Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Similar documents
Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

THE INFLUENCE OF THE TRANSDUCER BANDWIDTH AND DOUBLE PULSE TRANSMISSION ON THE ENCODED IMAGING ULTRASOUND

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

18th World Conference on Nondestructive Testing, April 2012, Durban, South Africa. Joanna X.Qiao 1, Matthias Jobst 2

Linear arrays used in ultrasonic evaluation

Choosing an Ultrasonic Sensor for Ultrasonography

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

CHAPTER 1 INTRODUCTION

Further development of synthetic aperture real-time 3D scanning with a rotating phased array

Designing Non-linear Frequency Modulated Signals For Medical Ultrasound Imaging

Virtual ultrasound sources

Ultrasonic Linear Array Medical Imaging System

White Rose Research Online URL for this paper: Version: Accepted Version

Real Time Deconvolution of In-Vivo Ultrasound Images

Introduction. Parametric Imaging. The Ultrasound Research Interface: A New Tool for Biomedical Investigations

FPGA-Based Control System of an Ultrasonic Phased Array Keywords: ultrasonic imaging, phased array, B-scan, FPGA

APPLYING SYNTHETIC APERTURE, CODED EXCITATION, AND TISSUE HARMONIC IMAGING TECHNIQUES TO ALLOW ULTRASOUND IMAGING WITH A VIRTUAL SOURCE ROBYN T.

Optimization of Axial Resolution in Ultrasound Elastography

FPGA-BASED CONTROL SYSTEM OF AN ULTRASONIC PHASED ARRAY

Original Article Barker-coded excitation in ophthalmological ultrasound imaging

Transducer product selector

Reconfigurable Arrays for Portable Ultrasound

Session: 2A NEW ULTRASOUND SYSTEMS Chair: H. Ermert University of Bochum 2A-1 10:30 a.m.

An Overview Algorithm to Minimise Side Lobes for 2D Circular Phased Array

Parametric Beamformer for Synthetic Aperture Ultrasound Imaging

ULTRASONIC IMAGING of COPPER MATERIAL USING HARMONIC COMPONENTS

Medical Imaging (EL582/BE620/GA4426)

Simulation of advanced ultrasound systems using Field II

Motion Compensation Improves Medical Ultrasound Image Quality

12/26/2017. Alberto Ardon M.D.

Retrospective Transmit Beamformation. Whitepaper. ACUSON SC2000 Volume Imaging Ultrasound System. Answers for life.

DESIGN & VALIDATION OF A SEMI-FLEXIBLE PAUT PROBE FOR THE MANUFACTURING INSPECTIONS OF LARGE FORGED ROTORS

System Architecture of an Experimental Synthetic Aperture Real-time Ultrasound System

3D synthetic aperture imaging using a virtual source element in the elevation plane

A Delta-Sigma beamformer with integrated apodization

Abstract. 1 Introduction. 1.2 Concept. 1.1 Problematic. 1.3 Modelling

High-Resolution Corrosion Monitoring for Reliable Assessment of Infrastructure

A Modified Synthetic Aperture Focussing Technique Utilising the Spatial Impulse Response of the Ultrasound Transducer

Ultrasound Nondestructive Evaluation (NDE) Imaging with Transducer Arrays and Adaptive Processing

ECHO-CANCELLATION IN A SINGLE-TRANSDUCER ULTRASONIC IMAGING SYSTEM

Mathematical Modeling of Ultrasonic Phased Array for Obstacle Location for Visually Impaired

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Resolution Enhancement and Frequency Compounding Techniques in Ultrasound.

VLSI Architecture for Ultrasound Array Signal Processor

AIP (2015) 34. AIP ISBN

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

15 th Asia Pacific Conference for Non-Destructive Testing (APCNDT2017), Singapore.

Coded excitations NINE. 9.1 Temporal coding

SIMULATION OF B-SCAN IMAGES FROM TWO-DIMENSIONAL TRANSDUCER ARRAYS: PART II - COMPARISONS BETWEEN LINEAR AND TWO-DIMENSIONALPHASED ARRAYS

Further Developments in Ultrasonic Phased Array Inspection of Aging Aircraft

BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM

Parameter Estimation Techniques for Ultrasound Phase Reconstruction. Fatemeh Vakhshiteh Sept. 16, 2010

Beamforming in ultrasound

Ultrasound physical principles in today s technology

ENHANCEMENT OF SYNTHETIC APERTURE FOCUSING TECHNIQUE (SAFT) BY ADVANCED SIGNAL PROCESSING

A COST-EFFECTIVE METHOD FOR ULTRASOUND VOLUMETRIC IMAGING

Optimal Control by Transmit Frequency in Tissue Harmonic Imaging

DACON INSPECTION SERVICES. Phased Array Ultrasonic Testing

On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size

The Middle East Distributor for AMBISEA Technology Corp. Electro-Medical Product Line

Integrated 16-channel Transmit and Receive Beamforming ASIC for Ultrasound Imaging

Optimized Semi-Flexible Matrix Array Probes for Large Rotor Shafts and DGS Sizing Diagram Simulation Tool

3. Ultrasound Imaging(2)

Encoding of inductively measured k-space trajectories in MR raw data

DIGITAL BEAM-FORMING ANTENNA OPTIMIZATION FOR REFLECTOR BASED SPACE DEBRIS RADAR SYSTEM

Abstract. methods. was degraded due to. realized. and M-mode. based on B-mode. rate. myocardial. diagnosis of the. vibration useful for.

Ultrasound Research Scanner for Real-time Synthetic Aperture Data Acquisition

Synthetic-aperture based photoacoustic re-beamforming (SPARE) approach using beamformed ultrasound data

3-D Imaging using Row--Column-Addressed 2-D Arrays with a Diverging Lens

Simulation of Ultrasonic Testing of Rail Wheel Face using Phased Array and DDF technique

B-mode imaging components

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

PHYSICALLY, the speed of sound in human tissue limits

Improving the Quality of Photoacoustic Images using the Short-Lag Spatial Coherence Imaging Technique

Theory and Applications of Frequency Domain Laser Ultrasonics

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

The Implementation of the New Standard EN ISO for Ultrasonic Phased-Array Systems at the Manufacturer

A complete COMSOL and MATLAB finite element medical ultrasound imaging simulation

A miniature all-optical photoacoustic imaging probe

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Wideband Focused Transducer Array for Optoacoustic Tomography

ISSN: ISO 9001:2008 Certified International Journal of Engineering Science and Innovative Technology (IJESIT) Volume 2, Issue 4, July 2013

Broadband Minimum Variance Beamforming for Ultrasound Imaging

Beamforming of Frequency Diverse Array Radar with Nonlinear Frequency Offset Based on Logistic Map

Kirchhoff migration of ultrasonic images

INSPECTION OF THERMAL BARRIERS OF PRIMARY PUMPS WITH PHASED ARRAY PROBE AND PIEZOCOMPOSITE TECHNOLOGY

Multiplierless sigma-delta modulation beam forming for ultrasound nondestructive testing

REAL-TIME B-SCAN ULTRASONIC IMAGING USING A DIGITAL PHASED. Robert Dunki-Jacobs and Lewis Thomas General Electric Company Schenectady, New York, 12301

Time Reversal Ocean Acoustic Experiments At 3.5 khz: Applications To Active Sonar And Undersea Communications

3. give specific seminars on topics related to assigned drill problems

Effect of coupling conditions on ultrasonic echo parameters

Introduction to Ultrasound Physics

Spatial Encoding Using a Code Division Technique for Fast Ultrasound Imaging

Detection of Multipath Propagation Effects in SAR-Tomography with MIMO Modes

Transcription:

ARCHIVES OF ACOUSTICS 33, 4, 573 580 (2008) LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI Institute of Fundamental Technological Research Polish Academy of Sciences Świętokrzyska 21, 00-049 Warszawa, Poland e-mail: igortr@ippt.gov.pl (received June 15, 2008; accepted November 6, 2008) The paper describes the synthetic transmit aperture (STA) imaging system with a single element transmitting and multi-element reception in medical ultrasound. Synthetic aperture method allows to achieve high electronic signal-to-noise ratio and good contrast resolution. A laboratory setup for acquisition of RF signals from linear transducer array was built. Simulated multichannel acquisition by multiplexing individual transducer was performed. In experiments 32-element linear transducer array with 0.48 mm inter-element spacing and a burst pulse with time duration 100 ns was used. Single element in the transducer transmitting aperture was used to generate a spherical wave covering the full image region. The echo signals were sampled independently by individual elements for each transmission. The comparison of 2D ultrasound images of wire phantom obtained using STA method and standard linear array scanning with commercial ultrasonograph is given. The results show excellent image resolution of the STA method and its robustness to refraction, attenuation and multiple reflection of ultrasound waves. Keywords: ultrasound imaging, synthetic aperture, contrast resolution. 1. Introduction Medical ultrasound imaging is a technique that has become prevalent medical imaging techniques since it is easily accessible, less expensive, safe, simple to use and produces images in the real time. However, images produced by an ultrasound imaging system, must be of sufficient quality to provide accurate clinical interpretation. The most commonly used image quality measures are spatial resolution and image contrast which can be determined in terms of beam characteristics of an imaging system: beam width and side-lobe level. In the design of an imaging system, the optimal set of system parameters is usually found as a trade-off between the side-lobe level and the beam width of an imaging system.

574 I. TROTS, A. NOWICKI, M. LEWANDOWSKI In conventional ultrasound imaging system, when only one transducer is used, the quality of images directly depends on the transducer acoustic field. Also in conventional ultrasound imaging the data are acquired sequentially one image line at a time that puts a strict limit on the frame rate that is important in real-time imaging system. Low frame rate means that moving structures (e.g. heart valves) are not perfectly reproduced and diagnosis may be impaired. This limitation can be lifted by employing synthetic aperture (SA) imaging. Here data are acquired simultaneously from all directions over a number of emissions, and the full image can be reconstructed from this data. The SA method aspects are discussed in [2, 8]. In SA method as well as in conventional method the peak acoustic power limits the signal-to-noise ratio (SNR) in ultrasound images. This limitation can be overcome by using long wide band coded transmitting sequences and compression techniques on the receiver side [1]. Longer signals allow to obtain the results similar to that obtained using single short pulses but with much higher amplitude [3, 9]. There are several papers in literature concerning similar boundary-condition problem of signal compression in medical diagnostic imaging [4 7]. In comparison with other coded excitation schemes, such as chirp, pseudo-random sequences and Barker codes, the CGS allowed virtually side-lobe free operation. In this paper a STA method for medical ultrasound imaging system was implemented and tested by using radio frequency (RF) data from wire phantom. The results obtained using STA method were compared with the results obtained using conventional composite transmit focusing method. The investigated STA method increased the frame rate and allowed to obtain the significantly better image quality. 2. Synthetic aperture ultrasound imaging One of the important processes in ultrasound imaging systems is beamforming. There are many different beamforming methods. In this paper the STA method is discussed only. In STA imaging, at each time one array element transmits a pulse and all elements receive the echo signals. The advantage of this approach is that a full dynamic focusing can be applied to the transmission and the receiving, giving the highest quality of an image. The method for acquiring synthetic aperture ultrasound images is shown in Fig. 1. A single element in the transducer aperture is used for transmitting a spherical wave covering the full image region. The received signals for all or part of the elements in the aperture are sampled for each transmission. Using the STA, dynamic focusing can be performed on the synthetic transmit aperture and the receive aperture, as well. In STA method focusing is performed by finding the geometric distance from the transmitting element to the imaging point and back to the receiving element. The structure of the synthetic aperture and geometric relation between the transmit and receive element combination is shown in Fig. 2.

LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING 575 Fig. 1. Principle of STA imaging method. Fig. 2. Geometric relation between the transmit and receive element combination and the focal point. When a short pulse is transmitted by element m and the echo signal is received by element n, as shown in Fig. 2, a round-trip delay is τ m,n = τ m + τ n, (1) where (m, n) is a transmit and receive element combination, 0 m, n N 1. According to Law of Cosine, the distances from the point to the m-th element, r m, and to the n-th element, r n, are r m = x 2 m + r 2 2x m r sinθ, r n = x 2 n + r 2 2x n r sinθ, (2) where x m, x n are the positions of the m-th and n-th elements, respectively and r is distance between synthetic aperture centre and the point (r, θ). The delays for m-th element and n-th element are τ m = 1 c (r r m), τ n = 1 c (r r n). (3) For an N-element array for each point in an image, the A-scan signal can be expressed as N 1 N 1 ( ) 2r a(r, θ) = y m,n c τ m,n, (4) m=0 n=0

576 I. TROTS, A. NOWICKI, M. LEWANDOWSKI where y m,n (t) is the echo signal and τ m,n is beamforming delay for the (m, n) receive and transmit element combination given in (1). The first and second summations correspond to transmit and receive beamforming. 3. Ultrasound imaging system A simplified ultrasound imaging system is shown in Fig. 3. Fig. 3. Block diagram of the ultrasound imaging system. Transducer transmits pulsed ultrasound waves, and receives reflected echo signals. Echoson SG3 (Echoson, Puławy, Poland) analog beamformer enables full control of selected 32 consecutive channels/transducers of a linear array transducer. Parameters of transmission and reception are programmable from a PC using a serial port (RS-232). Using the beamformer one can switch on arbitrary transmit and receive channels in the selected 32 channels aperture. The second block, real time scanline processing, is responsible for steering of the ultrasound waves. This block extracts the RF data, acquires it and send to the PC. Next, the collected digital data were processed offline and displayed on the monitor. All post processing and display is done on the computer using Matlab R routines. The processing creates 2D ultrasound imaging focused in every point of image. The system enabled us to perform simulated multichannel acquisition for synthetic aperture imaging. Using a single channel digitizer and switching receiving transducers we were able to collect RF data for up to 32 lines. The data acquisition in synthetic aperture imaging is radically different from a normal ultrasound system since data have to be stored for all receiving transducers and for a number of emissions. Repeating this procedure for each of the 32 transducers we obtain 1024 RF lines that are the input to the synthetic aperture algorithm, Eq. (4). 4. Experimental results and discussion The 32-element linear transducer array with 0.48 mm inter-element spacing and a burst pulse with time duration 100 ns (a half-cycle at nominal frequency 5 MHz) were used. The inter-element space is about 3λ. All elements are used for both transmitting

LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING 577 and receiving. One single element in the transducer transmitting aperture was used to generate a spherical wave covering the full image region. Each time only one element transmits the probing signal and all elements receive the echoes. The transmit and receive elements combination gives a total of 32 32 possible RF A-lines. All these possible A-lines echo signals were sampled independently at 50 MHz and stored. The wire phantom, used in the experiments, consisted of 24 wires 0.1 mm in diameter positioned every 2 mm axially and at an angle of 75 degrees. This phantom allows to examine the axial and lateral resolution at various depths in the ultrasound image as well as focal and dead zone registration. Three of 1024 (32 32) received RF echo signals which were digitized and stored in the PC are shown in Fig. 4. Fig. 4. The recorded by PC RF echo signals: a) element #9 is transmitting element #12 is receiving, b) #18 transmitting #24 receiving, c) #32 transmitting #32 receiving. All these RF echo signals are different and echo time position and signal amplitude in every case depends on sound field and geometrical position of transmitted and received transducers. After all emmissions the full set of the RF A-lines echo signals needed to reconstruct one 2D B-mode ultrasound image is obtained. For this aim the RF lines are input to the synthetic aperture algorithm which calculate the time delay for every imaging point. As was mentioned above, a single element in the transducer aperture is used for transmitting a spherical wave covering the full image region. The received signals for all or part of the elements in the aperture are sampled for each transmission. This data can be used for making a low resolution images (Fig. 5a 5c), which is only focused in receive due to the unfocused transmission. Focusing is performed by finding

578 I. TROTS, A. NOWICKI, M. LEWANDOWSKI the geometric distance from the transmitting element to the imaging point and back to the receiving element and can be obtained from Eq. (2). These low resolution images need to be added coherently to form the final high resolution image (Fig. 5d). a) b) c) d) Fig. 5. Low resolution images combined to produce a high resolution image. One element transmit at the time, while all is used to receive. The images are then added into one high resolution image. The comparison of the reconstructed wire phantom images obtained using STA method and standard linear array scanning with commercial ultrasonograph Antares (Siemens, Mountain View, CA, USA) is shown in Fig. 6. The 128-elements linear transducer array with 0.3 mm pitch (VF13-5) and a burst cycle pulse at nominal frequency 10 MHz were used in ultrasonograph. The maximum quantity of focal points, which is equal to 8, was chosen in the ultrasonograph, they are marked by triangles in the bar. It needs to be noted, that frame rate in that case dramatically decreases down to 4 fps (in the case of one focal point it is equal to 30 fps). Such frame rate is definitely insufficient to normal examine the dynamically moved organs, such as heart, where high frame rate even up to 50 fps is requirement. In Fig. 6 it can be easily seen, that axial and lateral resolutions at the top and at the bottom parts of the image are different and depend on the focal point quantity in these regions.

LABORATORY SETUP FOR SYNTHETIC APERTURE ULTRASOUND IMAGING 579 Fig. 6. 2D B-mode images of the wire phantom obtained using STA method (left) and commercial ultrasonograph Antares (right). The bar shows the focal points. The images show the performance of the STA method. It allows to obtain better resolution than the conventional ultrasound system with an increased frame rate. The results obtained show the effectiveness of the STA method and its robustness to the refraction, attenuation, and multiple reflection of ultrasound waves. 5. Conclusion The work concerns the development and investigation of the STA method that allows to increase system frame rate and thus to improve the image quality. The paper explains how SA ultrasound imaging can be acquired and processed. The STA method was investigated experimentally. The phantoms, which contain wires, were used to test general image quality. The image reconstructed using the STA method gives a better image resolution than standard ultrasonograph, while frame rate is not decreasing. The disadvantage of the STA system is that storage and processing requirements are higher than in conventional beamforming, because the RF data must be stored for every combination of transmit and receive elements, and later recombined. The synthetic aperture method can be applied in standard ultrasonography. Introduction of STA method in medical ultrasound increases effectiveness and quality of the ultrasound diagnostic. Acknowledgments The paper is extended version of the paper presentaed at Open Seminar Acoustics 2008, September 8 12, Wrocław Piechowice.

580 I. TROTS, A. NOWICKI, M. LEWANDOWSKI References [1] FURGASON E.S., NEWHOUSE V.L., BILGUTAY N.M., COOPER G.R., Application of random signal correlation techniques to ultrasonic flow detection, Ultrasonics, 13, 11 17 (1975). [2] JENSEN J.A., NIKOLOV S.I., GAMMELMARK K.L., PEDERSEN M.H., Synthetic aperture ultrasound imaging, Ultrasonics, 44, Suppl., e5 e15 (2006). [3] KLIMONDA Z., LEWANDOWSKI M., NOWICKI A., TROTS I., Direct and post-compressed sound fields for different coded excitations experimental results, Archives of Acoustics, 30, 4, 507 514 (2005). [4] LITNIEWSKI J., NOWICKI A., KLIMONDA Z., LEWANDOWSKI M., Sound fields for coded excitations in water and tissue: experimental approach, Ultrasound in Med. & Biol., 33, 4, 601 607 (2007). [5] MISARIDIS T., Ultrasound imaging using coded signals, Center for Fast Ultrasound Imaging, Technical University of Denmark, August 2001. [6] NOWICKI A., KLIMONDA Z., LEWANDOWSKI M., LITNIEWSKI J., LEWIN P.A., TROTS I., Comparison of sound fields generated by different coded excitations Experimental results, Ultrasonics, 44, 1, 121 129 (2006). [7] NOWICKI A., TROTS I., LEWIN P.A., SECOMSKI W., TYMKIEWICZ R., Influence of the ultrasound transducer bandwidth on selection of the complementary Golay bit code length, Ultrasonics, 47, 64 73 (2007). [8] O DONNELL M., YAO W., Coded excitation for synthetic aperture ultrasound imaging, IEEE Trans. Ultrason. Ferro. Freq. Contr., 52, 2, 171 176 (2005). [9] TROTS I., NOWICKI A., SECOMSKI W., LITNIEWSKI J., Golay sequences side-lobe canceling codes for ultrasonography, Archives of Acoustics, 29, 1, 87 97 (2004).