SIGNAL CONDITIONING FOR CRYOGENIC THERMOMETRY IN THE LHC

Similar documents
CERN (The European Laboratory for Particle Physics)

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems

Residual Resistivity Ratio (RRR) Measurements of LHC Superconducting NbTi Cable Strands

Special-Purpose Operational Amplifier Circuits

Measurement of SQUID noise levels for SuperCDMS SNOLAB detectors

Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

6. Bipolar Diode. Owing to this one-direction conductance, current-voltage characteristic of p-n diode has a rectifying shape shown in Fig. 2.

About the Tutorial. Audience. Prerequisites. Copyright & Disclaimer. Linear Integrated Circuits Applications

Chapter IX Using Calibration and Temperature Compensation to improve RF Power Detector Accuracy By Carlos Calvo and Anthony Mazzei

Operational amplifiers

Radiation Test Report Paul Scherer Institute Proton Irradiation Facility

LINEAR IC APPLICATIONS

Scheme I Sample Question Paper

CHAPTER 6 DIGITAL INSTRUMENTS

THE BENEFITS OF DSP LOCK-IN AMPLIFIERS

Interface Electronic Circuits

Using Optical Isolation Amplifiers in Power Inverters for Voltage, Current and Temperature Sensing

SENSOR AND MEASUREMENT EXPERIMENTS

Section 4: Operational Amplifiers

Operational Amplifiers

Analog Electronics. Lecture. Op-amp Circuits and Active Filters. Muhammad Amir Yousaf

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH European Laboratory for Particle Physics

Thermocouple Conditioner and Setpoint Controller AD596*/AD597*

Application Note (A12)

Physics 303 Fall Module 4: The Operational Amplifier

Phase Control Circuit Tacho Applications. Automatic retriggering. Phase control unit = f (V 12 ) Soft start 11(11) 12(12) 8(8) 7(7)

WebSeminar: Signal Chain Overview

AN-742 APPLICATION NOTE

ELECTRONIC SYSTEMS FOR THE PROTECTION OF SUPERCONDUCTING ELEMENTS IN THE LHC

RAZTEC LINK CURRENT SENSOR TECHNICAL INFORMATION

MAE334 - Introduction to Instrumentation and Computers. Final Exam. December 11, 2006

AC Resistance Thermometry Bridges and their Advantages By Peter Andrews

AC/DC Current Probe CT6844/CT6845/CT6846

Keywords: ISM, RF, transmitter, short-range, RFIC, switching power amplifier, ETSI

From Narrow to Wide Band Normalization for Orbit and Trajectory Measurements

AD596/AD597 SPECIFICATIONS +60 C and V S = 10 V, Type J (AD596), Type K (AD597) Thermocouple,

EE320L Electronics I. Laboratory. Laboratory Exercise #2. Basic Op-Amp Circuits. Angsuman Roy. Department of Electrical and Computer Engineering

Module 1: Introduction to Experimental Techniques Lecture 2: Sources of error. The Lecture Contains: Sources of Error in Measurement

ELECTRICAL CIRCUITS 4. OPERATIONAL AMPLIFIERS INPUT/OUTPUT CHARACTERISTICS

TL494M PULSE-WIDTH-MODULATION CONTROL CIRCUIT

Lecture 6: Digital/Analog Techniques

The simplest DAC can be constructed using a number of resistors with binary weighted values. X[3:0] is the 4-bit digital value to be converter to an

MECE 3320 Measurements & Instrumentation. Data Acquisition

Single-channel power supply monitor with remote temperature sense, Part 1

Semiconductor Detector Systems

William Thomson, Lord Kelvin, CAS2004. High Precision Measurements - Gunnar Fernqvist/CERN 1

LM231A/LM231/LM331A/LM331 Precision Voltage-to-Frequency Converters

Input Limiter for ADCs

Chapter X Measuring VSWR and Gain in Wireless Systems By Eamon Nash

MEMS Optical Scanner "ECO SCAN" Application Notes. Ver.0

Signal Characteristics and Conditioning

TL594 PULSE-WIDTH-MODULATION CONTROL CIRCUITS

AUTOMOTIVE CURRENT SENSOR DHAB S/25. Datasheet

Analog I/O. ECE 153B Sensor & Peripheral Interface Design Winter 2016

EE 210 Lab Exercise #5: OP-AMPS I

Testing Power Sources for Stability

Concepts to be Reviewed

A new hybrid protection system for high-field superconducting magnets

ANALOG TO DIGITAL CONVERTER ANALOG INPUT

Introduction to Measurement Systems

Isolated, Linearized RTD Input 7B34 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Analytical Chemistry II

How to Monitor Sensor Health with Instrumentation Amplifiers

Unit WorkBook 1 Level 4 ENG U22 Electronic Circuits and Devices 2018 UniCourse Ltd. All Rights Reserved. Sample

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil


ELT 215 Operational Amplifiers (LECTURE) Chapter 5

AUTOMOTIVE CURRENT TRANSDUCER HC6F600-S

Technical Brief FAQ (FREQUENCLY ASKED QUESTIONS) For further information, please contact Crystal Semiconductor at (512) or 1 (800)

MAX6675. Cold-Junction-Compensated K-Thermocoupleto-Digital Converter (0 C to C) Features

TL494 PULSE-WIDTH-MODULATION CONTROL CIRCUITS

Dev Bhoomi Institute Of Technology Department of Electronics and Communication Engineering PRACTICAL INSTRUCTION SHEET REV. NO. : REV.

TL594C, TL594I, TL594Y PULSE-WIDTH-MODULATION CONTROL CIRCUITS

AUTOMOTIVE CURRENT TRANSDUCER HC6F700-S

Dimensions in inches (mm) .268 (6.81).255 (6.48) .390 (9.91).379 (9.63) .045 (1.14).030 (.76) 4 Typ. Figure 1. Typical application circuit.

saac ewton roup ed maging etector

Practice questions for BIOEN 316 Quiz 4 Solutions for questions from 2011 and 2012 are posted with their respective quizzes.

User s Manual for Integrator Short Pulse ISP16 10JUN2016

INSTRUMENTATION AND CONTROL SYSTEM FOR THE INTERNATIONAL ERL CRYOMODULE

HC2F100-SN CLIPS AUTOMOTIVE CURRENT TRANSDUCER HC2F100-SN CLIPS. Datasheet

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

When input, output and feedback voltages are all symmetric bipolar signals with respect to ground, no biasing is required.

Using Magnetic Sensors for Absolute Position Detection and Feedback. Kevin Claycomb University of Evansville

Experiment 2: Transients and Oscillations in RLC Circuits

Scintillators as an external trigger for cathode strip chambers

CAV444 C/V transmitter IC with adjustable output voltage for capacitive input signals

IT.MLD900 SENSORS AND TRANSDUCERS TRAINER. Signal Conditioning

Cold-Junction-Compensated K-Thermocoupleto-Digital Converter (0 C to +128 C)

AUTOMOTIVE CURRENT SENSOR DHAB S/18. Datasheet

Proposal for instrumentation to calibrate DCCT s up to 24 ka

This Errata Sheet contains corrections or changes made after the publication of this manual.

Fundamentals of Data Converters. DAVID KRESS Director of Technical Marketing

LM13600 Dual Operational Transconductance Amplifiers with Linearizing Diodes and Buffers

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

llillllwlllll 1! lllllllllllllilllilllll PBE1BE158S

Chapter 2 Analog-to-Digital Conversion...

Logarithmic Circuits

Chapter 5: Signal conversion

Precision amplifier for bridge circuits AM467 PRINCIPLE FUNCTION

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HAH3DR 1100-S07

Transcription:

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH European Laboratory for Particle Physics Large Hadron Collider Project LHC Project Report 333 SIGNAL CONDITIONING FOR CRYOGENIC THERMOMETRY IN THE LHC J. Casas, P. Gomes, K.N. Henrichsen, U. Jordung and M.A. Rodriguez Ruiz Abstract Temperature measurement is a key issue in the Large Hadron Collider (LHC), as it will be used to regulate the cooling of the superconducting magnets. The compromise between available cooling power and the coil superconducting characteristics leads to a restricted temperature control band, around.9 K. An absolute accuracy of mk below. K, and 5 K above 5 K, is necessary. For resistive thermometers covering the full temperature range, and having a negative dr/dt sensitivity, this is typically equivalent to a relative accuracy DR/R of 3-3 over 3 resistance decades. Also, to limit the thermometer's self-heating, the sensing current must be limited to few ma. Furthermore, the radiation levels next to the accelerator are expected to significantly degrade the performance of conventional analog electronics. As these stringent requirements are not met by commercial conditioners, three different architectures have been developed at CERN. The first compresses the input dynamic range using a logarithmic transfer function; the second partitions the input range into three linear regions; the third converts resistance linearly into the frequency of a square wave. They fulfill the above specifications and provide industrial robustness in terms of thermal drift, galvanic protection, and compact packaging, while optimizing cost-to-performance ratio. This paper describes the principles of their design, compares their characteristics and shows results of field tests. Future developments include Application Specific Integrated Circuit versions, Fieldbus interfacing, and radiation tolerant re-design. LHC Division Presented at the 999 Cryogenic Engineering and International Cryogenic Materials Conference (CEC-ICMC'99), -6 July 999, Montreal, Canada Administrative Secretariat LHC Division CERN CH - Geneva 3 Switzerland Geneva, December 999

SIGNAL CONDITIONING FOR CRYOGENIC THERMOMETRY IN THE LHC J. Casas, P. Gomes, K.N. Henrichsen, U. Jordung, M.A. Rodriguez Ruiz CERN, European Organization for Nuclear Research Geneva 3, Switzerland ABSTRACT Temperature measurement is a key issue in the Large Hadron Collider (LHC), as it will be used to regulate the cooling of the superconducting magnets. The compromise between available cooling power and the coil superconducting characteristics leads to a restricted temperature control band, around.9 K. An absolute accuracy of mk below. K, and 5 K above 5 K, is necessary. For resistive thermometers covering the full temperature range, and having a negative dr/dt sensitivity, this is typically equivalent to a relative accuracy 'R/R of 3-3 over 3 resistance decades. Also, to limit the thermometer s self-heating, the sensing current must be limited to few PA. Furthermore, the radiation levels next to the accelerator are expected to significantly degrade the performance of conventional analog electronics. As these stringent requirements are not met by commercial conditioners, three different architectures have been developed at CERN. The first compresses the input dynamic range using a logarithmic transfer function; the second partitions the input range into three linear regions; the third converts resistance linearly into the frequency of a square wave. They fulfill the above specifications and provide industrial robustness in terms of thermal drift, galvanic protection, and compact packaging, while optimizing cost-toperformance ratio. This paper describes the principles of their design, compares their characteristics and shows results of field tests. Future developments include Application Specific Integrated Circuit versions, Fieldbus interfacing, and radiation tolerant re-design. INTRODUCTION The various components of the LHC cryogenic system work at temperatures from ambient down to.6 K. Depending on the actual temperature value, different accuracies are on its measurement. Between 3 K and 5 K, an uncertainty of 5 K can be tolerated to monitor the warmer components and the general cool-down. However, at the nominal operation of superconducting magnets (below. K) only mk inaccuracy is allowed, to give enough room for the regulation band of the cryogenic controller, while avoiding magnet quench and minimizing the cooling effort of the cryogenic system. Table shows the allowed uncertainty on temperature measurement on the LHC machine, for the different temperature ranges. The accuracy budget is to be evenly shared between the sensor and the signal conditioning (Figure ). The aimed resolution has to be ten times better than the overall accuracy (dt < mk, below. K).

Table. Required overall T accuracy and resolution T span [K] accuracy [mk] resolution [mk].6 l.. l 4. 4. l 6. 3 3 6. l 5 5 l 3 5 5 T [mk] Sensors Figure. Required conditioning temperature accuracy Cryogenic temperature sensors currently used at CERN can be classified according to four main attributes, as shown in Table. CERNOX (CX), TVO and RhFe cover the full temperature range with a single sensor. AllenBradley (AB) and Pt can be combined to cover respectively low and high temperature scales, or used alone in applications not requiring full range measurements. In terms of resistive values, CX s span is the largest among all sensors (3 decades), requiring wide dynamic range signal conditioning. Also covering the whole temperature range, RhFe spans over only decades of resistance, with the advantage of less demanding dynamic range, but with the consequence of limited sensitivity. Sensors with negative dr/dt (Figure ), like CX, TVO and AB, show high resistance and high sensitivity (dr/r / dt/t) (Figure 3) at low temperatures, where measurement accuracy has to be at its best. This semiconductor behavior relaxes the constraints on conditioner accuracy for low temperature measurement. On the other hand, at low temperature metallic sensors like RhFe exhibit a sensitivity one order of magnitude worse, demanding much more accurate signal conditioning. Signal conditioning The basic principle used to read resistive sensors consists in sending a known sensing current, over a pair of wires, and reading the voltage developed at the resistor leads, via another pair of wires. After amplification and correction, the read signal can be sent to the process controller under different formats, either analog or digital. Table. Typical characteristics of cryogenic temperature sensors currently used at CERN T span [K] R span [:] dr/dt [:/K] (dr/r) / (dt/t) CX.6 l 3 3 l 3-4 l -. -.7 l -. TVO.6 l 3 9 l 9-7 l -.7 -.3 l -. RhFe.6 l 3 6 l +.7 l +.4 +. l +. AB.6 l l - l -.3-3. l -. Pt 73 l 3 8 l +.4 l +.4 +. l +. R [ ] CERNOX TVO RhFe AB Pt (dr/r) / (dt/t).. CERNOX TVO RhFe AB Pt. Figure. Typical thermometers transfer function R(T) Figure 3. Dimensionless sensititvity vs temperature

Table 3. Conditioner s accuracy and resolution by each type of sensor, below and above T = 6 K T < 6 K T > 6 K sensor R span [:] 'R/R [ -3 ] dr [m:] R span [:] 'R/R [ -3 ] dr [m:] CX 5 l 3 3.3 9 3 l 5 8.9 6 TVO 6 l 9.3 7 9 l 6.8 3 RhFe 6 l 8.5.5.6 8.5 l 6. 5 AB 5 l 3. 3 l 5 5.8 Pt 8 l 9.8.... CERNOX TVO RhFe AB Pt requirement dr [ ]..... CERNOX TVO RhFe AB Pt.. Figure 4. Required accuracy Figure 5. Required resolution For each type of thermometer, the signal conditioning accuracy and resolution requirements are summarized in Table 3 and plotted in Figure 4 and Figure 5. The RhFe sensor imposes quite severe constraints on the conditioning electronics, with 'R/R <.5-3 for R < 8.5:, thus requiring techniques beyond the scope of the conditioners described in this paper. A conditioner satisfying 'R/R < 3-3 for R > 5: and 'R/R < 6-3 for R < 5: is adequate for CX, AB and Pt thermometers. A TVO sensor would impose a three times better accuracy on the resistance measurement. As the sensor s resistance will be calculated linearly from the read voltage (Eq. ), the relative accuracy is the addition of the voltage and the sensing current uncertainties. R sensor = U read / I sense 'R/R = 'U/U + 'I/I () The read signal has to be digitized in order to be usable by computerized control or diagnostics. Industrial control equipment typically employs -bit Analog to Digital Converters (ADC) and the targeted temperature accuracy cannot be satisfied unless some type of signal compression is implemented. This paper presents two compression architectures that are mixed logarithmic+linear (LOG+LIN) and linear piecewise (LIN pw). The resistance measurement accuracy must be split between the accuracy of the analog circuitry and that of the ADC, which can be expressed in terms of the ADC Least Significant Bit (LSB). Figure 6 and Figure 7 show respectively the accuracy and resolution imposed by the resistance measurement, and also the accuracy and resolution attained by the combination of different compression algorithms and a minimum size ADC..... LSB/R (R) 8b LIN b LOG 3b LOG+LIN 3b LIN pw dr [ ]..... 9b LIN 4b LOG 5b LOG+LIN 5b LIN pw LSB (R).. Figure 6. Required ADC accuracy Figure 7. Required ADC resolution 3

Table 4. ADC accuracy and resolution for each type of conditioner transfer function for CX, AB, Pt and TVO for CX, AB, and Pt only conditioner accuracy [bit] resolution [bit] accuracy [bit] resolution [bit] LIN 8 9 8 9 LOG 3 5 4 LOG+LIN 4 6 3 5 LIN pw 4 6 3 5 Table 4 summarizes these requirements. Due to its lower sensitivity, the TVO typically needs more bit than other sensors. For all sensors, the resolution normally requires more bits than the accuracy. It is clear that a -bit resolution ADC is not sufficiently accurate (above 4.5 K) for all the mentioned conditioning architectures. We are thus considering the use of larger ADC, with a suitable network interface. LOGARITHMIC CONDITIONER The first conditioner (Figure 8) incorporates a logarithmic compression of the input dynamic range. A DC sensing current of PA develops a voltage (. mv l mv) across the thermometer, which is raised by an instrumentation amplifier. The offset and gain, before and after the logarithmic amplifier, have to be manually adjusted. Power supply, input and output are galvanically isolated to 75 V. The overall performance against ambient temperature variation is limited by the Logarithmic IC. This explains the improvement in temperature drift at higher resistances, for LOG+LIN in comparison with LOG (Figure 9 and Figure ). The relationship between input and output is given by the following equations for LOG and LOG+LIN respectively: I loop = 4 + 6/3 Log (R/) [ma] () I loop = 4 + 8/3 Log (R/) + 8/(k-) (R-) [ma] (3) The rejection of power supply drift is good for R > 5 : ( 'R/R /'V <.7-3 /V, Figure ), but quite poor for small R. Several tens of these conditioners have been installed in cryogenic experiments at CERN, yielding good results (Figure ), provided the ambient temperature does not change more than a few degrees. Figure 8. Logarithmic conditioner block diagram 5 5 Tamb drift, referred to 8C (LOG) T = 8C T = 8C T = 8C T = 38C T = 48C T = 58C 5 5 Tamb drift, referred to 3C (LOG+LIN) T 3C T 3C T 3C T 34C T 45C T 54C -5-5 -5 R[:] Figure 9. Thermal drift (LOG function alone) -5 Figure. Thermal drift (combined LOG+LIN) 4

5 5-5 -5 Vcc drift, referred to 4V (LOG+LIN) Vcc = V Vcc = 4V Vcc = 6V Figure. Power supply drift.. DT [mk]..... accuracy Figure. Field measurements TT TT TT TT4 TT5 TT6 TT7 TT8 TT9 TT LINEAR MULTI-RANGE CONDITIONER In order to reduce the ADC size, a linear multi-range conditioner has been developed. The input span is partitioned into three regions (Table 5), one decade wide each, with the gain proportional to the sensing current. The thermal drift due to the logarithmic IC is thus avoided, at the expense of more complex circuitry for controlling the gain (Figure 3). Furthermore, the sensing current can be lower for high R values (low T), reducing the self-heating of the sensor, and higher for small R values, increasing the voltage developed at the sensor and its signal to noise ratio. The thermocouple voltages are cancelled by the use of a bipolar sensing current, oscillating at 4 Hz. The sensing current creates a voltage (r3.5 mv l r4 mv) across the thermometer, which is increased, by an instrumentation amplifier, (to r.875 V l r V), and then rectified and smoothed. When this voltage raises above V, the gain controller selects a sensing current times lower, leading to the same reduction in voltage. If the voltage drops below.875 V, a times higher current is selected. The gap between.875 V and V corresponds to a hysteresis in the transfer function (Figure 4), that prevents oscillation between consecutive ranges when close to the switching point. The selected range is indicated by two bit or a three level analog signal. Manual adjustments are necessary to correct general gain and offset and differential offset of the rectifier. Power supply, input and output are galvanically isolated to 75 V. Figure 3. Linear multi-range conditioner block diagram hysteresis (at Vout) Table 5. Sensing current and conditioner transfer function, for each input range R span [:] I sense [PA] I loop [ma] 35 l 4 4 + 6 R / 4 35 l 4 4 + 6 R / 4 3 5 l 4 4 + 6 R / 4 U [V] Figure 4. Hysteresis in the transfer function 5

The sensitivity to thermal drift ( 'R/R /'T amb <. -3 /qc, Figure 5) is very low compared to LOG conditioner. The thermal performance is further improved (Figure 6) at the voltage output, without galvanic isolation. The insensitivity to power supply drift is good for the whole input span, both at the current output ( 'R/R /'V <. -3 /V, Figure 7) and the voltage output ( 'R/R /'V <.4-3 /V, Figure 8). Figure 9 shows the excellent reproducibility of the conditioners characteristics between three samples. Several tens of these conditioners have been installed in cryogenic experiments at CERN, behaving in accordance with the requirements (Figure ). R/R [ e-3] - Tamb drift, referred to 5C (at Iout) T = C T = C T = 5C T = 4C T = 5C - Tamb drift, referred to 5C (at Vout) T = C T = C T = 5C T = 4C T = 5C - Figure 5. Thermal drift at the current output - Figure 6. Thermal drift at the voltage output Vcc drift, referred to 4V (at Iout) Vcc = 5V Vcc = 4V Vcc = 3V Vcc drift, referred to 4V (at Vout) Vcc = 5V Vcc = 4V Vcc = 3V - - - Figure 7. Power supply drift at current output - Figure 8. Power supply drift at voltage output - - exchangeability (at Vout) SN SN SN 3 Figure 9. Exchangeability.. T [mk].... accuracy (at Vout) Figure. Field measurements TT TT TT3 TT4 TT5 TT6 TT7 TT8 TT9 6

Figure. TF conditioner block diagram TF CONDITIONER This conditioner converts the voltage across the sensor resistance linearly into a frequency (Figure ). The requirements on accuracy and resolution are no longer put on voltage amplitude measurement but rather on frequency measurement. The output frequency spans over three decades of dynamic range, like the measured resistance. The logarithmic compression nuisances or the complexity of a multi-range controller are traded by a simpler concept. The sensing current follows a ramp of fixed slope (r PA/s). Once reached the upper ( mv) or lower (- mv) threshold referred to the input, the ramp sign is inverted. This fixed voltage amplitude restricts the thermometer self-heating to a relatively low value. Despite the fact that the working frequencies are below 5 Hz, the current control loop easily picks-up the mains noise. This effect is tackled by the use of a 5 Hz notch filter. In order to avoid spurious trigger of the flip-flop, a glitch detector is implemented, which introduces an extra delay on the oscillation period (Eq. 4). f = / ( -3 + 3 /R) [Hz] (4) Manual adjustments are necessary for notch filter width and center frequency, for the glitch filter delay and for the current slope. Figure shows the deviation ( 'R/R < -3 ) between two conditioners, using the same transfer function. The accuracy obtained in field measurements (Figure 3) is well within the requirements. Further tests are under way to investigate temperature and power supply drifts. - exchangeability SN... accuracy.8 K.8 K 3. K K - Figure. Exchangeability relative to SN. Figure 3. Field measurements 7

CONCLUSIONS The cryogenic thermometer signal conditioners presented in this paper are used in a variety of applications. When fast response is the LOG or LOG+LIN conditioner can be used. For higher accuracy at low temperature it would be necessary to reduce the sensing current. However, in this case the thermal drift of the input amplifier will be the dominant source of error for low values of the thermometric resistance. The LINpw and TF conditioners satisfy the LHC accuracy requirements. However their installation inside the LHC tunnel impose their redesign by using radiation tolerant integrated components, both passive and active. This problem is being solved by performing irradiation tests on commercial devices, and by porting the conditioners design into a radiation hardened Application Specific Integrated Circuit (ASIC). Once the design of a suitable signal conditioner for thermometers has been finalized, it will be necessary to adapt its characteristics in order to being able to use it as front end of other instruments like pressure sensors, liquid helium level gauges, etc. The cryogenic control system requires temperature measurements distributed around the 7 km circumference LHC machine. The data exchange can be done with either pointto-point analog signal transmission or by using an industrial field network. Analog transmission implies a large investment in cabling and installation. In order to reduce this cost we are also investigating the performance of network components and ADCs, to integrate them together with the signal conditioners. 8