNO 2.1-m

ISIS A beginner s guide

CCD Image Processing of M15 Images Estimated time: 4 hours

Phase-2 Preparation Tool

SBIG ASTRONOMICAL INSTRUMENTS

Mimir / LOIS Quick Start Guide. Table of Contents

Using CCDAuto (last update: 06/21/05)

SOAR Integral Field Spectrograph (SIFS): Call for Science Verification Proposals

Improved Spectra with a Schmidt-Czerny-Turner Spectrograph

XTcalc: MOSFIRE Exposure Time Calculator v2.3

Southern African Large Telescope. RSS CCD Geometry

Information for users of the SOAR Goodman Spectrograph Multi-Object Slit (MOS) mode. César Briceño and Sean Points

Guide to Processing Spectra Using the BASS Software

Presented by Jerry Hubbell Lake of the Woods Observatory (MPC I24) President, Rappahannock Astronomy Club

Sabeeh Irfan Ahmad, Physlab, 23 July 2016

Improving the Collection Efficiency of Raman Scattering

"Internet Telescope" Performance Requirements

Phoenix High-Resolution Infrared Spectrograph

Observational Astronomy ASTR 2401 Texas Tech University OBSERVING MANUAL

Phoenix High-Resolution Infrared Spectrograph

Make Your Own Digital Spectrometer With Diffraction Grating

Phase-2 Preparation Tool

UNIVERSITY OF HAWAII Institute for Astronomy. f/31 High Angular Resolution Imaging Spectrograph HARIS USER MANUAL update June 10, 1997

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

We recommend downloading the latest core installer for our software from our website. This can be found at:

MS260i 1/4 M IMAGING SPECTROGRAPHS

SpectraPro 2150 Monochromators and Spectrographs

A Stony Brook Student s Guide to Using CCDSoft By Stephanie Zajac Last Updated: 3 February 2012

Instruction Manual for HyperScan Spectrometer

Padova and Asiago Observatories

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Oriel MS260i TM 1/4 m Imaging Spectrograph

Guide to the Imaging Observation with MOIRCS

UV/Optical/IR Astronomy Part 2: Spectroscopy

WITec Alpha 300R Quick Operation Summary October 2018

University of Wisconsin Chemistry 524 Spectroscopic Components *

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE

arxiv: v1 [astro-ph.im] 26 Mar 2012

Image Slicer for the Subaru Telescope High Dispersion Spectrograph

FocusMax V4 Tutorials

INTRODUCTION TO CCD IMAGING

NIRSPEC Data Reduction Pipeline Data Products Specification

CCD User s Guide SBIG ST7E CCD camera and Macintosh ibook control computer with Meade flip mirror assembly mounted on LX200

TIRCAM2 (TIFR Near Infrared Imaging Camera - 3.6m Devasthal Optical Telescope (DOT)

CCD reductions techniques

SPECTRAL SCANNER. Recycling

UltraGraph Optics Design

Descriptions for Each Test

Nonlinearity in the Detector used in the Subaru Telescope High Dispersion Spectrograph

Observing a colour and a spectrum of light mixed by a digital projector

WFC3 Thermal Vacuum Testing: UVIS Broadband Flat Fields

Olivier Thizy François Cochard

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Instrument Acceptance Testing Plan

Southern African Large Telescope. Prime Focus Imaging Spectrograph. Grating and Filter Specification Document

DESIGN NOTE: DIFFRACTION EFFECTS

TriVista. Universal Raman Solution

Application Note (A11)

LACERTA M-GEN Stand-Alone AutoGuider

Miniature Spectrometer Technical specifications

STIS CCD Saturation Effects

Demetra User documentation DC0029A

VS7550 VUV/UV Mini Spectrograph Operating Manual

Observational Astronomy

Point Calibration. July 3, 2012

NICMASS/COUDE FEED OPERATING MANUAL

RENISHAW RAMAN MICROSCOPE STANDARD OPERATING PROCEDURE

SpeX Observing Manual

Astronomy and Image Processing. Many thanks to Professor Kate Whitaker in the physics department for her help

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

HR2000+ Spectrometer. User-Configured for Flexibility. now with. Spectrometers

What is CCD Commander?

LEICA TCS SP5 AOBS TANDEM USER MANUAL

F/48 Slit Spectroscopy

ARRAY CONTROLLER REQUIREMENTS

IBIL setup operation manual for SynerJY software version

Photometry. Variable Star Photometry

The only simultaneous absorbance and f uorescence system for water quality analysis! Aqualog

Confocal Raman Microscopy (WITec Alpha 300R)

CCD Commander. Automation of CCD Imaging. ...a User s Perspective. by Mike Sherick

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

APO TripleSpecTool User s Guide

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

The 0.84 m Telescope OAN/SPM - BC, Mexico

Cousins/Bessell vs. Johnson Filter Standards

NATIONAL OPTICAL ASTRONOMY OBSERVATORY. WIYN Consortium. WTTM User Manual

Image Processing Tutorial Basic Concepts

Technical Note How to Compensate Lateral Chromatic Aberration

You, too, can make useful and beautiful astronomical images at Mees: Lesson 1

SpeX Observing Manual

The DECam System: Technical Characteristics

First steps with a Slit Spectroscope

Spectrophotometric calibration system for DECam

Specifications. Offers the best spatial resolution for multi-stripe spectroscopy. Provides the user the choice of either high accuracy slit mechanism

Infinity Software Manual

Transcription:

VATTSpec Instructions Rev. 10/23/2015 Introduction VATTSpec is a medium resolution CCD range spectrograph with a skinny chip having excellent cosmetics. Its UA ITL chip, Serial Number 8228, has a gain = 1.3 e/dn with setting 10, noise = 3.4 electrons, and the full well is 65k DN. Wavelength range: 360 950 nm; with peak efficiency at 600 nm. Gratings available: Grating Resolution R Spectral Blaze wavelength Coverage 300 g/mm 1500 400 nm 500 nm 600 g/mm 3000 200 nm 400 nm 600 g/mm 3000 200 nm 750 nm 1200 g/mm 6000 100 nm 500 nm 1200 g/mm 6000 100 nm 750 nm The resolutions are for the 1 arcsec slit at 600nm. Slits of 1.5 and 2 arcsec are also available. The slit length is 30 arcsec. Bandpass filters: S8612 (red blocking); ZUL0350, GG-400, GG-475, OG-550 (cut-offs to the blue in nm). Sensitivity: curve is for efficiency of both telescope and spectrograph. Spectrograph efficiency alone is about 5 % higher throughout the range. This determination is courtesy of Mark Wagner and used the 300 g/mm grating.

Setup for AzCam and VATT Spectrograph GUI Connect to instrument computer, and log in as ʻazcamʼ Open AzCam and the VATTSpec GUI NB reload AzCam each day. With TestImage checked, it writes spectrum to test.fits, overwriting previous image Monitor temperature here (-110 for camera, -190 for dewer). Dewer lasts ~10 hours, and longer in winter. Focus value Grating value - to select central wavelength of spectra see Appendix A Open/close cover (close at end of night) Slit width (narrow =1 arcsec, medium=1.5 arcsec, wide=2 arcsec) Calibration lamps and flat field lamp (Qu) Create a directory for your images using the IRAF terminal within (/mnt/tbarray/images)

Click on ʻFilenameʼ in AzCam, enter the new directory name, the root name, and the sequence start number. Click on ʻDetectorʼ in AzCam, from here the binning and the specific region of the detector read out can be changed, and the level of binning in each direction. The slit length, with row binning equal to 1, corresponds to only 90 rows. To allow for the current small shift in the grating perpendicular to dispersion as elevation changes, 10 rows either side of the image of the slit on the detector should be added to make a total of 110 rows. The actual First Row and Last Row will depend on the grating used. Click on Preferences in AzCam. Uncheck Enable Instrument for the present, check Enable Telescope, Clear Array, Display image, and others as desired.

On the IRAF terminal, within the DS9 window, ensure the X axis is inverted (in top line Zoom menu). This places the blue end of the spectrum on the left. On the IRAF terminal, open an xgterm widow. Navigate to the home directory (~), and issue cl to start IRAF. Then cd into the directory which contains your images. Issue imexam to gain access to the various analysis procedures (e.g., j to measure the FWHM of a calibration lamp line, or k to measure the FWHM of the spectrum), or use implot on a particular file (i.e. implot vatt.0001), to plot particular rows/columns etc. Implot command :x 2680 10 inverts X axis, when plotting line(s), and so puts blue to the left of plot. NB. (1) resizing Implot window is only possible after doing a q. (2) reload imexam (after q ) for each new image. Initializing Coordinates, and Acquiring an Object To see an object in the guide camera, remember to open the slit cover! And put in the U-Mirror! Focusing VATTSpec and the telescope Replace the U-mirror with the Center mirror on the guider GUI (do not use calibration mirror Calib., since the mirror stage gets stuck there). Turn on the Mercury and/or Neon lamp, and allow it a minute to warm up. NB: Mercury and Neon warm up twice as fast as Argon, and you may want to delay their turn-on when taking wavelength calibration exposures with all three lamps. Take a ~10 second exposure, and measure the FWHM of a particular line. Change the spectrograph focus on the VATTSpec GUI in steps of ~100, and take another exposure. Measure the FWHM of the same line as before. Search for the focus value which minimizes the width. For the narrow slit, this value is near 2.0 (pixels). Turn off the lamps by clicking on the ʻMeʼ and/or ʻNeʼ buttons. [The above steps can be done in the afternoon before observing, with due allowance of the best focus for the expected drop in temperature during the night.]

Replace the center mirror with the U-mirror on the guider GUI. The dome floodlights can be turned on, and the guider exposure time increased for the slit to be visible on the guider camera, or earlier with the center mirror in place, turn on the Me lamp. Place the bright star slightly above the slit using the paddles.. Use an exposure time of 100 sec or so, and once Exposing, track the star down across the slit at a rate of 1 /sec until it appears below the slit. Then push Pause and Readout. (This saves any need to keep the star perfectly centered to get exact repeat exposures.) Use implot to measure a particular column of the spectrum, for example column 1300 (command :c 1300, and use p on either side of the base of the profile to find its FWHM.).The goal of telescope focusing is to make the spectrum as narrow as possible in the direction perpendicular to the dispersion. Adjust the telescope focus at the Guider GUI and take another exposure. Plot the same column again, and adjust the focus until the spectrum is as narrow as possible, or equivalently, the peak is as high as possible. Guiding, and taking science data Slew to the first target using the XEPhem GUI. Once the target is acquired and it has been maneuvered into the slit using the paddle, the autoguider can be started in either on-slit or field star mode. As from April 2015, see the PHD2 Guiding Manual, with VattSpec specific Notes. Chris Corbally & Rob DeRosa

Appendix A: Setting Grating Value for a Central Wavelength (approximate guide only) Note: the 1354 pixel position is for when the data is not binned; it would be 1354/2 for 2x binned data.