THE DECI REPORT. H. L. Dunegan. August, 2000 AN ALTERNATIVE TO PENCIL LEAD BREAKS FOR SIMULATION OF ACOUSTIC EMISSION SIGNAL SOURCES.

Similar documents
ELIMINATION OF EXTRANEOUS NOISE SOURCES FROM ACOUSTIC EMISSION BASED TERMITE DETECTION INSTRUMENT BY USE OF MODAL RATIOS H.L. DUNEGAN AUGUST 15, 2001

EFFECTS OF LATERAL PLATE DIMENSIONS ON ACOUSTIC EMISSION SIGNALS FROM DIPOLE SOURCES. M. A. HAMSTAD*, A. O'GALLAGHER and J. GARY

ACOUSTIC EMISSION MEASUREMENTS ON SHELL STRUCTURES WITH DIRECTLY ATTACHED PIEZO-CERAMIC

EWGAE 2010 Vienna, 8th to 10th September

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING

Acoustic Emission Basic Process and Definition

ISO INTERNATIONAL STANDARD. Non-destructive testing Acoustic emission inspection Secondary calibration of acoustic emission sensors

USB-UT350(T) Portable Ultrasonic Pulser/Receiver and Analog to Digital Converter. User s Guide

DETECTION AND SIZING OF SHORT FATIGUE CRACKS EMANATING FROM RIVET HOLES O. Kwon 1 and J.C. Kim 1 1 Inha University, Inchon, Korea

Section 7 - Measurement of Transient Pressure Pulses

DEVELOPMENT OF HEAT-RESISTANT OPTICAL FIBER AE SENSOR

A Detailed Examination of Waveforms from Multiple Sensors on a Composite Pressure Vessel (COPV)

Frequency Considerations in Air-Coupled Ultrasonic Inspection.

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term...

Transducer product selector

Recommendation of RILEM TC 212-ACD: acoustic emission and related NDE techniques for crack detection and damage evaluation in concrete*

CHAPTER 3 ACOUSTIC EMISSION TECHNIQUE FOR DETECTION AND LOCATION OF PD


Resonance Tube Lab 9

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

Improving CDM Measurements With Frequency Domain Specifications

D.C. Emmony, M.W. Godfrey and R.G. White

JOURNAL OF ACOUSTIC EMISSION

Acoustic emission signal attenuation in the waveguides used in underwater AE testing.

First and second order systems. Part 1: First order systems: RC low pass filter and Thermopile. Goals: Department of Physics

Guide to OPKUD and OPBOX Ultrasonic testing units Software Revision 3.0 / 2003

RODS AND TUBES AS AE WAVEGUIDES

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

A System to Measure Instantaneous Compressor Crankshaft Speed

COMPOSITES FROM PIEZOELECTRIC FIBERS AS SENSORS AND EMITTERS FOR ACOUSTIC APPLICATIONS*

INTERFERENCE OF SOUND WAVES

Improving the Noise Discrimination for an Ammonia Tank AE Test

Time Reversal FEM Modelling in Thin Aluminium Plates for Defects Detection

A GENERIC TECHNIQUE FOR ACOUSTIC EMISSION SOURCE LOCATION

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2

EXPERIMENTAL TRANSFER FUNCTIONS OF PRACTICAL ACOUSTIC EMISSION SENSORS

DETECTION OF PAD CRATER INITIATION IN SHOCK USING ACOUSTIC EMISSION DETECTION

MATERIALS CHARACTERIZATION USING LASER ULTRASONIC GUIDED WAVES

FUNCTION GENERATOR LCE NR WITH MODULATION

A COMPACT EMAT RECEIVER FOR ULTRASONIC TESTING AT ELEVATED TEMPERATURES* L. R. Burns, G. A. Alers, and D. T. MacLauchlan

ON LAMB MODES AS A FUNCTION OF ACOUSTIC EMISSION SOURCE RISE TIME #

Acoustic Transducer*

SELECTION OF MATERIALS AND SENSORS FOR HEALTH MONITORING OF COMPOSITE STRUCTURES

An experimental investigation of cavity noise control using mistuned Helmholtz resonators

Description of a device and software for precise sound velocity measurement

Electronic Noise Effects on Fundamental Lamb-Mode Acoustic Emission Signal Arrival Times Determined Using Wavelet Transform Results

Absolute Calibration of Acoustic Emission Transducers as per CEN ISO/TR in Disuse of Mechanical Sound Sources or Reference Transducers

Developments in Ultrasonic Phased Array Inspection III

PIEZOELECTRIC WAFER ACTIVE SENSORS FOR STRUCTURAL HEALTH MONITORING STATE OF THE ART AND FUTURE DIRECTIONS

CONTINUOUS DAMAGE MONITORING TECHNIQUES FOR LAMINATED COMPOSITE MATERIALS

Exercise 2: Distance Measurement

Standard Guide for Evaluating Performance Characteristics of Phased-Array Ultrasonic Testing Instruments and Systems 1

NONDESTRUCTIVE EVALUATION OF CLOSED CRACKS USING AN ULTRASONIC TRANSIT TIMING METHOD J. Takatsubo 1, H. Tsuda 1, B. Wang 1

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD 213, SHOCK (SPECIFIED PULSE)

Guided Wave Travel Time Tomography for Bends

A Numerical study on proper mode and frequency selection for riveted lap joints inspection using Lamb waves.

The Use of a Floating Threshold for Online Acoustic Emission Monitoring of Fossil High Energy Piping

SOME OBSERVATIONS ON RAYLEIGH WAVES AND ACOUSTIC EMISSION IN THICK STEEL PLATES #

DEVELOPMENT OF MEASUREMENT SYSTEM USING OPTICAL FIBER AE SENSORS FOR ACTUAL PIPING

DAMAGE DETECTION IN PLATE STRUCTURES USING SPARSE ULTRASONIC TRANSDUCER ARRAYS AND ACOUSTIC WAVEFIELD IMAGING

F. N. HOOGE and A. M. H. HOPPENBROUWERS Philips Research Laboratories, N. V. Philips Gloeila~penfabrieken, Eindhoven, Nederland

ULTRASOUND IN CFRP DETECTED BY ADVANCED OPTICAL FIBER SENSOR FOR COMPOSITE STRUCTURAL HEALTH MONITORING

Testing Sensors & Actors Using Digital Oscilloscopes

Retuning of FMIT #3 from 70.0 MHZ to 78.0 MHZ Charley Schwartz 10/28/02 DRAFT

Title: Reference-free Structural Health Monitoring for Detecting Delamination in Composite Plates

Exam Booklet. Pulse Circuits

Ultrasound Brain Imaging System

MEASUREMENT OF SURFACE DISPLACEMENT EXCITED BY EMAT TRANSDUCER

Lab 1: Basic Lab Equipment and Measurements

Investigation on Acoustic Emission of Hydrogen Induced Cracking for Carbon Steel Used in Pressure Vessel

MEASURING HUM MODULATION USING MATRIX MODEL HD-500 HUM DEMODULATOR

Features. Haltronics Ltd (

Ultrasonic Level Detection Technology. ultra-wave

IPC-TM-650 TEST METHODS MANUAL

FATIGUE CRACK CHARACTERIZATION IN CONDUCTING SHEETS BY NON

Capacitive Touch Sensing Tone Generator. Corey Cleveland and Eric Ponce

IMAGING OF DEFECTS IN CONCRETE COMPONENTS WITH NON-CONTACT ULTRASONIC TESTING W. Hillger, DLR and Ing. Büro Dr. Hillger, Braunschweig, Germany

Intelligent Acoustic Emission System

Validated AE Application for Continuous Monitoring of the Structural Condition of the Supporting Structure of Offshore Wind Turbines

RECENT ADVANCEMENTS IN THE APPLICATION OF EMATS TO NDE

A Wire-Guided Transducer for Acoustic Emission Sensing

FISCHER CUSTOM COMMUNICATIONS, INC.

Department of Electrical Engineering and Computer Science

Kistler portable triaxial Force Plate

Television and video engineering

MULTIPLEX SENSOR TELEMETRY

NOVEL ACOUSTIC EMISSION SOURCE LOCATION

Underwater Acoustics. A Brief Introduction. Ethem Mutlu Sözer Research Engineer MIT Sea Grant College Program

Easy Ultrasonic Phased Array Inspection of Corrosion - Resistant Alloys and Dissimilar Weld Materials

Laboratory Exercise 6 THE OSCILLOSCOPE

Measuring of small AC signals using lock-in amplifiers. Narrow band selective amplifiers + amplitude detector. Lock-in amplifiers

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

3. Sound source location by difference of phase, on a hydrophone array with small dimensions. Abstract

X. SPEECH ANALYSIS. Prof. M. Halle G. W. Hughes H. J. Jacobsen A. I. Engel F. Poza A. VOWEL IDENTIFIER

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

ULTRASONIC GUIDED WAVES FOR AGING WIRE INSULATION ASSESSMENT

-60- XVI. VISUAL REPLACEMENT PROJECTS. Dr. C. M. Witcher L. Washington, Jr. A. STEP-DOWN DETECTOR

ARTICLE 29 ACOUSTIC EMISSION STANDARDS

Ultrasonic Testing. Portable Instruments Industrial inspection systems Air-coupled Testing Transducers

Transcription:

THE DECI REPORT H. L. Dunegan August, 2000 AN ALTERNATIVE TO PENCIL LEAD BREAKS FOR SIMULATION OF ACOUSTIC EMISSION SIGNAL SOURCES. INTRODUCTION Over 25 years ago Nelson Hsu while working with Cliff Bailey at Lockheed Georgia discovered that breaking of a pentel pencil lead on the surface of an aircraft panel would act as a simulated source of acoustic emission signals. Since that time the pentel pencil lead break has been in common use all over the world for simulating sources of acoustic emission for calibration purposes. We have found in our research that the pencil lead break on the surface of a plate replicates the type of source one encounters from noise sources such as impact and friction and pencil lead breaks on the edge of a plate replicates the type of source produced by a growing crack in a plate. One of the main problems encountered when using the pencil lead break is the difficulty in reproducing the same amplitude and phase of the signal. Another drawback is the difficulty in producing this type of simulated signal on a large structure. Another method used for simulating acoustic emission sources for many years is the use of a resonant transducer (usually in the 150Khz range of frequencies) used as a pulser. As will be shown the use of the 150Khz transducer as a transmitter on the surface of the plate is not a good simulation of the type of signal produced by the pencil lead break or crack growth in the plate, due to the large aperture of the 150Khz transducer. It was our desire to design a repeatable source of crack like signals that could be placed anywhere on a large structure. Since the pencil lead break has been shown to adequately replicate the growth of a crack, it was our desire to design a pulser-transducer combination that would simulate the type of signals produced by the pencil lead break. After many design changes and tests on many types of transducers used as transmitters we produced the Model 600 pulser working into the SE25-P miniature small aperture transducer as a transmitter as the best combination for simulating pencil lead breaks on plates. This report shows some of the experimental results from our study. 1

Model 600B battery supply Model 600 Pulser Model SE9125- MI Transducer Figure 1 EXPERIMENTAL PROCEDURE SE25-P Transducer SE375-MI Transducer 0.5mm pentil pencil A 500mm X 500mm aluminum plate 4.6mm thick (figure 1) was utilized for the experimental work. A trigger transducer (SE375-MI) was placed at one edge of the plate, and a data transducer (SE9125-MI) was placed at approximately 250mm from the trigger transducer. Signals from the data transducer were input to the AESMART 2000 instrument where they were split into a 20 to 60Khz band pass (LF) and a 100-500Khz bandpass (HF). After 40dB of amplification, signals in both frequency ranges were captured for display on a digital oscilloscope. Signals from the trigger transducer were also input into the AESMART 2000 and amplified by 40dB. The SYNC output of the trigger channel was used to externally trigger the sweep of the digital oscilloscope. Both the SE9125-MI and SE375-MI transducers contain internal preamplifiers with 20dB of fixed gain. 0.5mm pentel pencil lead breaks were made adjacent to the trigger transducer on top of the plate (out-of-plane (OOP) source representing noise sources) and on the edge of the plate (in-plane IP source representing crack growth sources). These signals were captured and used as a reference source for experiments utilizing the pulser and different transmitting transducers to find the best match. EXPERIMENTAL RESULTS Figure 2 shows the results of breaking a 0.5mm pencil lead on the top surface of the plate adjacent to the trigger transducer. Note in the HF (high frequency) channel the small signal at approximately 50 microseconds. This is the arrival of a very low level extensional wave created by the OOP source. The large signal arriving at approximately 80 microseconds is a high frequency shear wave with a calculated velocity of 3,200M/sec. The LF (low frequency) channel shows a large low frequency signal arriving at approximately 100 microseconds. This is the flexure wave created by the OOP pencil lead break. Note that the peak amplitude of the HF signal divided by the peak amplitude of the LF signal (HF/ LF) is approximately one. 2

Figure 3 shows the results obtained when the pencil lead is broken on the edge of the plate (IP source). Note that the first arrival in the HF channel occurs at approximately 50 microseconds. This is the arrival of the extensional wave (crack like source) at the data transducer. A value of 5,000M/sec is observed for this extensional wave. Note that the signal received in the LF channel is much lower than observed in figure 2 for an OOP source. The amplitude of this LF signal is proportional to the depth in the plate where the pencil lead is broken. This feature allows one to estimate the depth of a growing crack by observing the value of the HF/LF ratio. If broken exactly in the center of the plate, no flexure wave component is observed and the ratio of the peak amplitudes-hf/lf is infinity. The software in the AESMART 2000 gives a default value of 60 for these cases. Note that the peak amplitude ratio (HF/LF) is greater than one for this IP source. Figure 4 shows the results for the SE25-P miniature transducer being used as a transmitter and driven by a 150 volt pulse from the model 600 pulser. For this data the transducer was placed on top of the plate OOP source (simulated noise source). Instrumentation gain was kept constant throughout these tests. We see from figure 3 in comparison to figure 2 that a higher amplitude signal is produced by the pulser as opposed to the 0.5mm pencil lead break, but similar HF/LF ratios are observed. Figure 5 shows the results for the SE25-P transducer for identical conditions for figure 4 with the exception that the transducer was placed on the edge of the plate adjacent to the trigger transducer to produce an IP (crack like) source. Note the similarity between figure 3 and figure 5. Again we observe a higher amplitude signal from the pulsed SE25-P as opposed to the 0.5mm pencil lead break. 3

Many researchers for years have used a 150Khz resonant transducer as a pulser for calibration of multiple channel AE systems. We decided to compare the SE150-M transducer with the results achieved from the pencil lead breaks and the SE25-P. Therefore a SE150-M transducer was placed on top of the plate adjacent to the trigger transducer and pulsed with the 150 volt pulse from the model 600 pulser to act as an OOP source. All other instrumentation settings were identical to those used for the data in figures 2 through 5. Figure 6 shows the results of this experiment. An extensional wave arrival is shown at 50 microseconds in the HF channel. Due to the high amplitude of this signal, no separation is shown between the extensional wave and shear wave. The 15 volt peak signal seen at 150 microseconds is a reflection from the back surface of the plate. The signal in the LF channel is very low level in comparison to the HF channel and shows very incoherent wave properties. This is due to the large aperture of the SE150-M in comparison to the SE25-P transducer and pencil lead breaks used in the previous figures. Figure 7 shows the results of placing the SE150-M on the edge of the plate to produce an IP (crack like) source. For this situation there is a large amplitude signal arriving at the extensional wave velocity and a small separation between the extensional and shear wave. The signals produced in the LF channel have similar properties to those produced by this sensor used as an OOP source. It is obvious that the 150 Khz resonance transducer is a poor choice to use as a transmitter for ratio analysis to identify the type of source producing an AE signal. The next experiment performed was designed to determine the difference between the pencil lead break and the pulser performance of the SE25-P. This was accomplished by first breaking a 0.5mm pencil lead break directly on the 3.1mm aperture of the SE25-P. The signal produced was large enough that the output of the transducer could be input directly into the digital oscilloscope without any amplification. Another SE25-P transducer used as a pulser was coupled with petroleum jelly face to face with the transducer used in the pencil lead break above. Again the 150 volt selection was made on the model 600 pulser. 4

Figure 8 0.5mm pencil lead break on SE250-P Two SE25-P face to face with Mod. 600 pulser 150V position on 600 pulser SE25-P output directly into scope in both cases - no amplification Figure 8 shows the results of this experiment. The upper trace in figure 8 is the signal produced by breaking the pencil lead directly on the face plate of the 3.1mm diameter aperture of the transducer. The lower trace shows the results produced by face to face coupling with one transducer used as a transmitter. Both traces show a very fast rise time to peak amplitude and some high frequency ringing of the transducer with the peak response of the pulsed transducer reaching a higher level than the pencil lead break. This higher amplitude signal from the pulser was also observed from the plate waves generated in the previous figures. CONCLUSIONS These results show that the model 600 pulser with its switch in the 150 volt position in combination with a SE25-P miniature small aperture transducer produces signals in plates that are very similar to those produced by 0.5mm pencil lead breaks. It is further shown that the traditional use of a 150Khz transducer as a transmitter does not have similar characteristics to pencil lead breaks, and the distortion produced by the large aperture of this transducer in the low frequency range makes it unsuitable for modal ratio analysis to determine the source of an AE signal. The pulser-transducer combination described in this report will be used by DECI clients in the field on large structures such as bridges and roller-coasters. The SE25-P coupled to the edge of a plate on a structure will give a repeatable crack-like AE signal to assure that ratio filtering of noise sources will still allow detection of a growing crack. NOTICE I desire to send out the DECI report as email in the future. If you would still like to receive the report, send me an email stating so. My email address is hal@deci.com. If you cannot send or receive email, fax me at 949-661-3723 with your desire to remain on the mailing list. DUNEGAN ENGINEERING COMPANY, INC. P.O. Box 1749 San Juan Capistrano, CA 92693 Telephone: (949) 661-8105 FAX: (949) 661-3723 email: hal@deci.com Homepage: www.deci.com 5