Safelight Fog does what to contrast and density on film?

Similar documents
Nuclear Associates

Nuclear Associates

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Nuclear Associates

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Beam-Restricting Devices

LECTURE 1 The Radiographic Image

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

Mammography is a radiographic procedure specially designed for detecting breast pathology Approximately 1 woman in 8 will develop breast cancer over

Compliance Guidance for RADIOGRAPHIC QUALITY CONTROL (5 th Edition)

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Nuclear Associates

RADIOGRAPHIC EXPOSURE

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

MAMMOGRAPHY - HIGH LEVEL TROUBLESHOOTING

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

1. Carlton, Richard R., and Arlene M. Adler. Principles of Radiographic Imaging: An Art and a Science, 5th edition (2013).

NJDEP Medical Physicist s Radiographic QC Survey Registration Number:

Essentials of Digital Imaging

10/15/2012 SECTION III - CHAPTER 6 DIGITAL FLUOROSCOPY RADT 3463 COMPUTERIZED IMAGING

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

A Practical Overview of the Clinical and Operational Impact of Computed Radiography(CR) Implementations. Shirley Weddle, RT(R)(M), CIIP, BBA

Quality Control of Full Field Digital Mammography Units

SECTION I - CHAPTER 1 DIGITAL RADIOGRAPHY: AN OVERVIEW OF THE TEXT. Exam Content Specifications 8/22/2012 RADT 3463 COMPUTERIZED IMAGING

Radiology. Radiograph: Is the image of an object made with use of X- ray instead of light.

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Digital Imaging started in the 1972 with Digital subtraction angiography Clinical digital imaging was employed from the 1980 ~ 37 years ago Amount of

Ansur TNT Users Manual. Plug-In

Implementation of the Theory of Terri L. Fauber on Various Tube Voltage (KV) and Current (mas) on Radiographic Density

Contrast. Contrast: the difference in density on adjacent areas of a radiograph or other image receptor. Subjective. Long Scale (Low Contrast)

X-RAY. Lecture No.4. Image Characteristics:

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

CHAPTER 6 QC Test For Fluoroscopic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

RAD 150 RADIOLOGIC EXPOSURE TECHNIQUE II

X-RAY IMAGING EE 472 F2017. Prof. Yasser Mostafa Kadah

Nuclear Associates EZ CR-DIN Phantoms

X-ray Tube and Generator Basic principles and construction

X-RAYS - NO UNAUTHORISED ENTRY

10/3/2012. Study Harder

10/26/2015. Study Harder

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

Learning Objectives: What s my motivation? (unknown screen actor) Workshop Overview

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

Century focus and test chart instructions

Exposure System Selection

Photomultiplier Tube

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

FMT18 FLOOR MOUNTED SYSTEM

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

DIAGNOSTIC ACCREDITATION PROGRAM. Radiology and CT Quality Control Procedures Workbook

Introduction. Digital Mammography QA: Comparing the Manufacturers Recommendations. What is QC and why is it important? Review & compare QC tests

KODAK Simulation Film / 7157

SPRINGFIELD TECHNICAL COMMUNITY COLLEGE ACADEMIC AFFAIRS

JEFFERSON COLLEGE. Radiographic Exposures

Digital radiography (DR) post processing techniques for pediatric radiology

X-ray Tube and Generator Basic principles and construction

P UBLIC Law passed by the Congress in 1969 and often referred to

- KiloVoltage. Technique 101: Getting Back to Basics

Do you have any other questions? Please call us at (Toll Free) or , or

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

P R E S E N T E D B Y. K A M A R U L A M I N A B D U L L A H Dip. MED. IMG., BSc. MED. IMG. (UiTM)

Overview of Safety Code 35

Density vs. Contrast

(CR) PORTABLE CHEST EXAM Baseline Definition Worksheet

Acceptance Testing of a Digital Breast Tomosynthesis Unit

RADIOGRAPHIC EXAMINATION PROCEDURE

Dedicated Veterinary Imaging Solutions Digital, CR and Analog Imaging Solutions for any size patient and any size budget.

NOTES ON PRINTING 08/06

Outline. Digital Radiography. Understanding Digital Modalities: Image Quality and Dose. Image Quality. Dose Control

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Digital Imaging Considerations Computed Radiography

Third Edition DECEMBER Collaborating institutes: QARAD, BE LUCK, BE ARCADES, FR LRCB, NL EMIFMA, BE NCCPM, UK. Contributors:

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Principle of X-Ray Systems

FOUR CATEGORIES OF SAFETY

Distributed by. Ecotron Anyview Series. The Intelligent C-Arm System

KODAK Medical X-ray Film / General Purpose / Blue / 3528 / MXB

Acquisition, Processing and Display

Essentials of Digital Imaging

C-ARM FLUOROSCOPIC AND SPOT-FILM SYSTEMS

KODAK MIN-R S Film / 4906

Visibility of Detail

Film Replacement in Radiographic Weld Inspection The New ISO Standard

FFDM in the Field: Physicist's Role in the QC of Mammography Laser Printers May Carl R. Keener, Ph.D., DABMP, DABR

7/20/2014. Outline. Outline. Disclosures. Learning Objectives. SBB: Practical Aspects of ACR Accreditation, QC and ACR On Site Surveys

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

Outline ASRT Changes Impact on current curriculum Potential new courses WECM Changes Last update Resources and needs

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

X-RAY FLUOROSCOPY IMAGING SYSTEMS. Dr Slavik Tabakov. Luminescence: Dept. Medical Eng. & Physics King s College London

Enhanced Functionality of High-Speed Image Processing Engine SUREengine PRO. Sharpness (spatial resolution) Graininess (noise intensity)

NEXT Protocol for the Survey of Pediatric Chest Radiography

Transcription:

Terri Jurkiewicz

Safelight Fog does what to contrast and density on film?

ANSWER INCREASES DENSITY DECREASES CONTRAST

Explain how you determine if the focal spot size is within appropriate limits. <.8 =50%.8-1.5 = 40% > 1.6 = 30%

ANSWER YOU MUST HAVE THE MANUFACTURERS FOCAL SPOT SIZES. FOR OUR EQUIPMENT THE FOCAL SPOT SIZES WERE:.6 SMALL FOCAL SPOT 1.5 LARGE FOCAL SPOT ADD X+Y AND DETERMINE THE ENLARGEMENT FACTOR WHICH IS 2 FOR FS LESS THAN 2.5 (THE MAJORITY OF FS WILL BE LESS THAN 2.5) DIVIDE BY THE ENLARGEMENT FACTOR AND COMPARE TO MANUFACTURERS FOCAL SPOT SIZE.

When running a test strip for sensitometry, if when it is being plotted it falls above the upper or lower limits, what does that mean you should do?

ANSWER RUN TEST STRIP AGAIN. IF IT REMAINS OUT OF LIMITS. CONTACT SERVICE ENGINEER

17. What are the densities used to set up a control chart?

ANSWER BASE + FOG NEVER MORE THAN.25 SHOULD BE BETWEEN.1-.2 MINIMUM DENSITY IS THE STEP CLOSEST TO.25 MAXIMUM DENSITY IS STEP CLOSEST TO 2.0 ABOVE B+F SPEED INDICATOR IS TH MID DENSITY STEP CLOSEST TO 1. ABOVE THE BASE + FOG. SO IF B+F IS.1 THEN THE STEP CLOSEST TO 1.1 IS THE SPEED INDICATOR. CONTRAST INDICATOR IS THE DIFFERENCE BETWEEN THE MAXIMUM AND MINIMUM VALUES THIS SHOULD NOT VARY BY MORE THAN.15

20. WHAT IS RECIPROCITY AND How many exposures should be made to determine reciprocity?

ANSWER NO MATTER WHAT THE MA/TIME SETTINGS ARE AS LONG AS THE MAS STAYS THE SAME WITH THE SAME KVP YOUR EXPOSURE SHOULD ALWAYS BE THE SAME WITH MINIMUM OF 3

THEN WHAT IS REPRODUCIBILITY?

ANSWER IT IS TAKING IMAGES WITH THE SAME TECHNIQUE TO ENSURE THE TECHNIQUE REMAINS THE SAME WITH EACH EXPOSURE WITH NO MORE THAN 5% VARIABILITY

WHAT IS MA/TIME LINEARITY?

ANSWER SEQUENTIAL INCREASES OR DECREASES IN MAS SHOULD PRODUCE THE SAME SEQUENTIAL INCREASE IN EXPOSURE MEASURED.

If 50 radiographs are rejected due to positioning errors and there are 250 radiographs rejected in a month, what percentage is due to positioning errors?

ANSWER 50/250=.2 OR 20%

When performing safelight fog test, one half of the film must be exposed because??

ANSWER LATENSIFICATION FILM IS MUCH MORE SENSITIVE TO LIGHT EXPOSURE AFTER EXPOSURE TO RADIATION.

WHAT IS VOLTAGE RIPPLE?

ANSWER IT IS A GRAPH OF THE VOLTAGE WAVEFORM GIVEN OFF FROM EACH GENERATOR

What is the ripple effect in the waveform of a high frequency generator? What of a 3 phase 12 pulse generator?

ANSWER 1-15 % High Frequency generator 3.5% Three phase 12 pulse generator

What is the name of this object?

ANSWER PENTROMETER

Electronic transmission of radiographic images from one location to another is what?

ANSWER TELERADIOLOGY

Verification of equipment on installation is called?

ANSWER ACCEPTANCE TESTING

What is veiling glare and what causes it?

ANSWER IT IS GLARE OR FLARE CAUSED BY LIGHT REFLECTION FROM THE WINDOW OF THE OUTPUT PHOSPHOR. REDUCING CONTRAST. YOU SEE THIS IN FLUORO. WHEN MOVING THE UNIT FROM AREA OF THE BODY TO ANOTHER.

How often and how long should QC of a flat panel display take by QC technologist?

ANSWER 1 MINUTE DAILY

What might cause and increased base + fog?

Repeat rates should have at least patients as a sample for valid results.

ANSWER 250 PATIENTS

Your repeat rate is 6.4 % what does this mean?

ANSWER YOU ARE SLIGHTLY HIGHER THAN THE APPROVED REPEAT RATE. BUT, RATES 10-12% SHOULD BE SERIOUSLY EXAMINED AS THOSE ARE WAY TOO HIGH COSTING DEPARTMENTS MONEY AND EXTRA RADIATION EXPOSURE TO PATIENTS.

Why do you want to do a repeat/reject analysis?

ANSWER IT ALLOWS THE DEPARTMENT TO ANALYSE WHERE THE FACTORS WHICH CONTRIBUTE TO THE OVERALL REPEAT RATE. IF CORRECTIVE ACTION NEEDS TO BE TAKEN, THEN THE KNOWLEDGE OF WHERE THE ISSUES ARE ARE EVIDENT.

WHAT IS THE ACCURACY OF THE COLLIMATION %? PLEASE DISCUSS PERPENDICULATIRY AND COLLIMATION.

ANSWER COLLIMATION MUST BE WITHIN 2% OF SID WHICH IS THE WIDTH OF A PENNY AT 40 INCHES. PERPENDICULARITY IS REQUIRED TO LIMIT DISTORTION AND SHOULD BE WITHIN 1 DEGREE.

WHAT DOES POOR FILM SCREEN CONTACT PROJECT ON AN IMAGE. WHAT IS THE TEST TOOL THAT IS USED TO DEMONSTRATE POOR CONTACT.

ANSWER TOOL: WIRE MESH TEST TOOL IMAGE IS DARKER AND BLURRED.

HOW OFTEN SHOULD A FILM SCREEN CASSETTE BE CLEANED?

ANSWER WEEKLY

Repeat/Reject Analysis What percent of repeats are acceptable in a Radiology Department? At what point should a Radiology department take a serious look at their repeats and make some changes?