P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Similar documents
P5100A & P5150 High Voltage Probes Performance Verification and Adjustments

Instruction Manual CT-6 High Frequency AC Current Probe

*P * ZZZ

P7313SMA 13 GHz Differential Probe

Instruction Manual. TDS 420A, TDS 430A & TDS 460A Digitizing Oscilloscope Performance Verification and Specifications

RMD5000 Rackmount Kit Instructions

AMP-13 OPERATOR S MANUAL

RM5 Rackmount Kit Instructions

TCPA300/400 Amplifiers & TCP300/400 Series AC/DC Current Probes Service Manual

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes

TCA-1MEG High Impedance Buffer Amplifier Instruction Manual

RMD2000 DPO2000 and MSO2000 Series Rackmount Kit Instructions

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

User Manual. TDS 420A, TDS 430A, TDS 460A & TDS 510A Digitizing Oscilloscopes

TETRIS 1000 High Impedance Active Probe. Instruction Manual

Instruction Manual. P6015A 1000X High Voltage Probe

Instruction Manual. P GHz Differential Probe

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

MODEL W Power Amplifier

TEGAM, INC. SINGLE/DUAL CHANNEL HIGH VOLTAGE AMPLIFIER MODEL 2340/2350. Instruction Manual PN# CD Publication Date: June 2006 REV.

VT1586A Rack Mount Terminal Panel Installation and User s Manual

2015 RIGOL TECHNOLOGIES, INC.

Advanced Test Equipment Rentals ATEC (2832)

TLA72X, TLA7XM, & TLA7016 Benchtop Mainframe Rackmount Installation

Instruction Manual. P7380SMA 8 GHz Differential Probe

Operator s Manual. PP016 Passive Probe

P GHz 10X Active Probe Service Manual

TETRIS User's Guide. High Impedance Active Probe DO177-1

Service Manual. P5210 High Voltage Differential Probe

AMP-12 OPERATOR S MANUAL

RM4000 Tektronix 4000 Series Rackmount Kit

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

DS1000E, DS1000D Series Digital Oscilloscope

BROADBAND LINEAR AMPLIFIER Model P150

PHV RO High Voltage Passive Probe. Instruction Manual

DPO3PWR, MDO3PWR and DPO4PWR Power Analysis Modules ZZZ User Manual

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

PKT 512A-RO High Impedance Passive Cable Divider

BROADBAND'LINEAR'AMPLIFIER WITH'A±B'INPUTS'AND'DC5OFFSET Model&A400X

BROADBAND LINEAR AMPLIFIER Model F1020 (models F10A and F20A combined)

RMD3000 DPO3000 Series Rackmount Kit

DUAL%CHANNEL%LINEAR%AMPLIFIER %WITH%PHASE%INVERTER Model&A400DI

BROADBAND'LINEAR'AMPLIFIER Model&F10AD

RM4000 DPO4000 Series Rackmount Kit

Operator s Manual. PP022 Passive Probe

N2792A and N2793A Differential Probes User s Guide

E2621A and E2622A Probe Adapters for Infiniium Oscilloscopes. User s Guide. Publication number E September 2002

Operator s Manual. PP017 and PP018 Passive Probes

DS1000B Series Digital Oscilloscope

34134A AC/DC DMM Current Probe. User s Guide. Publication number April 2009

N2790A Differential Voltage Probe

Instruction Manual. P7350SMA 5 GHz Differential Probe

INSTALLATION GUIDE. Video Balun Transceiver with fixed BNC for twisted pair operation with other balun transceivers or active receivers.

Instruction Manual. P5200 High Voltage Differential Probe

PCO-7114 Laser Diode Driver Module Operation Manual

Glass Electrode Meter

TA MHz oscilloscope probe TA MHz oscilloscope probe

TBS2000 Series Oscilloscopes Specification and Performance Verification

TK 100 Probe Calibrator. Instruction Manual

User Manual. TPS2PWR1 Power Analysis Application

Instruction Manual. TCA-1MEG High Impedance Buffer Amplifier

TDS1000 & TDS2000 Series Oscilloscopes Operator Training Kit Manual

TA MHz ±700 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

TCPA300/400 Amplifiers & TCP300/400 Series AC/DC Current Probes Instruction Manual

AFG2021 Arbitrary Function Generator Specifications and Performance Verification Technical Reference

TDS2000C and TDS1000C-EDU Series Digital Storage Oscilloscopes User Manual

TPS2000 Series Digital Storage Oscilloscope User Manual

Instruction Manual. P5205 High Voltage Differential Probe

TT-SI MHz Active Differential Probe

Model 4007DDS. 7 MHz Sweep Function Generator

TT-SI MHz DIFFERENTIAL PROBE

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

TDS3000 & TDS3000B Operator Training Kit Manual

Model 113 Scintillation Preamplifier Operating and Service Manual

P7700 Series TriMode Probes Technical Reference

Model 7000 Low Noise Differential Preamplifier

TT-SI MHz Active Differential Probe

TA MHz x10/x100 Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

MODEL 3810/2 Line Impedance Stabilization Network

Agilent N2780A, N2781A, N2782A, and N2783A Current Probes

Model 9302 Amplifier-Discriminator Operating and Service Manual

TT-SI 9001 / TT-SI MHz Active Differential Probes

400 MHz Passive High-Voltage Probe R&S RT-ZH

AP034-OM-E Rev D ISSUED: January 2000 ²

TBS1000B and TBS1000B-EDU Series Digital Storage Oscilloscopes User Manual

Model 9305 Fast Preamplifier Operating and Service Manual

Technical Reference. TDS 500D, TDS 600C, TDS 700D & TDS 714L Digitizing Oscilloscopes Performance Verification and Specifications

Directed Energy, Inc Oakridge Dr., Suite 100, Fort Collins, CO

RIGOL. User s Guide. RP1003C/RP1004C/RP1005C Current Probe. Mar RIGOL Technologies, Inc

Model 533 Dual Sum and Invert Amplifier Operating and Service Manual

Supplement. TDS5032 and TDS5034 Digital Phosphor Oscilloscopes

CLEANING CALIBRATION INTERVAL

ELECTROSURGICAL UNIT ANALYZER

TBS1000B and TBS1000B-EDU Series Digital Storage Oscilloscopes User Manual

TA MHz ±30 V Differential Probe User s Manual. This probe complies with IEC , IEC CAT I, Pollution Degree 2.

TDS2000C Series Digital Storage Oscilloscopes User Manual

Keysight N2790A Differential Voltage Probe. User Guide

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95

Signal Forge 2500M Frequency Expansion Module. 1.5 GHz to 2.6 GHz. User Manual

Bench Test Instruments and Handheld Oscilloscopes Basic Service

Transcription:

x P5100A & P5150 High Voltage Probes Performance Verification and Adjustments ZZZ Technical Reference *P077053002* 077-0530-02

xx P5100A & P5150 High Voltage Probes Performance Verification and Adjustments ZZZ Technical Reference Revision A www.tektronix.com 077-0530-02

Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. Contacting Tektronix Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA For product information, sales, service, and technical support: In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to find contacts in your area.

Warranty Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix. In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product. THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES. [W2 15AUG04]

Table of Contents Table of Contents General safety summary.............................................................................................................. iii Service safety summary... v Preface... vi P5100A Performance Verification... 1 Test Equipment... 1 Test Setup... 1 System DC Attenuation Accuracy Check....................................................................................... 2 Test Record... 4 P5100A Adjustments................................................................................................................. 5 DC Gain and High Frequency (HF) Compensation............................................................................. 7 Reassemble the Probe... 12 P5150 Adjustments.................................................................................................................. 13 P5100A & P5150 High Voltage Probes Technical Reference i

Table of Contents ii P5100A & P5150 High Voltage Probes Technical Reference

General safety summary General safety summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified. Only qualified personnel should perform service procedures. While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system. To avoid fire or personal injury Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source. Connect and disconnect properly. Connect the probe output to the measurement instrument before connecting the probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input. Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from the measurement instrument. Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. Observe all terminal ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Connect the probe reference lead to earth ground only. Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal. Do not operate without covers. Do not operate this product with covers or panels removed. Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualified service personnel. Avoid exposed circuitry. Do not touch exposed connections and components when power is present. Do not operate in wet/damp conditions. Do not operate in an explosive atmosphere. Keep product surfaces clean and dry. Provide proper ventilation. Refer to the manual's installation instructions for details on installing the product so it has proper ventilation. P5100A & P5150 High Voltage Probes Technical Reference iii

General safety summary Termsinthismanual These terms may appear in this manual: WARNING. Warning statements identify conditions or practices that could result in injury or loss of life. CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property. Symbols and terms on the product These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. The following symbol(s) may appear on the product: iv P5100A & P5150 High Voltage Probes Technical Reference

Service safety summary Service safety summary Only qualified personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures. Do not service alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present. Disconnect power. mains power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. To avoid electric shock, do not touch exposed connections. P5100A & P5150 High Voltage Probes Technical Reference v

Preface Preface This manual contains procedures for checking the P5100A probe performance and for making adjustments to achieve the warranted specifications. The P5150 probe section describes adjustment procedures only. WARNING. The procedures included in this manual are for use by qualified service personnel only. vi P5100A & P5150 High Voltage Probes Technical Reference

P5100A Performance Verification P5100A Performance Verification Use these performance verification procedures to verify that the probe meets the warranted specifications. These procedures should be performed in the order given using a DPO4104B oscilloscope with the equipment recommended in the following table. (See Table 1.) WARNING. The following instructions are for use by qualified service personnel only. To avoid electrical shock, do not disassemble or maintain the probe while it is connected to a signal source other than those specified in these procedures. NOTE. The probe must be low-frequency compensated before performance can be verified. If your probe fails to qualify under these conditions, service-level adjustments (DC gain and high-frequency compensation) may be required. These adjustments must be done by a trained service person. Test Equipment Table 1: Recommended test equipment Equipment Minimum requirements Recommended test equipment 1 Oscilloscope 500 MHz BW, 1 MΩ ±1% input R Tektronix DPO4104B Calibration generator 0.5% accuracy, 10 Hz square wave, 5V p p output Adapter, BNC Female-to-female barrel 103-0028-xx Adapter, probe tip-to-bnc Probe tip-to-male BNC 013-0291-xx Pulse generator 2 500 ps rise time, 100 ns pulse width, 3 V p p into 50 Ω Termination, 50 Ω feedthrough 2 50 Ω male-to-female BNC 011-0049-xx BNC cable, precision 2 50 Ω, 36in 012-0482-xx Adjustment tool 2 Insulated, straight edge 003-1433-xx Screwdriver 2 Knife 2 Flat blade Razor edge 1 Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers 2 Required for adjustment procedures only Fluke 9500B with 9530 active head Picosecond Labs 2600C or Fluke 9500B with 9530 active head Test Setup 1. Turn on the oscilloscope. 2. Connect the probe to any channel (1 4) of the oscilloscope. 3. Turn on the generators and let the probe and test equipment warm up for 20 minutes. 4. Make a copy of the test record to tabulate the test results. (See Table 2 on page 4.) P5100A & P5150 High Voltage Probes Technical Reference 1

P5100A Performance Verification System DC Attenuation Accuracy Check 1. Connect the test setup as shown. Figure 1: Setup for oscilloscope gain check 2. Set the oscilloscope to: vertical to 10 mv/division horizontal to 20 ms/division averaging to 64 3. Set the calibration generator to: output frequency = 10 Hz 50 mv p-p square wave output output into 1 MΩ load 4. Enable the calibration generator output. 5. Trigger the oscilloscope and center the waveform on the screen. 6. Use the amplitude measurement selection on the oscilloscope to measure the generator output voltage. 7. Divide the measured output from the previous step into 50 mv to derive a correction factor, CF. For example, if you measured 50.13 mv, then the CF = 50/50.13 = 0.997. The correction factor will be used in step 14 of this procedure to calculate the corrected amplitude, V corr. 2 P5100A & P5150 High Voltage Probes Technical Reference

P5100A Performance Verification 8. Disconnect the calibration generator output from the oscilloscope and connect the probe to the same channel, as showninthetest setup below. You must use the same oscilloscope channel throughout this procedure for the test to be accurate and valid. Figure 2: Setup for system accuracy check 9. Set the oscilloscope vertical to 1 V/division. 10. Set the calibration generator to 5 V p-p square wave output. 11. Enable the calibration generator output. 12. Trigger the oscilloscope and center the waveform on the screen. 13. Use the amplitude measurement selection on the oscilloscope to measure the voltage amplitude. 14. Calculate the corrected amplitude, V corr by multiplying the measured output from the previous step by the correction factor (CF) derived in step 7. For example, a measured output of 5.04 V multiplied by a correction factor of 0.997 yields a corrected amplitude of 5.025 V. 15. Verify that the corrected amplitude, V corr, is 5 V, ±0.088 V. Record the results in the test record. If your probe does not meet the specification, service-level adjustments (DC gain and high-frequency compensation) may be required. (See page 5, P5100A Adjustments.) The procedures to make adjustments and to optimize the bandwidth and high frequency response must be done by a qualified service technician. P5100A & P5150 High Voltage Probes Technical Reference 3

P5100A Performance Verification Test Record Photocopy this test record for recording the results of the performance verification procedures. Table 2: P5100A probe test record Probe Model Number: P5100A Probe Serial Number: Temperature: Certificate Number: RH %: Technician: Date of Calibration: Probe test Minimum Incoming Outgoing Maximum System DC Attenuation Accuracy 4.912 V 5.088 V 4 P5100A & P5150 High Voltage Probes Technical Reference

P5100A Adjustments P5100A Adjustments To gain access to the adjustments, you must disassemble the probe compensation box as described below. 1. Using a razor knife, cut in half the two labels on the sides of the compensation box, along the top and bottom cover seams. Be careful to not damage the labels; they must remain intact and on the probe due to regulatory requirements. Figure 3: Cut the two side labels in half 2. Separate the covers by inserting a thin flat-blade screwdriver into the one of the slots on the side of the compensation box. The slots are hidden beneath the labels. Figure 4: Separate the covers P5100A & P5150 High Voltage Probes Technical Reference 5

P5100A Adjustments 3. Lift the cable end of the top cover away from the compensation box and then pull the cover out of the connector housing. Figure 5: Remove the top cover from the probe The five adjustments that you will likely need to access are available through the openings in the metal shield. If you need to access the sixth adjustment (fine LF comp), you must slide the metal shield away from the connector end of the probe. Figure 6: P5100A probe adjustment locations 6 P5100A & P5150 High Voltage Probes Technical Reference

P5100A Adjustments DC Gain and High Frequency (HF) Compensation Precision or high-frequency compensation is not a common user adjustment. However, precision compensation should be checked if the probe was repaired or if any of the following conditions are true: The probe has excessive high-frequency aberrations. You have installed the probe on an oscilloscope having an input capacitance near the limits of the probe compensation range. To perform the high-frequency compensation adjustment you will need a signal source that has all of the following characteristics: Fast rise output with rise time 500 ps Output properly terminated The Picosecond Labs 2600C Pulse Generator output meets these requirements when properly terminated. P5100A & P5150 High Voltage Probes Technical Reference 7

P5100A Adjustments Gain Adjustment (DC Accuracy) 1. Connect the probe as shown in the System DC Attenuation Accuracy Check setup. (See Figure 2 on page 3.) 2. Set the oscilloscope to: vertical to 1 V/division horizontal to 20 ms/division averaging to 64 3. Set the calibration generator to: output frequency = 10 Hz 5V p-p square wave output output into 1 MΩ load 4. Locate R9 (the DC gain adjustment). WARNING. To reduce risk of electric shock, use only the provided insulated adjustment tool when making adjustments. Hazardous voltages may be present inside the compensation box. Figure 7: R9 (DC gain) adjustment location 5. Adjust the DC gain (R9) to a probe output level of 5 V, ±0.088 V, as displayed on the oscilloscope. 6. Record the output level in the outgoing column of the System DC Attenuation Accuracy check in the test record. 8 P5100A & P5150 High Voltage Probes Technical Reference

P5100A Adjustments Low-Frequency Compensation (LF) Adjustment 1. Connect the probe to the Probe Compensation output of the oscilloscope and check the LF comp of the probe. If it needs adjustment, do the following: Adjust LF compensation with C2, and if necessary, do a minor adjustment to C116 for best LF flatness. Figure 8: C2 (LF) and C116 (fine LF) adjustment locations Use C2 to check LF compensation range over and under compensation range. Use C2 for a coarse LF adjust and use C116 to get a consistent roll up or roll off. Finally, go back and use C2 one last time to flatten the overall response. 2. Proceed to the High Frequency (HF) Adjustment. (See page 10.) P5100A & P5150 High Voltage Probes Technical Reference 9

P5100A Adjustments High Frequency (HF) Adjustment 1. Connect the test setup as shown. Figure 9: Setup to adjust HF response 2. Set the output level of the pulse generator to 5 volts (20 db) and the pulse width to 100 ns. 3. Set the oscilloscope to: vertical to 1 V/division horizontal to 4 ns/division averagingto64 input impedance to 1 MΩ 4. Adjust the HF response as follows: 10 P5100A & P5150 High Voltage Probes Technical Reference

P5100A Adjustments Figure 10: Areas of the waveform affected by adjustments R1, R2, and R3 First adjust R1 to flatten out to 20-30 ns. Next, adjust R3 for the 10-20 ns area. Adjust R2 for the last ~5 ns. If the overshoot is too great, then adjust R2 to roll off the front corner and readjust R3 followed by R1. The procedure is done when the 3 adjustments have been optimized for the best overall response. 5. Reassemble the probe. (See page 12, Reassemble the Probe.) P5100A & P5150 High Voltage Probes Technical Reference 11

P5100A Adjustments Reassemble the Probe If you removed the metal shield to access the fine LF adjustment, it is easier to reinstall the shield with the bottom cover removed, so begin with step 1. Otherwise, skip to step 3. 1. Remove the bottom cover. 2. Install the shield over the circuit board so that the shield covers the metal ends of the board housing. (See Figure 11.) 3. Using your thumb or finger to hold the shield in place, align and seat the probe into the bottom cover, beginning at the cable end. As you seat the probe into the connector end of the cover, pull out the spring-loaded connector to allow it to clear the front edge of the cover and fully seat into the bottom cover. Figure 11: Align and seat the probe into the bottom cover 4. Verify that the metal shield is properly located. (See Figure 11.) Align the top cover onto the probe, beginning at the cable end. As you seat the cover onto the connector end of the probe, pull out the connector to allow the cover to clear the connector and to engage with the bottom cover. Figure 12: Install the top cover 5. Recheck that the probe functions properly. End of procedure. 12 P5100A & P5150 High Voltage Probes Technical Reference

P5150 Adjustments P5150 Adjustments These procedures describe how to make low- and high-frequency compensation adjustments to the P5150 probe. Use the equipment listed below for the test setup. Table 3: Recommended test equipment Equipment Minimum requirements Recommended test equipment 1 Oscilloscope 2 200 MHz bandwidth (floating input) 500 MHz bandwidth Tektronix TPS2024B or THS3024 Tektronix TDS3054C (ground-referenced input) Pulse generator 500 ps rise time, 100 ns pulse width, 3 V p p into 50 Ω Picosecond Labs 2600C or Fluke 9500B with 9530 active head Termination, 50 Ω feedthrough 50 Ω male-to-female BNC 011-0049-xx Adapter, BNC Female-to-female barrel 103-0028-xx Adapter, probe tip-to-bnc Probe tip-to-male BNC 013-0291-xx BNC cable, precision 50 Ω, 36in 012-0482-xx Adjustment tool Insulated, straight edge 003-1433-xx Screwdriver Flat blade 1 Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers 2 P5150 probes are intended for use with oscilloscopes that have isolated (floating) inputs. The Tektronix TPS2024B and THS3024 oscilloscopes are the highest bandwidth floating-input oscilloscopes currently available. If you do not have access to one of these oscilloscopes, use the recommended ground-referenced model. Access the Adjustments To gain access to the adjustments, you must slide the cover off of the probe body. Insert a thin flat-blade screwdriver into the slot on the top of the compensation box and slide back the cover. (See Figure 13.) Figure 13: Depress the tab and slide back the cover P5100A & P5150 High Voltage Probes Technical Reference 13

P5150 Adjustments Low-Frequency Compensation (LF) Adjustment 1. Connect the probe to the compensation output of the oscilloscope and check the LF comp of the probe. If it needs adjustment, do the following: Use C2 to check that the LF compensation range spans above and below the optimized response. Adjust LF compensation with C2, and if necessary, do a minor adjustment to C116 for best LF flatness. Figure 14: LF and fine LF adjustment locations Use C2 for a coarse LF adjust and use C116 to get a consistent roll up or roll off. Finally, go back and use C2 one last time to flatten the overall response. 2. Proceed to the High Frequency (HF) Adjustment. (See page 15, High Frequency (HF) Adjustment.) 14 P5100A & P5150 High Voltage Probes Technical Reference

P5150 Adjustments High Frequency (HF) Adjustment 1. Connect the probe and equipment as shown in the High Frequency (HF) Adjustment setup for the P5100A probe. (See Figure 9 on page 10.) 2. Set the oscilloscope horizontal to 4 ns/div. (Adjust the oscilloscope as necessary to view the leading edge of the waveform.) 3. Adjust the HF response as follows: Figure 15: Areas of the waveform affected by adjustments R1, R2, and R3 First adjust R1 to flatten out to 20-30 ns. Next, adjust R3 for the 10-20 ns area. Adjust R2 for the last ~5 ns. If the overshoot is too great, then adjust R2 to roll off the front corner and readjust R3 followed by R1. The adjustments are complete when you have optimized the waveform. 4. Slide the probe cover back over the probe body until the body tab engages with the cover slot. End of procedure. P5100A & P5150 High Voltage Probes Technical Reference 15