EnSite Precision Cardiac Mapping System ACCURACY OF THE ENSITE NAVX NAVIGATION AND VISUALIZATION TECHNOLOGY, SENSOR ENABLED

Similar documents
Sensor Array. Multi-Electrode Mapping Vest. Technical Manual

Initial setup and subsequent temporal position monitoring using implanted RF transponders

RD SET. Masimo Sensors. Maximizing Patient Comfort Optimizing Clinician Workflow Helping Hospitals Meet Their Green Initiatives

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

SJM MRI Activator. Handheld Device. User's Manual. Model EX4000

RD SET. Masimo Sensors. Maximizing Patient Comfort Optimizing Clinician Workflow Helping Hospitals Meet Their Green Initiatives

Electrocardiogram (ECG)

RD SET. Masimo Sensors. Maximizing Patient Comfort Optimizing Clinician Workflow Helping Hospitals Meet Their Green Initiatives

Potential Risks of MRI in Device Patients

MRI with the VNS Therapy System October 2017

RD SET. Masimo Sensors. Maximising Patient Comfort Optimising Clinician Workflow Helping Hospitals Meet Their Green Initiatives

Specification for Radiated susceptibility Test

EE 230 Experiment 10 ECG Measurements Spring 2010

Exercise 5 EnSite NavX OneMap. Objectives

EG medlab. Three Lead ECG OEM board. Version Technical Manual. Medlab GmbH Three Lead ECG OEM Module EG01010 User Manual

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Image guidance reinvented. Foresight. Intracardiac Echo System

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes

VivoSense. User Manual - Equivital Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248)

2015 Spin echoes and projection imaging

ME6000 Sensor guide for MegaWin

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

POWERFULLY SIMPLE SEEQ. Mobile Cardiac Telemetry System SHORT-TERM CARDIAC MONITORING FOR UP TO 30 DAYS 95 % PATIENT SATISFACTION 1

Technical Explanation for Displacement Sensors and Measurement Sensors

4GHz / 6GHz Radiation Measurement System

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

Antennas and Propagation for Body-Centric Wireless Communications

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Application Note 175 Using the STMISOC Stimulus Isolator

EMC standards. Presented by: Karim Loukil & Kaïs Siala

BENG 186B Winter 2012 Quiz 3. March 7, NAME (Last, First): This quiz is closed book and closed note. You may use a calculator for algebra.

ENGLISH. Spektikor disposable ECG-indicator

Control and confidence all around. Philips EP cockpit people focused solutions for heart rhythm care

Procedure, Field Replacement, PCU Kit, 6003A/6004, 2406 & 4003A

CyberKnife Iris Beam QA using Fluence Divergence

Lab E5: Filters and Complex Impedance

Kennedy Thorndike on a small satellite in low earth orbit

Health and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015

Specification of APS Corrector Magnet Power Supplies from Closed Orbit Feedback Considerations.

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

RF/High Frequency Meter HF-A3G

from signals to sources asa-lab turnkey solution for ERP research

DEVELOPMENT OF CAPACITIVE LINEAR-CUT BEAM POSITION MONITOR FOR HEAVY-ION SYNCHROTRON OF KHIMA PROJECT

Central Blood Pressure Meter Model cbp301. Operating Manual

troubleshooting LEVEL 1 training

ECSEL Smart Health Project Portfolio and Strategic Agenda. Patrick Vandenberghe Programme Officer

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

Lecture 4 Biopotential Amplifiers

AD8232 EVALUATION BOARD DOCUMENTATION

A Design Of Simple And Low Cost Heart Rate Monitor

GPS-Aided INS Datasheet Rev. 2.6

NOW. Approved for NTIOL classification from CMS Available in Quar ter Diopter Powers. Accommodating. Aberration Free. Aspheric.

ECG & Respiration Transmitter LX Operation Manual

PHYSIOFLOW Q-LINK TM

ME7220A. Radar Test System (RTS) Target Simulation & Signal Analysis for Automotive Radar Exceptional Performance at an Affordable Price.

Standard Configuration

v tome x m microfocus CT

RAISING THE LEVEL OF PROTECTION IN YOUR ORS

Accurate Utility Depth Measurements Using the Spar 300

Existing and Emerging Opportunities in Printed Electronics For Printers

Integrated Dual-Axis Gyro IDG-1004

Test and Measurement for EMC

EXPERIMENT 8 Bio-Electric Measurements

Physics Spring 2006 Experiment 9 TRAVELING WAVES

Delivering Modulation Solutions

ABSTRACT 1 CEBAF UPGRADE CAVITY/CRYOMODULE

OT 528: OCCUPATION-BASED TREATMENT KIT

Validation / Calibration Incyte. Quick Guide

Dosimetry and Position Sensing Ionization Chamber for Ion Beam Tracking

ULP Wireless Technology for Biosensors and Energy Harvesting

Material Parameter Measurement (MPM)

Analysis of static and time-variable magnetic fields

Fax Japan São Paulo (SP) Tokyo Minato-ku , Minami-Aoyama. St. Jude Medical Brasil Ltda.

Kit for building your own THz Time-Domain Spectrometer

MR Advance Techniques. Flow Phenomena. Class II

ALIGNMENT METHODS APPLIED TO THE LEP MAGNET MEASUREMENTS. J. Billan, G. Brun, K. N. Henrichsen, P. Legrand, 0. Pagano, P. Rohmig and L. Walckiers.

Fundamentals of Electromagnetics With Engineering Applications by Stuart M. Wentworth Copyright 2005 by John Wiley & Sons. All rights reserved.

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

Field Service Spares Procedure Replacement Coax Switch Kit, 8897B, 9497B, 14400B, ST88, ST94 & ST144

The evolution of Medical implant telemetry and Body Area Network

D C 01/2019 3

Wireless In Vivo Communications and Networking

Submersion/Insertion Toroidal Sensor

GPS-Aided INS Datasheet Rev. 2.3

Paceart Optima System

Physician s Manual. June VNS Therapy Programming Software Model 250 Version 11.0 Programming Computer. For Healthcare Professionals.

Lab 3 Swinging pendulum experiment

BARTINGTON INSTRUMENTS. How to use this Manual... 3 Symbols Glossary Introduction to the Mag Vector Measurements and Conventions...

EXHIBIT 7: MEASUREMENT PROCEDURES Pursuant 47 CFR 2.947

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Your Reliable and Competent Partner for Complex Sensor Systems

Exercise Heart Rate Sensor

Fundamentals of Pacing Therapy

BENNING. CM7 Electrical Tester. Service Information

Bio-Potential Amplifiers

The Fastest, Easiest, Most Accurate Way To Compare Parts To Their CAD Data

System description 4. SERVICES ONSITE INSTALLATION AND TRAINING SYSTEM ACCEPTANCE MAINTENANCE... 7

Phantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging

Transcription:

EnSite Precision Cardiac Mapping System ACCURACY OF THE ENSITE NAVX NAVIGATION AND VISUALIZATION TECHNOLOGY, SENSOR ENABLED

all trodes IN using a or using rep) that gritty l the gel he skin. ith soap as been ning with skin is ith lotion. that all fore es. Note: r of the circular ontact. PRINCIPLE OF OPERATION ION l to all ng to he noise nt during ures. The EnSite Precision System is a 3-D cardiac mapping system that uniquely combines impedance and magnetic field technology to enable precise navigation and accurate 1 tracking of conventional and Abbott Sensor Enabled electrophysiology catheters during the creation of three-dimensional maps based on the anatomy of the cardiac chamber. An EnSite Precision Cardiac Mapping System procedure using EnSite NavX Navigation and Visualization Technology, Sensor Enabled is performed using hardware components and disposable products. Hardware The EnSite Precision field frame universally mounts underneath the patient bed and generates a low-powered magnetic field within which the position of a Sensor Enabled device can be detected. The EnSite Precision link, Sensor Enabled connects Sensor Enabled tools, including catheters, patient reference sensors (PRSs) and the field frame and relays the information to the EnSite amplifier via fiber-optic connection. System Reference Electrode Defibrillator Patch Disposables Patient Reference Sensor FIGURE 1: Surface electrodes placed in three transthoracic pairs and two patient reference sensors, along with other surface electrodes, for an EnSite RF Dispersive NavX Patch Navigation and Visualization Technology, Sensor Enabled CAUTION: The system reference surface electrode must be the first patient electrode connected to the EnSite study. Amplifier at the beginning of a study, and the last patient electrode to be disconnected at the end of the study. The EnSite Precision surface electrode kit consists of a system reference surface electrode and six surface electrodes that are placed on the patient in pairs: anterior to posterior, left to right lateral, and superior (neck) to inferior (leg) (Figure 1). The three electrode pairs form three orthogonal axes (X-Y-Z), with the heart at the center. Two PRSs are also used with the EnSite Precision System: a PRS anterior, and a PRS posterior. The PRSs are connected to the patient using disposable, self-adhesive patches, and primarily function as sensors for metal distortion and patient movement. During model collection, both impedance-based points and magnetic-based points are collected from a Sensor Enabled tool. EnSite Precision Surface Electrode re ECG e placed : n abrader System Reference Electrode Defibrillator Patch EnSite Precision Surface Electrode Patient Reference Sensor RF Dispersive Patch CAUTION: The system reference surface electrode must be the first patient electrode connected to the EnSite Amplifier at the beginning of a study, and the last patient electrode to be disconnected at the end of the study.

Impedance Data Catheter location and navigation of all compatible tools, both conventional and Sensor Enabled, is based on the impedance field generated by the EnSite surface electrodes. When the surface electrodes are connected to the EnSite Precision System, an 8 khz signal is sent alternately through each pair of surface electrodes to create a voltage gradient along each axis, forming a transthoracic electrical field. Conventional or Sensor Enabled electrophysiology catheters are connected to the EnSite Precision System and advanced to the heart. As a catheter enters the transthoracic field, each catheter electrode senses voltage, timed to the creation of the gradient along each axis (Figure 2). Using the sensed voltages compared to the voltage gradient on all three axes, the EnSite Precision System calculates the three-dimensional position of each catheter electrode for all electrodes simultaneously. FIGURE 3: The EnSite Precision System displays the catheter bodies in real time FIGURE 2: Each intracardiac electrode measures voltage along the gradient of each axis The EnSite Precision System displays the located electrodes as catheter bodies with real-time navigation. It permits the simultaneous display of multiple catheter electrodes (Figure 3) and also reflects real-time motion of ablation and diagnostic catheters in the heart. 2 By tracking the position of the catheters, the system enables the creation of 3-D electroanatomical models of the cardiac chambers. Magnetic Data When a Sensor Enabled catheter is introduced and EnSite NavX Navigation and Visualization Technology, Sensor Enabled Field Scaling is applied, the EnSite Precision System dynamically optimizes the model by adjusting the dimensions of the navigation field using known offsets between the position and orientation of magnetic sensor(s) and electrodes. The EnSite Precision System also uses magnetic information as an input for the EnGuide stability monitor to monitor field stability for unexpected changes. This feature can be enabled and used to monitor the location of a Sensor Enabled tool real time within an EnSite NavX Navigation and Visualization Technology, Sensor Enabled field scaled mode.

ACCURACY TESTING AND VERIFICATION 3 Performance bench testing was conducted to verify the tracking and navigation accuracy of the magnetic navigation and visualization technology utilized in the EnSite Precision System. Accuracy Dynamic Wet Lab Study This study tested the tracking accuracy and the navigation accuracy of the EnSite NavX Sensor Enabled technology in the EnSite Precision module with the FlexAbility ablation catheter, Sensor Enabled and the Advisor FL circular mapping catheter, Sensor Enabled. Testing was performed using a calibrated robotic dynamic wet lab (DWL) to create a 10 cm 3-D geometric cube phantom model using sensor enabled catheter(s) held in an orthogonal fixture (Figure 4). Each catheter movement between 14 horizontal and 10 vertical displacement locations provided multiple electrode measurements during model creation (Figure 5a). Cube models were created using EnSite NavX Navigation and Visualization Technology, Sensor Enabled field scaling in static mode, and in dynamic mode with simulated cardiac (30 120 BPM) and respiration (10 12 RPM) motion enabled. System error was calculated based on the differences in measurements between the cube models created by the EnSite Precision System and the known dimensions of the phantom cube geometry (Figures 5b and 6). FIGURES 5A (LEFT) AND 5B (RIGHT): DWL a. cube displacement levels and b. measurements of Tracking Accuracy (from position A to position B) and Navigation Accuracy (return to position A) FIGURES 6: EnSite Precision System display during DWL cube model creation, with automatic EnSite NavX Navigation and Visualization Technology, Sensor Enabled magnetic field scaling enabled FIGURE 4: DWL fixture showing two FlexAbility ablation catheters, Sensor Enabled and one Advisor FL circular mapping catheter, Sensor Enabled Tracking Accuracy was measured as the error between an induced catheter displacement and the corresponding measured displacement, in this case as catheters were displaced by 50 mm from Position A to B. The specified maximum allowable Tracking Accuracy error in a field scaled model is 10%, or 5.0 mm in this test scenario. Navigation Accuracy was measured as the error in navigating a catheter back to a defined location (i.e., as catheters were then returned to Position A). The specified maximum catheter Navigation Accuracy error is 2.0 mm with a field scaled model. Mean tracking accuracy error was under 0.5 mm and maximum tracking accuracy error was well under the specified maximum of 10% over the 50-mm tested displacement value in both static and dynamic test modes. Mean navigation accuracy was within 0.4 mm for all catheter electrodes tested using the DWL system, with a maximum error of 0.9 mm, also well below the specified maximum value of 2.0 mm (see Table 1 for details).

ACCURACY TEST RESULTS 2 Table 1: Accuracy of EnSite Precision System (v2.0.1) with EnSite NavX Navigation and Visualization Technology, Sensor Enabled Magnetic Field Scaling SENSOR ENABLED CATHETERS TESTED TRACKING ACCURACY ERROR (NO. SAMPLES) NAVIGATION ACCURACY ERROR (NO. SAMPLES) ADVISOR FL CIRCULAR MAPPING CATHETER, SENSOR ENABLED AND FLEXABILITY CATHETER, SENSOR ENABLED (DWL PHANTOM MODEL) Dynamic: Mean: 0.44 ± 0.45 mm Max: 2.59%, Min: 4.11% (n = 4032) Static: Mean: 0.45 ± 0.49 mm Max: 2.80%, Min: 4.37% (n = 4032) Dynamic: Mean: 0.18 ± 0.07 mm Max: 0.38 mm (n = 2688) Static: Mean: 0.34 ± 0.16 mm Max: 0.92 mm (n = 2688) SUMMARY Tracking and navigation accuracy are important in advanced cardiac mapping and ablation procedures, such as when it becomes necessary to guide a second ablation catheter to the same location as the first. 4 The accuracy of the EnSite NavX navigation and visualization technology, Sensor Enabled in the EnSite Precision System has been verified using dynamic bench testing to provide reliable catheter navigation and localization in electrophysiology procedures.

REFERENCES 1. Abbott. Data on File. Report 90212729. 2. Packer, D. L. (2005). Three-dimensional mapping in interventional electrophysiology: techniques and technology. Journal of Cardiovascular Electrophysiology, 16(10), 1110-1116. 3. Abbott. Data on file. Report 90237452. 4. Wittkampf, F. M., Wever, E. F., Derksen, R., Wilde, A. A., Ramanna, H., Hauer, R. N., & Robles de Medina EO. (1999). LocaLisa: new technique for real-time 3-dimensional localization of regular intracardiac electrodes. Circulation, 99, 1312-1317.

Abbott One St. Jude Medical Dr., St. Paul, MN 55117 USA, Tel: 1 651 756 2000 SJM.com St. Jude Medical is now Abbott. Rx Only Brief Summary: Prior to using these devices, please review the Instructions for Use for a complete listing of indications, contraindications, warnings, precautions, potential adverse events and directions for use. Indicates a trademark of the Abbott group of companies. Indicates a third party trademark, which is property of its respective owner. 2018 Abbott. All Rights Reserved. 26076-SJM-ENS-0916-0097a(2) Item approved for global use.