CALIBRATION PROCEDURE FOR STABILIZED LASERS USING THE METHOD OF OPTICAL BEATS MEASUREMENT UNCERTAINTY

Similar documents
FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM

Effect of frequency modulation amplitude on Iodine stabilized He-Ne Laser, at λ 633nm

A Multiwavelength Interferometer for Geodetic Lengths

Option G 4:Diffraction

Long-term Absolute Wavelength Stability of Acetylene-stabilized Reference Laser at 1533 nm

Absolute frequency measurement of wavelength standards

Frequency stability and reproducibility of iodine stabilised He-Ne laser at 633 nm

INTERFEROMETRIC VIBRATION DISPLACEMENT MEASUREMENT

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

A transportable optical frequency comb based on a mode-locked fibre laser

Chapter 7. Optical Measurement and Interferometry

TIME AND FREQUENCY ACTIVITIES AT THE CSIR NATIONAL METROLOGY LABORATORY

Lab 12 Microwave Optics.

Absolute distance interferometer in LaserTracer geometry

Module 5: Experimental Modal Analysis for SHM Lecture 36: Laser doppler vibrometry. The Lecture Contains: Laser Doppler Vibrometry

Optical generation of frequency stable mm-wave radiation using diode laser pumped Nd:YAG lasers

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Research Activities on Time and Frequency National Metrology Institute of Japan (NMIJ)/AIST

레이저의주파수안정화방법및그응용 박상언 ( 한국표준과학연구원, 길이시간센터 )

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Chapter 17: Wave Optics. What is Light? The Models of Light 1/11/13

Exam 3--PHYS 102--S10

PHYS General Physics II Lab Diffraction Grating

Fiber Pigtailed Variable Frequency Shifters Acousto-optic products

Swept Wavelength Testing:

Single Photon Interference Laboratory

A Fast Phase meter for Interferometric Applications with an Accuracy in the Picometer Regime

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

The Henryk Niewodniczański INSTITUTE OF NUCLEAR PHYSICS Polish Academy of Sciences ul. Radzikowskiego 152, Kraków, Poland.

Laser Telemetric System (Metrology)

RADIOMETRIC AND PHOTOMETRIC MEASUREMENTS AT THE LNE-INM/CNAM

Timing Noise Measurement of High-Repetition-Rate Optical Pulses

Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, 77. Table of Contents 1

3.003 Lab 3 Part A. Measurement of Speed of Light

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

Theory and Applications of Frequency Domain Laser Ultrasonics

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Spectrally resolved frequency comb interferometry for long distance measurement

Unit-23 Michelson Interferometer I

Zeeman Shifted Modulation Transfer Spectroscopy in Atomic Cesium

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Measuring the speed of light

Supplementary Figures

NCSL International 2995 Wilderness Place, Suite 107 Boulder, Colorado Office: (303) Fax: (303)

Optical phase-coherent link between an optical atomic clock. and 1550 nm mode-locked lasers

Interferometric OSEM Sensor Development

Introduction to the operating principles of the HyperFine spectrometer

Using GNSS for optical frequency and wavelength measurements

PoS(PhotoDet 2012)058

Exercise 8: Interference and diffraction

COATS: compact optical 5DoF attitude sensor for space applications

Absolute frequency measurement of the iodine-stabilized He Ne laser at 633 nm

Stability of a Fiber-Fed Heterodyne Interferometer

Microwave Diffraction and Interference

TIME & FREQUENCY SERVICES IN MALAYSIA

It s Our Business to be EXACT

Length and Position Measurement

Femtosecond Synchronization of Laser Systems for the LCLS

Two-Mode Frequency Stabilization of an Internal-Mirror 612 nm He-Ne Laser

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

Electro-Optic Sensors for RF Electric Fields: a Diagnostic Tool for Microwave Circuits and Antennas

THz Components and Systems

9. Microwaves. 9.1 Introduction. Safety consideration

Holography Transmitter Design Bill Shillue 2000-Oct-03

Improving a commercially available heterodyne laser interferometer to sub-nm uncertainty

Initial Results from the C-Mod Prototype Polarimeter/Interferometer

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 & ANSI/NCSL Z

Comparison of the NIST and NRC Josephson Voltage Standards (SIM.EM.BIPM-K10.b)

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Diffraction Single-slit Double-slit Diffraction grating Limit on resolution X-ray diffraction. Phys 2435: Chap. 36, Pg 1

Length section: New calibration and research services

Observation of Rb Two-Photon Absorption Directly Excited by an. Erbium-Fiber-Laser-Based Optical Frequency. Comb via Spectral Control

Microwave Circuit Design and Measurements Lab. INTRODUCTION TO MICROWAVE MEASUREMENTS: DETECTION OF RF POWER AND STANDING WAVES Lab #2

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

Practice Problems for Chapter 25-26

General Physics Laboratory Experiment Report 2nd Semester, Year 2018

Chapter Ray and Wave Optics

Ph 3455 The Photoelectric Effect

Detecting the Ratio of I ac. /I ave. photoelastic modulators

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Spectroscopy of Ruby Fluorescence Physics Advanced Physics Lab - Summer 2018 Don Heiman, Northeastern University, 1/12/2018

PhysicsAndMathsTutor.com 1

Photometry for Traffic Engineers...

Week IX: INTERFEROMETER EXPERIMENTS

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.

Wavelength Control and Locking with Sub-MHz Precision

MODERN INSTRUMENTS FOR ELECTRICITY MEASUREMENT UNDER PRESENT SITUATION

Laser Beam Analysis Using Image Processing

Understanding Optical Specifications

Physical Optics. Diffraction.

(51) Int Cl.: G01B 9/02 ( ) G01B 11/24 ( ) G01N 21/47 ( )

Report to the 20th Meeting of CCTF Research Activities on Time and Frequency National Metrology Institute of Japan (NMIJ)/AIST

Measure the roll-off frequency of an acousto-optic modulator

Frequency stabilized three mode HeNe laser using nonlinear optical phenomena

Agilent 83440B/C/D High-Speed Lightwave Converters

Agilent 71400C Lightwave Signal Analyzer Product Overview. Calibrated measurements of high-speed modulation, RIN, and laser linewidth

Theoretical Approach. Why do we need ultra short technology?? INTRODUCTION:

EE119 Introduction to Optical Engineering Spring 2002 Final Exam. Name:

First and second order systems. Part 1: First order systems: RC low pass filter and Thermopile. Goals: Department of Physics

Transcription:

U.P.B. Sci. Bull., Series A, Vol. 69, No. 1, 007 ISSN 13-707 CALIBRATION PROCEDURE FOR STABILIZED LASERS USING THE METHOD OF OPTICAL BEATS MEASUREMENT UNCERTAINTY Elena DUGHEANU 1 Lucrarea descrie metoda de etalonare a laserelor stabilizate prin metoda absorbţiei saturate în cazul 17 I prin metoda bătăilor optice, aplicată în laboratorul Lungimi al Institutului Naţional de Metrologie, Bucureşti, Romania şi estimarea bugetului de incertitudine a instalaţiei de comparare a frecvenţelor laser, în conformitate cu specificaţiile standardului SR ENV 13005:003 Ghid pentru exprimarea incertitudinii de măsurare. The paper describes the calibration method used for the calibration of stabilized lasers by saturated absorption in molecular iodine 17 I and the estimation of the uncertainty budget of the installation used for the comparison of two laser frequencies in Length Laboratory, National Institute of Metrology, Bucharest, Romania. The contribution of the author consists of the analyses of the sources of error which appear during the measurement, due to laser and to the equipments that compound the installation, according to the specifications of Guide to the expression of uncertainty in measurements, SR ENV 13005:003 Key words: stabilized lasers, installation for calibration of stabilized lasers, uncertainty budget 1. Introduction Before the invention of the laser, the infrared and visible spectrum were studied using interference techniques whose accuracy was limited by the nature of radiation source as spectral width and by method itself (errors due to diffraction, mirrors phase, etc.). The measurement of frequencies is in principle not limited by stability and accuracy of the oscillators used. The measurement method of frequency is more accurate with several orders than the measurement of wavelength. A laser is a monochromatic source radiation having a spectral width comparable to radio frequency and microwaves oscillator. From the second resolution of CIPM Comité International des Poids et Mesures adopted in 1983, the length may be defined as the interval of time in which the light passes a distance. This principle is put into practice by a 1 Phys.; National Institute of Metrology Bucharest, ROMANIA

64 Elena Dugheanu technique of interferential measurement method where the displacement of the mobile retroreflector of the interferometer is given by the following formula: L = (k+ε)λ / (1) c using λ =, equation (1) becomes: f c L = ( k + ε ) f () where: k is a whole number given by numbers of interferential fringes, counted by interferometer counter, ε represents the fraction fringe given by known radiation and c is the light velocity. The optic interferometer could be thus considered as a length measurement device, based on its measurements of the time necessary for the light to pass through a given distance. This interpretation used the light s frequency as a clock to measure the period of time, from where the length results according to equation (). The generation and the measurement of optical frequencies were possible due to the achievement of stabilized lasers based on saturated absorption and to the development in optical range of a heterodyne technique where one or more oscillators (from which one has a known frequency) are mixed in an adapted element in order to give a difference of low frequency which can be measured directly (fig 1). Fig. 1: Schematic of principle of heterodyne technique

Calibration of frequency stabilized helium - neon lasers... 65 Known frequency can be synthesized as an exact harmonic closed by the frequency that is mixed. This harmonic is mixed with unknown frequency, in an avalanche photo diode, so the frequencies of the heterodyne beats are close to MHz order. In this case, the unknown frequency f x is given by formula [3]: Δf = f x lf 1 mf nf - nf μu (3) where: Δf are the frequency beats, f x is the unknown frequency, f 1, f are the known frequencies of the two lasers and f μu is the frequency of microwaves oscillator. In this case the mix order is given by formula: p = l + m + n (4) In a particular case in which a laser having a known frequency is compared with a laser having an unknown frequency, the following formula can be used: f x = kf e ± Δf (5) where k is a harmonic number that can be determined by measuring of a known length.. Measuring of the laser frequencies using the method of beats frequencies A known method of measuring lasers frequencies is the method of optical beats. Due to the very high stability and very fine spectral width of radiations the method permits the measurement of the beats between two different lasers with closed frequencies, f 1 şi f. The method consists of the simultaneous illumination of the same avalanche photo diode and two laser beams using a beam splitter and several optical elements for alignment. The illumination of the fotodetector and the output current are modulated with beats frequency Δf = f 1 - f. Modulation frequency Δf of the current is measured by usual radio electronic methods. This installation consists of (fig:): lasers (I and II) between which is measured the difference in frequency, optical elements for beam lasers alignment (1 and 3), mirror for the reflection of laser beam with the possibility of rotation in two rectangular axes (), beam splitter with the possibility of rotation in two rectangular axes (4), lens with a focal distance of 1 cm, mounted on a magnetic support and provided with a sliding bar (5), avalanche photo diode mounted on a support adjustable on x, y, z axes (6), impedance adaptor, decade attenuator (7), broadband amplifier (1 GHz, G = 0) (8), frequency meter 1,3 GHz with low level of automatic trigger of counting (9), 1GHz band spectrum analyzer (10) and computer (11).

66 Elena Dugheanu Fig. : Installation for measuring the laser frequency using the optical beat method The whole installation is mounted on a stable metal base equipped with pneumatic suspension to reduce the level of vibrations under 10 Hz. This method permits the comparison of two closed laser frequencies with a maxim difference of 1Ghz. The ranges of frequency, which are measured in case of stabilized lasers in frequency on the same nominal value, are between 0.1 MHz and 1 GHz. The unknown frequency f x which depends on standard frequency f e is given by the formula: f x = f e ± Δf (6) The photo diode behaviour is in avalanche mode, for a high frequency (manufactured by RCA Electro-Optics, Canada, type C 3090 E) and has the following characteristics: - spectral range of measurement: λ = (400 1000) nm; - small responding time: 0,5 ns - current density at 0 C: 5 ma/mm - photosensitive layer: circular surface equal to 0. mm ; - dark current: (1.5 10-8 3 10-8 ) A; - noise current: : (.3 10-13 5 10-13 ) A/Hz 1/ - power supply: (180 50) V. A schematic of the photodiode is showed in fig. 3. The electronic components have the following values: R 1 =.7 kω, R = 510 Ω, C 1 = 4.7nF/500V, C = 4.7nF/500V, C 3 = 4.7nF/500V, FD photodiode C 3090 E, RCA.

Calibration of frequency stabilized helium - neon lasers... 67 Fig. 3: Schematic of the photodiode 3. Program for the measuring of laser frequencies The program used for measuring laser frequencies and processing the experimental data is a computer based program which has the following set of instructions: a) initialize the HP-IB interface b) order HP 538 B frequency meter to set the data (resolution, interval between two readings, etc) c) permit the user to set up the following parameters: - sampling interval; - n number of readings at a time t ; - determination of average value of frequency f ; standard deviation of frequency f ; Allan s(,t) variance; - displaying of results; - printing the results. d) processing of the experimental data permits the determination of the following characteristics: - determination of f x frequency of laser which is calibrated according to formula (6): - standard deviation (s) for one measurement for an establish sample of time t and number of readings n : where: n ( Δ fi Δf ) i= 1 s = (7) n 1 n i= 1 Δ f = Δf n i (8)

68 Elena Dugheanu is the frequencies average. The stability of frequency s (,t), is determined using Allan s variance formula: N ( Δfi Δfi 1 ) 1 i = (, t) = n 1 s (9) where: t is the sample time that be between (0.1 1000) s, Δf i, Δf i-1 represent the measurements of successive differences in frequency between the two frequency between the two lasers and n is the number of readings. The relative value of stability of frequency is determined by formula: s(, t) s(, t) = (10) rel where f x represents the frequency of the laser which is tested. f x 4. Measurement uncertainty of installation for the measuring of the laser frequencies Uncertainty measurement u is an estimator that establishes the limits of the result values. That includes, with a probability, the true value of measurand. The measurement uncertainty consists of several components. One of them (u A component) can be estimated from statistic distributions of values and can be characterized from experimental standard deviation; other ones (u B components) are estimated from previous experiments or from other information. The determination of uncertainty of measurement of laser frequency using the special installation for measuring the beats frequencies is based on the evaluation of type A component for repeatability of measurement of laser stabilized frequency and of type B component, achieved by evaluating the measurement uncertainty of equipments which compound the installation. 4.1 Value of type A uncertainty, determined from repeatability of measurement of laser stabilized frequency is calculated by formula: s u A = n (11) From experimental data using a standard frequency generator, results u A = 0,4 khz 4. Value of type B uncertainty is given by: the uncertainty of frequency meter, u E1, the uncertainty of avalanche photo diode, u E, the uncertainty of absolute laser frequency, according the BIPM certificate no. 64 / August 3 005, u l (f f ) = 0, khz, for a frequency: f s = 473 61 353 61.7 khz.

Calibration of frequency stabilized helium - neon lasers... 69 a) The frequency meter error (E 1 ), based on the technical documentation of apparatus is given by the following formula: E 1 = ± resolution ± time base error frequency (1) Using the following data: the resolution = 1 Hz, the time base error = 10-7, the frequency, f 5 10 5 khz one obtains E 1 = ± 0,051 khz. For a rectangular distribution, the value of uncertainty is u E1 = 0.09 khz. b) the error determined by the noise of avalanche photo diode is, based on the documentation, equal to E = ± 10-3 khz. (13) For a rectangular distribution, the value of uncertainty is u E = 0,001 khz. The B component of uncertainty is: u B = u ( f f ) + u ( E1) + u ( E ) = 0,0 khz (14) 4.3 Composed uncertainty is: u C = u A + ub = 0.45 khz (15) The relative uncertainty of the installation is given by formula: u r = where: u 0.45 1 r = = 0.95 10. 4736135361.7 u f c l (16) 4.4 Extended uncertainty 4.5 For an extended factor k =, the relative extended uncertainty U of the installation is given by formula: U r = u r =1,90 10-1 (17) 5. Conclusions In Romania, the transmission of unit of length to secondary standards is made using a stabilized laser by saturated absorption in molecular iodine 17 I as national standard of length in association with a special installation for calibration of others lasers He-Ne, as secondary standards of length. The installation which is unique in our country offers the following advantages: transfers the measurement unit, the wavelength of He-Ne laser stabilized in frequency, directly from the national standard of length as He-Ne laser stabilized in frequency which is connected to the cesium standard of frequency and the light velocity in vacuum, has a high reproducibility of measurement unit (10-10 10-8 ).

70 Elena Dugheanu The value of the measurement uncertainty of the installation is a very important parameter in determining the metrological characteristics of stabilized laser in frequency and assures the dissemination of length unit from the National Standard of Length to all secondary and reference standards of length with high accuracy, at a very closed level provided by metrological laboratories of length from developed countries, as it is showed in recent international key comparison BIPM.L K11/005. R E F E R E N C E S [1] Guide to the expression of uncertainty in measurements, SR ENV 13005:003 [] I.M. Popescu, A.M. Preda, G.F. Cone - Aplicaţii ale laserilor, Ed. Tehnică 1979 [3] D. Boiciuc, Teza de doctorat: Laserul în metrologia lungimilor, 1996 [4] C. S. Edwards, G. P. Barwood, P. Gill and W. R. C. Rowley, "A 633 nm iodine-stabilized diode-laser frequency standard", Metrologia 36, 41 (1999). [5] J.A.Stone, S.D.Phillips and G.A.Mandolfo, Corrections for Wavelength Variation in Precision Interferometric Displacement Measurements, J.Res.Natl.Inst.Std and Technol. (US) pg. 101 (5) (1996) [6] Journal of Research of the National Institute of Standards and Technology, Volume 104, Number 3, May-June 1999 [7] A. Duta, D. Boiciuc, D.Georgescu, Results from the CI-005 campaign at the BIPM of the BIPM.L-K11 ongoing key comparison.