SmartLine Devices FDM Offline Configuration User s Manual 34-CT Revision 1.0 July 2017

Similar documents
900 ControlEdge Remote Termination Panel (RTP) for:

SmartLine Multivariable Transmitter SMV800 Quick Start Installation Guide

HART Mode. Users Guide. July Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.

STT850 SmartLine Temperature Transmitter Quick Start Guide

Field Device Manager Express

MI June Instruction. I/A Series Intelligent Pressure Transmitters. Operation, Configuration, and Calibration Using a HART Communicator

RTT20 I/A Series Temperature Transmitter

NWS300 Field Type Smart Meter

Wireless Acoustic Transmitters Series XYR 5000, Model WN571 Specifications - Americas

SMARTLINE LEVEL SLG700

815LT Submersible Smart Level Transmitter

4590 Tank Side Monitor. Service Manual. Mark/Space Communication Protocol. Software Version v2.03 SRM009FVAE0808

STT700 SmartLine Temperature Transmitter User s Manual 34-TT Revision 2 February 2018

Rosemount 5408 and 5408:SIS Level Transmitters

27 June 2017 Peter Overgaauw BENEFITS OF INTELLIGENT FIELD DEVICE MANAGEMENT

2012 Honeywell Users Group EMEA Tony Alexander Instrument Asset Management

815PT Smart Pressure Transmitter

Calibration Instruction Manual Emerson 475 Field Communicator. Manual Revision FC.2

Wireless Transmitter Series XYR 6000 Absolute Pressure Model STAW94L 0 to 500 psia 0 to 35 bara Specifications

Sustain.Ability. Kevin Boatright & John Yingst Field Device Manager: Instrument Nourishment through Smart Device Management

Instruction Manual ABM HART Gateway Software. Instruction Manual Revision A.1

Rosemount 333 HART Tri-Loop

TEK-TROL HART GATEWAY SOFTWARE. Operating Instruction Manual.

FLXA402 4-Wire Converter HART Communication. FLEXA, FLXA, SENCOM and FieldMate are registered trademarks of Yokogawa Electric Corporation.

Technical Bulletin, Communicating with Honeywell TM ST3000/STT3000 Smart Transmitters

Honeywell s Experion PKS and Field Device Manager Provide Award-winning HART Solution for Shell Upgrader

Smart Instruments Graduate Top of their Class with FDM Peter Overgaauw, Honeywell

XYR 5000 Wireless Temperature Transmitters Model WT530 Specifications - Europe

AN UCODE I2C PCB antenna reference designs. Application note COMPANY PUBLIC. Rev October Document information

SFC Smart Field Communicator Model STS103. Specification 34-ST /24/06. Description. Function. Page 1 of 5

Global Water Instrumentation, Inc.

STT Series STT250 Smart Temperature Transmitter

Micro Motion Model 2700 Transmitters with Configurable Input/Outputs 2700***B 2700***C. Configuration and Use Manual Supplement

Micro Motion Model 1700 Transmitters with Analog Outputs 1700***A. Configuration and Use Manual Supplement. Configuration Operation Maintenance

Contents. Overview Introduction...3 Capabilities...3 Operating Instructions Installation...4 Settings... 5

AT200 External Mount Magnetostrictive Level Transmitter

Aimetis Outdoor Object Tracker. 2.0 User Guide

STT850 SmartLine Temperature Transmitter User s Manual 34-TT Revision 5 November 2017

THYRISTORS T3-3PH from 35A to 2700A Specifications Universal 3 Phase Thyristor Unit , March 2012

EchoSonic II Ultrasonic Level Transmitter LU23, LU28 & LU29 Series Quick Start NEMA 4X Enclosure

SLG 700 SmartLine Level Transmitter Guided Wave Radar FOUNDATION TM Fieldbus Option Manual

Global Navigation Satellite System for IE 5000

Wireless Pressure Transmitter User Manual Americas

ST 3000 Fieldbus Transmitter Series 100 Models

Instruction Manual TRICOR TCE HART Communication TCM COHA_S_EN_161220_E002

DeltaV SIS Logic Solver

INSTRUCTION MANUAL FOR ULTRASONIC/MICROWAVE SENSORS

Smith Meter AccuLoad. Electronic Preset Delivery System. Tank Proving Guide. Bulletin MN Issue/Rev. 0.1 (7/10)

SAP Dynamic Edge Processing IoT Edge Console - Administration Guide Version 2.0 FP01

EchoSonic II Ultrasonic Level Transmitter LU27 Series Quick Start NEMA 4X Enclosure

Automation 2013 Mumbai 2013 Fieldbus Foundation

Rosemount 248 Wireless Temperature Transmitter

RAZER GOLIATHUS CHROMA

IVI STEP TYPES. Contents

User manual Automatic Material Alignment Beta 2

CompoNet Gateway Unit for CC-Link

Accessories for EST4 Series

Lifetime Power Energy Harvesting Development Kit for Wireless Sensors User s Manual - featuring PIC MCU with extreme Low Power (XLP) Technology

Uplink 5500EZ. Installation and User Guide. S e pte m be r 1 2,

KGX 150/130 ADS -B Certified Transceivers & Receivers

VITO LT. Global Experience. Locally Applied. A dependable choice.

MINIMUM SYSTEM REQUIREMENTS

Rosemount 648 Wireless Temperature Transmitter

Smart Radar Level Transmitter ALT6200 Overview Features Specification

Rosemount 3101, 3102, and 3105

Quick start guide. Océ Touchstone Extensions

ROC800-Series HART Module

LD2342 USWM V1.6. LD2342 V1.4 Page 1 of 18

General Specifications

LT3000 ULTRASONIC LEVEL TRANSMITTER

Remote Monitoring of Environmental Sites using Solar Powered Wi-Fi Enabled Systems

Model 144H NEW. PC-Programmable Temperature Transmitter. Product Data Sheet

User s Guide. Model PX2088 Pressure Transmitter. PX2088 Pressure Transmitter

User s Manual DPharp HART 5/HART 7 Communication Type (EJX A, EJA E)

KENCO Loop Powered Magnetostrictive Transmitter. KMD SERIES Operation and Installation Manual. Table of Contents

CrystalControl User s Manual. for nvision Reference Recorder and HPC40 Series Pressure calibrator

SmartLine. STD700 SmartLine Differential Pressure Specification 34-ST Technical Information. Introduction

USER MANUAL. EPP Intelligent Positioner Control Unit 1/22.

FLOW SWITCH 600 Series Velocity Flow Sensor. Instruction Manual

Rosemount 848L Logic Transmitter with FOUNDATION Fieldbus

F210 Vision Sensor Flow Menus and Macro Capability

TRBOnet Guard Tour Configuration and Operation Guide

SRT Marine Technology. LD2342 V1.4 Page 1 of 22

User s Guide RF Shield. User s guide Version boosting wireless efficiency

Servo Gauge 854 XTG. Global Experience. Locally Applied. Cut costs, not repeatability.

AN Extended Range Proximity with SMSC RightTouch Capacitive Sensors

TS16 - Pre-Instructional Survey

Rosemount Specifications. Product Data Sheet , Rev BA February 2009 PERFORMANCE SPECIFICATIONS

SITRANS L Level instruments

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

Quad single-pole single-throw analog switch

Rosemount 3300 Series Guided Wave Radar Level & Interface Transmitter

Micro Motion Filling Mass Transmitters with Modbus

AUTOTUNE USER GUIDE. R8000 Series Communications Systems Analyzer

CCE Image may differ from the actual product By Martin Labbé, eng., Jasmin Goupil & Louis Perreault

Rosemount 5400 Series Radar Level Transmitter

UM User manual for di2c demo board. Document information

Rosemount 753R Remote Web Based Monitoring Indicator

Positioner with HART Communication Type 3780

Best Practices for Power and Transient Protection on Rosemount Radar Transmitters

Transcription:

SmartLine Devices FDM Offline Configuration User s Manual 34-CT-25-01 Revision 1.0 July 2017 Honeywell Process Solutions

Copyrights, Notices and Trademarks Copyright 2017 by Honeywell International Revision 1.0, July 2017 While the information in this document is presented in good faith and believed to be accurate, Honeywell disclaims any implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in the written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special, or consequential damages. The information and specifications in this document are subject to change without notice. While the information in this document is presented in good faith and believed to be accurate, Honeywell disclaims any implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in the written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special, or consequential damages. The information and specifications in this document are subject to change without notice. Honeywell Process Solutions 1250 W Sam Houston Pkwy S Houston, TX 77042 Page ii FI LEAP Offline Configuration Tool User s Manual Revision 1

About This Manual Honeywell s LEAP for FI is a program with the goal of increasing the effectiveness of project execution and operations by making SmartLine instruments a key enabler to provide enhanced end to end user experience from Field Instruments to control room. LEAP refers to Lean Execution of Automation Projects This document lists the offline configuration guidelines for the user who creates the FDM Offline Configuration that will be used in Experion Automated Device commissioning. This Document contains the guidelines for the SmartLine devices: SLG700, ST800 / 700, STT850/750/STT700 and SMV800 Revision History FI LEAP Offline Configuration Tool User s Manual, Document #34-CT-25-01 Rev. 1.0 July 2017 First release (R100) Support and Contact Information For Europe, Asia Pacific, North and South America contact details, refer to the back page of this manual or the appropriate Honeywell Support web site: Honeywell Corporate www.honeywell.com Honeywell Process Solutions https://www.honeywellprocess.com/* Telephone and Email Contacts Area Organization Phone Number United States and Canada Global Email Support Honeywell Inc. Honeywell Process Solutions 1-800-343-0228 Customer Service 1-800-423-9883 Global Technical Support ask-ssc@honeywell.com Revision 1 FI LEAP Offline Configuration Tool User s Manual Page iii

Definitions, Acronyms, and Abbreviations The following are of special significance to this version of the document Term Definition ADC AVT BF (Brownfield) Cloud Design Tool CDA DCS DD DTM EDDL End User EPC EPKS EPKS STAC Project FDI FDM FDT FI GF (Greenfield) HART 7.x HART Protocol LEAP Automated Device Commissioning Application and Validation Tool. Also referred as Cloud Design tool in this document Brownfield is a term used in urban planning to describe land previously used for industrial purposes or some commercial uses. Same as AVT Control Data Access: proprietary protocol between embedded process nodes (C300, FIM) and Experion system Distributed Control System (DCS) is the process control system that accepts the output of transmitters and provides output to the final control elements in the field. A Device Description (DD) provides an extended description of each device and includes information needed for a control system or host to understand the meaning of data in the device. Device Type Manager: A driver that plugs into a host system that allows a user to configure devices without another configuration device. Electronic Device Description Language (EDDL): Allows users to interact with their intelligent devices in new ways, including graphs, charts and calculations, among other benefits. End customer using Honeywell control and Instrumentation offering Engineering, Procurement, Construction Experion Process Knowledge System (EPKS) is Honeywell s flagship Distributed control system Experion STAC project is part of the larger LEAP program. LEAP is based on UIO, Virtualization and VEP targeted to the System Test and Commissioning phase of the project. Field Device Integration Technology that combines the benefits of Device Description by means of UID (User Interface Descriptor) and the DTM by means of Optional UIP (User Interface Plugin) Field Device Manager (Experion): The configuration Tool for Experion Field Device Tool: An application program technology for a host that allows a user to navigate between devices for configuration purposes using a device driver called DTM Field Instrumentation Greenfield (GF) is a description of a customer order segment type. Greenfield orders refer to orders that are placed from customer plants that are newly constructed. HART release 7.x Specification, includes Wireless HART option HART Communications Protocol: Created by Rosemount and now supported by the FieldComm Group, previously, HART Communications Foundation (HCF). Lean Execution of Automation Projects Page iv FI LEAP Offline Configuration Tool User s Manual Revision 1

MTC Parameter Selected In Offline view SmartLine device or Transmitter STAC VEP Minimum To Compete This means, in FDM offline view, only if the checkbox next to the parameter is checked, this parameter will be included in the offline download. If this parameter is not Selected, and is part of a multiple parameters command, then the host will use the last value read from the device for that parameter while sending the relevant command to the device. Honeywell Pressure, Temperature, Level and Multivariable Transmitters Only HART devices are in scope for this project Smart Test and Automated Commissioning Virtual Engineering Platform that supports Testing virtually Revision 1 FI LEAP Offline Configuration Tool User s Manual Page v

Contents 1 Introduction... 1 Overview... 1 FDM Offline Configuration overview... 1 Offline Configuration Creation / Import / Export... 1 Common parameters to unselect for all SmartLine... 4 SLG700 device... 5 Unit Relations... 5 Power Cycle requirement on the device on downloading some parameters... 8 Reflection Model parameters and Measurement related parameters... 9 Mounting Location and dependent parameters... 10 Probe Type and dependent parameters... 10 Latching Mode or Latching Alarm... 10 ST 700 standard and ST 700 basic Models... 11 ST 700 Standard and Basic models... 11 STT850/750/700... 13 Latching... 13 Parameters that can be downloaded from Offline template... 13 SMV800... 15 Parameters that can be downloaded from Offline template... 15 2 Security... 17 How to report a security vulnerability... 17 List of Figures Figure 1 - FDM Offline User Interface... 1 Figure 2 - FDM Offline parameter listing for selected device type... 2 Figure 3 - FDM Offline parameter editing, and saving... 3 List of Tables Table 1 - Common Parameters across SmartLine devices... 4 Table 2 - Parameter Relationships... 5 Table 3 Parameters that require device power cycling when updated... 8 Table 4 Reflection model parameters... 9 Table 5 Mounting Location and dependent parameters... 10 Table 6 - Probe Type and dependent parameters... 10 Table 7 ST 700 Pressure device model, Display type and dependent parameters... 11 Page vi FI LEAP Offline Configuration Tool User s Manual Revision 1

1 Introduction 1.1 Overview This document assumes that the User is familiar with the FDM Tool, Online and Offline Configuration, Creation of Templates, Import and Export. Section 1.2 gives a brief overview of FDM offline feature. Section 1.2.2 and above provide guidelines to follow for individual SmartLine devices while creating Offline Configuration Templates. 1.2 FDM Offline Configuration overview 1.2.1 Offline Configuration Creation / Import / Export After the FDM tool is Installed, Launch FDM. Select the Offline View, right click on the HART Node, and select Create Configuration Figure 1 - FDM Offline User Interface Revision 1.0 FI LEAP Offline Configuration Tool User s Manual Page 1

Select the Protocol, Manufacturer, Device Type, Device Revision for this device, select Create Figure 2 - FDM Offline parameter listing for selected device type Page 2 FI LEAP Offline Configuration Tool User s Manual Revision 1

Offline configuration screen areas are labelled below: FDM Offline User Interface Help Text PRE EDIT ACTION Download, Save buttons Figure 3 - FDM Offline parameter editing, and saving Make sure you select only needed parameters when creating FDM Offline Templates. If you use the Select All feature in Offline Template creation, make sure to unselect the listed parameters per the Design requirement of the individual SmartLine device as detailed in section 1.2.2 and above. Help Text: Hover the mouse over the parameter to see brief help Download and Save buttons: In Offline mode, the selected configuration can be saved to the database that can be exported to an external file. This file is the Input for one of the steps in Experion Auto Device Commissioning. In Online mode, selected configuration can be downloaded to the connected device by selecting Download button. Based on the Context whether the mode is Online or Offline, the button shows Download or Save. PRE EDIT ACTION: On some parameters a helpful tip will appear to guide the user on accepted values for that parameter. Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 3

1.2.2 Common parameters to unselect for all SmartLine Table 1 - Common Parameters across SmartLine devices Parameter to Unselect in Offline Configuration Long Tag Loop current Details Factory ordered device should already have the Long Tag name programmed into it. Experion Channel Name would already match this Long Tag Name. This will help automatically match the channel name with the Long Tag name in the device during Auto Device Commissioning. Selecting this parameter in offline Template will download this parameter and thus, will overwrite the device Long Tag. If the new Log Tag is different from the Channel name, then the device will not be found during Commissioning step. Loop Current parameter used during Loop Test / Output Mode to set the value other than 4 or 20mA. Device will reject this parameter download in other scenarios. Install Date Transmitter Install Date Pressure Sensor Install Date Temperature Sensor Install Date Communication Module Install Date This parameter is Writable only once (usually factory configures it to start accruing advanced diagnostics). So in the field user cannot configure this anymore. Write of this parameter 2nd time on will fail. Tooltip provided on this parameter. Note: Depending upon the SmartLine device type, any or all of the above Install Dates might be supported. Unselect them all. Poll Addr During auto detection of device in ADC, polling address has to be 0. Unselect this parameter from inadvertently setting a different polling address to the device during download. % Unit Unselect this parameter in Offline Template. This unit is applicable for download % type of device variables, and unit itself is not writable to the device. However, Unit relationship with device variables makes this parameter show up in Offline. Page 4 FI LEAP Offline Configuration Tool User s Manual Revision 1

1.3 SLG700 device Follow the guidelines below while configuring the device specific offline parameters 1.3.1 Unit Relations When changing any or all of the parameters listed below, follow the hierarchy as shown below and make sure valid selections are made or valid values are entered: Measured Product o Volume Calc. Type Change the units if needed. Mapped variables to PV, SV, TV, QV will inherit one of the units listed here based on the type of the mapped variables Length Unit or Volume Unit or Velocity Unit o o o PV LRV (for the mapped PV) PV URV (for the mapped PV) Allowed selections for the above listed parameters are given in Table 2. Note: If you are not sure about mapped PV, unselect PV LRV, URV in offline template to avoid any out of range values during download. The Unit of Measurement for PV LRV, URV will not be downloaded regardless of what variable is mapped to the PV. Unit 'ft (feet)' appearing next to PV LRV, PV URV is just an indication that the default length unit is ft (feet), but this unit is not downloaded to the device therefore the user should ignore this unit while entering PV LRV, URV values. Note that Table 2 includes the changing parameter lists based on Measured Product and Volume Calculation parameter selections. Table 2 - Parameter Relationships Measured Product Volume Calculation Type Variables that are allowed for mapping to PV, SV, TV, QV (Note that variable mapping is not supported for download from Offline Template) URV, LRV of mapped PV Single Liquid None Product Level Must be within the upper/ lower limits of mapped variable Product Level % Distance To Product Product Level Rate Vapor Thickness Vapor Thickness % Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 5

Measured Product Volume Calculation Type Variables that are allowed for mapping to PV, SV, TV, QV (Note that variable mapping is not supported for download from Offline Template) URV, LRV of mapped PV Single Liquid Ideal Tank Shape or Strapping Table Calculation Product Level Must be within the upper / lower limits of mapped variable Product Level % Distance To Product Product Level Rate Vapor Thickness Vapor Thickness % Product Volume Vapor Volume Two Liquids, Flooded None Interface Level Must be within the upper / lower limits of mapped variable Interface Level % Distance To Interface Interface Level Rate Upper Product Thickness Two Liquids, Flooded Ideal Tank Shape Or Strapping Table Calculation Interface Level Must be within the upper / lower limits of mapped variable Interface Level % Distance To Interface Interface Level Rate Upper Product Thickness Lower Product Volume Upper Product Volume Page 6 FI LEAP Offline Configuration Tool User s Manual Revision 1

Measured Product Volume Calculation Type Variables that are allowed for mapping to PV, SV, TV, QV (Note that variable mapping is not supported for download from Offline Template) URV, LRV of mapped PV Two Liquids, non-flooded None Product Level Must be within the upper / lower limits of mapped variable Product Level % Distance To Product Product Level Rate Vapor Thickness Vapor Thickness % Interface Level Interface Level % Distance To Interface Interface Level Rate Upper Product Thickness Two Liquids, non-flooded Ideal Tank Shape Or Strapping Table Calculation Product Level Must be within the upper / lower limits of mapped variable Product Level % Distance To Product Product Level Rate Vapor Thickness Vapor Thickness % Interface Level Interface Level % Distance To Interface Interface Level Rate Upper Product Thickness Product Volume Vapor Volume Lower product volume Upper Product Volume Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 7

1.3.2 Power Cycle requirement on the device on downloading some parameters Parameters with * : Power cycle required after writing these parameters to the device. So, if the Offline Template had these parameters selected, after the Configure Step in ADC, device power cycle is required to clear the Critical status. Device will be in burnout (output at 21.5mA and all device variables reading NaN). If it is not required to change these parameters during ADC Configure step, unselect these parameters while creating the offline template. Parameters are listed below: Table 3 Parameters that require device power cycling when updated Measured Product(*) Vapor DC (*R) Upper Product DC (*R) Lower Product DC(*) Mounting Height(*) Mounting Location(*) Mounting Diameter(*) Mounting Angle (*) Probe Type(*) Probe End Type(*) Centering Disc Material(*) Centering Disc Diameter(*) If you have to make changes to the parameters listed in section 1.3.3 in addition to the parameters in section 1.3.2, follow the sequence below. But, do not download the parameters in section 1.3.2 and 1.3.3 together in one download. Note1: Parameters in section 1.3.3 should not be changed/sent unless you have a known valid set from an installed transmitter on the same process Follow this sequence: 1. Change any of the parameters in section 1.3.2 2. Download parameters to the device 3. Power cycle the device 4. Change any of the parameters in section 1.3.3 (Follow Note 1. If not sure, do not perform step 4 and 5) 5. Download parameters to the device Page 8 FI LEAP Offline Configuration Tool User s Manual Revision 1

1.3.3 Reflection Model parameters and Measurement related parameters Select the below listed parameters for download only under special circumstances where you need to preserve non-default values. These should not be changed/sent unless you have a known valid set from an installed transmitter on the same process. Reference Model Width Table 4 Reflection model parameters Reference Model Attenuation Reference Model Gain Reference Model Objective Threshold Process Connector Model Width Process Connector Model Attenuation Process Connector Model Gain Process Connector Model Objective Threshold Surface Model Width Surface Model Attenuation Surface Model Gain Surface Model Objective Threshold Interface Model Width Interface Model Attenuation Interface Model Gain Interface Model Objective Threshold End Of Probe Model Width End Of Probe Model Attenuation End Of Probe Model Gain End Of Probe Model Objective Threshold When the user downloads any of the parameters listed in section 1.3.2, sensor automatically calculates the Reflection parameters and user does not need to adjust the above parameter values manually. Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 9

1.3.4 Mounting Location and dependent parameters Based on the Mounting Location, relevant dependent parameters will be made Read only or Read Write as applicable. Table 5 Mounting Location and dependent parameters Mounting Location Tank Bracket Nozzle Bypass Still Well Dependent parameters that will be Read/Write Mounting Angle Mounting Angle Mounting Angle Mounting height Mounting diameter Mounting Angle Mounting height Mounting diameter Mounting Angle Mounting height Mounting diameter 1.3.5 Probe Type and dependent parameters Based on the Probe Type, Centering Disk Material, Disk Diameter is made Read only or Read/Write as applicable. Table 6 - Probe Type and dependent parameters Probe Type Centering Disk Material Dependent parameters Wire, Rod, Custom, Multi Twist Wire None Centering disk diameter is N/A, so this will be shown as read only in Offline configuration 316/316L Stainless Steel Centering disk diameter applicable PTFE C-276 Nickel Alloy Coax All disk materials N/A Centering disk diameter N/A, so this will be shown as read only in Offline configuration 1.3.6 Latching Mode or Latching Alarm Set Latching mode parameter value to: Non-Latching so that during the configuration download any device statuses are not Latched. Page 10 FI LEAP Offline Configuration Tool User s Manual Revision 1

1.4 ST 700 standard and ST 700 basic Models 1.4.1 ST 700 Standard and Basic models Follow the table below to select the right Display Type, Meter Body (MB) Type and dependent parameters based on the ST 700 model NOTE: The distinctions between 2 display models here: ST 700 Standard Model (HART Device Revision 5): Supports Basic Display with 3 External buttons or Standard Display with 2 internal or 2 external buttons ST 700 Basic Model (HART Device Revision 4): Supports Standard Display with 2 internal or 2 external buttons. Table 7 ST 700 Pressure device model, Display type and dependent parameters For either of ST 700 models, Advanced Display is not applicable ST 700 Model Display Type (None, Basic, Standard) MB Type Screen Format PV Selection (Disp. PV Type) Units and related parameters PV Scaling ST 700 Standard Rev 5 & ST 700 Basic Rev 4 None N/A N/A N/A N/A N/A Pressure Pressure units Linear Percent Output % (auto selected, nothing to configure in Offline) N/A DP PV Loop Output ma (auto selected, nothing to configure in Offline) N/A ST 700 Standard Rev 5 Basic Display MBT degc N/A Static Pressure Psi N/A Pressure Pressure Units Linear AP or GP PV Percent Output Loop Output % (auto selected, nothing to configure in Offline) ma (auto selected, nothing to configure in Offline) N/A N/A MBT degc N/A Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 11

ST 700 Model Display Type (None, Basic, Standard) MB Type Screen Format PV Selection (Disp. PV Type) Units and related parameters PV Scaling Pressure Pressure Units N/A Percent Output % (auto selected, nothing to configure in Offline) N/A ST 700 Standard Rev 5 & ST 700 Basic Rev 4 Standard Display DP N/A Loop Output Flow ma (auto selected, nothing to configure in Offline) Disp. Scaling High Disp. Scaling Low Disp. Flow Units N/A N/A Pressure Pressure Units N/A AP or GP N/A Percent Output % (auto selected, nothing to configure in Offline) N/A Loop Output ma (auto selected, nothing to configure in Offline) N/A Page 12 FI LEAP Offline Configuration Tool User s Manual Revision 1

1.5 STT850/750/700 STT700 does not support mixed sensor types: TC/RTD, RTD/TC. Both the types should be the same such as TC/TC or RTD/RTD. 1.5.1 Latching Set Latching Alarm parameter value to: Disabled so that during the configuration download any device statuses are not Latched. 1.5.2 Parameters that can be downloaded from Offline template Please refer the table below while creating Offline templates in FDM. Parameters that are not listed here are not supported for download during Auto device commissioning and thus, user should not select these parameters in offline template. Offline parameters that are supported in STT850/750/700 STT850 STT750 STT700 Remarks tag tag tag long tag long tag long tag Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 date date date desc desc desc PV unit PV unit PV unit pv damp pv damp pv damp SV unit SV unit SV unit Loop current mode Loop current mode Loop current mode Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 Sensor scratch pad Sensor scratch pad Sensor scratch pad Message Message Message Tamper mode Tamper mode n/a Tamper Latency Tamper Latency n/a Max allowable attenpts Max allowable attenpts n/a NAMUR selection NAMUR selection NAMUR selection poll addres poll addres poll addres Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 Rotation Time Rotation Time n/a Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 13

Offline parameters that are supported in STT850/750/700 STT850 STT750 STT700 Screen Rotation Screen Rotation n/a Contrast level Contrast level n/a Sesnor1 Bias n/a Sesnor1 Bias PV Delta n/a Delta Delta Limit n/a Delta Limit Sensor scratch pad Sensor scratch pad n/a Damp bumpless transfer n/a Damp bumpless transfer Loop Control Mode n/a n/a Excess delta detect n/a Excess delta detect Hysterisys band n/a n/a Breakdetect Breakdetect Lead wire n/a n/a Match PV n/a n/a Compansation external n/a n/a Latching Alarm Latching Alarm n/a Alarm 1 type n/a n/a Alarm 2 Type n/a n/a Alalrm Latching Alalrm Latching n/a Allarm Blocking Allarm Blocking n/a Advance diag option n/a n/a Deviation limit n/a n/a n/a Sensor Bias n/a n/a n/a Sensor 2 Bais n/a n/a Sensor1 Install Date n/a n/a Sensor2 Install Date Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 Page 14 FI LEAP Offline Configuration Tool User s Manual Revision 1

1.6 SMV800 1.6.1 Parameters that can be downloaded from Offline template Please refer the table below while creating Offline templates in FDM. Parameters that are not listed here are not supported for download during Auto device commissioning and thus, user should not select these parameters in offline template. Parameters Remarks Tag Descriptor Long Tag Do not select this parameter in FDM Offline template. Refer section 1.2.2 Table 1 Message Final asmbly num Namur Selection PV is SV is TV is QV is SP Unit DP Unit If you select this parameter, select SV is, TV is, QV is parameters also in the offline Template. If you select this parameter, select PV is, TV is, QV is parameters also in the offline Template. If you select this parameter, select PV is, SV is, and QV is also in the offline template If you select this parameter, select PV is, SV is, and TV is also in the offline template If you select this parameter, select all 4 device variable parameters: PV is, SV is, TV is, and QV is. If you select this parameter, select all 4 device variable parameters: PV is, SV is, TV is, and QV is. DP LRV DP URV SP LRV SP URV Flow Damp Sensor Scratch Pad Break Detect Latching Alarm Set this Disabled or unselect it in offline Template so that any intermediate statuses during bulk download are not Latched Lower Calib Point Upper Calib Point Sensor Bias Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 15

KUser/Flow Coeff/Fc Pipe Diamter_D Bore Dia_d/APT Prob Width d Pipe Dia Mes Temp Bore Dia Mes Temp Atmospheric Pressure Max Flow Rate Max Diff Pressure Flow Cutoff Low Flow Cutoff High Tamper Mode Tamper Latency Max Allowable Attempts Language Rotation Time Screen Rotation Contrast Level Page 16 FI LEAP Offline Configuration Tool User s Manual Revision 1

2 Security 2.1 How to report a security vulnerability For the purpose of submission, a security vulnerability is defined as a software defect or weakness that can be exploited to reduce the operational or security capabilities of the software or device. Honeywell investigates all reports of security vulnerabilities affecting Honeywell products and services. To report potential security vulnerability against any Honeywell product, please follow the instructions at: https://honeywell.com/pages/vulnerabilityreporting.aspx Submit the requested information to Honeywell using one of the following methods: Send an email to security@honeywell.com. or Contact your local Honeywell Process Solutions Customer Contact Centre (CCC) or Honeywell Technical Assistance Centre (TAC) listed in the Support and Contact information section of this document. Revision 1 FI LEAP Offline Configuration Tool User s Manual Page 17

Sales and Service For application assistance, current specifications, pricing, or name of the nearest Authorized Distributor, contact one of the offices below. ASIA PACIFIC Honeywell Process Solutions, Phone: + 800 12026455 or +44 (0) 1202645583 (TAC) hfs-tacsupport@honeywell.com Australia Honeywell Limited Phone: +(61) 7-3846 1255 FAX: +(61) 7-3840 6481 Toll Free 1300-36-39-36 Toll Free Fax: 1300-36-04-70 China PRC - Shanghai Honeywell China Inc. Phone: (86-21) 5257-4568 Fax: (86-21) 6237-2826 Singapore Honeywell Pte Ltd. Phone: +(65) 6580 3278 Fax: +(65) 6445-3033 South Korea Honeywell Korea Co Ltd Phone: +(822) 799 6114 Fax: +(822) 792 9015 EMEA Honeywell Process Solutions, Phone: + 800 12026455 or +44 (0) 1202645583 Email: (Sales) FP-Sales-Apps@Honeywell.com or (TAC) hfs-tac-support@honeywell.com AMERICAS Honeywell Process Solutions, Phone: (TAC) (800) 423-9883 or (215) 641-3610 (Sales) 1-800-343-0228 Email: (Sales) FP-Sales-Apps@Honeywell.com or (TAC) hfs-tac-support@honeywell.com Specifications are subject to change without notice. For more information To learn more about LEAP visit www.honeywellprocess.com Or contact your Honeywell Account Manager Process Solutions Honeywell 1250 W Sam Houston Pkwy S Houston, USA, TX 77042 Honeywell Control Systems Ltd Honeywell House, Skimped Hill Lane Bracknell, England, RG12 1EB Shanghai City Centre, 100 Jungi Road Shanghai, China 20061 www.honeywellprocess.com 34-CT-25-01 Rev.1 July 2017 2017 Honeywell International Inc.