OPERATOR S INSTRUCTION MANUAL

Similar documents
DVM98. True RMS Digital Multimeter. 1 Safety information. 1.1 Preliminary. 1.2 During use

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

1. SAFETY 1.1. SAFETY INFORMATION 1.2. SAFETY SYMBOLS

DIGIT & POINTER MULTIMETER

Model DIGITAL MULTIMETER

DIGITAL MULTIMETER INSTRUCTION MANUAL

AMM-1022 Digital Multimeter USER`S MANUAL

Digital Multimeter with Backlight

1. SAFETY INFORMATION.1 2. DESCRIPTION SPECIFICATIONS.6 4. OPERATING INSTRUCTION Voltage measurement Current measurement 10

MS8216 DIGITAL MULTIMETER OPERATOR S MANUAL

DMM8900 SERIES USERS MANUAL

INSTRUCTION MANUAL DIGITAL MULTIMETER

MS8211 DIGITAL MULTIMETER INSTRUCTION MANUAL

DVM1190 DIGITAL MULTIMETER

Sales: Technical: Fax:

PEN TYPE DIGITAL MULTIMETER OPERATION MANUAL T8211D

MS6231 DIGITAL ENGINE ANALYZER OPERATOR S MANUAL CONTENTS CONTENTS

EM420A/420B DIGITAL MULTIMETER OWNERS MANUAL Read this owners manual thoroughly before use

DIGITAL MULTIMETER OPERATING INSTRUCTIONS MODEL CDM-35. Part No

AC/DC CLAMP METER USER S MANUAL

Item ref: UK MTM01 DIGITAL MULTITESTER. User Manual

Thank you again for choosing AstroAI, if you have any questions or concerns regarding your product, please contact us at

Model ST Instruction Manual. True RMS Autoranging Digital Multimeter. reedinstruments. www. com

DIGITAL MULTIMETER OPERATOR S MANUAL

DIGITAL MULTIMETER OPERATOR'S INSTRUCTION MANUAL HOLD 10A COM LIGHT MS8265 ON/OFF. 200M KHz 2K 20K μ μ μ n.

Digital Multimeter AX-572. Instruction manual

M-1000D DIGITAL MULTIMETER

AUTO RANGING DIGITAL MULTIMETER

OWNER S MANUAL HH0308C. AUTO-RANGING DC/True RMS AC DIGITAL MULTIMETER

NOTE: Fully read and understand this manual before using this Digital Multimeter.

AC/DC DIGITAL CLAMP METER OPERATION MANUAL

ProfiScale MULTI Multimeter

OPERATOR S INSTRUCTION MANUAL M-2625 AUTO RANGING DIGITAL MULTIMETER

HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

DVM645BI BENCH MULTIMETER TAFELMULTIMETER MULTIMETRE DE TABLE BANCO MULTÍMETRO TISCHMULTIMETER. User Manual. Gebruikershandleiding

770E DIGITAL MULTIMETER OPERATOR S MANUAL

DIGITAL MULTIMETER CONTENTS DIGITAL MULTIMETER CONTENTS

Analog Technologies VC99. Multimeter FEATURES

1.General instructions Specifications Description...7

VC835 DIGTAL MULTIMETER Operation Manual

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.

MS8223A. CAT.IIi 600 V

DIGITAL MULTIMETER. MODEL CDM-50 Part No OPERATING INSTRUCTIONS

Item ref: UK MTTR01 TRUE RMS DIGITAL MULTITESTER WITH USB INTERFACE. User Manual

Multifunction Digital

MM V 10A ENGLISH. INSTRUCTION MANUAL Auto-Ranging DATA HOLD AUDIBLE CONTINUITY MIN / MAX TEMPERATURE DIODE TEST CAPACITANCE

User s Guide. MultiView Series Digital MultiMeters Models: MV110 MV120 MV130

MS8268 HANDHELD DIGITAL MULTIMETER OPERATOR S INSTRUCTION MANUAL

MS8250D DUAL DISPLAY DIGITAL MULTIMETER User s Manual

MS8250A/B OPERATION MANUAL MS8250A. Hz% FUNC REL RANGE REL HOLD OFF 10A. Hz% A NCV. Hz% COM. A ma 10A FUSED 600V CAT IV.

DIGITAL MULTIMETER AX-588B

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

ETHOS 5030 TRUE-RMS DIGITAL MULTIMETER OPERATION MANUAL

INSTRUCTION MANUAL MODEL 21 MULTI-TESTER

RAGU 81D DIGITAL MULTIMETER OPERATION MANUAL

AVM360 Analog multimeter OPERATION MANUAL GEBRUIKERSHANDLEIDING MANUEL D UTILISATEUR

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL-860A OPERATION MANUAL

Pocket Size Digital Multimeter Models: and

4.5 Auto Power Off During use Preparation For Measurement Symbols...3

MM700. INSTRUCTION MANUAL Auto-Ranging Digital Multimeter True RMS

USER'S MANUAL DMR-4300

User Manual. All rights reserved. Specifications are subject to change without notice.

The Meter Structure The Figure 2-1 shows the Meter structure.

User Manual Digital Multimeter

User Manual Digital Multimeter. model no.: MSR-U1000

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

User s Guide. 400A AC/DC Clamp Meter. Model MA Washington Street Melrose, MA Phone Toll Free

3B SCIENTIFIC PHYSICS

User s Guide. 400A AC/DC Clamp Meter. Model MA220

UT207A/208A/209A Operating Manual. Table of Contents

DC/AC CLAMP METER ISO-TECH 2000 USERS MANUAL

Digital Clamp Meter Model: &

DIGITAL MULTIMETER AUTORANGING

KMD-S04 Multímetro de bolsillo

OPERATOR S INSTRUCTION MANUAL

Table of Contents Title Page

EX350 Series USER GUIDE. True RMS Digital Multimeters. EX350 True RMS Digital Multimeter EX355 True RMS Digital Multimeter with Temperature

INSTRUCTION MANUAL. Model Autoranging DMM ProbeMeter TM. Measures voltage, resistance, frequency, capacitance, temperature, and duty cycle.

USER'S MANUAL DMR-6700

MULTIMETRE DIGITAL 3 3/4 AVEC ECHELLE AUTOMATIQUE & EXTINCTION AUTOMATIQUE MANUEL D UTILISATION

Handheld Digital Multimeter PRO-50A

Model UT39A/B/C: OPERATING MANUAL

User s Manual. MiniTec TM Series. Model MN26 (Model MN26T includes temperature probe) Mini Autoranging MultiMeter

METRAVI. Table of Contents

Pen-type DIGITAL MULTITESTER

Accuracy : ± (0.8% rdg + 1dgt) on all ranges except Ranges : 200mV, 2V, 20V, 200V, 1000V

MS8250D DUAL DISPLAY DIGITAL MULTIMETER User s Manual

Digital Clamp Meter (TRMS) Model : 2727

Dawson DDM181. Pocket-Size Autorange Digital Meter User s Manual

OPERATING INSTRUCTION

400Amp True RMS AC/DC Clamp Meter Model EX613

DIGITAL DUAL DISPLAY AC/DC CLAMP METER MODEL- KM 2782 / 2782-T

Model : OPERATING MANUAL Table of Contents (1)

BENCH METER Model>9803. Wavecom Instruments

Dawson DDM190. Digital Multimeter User s Manual

Pocket Size Digital Multimeter with USB Model: A & A

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

R/C SMD Tweezers Multimeter Model RC200

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

EX360 Series USER MANUAL. True RMS Digital Multimeters. EX360 True RMS Digital Multimeter EX363 True RMS DMM with Temperature and µa AC/DC

Transcription:

OPERATOR S INSTRUCTION MANUAL AUTO-RANGE DUAL DISPLAY CONFORMED IEC1010 DIGITAL MULTIMETER

CONTENTS PAGE SAFETY INFORMATION..... DESCRIPTION.. OPERATING INSTRUCTION.. SPECIFICATIONS.... ACCESSORIES. BATTERY AND FUSE REPLACEMENT... 1 5 10 17 26 27

1. SAFETY INFORMATION The meter has been designed according to IEC-1010 concerning electronic measuring instruments with an overvoltage category (CAT II) and pollution 2. Follow all safety and operating instructions to ensure that the meter is used safely and is kept in good operating condition. 1.1 PRELIMINARY When using the meter, the user must observe all normal safety rules concerning : -Protection against the dangers of electrical current. -Protection of the meter against misuse. Full compliance with safety standards can be guaranteed only if used with test leads supplied. If necessary, they must be replaced with the same model or same -1-

electric ratings. Measuring leads must be in good condition. 1.2 DURING USE Never exceed the protection limit values indicated in specifications for each range of measurement. When the meter is linked to a measurement circuit, do not touch unused terminals. When the value scale to be measured is unknown beforehand, set the range selector at the highest position. Before rotating the range selector to change functions, disconnect test leads from the circuit under test. When carrying out measurements on TV or switching power circuits, always remember that there may be high amplitude voltage pulses at test points, which -2-

can damage the meter. Never perform resistance measurements on live circuits. Always is careful when working with voltages above 60V dc or 30V ac rms. Keep fingers behind the probe barriers while measuring. 1.3 SYMBOLS -3- Important safety information, refer to the operating manual. Dangerous voltage may be present. Earth ground. Double insulation (Protection class II).

1.4 MAINTENANCE Before opening the meter, always disconnect test leads from all sources of electric current. For continue protection against fire, replace fuse only with the specified voltage and current ratings: F 15A/250V If any faults or abnormalities are observed, the meter can not be used any more and it has to be checked out. Never use the meter unless the back cover is in place and fastened fully. To clean the meter, use a damp cloth and mild detergent only, do not use abrasives of solvents on it. -4-

2. DESCRIPTION This meter is an autoranging professional measuring instrument. Digit reading is 3999 counts and the bar graph consists of 38 segments for LCD, capable of performing functions: -DC voltage measuring (Auto Ranging) -AC voltage measuring (Auto Ranging) -DC current measuring -AC current measuring -Temperature measuring -Resistance measuring (Auto Ranging) -Capacitance measuring -5-

-Diode testing -Transistor testing -Audible continuity testing 2.1 ON/OFF SWITCH A push push switch is used to turn the meter on or off. When the switch is push the beeper will sound. 2.2 DC/AC OR Ω / BUTTON This button is used to select DCV or ACV on voltage range, DCA or ACA on current range, Ω or on Ω / range. When the button is push the beeper will sound. 2.3 R-H BUTTON In auto-range, press this button to select manual range and R-H symbol will appear on -6-

the LCD. Press the button again for more than 1 second will come back the auto-range mode. When the R-H button is pushed, voltage ranges will change as: 4V 40V 400V 1000V Or resistance ranges change as: 40MΩ 4MΩ 400kΩ 40kΩ 4kΩ 400Ω 2.4 B/L BUTTON Press this button to toggle back-lit ON. The screen of the meter will light. Press the button again, back-lit turn off. -7-

2.5 INPUT JACKS This meter has four input jacks that are protected against overload to the limits. During use, connect the black test lead to the COM jack and the red test lead as shown below: FUNCTION RED LEAD CONNECTION INPUT LIMITS DCV/ACV Ω ma 10A V/Ω V/Ω V/Ω ma 10A 1000V dc or 750V rms ac 250V dc or rms ac 250V dc or rms ac 400mA dc or rms ac 10A dc or rms ac -8-

2.6 FRONT PANEL 1. DISPLAY 2. ON/OFF SWITCH 3.TRANSISTOR TESTING SOCKET 4. FUNCTION SWITCH 5. CAPACITANCE OR TEMPERATURE TESTING SOCKET 6. INPUT JACKS 7. RS232C SOCKET 8. B/L BUTTON 9. R-H BUTTON 10. DC/AC OR Ω / BUTTON

-9-3. OPERATING INSTRUCTION 3.1 VOLTAGE MEASURING 1. 2. 3. 4. Connect the black test lead to the COM jack and the red test lead to the V/Ωjack. Set the function switch at V range and connect test leads across the source or load under measurement. Push the DC/AC button to select DCV or ACV measuring mode. (Push the R-H button to select manual range.) Read LCD display. The polarity of red test lead connection will be indicated when making a DC measurement. 3.2 CURRENT MEASURING

1. Connect the black test lead to the COM jack and the red test lead to the ma jack -10- for a maximum of 400mA. For a maximum of 10A, move the red lead to the 10A jack. 2. 3. 4. Set the function switch at ma or 10A range and push the DC/AC button to select DCA or ACA measuring mode. Connect test leads in series with the load in which the current is to be measured. Read LCD display. The polarity of red test lead connection will be indicated when making a DC measurement. 3.3 RESISTANCE MEASURING 1. Connect the black test lead to the COM jack and the red teas lead to the V/Ω jack. (NOTE: The polarity of the red test lead connection is positive + )

2. Set the function switch at Ω range to be used and connect test leads across the resistance under measurement. -11- NOTE: 1. For resistance above 1MΩ, the meter may take a few seconds to stabilize reading. This is normal for high resistance measuring. 2. When the input is not connected, i.e. at open circuit, the figure OL will be displayed and full bar graph appears for the overrange condition. 3. When checking in-circuit resistance, be sure the circuit under test has all power removed and all capacitors fully discharged. 3.4 CAPACITANCE MEASURING

1. 2. Set the function switch at nf range. Before inserting capacitor under measurement into capacitance testing socket, be sure that the capacitor has been discharged fully. WARNING: Before attempting to insert capacitor for testing, always be sure that test leads have been disconnected from any measurement circuits. Components should not be connected to the capacitor socket when making voltage measurements with test leads. 3.5 TEMPERATURE MEASURING 1. Set the function switch at TEMP range. -12-

2. Insert K type thermocouple into the temperature measuring socket on front panel and contact the object to be measured with the thermocouple probe. -13- WARNING: To avoid electric shock, be sure the thermocouple has been removed before changing to another function measurement. Before attempting to insert thermocouple for testing, always be sure that test leads have been disconnected from any measurement circuits. 3.6 AUDIBLE CONTINUITY TESTING

1. 2. 3. Connect the black test lead to the COM jack and the red test lead to the V/Ω jack. (NOTE: The polarity of red test lead connection is positive + ) Set the function switch at position and push the Ω/ button. Connect test leads to two points of the circuit under measurement. If continuity exists (i.e., the resistance is lower than 30Ω), built-in buzzer will sound. 3.7 DIODE TESTING 1. Connect the black test lead to the COM jack and the red test lead to the V/Ω jack. -14-2. 3. (NOTE: The polarity of the red test lead connection is positive + ) Set the function switch at position. Connect the red test lead to the anode of the diode to be tested and the black test lead to the cathode.

4. Read LCD display to get the forward voltage drop of the diode under testing. 3.8 TRANSISTOR TESTING 1. Set the function switch at hfe position. 2. Determine whether the transistor is NPN or PNP type and locate emitter, base, and collector lead Insert leads of the transistor into proper holes of the transistor testing socket on the front panel. 3. The meter will show the approx. hfe value at test condition of base current 10µA and Vce 3.0V. -15- WARNING: Before attempting to insert transistor for testing, always be sure that test leads have been disconnected from any measurement circuits.

Components should not be connected to the transistor socket when making voltage measurements with test leads. 3.9 INTERFACING THE METER WITH A PC (OPTION) 1. 2. Connect the RS232C cable between the meter and computer s serial ports. Press ON/OFF to turn on the meter. Turn on the computer. WARNING: 1. Use only the type of serial interface cable RS232C cable exclusively designed -16- for your meter. Never attempt to modify of extend the length or RS232C cable. 2. Please read this file of readme from Disks. 4. SPECIFICATIONS

Accuracy is specified for a period of one year after calibration and at 18 to 28 (64 F to 82 F) with relative humidity to 75%. 4.1 GENERAL Max. Voltage Between Terminals and Earth Ground: 1000V dc or 750V rms ac (sine) Power Supply: 9V NEDA1604 6F22 006P Ranging Method: Auto / Manual Display: LCD, 3999 counts max and bar graph consists of 38 segments -17-

Overrange Indication: OL displayed Polarity Indication: - displayed automatically Low Battery Indication: displayed Operating Temperature: 5to 35(41 F to 95 F) Storage Temperature: -10 to 60(14 F to 140 F) Dimension: 78mm 186mm 35mm Weight: 300g(Including battery) -18-

4.2 DC VOLTAGE Range Resolution Accuracy 4V 1mV 40V 10mV ±0.5% of rdg±3 digits 400V 0.1V 1000V 1V ±0.8% of rdg±3 digits Input Impedance: 10MΩ. -19-

4.3 AC VOLTAGE Range Resolution Accuracy 4V 1mV 40V 10mV ±1.2% of rdg±5 digits 400V 0.1V 750V 1V ±1.5% of rdg±5 digits Input Impedance: 10MΩ Frequency Range: 40 to 400Hz Response: Average, calibrated in rms of sine wave. -20-

4.4 RESISTANCE Range Resolution Accuracy 400Ω 0.1Ω 4kΩ 1Ω 40kΩ 10Ω ±1.2% of rdg±3 digits 400kΩ 0.1kΩ 4MΩ 1kΩ 40MΩ 10kΩ ±3.0% of rdg±5 digits Maximum Open Circuit Voltage: 3.0V Overload Protection:250V dc or rms.ac for all ranges. -21-

4.5 DC CURRENT Range Resolution Accuracy 4mA 1µA ±1.2% of rdg±3 digits 400mA 0.1mA 10A 10mA ±2.0% of rdg±8 digits Overload Protection: F 15A/250V fuse for 10A range. -22-

4.6 AC CURRENT Range Resolution Accuracy 4mA 1µA ±1.5% of rdg±8 digits 400mA 0.1mA 10A 10mA ±3.0% of rdg±8 digits Overload Protection: F 15A/250V fuse for 10A range. Frequency Range: 40 to 400Hz Response: Average, calibrated in rms of sine wave. -23-

4.7 CAPACITANCE Range Resolution Accuracy 4nF 1pF 400nF 0.1nF ±4.0% of rdg±5 digits 4.8 TEMPERATURE Range Resolution Accuracy 0 to 400 1 ±3.0% of rdg±3 digits 401 to 750 1 ±3.0% of rdg±5 digits -24-

4.9 AUDIBLE CONTINUITY Function Description Built-in buzzer will sound, if resistance is lower than 30Ω. 4.10 DIODE Function Resolution Test Current Open Circuit Voltage 1mV 25µA 3.0V 4.11 TRANSISTOR Function Range Base Current Vce hfe 1 to 1000 10µA 3.0V -25-

5. ACCESSORIES 5.1 SUPPLIED WITH THE METER Operating Manual Set of test leads 9V battery. NEDA1604 6F22 006P K type thermocouple 5.2 OPTIONAL ACCESSORY Holster RS232C Cable Set of 3.5 1.44MB disk -26-

6. BATTERY AND FUSE REPLACEMENT If the sign appears on the LCD display, it indicates that the battery should be replaced. Remove screws on the back cover and open the case. Replace the exhausted battery with a new one. Fuse rarely need replacement and blow almost always as a result of the operator s error. Open the case as mentioned above and take the PCB assembly out from the case. Replace the blown fuse with ratings specified. WARNING: Before attempting to open the case, be sure that test leads have been disconnected from measurement circuit to avoid electric shock hazard. For protection against fire, replace fuses only with specified ratings: F 15A/250V -27-

F=( C 1.8+32 ) HYS004695 A0