Memosens CPS96D. Technical Information

Similar documents
Technical Information Memosens CPS96D

Technical Information Orbipore CPS91D and CPS91

Technical Information Orbipac CPF81D/CPF82D and CPF81/CPF82

Ceramax CPS341D. Technical Information

Orbipac CPF81/CPF82 and CPF81D/CPF82D

Condumax CLS21 and CLS21D

Technical Information Condumax CLS21D/CLS21

Condumax CLS15 and CLS15D

Technical Information Condumax CLS15D/CLS15

Condumax W CLS15 and CLS15D

Memosens New technology for contactless, inductive, digital transmission of ph signals

Technical Information Memosens CLS82D

Condumax CLS15 and CLS15D

Condumax CLS21 and CLS21D

Tophit CPS491 and CPS491D

Condumax CLS16 and CLS16D

Condumax W CLS15 and CLS15D

Indumax CLS50/CLS50D. Technical Information

Technical Information Indumax CLS50D/CLS50

Technical Information Memosens CCS50D

Tophit CPS441 and CPS441D

Tophit CPS491 and CPS491D

Turbimax CUS71D. Technical Information. Ultrasonic interface sensor Immersion sensor for interface measurement

Liquisys M CPM223/253

ph or ORP Transmitter

Indumax CLS54D. Technical Information

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

OPTISENS ORP 8590 Technical Datasheet

IsFET sensor for ph Measurement TopHit CPS 491

Technical Information Smartec CLD18

OPTISENS PH 8390 Technical Datasheet

ph / ORP Measurement mypro CPM 431

Technical Information Smartec CLD18

SMARTPAT PH 1590 Technical Datasheet

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

Digital ORP sensor for the industrial wastewater industry

Bulb ph and ORP Electrodes

Conductive conductivity measurement mypro CLM 431 / CLD 431 conductive

OPTISENS PH 8100 Technical Datasheet

Sterilizable IsFET sensor for ph Measurement TopHit CPS 441

Technical Information Smartec CLD18

Signet ph/orp Electrodes General Purpose

Signet ph/orp Electrodes

Model 50 Ground Loop ph Sensor Model 55 Solution Ground ph Sensor

OPTISENS PH ph sensor. Technical Datasheet

SMARTPAT ORP 8150 Technical Datasheet

ABB ph/orp analyzer and probe package

Measuring and Controlling Unit Multronic OC

Digital ORP sensor for the water and wastewater industry

Conductivity Transmitter

SMARTPAT ORP 8510 Technical Datasheet

Technical Information OUSBT66

ph/orp Measurement OPM223/253

Inductive conductivity measurement mypro CLM 431 / CLD 431 inductive

Intrinsically safe pressure transmitter MBS 4201, MBS 4251, MBS 4701 and MBS 4751

B Operating Instructions V1.00/EN/

OEM radar transmitter, for aggressive media level measurement

Submersible pressure sensor For water and wastewater Model LW-1

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

The transducer component converts the measured signal and displays the actual value.

TUpH 396RVP and 398RVP Retractable ph/orp Sensors

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

SITRANS F flowmeters SITRANS F M. Flow sensor MAG /48 Siemens FI Overview

Technical Information Indumax CLS54

Flow Measurement SITRANS F M

Measuring and Controlling Unit Multronic

OPTIFLUX 5000 Technical Datasheet

Flow Measurement SITRANS F M. Flow sensor MAG /49. Overview

Submersible pressure sensor For oils and fuels Model LF-1

MJK Expert 700 / 800 / 900 / 1100 / 2100 MJK Expert 1400 / 3400

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

SALES SERVICE +358 (0)

OPTIFLUX 5000 Technical Datasheet

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

ph Measurement Good Operating Procedures for ph Sensors Mettler-Toledo AG Process Analytics Quality Measurements for In-Line Applications

Conductivity Sensor for High-Temperature Applications ConduMax W CLS 13

Remote Controlled FOUNDATION Fieldbus Two-Wire ph/orp Transmitter

Ultrasonic level measuring device, non-contact

C4E NUMERICAL SENSOR User manual

Functional Safety Manual Memosens Cable CYK10

SITRANS F flowmeters. SITRANS F System information MAGFLO electromagnetic flowmeters 4/9

Technical Data. General specifications. Rated operating distance s n 3 mm

Digital inductive conductivity transmitter

itemp HART DIN rail TMT112

Temperature transmitter itemp PCP DIN rail TMT 121

A Higher Level of Performance. Data Sheet. Series Pressure and Level Transmitters. For more information, please visit >

Submersible pressure sensor For superior applications Model LF-1

Pressure transmitter with flameproof enclosure For applications in explosion-protected areas Models E-10 and E-11

600T Series Transmitters. Specification Sheet. Model 653T Electronic temperature transmitter (PC programmable)

0 2 1 â Ó. itemp HART TMT182. Technical Information

Automation of ph/orp Measurements TopCal S CPC 300

oxy.iq Panametrics Oxygen Transmitter Applications Features bhge.com Two-wire, loop-powered 4 to 20 ma transmitter An oxygen transmitter for use in:

B Operating Instructions /

phix Compact 4.10 PH, TEMPERATURE AND REDOX MEASUREMENT DATASHEET EN 4.10 PHIX DATASHEET 1812

Conductivity measurement

Valve Series V2001 Globe Valve Type 3321 with Pneumatic or Electric Actuator

SMARTPAT COND 1200 Technical Datasheet

600T Series Transmitters Model 653L Electronic temperature transmitter (PC programmable for RTD and resistance) Specification Sheet

Process Protection Acoustic sensors for pump monitoring

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

Transcription:

Technical Information Memosens CPS96D Combination of ph and ORP electrode with Memosens technology With open aperture junction for very contaminated media and suspended solids Resistent to poisoning reference with ion trap Application Chemical processes Pulp and paper industry Contaminated media and suspended solids Dispersions Precipitation reactions Emulsions With ATEX and IEC approval for use in hazardous areas Your benefits Simultaneous measurement of ph, ORP and rh values (in rh mode) Platinum additional electrode to measure reference impedance Open aperture junction enables use in processes containing solids Low maintenance owing to firm, stabilized gel with good chemical resistance properties Not affected by fluctuations in pressure and temperature Quick response time Integrated temperature sensor for effective temperature compensation Long service life thanks to resistent to poisoning reference with ion trap Glass breakage and blockage detection by measuring: Resistance of glass membrane Reference impedance Additional advantages offered by Memosens technology Maximum process safety thanks to non-contact, inductive signal transmission Digital data transmission ensures data security Very easy to use as sensor-specific data are stored in the sensor Recording of sensor load data in the sensor enables predictive maintenance TI00507C/07/EN/01.12 71163968

Function and system design Measuring principle Measuring system ph measurement The ph value is a measure of the acid or base character of a medium. Depending on the ph value of the medium, the electrode's membrane glass provides an electrochemical potential. This is the result of H + ions selectively penetrating the outer layer of the membrane. As a result, an electrochemical boundary layer forms here with an electric potential. An integrated Ag/AgCl reference system forms the required reference electrode. The transmitter converts the measured voltage into the corresponding ph value according to the NERNST equation. ORP measurement The oxidation-reduction potential (ORP) is an indicator of the balance between oxidizing and reducing components in the medium. The ORP is measured using a platinum electrode. As in the case of ph measurement, the integrated Ag/AgCl reference system is the reference electrode used. rh measurement The rh value is defined as the negative logarithm of partial pressure of hydrogen in a solution. The ph value and ORP value of a solution must be measured simultaneously to calculate the rh value. The value is calculated using the following equation: rh = 2 x (mv/s) + 2 ph ph = measured ph value mv = measured ORP value in mv + 207 mv (Ag/AgCl system) S = slope of ph electrode The rh value is an indicator of the oxidation or reducing ability of a process solution. The measuring range is from 0 to 42. A process medium with rh values between 0 and 9 has a high reducing ability; a low reducing ability with rh 9-17; an indeterminate ability with rh 17-25; a low oxidation ability with rh 25-34, and a high oxidation ability with rh 34-42. A complete measuring system comprises: ph/orp combined electrode CPS96D Transmitter, e.g. Liquiline M CM42 Measuring cable CYK10 Immersion, flow or retractable assembly, e.g. Cleanfit P CPA473 Example of a measuring system 1 ph/orp combined electrode CPS96D 2 Transmitter Liquiline M CM42 3 Measuring cable CYK10 4 Cleanfit CPA473 retractable assembly a0017200 2 Endress+Hauser

Communication and data processing Reliability Communication with the transmitter Only connect digital Memosens sensors with a combined ph and ORP electrode to a transmitter with appropriate functionality and Memosens technology. These are the transmitters Liquiline CM42 (software package 9 or above) and Liquiline CM44x (software version 01.04.00, from 4th quarter 2012). To upgrade a Liquiline CM42 to software package 9 or higher, the digital sensor input module FSDG1 with firmware version 1.01.01 or higher is required. If the firmware version of the FSDG1 module is lower than 1:01:01, the module must be replaced. Order numbers are 51518007 (non-ex) and 51517469 (Ex). To upload software package 9 to a CM42 use the DAT CY42-S1. Data transmission to older, Memosens-capable transmitters is limited (ph measurement only). In some cases it may occur that no data is transmitted. Measuring system data which digital sensors can save in the sensor include: Manufacturer data Serial number Order code Date of manufacture Calibration data Date of calibration Calibrated slope at 25 C (77 F) Calibrated zero point at 25 C (77 F) Calibration offset (ORP mv measuring mode) Slope in % (ORP % measuring mode) Temperature offset Number of calibrations Serial number of the transmitter used to perform the last calibration Calibration database (saves the last 8 calibrations in the Memosens head) Operating data Temperature application range ph and ORP application range Date of initial commissioning Maximum temperature value Operating hours at temperatures above 80 C / 100 C (176 F / 212 F) Operating hours at very low and very high ph values (Nernst voltage under -300 mv, over +300 mv) Number of sterilizations Open aperture junction Thanks to its open aperture junction, the electrode is suitable for use in very contaminated media. Temperature compensation An NTC 30k temperature sensor is integrated in the electrode. Long service life The reference is immune to temperature and pressure fluctuations and is protected against poisoning by the ion trap. This results in a far longer sensor service life. Maximum process safety With its inductive transmission of the measured value via a non-contact plug-in connection, Memosens guarantees maximum process safety and offers the following advantages: All problems caused by moisture are eliminated: The plug-in connection is free from corrosion Moisture cannot corrupt the measured value Plug-in system can even be connected under water The transmitter is galvanically decoupled from the medium. EMC safety is guaranteed by screening measures in the digital measured value transmission. Can easily be used in hazardous areas thanks to intrinsically safe electronics. Endress+Hauser 3

Integrity Digital data transmission ensures data security Memosens technology digitizes the measured values in the sensor and transmits them to the transmitter via a non-contact connection in a way that is free from any potential interference. The result: Automatic error message generation if the sensor fails or the connection between sensor and transmitter is interrupted Immediate error detection increases measuring point availability Ease of use Sensors with Memosens technology have integrated electronics that save calibration data and other information, such as total hours of operation and operating hours under extreme measuring conditions etc. When the sensor is connected, the sensor data are automatically sent to the transmitter and used to calculate the current measured value. Saving the calibration data makes it possible to calibrate the sensor irrespective of the measuring point. The result: Convenient calibration in the measuring lab under optimum external conditions improves the quality of the calibration. Measuring point availability is dramatically increased by the quick and easy replacement of precalibrated sensors. Installing the transmitter in the measuring container with integrated measuring devices reduces the amount of fastening material and cabling work required. The availability of the sensor data makes it possible to accurately determine the maintenance intervals of the measuring point and enables predictive maintenance. The sensor history can be documented with evaluation programs such as Memobase Plus. Input Measured variables ph value ORP rh value Temperature Measuring range ph ORP rh Temperature BT version 0 to 14-1500 to 1500 mv 0 to 42 0 to 110 C (32 to 230 F) Wiring Connecting to the transmitter The sensor is connected to the transmitter via the measuring cable CYK10. Measuring cable CYK10 a0003350 4 Endress+Hauser

Installation Installation instructions Installation angle Before screwing in the electrode, make sure the assembly thread is clean and runs smoothly. Screw in the electrode finger-tight (3 Nm)! (data only applicable if installing in Endress+Hauser assemblies.) Also pay attention to the installation instructions provided in the Operating Instructions of the assembly used. Do not install the electrodes upside down. The inclination angle must be at least 15 from the horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the glass sphere and prevent the inner electrolyte from completely wetting the ph diaphragm. A 15 15 B Installation angle A Permitted orientation B Incorrect orientation a0003133 Environment Ambient temperature range Storage temperature Degree of protection NOTICE Frost can break the glass Sensor failure The sensor cannot be used at temperatures below -15 C (5 F). 0 to 50 C (32 to 120 F) IP 68 (10 m (33 ft) water column, 25 C (77 F), 45 days, 1 M KCl) Endress+Hauser 5

Process temperature range Process BT version: 0 to 110 C (32 to 230 F) Process pressure range BT version: 0 to 13 bar (0 to 190 psi) Pressure-temperature ratings Pressure-temperature ratings CPS96D a0017204 Minimum conductivity 500 μs/cm 6 Endress+Hauser

Mechanical construction Design, dimensions Design 1 Memosens plug-in head, Pg 13.5 2 Viton O-ring with thrust collar 3 Glass shaft 4 Platinum round plate (ORP element) 5 ph glass membrane 6 Ag/AgCl reference lead - ph 7 Reference with ion trap a0017212 Dimensions a0017211 Weight 0.1 kg (0.22 lbs) Materials Electrode shaft ph membrane glass ORP measuring element Junction Metal lead Gel Glass to suit process Type B Platinum round plate Open aperture junction Ag/AgCl Stabilized and firm Process connection Pg 13.5 Temperature sensor Plug-in head Reference system NTC 30K Memosens plug-in head for digital, non-contact data transmission Ag/AgCl reference lead with stabilized gel and ion trap in reference Endress+Hauser 7

Certificates and approvals Ex approval ATEX II 1G Ex ia IIC T4/T6 Ga IECEx Ex ia IIC T4/T6 Ga ATEX/Nepsi II 3G Ex nl IIC Digital sensors with Memosens technology that are suitable for use in hazardous areas have an orange/ red ring in the plug-in head. Electromagnetic compatibility Interference emission and interference immunity as per EN 61326: 2006 Ordering information Product page You can create a valid and complete order code on the Internet with the Configurator tool. Link to product page: www.products.endress.com/cps96d Product Configurator 1. You can find the following options on the right-hand side of the product page: 2. Click "Configure this product". 3. The Configurator opens in a separate window. You can now configure your device and receive the complete and valid order code. 4. Now export the order code as a PDF or Excel file. To do so, click the corresponding button at the top of the page. 8 Endress+Hauser

Accessories The most important accessories available at the time this document went to print are listed below. Please contact your Service Team or Sales Center for accessories that are not listed here. Assemblies (selection) Dipfit CPA111 Immersion and installation assembly for open and closed containers Order according to product structure (--> Online Configurator, www.products.endress.com/cpa111) Technical Information TI00112C/07/EN Dipfit CPA140 ph/orp immersion assembly with flange connection for very demanding processes Order according to product structure (--> Online Configurator, www.products.endress.com/cpa140) Technical Information TI00178C/07/EN Cleanfit CPA471 Pneumatic or manual retractable assembly for installation in tanks and pipes Order according to product structure (--> Online Configurator, www.products.endress.com/cpa471) Technical Information TI00217C/07/EN Cleanfit CPA472 Compact plastic retractable assembly for installation in tanks and pipes, for manual or pneumatically remotecontrolled operation Order according to product structure (--> Online Configurator, www.products.endress.com/cpa472) Technical Information TI00223C/07/EN Cleanfit CPA472D Robust retractable assembly for ph, ORP and other industrial sensors, for manual or pneumatically remotecontrolled operation, heavy-duty version made from very durable materials Order according to product structure (--> Online Configurator, www.products.endress.com/cpa475) Technical Information TI00403C/07/EN Cleanfit CPA473 Stainless steel process retractable assembly with ball valve shutoff for particularly reliable separation of the medium from the environment Order according to product structure (--> Online Configurator, www.products.endress.com/cpa473) Technical Information TI00344C/07/EN Cleanfit CPA474 Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of the medium from the environment Order according to product structure (--> Online Configurator, www.products.endress.com/cpa474) Technical Information TI00345C/07/EN Cleanfit CPA450 Manual retractable assembly for installation of 120 mm sensors in tanks and pipes Order according to product structure (--> Online Configurator, www.products.endress.com/cpa450) Technical Information TI00183C/07/EN Flowfit CPA250 Flow assembly for pipe installation of ph/orp sensors with Pg 13.5 and 120 mm (4.72") installation length Order according to product structure (--> Online Configurator, www.products.endress.com/cpa250) Technical Information TI00041C/07/EN Flowfit CPA240 ph/orp flow assembly for very demanding processes Order according to product structure (--> Online Configurator, www.products.endress.com/cpa240) Technical Information TI00179C/07/EN Ecofit CPA640 Adapter for 120 mm ph sensors Order according to product structure (--> Online Configurator, www.products.endress.com/cpa640) Technical Information TI00264C/07/EN Endress+Hauser 9

Buffer solutions Quality buffers from Endress+Hauser Solutions which are traced by a DAkkS-accredited Endress+Hauser buffer laboratory (DAkkS = German Accreditation Body) to a primary reference material of the PTB and to standard reference material of the National Institute of Standards and Technology (NIST) in accordance with DIN 19266 are used as secondary reference buffer solutions. ph value A ph 2.00 (accuracy ± 0.02 ph) C ph 4.00 (accuracy ± 0.02 ph) E ph 7.00 (accuracy ± 0.02 ph) G ph 9.00 (accuracy ± 0.02 ph) I ph 9.20 (accuracy ± 0.02 ph) K ph 10.00 (accuracy ± 0.05 ph) M ph 12.00 (accuracy ± 0.05 ph) Quantity 01 20 x 18 ml (0.68 fl.oz) only buffers ph 4.00 and 7.00 02 250 ml (8.45 fl.oz) 10 1000 ml (0.26 US gal) 50 5000 ml (1.32 US gal) canister for Topcal S Certificate A Buffer analysis certificate Version 1 Standard CPY20- Complete order code Technical ORP buffer solutions +220 mv, ph 7, 100 ml (3.4 fl.oz.); Order No. CPY3-0 +468 mv, ph 0.1, 100 ml (3.4 fl.oz.); Order No. CPY3-1 Measuring cable Memosens data cable CYK10 For digital sensors with Memosens technology Order according to product structure (--> Online Configurator, www.products.endress.com/cyk10) 10 Endress+Hauser

Endress+Hauser 11

TI00507C/07/EN/01.12 71163968 FM9