Large Signal Displacement Measurement with an Asylum SA Atomic Force Microscope Rev B

Similar documents
Large Signal Displacement Measurement with an MTI Photonic Sensor Rev B

Application Note: Precision Displacement Test Stand Rev A

Large Signal Displacement Measurement With Piezo Jena vibrometer Rev A

Technical Report Preventing Air Gap Breakdown Rev -

Application Note Measuring Small Signal Capacitance vs Magnetic Field Rev A

Technical Report MFIS and TFFT Testing with the Premier II Tester Rev C

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

Standard Configuration

Precision Non-Linear Materials Testers. New Precision PiezoMEMS Analyzer

Table of Contents TABLE OF CONTENTS...I TABLE OF FIGURES...III C - QUIKLOOK SETUP...22

PFM Experiments with High Voltage DC/AC Bias

Prepare Sample 3.1. Place Sample in Stage. Replace Probe (optional) Align Laser 3.2. Probe Approach 3.3. Optimize Feedback 3.4. Scan Sample 3.

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Using the EDU Oscilloscope to Operate Sensors

Agilent 4070 Series Accurate Capacitance Characterization at the Wafer Level

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes

SENSOR AND MEASUREMENT EXPERIMENTS

Measurement of Microscopic Three-dimensional Profiles with High Accuracy and Simple Operation

Atomic Force Microscopy (Bruker MultiMode Nanoscope IIIA)

CHAPTER 2 D-Q AXES FLUX MEASUREMENT IN SYNCHRONOUS MACHINES

MEC751 Measurement Lab 2 Instrumented Cantilever Beam

Product Note 73 Vibration Tester for On-Wafer Tuner Operation

Basic methods in imaging of micro and nano structures with atomic force microscopy (AFM)

V2018 SPINSTAND AND NEW SERVO-8 SYSTEM

Active Vibration Isolation of an Unbalanced Machine Tool Spindle

Optical Microscope. Active anti-vibration table. Mechanical Head. Computer and Software. Acoustic/Electrical Shield Enclosure

STJ-100 TMR Magnetic Microsensor Dual In-line Package

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Non-Contact AC Voltage Probe SP3000, AC Voltage Probe SP9001

DesignCon Noise Injection for Design Analysis and Debugging

Load Cells, LVDTs and Thermocouples

CONNECTING THE PROBE TO THE TEST INSTRUMENT

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

Sweep / Function Generator User Guide

Optical Monitoring System Enables Greater Accuracy in Thin-Film Coatings. Line Scan Cameras What Do They Do?

Managing Complex Impedance, Isolation & Calibration for KGD RF Test Abstract

Unit-25 Scanning Tunneling Microscope (STM)

Internship report submitted in partial fulfilment of the requirements for the degree of Bachelor of Science in Applied Physics and Electronics

Characterization of Silicon-based Ultrasonic Nozzles

Surface Finish Measurement Methods and Instrumentation

LION PRECISION. TechNote LT February, Capacitive Sensor Operation and Optimization

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

Name Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits

Vibration-compensated interferometer for measuring cryogenic mirrors

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

10 Things to Consider when Acquiring a Nanopositioning System

Aero Support Ltd, 70 Weydon Hill Road, Farnham, Surrey, GU9 8NY, U.K.

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS

A Project Report Submitted to the Faculty of the Graduate School of the University of Minnesota By

Nanonics Systems are the Only SPMs that Allow for On-line Integration with Standard MicroRaman Geometries

PACS Nos v, Fc, Yd, Fs

Configuration of PNA-X, NVNA and X parameters

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2

Electric polarization properties of single bacteria measured with electrostatic force microscopy

MEASUREMENT OF STRAIN AND POLARIZATION IN PIEZOELECTRIC AND ELECTROSTRICTIVE ACTUATORS

Optical Pumping Control Unit

First Time User Manual

Impedance 50 (75 connectors via adapters)

ACCUMEASURE. Non-contact Capacitance Position Measurement with Nanometer Accuracy. A worldwide leader in precision measurement solutions

E84 Lab 3: Transistor

PeakForce SECM with Bio-Logic SP-300 Potentiostat

Radial Polarization Converter With LC Driver USER MANUAL

There is a twenty db improvement in the reflection measurements when the port match errors are removed.

Figure for the aim4np Report

Theory: The idea of this oscillator comes from the idea of positive feedback, which is described by Figure 6.1. Figure 6.1: Positive Feedback

LAB 4: OPERATIONAL AMPLIFIER CIRCUITS

total j = BA, [1] = j [2] total

Model LIA100. Lock-in Amplifier

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

Experiment 2: Electronic Enhancement of S/N and Boxcar Filtering

Instruction manual for T3DS software. Tool for THz Time-Domain Spectroscopy. Release 4.0

Specifying and Measuring Nanometer Surface Properties. Alson E. Hatheway

The shunt capacitor is the critical element

The Active Bridge 11/20/09

IR-183A & IR-283A Intracellular Recording Amplifier Instruction Manual

Operational Amplifiers

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

Experiment 1: Instrument Familiarization (8/28/06)

CO2 laser heating system for thermal compensation of test masses in high power optical cavities. Submitted by: SHUBHAM KUMAR to Prof.

Rapid Array Scanning with the MS2000 Stage

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

Experiment 1: Instrument Familiarization

Experiment P-10 Ohm's Law

High Power Piezo Driver

Transient Current Measurement for Advance Materials & Devices

Profile Measurement of Resist Surface Using Multi-Array-Probe System

ANALOG TO DIGITAL CONVERTER ANALOG INPUT

Mach 5 100,000 PPS Energy Meter Operating Instructions

Manual CBT current preamplifier PA1 and CBT Reader software Version 1.5

Body-Biased Complementary Logic Implemented Using AlN Piezoelectric MEMS Switches

Description of a Function Generator Instrument

TD250 6 Channel 250V Amplifier Manual and Specifications

XYZ Stage. Surface Profile Image. Generator. Servo System. Driving Signal. Scanning Data. Contact Signal. Probe. Workpiece.

SAE AE-2 Lightning Committee White Paper

Description of options, upgrades and accessories for the laser beam stabilization system Compact

CAPACITIVE SENSORS. Series KS. Key-Features:

PIEZO FILM LAB AMPLIFIER

EXPERIMENT 5 Bioelectric Measurements

SA210-Series Scanning Fabry Perot Interferometer

University of Utah Electrical Engineering Department ECE 2100 Experiment No. 2 Linear Operational Amplifier Circuits II

Transcription:

Radiant Technologies, Inc. 2835D Pan American Freeway NE Albuquerque, NM 87107 Tel: 505-842-8007 Fax: 505-842-0366 e-mail: radiant@ferrodevices.com www.ferrodevices.com Large Signal Displacement Measurement with an Asylum SA Atomic Force Microscope Rev B Date: January 8, 2009 Author: Joe Evans Introduction: At the invitation of Dr. Roger Proksch, I visited Asylum Research in Santa Barbara, CA on Tuesday, October 21 and Wednesday, October 22, 2008. Asylum has an experimental lab with an SA Atomic Force Microscope for use in developing hardware and software for its products. I connected a Precision Premier II to the AFM and captured the butterfly loop of a 1µ thick PNZT capacitor at 1Hz. The Asylum instruments may be found at www.asylumresearch.com. Test Setup: I brought to Asylum a 100V Precision Premier II for acquiring the measurements. Using its built-in USB communications port, I ran the tester from my laptop. Roger programmed the SA to route the output of the quad-cell detector for the AFM to an external BNC connector on the AFM. I connected that signal to the SENSOR input of the Premier II. Roger calibrated the output of the AFM Z displacement to a sensitivity of 70nm/V or 700Å/V. The Premier II SENSOR input has a resolution of 76µV/bit and a single pass noise floor of 1mV. With a 1mV noise floor on a single pass, the Premier II with the Asylum SA AFM sensitivity has a useful displacement resolution of approximately 0.7Å. The effect of the noise floor is reduced by a factor of four or more using averaging. (See the document Displacement Sensor Operations from Radiant Technologies, Inc. for more information.) The resolution limit calculated in the previous paragraph is that of the Premier II alone. The sensor, in this case an AFM, adds its own noise to the measurement. This particular AFM generated about 15Å to 20Å of noise peak-to-peak, a value much larger than expected for an Asylum AFM. No doubt, the use of this instrument as a test bed by Asylum contributed to its out-of-specification noise level. Radiant has a Premier II at its office abused in a similar manner. Roger attached a Stanford Research Systems programmable filter on the output of the AFM set for 1KHz ( www.thinksrs.com ). This modification significantly reduced the noise level of the AFM signal to 3Å peak-to-peak. Averaging x10 of the 1Hz measurement reduced the noise level to approximately 1Å peak-to-peak. The Precision Premier II and Precision LC generate a 5V SYNC signal when they are capturing data. (Again, see the document Displacement Sensor Operations from Radiant Technologies, Inc. for more information.) The SYNC signal is accessible at a BNC on the rear face of the testers. With some sensors, the SYNC signal can be used to coordinate the operations of the tester and the sensor. While the SA AFM does allow controlling digital inputs, it was not feasible to use the SYNC signal to control the Z integration function on the SA for our measurements this time. We thus had two options: 1) Put a PAUSE task in the program at every point where the AFM must be manually manipulated prior to a measurement loop being executed or Property of Radiant Technologies, Inc. 1

2) Free the AFM prior to starting the test program on the Premier II with the knowledge that the AFM output will remain stable for the duration of the program execution. The integrator on the SA AFM was stable long enough for us to take 10 sequential loops under program control on the Premier II so we chose option 2. Sample Description: The data in this report were acquired from a Radiant Type AC capacitor. The Type AC die has 1µ of 4/20/80 PNZT on a 1500Å platinum bottom electrode. The top platinum electrode is 1000Å thick. It is covered by 800Å of PNZT followed by 400Å of TiOx and 2300Å of silicon dioxide. The metal connections are in chrome/gold. The dimensions of the targeted capacitor were 80µ x 125µ. Both of these dimensions are less than the 550µ thickness of the silicon substrate so there should be little bending moment of the substrate to contribute to an amplified piezoelectric piston movement. NOTE: 1µ thick 4/20/80 PNZT will bend a 4 diameter, 550µ thick <100> silicon wafer 1000Å at a distance of 5 millimeters from the center of the capacitor. Bending moments are important to consider when measuring converse piezoelectric constants. This particular die had a well etched under the left-hand side of the die, creating a membrane capacitor. We measured the displacement of the 10,000µ 2 capacitor on the right-hand side of the die as marked by the red dots in Figures 1 and 2 below. The measured capacitor was on solid silicon. Figure 1 Type AC Die without Well Figure 2 The Tested Die in Package For the Asylum visit, Carl Montross at Radiant prepared the Radiant Type AC die on a TO-18 header without a lid. At Asylum, Roger Proksch mounted the package on a blank electronic perf board and soldered small-gauge wires to the package leads. He placed the board on the test frame of the AFM and placed the AFM probe head over the sample. I connected the capacitor leads to the Premier II using minigrabbers at the end of coax cables. Photographs of the AFM with the sample are shown below. Property of Radiant Technologies, Inc. 2

Perf Board and Sample on the AFM Base Figure 3 Close-up of Sample on the Perf Board with Wires Figure 4 Property of Radiant Technologies, Inc. 3

The AFM Probe Head over the Sample Figure 5 The Sample under the AFM Probe Head Figure 6 Once the sample was in place and the AFM test fixture assembled, Roger used a built-in camera and motorized X/Y stage to position the AFM probe tip in the center of the capacitor to be tested. Property of Radiant Technologies, Inc. 4

Parasitics Affecting Atomic Force Microscopes: Atomic Force Microscopes can measure extremely small displacements. Nevertheless, they are subject to several unique parasitic effects which can distort the results. 60Hz/50Hz/Control System Electrical Noise: Atomic force microscopes are physically large systems with high-voltage control circuits. When connecting a Precision tester to a sample mounted in an AFM, it is highly likely that the RETURN lead will pick up 60Hz or 50Hz from the AFM power supply as well as EMF from the high voltage signal that controls the Z-position of the AFM stage. Coax cables should be used to connect the DRIVE and RETURN BNCs of the tester to the sample. The sample should be isolated electrically from the AFM chuck and the AFM chuck should be grounded. Finally, and most important, the shields of the coax cables from the tester must be connected to the frame ground of the AFM. This connection makes the EMF noise common mode for both machines, allowing the tester measurement circuits to subtract it out automatically. If all else fails, connect both the tester and the AFM to the same wall socket! Mechanical Amplification The purpose of this experiment was to measure the piston displacement of the surface of the thin ferroelectric film capacitor. The total displacement of the capacitor surface should be in the range of 20Å. One micron of PZT is capable of bending a 550µ thick silicon wafer, overwhelming the piston displacement of the capacitor. To avoid this effect, it is best to measure a capacitor with dimensions much smaller than the thickness of the substrate thereby reducing the length of the lever arm which the capacitor can bend. Additionally, placing the measurement point as close as possible to the geometric center of the capacitor top electrode where the capacitor cups further reduces amplification effects. You can have the AFM scan the capacitor surface to find the geometric center for you. In this experiment, I measured the center of a capacitor with dimensions of 80µ by 125µ on a 550µ thick silicon substrate. The capacitor was large enough in dimensions that it could have shown slight amplification at the outer edges of the capacitor top electrode. An additional factor here was that the die was epoxied to the package header, further reducing its ability to move when the capacitor was actuated. Such clamping of the sample substrate helps reduce the lever arm of the mechanical amplification but it does not totally eliminate the possibility of wrinkling by the substrate while under restraint. Were my purpose to determine the converse piezoelectric constant for the ferroelectric material, I would measure a much smaller capacitor, possibly 10µ x 10µ or even 5µ x 5µ. Since the ferroelectric film in this case is 1µ thick, measuring the displacement of a capacitor with the dimensions of 1µ on a side could lead to other errors because fringe electric fields between the electrodes could contribute to the sample response. Starting Point Alignment All test fixtures, not just AFMs, can have drift in the Z-position of the frame holding the test fixture. Consequently, the first point of each sequential measurement might end up at a different location in Z space, causing inaccurate reproduction of the measured signal when averaging multiple measurements. A solution to this problem is to mathematically zero the first point of each sequential measurement before averaging. A measurement is zero d by subtracting the value of the first point from all points in that measurement, causing a translation along the Y-axis. Maximum Sensitivity Normally, AFMs acquire the location of the AFM probe tip in space and then move the chuck piezoelectrically to maintain the prove tip in a fixed position in space. The frequency response of this control circuit is usually in the hundreds of Hertz so measurements of the butterfly loop taken with the control signal active must be taken at 1Hz or slower. When in this mode, the error signal from the Z Property of Radiant Technologies, Inc. 5

control system is the desired source of information for the tester SENSOR input and it may need to be integrated using the Single-Trace Math filter in the Vision Library. For maximum sensitivity and frequency response, disable the control system and feed the output of the position sensors directly to the tester SENSOR input. The position information is very fast, in the hundreds of kilohertz range, and it is a direct measure of the position of the AFM probe tip in space. Theoretically, this free position signal is the best one to use for measuring the large signal butterfly loop. The problem with this approach is that without the control system actively controlling the position of the chuck, the AFM chuck will begin a relatively constant-rate drift in one direction or the other until it hits an internal stop. Please note that all AFMs have this issue, not just the Asylum instruments. This drift is the source of the Z-drift error described next so if you use this mode of operation, be prepared to characterize the rate of Z-drift and subtract it from your results. Z-drift When using an AFM at maximum sensitivity in the free mode, the constant drift rate of the chuck will add to the displacement data. The amount of drift differs from machine to machine and from day to day. This effect will not only cause the starting point of each measurement to be different, as noted earlier, but will also cause a slope in the resulting measurement so the starting point and ending point of a butterfly loop do not line up. We captured a classic example of this effect on the SA AFM as shown in Figure 7 below. The data in Figure 7 is the same as that analyzed in the next section of this report. Figure 7 is a plot of the absolute measured values of 10 sequential 1-second butterfly loops on the SA AFM after the AFM control system was turned off. Displacement [ Type AC Solid ] HYST 1s 20v: 1 HYST 1s 20v: 2 HYST 1s 20v: 3 HYST 1s 20v: 4 HYST 1s 20v: 5 120 HYST 1s 20v: 6 HYST 1s 20v: 7 HYST 1s 20v: 8 HYST 1s 20v: 9 HYST 1s 20v: 10 110 100 90 80 70 60 50 Z D R I F T 40 30-20 -15-10 -5 0 5 10 15 20 Volts Sequential Measurements of the Same Sample on the SA AFM Property of Radiant Technologies, Inc. 6

Figure 7 The drift rate was high enough that it introduced a slope to each butterfly. Two of the loops are shown in Figure 8 as raw data. The drift between the first and last points for each butterfly loop is readily apparent. 42.5 Displacement [ Type AC Solid ] HYST 1s 20v: 1 HYST 1s 20v: 2 40.0 37.5 35.0 32.5 30.0-20 -15-10 -5 0 5 10 15 20 Volts Two of the Loops from Figure 7 Showing Drift Effects Figure 8 The distortion of the measured butterfly loops can be corrected by mathematically introducing a drift of equal magnitude in the opposite direction. To accomplish this requires careful characterization of the AFM over a time period equivalent to or longer than that required to capture all of the measurements to be averaged. Luckily, each measurement record in the Vision Archive contains a time stamp. I copied the time stamp for each measurement in Figure 7 into a spreadsheet along with the value of the first point of each loop in Figure 7. I found that the drift rate for this particular AFM was consistently 0.54 nanometers, or 5.4Å, on each test loop of 1 second. Each test consisted of 1017 points, a number also accessible from the Vision Archive, so the drift rate was 0.00053nm per measurement point. With this information, it was relatively easy in the spreadsheet to construct a sloped line to subtract from each measurement in order to eliminate the Z-drift distortion. Correcting Systematic Distortions At present, the PIEZO task in Vision will accept a Z-drift value from the operator and subtract that drift from the measurement before it is stored in the Archive. PIEZO will also plot individual measurements with the data translated so the first point is zero. The PIEZO SENSOR data cannot be readily submitted to a loop averaging task. Radiant is presently constructing two filters for the Vision Library that will allow Property of Radiant Technologies, Inc. 7

the correction operations to be performed on PIEZO SENSOR data as well as on other measurement tasks before the data is sent to a loop averaging filter. The new tasks will also provide a smoothing filter. Results: The test consisted of 10 sequential 20V hysteresis loops in a row with 1s periods. The SRS programmable filter was connected between the AFM Z output and the SENSOR input. The filter cutoff frequency was set to 1kHz. Being 1000 times faster than the measurement period, the filter introduced negligible phase distortion to the final results. The absolute values of the ten measurements have already been shown above in Figure 7. I exported the ten loops to a spread sheet where I 1) moved each measurement vertically so that each starting point was 0, 2) calculated the Z-drift rate and subtracted it from each loop, 3) averaged the ten loops, and 4) applied a smoothing filter to the averaged butterfly loop. As described earlier in the Test Setup section of this report, we did not use the SYNC signal of the Premier II to coordinate the tester and the AFM operations. Instead, Roger disabled the control system and fed the output of the position sensors directly to the tester. Roger found that the chuck on the SA AFM would take several minutes to hit its stop after the control system was turned off, long enough for the Premier II to execute at least 10 measurement loops for averaging. Therefore, the procedure was for me to prepare to run the test program and, just prior to execution, for Roger to turn off the Z-position control system to release the chuck. After the program execution, Roger would turn on the control system in preparation for the next test. The raw polarization and displacement data for one of the ten measurements of this sample is shown below: Property of Radiant Technologies, Inc. 8

75 Hysteresis Data Polarization (µc/cm2) 50 25 0-25 -50-75 33.0 32.5 32.0 31.5 31.0 30.5-20 -15-10 -5 0 5 10 15 20 Voltage Raw Measurement of Polarization and Piston Displacement of Type AC Capacitor Figure 9 Note that the negative saturation point of the polarization loop shows leakage. The voltage at which the leakage started was three times the negative coercive voltage for the sample. This is not unexpected for a thin PZT capacitor when exposed to strong electric fields at the high illumination levels typical in an AFM. Also note that the leakage of the capacitor did not appear to affect the shape of the butterfly loop. Finally, you can see the discontinuity between the butterfly loop start/stop points due to the Z-drift. The plot below shows the ten loops after zeroing and correction for Z-drift. Property of Radiant Technologies, Inc. 9

Displacement - All Loops [ Corrected for Z-drift ] 3 2 2 1 1 0-30 -20-10 -1 0 10 20 30-1 -2 d1 d2 d3 d4 d5 d6 d7 d8 d9 d10 Volts All Loops Zero d and Corrected for Z-drift Figure 10 Property of Radiant Technologies, Inc. 10

Figure 11 shows the average of the loops in Figure 10. Displacement - Averaged [ Corrected for Z-drift ] 2.0 1.5 1.0 0.5 0.0-30 -20-10 0 10 20 30-0.5-1.0 Average Corrected Volts Averaged Butterfly Loop Corrected for Distortions Figure 11 For comparison, I plot below the zero d and averaged butterfly loop without the Z-drift correction. Displacement - Averaged [ Uncorrected for Z-drift ] 75.0 74.5 74.0 73.5 73.0 72.5 72.0 71.5 Average Uncorrected -30-20 -10 0 10 20 30 Volts Butterfly Loop Uncorrected for Z-drift Figure 12 Property of Radiant Technologies, Inc. 11

Finally, Figure 13, the lucky figure, shows the butterfly loop of Figure 11 after the application of a triangular 9-point smoothing filter. Displacement - Averaged [ Smoothed and Z-corrected ] 2.0 1.5 1.0 0.5 0.0-30 -20-10 0 10 20 30-0.5-1.0 Smoothed Average Nice! Volts The Butterfly Loop Figure 13 Conclusion The piston-only butterfly displacement loop of a thin ferroelectric film capacitor can be measured accurately with an atomic force microscope. Given the resolution of AFMs, it is possible to capture the displacement of the surface of the capacitor to a resolution less than 1Å. The resolution can be further improved to almost 0.1Å with a smoothing filter. The parasitic effect of mechanical amplification may be significantly reduced by measuring capacitors with very small areas. Radiant is preparing new filter tasks for the Vision Library to be used in displacement measurements to correct acquired data for parasitic distortions from attached displacement sensors. Property of Radiant Technologies, Inc. 12