Calibration of ARM Spectral Shortwave Radiometers

Similar documents
metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

Influence of the light spectral distribution used in the radiometers calibration

REPORT OF CALIBRATION of One Standard of Spectral Irradiance ( nm)

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

SPECTRAL IRRADIANCE DATA

The Standard for over 40 Years

LI-193 Spherical Quantum Sensor

Instrumental and Methodological Developments in UV Research

Report on BLP Spectroscopy Experiments Conducted on October 6, 2017: M. Nansteel

LI-192 Underwater Quantum Sensor

A Quantix monochrome camera with a Kodak KAF6303E CCD 2-D array was. characterized so that it could be used as a component of a multi-channel visible

CU-LASP Test Facilities! and Instrument Calibration Capabilities"

Comparison of Quantum Sensors with Different Spectral Sensitivities

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

Capabilities of NIST SIRCUS for Calibrations of SSI Vis-IR Instruments

RADIATION BUDGET INSTRUMENT (RBI): FINAL DESIGN AND INITIAL EDU TEST RESULTS

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

Pulsed Laser Power Measurement Systems

Measurements of Infrared Sources with the Missile Defense Transfer Radiometer

White Paper on SWIR Camera Test The New Swux Unit Austin Richards, FLIR Chris Durell, Joe Jablonski, Labsphere Martin Hübner, Hensoldt.

TracQ. Basic Data Acquisition and Spectroscopy Software

The TSIS Spectral Irradiance Monitor: Prism Optical Degradation Studies

Calibration of a High Dynamic Range, Low Light Level Visible Source

Development of 2 Total Spectral Radiant Flux Standards at NIST

Total solar irradiance measurements with PREMOS/PICARD

Spectroradiometer characterisations for traceable solar radiation measurements

New automated laser facility for detector calibrations

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST

Silicon Photodiodes - SXUV Series with Platinum Silicide Front Entrance Windows

Part 1: New spectral stuff going on at NIST. Part 2: TSI Traceability of TRF to NIST

The Issues of Measurement of Optical Hazard Using Photometers EMRP JRP ENG05 Metrology for Solid State Lighting

Developement of the method for realization of spectral irradiance scale featuring system of

StarBright XLT Optical Coatings

771 Series LASER SPECTRUM ANALYZER. The Power of Precision in Spectral Analysis. It's Our Business to be Exact! bristol-inst.com

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED.

The FTNIR Myths... Misinformation or Truth

\Ç à{x ÇtÅx Éy ALLAH à{x `xüv yâä

2. Instrumentation Design, Specifications, and Installation of the SUV-100

A Laser-Driven Light Source (LDLS) as a portable spectral irradiance calibration source in the UV range and other radiometric applications.

LI-1500 Light Sensor Logger

Color Measurement with the LSS-100P

Black Body Radiation. References: P.A. Tipler, Modern Physics, pp (Worth Publishers, Inc., NY, 1978).

Miniature Spectrometer Technical specifications

Spectral responsivity uncertainty of silicon photodiodes due to calibration spectral bandwidth

Improved Radiometry for LED Arrays

Contact person Date Reference Page Stefan Källberg MTt6F (4) Measurement Technology

UV PROCESS SUPPLY, INC. CON-TROL-CURE COMPACT RADIOMETER INSTRUCTION MANUAL PART # M

Irradiance Calibration Using a Cryogenic Radiometer and a Broadband Light Source

Accepted for publication in Journal of Geophysical Research. Copyright 1999 by the American Geophysical Union. Further electronic distribution is not

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

Fire testing: Calibration of smoke opacity measuring systems

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Application Note (A16)

INTERNATIONAL INTERCOMPARISON OF WAVELENGTH SCALE AND PHOTOMETRIC SCALE OF SPECTROPHOTOMETRY LABORATORIES CENAM - NRC - INMETRO - NIST.

Instruction Manual of Luminance and Spectral Radiance Calibrations

Measurement and Measurement Error of Light Used for Photosynthesis & Plant Growth Richard Garcia April 20, 2010

Introduction to the operating principles of the HyperFine spectrometer

Agilent 86030A 50 GHz Lightwave Component Analyzer Product Overview

apogeeinstruments.com Introducing the PV-100 PV Monitoring Package

Experimental Analysis of Luminescence in Printed Materials

The PMOD/WRC Precision Spectroradiometer PSR

NEAR EYE DISPLAY (NED) SPECTRORADIOMETER SYSTEMS DATASHEET

BLACKBODY RADIATION PHYSICS 359E

EDGES. Judd D. Bowman, Arizona State University Alan E. E. Rogers, Haystack Observatory

Solid State Luminance Standards

Coating Thickness Measurement System

S.M. Vaezi-Nejad, M. Cox, J. N. Copner

Voltage-dependent quantum efficiency measurements of amorphous silicon multijunction mini-modules

Pyranometers SP- 100 Series SP -200 Series

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum.

RAMSES. A modular multispectral radiometer for light measurements in the UV and VIS

The University of Toledo R. Ellingson and M. Heben

Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting

Program for UV Intercomparison 2014 in Davos:

Typical LED Characteristics

Instruction manual for Ocean Optics USB4000 and QE65 Pro spectroradiometers

Spectral Reflectance Sensor SRS-NDVI

Laser Beam Analysis Using Image Processing

LED Lighting Flux and Color Measurement System (LFC)

GSM OPTICAL MONITORING FOR HIGH PRECISION THIN FILM DEPOSITION

Performance Comparisons of InGaAs, extended InGaAs, and Short-wave HgCdTe Detectors between 1 µm and 2.5 µm

GEO-SolarSIM-D2 and SunTracker-2000/3000

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

FREQUENCY COMPARISON AT 633 NM WAVELENGTH: DETERMINATION OF DIAGONAL ELEMENTS OF MATRIX MEASUREMENTS BY USING A MASTER-SLAVE He-Ne LASER SYSTEM

TRACEABLE SOLAR RADIATION MEASUREMENTS

BTS256-EF. Product tags: VIS, Spectral Measurement, Waterproof, WiFi. Gigahertz-Optik GmbH 1/7

Stable OES System for Fault Detection and Process Monitoring

Lambda 1050 / 950 UV/Vis/NIR

BYK-mac i Multi-angle color, effect and fluorescence measurement. Gabriele Kigle-Böckler, BYK-Gardner GmbH, 2013

Disclosure to Promote the Right To Information

Railroad Valley Playa for use in vicarious calibration of large footprint sensors

A stray light corrected array spectroradiometer for complex high dynamic range measurements in the UV spectral range.

ROTATING SHADOWBAND SPECTRORADIOMETER MODEL RSS-1024/UVRSS-1024 BULLETIN RSS/UVRSS-1024

PRECISE MEASUREMENTS OF SOLAR BEAM IRRADIANCE THROUGH IMPROVED SENSOR CALIBRATION

YOUNGS MODULUS BY UNIFORM & NON UNIFORM BENDING OF A BEAM

Pre-Lab 10. Which plan or plans would work? Explain. Which plan is most efficient in regard to light power with the correct polarization? Explain.

Southern African Large Telescope. RSS CCD Geometry

Improving the Collection Efficiency of Raman Scattering

Ensure Optimal Instrument Performance with Genuine Agilent Long-life Deuterium Lamps

Transcription:

Calibration of ARM Spectral Shortwave Radiometers J. J. Michalsky, J. L. Berndt, P. W. Kiedron, and L. C. Harrison Atmospheric Sciences Research Center State University of New York at Albany Albany, New York Abstract Absolute calibration of spectral shortwave radiometers is typically performed using National Institute of Standards & Technology (NIST) or NIST-traceable spectral lamps. In this paper, we compare 18 spectral irradiance lamps from NIST and three commercial vendors all using the same spectrometer to assess their agreement with each other and the NIST standards. The NIST procedure is followed for the 1000W FEL lamps from NIST, Optronics, and EG&G. A modified calibration procedure developed by LI-COR is followed for their 200W tungsten-halogen lamps. Results are reproducible from one day to the next to about 0.2% using the same spectrometer. Measurements taken four months apart using two similar, but different spectrometers were reproducible to 0.5%. The comparisons suggest that even NIST standards may disagree with each other well beyond their stated accuracy. Some of the 1000W commercial lamps agreed with the NIST lamps to within their stated accuracy, but not all. Surprisingly, the lowest cost lamps from LI-COR agreed much better with the NIST lamps than their stated accuracy of 4%, typically within 2%. We conclude that we can transfer the calibration from a standard lamp to a spectral radiometer with about 2% added uncertainty. Introduction The broadband measurements in the Atmospheric Radiation Measurement (ARM) Program and elsewhere have painted a fairly consistent picture of an overestimation of shortwave radiation by clear-sky models relative to the best shortwave measurements available (e.g., Kato et al. 1997). Key to resolving this discrepancy is to find where in the shortwave spectrum the model overestimates occur. To this end, we require very accurate spectral irradiance measurements throughout the shortwave. Central to good spectral measurements is careful calibration of the instruments and then proper operation in the field. A first attempt to compare ten calibration lamps of two different types from three manufacturers yielded results that were typically far apart in some parts of the spectral range between 400 nm and 1050 nm, even though the lamps were stated to be accurate to 4% or better in this spectral range. Figure 1 is a comparison of these ten lamps by ratioing the stated spectral irradiance of the one NIST lamp that we had in August 1997 to the manufacturer s stated spectral irradiances of their lamps. One EG&E lamp (GS0939) is within 1% of the NIST standard, but only for the 400 to 700 nm range. LI-COR 200W lamps operated in the ASRC LI-COR calibrator show a nearly constant offset with wavelength, but are close to their stated 4% accuracy. We report here on a second, more careful and extensive study to resolve the issues regarding absolute spectral calibration in December 1997. We compared thirteen 1000W FEL lamps, including four NIST standards, four from EG&G, a manufacturer of secondary standards, and five secondary standards from Optronics. We also compared to the NIST group a set of five LI-COR 200W quartz-halogen lamps operating at conditions different from the NIST recommended configuration, but presumably equivalent to it. Comparison of 1000W FEL Lamps At the beginning and end (and sometimes in the middle) of each day of the December comparison, we measured the sensitivity of our 512-channel diode array spectrometer to the spectral irradiance of one or two lamps. Figure 2 contains plots of these ratios for the 4 days of the comparison. In general, the ratios are within 0.1 to 0.2%. This demonstrates the stability of the power supply and the output of the lamp, the stability of the spectrometer, and our ability to position the lamps at the same distance with the same orientation. Based on the relative agreement among three of the NIST lamps (403, 404, and 405), we took their average responsiveness in counts/wm -2 nm -1 to develop a working standard for these comparisons. Figure 3 shows the ratio of the working standard irradiance to each of the four NIST 471

Figure 1. Ratio of NIST standard lamp F340 irradiance to irradiance of nine secondary sources in August 1997 run. (For a color version of this figure see http://www.arm.gov/docs/documents/technical/ conf_9803/michalsky- 98.pdf). Figure 2. Ratio of spectral sensitivity in counts/wm -2 nm -1 at beginning and end of each day s run for one or two lamps. (For a color version of this figure see http://www.arm.gov/docs/documents/technical/conf_9803/ michalsky-98.pdf). 472

Figure 3. Ratio of NIST working standard irradiance to each NIST lamp irradiance. Two colors for two different days. (For a color version of this figure see http://www.arm.gov/docs/documents/technical/conf_9803/michalsky- 98.pdf). lamps. Lamps 403, 404, and 405 are almost within 1% of the working standard. This should not be much of a surprise because these three lamps serve as the basis for the working standard, but it does confirm that they are consistent. Since the NIST uncertainty at the two standard deviation levels is about 1.1% or smaller over these wavelengths, lamp 340 is clearly an outlier with variations with wavelength between 4% and 6%. This lamp was the standard adopted in Figure 1 of this paper, thus explaining many of the discrepancies in that figure. The red and blue lines represent the experiment on two different days, leading one to conclude that these results are repeatable to about 0.2%. Optronics and EG&G buy NIST standard lamps and use them to calibrate 1000W FEL lamps to sell as secondary standards to customers. These are calibrated following the NIST recommended procedures of positioning the lamps at 50 cm from the spectrometer used for calibration transfer. Conditions are the same other than the fact that the lamps from these two companies operate at a slightly lower constant current of 8.000 amps versus the NIST lamps that operate at 8.200 amps. The company states a transfer accuracy of about 1%, which yields an overall uncertainty of about 1.6% when added to the NIST uncertainty. Figures 3 and 4 are similar in that Optronics lamps are compared to the working standard based on the three NIST lamps. Three of the lamps are within 2% of the NIST and show a constant offset with wavelength. A fourth and a fifth lamp show a similar wavelength dependence, i.e., almost neutral, but with a 3.5% and 8% offset. One could conclude that three lamps are close enough although they fall outside the expected 1.6% uncertainty, but the other two are clearly outliers. It would seem that if one buys a single lamp from NIST or Optronics, one cannot conclude with much conviction the accuracy about measurements because there are outliers from both sources. Figure 5 is a plot similar to Figures 3 and 4, except that it covers the four FEL lamps that we tested from EG&G Gamma Scientific. GS937 falls within the expected uncertainty limits at all wavelengths and is nearly independent of wavelength. GS911 and GS938 are within the uncertainty limits at some wavelengths, but not others and their wavelength dependencies are similar, but offset from one another by about 4%. GS939 is clearly an outlier, reinforcing the previous conclusion that a single lamp does not give one confidence in the verity of that lamp. The blue and red lines represent the experiment repeated with a 4-month interval between runs with different spectrometers of the same design. The results are reproducible at the 0.5% level on average. 473

Figure 4. Ratio of NIST working standard to each Optronics 1000W FEL lamp tested. (For a color version of this figure see http://www.arm.gov/docs/ documents/technical/conf_9803/michalsky-98.pdf). Figure 5. Ratio of NIST working standard to each EG&G 1000W FEL lamp tested. Two colors are from tests 4 months apart using different, but similar spectrometers. (For a color version of this figure see http://www.arm. gov/docs/documents/technical/conf_9803/michalsky-98.pdf). 474

Figure 6. Ratio of NIST working standard to several LI-COR 200W tungsten-halogen lamps in two different LI-COR calibrators. (For a color version of this figure see http://www.arm.gov/docs/documents/technical/ conf_9803/michalsky-98.pdf). Comparison of 200W LI-COR Lamps LI-COR has developed a self-contained optical radiation calibrator. The lamp used is a 200W tungsten-halogen cycle lamp that operates at constant power. The operating distance from the filament of the lamp is 20 cm, which is about 11 mm beyond the front surface of the box allowing adaptor plates to be built that will hold the detector rigidly at the fixed 20-cm distance. Within the box are the power supplies, electronics, and baffles to reject extraneous light. As seen in Figure 1, the LI-COR calibrator labeled Atmospheric Sciences Research Center (ASRC) and the one labeled ARM gave very different results. It was found upon return to the factory that the shunt resistor used to measure current in the ARM unit was out of tolerance and was subsequently replaced. The LI-COR lamps operating in two different calibrator housings compared to the working standard is displayed in Figure 6. The mean bias error from the working standard is only about 1% with all values within 3% for this wavelength range. The stated uncertainty of these lamps is 4%. There is no outlier and the lamps seem to operate equivalently in either calibrator housing. Summary and Conclusions Adopting three reasonably consistent NIST lamps as a working standard, we have compared 18 irradiance standards from four manufacturers using a diode-array spectrometer. One NIST lamp was outside the uncertainty limits set by NIST using this working standard. All of the Optronics lamps were outside the uncertainty using this working standard, although three of the five were close and were almost independent of wavelength. Three of the four EG&G lamps were outside the uncertainty limits at some or all the wavelengths. Finally, the LI-COR lamps were all within their stated uncertainty using both calibrators. We also addressed a few other issues that contribute uncertainty at the 0.5% or smaller level. Out-of-band rejection was estimated using laser lines to measure the slit function. We estimate less than 0.2% error from this source. Linearity was not tested. We noticed a very slight instability in the wavelength registration of the spectrometer that is correctable to 0.5% or better. We looked at temperature effects on the spectrometer foreoptic, which appears to be around a 0.2% effect. Based on these effects and those discussed in the paper, we estimate our ability to transfer a calibration to a field instrument at the 2% level. The measurement uncertainty is larger since the uncertainty 475

of the lamp output must be added to this. Finally, we are investigating the accuracy of field calibration based on a robust estimate using a number of Langley calibrations that may rival or exceed lamp calibration accuracy. Reference Kato, S., T. P. Ackerman, E. E. Clothiaux, J. H. Mather, G. G. Mace, M. L. Wesely, F. Murcray, and J. Michalsky, 1997: Uncertainties in modeled and measured clear-sky surface shortwave irradiances. J. Geophys. Res., 102, 25,881-25,898. 476