Introduction of a Single Chip TLD System for Patient Dosimetry

Similar documents
Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

2012 :15th SESSION of ESMP

Performance of Thermoluminescent Dosimeters Under As-Deployed Conditions by John A. Harvey

Nuclear Associates , , CT Head and Body Dose Phantom

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

An Activity in Computed Tomography

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

First EURADOS intercomparison exercise of eye lens dosemeters for medical applications

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

Determination of the tube voltage from clinic mammographic system using two types of detectors

Y11-DR Digital Radiography (DR) Image Quality

Preliminary studies of a new monitor ionization chamber

Summer Student project report

An Activity in Computed Tomography

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

UT-ONE Accuracy with External Standards

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Half value layer and AEC receptor dose compliance survey in Estonia

Wide beam CT dosimetry. Elly Castellano

Calibration of KAP meters

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Dosimetry for spectral molecular imaging of small animals with MARS-CT

TECHNICAL DATA. benefits

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Cylindrical Ion Chambers

Test Equipment for Radiology and CT Quality Control Contents

K-edge subtraction X-ray imaging with a pixellated spectroscopic detector

Overview of Safety Code 35

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Cylindrical Ion Chambers Victoreen Model 550 Series

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

DETECTORS. Including Codes of Practice IONIZING RADIATION

Electronic Brachytherapy Sources. Thomas W. Rusch

I. Introduction.

- Water resistant. - Large size.

New Detectors for X-Ray Metal Thickness Measuring

When small things matter. Small Field Dosimetry Application Guide

Investigation of the film FUJIFILM IX 80 in TETENAL machine processing

TOPICS: CT Protocol Optimization over the Range of Patient Age & Size and for Different CT Scanner Types: Recommendations & Misconceptions

QC Testing for Computed Tomography (CT) Scanner

NOT FOR DISTRIBUTION JINST_083P_0914 v1

Unfors EDD-30 Radiation Protection in Fluoroscopy

Post-irradiation colouration of Gafchromic EBT radiochromic film

ELECTRONIC CONTROL CONCEPTS 160 Partition Street Saugerties, NY or local phone

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

High collection efficiency MCPs for photon counting detectors

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Veraviewepocs 2D High Speed Panoramic X-Ray Crystal Clear Images with Reduced Radiation

Development of new dosimeter for measuring dose distribution in CT

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

SCINTILLATING FIBER DOSIMETER ARRAY

Use of Back Scattered Ionizing Radiation for Measurement of Thickness of the Catalytic Agent Active Material

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE

PD233: Design of Biomedical Devices and Systems

GAFCHROMIC HD-810 Radiochromic Dosimetry Film Configuration, Specifications and Performance Data

Features and Weaknesses of Phantoms for CR/DR System Testing

RADIATION SAFETY DIRECTORATE

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

NM Module Section 2 6 th Edition Christian, Ch. 3

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

PEDIATRIC AND STAFF DOSE EVALUATION IN FLUOROSCOPY UPPER GASTROINTESTINAL SERIES

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

Computerized transverse axial scanning (tomography): Part I. Description of system

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

diagnostic examination

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Photomultiplier Tube

APPENDIX K UNF RADIATION GENERATING DEVICE SAFETY PROCEDURES

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

Investigation of the film Kodak MX 125 in Agfa NDT S eco processing

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose

Nuclear Associates

ISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Results of FE65-P2 Pixel Readout Test Chip for High Luminosity LHC Upgrades

COMPUTED TOMOGRAPHY 1

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Nuclear Associates

Fully Automatic X-ray meter kvp, dose, rate, HVL and time Extremely easy to use. Unfors ThinX RAD for Radiographic Quality Control

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Transcription:

Introduction of a Single Chip TLD System for Patient Dosimetry C. Hranitzky a, M. Halda a, G. Müller a, B. Obryk b, H. Stadtmann a* a Austrian Research Centers GmbH ARC, 2444 Seibersdorf, Austria. b Institute of Nuclear Physics (IFJ), ul. Radzikowskiego 152, 31-342 Kraków, Poland. Abstract. A thermoluminescence dosimetry system with single detector chips was developed for patient dosimetry applications. LiF:Mg,Cu,P detector chips, dosimetry protocol, calibration, and dose calculation were prepared for measurements inside phantoms for determining organ and effective doses in medical diagnostic examinations. The first step was optimizing the readout time-temperature-profile for reaching a well resolved dosimetric peak and stability of the glow curves. A number of parameters was varied for the optimization process, e.g. preheating and heating rate. Individual chip sensitivities, residual dose and dose linearity were studied for establishing a reliable and accurate TL dosimetry system. KEYWORDS: patient dosimetry, thermoluminescence, TLD, single chips, glow curve, calibration. 1. Introduction Detailed knowledge about patient dose accumulation during medical exposure is of increasing importance especially since the introduction of the European Council Directive 97/43/EURATOM [1]. A TL (thermoluminescence) dosimetry system with single detector chips is being developed mainly for new or special patient dosimetry applications. A number of TL chips is positioned inside anthropomorphic phantoms for determining organ and resulting effective dose values. First applications are planned for diagnostic radiology examinations with new 3D dental imaging scanners. LiF:Mg,Cu,P detector material has been selected for medical dosimetry applications due to a number of reasons, especially due to its dose detection accuracy and dose linearity in the typical diagnostic dose range up to several tens of mgy. Higher accuracy may be achieved by special treatments or higher dose levels as in radiotherapy TL dosimetry [2]. The main work of introducing a new single chip TL system was optimizing the readout parameters for reaching a number of goals such as stability of the readout glow curves. 2. Materials and Methods 2.1 Detector Chips and TL Reader TLD-7H detector chips (from Thermo Fisher Scientific) of 3.2 x 3.2 x.89 mm size are read out in a Harshaw 35 manual TL reader with planchet heating system and WinREMS readout software. TLD-7H detector material ( 7 LiF:Mg,Cu,P) has properties well suited for a number of medical diagnostic dosimetry applications, e.g. small sizes for measurements inside phantoms tissue equivalence for detecting tissue and organ dose improved sensitivity compared to standard materials such as TLD-1 independence of dose rate for pulsed radiation fields * Presenting author, E-mail: hannes.stadtmann@arcs.ac.at 1

Two groups with 2 and 14 chips were used for testing purposes and for measurement purposes, respectively. TL light is detected by a photomultiplier tube positioned close to the heated detector chip. Glow curves represent detected light signals as a function of readout time. Two important requirements for optimizing the dosimetry protocol were: long term stability in glow curve shapes total readout time less than 3 hours The main glow curve peak, referred to as peak 4 for this TL material, has to be fully visible and well resolved in the glow curve. The requirement of total readout times less than three hours allow fast measurements within half a day following the protocol steps: oven annealing irradiation (except background chips) simultaneous calibration irradiation (only calibration chips) oven pre-readout tempering readout (including empty readouts, test chip readouts, QC noise and reference light readouts) 2.2 Readout Time-Temperature-Profile The TTP (Time-Temperature-Profile) defines the readout temperature as a function of readout time. The TTP consists of a preheating phase, a linear heating phase up to the maximum readout temperature and a final constant readout period at the maximum temperature. The following five parameters are describing the TTP: preheating temperature ( C) preheating time (s) at the preheating temperature heating rate ( C/s) maximum readout temperature ( C) total readout time (s) The recording of the glow curve starts directly after the preheating phase at the preheat temperature. The actual heating time is defined by the ratio of temperature increase up to the maximum readout temperature and the heating rate. The remaining readout time at the maximum temperature is determined by the difference of the total readout time and the heating time. This final phase is important for the chips to reach the adjusted maximum temperature and to achieve a complete readout the TL signal of the main dosimetric peak 4. 2.3 Glow Curve Optimization The main goals of the optimization process were achieving long-term stability in glow curve shape and reproducible TL signals. The following parameters were investigated with varying parameter values close to manufacturer or literature recommendations: annealing process, recommended 24 C for 1 min pre-readout tempering, recommended approx. 1 C for 1 or 2 min preheating temperature, varied between 135 C and 165 C preheating time, varied between 5 s and 2 s heating rate, varied 5 C/s to 3 C/s The maximum temperature was set close to the recommended 24 C limit for this TL detector material avoiding sensitivity loss. The total readout time depends mainly on the heating rate and was set accordingly to get a full readout of main peak 4. Time intervals between annealing, i.e. zeroing of the TL chips, and readout are typically between half a day and one day. Fading effects are therefore expected to be negligible. 2

2.4 Individual Chip Sensitivity Differences in individual chip sensitivities of a production batch are mainly caused by differences in chip masses. k i (individual chip sensitivity correction factor) measurements are carried out regularly for determining long term changes in chip sensitivities. All detector chips are irradiated under reference conditions (1 mgy 137 Cs free-air calibration) and k i values are calculated as ratios of the average TL signal to the individual TL signal. 2.5 Dose Determination Dose calculation is based on summing up the whole glow curve resulting in a single TL signal value. This simple analysis is possible due to a number of reasons, especially due to the stability of main peak 4, due to the low signal tails at the beginning and end of the glow curves and due to the stabilization of peak 3 which showed the biggest variations during the optimization process. Extracting only peak 4 signal values by simple glow curve analyses showed no significant improvements in calculated dose values. The dosimetry model for determining dose D i of chip number i is based on a simple linear equation: D i = N k k TL TL ) (1) Q ( i i bg Individual chip signals TL i are k i sensitivity corrected and have an average background signal value TL bg of a group of unirradiated background chips subtracted from them. Typically a number of measurement chips are positioned together allowing the determination of a mean dose value D and a related standard deviation of the mean. At least ten percent of the chips are used as calibration chips for determining the current readout calibration factor N. The group of calibration chips are irradiated at reference conditions in a 137 Cs free-air calibration facility at 1 mgy air kerma. Finally the appropriate k Q quality correction factor has to be applied. A set of k Q values will be determined at the secondary standards dosimetry laboratory Seibersdorf in diagnostic reference X-ray radiation fields free-in-air and in-phantom. 3. Results 3.1 Final TTP The final Time-Temperature-Profile can be described by the parameter set 135/1/5/245/26.7, i.e. a preheating phase at 135 C for 1 s, a linear heating phase with 5 C/s heating rate up to the maximum readout temperature of 245 C with a total readout time for the glow curve of 26.7 s. Examples of glow curves can be seen e.g. in figure 5 for a single chip calibrated with a constant dose of 1 mgy. Temperature shifts in the glow curves may depend on the thermal contact and heat transfer between planchet and chip. It has to be noted that the displayed temperature is not equal to the real chip temperature but close to the planchet temperature. 3.2 Variation of Parameters One goal of the optimization of glow curve structure was to reduce peak 3 compared to the main peak (peak 4). Additionally peak 3 should be as stable as possible, i.e. the ratio of peak 3 to peak 4 maximum should not vary due to small variations in measurement parameters. At the beginning of the optimization process peak 3 magnitude was up to one half of main peak magnitude for TTPs similar to the final one. 3.2.1 Annealing Annealing at 24 C for 1 minutes is recommended in literature. The manufacturer recommends oven annealing only after dose applications above certain limits. Zeroing may be done by individual readout 3

cycles but result in longer preparation times. To study the influence of the annealing process on the reproducibility of glow curves, chips with and without annealing have to be compared over longer time periods. As in routine radiation protection dosimetry, TL detectors are stabilized before readout by a prereadout oven tempering process at about 1 C for 1 or 2 minutes. The results of this tempering process were investigated. As expected, the influences were quite small for glow curves with preheating temperatures above 135 C. In figure 1, glow curves without any preheating are compared. Figure 1: Comparison of glow curves with and without pre-readout tempering (tempering was at 1 C for 1 minutes). Both glow curves are without a preheating phase i.e. starting from room temperature. The glow curve no pre-readout tempering clearly shows peaks 2, 3, and 4 from left to right i.e. from lower to higher readout temperatures. 1.2 1 normalized TL glow curve.8.6.4.2 no pre-readout tempering pre-readout tempering 1 C / 1 min 3.2.2 Preheating Glow curves without preheat may show pronounced low temperature peaks 2 and 3 depending on the time interval between irradiation and readout due to different fading characteristics of the peaks. A drawback of preheated glow curves is that signal information at low temperatures is not visible e.g. irregularities due to dirt effects. Preheat temperatures were varied between 135 and 165 C. The applied preheat temperature should not reduce the main peak but only reduce the lower temperature peaks. Preheat time has to be long enough not to produce any changes at the beginning of the glow curve. Further increase of the preheat time should not change the glow curve. Preheat times were varied between 5 and 2 s. Preheating temperatures below 135 C resulted in no differences in the glow curves but in longer readout times per chip, e.g. 5 s instead of 26.7 s at 5 C/s heating rate without preheating. Preheating shows a better resolved, i.e. broader peak 4. Finally, a preheating phase at 135 C preheat temperature and 1 s preheat time was chosen, similar to the manufacturer s recommendations. 3.2.3 Heating Rate Increasing the heating rate allows faster readout times but results in increasing variations of temperature shifts of the glow curves. Heating rates were varied between 5 and 25 C/s. Lower rates were not considered because of too long total readout times. Figure 5 shows temperature shifts for the final heating rate of 5 C/s. Higher heating rates produce glow curves with narrower peak shapes resulting in unwanted overlaps of peak 3 and peak 4, as can be seen in figure 2. 4

Figure 2: Comparison of glow curves with heating rates of 5, 15, and 25 C/s. Curves are peak maximum normalized. Original curves have similar areas because of similar applied dose values. 1.2 1 normalized TL glow curve.8.6.4.2 linear heating rate ( C/s) 5 15 25 3.3 Second Readout Residual The residual signal can be determined as the ratio of the second readout TL signal to first readout TL signal. Irradiations at different dose levels showed residuals in the range of 1 to 2%. Increasing residuals, up to 8%, were measured below 1 µgy down to the lower detection limit of about 1 µgy. Individual second readout signals were not taken into account for dose determination because dose linearity calculations showed no improvement compared to subtracting an average background signal. Therefore, for each measurement a group of background chips (typically 2% of all chips) is used. Additionally the performance of the annealing cycle for removing all residual TL signals, especially at higher dose levels is necessary. Figure 3: Glow curve examples of 1 mgy and.1 mgy 137 Cs irradiation and their second readouts immediately after the first readout glow curves. 2 TL glow curve (cts) 18 16 14 12 1 8 6 1 mgy 1 mgy 2nd readout 1 µgy 1 µgy 2nd readout 4 2 3.4 k i Correction Factors Five k i (individual chip sensitivity correction) experiments are currently being carried out. All k i values of this production batch are between.93 and 1.3. Standard deviations of individual chips vary between.2% and 2.3%. The average standard deviation of the 14 chips in these five measurements 5

is 1.1%. It has to be noted that differences between consecutive k i values for a given chip are up to 8%. Typically, an increase in k i is followed by a decrease and vice versa. Therefore it is important to use average sensitivity correction factors instead of current values. No criteria are currently applied to remove chips with high k i values, fluctuating behaviour or long term drifts. Frequency of k i measurements is now increased for faster identification of problems in the chip sensitivities. 3.5 Dose and its Uncertainty Relative standard deviations for investigated dose levels (between 1 µgy and 3 mgy) are on average 1.1%, varying between.4% and 1.9%. These fluctuations are similar to the fluctuations in the k i values and are therefore inherent to the accuracy of the TL dosimetry method. Techniques may be introduced to reduce this uncertainty contribution, e.g. to eliminate chips with obvious instabilities in their k i history. Currently no criteria are applied, because the accuracy achieved is sufficient for most diagnostic applications. The only statistical test which is taken into account is to remove outliers whose values are more than three times off the standard deviation. At least three chips should be grouped together at a measurement point to get statistically trustable results. Dose dependent coefficients of variation values (relative standard deviations determined during the linearity experiment) may be applied for estimating statistical uncertainties of measurements with low number of detectors. A detailed uncertainty analysis based on the dose model of equation (1) is currently under development. The biggest contribution is expected to result from the energy response of the TL detector material in the range of diagnostic X-ray energies. Narrow filtered X-ray radiation showed energy responses up to 2% below the reference quality of 137 Cs radiation (662 kev). Precise values of energy correction factors k Q have to be determined for diagnostic X-ray radiation fields, a single correction factor for photon energies up to 15 kev may be useful. The total measurement uncertainty is expected to be within 1% at 95% confidence level. 3.6 Calibration Calibration chips are handled similarly to the other chips. Calibration chips readouts are distributed over the whole readout process to get the best average calibration factor taking possible drifts of the reader into account. It has been shown that there is a tendency of decreasing signal output over typical readout periods of about two hours taking the warm-up phase of the reader as well as empty and test chip readouts into account. Readouts of the linearity experiment presented in figure 4 show linear decreases of signals of about 1 to 2%. Figure 4: TL readout signals of 9 individual chips irradiated in a linearity test between 3 µgy and 3 mgy. Linear interpolations are shown at 1 and 3 mgy readouts over 4 hour total readout time. Second readouts are carried out but not shown here. 1.6E+8 1.4E+8 3 mgy 1.2E+8 TL signal (cts) 1.E+8 8.E+7 6.E+7 4.E+7 1 mgy 2.E+7 3 mgy 3 µgy.e+ 14: 15: 16: 17: 18: read-out time 6

A time dependent calibration function may be introduced for high accuracy measurements such as in radiotherapy applications. In our dosimetry system the small drift in the calibration procedure will be taken into account in the uncertainty of the average calibration factor. Figure 5 shows the glow curves of a single chip after a number of 1 mgy 137 Cs irradiations over a 7-month period. Glow curve shapes are almost identical proving stability of the dosimetry parameters over the time. Temperature shifts can be seen for the single chip over a number of readouts similar to different chips over a single readout. Shifts may depend on the precise chip positioning inside the reader, but may not influence the TL signal which corresponds to the area under the glow curve. Typical standard deviations of 1 calibration chips are close to 2%. Differences in TL signals of individual calibration chips are well below 2%. These fluctuations have not to be taken into account in the uncertainty analysis because calibration factors are calculated for each individual measurement experiment. It has to be mentioned that the calibration factor shows no long term drift up to now. Sensitivity changes due to inappropriate readout parameters such as a too high maximum temperature can therefore be excluded. It seems possible to use an average calibration factor without too large an increase in uncertainty, thus avoiding efforts to irradiate calibration chips at the 137 Cs nuclide facility at the very same time as the measurement chips. Figure 5: Comparison of uncorrected glow curves of chip number #1 which was used as a calibration chip (1 mgy 137 Cs) for a number of measurements. 14 TL glow curve (cts) 12 1 8 6 4 5.11.7 calib. chip #1 6.12.7 calib. chip #1 13.12.7 calib. chip #1 26.3.8 calib. chip #1 3.4.8 calib. chip #1 24.4.8 calib. chip #1 5.5.8 calib. chip #1 14.5.8 calib. chip #1 2.5.8 calib. chip #1 12.6.8 calib. chip #1 2 3.7 Dose Linearity Linearity of measured dose was investigated in the typical dose range of diagnostic applications. Ten chips were irradiated together at air kerma values up to several tens of mgy in a 137 Cs calibration field. The mean value of unirradiated background chips was subtracted. Variations of dose values were within 1% for dose levels greater than 1 µgy. Differences of determined low dose values compared to applied doses were 3%, 4%, and 1% for 3 µgy, 1µGy and 3 µgy, respectively. Lowest doses in this linearity study are close to the background level of unirradiated chips of about 2.4 µgy ( 137 Cs equivalent). (a) is chosen from the four listed possibilities of taking non-linearity into account:. (a) no linearity correction is applied, a relative uncertainty contribution due to non-linearity of up to 4%/Sqrt(1) at the lower dose limit of 1 µgy is taken into account (b) a non-linearity correction factor may be applied in the dose formula (1) depending on the measured dose, e.g. at 1 µgy equal 1/1.4=.96 (c) a background correction may be introduced, the optimized background dose value for this readout is 2.7 µgy resulting in a background correction factor of 1.12 (d) individual glow curves may be corrected by subtracting zero-dose glow curves or by applying computerized glow curve deconvolution. 7

Figure 6: Linearity graph relating measured dose to dose applied at the 137 Cs facility. Details in differences to the ideal linear dose dependence may not be visible due to the logarithmic scaling. 1. air kerma measured (mgy) 1. 1..1.1.1.1.1.1 1. 1. 1. air kerma applied (mgy) Figure 7: Examples of glow curves from the linearity experiment normalized to 1 mgy. Dose levels are between 3 and 3 µgy (upper figure) and.3 and 3 mgy (lower figure). Temperature shifts are corrected in these presentations for better comparability. Glow curves at lower doses show increasing signal contributions outside main peak 4. 14 normalized TL glow curve (cts/mgy) 12 1 8 6 4 2 3 µgy exposed chip 3 µgy exposed chip 3 µgy exposed chip 14 normalized TL glow curve (cts/mgy) 12 1 8 6 4 2 3 µgy exposed chip 3 mgy exposed chip 3 mgy exposed chip 8

4. Applications Detailed patient dose measurements are planned in different digital volume tomographs which are now introduced for 3D imaging of patient s jaws and teeth in dental practise in Austria. First free-air and in-phantom measurements were carried out in the rotating X-ray cone beam fields using adult patient examination settings [3]. The phantom positions were like the one of patients inside the scanner, see figure 8. Free-air exposure resulted in a mean air kerma values of about 4 mgy for a single scan. Phantom measurements revealed absorbed dose to water values of about 3 to 6 mgy depending on the position inside the phantom. Differences between ionisation chamber measurements and TL dose results were within several percent. Future experiments will include organ dose measurements with an anthropomorphic phantom. Studies of different volume tomographs and alternative examination devices such as panoramic units and CTs are planned comparing organ and resulting effective doses. Figure 8: Adult head sized phantoms positioned inside a digital volume tomograph, cylindrical PMMA calibration phantom (left) and Rando-Alderson head phantom borrowed from IAEA (right). REFERENCES [1] Council of the European Union, Directive on health protection of individuals against the dangers of ionizing radiation in relation to medical exposure, Council Directive 97/43/EURATOM, 3 th June 1997. [2] DIN Deutsches Institut für Normung e.v., Procedures of dosimetry with probe-type detectors for photon and electron radiation Part 5: Thermoluminescence dosimetry, DIN 68-5, Berlin, April 25 (in German). [3] Halda, M., Hranitzky, C., Neuwirth, J., Hefner, A., Stadtmann, H., Kindl, P., Dose measurements in 3D imaging systems, ÖGMP annual meeting and 3 rd AISCMP meeting, Graz, Austria, 3-31 May, 28. 9