A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING

Similar documents
Acoustic Emission Preamplifiers Specification

ACOUSTIC EMISSION MEASUREMENTS ON SHELL STRUCTURES WITH DIRECTLY ATTACHED PIEZO-CERAMIC

EWGAE 2010 Vienna, 8th to 10th September

ISO INTERNATIONAL STANDARD. Non-destructive testing Acoustic emission inspection Secondary calibration of acoustic emission sensors

COMPOSITES FROM PIEZOELECTRIC FIBERS AS SENSORS AND EMITTERS FOR ACOUSTIC APPLICATIONS*

Maximizing LPM Accuracy AN 25

3D Distortion Measurement (DIS)

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

LIMITATIONS IN MAKING AUDIO BANDWIDTH MEASUREMENTS IN THE PRESENCE OF SIGNIFICANT OUT-OF-BAND NOISE

Effect of coupling conditions on ultrasonic echo parameters

AN5E Application Note

Balanced Armature Check (BAC)

JOURNAL OF ACOUSTIC EMISSION

Transfer Function (TRF)

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

Recommendation of RILEM TC 212-ACD: acoustic emission and related NDE techniques for crack detection and damage evaluation in concrete*

Pre-Amplifier SPA Series

Location of Leaks in Liquid Filled Pipelines under Operation

MODEL 5002 PHASE VERIFICATION BRIDGE SET

1. Introduction. 2. Mobile Ultrasonic Inspection System MUSE

EFFECTS OF LATERAL PLATE DIMENSIONS ON ACOUSTIC EMISSION SIGNALS FROM DIPOLE SOURCES. M. A. HAMSTAD*, A. O'GALLAGHER and J. GARY

DETECTION AND SIZING OF SHORT FATIGUE CRACKS EMANATING FROM RIVET HOLES O. Kwon 1 and J.C. Kim 1 1 Inha University, Inchon, Korea

Spectral Distance Amplitude Control for Ultrasonic Inspection of Composite Components

Effectively Using the EM 6992 Near Field Probe Kit to Troubleshoot EMI Issues

Transducer product selector

Thermal Johnson Noise Generated by a Resistor

MEASUREMENT OF SURFACE DISPLACEMENT EXCITED BY EMAT TRANSDUCER

1818. Evaluation of arbitrary waveform acoustic signal generation techniques in dispersive waveguides

IEC Electrical fast transient / Burst immunity test

MEASUREMENT OF SURFACE ACOUSTIC WAVE USING AIR COUPLED TRANSDUCER AND LASER DOPPLER VIBROMETER

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic thickness measurement

EE 422G - Signals and Systems Laboratory

Time Reversal FEM Modelling in Thin Aluminium Plates for Defects Detection

THE DECI REPORT. H. L. Dunegan. August, 2000 AN ALTERNATIVE TO PENCIL LEAD BREAKS FOR SIMULATION OF ACOUSTIC EMISSION SIGNAL SOURCES.

Aero Support Ltd, 70 Weydon Hill Road, Farnham, Surrey, GU9 8NY, U.K.

Lab 4. Crystal Oscillator

Description of options, upgrades and accessories for the laser beam stabilization system Compact

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

ISO INTERNATIONAL STANDARD. Non-destructive testing Acoustic emission inspection Primary calibration of transducers

ON LAMB MODES AS A FUNCTION OF ACOUSTIC EMISSION SOURCE RISE TIME #

not overpower the audience just below and in front of the array.

MIL-STD-202G SHOCK (SPECIFIED PULSE)

ISO INTERNATIONAL STANDARD. Non-destructive testing Ultrasonic inspection Evaluating electronic characteristics of ultrasonic test instruments

Attenuation and velocity of ultrasound in solid state materials (transmission)

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

Frequency Dependence of Receiving Sensitivity of Ultrasonic Transducers and Acoustic Emission Sensors

A R T A - A P P L I C A T I O N N O T E

Texas Components - Data Sheet. The TX53G1 is an extremely rugged, low distortion, wide dynamic range sensor. suspending Fluid.

80 khz Cylindrical Ultrasound Transducer

Dynamic Generation of DC Displacement AN 13

Techniques to reduce electromagnetic noise produced by wired electronic devices

SIGNAL RECOVERY. Model 7265 DSP Lock-in Amplifier

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

SENSOR AND MEASUREMENT EXPERIMENTS

EXPERIMENTAL TRANSFER FUNCTIONS OF PRACTICAL ACOUSTIC EMISSION SENSORS

Keywords: Ultrasonic Testing (UT), Air-coupled, Contact-free, Bond, Weld, Composites

OPVibr Ultrasonic vibration measurement system Ultrasonic vibrometer INSTRUCTION MANUAL

Reduction of Dispersive Wave Modes in Guided Wave Testing using Split-Spectrum Processing

ELEC 391 Electrical Engineering Design Studio II (Summer 2018) THE UNIVERSITY OF BRITISH COLUMBIA Department of Electrical and Computer Engineering

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler

Agilent N9320B RF Spectrum Analyzer

R3477. Ideal for mobile communication applications including base stations and handsets, from the development stage to production and installation

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

MGA Magnetic field system. DC khz

Improving Amplitude Accuracy with Next-Generation Signal Generators

55:041 Electronic Circuits The University of Iowa Fall Exam 3. Question 1 Unless stated otherwise, each question below is 1 point.

PDu150CL Ultra low Noise 150V Piezo Driver with Strain Gauge Feedback

ENGINEERING STAFF REPORT. The JBL Model L40 Loudspeaker System. Mark R. Gander, Design Engineer

Transducer for Measurement Bridges SCM90, SIGS15

EE12: Laboratory Project (Part-2) AM Transmitter

A Study on Correlation of AE Signals from Different AE Sensors in Valve Leakage Rate Detection

Where DSP meets Measurement Science: A Sound Example. By Andrew Hurrell PhD

SYSTEM ONE * DSP SYSTEM ONE DUAL DOMAIN (preliminary)

Magnetic-Field Test System / Low-Frequency Test System for Emission and Immunity Tests / MTS-800

STEREO IMPACT Solar Energetic Particles Package (SEP) Dynamic Test Plan

UNIT-3. Electronic Measurements & Instrumentation

PDu150CL Ultra-low Noise 150V Piezo Driver with Strain Gauge Feedback

A minimum hydrophone bandwidth for undistorted cavitation noise measurement

Electronic Noise Effects on Fundamental Lamb-Mode Acoustic Emission Signal Arrival Times Determined Using Wavelet Transform Results

Vibration Transducer Calibration System

Sampling and Reconstruction

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

Measurement at defined terminal voltage AN 41

Quadra 10 Available in Black and White

The Impact of Very High Frequency Surface Reverberation on Coherent Acoustic Propagation and Modeling

200 MHz Photoreceiver with Si PIN Photodiode

8 Hints for Better Spectrum Analysis. Application Note

Measuring Batteries using the Right Setup: Dual-cell CR2032 and Battery Holder

Causes for Amplitude Compression AN 12

OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING

8 Hints for Better Spectrum Analysis. Application Note

Data and Computer Communications Chapter 3 Data Transmission

Standard Guide for Evaluating Characteristics of Ultrasonic Search Units 1

EFFECT OF INTEGRATION ERROR ON PARTIAL DISCHARGE MEASUREMENTS ON CAST RESIN TRANSFORMERS. C. Ceretta, R. Gobbo, G. Pesavento

Technical Documentation

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

High-temperature Ultrasonic Thickness Gauges for On-line Monitoring of Pipe Thinning for FAC Proof Test Facility

Kent Bertilsson Muhammad Amir Yousaf

Transcription:

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING HARTMUT VALLEN, JOCHEN VALLEN and JENS FORKER Vallen-Systeme GmbH, 82057 Icking, Germany Abstract AE testing of atmospheric storage tanks filled with liquid becomes more and more accepted. Sensors are attached to the tank wall and shall differentiate smallest amplitudes in the 25-45 khz frequency range from the background noise. Different sensor models are offered for this application. Are these comparable regarding sensitivity and signal-to-noise ratio? This paper describes a reproducible method to compare such sensors. For continuous as well as for pulse excitation, a suitable setup and obtained results are described. Measuring the noise spectra helps to explain the influence of the frequency filtering on the signal-to-noise ratio. Introduction This paper addresses the needs of AE testers of atmospheric storage tanks of various sizes, who want to compare the sensitivity of different sensor models or sensors of same model using a simple method. The most important standards for AE sensor calibration are ISO 12713 (primary calibration) [1] and ISO 12714 (secondary calibration) [2]. Both standards are designed for calibration laboratories. They require a large and heavy calibration block. AE testers usually do not have access to such a calibration block. Additionally, ISO 12714 is recommended for the frequency range 100 khz 1 MHz, whereas tank testing is usually done in the frequency range 25-45 khz. Hence, these standards are of no help for the AE tester. Other standards like EN 13477 [3] and ASTM E976 [4] are general and do not consider the fact that plate waves propagate in different modes. To detect smallest AE signals, the sensor should have the largest possible signal-to-noise ratio (SNR). This means that for a certain excitation the sensor shall provide the largest possible signal voltage superposed with the smallest possible inherent noise. For a sensitivity comparison one has to proceed as follows: 1. Excite both sensors in identical manner (frequency sweep) and strongly enough so that the noise can be neglected against the signal from the excitation. The output signal shall be U A (f). 2. Measure the inherent noise of the sensors (no excitation!). This is the output voltage U R. 3. The SNR at given excitation is then: SNR(f) = U A (f)/u R. It might be of interest to calculate a SNR for peak values and one for RMS values. 4. The frequency response as well as the noise measurement shall be made with the same frequency filters as used for the tank test. 5. The sensor with the higher SNR can distinguish smaller excitations from the background noise. Hence, this is the more sensitive sensor. Setup For the comparison the following setup was used: J. Acoustic Emission, 25 (2007) 132 2007 Acoustic Emission Group

Function Generator (FG) The function generator creates a sine wave voltage with adjustable frequency and amplitude or a pulse with selectable duration and amplitude, respectively. In our test we used model 33220A (Agilent). Sensor Excitation For a comparison, the sensors under test (SUT) need to be excited acoustically in exactly the same way. As emitter an ultrasonic transducer model V101 (Panametrics) was coupled face to face to one SUT using light machine oil as coupling agent. The emitter was driven by a frequency-swept sine wave and the AE signal amplitude from SUT was measured. This method provides well reproducible results and is well suited for routine sensor verification. But the following objections could be raised: a) AE tank-floor testing analyzes burst AE and not continuous (sine wave) AE. A comparison should also consider burst excitation. b) Different SUT models could have different feedback on the V101 emitter and thereby tamper the comparison result. Considering these objections, a second comparison was made using a pulse excitation via an aluminum rod of 610-mm length and 19-mm diameter with polished ends. Both excitation methods led to almost the same results for the frequency range 25-45 khz. Compared Sensor Models We compared a Vallen VS30-SIC-46dB sensor (S/N 120) with another sensor, hereafter called XXX. Both sensors have an integral preamplifier requiring 28V DC supply voltage on the signal wire and 20.6 mm diameter. The following lists the differences. Model Length Face Connector VS30-SIC-46dB 52.8mm isolated ceramic plate BNC at case XXX 38.8mm non-isolated metal plate BNC with 1 m cable Measurement Chain with Various Filters For measurements we used Vallen AMSY-5 AE system with dual-channel AE processor ASIP-2, a 25-45 khz band-pass for the first test, and a 25-100 khz band-pass for a second test. These band-pass filters consist of digital high- and low-pass filters each with 48 db/octave steepness. Figure 1 shows the response curves of the band-pass at 1 V PP continuous sine wave at ASIP-2 input. Frequency sweep and RMS measurement were controlled by Vallen Sensor Tester software. Sensor Frequency Response To obtain Figs. 2 and 3, the FG output (50 mv PP, terminated externally with 50 ) was connected to the V101, face-to-face with SUT. The red curves were taken with 25-45 khz filter and the blue curves with 25-100 khz filter in ASIP-2. For the determination of the inherent noise of the SUT, we removed the acoustic excitation by disconnecting the V101 from the FG. We amplified the sensor output with an auxiliary amplifier by 40 db, which allowed one to ignore the noise added by subsequent measurement stages. In this way the horizontal lines in orange (25-45 khz) and green (25-100 khz) were recorded. 133

Fig. 1 Frequency response of the used filters. During this noise measurement, we ensured that no acoustic noise sources like fans, human voices, or others could cause a distortion within the frequency range under evaluation. Both SUT were treated in exactly the same way. Fig. 2 Frequency response and noise of VS30-SIC-46 db. 25-45 khz: red and orange lines; 25-100 khz: blue and green lines. Fig. 3 Frequency response and noise of XXX. Line colors same as in Fig. 2. The results are summarized in Table 1. They were deduced from the 4 lines of Figs. 2 and 3, whereby the noise has been corrected by the 40 db post-amplification. As can be seen from noise and maximum amplitude, the gain of the integral preamplifier of XXX is lower than that of the VS30-SIC-46dB, but the deciding factor is the SNR as this is independent of the gain. Results for continuous excitation and 25-45 khz filter: VS30-SIC-46dB provides 17 db more signal amplitude and 9.5 db better SNR. 134

Table 1 Results for continuous sensor excitation. Sensor: VS30-SIC-46dB XXX Filter [khz]: 25-45 25-100 25-45 25-100 Maximum amplitude (RMS): line color: 87 db red 90.5 db blue 70 db red 72 db blue Frequency at max. ampl.: 35 khz 53 khz 35 khz 78 khz RMS noise: line color: 10 db orange 20 db green 2.5 db orange 5 db green Signal-to-noise ratio SNR: 77 db 70.5 db 67.5 db 67 db Difference wrt. XXX at 25-45 khz: 9.5 db 3 db 0 db -0.5 db Using Burst Excitation For XXX the exciting pulse amplitude was 5 V P. For VS30-SIC-46dB the amplitude was reduced to 1 V P because 5 V P led to saturation due to the larger gain. Figure 4 shows the response of the VS30-SIC-46dB with 1-V P excitation amplitude at V101, and Fig. 5 shows the response of XXX with 5 V P at V101, both with 25-45 khz filter. For the FFT, a 190- s long Hamming window was used. Table 2 lists the maximum amplitude in the time domain (line 2), converted to db (line 3), the maximum amplitude in the frequency domain (line 4). To compensate for the 5-V P excitation of XXX, its line-2 value is divided by 5 (600/5 = 120 mv) and line 4 is reduced by 14 db (97 14 = 83 db). Table 2 also lists the noise in mv P (line 5) and converted to db (line 6). The noise values were recorded separately, in reference to the SUT output (before 40-dB amplification) and are maximum values (peaks), which occurred in a frequency of 1/s or less. The resulting signal-noise-ratios are listed in line 7 (time domain) and line 8 (frequency domain). Scaling in Figs. 4-7 refers to the input voltage at the AE signal processor (ASIP-2). Fig. 4 Pulse response VS30-SIC-46dB, filter: 25-45 khz, excitation 1 V P x 1 s. max. 940 mv in time domain or 100 db in FFT. Noise: 0.58 mv P. Figures 6 and 7 were taken with band-pass of 25-300 khz. Figure 6 shows at approximately 30 s the arrival of the s 1 mode, which travels with ~4000 m/s at 250 khz according to Fig. 8. Considerable differences between the two sensor models are seen: VS30-SIC-46dB exhibits a 135

resonance at ~60 khz, where XXX exhibits resonances at 40 and 80 khz and an anti-resonance at 60 khz. Due to the obscure influence of the s 1 mode, a direct comparison of Figs. 6 and 7 is not recommended. Fig. 5 Pulse response XXX, filter: 25-45 khz, excitation 5 V P x 1 s. max. 600 mv in time domain, 97 db in FFT. Noise: 0.20 mv P. Fig. 6 Pulse response VS30-SIC-46dB, filter: 25-300 khz, excitation 1 V P x 1 s. max. 1900 mv/103 db at 60 khz. Noise: 2.18 mv P. Arrival of s 1 mode at t = 30 s. Result for burst excitation with 25-45 khz filter: Table 2, line 3 indicates that VS30 delivers 17.8 db more amplitude and line 8 shows 8.6 db more SNR in time domain and 7.8 db more in frequency domain. This result is very similar to continuous excitation. 136

Fig. 7 pulse response XXX, filter : 25-300 khz, excitation 5 V P x 1 s. Max. 1750 mv/98 db at 80 khz. Noise: 0.33 mv P. Fig. 8 Dispersion curves for 19 mm aluminum rod, according to [5]. Table 2 Burst excitation results with 25-45 khz filter. 137

Noise Spectra, Impedance and Natural Frequency The frequency, at which a sensor shows a sudden jump in its impedance, is called natural frequency [6]. For obtaining an impedance curve (Figs. 9 and 10), a passive sensor must be used. A sine wave of 100 mv PP from a function generator in series with 10 pf was fed in parallel to a sensor VS30-V (same piezo-element as VS30-SIC-46dB) connected to a preamplifier AEP4 (40 db). Figure 9 shows the lowest impedance at 51 khz, and the highest at 58 khz. Peculiar with this frequency is, that the amplitude measured with sensor connected (Fig. 9: 94 db) is higher than measured with the sensor disconnected (89 db)! The impedance combination of both, sensor and preamplifier, generate a sharp resonance peak. Fig. 9 Impedance jumps of VS30-V (no integrated preamplifier). Fig. 10 Zoom of Fig. 9 around the natural frequency (58 khz). This peak can also be seen in the noise spectrum (Fig. 11) of a sensor with integrated preamplifier. This dominating peak in the noise spectrum is the reason for the increase of inherent noise when using the 25-100 khz band-pass filter instead of 25-45 khz. In both Figs. 9 and 11, further peaks at 112 khz, 175 khz and 270 khz can be identified. XXX has its dominating peak in the noise spectrum below 25 khz (Fig. 12). This explains why the noise of the XXX does not substantially increase with a 25-100 khz band-pass filter. The determination of a reproducible noise spectrum requires averaging the FFT over many measurements as the individual spectra of noise records scatter considerably. For Figs. 11 and 12, we averaged 1000 noise records using the Vallen FFT-Averager. The absolute scaling of Figs. 11 and 12 must not be compared. These figures shall just illustrate the different natural frequencies and the effect of filter bandwidth on the noise of the filtered signal. 138

Fig. 11: Noise spectrum VS30-SIC-46dB (average of 1000 FFTs). Fig. 12: Noise spectrum XXX (average of 1000 FFTs). Conclusion This report describes two setups to compare the sensitivity of different sensor models. One is with face-to-face coupling and continuous excitation, and the other with burst excitation via an aluminum rod. In both cases, the excitation is perpendicular to the sensitive area. When looking at the SNR (signal-to-noise ratio), the inherent noise and the noise spectra have to be considered. Two sensor models have been compared. The result depends strongly on the used frequency range. For 25-45 khz and identical excitation, one sensor model provides 17 db more signal and about 9 db better SNR than the other. For tank floor testing, the frequency range of 25-45 khz is suited best. The more sensitive sensor model shows a natural frequency of 60 khz. This causes a peak in the noise spectrum, which is excluded effectively by the 25-45 khz band-pass filter. References [1] Non-destructive testing Acoustic emission inspection Primary calibration of transducers, ISO 12713, ISO/TC135, 1998. [2] Non-destructive testing Acoustic emission inspection Secondary calibration of acoustic emission sensors, ISO 12714, ISO/TC135, 1999. [3] Non-destructive testing Acoustic emission Equipment characterisation- Part 2: Verification of operating characteristics, EN13477-2, CEN/TC138, 2001. [4] Standard Guide for Determining the Reproducibility of Acoustic Emission Sensor Response, ASTM E976 ASTM/E07.04, 2001. [5] J.L. Rose, Ultrasonic Waves in Solid Media, Cambridge University Press, ISBN 0 521 54889 6, 2004, p. 149+. [6] G. Gautschi, Piezoelectric Sensorics, Springer Verlag, ISBN 3-540-42259-5, 2002, p. 195+. 139