EMS PRODUCT INFORMATION: PORTABLE NOISE MONITORING TERMINAL TYPE 3655-C-DMO with Noise Sentinel On Demand Weather Station Kit MM-0256-A-DMO

Similar documents
PRODUCT DATA. Sound Intensity Calibrator Type Uses and Features

PRODUCT DATA. Applications. Uses

PRODUCT DATA. Noise Monitoring Terminal Plus Type 3639-E Noise Monitoring Terminal Advanced Type 3639-G. Uses and Features

PRODUCT DATA. Noise Monitoring Terminals Types 3597 C, 3637 A/B and 3631

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 including Sound Intensity Microphone Pair Type 4197 and Dual Preamplifier Type 2683

PRODUCT DATA. Noise Monitoring Terminal Plus Type 3639-E Noise Monitoring Terminal Advanced Type 3639-G

PRODUCT DATA USES. BENEFITS Normal incidence parameters are determined Fast and accurate measurements. Type 4206A. Type Type 4206T 50 Hz 1.

PRODUCT DATA. Noise Monitoring Terminal Types 3639-A and 3639-B

PRODUCT DATA. Sound Intensity Calibrator Type 4297

PRODUCT DATA. Falcon Range 1/4-inch Microphone Preamplifier Type 2670

PRODUCT DATA. Sound Intensity Probe Kit Type 3599 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683

Technical Documentation

Environmental Noise Measurement Kits

PRODUCT DATA. Sound Intensity Probe Kit for 2260 Type 3595 Sound Intensity Microphone Pair Type 4197 Dual Preamplifier Type 2683

C1.1 Specification of the SVAN 971 as SLM in the standard configuration

PRODUCT DATA. 16-channel DeltaTron Conditioning Amplifiers Types 2694-A/B/C/D. Features and Uses

PRODUCT DATA. NEXUS Conditioning Amplifiers for Very High Input Types 2692-C and 2692-D

2260 INVESTIGATOR MODULAR PRECISION SOUND ANALYZER WITH APPLICATION SOFTWARE BZ 7210, BZ 7203, BZ 7204, BZ 7205, BZ 7206, BZ 7207, BZ 7208

PRODUCT DATA. 3-ch. Input Module LAN-XI khz Type Uses and Features

PRODUCT DATA. Uses and Features

Precision Integrating Sound Level Meter

CAN Bus Module LAN-XI 25.6 khz Type 3058

CAN Bus Module LAN-XI 25.6 khz Type 3058

PRODUCT DATA. The NEXUS Range of Conditioning Amplifiers Types 2690, 2691, 2692 and Uses And Features

Fluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease

CR:247 Invictus. Portable Noise Monitor. Features. Features. Applications

SLM-52N INSTRUCTION MANUAL

NoiseMeters. The Practical & Easy To Use Solution for Industrial & Environmental Noise Measurements.

R&S ZVT Vector Network Analyzer Specifications

FM Modulation. Accuracy. Total. Harmonics. Input Range. Frequency. Range. Deviation. Range. Modulation. Accuracy. Audio In Switchable Loads.

CEL63x Sound Level Meter CEL630, CEL631, CEL632, CEL633 Class 1 or Class 2. Integrating. Advanced Logging. Audio Recording

Portable Multi-Channel Recorder Model DAS240-BAT

Fluke 1740 Series. Three-Phase Power Quality Loggers Memobox. Assess power quality and conduct long-term studies with ease

Submersible pressure sensor For water and wastewater Model LW-1

2302 Battery Simulator 2306, 2306-PJ Battery/Charger Simulators

NI PXIe-5601 Specifications

PRODUCT DATA. PULSE Multichannel Sound Power Type 7748 Version 1.2

Key Features. Base-line surveys for planning Construction site monitoring Road traffic noise monitoring

Microphones & Accessories

SAS-562B Active Loop Antenna Operation Manual

Model Microphone Preamplifier Power Supply Manual

Environmental Noise Management Sofia December Brüel & Kjær S&V Torben Munk

PM 6669 High-Precision Frequency Counter Specifications

SC101. Class 1 integrating sound level meter with measurement. Applications Includes measurement protocols for: Noise generated by motor vehicles

Advanced Test Equipment Rentals ATEC (2832)

Submersible pressure sensor For oils and fuels Model LF-1

Testing to IEC

OtoRead - Technical Specifications Page 0. Technical Specifications. OtoRead D A 2017/06

PRODUCT DATA. PULSE Vibration Check System for Aircraft Engines Type 3641 including PULSE Standard Configuration 3560-B-T57

02/11/2015

R&S ZVT Vector Network Analyzer Specifications

SoundExpert LxT Sound Level Meter:

P R O D U C T D A T A

SoundExpert TM LxT. Display Navigation. Dual Purpose Start/Stop Reset/Clear Memory. Recessed On/Off Button. USB Host (Thumb Drive Storage)

PRODUCT DATA. Piezoelectric Charge Accelerometer Types 4326-A and 4326-A-001. Uses and Features

PRODUCT DATA. DC Response Accelerometers Types 4570, 4571, 4572, 4573, 4574, and 4575

2. Measurement Range / Further specifications of the LOG_aLevel system

Continuously monitors and stores the levels of Electromagnetic fields Up to four simultaneous bands: GSM 900 / 1800 MHz / UMTS / Broadband 100 khz 3

METRISO. Analog Insulation, Low Resistance and Voltage Measurement Instrument /9.14

SPECIFICATION SWITCHING POWER SUPPLY SINGLE POWER 250W BEST POWER SOLUTIONS, INC. 9F, NO.196-7, SEC.3, TA-TUNG RD SHI CHIH, TAIPEI, TAIWAN, R.O.C.

ARCO 400. Technical Data

Accutech AM20. Wireless acoustic monitor field unit

SPECIFICATION SWITCHING POWER SUPPLY BPS-300SP/2U 6-OUTPUT WITH ACTIVE PFC FUNCTION SINGLE POWER 300W BEST POWER SOLUTIONS, INC.

SAS-563B Active Loop Antenna Operation Manual

PRODUCT DATA. Type channel Human Vibration Front-end Type 1700

SPECIFICATION REDUNDANT POWER SUPPLY HOT-SWAPPABLE BPS-460RDX 6-OUTPUT 460W+460W BEST POWER SOLUTIONS, INC.

GRAS 40BD-FV 1/4" Prepolarized Pressure Microphone, Front Vented

SPECIFICATION REDUNDANT POWER SUPPLY HOT-SWAPPABLE 370W+370W

JD723A/JD724B/JD726A Cable and Antenna Analyzers

Tender NO. PGVCL/PROC/Cable & Pipeline Locator/832

Appendix A: Specifications

SoundExpert TM LxT. n Site Assessment Run/Pause Control. Display Navigation. Dual Purpose Start/Stop Reset/Clear Memory. Recessed On/Off Button

SPAC265-8W. AC-DC power supply module. Features. Description. Applications

High Sensitivity Triaxial CCLD Accelerometer Type 4506-B-003

Sound Level Meter C.A 832

Submersible pressure sensor For superior applications Model LF-1

02/11/2015

Wireless Acoustic Transmitters Series XYR 5000, Model WN571 Specifications - Americas

02/11/2015

Instrument. Specifications OR34/35. SmartRouter. 2 to 8 Channels Multi-Analyzers. Controller Unit TECHNICAL SPECIFICATIONS

WIRELESS INSULATOR POLLUTION MONITORING SYSTEM

Precipitation Monitor ,

Battery Simulator Battery/Charger Simulators

Protocol for Ambient Level Noise Monitoring

vscan Data Sheet V1.0

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one.

USER MANUAL. Hand-held Analyzer Types 2250 and 2270

Product Data Sheet 3252J/2H3PU

Custom "Compact" range with display CD20 Custom Part number

PART NUMBERING SYSTEM

Vector Network Analyzers ZVB

Agilent U1273A/U1273AX Handheld Digital Multimeter. Quick Start Guide

Technical Documentation

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

Fluke 3540 FC Three-Phase Power Monitor

HMC-T2100B. Portable & Battery Operated! OBSOLETE PRODUCT. Portable Synthesized Signal Generator, 10 MHz to 20 GHz

2.1 Performance Standards The UPS is designed with the applicable sections of UL, CUL, and ISO The UPS has UL and CUL listing.

Conductivity Transmitter

Fluke Three-Phase Power Quality Recorder Topas. Class-A compliance for the most demanding power quality tests

Product Data Sheet 3254 J/2H3P

Transcription:

PRODUCT INFORMATION: PORTABLE NOISE MONITORING TERMINAL TYPE 3655-C-DMO with Noise Sentinel On Demand Weather Station Kit MM-0256-A-DMO Portable Noise Monitoring Terminal (NMT) Type 3655-C-DMO is a complete solution delivered as part of Noise Sentinel On Demand. This enables you to monitor and manage noise through an end-to-end online experience: book and obtain instrumentation, collect and report data and maintain data indefinitely in the cloud. Portable NMT Type 3655-C-DMO comes complete with tripod, router, SIM card and airtime contract, antenna, sound level calibrator and all accessories required for deployment. In addition, a smart phone App eases setup and configuration, including easy location based on GPS, as well as entering location notes and photographs. The NMT is delivered in two transport cases, one for deployment and one for the accessories. The NMT can operate off its own internal batteries for 32 hours (typical) with the router. In addition, it can operate off AC power or with external DC power. Fig. 1 Portable NMT in use on a construction site With Noise Sentinel On Demand Weather Station Kit MM- 0256-A-DMO you can add a weather station that collects weather data simultaneously with your noise data. This kit is delivered in a single transport case that contains everything you need to set up the weather station. USE AND FEATURES Uses Measurements made: - Outdoors - Unattended - In workplaces and on construction sites - In remote locations Measurements for: - Area planning - Noise control - Complaint investigation - Venue licensing Simultaneous measurement of noise and weather data Features Weather protection to IP 43 Easy to carry and transport Tamper protection All accessories needed to deploy and use from analyzer to router with pre-paid SIM card to sound level calibrator Hot swap of batteries Type approved to IEC 61672 Class 1 specifications; uniquely, including windscreen effects 120 db dynamic range Built-in facilities to minimise gaps in data Charge Injection Calibration (CIC) for remote verification of the entire measurement chain Safe and reliable live data streaming 3G communication capabilities for remote operation Industry-standard Internet and security protocols for safe and reliable data transfer PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 01

INSIDE THE CASE 1. Foam support for batteries and analyzer 2. Built-in external antenna 3. Rain shield 4. Cable channel 5. Router with velcro straps 6. Cable strain relief posts 7. Cable entry exit (ext. microphone) 8. Two batteries 9. Hand-held analyzer (2250-L) 10. Warning label 11. Power panel 12. DC out on/off switch 13. Foam support 14. Two eyes for chain and lock 15. Two chargers with straps 16. Sound calibrator (4231) 17. Accesories: Microphone, cable, cable lock, etc. 18. Two eyes for lock 19. User manual 20. Tripod 21. External microphone (4952) 22. Tripod mounting adaptor 23. Mains power cable Fig. 2 The contents of the Portable NMT Type 3655-C-DMO PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 02

A Complete Noise Monitoring Solution Portable Noise Monitoring Terminal Type 3655-C-DMO comes complete with tripod, router, SIM card and airtime contract, antenna and sound level calibrator and all accessories required for deployment in the field. The NMT is delivered in two transport cases, one for deployment and one containing the accessories. Type 3655-C-DMO includes everything needed to take measurements: NMT with microphone and cable Tripod with guy ropes and pegs 3G router, SIM card and airtime contract Rechargeable 32 hr battery pack Mains power supply Cable for external batteries Security cable and locks Calibrator Assembly toolkit Shipping cases Shipment to and from site Assembly and testing instructions Protection The cases are light and robust, and their contents are secured in the case with high-density machined foam inlays. The cases are designed for unattended noise monitoring. They protect the measurement system from the weather and unauthorised access while providing power and remote data retrieval. Fig. 3 Protective high-density machined foam inlays POWER Type 3655-C-DMO comes with two powerful batteries and two chargers. Power for the system may be provided by any or all or the following: One of the included batteries (QB-0079) Both of the included batteries (QB-0079) One of the included chargers (ZG-0857), provided mains power is available Both of the included chargers (ZG -0857), provided mains power is available External DC power 12 24 V The internal batteries can power the system for 32 hours (typical) with the router. In addition, the hand- held analyzer has an internal Li-Ion battery, that provides a further 8 hours of power for the analyzer, itself. The case s batteries use lithium-ion (Li-Ion) technology, well known from mobile phones, portable PCs and hand tools. These lightweight batteries make Type 3655-C-DMO compact and completely portable. Li-Ion batteries are renowned for their excellent charge retention, lack of memory, and very high energy efficiency (5 times better power to weight ratio than traditional lead-acid batteries). For reliable long-life operation, the batteries have internal circuitry to protect against shorts and over-discharge. The chargers are used to charge the batteries prior to measurements but may also be used to power the system during measurement if mains power is available. The NMT comes with a 1.5 m mains power cable for use with AC power (100 240 V AC, 0.6 A, 50 60 Hz). You can also use an external 12 24 V DC power source that includes a suitable power guard to enable correct and safe restart after external power has been lost. All power sources are connected to the case s power panel, which always directs power from the source with the highest voltage to supply the measurement system and optional modems. You may connect or disconnect power sources at any time during measurement as long as just one power source remains in operation, including during hot-swapping of batteries. PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 03

RELIABLE UNATTENDED MEASUREMENTS Based on Hand-held Analyzer Type 2250-L with Noise Monitoring Software BZ-7232, it functions exactly as Noise Monitoring Terminal Type 3639-A in a portable case rather than a mounted cabinet. The noise monitoring functionality includes CIC checks, optional event detection and sound recording for extended periods of monitoring. The included router enables remote viewing of realtime results and to remotely control the unit from Noise Sentinel On Demand. Data can be automatically downloaded real- time or on connection with the server. Measurement Integrity Measurement integrity is of primary importance in noise measurement situations, whether for inside or outside or for attended or unattended measurements. Class 1, as described in the current sound level meter standard IEC 61672 1:2002, is the grade of accuracy often required for hand held measurements. In addition nothing less than Class 1 accuracy is required for outdoor measurement systems incorporating rain shields, wind shields or any other form of environmental protection used in microphone systems. Placing a small device, such as a rain guard, in close proximity to the microphone diaphragm may produce significant acoustic disturbance and thus measurement errors at mid and high frequencies. Outdoor Microphone Type 4952 The compact and lightweight Outdoor Microphone Type 4952 is suitable for long periods of unattended outdoor operation and the ideal choice for use with the Type 3655-C-DMO system. The microphone is protected against the effects of wind, rain and perching birds and, with the hand-held analyzer, fulfils IEC 61672 Class 1 requirements. The reference direction angle of incidence can be set to 0 or 90, dependent on the noise monitoring application. Inside the microphone is a highly stable pre- polarized free-field 1/2 2 microphone cartridge with a stainless steel diaphragm. Type 4952 is recommended for extended use in all kinds of weather. Up to 100 m of microphone extension cable can be used while maintaining measurement accuracy. You can mount the microphone on the supplied tripod using Tripod Adaptor UA-1707 or on a 1 2 threaded pole. Hand-held Analyzer Type 2250-L and Outdoor Microphone Type 4952 are a type approved system combination, independently approved to Class 1 accuracy by PTB in Germany (see type approval certificate: Fig. 2). This ensures that the measurement system complies with the minimum requirements of accuracy for unattended noise measurement, a consideration often overlooked in portable noise measurement systems. Type approval of the system to IEC 61672 1:2002 also permits the weather-protected measurement system to be laboratory calibrated in accordance with part III of this sound level meter standard. Fig. 4 Outdoor Microphone Type 4952 suitable for long periods of unattended monitoring PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 04

Measuring Weather Data Noise Sentinel On Demand Weather Station Kit MM-0256- A-DMO enables the simultaneous measurement of noise and weather data. This six- parameter station measures wind speed and direction, precipitation, temperature, humidity and pressure. The kit is delivered in a robust transport case that contains everything needed to set up the weather station in the field, including a tripod. Based on ultrasound, the weather stations operate silently, which allows close placement to the microphone position. Weather conditions affect the propagation of sound and therefore any measured noise levels. Particularly, wind speed and direction must be taken into account when measuring noise outdoors. Most environmental noise measurement standards define limits for wind speed and direction. Environmental noise measurements must document weather conditions during the measurement period. s Weather Station Kit MM-0256-A-DMO is based on Vaisala sensors and is designed to fully meet your needs. The weather station is lightweight and connects to the instrument s USB port, eliminating the need for separate batteries. Fig. 5 Noise Sentinel On Demand Weather Station Kit MM-0256-A-DMO COMPLIANCE WITH STANDARDS The CE marking is the manufacturer s declaration that the product meets the requirements of the applicable EU directives RCM mark indicates compliance with applicable ACMA technical standards that is, for telecommunications, radio communications, EMC and EME China RoHS mark indicates compliance with administrative measures on the control of pollution caused by electronic information products according to the Ministry of Information Industries of the People s Republic of China WEEE mark indicates compliance with the EU WEEE Directive Safety EMC Emission EMC Immunity EN/IEC 61010 1, ANSI/UL 61010 1 and CSA C22.2 No.1010.1: Safety requirements for electrical equipment for measurement, control and laboratory use EN/IEC 61000 6 3: Generic emission standard for residential, commercial and light industrial environments EN/IEC 61326: Electrical equipment for measurement, control and laboratory use EMC requirements CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits IEC 61672 1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards Note: The above is only guaranteed using accessories listed in this Product Data EN/IEC 61000 6 2: Generic standard Immunity for industrial environments EN/IEC 61326: Electrical equipment for measurement, control and laboratory use EMC requirements IEC 61672 1, IEC 61260, IEC 60651 and IEC 60804: Instrumentation standards Note: The above is only guaranteed using accessories listed in this Product Data PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 05

Temperature IEC 60068 2 1 & IEC 60068 2 2: Environmental Testing. Cold and Dry Heat. Operating Temperature: 10 to +50 C (14 to 122 F) Storage Temperature: 25 to +70 C ( 13 to 158 F) Humidity IEC 60068 2 78: Damp Heat: 93% RH (non-condensing at +40 C (104 F)). Recovery time 2 ~ 4 hours Mechanical Non-operating: IEC 60068 2 6: Vibration: 0.3 mm, 20 m/s2, 10 500 Hz IEC 60068 2 27: Bump: 1000 bumps at 400 m/s2 IEC 60068 2 27: Shock: 1000 m/s2, 6 directions Enclosure IEC 60529 (1989): Protection provided by enclosures: IP43 Additional Standards for Charger ZG-0857-001* Safety EMC Emission EN 60335 1 (EN 60335 1:1994 +A11:1995, A1:1996, A12:1996) EN 60335 2 29 (EN 60335 2 29:1996 +A11:1997) EN 60601 1 (EN 60601 1:1990 +A1:1993, A2:1995, A12:1993, A13:1996 +Corr.:1994) UL 2601 1, 2.ed:1997 and CAN/CSA C22.2 No.601.1-M90 EN 50081 1:1992 (EN 55022B:1998 +A1:2000, EN 61000 3 2:1995 +A1:1998, A2:1998, A14:2000) EN 50082 1:1997 (EN 61000 4 2:1995 +A1:1998, 4 3:1996, A1:1998, 4 4:1995, 4 5:1995, 4 6:1996, -4-8:1993, 4 11:1994, ENV 50204:1995) EN 60601 1 2:1993 (EN 55011B:1991, IEC 801 2:1991, 3:1988, 4:1988, 5:1994) Additional Standards for Router UL-1041* Safety EN 62311 EN 60950 1 EMC Emission EN 301 511 EN 301 908 1 EN 301 908 2 EN 301 489 1 EN 301 489 7 EN 301 489 24 Additional Standards for Noise Sentinel On Demand Weather Station Kits MM-0256-A and MM-0316-A EMC Emission EN61326 1 IEC 60945/61000 4 2, 4 3, 4 4, 4 5, 4 6 * From the Mascot declaration of conformance PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 06

SPECIFICATIONS PORTABLE NOISE MONITORING TERMINAL TYPE 3655-C-DMO All specifications are valid with Noise Monitoring Software BZ-7232 version 4.3.1 Portable Noise Monitoring Terminal Type 3655-C-DMO is supplied with Outdoor Microphone Type 4952, which includes Microphone Preamplifier ZC-0034. The microphone can only be connected to the analyzer through a microphone extension cable MICROPHONE Type: Prepolarized Outdoor Microphone Nominal Open Circuit Sensitivity: 31.6 mv/pa, (corresponding to 30 db re 1 V/Pa) ±2 db Capacitance: 12 pf (at 250 Hz) Reference Direction: Selectable between 0 (Top) and 90 (Side) MICROPHONE PREAMPLIFIER ZC-0034 Nominal Preamplifier Attenuation: 0.3 db Extension Cable between Microphone Preamplifier ZC-0034 and Analyzer: Up to 100 m without degradation of the specifications MEASURING RANGES (BROADBAND) Dynamic Range: From typical noise floor to max. level for a 1 khz pure tone signal, A-weighted: 20.0 141 db Linear Operating Range: In accordance with IEC 61672, A-weighted: 1 khz: 31.1 141 db Primary Indicator Range: In accordance with IEC 60651, A-weighted: 29.8 124 db Linearity Range: In accordance with IEC 60804, A-weighted: 27.7 141dB MEASURING RANGES (1/3 OCTAVE) Dynamic Range: From typical noise floor to max. level for a pure tone signal at 1 khz 1/3-octave: 2.9 141 db Linear Operating Range: In accordance with IEC 61260: 29.5 139.3 db SELF-GENERATED NOISE LEVEL Typical values at 23 C for nominal microphone opencircuit sensitivity: Weighting Microphone Electrical Total Z (5Hz 20 khz) A 14.0 db 18.7 db 20.0 db B 12.9 db 17.5 db 18.8 db C 13.0 db 18.7 db 19.7 db 14.4 db 24.8 db 25.2 db CORRECTION FILTERS Noise Monitoring Software BZ-7232 is able to correct the frequency response to compensate for sound field and accessories Sound Field: Free field or Diffuse field. Two Free-field reference directions: 0 (Top) and 90 (Side) MICROPHONE POLARIZATION VOLTAGE Selectable between 0 V and 200 V CALIBRATION Initial calibration is stored for comparison with later calibrations Acoustic Calibration: Using Sound Calibrator Type 4231, the calibration process automatically detects the calibration level Automatic checks: Up to 4 times per day with CIC FREQUENCY ANALYSIS 1/1- or 1/3-octave band analysis 1/1-oct. Band Centre Frequencies: 16 Hz 16 khz 1/3-oct. Band Centre Frequencies: 12.5 Hz 20 khz POWER PANEL ZH-0685 Please refer to the Power Panel schematic diagram in Chapter 3 of the Type 3535-A user manual Mains Input: 100 240 VAC, 0.6 A, 50 60 Hz Mains Outputs: AC Out 1, AC Out 2, max. 0.3 A. For Chargers ZG-0857-001 Mains Only Operation: Without batteries the charger(s) will supply the DC Outputs PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 07

Mains Only DC Outputs: 16.8 V, max. 0.9 A DC In: External supply, 12 24 V DC, 1 A. Cable AQ-0175 supplied Charger 1, Charger 2: For Chargers ZG-0857-001 Bat. 1, Bat. 2: For Batteries QB-0079 DC Outputs: DC Out 1: For Type 2250-L Ext. Power DC Out 3, DC Out4, DC Out 5: Power for auxiliary devices. Cables AQ-1782 and AQ-1783 supplied DC Output Voltage: Whichever is higher of the DC In, Bat.1/Charger 1 and Bat.2/Charger 2 DC Output Switch: Switches all DC Outputs on or off CLOCK Back-up battery powered clock. Drift <0.45 seconds per 24-hour period WARM-UP TIME From Power Off: <2 minutes From Standby: <10 seconds TEMPERATURE Operating Temperature: Battery powered: 10 to +50 C (+14 to 122 F) Charger powered: 10 to +40 C (+14 to 104 F) Charge Temperature: 0 to +40 C (+32 to 104 F) with case lid open Storage Temperature: 10 to +60 C (+14 to 140 F) BATTERY QB-0079 Weight: 560 g (1.23 lb) Nominal Voltage: 14.8 V Rated Capacity: 6.3 Ah minimum, 6.6 Ah typical Rated Energy: 89 Wh Expected Life Cycles: >300 cycles at >70% of initial capacity Charge Retention in Storage: 1 year at 20 to +20 C ( 4 to +68 F): >80% 1 month at 20 to +60 C ( 4 to +140 F): > 80% CHARGERS ZG-0857-001 Input Voltage: 90 264 V AC Output Current Max.: 0.9 A Output Voltage Max.: 16.8 V Charge Start: <16.4 V Step 1: Constant current 0.9 A, Lamp: orange Step 2: Constant voltage 16.8 V, Lamp: orange Step 3: Charge Termination <100 ma, Lamp: green Charging Time for QB-0079: 9 hours (typical) TYPE 2250-L OPERATING TIME (BACKLIGHT OFF, 2 BATTERIES) Offline: Approx. 72 hours (typical) With router: Approx. 32 hours (typical) MECHANICAL Environmental Protection: IP 43 WEIGHT AND DIMENSIONS NMT: Weight: 7.5 kg (16.5 lb) including hand-held analyzer, router and packing material for transport Dimensions: 500 400 250 mm (19.69 15.75 9.84 2 ) Accessory Case: Weight: 8.5 kg (18.7 lb) Dimensions: 650 450 250 mm (25.59 17.72 9.84 2 ) Weather Station Case: Weight: 8 kg Dimensions: 650 450 250 mm (25.59 17.72 9.84 2 ) WEATHER DATA (REQUIRES WEATHER STATION): Wind Direction Wind Speed Ambient Temperature Ambient Humidity Ambient Pressure Rainfall PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 08

SPECIFICATIONS FOR SOFTWARE CONTROLLED VIA REMOTE PC All specifications are valid with Noise Monitoring Software BZ-7232 version 4.3.1. Noise Monitoring Terminal Type 3639-A/C can be remote controlled from a PC running Environmental Noise Management System Software Type 7843, ANOMS or Noise Sentinel Type 7871. The specifications that can be fulfilled is dependent on the system software used. In some cases, the relevant system software is specified BASIC MEASUREMENTS Logging Rate: ½ or 1 s Detectors: Parallel detectors on every measurement: A- or B-weighted (switchable): Broadband detector channel with one exponential time weighting (Fast, Slow, Impulse), one linearly averaging detector and one peak detector C- or Z-weighted (switchable): As for A- or B-weighted Overload Detector: Monitors the overload outputs of all the frequency weighted channels Measurements: X = frequency weightings A or B Y = frequency weightings C or Z V = frequency weightings A, B, C or Z U = time weightings F,S or I LXeq LYeq LXE LYE LCeq LAeqk LXUmax LYUmax LXUmin LYUmin LXIeq LYIeq LAIeq LAeq Lvpeak REPORTS* Short Reports: Period: User-defined 1 to 30 minutes, whole number of reports each hour Data: Start time; Stop time; Minimum of L(SPL) over the period; Maximum of L(SPL) over the period; Total Leq over the period; Total LIeq over the period; 5 LN Values with user-defined percentile levels; Standard deviation; Wind speed and wind direction (Noise Sentinel only) One Hour Reports: Data: Start time; Stop time; Level distribution (per mil for L (instan- taneous)) in 110 1 db classes, plus an Overload class and a Below class; One hour minimum of L(SPL); One hour maximum of L(SPL); One hour total Leq; One hour minimum of Leq; One hour maximum of Leq; Leq Event value (total Leq for all the events during the one-hour period); Leq Background value (total Leq for all the periods between events during the one-hour period); Persistent overload for the one-hour period; Standard deviation System Check Reports: CIC or actuator (depending on configuration). Start time; Leq during check; Leq before check; Leq after check NMT Health Reports: One hour reports with Start time and 60 minute values of Internal temperature, Battery voltage, Mains voltage, External voltage (connection to Utility Unit ZH-0689 required), Internal/Storage disk capacity, Internal/Storage disk free space, Available physical memory, and Idle CPU SOUND RECORDING Triggered By: Events or Short Reports Duration: User-defined up to 3 min Format: WAV Sound Quality Sampling Rate (khz) Low 8 16 High 48 96 Memory (kbyte/s) CALIBRATION CHECK The calibration can be checked and reported using CIC Interval: Up to 4 times per 24 hour period Report: Start Time, Leq before check, Leq during check, Leq after check INTERNAL STORAGE Logged data are stored on an 8 Gbyte Secure Digital memory card (SD card) Capacity: Up to 30 days. After this time the oldest data are overwritten PRODUCT INFORMATION PORTABLE NOISE MONITORING TERMINALS 09

AUDIO STORAGE The NMT continuously records the audio in listening quality. The last 2 days of audio are stored kept in NMT for retrieval of audio by Noise Sentinel On Demand alert functionality. After this time the oldest data are overwritten. INTERFACE LAN or Cellular Router READOUTS Data Status: Overview of the number of reports generated and sent Streamer: Readout parameters displaying the status of the streamer engine and network connection ORDERING INFORMATION Type 3655-C-DMO General-purpose Portable Noise Monitoring Terminal Included with Type 3655-C-DMO: Type 2250-N-D00 Type 4231 Hand-held Analyzer Type 2250-L (G4) with Noise Monitoring Software BZ-7232 and selected accessories (no microphone or preamplifier) Sound Calibrator Type 4952-A Outdoor Microphone Type 3535-A AF-0004 AN-0041 AN-0042 AN-0041 AO-0645-D-100 AO-1449-D-005 AO-1494 AQ-0696 AQ-1785 All-weather Case Guy Rope Mains Cable EU, CEE 7 (M) to C13 (F), 2.5 m (8.2 ft) Mains Cable GB, BS 1363A (M) to C13 (F), 2.5 m (8.2 ft) Mains Cable, US NEMA 5-15 (M) to C13 (F), 2.5 m (8.2 ft) Microphone Extension Cable, 3 m (9.8 ft) LAN Cable USB-A (M) to Micro-B (M) Cable, 1.8 m (5.9 ft) Power Cable for Router DC Power Cable DH-0800 Guy Rope Fasteners (set of 10) DH-0801 Tent Pegs (set of 10) DK-1763 Padlock ( 2) with Keys ( 4) DK-1769 DS-1195 KE-4339 Steel Wire Padlock, 1.5 m (4.9 ft) Transport Protection for Antenna Accessory Case QB-0079 Li-Ion Battery 2 UA-0803 Tripod UA-1707 Tripod Adaptor for Type 4952 UL-1017 UL-1036 UL-1041 ZG-0426 Post-Processing Type 7872 Secure Digital Memory Card Antenna for Digi Router Airlink GX400 Router with SIM card Mains Power Supply for Hand-held Analyzer Noise Sentinel On Demand Web-based subscription service providing long-term unattended outdoor noise management and monitoring MM-0256-A-DMO Noise Sentinel On Demand Weather Station Kit (Optional Accessory) Included with MM-0256-A-DMO: MM-0256-002 Six-parameter Weather Station (and mounting kit) AF-0004 Guy Rope AO-0657 USB Cable AO-0659 Cable M12 8-pin (F) to Lemo 1-B 8-pin (M), 10 m (33.3 ft) BR 1779 Weather Station Field Guide DB-4364 Weather Station Pole Adapter DH-0800 Guy Rope Fasteners (set of 10) DH-0801 Tent Pegs (set of 10) KE-4340 Weather Station Carrying Case QX-0016 Screwdriver QX-1171 2.5mm Hex Wrench UA-0803 UA-1707-A ZH-0689 Tripod Weather Station Tripod Adaptor Weather Station USB Adaptor Cardboard packaging (re-used for returns) Although reasonable care has been taken to ensure the information in this document is accurate, nothing herein can be construed to imply representation or warranty as to its accuracy, currency or completeness, nor is it intended to form the basis of any contract. Content is subject to change without notice contact EMS for the latest version of this document EMS Bruel & Kjaer Melbourne Level 12, 432 St Kilda Road Melbourne Victoria Australia Tel: +61 3 9508 4900 emsbk.com info@emsbk.com Sacramento 2330 East Bidwell, Suite 210 Folsom CA 95630 USA Tel: +1 916 265 7700 Denmark Skodsborgvej 307 DK-2850 Nærum Denmark Tel: +45 8877 4560 Taiwan PRODUCT 13FL-1,#128, INFORMATION sec.3 Min Shen PORTABLE E. Road Taipei NOISE Taiwan Tel: MONITORING +886 2 25462988TERMINALS 010 Madrid C/ Teide 5 28703 San Sebastián de los Reyes Spain Tel: +34 91 659 08 20