Operating manual. Force transducer with strain gauge measuring system U9B. B 20. U9B.40 en

Similar documents
Operating manual. Force transducer with strain gage measuring system C9B. A en. hvordan man bruger astrologi med tibetanske medicin

SUNSTAR 传感与控制

U9C. Data sheet. Force Transducer. Special features. B en

U2B. Data sheet. Force Transducer. Special features. B en

U2B. Force Transducer. Special features. Data sheet

U15. Force transducer. Special features. Data sheet. Installed dimensions of connection variants

C9C. Force Transducer. Special features. Data sheet

Voice-Frequency Telegraph Modem HY101-SWT

Data sheet. P3 Top Class BlueLine. P3MBP BlueLine. Ultra-high pressure transducers to 15,000 bar. Special features

INSTRUCTIONS FOR QUALIFIED PERSONNEL ONLY!

Operating Manual Series Load Pins

KMT - Kraus Messtechnik GmbH

U10M. Force Transducer. Special features. Data sheet. Mounting dimensions of the connection variants in mm [inch]

MDEX. on customized sensors. Technical information. Torque transducer. CS-T-04-MDEX Special features

Operating manual. 600 Hz Measuring Amplifier ME30. A en

C10. Data sheet. Force transducer. Special features. B de

Specifications Type WA2 WA10 WA20 WA50 WA100 WA200 WA300 WA500 Nominal displacement mm Nomi

KMT - Kraus Messtechnik GmbH

Inductive Standard Displacement Transducers

S9M. Force Transducer. Special features. Data sheet

E *) F 15. H 12 h9. 5.5f7 8h6. a./f approx. 0.5mm Free stroke. Displacement. Initial stroke

T40FH. Torque flange. Special features. Data sheet

T22. Data Sheet. Torque transducer. Special features. Installation example with two bellows couplings. B en

KMT - Kraus Messtechnik GmbH

T40FM. Data Sheet. Torque flange. Special features. Overall concept. B en

T22. Torque transducer. Special features. Data sheet

T10F. Data Sheet. Torque Flange. Special features. Installation example T10F. B en

T10FS. Data Sheet. Torque Flange. Special features. Installation example T10FS. B en

alfa-clamp Tool Assembly Device

Data Sheet. Absolute pressure transducer Nominal (rated) pressure 10 bar to 3000 bar. Special features

Single point load cell

Load cell. Special features. Data sheet

KMT - Kraus Messtechnik GmbH

RSCC. Data Sheet. Load cells. Special features. Diagram of RSCC load cell. Design: 500 kg to 5 t. Design: 50 kg to 200 kg. B

MANUAL. GPS Antenne. Antenna/Cnverer Unit Meinberg Radio Clocks GmbH & Co. KG

Telemetrie-Messtechnik Schnorrenberg

Operating Instructions

T40B. Torque Flange. Special features. Data sheet. Overall concept

Telemetrie-Messtechnik Schnorrenberg

Single point load cell

Installation Guide. English. FS62 Surface Mountable Strain Sensor

100/200/400 t. Self-centering pendulum lead cells. Data sheet. Special features

Melt pressure Transmitter GREENLINE Type DTAI with Temperature element

electronics Data Sheet Industrial Amplifier Special features Dimensions (in mm; 1 mm= inches) B en

Operating Instructions for Pressure Sensor with Ceramic Sensor Element. Model: SEN-9601

Operating manual GTL 241 GTL 251. Temperature probe. Please keep the manual for future use.

Technical data Model F1301 F13C1 ATEX/IECEX 2) F13C1 (Option) Nominal load F nom 10 / 20 / 30 / 50 / 100 / 200 kn 10 / 20 / 30 / 50 / 100 / 200 kn 10

Original operating instructions Fail-safe inductive sensor GG507S

CANHEAD. Data Sheet. Special features. Distributed measurement acquisition. B en

Terms and expressions for specifying torque transducers

Installation and Operational Instructions for ROBA -switch Type 017._00.2

Precipitation Monitor ,

Multipoint measuring units UPM100 and Centipede100

Pressure transmitter for general industrial applications Model A-10

Precipition Sensor with analogue Intensity Output ,

MX1615. Data sheet. Bridge/strain gauge amplifier

XPC10 High temperature miniature pressure sensor

Tension/compression force transducer. With thin film technology up to 100 kn. Models F2301, F23C1, F23S1. Force. Applications.

Magnetic Inductive Flow Sensor induq

Instruction manual for STA 1 sectional door operator

Z6... Data Sheet. Load cell. Special features. Dimensions (in mm; 1 mm = inches) B en 1

Jetronic-BOX HV. R/C Controller Doorsequencer with Air Failsave [0288] 2013 evojet GmbH Am Parir 4A D Langerwehe/Germany

Tension/compression force transducer. With thin film technology up to 100 kn. Models F2301, F23C1, F23S1. Force. Applications.

Installation and Operational Instructions for ROBA -multiswitch Type 019._00.2

Specifications (in accordance with DIN 16086) Equipment family Digibar II Type 1-PE50A1/...; K-PE PE00A1/...; K-PE00... Class of accuracy 0.2 0

Original operating instructions Fail-safe inductive sensor GG507S / / 2013

SAUTER HD Version /2015 GB

Installation Guide. English. FS62 Embedded Strain Sensor

Installation and Operating Instructions for Phase Demodulator Type _.2

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

Pressure transmitter for general industrial applications Model A-10

Series 7000 Torque Sensor for PTO-shafts

Transducer transmitter BILT 4 II 2(1) G. Technical Manual

External brake resistor AX2090-BW5x

MX1615B R. Data Sheet. Ultra-rugged Bridge Amplifier

High-quality pressure transmitter For general industrial applications Model S-20

NOVOHALL Rotary Sensor non-contacting. Series RSC2800 analog

Oil lubrication The gears and bearings are lubricated by means of combined oil splash and mist lubrication. Caution! The unbalance exciters are not

Capacitive level meters CLM 40 INSTRUCTIONS MANUAL. průmyslová elektronika

GREEN PRODUCT It has been RoHS Compliant SDR-2716 WT-716/WM-716 VHF PLL 16 CHANNEL WIRELESS SYSTEM OPERATING MANUAL

Strain Gauge Measuring Amplifier BA 662

WATERFLUX 3000 Quick Start

Operating Instructions TIC 251/252. Profibus DP-Gateway PG 0009 BE/E (0903)

Instruction Manual. SENSORS & SYSTEMS Authority in Displacement Measuring. Non-contact eddy-current displacement and position measurement

E8EB-N0C2B E8EB-N0B2B

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

Glass-fibre-reinforced polycarbonate. Dead zone Adaptive or fixed from % 1.5 s adjustable. 20 ms. 60 ms. 60 ms. Approx. 1.

WATERFLUX 3000 Quick Start

Shear beam With thin-film technology from 2 kn Models F3301, F33C1, F33S1

Operating manual. Digital Transducer-Electronics AED 9301A. Part 1, Profibus basic device

MX1615B-R. Ultra-rugged Bridge Amplifier. Special features. Data sheet. Block diagram

Instruction Manual MSC710 MSC710-U MSC710-I

Resistance Thermometers Model Series TR7X0, Sheathed Design

Original operating instructions Fail-safe inductive sensor GM504S / / 2010

User Manual AED9501A. Digital Transducer Electronics Basic device CANOpen/DeviceNet. I en

T12. Data sheet. Digital Torque Transducer. Special features. Signal flow block diagram. B en

Flush pressure transmitter For viscous and solids-containing media Model S-11

ADDUCO Hydraulic Clamping Nut

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

Torque Sensor DML 500 S080S10X

Transcription:

Operating manual Force transducer with strain gauge measuring system U9B B 20. U9B.40 en

U9B Contents 3 Page Safety instructions.............................................. 4 1 Scope of delivery......................................... 7 2 Field of application and notes on use....................... 7 3 Structure and mode of operation........................... 8 3.1 Measuring element......................................... 8 3.2 Measuring procedure, output signal........................... 8 4 Conditions at the site of installation........................ 8 4.1 Ambient temperature....................................... 8 4.2 Humidity.................................................. 9 4.3 Deposits.................................................. 9 4.4 External pressure.......................................... 9 5 Mechanical installation.................................... 10 5.1 Important precautions during installation....................... 10 5.2 General guidelines for installation............................ 10 5.2.1 Mounting for tensile and compressive loads.................... 10 6 Electrical connection...................................... 11 6.1 Notes on wiring............................................ 11 6.2 Allocation of the cable cores................................. 11 6.3 Connection technique....................................... 12 6.4 Cable extensions........................................... 13 6.5 Parallel connection of more then one transducers............... 13 7 Technical data............................................ 14 8 Dimensions (in mm)....................................... 15 9 Copy of Declaration of Conformity.......................... 16

4 U9B Safety instructions Use in accordance with the regulations Force transducers in the U9B range are designed for measuring static and dynamic compressive forces. Use for any additional purpose shall be deemed to be not in accordance with the regulations. In the interests of safety, the transducer should only be operated as described in the Mounting Instructions. It is also essential to observe the appropriate legal and safety regulations for the application concerned during use. The same applies to the use of accessories. The transducer is not a safety element within the meaning of its use as intended. Proper and safe operation of this transducer requires proper transportation, correct storage, assembly and mounting and careful operation and maintenance. General dangers due to non observance of the safety instructions The U9B force transducer corresponds to the state of the art and is fail safe. The transducers can give rise to residual dangers if they are inappropriately installed and operated by untrained personnel. Everyone involved with the installation, commissioning, maintenance or repair of a force transducer must have read and understood the Mounting Instructions and in particular the technical safety instructions. Residual dangers The scope of supply and performance of the transducer covers only a small area of force measurement technique. In addition, equipment planners, installers and operators should plan, implement and respond to the safety engineering considerations of force measurement technique in such a way as to minimise residual dangers. Prevailing regulations must be complied with at all times. There must be reference to the residual dangers connected with force measurement technique.

U9B 5 In these mounting instructions residual dangers are pointed out using the following symbols: Symbol: DANGER Meaning: Highest level of danger Warns of a directly dangerous situation in which failure to comply with safety requirements will lead to death or serious physical injury. Symbol: WARNING Meaning: Possibly dangerous situation Warns of a potentially dangerous situation in which failure to comply with safety requirements can lead to death or serious physical injury. Symbol: ATTENTION Meaning: Possibly dangerous situation Warns of a potentially dangerous situation in which failure to comply with safety requirements could lead to damage to property, slight or moderate physical injury. Symbol: NOTE Refers to the fact that important information is being given about the product or its use. Symbol: Meaning: CE mark The CE mark signals a guarantee by the manufacturer that his product meets the requirements of the relevant EC directives (see Declaration of conformity at the end of this document).

6 U9B Prohibition of own conversions and modifications The transducer must not be modified from the design or safety engineering point of view except with our express agreement. Any modification shall exclude all liability on our part for any damage resulting therefrom. Qualified personnel This instrument is only to be installed by qualified personnel strictly in accordance with the technical data and with the safety rules and regulations which follow. It is also essential to observe the appropriate legal and safety regulations for the application concerned. The same applies to the use of accessories. Qualified personnel means persons entrusted with the installation, fitting, commissioning and operation of the product who possess the appropriate qualifications for their function. Conditions on site Protect the transducer from damp and weather influences such as rain, snow, etc. Maintenance The U9B force transducer is maintenance free. Accident prevention Although the specified nominal force in the destructive range is several times the full scale value, the relevant accident prevention regulations from the trade associations must be taken into consideration.

U9B 7 1 Scope of delivery Force transducer U9B; mounting instruction 2 Field of application and notes on use The transducers are intended for measuring static and dynamic tensile and compressive forces. Their very small dimensions make them ideal for use where little space is available (for example for measuring penetration forces in manufacturing processes). The force transducers are designed for harsh environmental conditions and stringent service in requirements. They are maintenance-free and can even be installed in places where access is difficult. Their electrical measuring signals can be transmitted to remote measuring stations and control rooms and displayed or recorded there and used for control and regulating duties. The transducers, as precision measuring instruments, require careful handling during transport and installation. Loading impacts (for example free fall application of the load) even during measurement operations can lead to unexpected overloads with permanent damage. Special care must therefore be given to the base of the housing, which is relatively thin. The limits for permissible mechanical, thermal and electrical stressing are given in the technical data. It is essential that these be respected.

8 U9B 3 Structure and mode of operation 3.1 Measuring element The force is transferred to the diaphragm measuring element in the U9B by a threaded bolt. This type of measuring element has the advantage in that it also acts as a housing, protecting the applied strain gauges. The transducer is hermetically sealed by an adapter with threaded bolt which is welded on the bottom. 3.2 Measuring procedure, output signal The measuring spring and strain gauges are elastically deformed by a load acting in the direction of measurement. The strain gauges modify their resistance proportionally to their change in length. The Wheatstone bridge is thus detuned. When a bridge excitation voltage is applied, the circuit supplies an output signal proportional to the change in resistance and hence also proportional to the force applied. A measuring amplifier, suitable for strain gaugs measurements, is needed to process the measurement signal. 4 Conditions at the site of installation 4.1 Ambient temperature To achieve optimum measuring results, the nominal temperature range from -10 C to +70 C must be maintained. Ideally, the temperatures should be constant or subject to slow variations. The temperature effects indicated are applicable in compliance with VDI/VDE Recommendation 2637 when the temperatures do not change faster than 5 K/h.

U9B 9 Temperature gradients in the transducer by cooling or heating on one side (radiant heat) cause problems. A screen to protect against radiant heat and thermal insulation on all sides produce marked improvements. Naturally, these should not cause any force shunt. NOTE: The transducer output signal can be affected by the warmth of the hand. If measurements are made at room temperature (about 20 C), then you should wait about 15 min. after handling the transducer before taking measurements. 4.2 Humidity Ambient humidity and a tropical climate do not affect the function of the transducer, according to the classification of the protection indicated, corresponding to protection class IP 67 to EN 60 529. IP 67 means protected against the ingress of dust. It is also protected against water, if the transducer is immersed in water under specified conditions with regard to pressure and time (0.5 h in 1 m depth of water). The housings of the transducers are manufactured entirely in stainless steel. They can be used in harsh environments. The sheath of the connection cable consists of polyurethane (PUR). 4.3 Deposits Dust, dirt and other foreign bodies should not be allowed to accumulate to the point where they convey part of the measuring force to the housing, thus falsifying the measured value (force shunt). 4.4 External pressure The external atmospheric pressure can lie between 0 and 5 bars. Please note that pressure fluctuations can displace the zero point. Nennlast kn 0.5 1 2 5 10 50 Change in zero point for a change of 10mbar in ambient pressure (related to the nominal load) % 0.01 0.006 0.01 0.004 0.002 0.001

10 U9B 5 Mechanical installation 5.1 Important precautions during installation Handle the transducers carefully. The force application surfaces on the transducer and on the mounting accessories must be absolutely clean and seat properly. Do not overload the transducer, even briefly. Also short-term overloads e.g. due to unevenly distributed support loads must be avoided. Each transducer should be shunted by a stranded copper cable (approx. 50mm 2 ) during or immediately after installation. For this purpose, HBM supplies very flexible earthing cable EEK 5.2 General guidelines for installation Force must act on the force transducer precisely in the direction of measurement as far as possible. Off-centre loading and side loading or lateral forces are interference variables, and therefore causes for measuring errors, and they can damage the transducer by exceeding the permissible limits. Side loads and lateral forces also include the relevant components of any measurement quantities which may be introduced at an oblique angle. 5.2.1 Mounting for tensile and compressive loads The transducer can measure axial forces in the tensile and the compressive directions. Even alternating loads are measured properly. To achieve this though, the transducer must be mounted without any axial play. Tensile and/or compressive forces are transferred via the two threaded bolts. The force transfer parts are locked against the transducer using the hexagon nuts included with the supplied items. To do this, the transducer is loaded with the nominal force and the nuts tightened with the relevant tightening torque: Nominal force (kn) Tightening torque (Nm) 0.5...1 8 2...20 40 50 200 As an installation aid, HBM recommends the knuckle eyes (see Chapter 9 Dimensions ) which prevent transfer to the transducer of torsion and bending moments as well as lateral forces. With dynamic loading pliable tie rods should be used.

U9B 11 6 Electrical connection 6.1 Notes on wiring Electrical and magnetic fields often cause the introduction of disturbing voltages into the measuring circuit. These disturbances are chiefly caused by heavy current conductors installed parallel to the measuring lines, but they can also be produced by contactors or electric motors in the vicinity. Also interference voltages can be occur along the electrical path, in particular through earthing the measuring system at a number of points, causing differences in potential. To avoid noise injection,please note the following: Use only screened, low capacitance measuring (cable from HBM fulfills these requirements). Do not lay the measuring cable parallel to power and control lines. If this is not possible (e.g. in cable ducts), the measuring cable can be protected, e.g. by steel conduit and a minimum distance of 50cm is maintained to the other cables. The stray fields of transformers, motors and contactors must be avoided. Do not wire the transducer, amplifier and display device to multiple earths. All equipment in the measuring system should be connected to the same earth conductor. Further information can be taken from our reprint Electromagnetic Compatibility EMC 6.2 Allocation of the cable cores The 1.5 m long connection lead of the transducer has colour-coded free core ends. The cable screen (YE) should be connected to the transducer ground. If the transducer is connected according to the information given in the table, the output voltage from the amplifier is positive for a compression load on the transducer. If a negative output voltage is required at the measuring amplifier during compressive loading, cores white and red of the transducer should be changed over at the amplifier input.

12 U9B 6.3 Connection technique The transducers are fitted with a four-core connection cable and calibrated as standard using the four-wire circuit wh (white) Measuring signal (+) bk (black) Bridge excitation voltage ( ) rd (red) Measuring signal ( ) bu (blue) Bridge excitation voltage (+) ye (yellow) Cable screen Fig. 6.1: Transducer with four-core cable wh (white) Measuring signal (+) Sensor circuit bk (black) Bridge excitation voltage ( ) rd (red) Measuring signal ( ) bu (blue) Bridge excitation voltage (+) Sensor circuit ye (yellow) Cable screen Fig. 6.2: Transducer with four-core cable, amplifier in six wire technique ATTENTION: There are balancing resistors under a coloured heat-shrink sleeve at the end of the cable that must not be removed (re-use when shortening the cable).

U9B 13 6.4 Cable extensions Extension cables must be of the screened low-capacitance type (HBM cables fulfil this requirement). For cable extensions, care must be tacken to ensure a satisfactory connection with the lowest transfer resistance and good insulation. The plug connectors from HBM fulfil these requirements. If special humidity protection is required, e.g. the KVM Cable Joining Sleeves (soldered and potted joints) or the VKK cable connection box (screwed connections in cast housing) can be used. 6.5 Parallel connection of more then one transducers The parallel wiring of a number of transducers is normally possible. This might lead to measurement errors through due to different output resistances. When wiring transducers in parallel, cores of the same colour are connected together. The overall arrangement of transducers having the same nominal force with even load distribution can be loaded with the sum of the individual forces. The sensitivity of the complete circuit corresponds to the sensitivity of a single transducer.

14 U9B 7 Technical data Type Nominal force kn 0.5 1 2 5 10 20 50 Accuracy class 0.5 Nominal sensitivity C nom mv/v 1 Relative sensitivity deviation d c % v"1 tens. / v"2 compress. Effect of temperature per 10K on TK C sensitivity in the nominal temperature range in the service temperature range % % v "0.5 v "0.8 Effect of temperature per 10K on zero signal in the nominal temperature range in the service temperature range TK 0 % % U9B v "0.5 v "0.8 Linearity % v "0.5 Hysteresis related to measuring v "0.5 range limit U % Span in fixed mounting orientation % v "0.5 Creep at nominal load and reference d crf+e temperature over 30 min % v "0.2 Input resistance (BK-BU) at reference R e temperature Ω u 345 Output resistance (RD-WH) at reference R a temperature Ω 300 400 Insulation resistance R Is GΩ u 1 Service range of supply voltage B U,G V 0.5...12 Reference supply voltage U ref V 5 Reference temperature t ref C [ F] +23 [+73.4] Nominal temperature range B t,nom C [ F] 10..70 [+14...+158] Service temperature range B t,g C [ F] 30...+85 [ 22...+185] Storage temperature range B t,s C [ F] 30...+85 [ 22...+185] Protection class to EN 60 529 IP 67 Nominal measurement displacement "15% S nom mm t 0,1 Natural frequency "15% khz 15.5 23.7 18.7 20 23 27.8 20 Service load (F G ) % 120 Breaking load (F B ) % u200 Relative static side-load limit 1) (F Q ) % 40 20 Permissible vibration amplitude to DIN 50 100 F rb % 70 40 Weight, approx. g 65 100 400 Cable length m 1.5 1 ) referred to the load transfer point 2mm above diaphragm

U9B 15 8 Dimensions (in mm) U9B; for nominal forces 2 50kN M R B F E Z C Cable Ø 3 Cable length 1,5 m A a.f 9 G F R M a.f Y Nominal force A 0.1 B C E F G M R SW Y Z 0.5...1 kn 26 44.5 20.5 13 9.5 13.5 M5 20 8 ca.5.5 2.5 2...20 kn 26 60 28.5 21 16 21 M10 40 17 ca.5.5 5 50 kn 46 84 40 28 21.5 28 M16 x 1.5 80 24 ca.5.5 8 Knuckle eye ZGW, stainless (Accessories) W H D L A F G B M AF J K for Nominal A B D F G H J K L M A.F W force 0.5...1 kn 10 5 H7 18 27 36 6 9 11 4 M5 9 8 2...20 kn 20 10 H7 28 43 57 10.5 15 19 6.5 M10 17 14 50 kn 28 16 H7 42 64 85 15 22 27 8 M16x1.5 22 21

16 U9B 9 Copy of Declaration of Conformity HOTTINGER BALDWIN MESSTECHNIK HOTTINGER BALDWIN MESSTECHNIK GMBH Im Tiefen See 45 D-64293 Darmstadt Tel. ++49/6151/803-0, Fax. ++ 49/6151/894896 Konformitätserklärung Declaration of Conformity Déclaration de Conformité Document: 065/05.1996 Wir, We, Nous Hottinger Baldwin Messtechnik GmbH, Darmstadt erklären in alleiniger Verantwortung, daß das Produkt declare under our sole responsibility that the product déclarons sous notre seule responsabilité que le produit Kraftaufnehmer der Typenreihe U9B (C9B und Sondertypen) auf das sich diese Erklärung bezieht, mit der/den folgenden Norm(en) oder normativen Dokument(en) übereinstimmt (siehe Seite 2) gemäß den Bestimmungen der Richtlinie(n) to which this declaration relates is in conformity with the following standard(s) or other normative document(s) (see page 2) following the provisions of Directive(s) auquel se réfère cette déclaration est conforme à la (aux) norme(s) ou autre(s) document(s) normatif(s) (voir page 2) conformément aux dispositions de(s) Directive(s) 89/336/EWG Richtlinie des Rates vom 3. Mai 1989 zur Angleichung der Rechtsvorschriften der Mitgliedsstaaten über die elektromagnetische Verträglichkeit, geändert durch 91/263/EWG, 92/31/EWG und 93/68/EWG Erstmalige Anbringung der CE- Kennzeichnung: 1995 Die Absicherung aller produktspezifischen Qualitätsmerkmale erfolgt auf Basis eines von der DQS (Deutsche Gesellschaft zur Zertifizierung von Qualitätsmanagementsystemen) seit 1986 zertifizierten Qualitätsmanagementsystems nach DIN ISO 9001 (Reg.-Nr. DQS-10001). Die Überprüfung der sicherheitsrelevanten Merkmale (Elektromagnetische Verträglichkeit, Sicherheit elektrischer Betriebsmittel) führt ein von der DATech erstmal 1991 akkreditiertes Prüflaboratorium (Reg.- Nr. DAT-P-006 und DAT-P-012) unabhängig im Hause HBM durch. Darmstadt, 10.05.96 First attachment of the CE mark: 1995 All product-related features are secured by a quality system in accordance with DIN ISO 9001, certified by DQS (Deutsche Gesellschaft zur Zertifizierung von Qualitätsmanagementsystemen) since 1986 (Reg.-No. DQS-10001). The safety-relevant features (electromagnetic compatibility, safety of electrical apparatus) are verified at HBM by an independent resting laboratory which has been accredited by DATech in 1991 for the first time (Reg Nos. DAT-P-006 and DAT-P-012). Première application de la marque CE: 1995 Chez HBM, la détermination de tous les critères de qualité relatifs à un produit spécifique est faire sur la base d un protocole DQS (Deutsche Gesellschaft zur Zertifizierung von Qualitätsmanagementsystemem) certifiant, depuis 1986, notre système d assurance qualité selon DIN ISO 9001 (Reg.-No. DQS-10001). De même, tous les critères de protection électrique et de compatibilité électromagnetique sont certifiès par un laboratoire d essais indépendant et accrèdiré depuis 1991 (Reg.-No. DAT-P-006 et DAT-P-012.)

U9B 17 Seite 2 zu Page 2 of Page 2 du Document: 065/05.1996 Diese Erklärung bescheinigt die Übereinstimmung mit den genannten Richtlinien, beinhaltet jedoch keine Zusicherung von Eigenschaften. Die Sicherheitshinweise der mitgelieferte Produktdokumentation sind zu beachten. Folgende Normen werden zum Nachweis der Übereinstimmung mit den Vorschriften der Richtlinie(n) eingehalten. This declaration certifies conformity which the Directives listed above, but is no asservation of characteristics. Safety directions of the delivered product documentation have to be followed. The following standards are fulfilled as proof of conformity with the provisions of the Directive(s) Cette déclaration atteste al conformité avec les directives citées mais n assure pas un certain charaktère. S.v.p. observez les indications de sécurité de la documentation du produit ajoutée. Pour la demonstration de la conformité aux disposition de(s) Directive(s) le produit satisfait les normes: EN 50082-2: 1995 Elektromagnetische Verträglichkeit (EMV); Fachgrundnorm Störfestigkeit; Teil 2: Industriebereich; Deutsche Fassung

18 U9B

U9B 19

Modifications reserved. All details describe our products in general form only.they are not to be understood as express warranty and do not constitute any liability whatsoever. Hottinger Baldwin Messtechnik GmbH B 20. U9B.40 en Postfach 10 01 51, D-64201 Darmstadt Im Tiefen See 45, D-64293 Darmstadt Tel.: +49/61 51/ 8 03-0; Fax: +49/61 51/ 8039100 E mail: support@hbm.com www.hbm.com