MANUAL Special Documentation Proline Promass F 3, 5 Low-Temperature Version for Cryogenic Media SD1966O/6/EN/2.17 Material Number: 71362224
Document function Proline Promass F 3, 5 1 Document function This manual is Special Documentation and does not replace the Operating Instructions included in the scope of supply. It is part of the Operating Instructions and provides additional information on the low temperature range. 2 Affected documentation This Special Documentation is an integral part of the following documentation: Documentation Measuring device HART FOUNDATION Fieldbus PROFIBUS PA Modbus RS485 Promass F 3 BA1663O BA1518D BA157D BA1668O Promass F 5 BA1673O BA1562D BA1551D BA1678O 2
Proline Promass F 3, 5 Performance characteristics 3 Performance characteristics o.r. = of reading; 1 g/cm 3 = 1 kg/l 3.1 Maximum measured error 3.1.1 Base accuracy Mass flow (cryogenic liquids) ±.35 % o.r. Density (cryogenic liquids) ±.5 g/cm³ 3.2 Zero point stability NOTICE Zero point confirmation and zero point adjustment are difficult to carry out in the field due to the vaporization of the cryogenic liquid. As a general rule, the factory-set zero point should not be changed. Please ensure that the medium is in the liquid phase if a zero point adjustment is to be carried out. DN Zero point stability [mm] [in] [kg/h] [lb/min] 8 ³ ₈.3.1 15 ½.2.7 25 1.54.19 4 1½ 2.25.83 5 2 3.5.129 8 3 9..33 1 4 14..514 15 6 32. 1.17 25 1 88. 3.23 3.3 Repeatability Mass flow (cryogenic liquids) ±.175 % o.r. Density (cryogenic liquids) ±.25 g/cm 3 3
Installation Proline Promass F 3, 5 4 Installation 4.1 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping). Orientation Recommendation A Vertical orientation A15591 B Horizontal orientation, transmitter at top A15589 C Horizontal orientation, transmitter at bottom D Horizontal orientation, transmitter at side A1559 A15592 4
Proline Promass F 3, 5 Process 5 Process 5.1 Medium temperature range Low-temperature version 196 to +15 C ( 32 to +32 F) Order code for "Measuring tube material", option LA NOTICE Material fatigue due to excessive temperature differential. The maximum temperature differential of the fluid used must not exceed 3 K. 5.2 Pressure-temperature ratings 5.2.1 Flange according to EN 192-1 (DIN 251) [psi] [bar] 14 13 1 9 8 PN1 12 7 8 6 5 PN63 6 4 2 4 3 2 1 PN4 PN16-2 5 1 15 [ C] -32 8 16 24 32 [ F] A32855-EN 1 With flange material 1.444 (F316/F316L) 5
Process Proline Promass F 3, 5 5.2.2 Flange according to ASME B16.5 [psi] 14 13 [bar] 1 9 8 Class 6 12 8 6 7 6 5 4 Class 3 4 2 3 2 1 Class 15-2 5 1 15 [ C] -32 8 16 24 32 [ F] 2 With flange material 1.444 (F316/F316L) A32856-EN 5.2.3 Flange JIS B222 [psi] 12 8 6 4 2 [bar] 7 6 63K 5 4 4K 3 2 2K 1 1K -2 5 1 15 [ C] -32 8 16 24 32 [ F] 3 With flange material 1.444 (F316/F316L) A32854-EN 6
Proline Promass F 3, 5 Process 5.3 Cold insulation It is not generally necessary to insulate the sensor. If insulation is required, the information in the original documentation applies. 7
Important All information and technical specifications in this documentation have been carefully checked and compiled by the author. However, we cannot completely exclude the possibility of errors. TechnipFMC is always grateful to be informed of any errors. Contact us on the website. Smith Meter is a registered trademark of TechnipFMC. Technical Support Contact Information: Field Service Response Center 24/7 Technical Support/Schedule a Technician: 1-844-798-3819 System Installation Supervision, Start-Up, and Commissioning Services Available Customer Support Contact Information: Customer Service TechnipFMC 162 Wagner Avenue Erie, Pennsylvania 1651 USA P: +1 814 898-5 F: +1 814 899-8927 measurement.solutions@technipfmc.com TechnipFMC.com Literature Library: http://fmctechnologies.com/en/measurementsolutions/onlineservices.aspx The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in effect. Otherwise, the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect. Contact information is subject to change. For the most current contact information, visit our website at TechnipFMC.com and click on the Contact Us link. TechnipFMC.com FMCTechnologies.com/MeasurementSolutions.com TechnipFMC 217 All rights reserved. SD1966O/6/EN/2.17 TechnipFMC FMC Technologies Measurement Solutions, Inc. 5 North Sam Houston Parkway West, Suite 1 Houston, Texas 7767 USA P:+1 281.26.219 USA Operation 162 Wagner Avenue Erie, Pennsylvania 1651 USA P:+1 814.898.5 Germany Operation Smith Meter GmbH Regentstrasse 1 25474 Ellerbek, Germany P:+49 411 34.