Multiport Solutions for E5071C ENA RF Network Analyzers Using External Switches. Application Note

Similar documents
Keysight Technologies Multiport Solutions for E5071C ENA RF Network Analyzers Using External Switches. Application Note

UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer

Advanced Measurement Techniques for RF Amplifiers Using Unique Functions of the Agilent E5071C ENA. Application Note

Solutions for Solar Cell and Module Testing

Agilent NFA Noise Figure Analyzer

Agilent N8480 Series Thermocouple Power Sensors. Technical Overview

Agilent Maximizing Measurement Speed Using P-Series Power Meters

Two-Way Radio Testing with Agilent U8903A Audio Analyzer

Agilent AN Balanced Circuit Measurement with an Impedance Analyzer/LCR Meter/Network Analyzer Application Note

Process Control Calibration Made Easy with Agilent U1401A

Agilent 87222C/D/E Coaxial Transfer Switches dc to 26.5, 40, 50 GHz

MEMS On-wafer Evaluation in Mass Production Testing At the Earliest Stage is the Key to Lowering Costs

Agilent MXG Signal Generators

A Time-Saving Method for Analyzing Signal Integrity in DDR Memory Buses

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Agilent U9397A/C FET Solid State Switches (SPDT)

Agilent Correlation between TDR oscilloscope and VNA generated time domain waveform

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

Agilent 87075C Multiport Test Set

Agilent N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz

Agilent J7211A/B/C Attenuation Control Units

Multipurpose Lab Station by Agilent Technologies

Agilent MXG Signal Generators

Discovering New Techniques of Creating, Editing, and Transferring Arbitrary Waveforms

Agilent 8762F Coaxial Switch 75 ohm

Agilent 87104/87106A, B, C Multiport Coaxial Switches dc to 4 GHz, dc to 20 GHz, dc to 26.5 GHz

Introduction. Part 1. Introduction...2

Agilent Technologies Noise Figure Selection Guide

Agilent 87405B. Preamplifier 10 MHz to 4 GHz. Technical Overview. Features. Benchtop/General Purpose Use

Advanced Test Equipment Rentals ATEC (2832)

Agilent 8761A/B Microwave Switches

Agilent E4438C/E8267D Option 422 Scenario Generator for GPS


Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz

Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes

Introduction. Part 1. Introduction...2


Agilent 4294A Precision Impedance Analyzer, 40 Hz to 110 MHz. Configuration Guide

Keysight Technologies Innovative Passive Intermodulation (PIM) and S-parameter Measurement Solution with the ENA. Application Note

Making a S11 and S21 Measurement Using the Agilent N9340A

Keysight E5063A ENA Series Network Analyzer

Agilent U2000 Series USB Power Sensors. Data Sheet

Keysight Technologies, Inc. UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer. Application Note

Agilent NFA Noise Figure Analyzer

Agilent Migration from the Agilent 34401A to the Agilent 34405A Digital Multimeter. Application Note

Agilent 87204/87206A, B, C Multiport Coaxial Switches dc to 4 GHz, dc to 20 GHz, dc to 26.5 GHz

Evaluating Oscilloscope Bandwidths for your Application

Techniques to Achieve Oscilloscope Bandwidths of Greater Than 16 GHz

Keysight M9485A PXIe Multiport Vector Network Analyzer

Agilent 4287A RF LCR Meter 1 MHz - 3 GHz. Technical Overview

I-V Curve Characterization in High-Power Solar Cells and Modules

Agilent InfiniiMax III probing system

Creating Arbitrary Waveforms in the U2300A Series and U2500A Series Data Acquisition Devices

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

9 khz to 4.5/6.5/8.5 GHz 100 khz to 4.5/6.5/8.5 GHz (with bias tees) 300 khz to 14/20 GHz (with bias tees)

Keysight Technologies 87104/6D Multiport Electromechanical Coaxial Switches

MIL-STD 1553 Triggering and Hardwarebased Decode (Option 553) for Agilent s InfiniiVision Series Oscilloscopes

Agilent HMMC-3124 DC-12 GHz Packaged High Efficiency Divide-by-4 Prescaler 1GC TR1-7" diameter reel/500 each 1GC BLK-bubble strip/10 each

Agilent U1700 Series Handheld LCR Meters

Solar Array Simulation System Integration

Agilent N5531S Measuring Receiver

Agilent 84904, 6, 7K/L Programmable Step Attenuators

Agilent N8700 Series System DC Power Supplies

Agilent 2-Port and 4-Port PNA-X Network Analyzer

Educator s Oscilloscope Training Kit for the InfiniiVision 2000 & 3000 X-Series

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V

Agilent 81180A Arbitrary Waveform Generator

Agilent ENA 2 and 4 Port RF Network Analyzers

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches

Keysight Technologies PXI Vector Network Analyzer Series. Drive down the size of test

Agilent 81150A Pulse Function Arbitrary Noise Generator Applications

Agilent NFA Noise Figure Analyzer

Agilent E4428C and E4438C ESG Signal Generators

Keysight Technologies RF & Microwave Attenuators. Performance you can count on

Introducing ENA Series E5061B-1x5/2x5/1x7/2x7 RF NA Options 100 khz to 1.5 / 3 GHz

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

EM Insights Series. Episode #1: QFN Package. Agilent EEsof EDA September 2008

Keysight 8762F Coaxial Switch 75 ohm

Agilent MIMO Manufacturing Solution. Application Note

Keysight Technologies Electronic Calibration (ECal) Modules for Vector Network Analyzers

MXG X-Series Signal Generator N5183B Microwave Analog

Keysight Technologies NFA Noise Figure Analyzer. Configuration Guide

Keysight Technologies Basics of RF Ampliier Measurements with the E5072A ENA Series Network Analyzer

Flexible Signal Conditioning with the Help of the Agilent 81134A Pulse Pattern Generator

Keysight Technologies USB Preamplifiers

Agilent ENA-L RF Network Analyzers

Keysight Technologies Achieving Accurate E-band Power Measurements with E8486A Waveguide Power Sensors. Application Note

Agilent U1240 Series Handheld Digital Multimeters

Essential Capabilities of EMI Receivers. Application Note

Agilent Spectrum Visualizer (ASV) Software. Data Sheet

Agilent 4-Port PNA-L Network Analyzers

Agilent U8903A Audio Analyzer

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

MXG Analog Signal Generator Express Configurations

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

Agilent EEsof EDA.

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Agilent Introduction to the Fixture Simulator Function of the ENA Series RF Network Analyzers: Network De-embedding/Embedding and Balanced Measurement

Agilent N8262A P-Series Modular Power Meter and Power Sensors. Data Sheet

Transcription:

Multiport Solutions for E5071C ENA RF Network Analyzers Using External Switches Application Note

Introduction The evolution of components and modules is unstoppable because of increasing requirements from the market. In the wireless communication industry, designers and manufacturers of cellular handsets desire smaller and less expensive components with multi-band capability in order to meet the increasing demands from their end users. More and more functionalities are integrated onto a single component, which introduces multiport devices with combined features and thus, overall multiport characterization becomes increasingly important than in the past. While there are an increasing number of measurement parameters for multiport devices, total testing time is aimed to be minimized by high volume manufacturers. For network analysis, multiport test sets designed to work with the network analyzers are benefi cial especially for reducing test times, as one way to boost production throughput is to reduce the time required for connecting devices under test (DUTs). For devices connected to power amplifi ers in a cellular handset, non-linearity characterization such as harmonic distortion measurements with a high-power signal applied to the DUT is required, as well as S-parameter measurements that are typically performed under small signal conditions. Because of the power handling capability, it is recommended to combine a multiport test set with external electromechanical (EM) es in order to achieve faster test times in a cost-effective manner. This application note will briefl y explain how to expand the capability of multiport solutions on the E5071C network analyzer, and provide insights for handling high-power and low-power signal measurements with a single connection. 2

Multiport Solution For Network Analysis One example of a multiport solution for the network analysis is shown in the Figure 1. Figure 1 describes the internal block diagram of the Agilent E5091A test set, a ing multiport test set that is designed to work with the ENA network analyzer. Signals from the 4-port ENA are diverted via internal es in the test set for characterization of multiport devices. This test set can be controlled with the fi rmware on the ENA, and high-speed measurements are achievable with communication via a USB interface between the two instruments. The E5091A offers two different options: option 009 (9-port confi guration in left) and option 016 (13- or 16-port confi guration in right). 1 2 3 4 1 2 3 4 SW1 SW3 SW4 SW5 SW1 SW4 SW7 SW10 SW2 SW6 SW7 SW2 SW3 SW5 SW6 SW8 SW9 SW11 SW12 ANT T1 T2 R1 + R2 + R3 + R1 R2 R3 SW13 SW14 SW15 A1 A2 A3 A4 X1 X2 Y1 Y2 Z1 Z2 T1 T2 T3 B1 B2 B3 B4 R1 + R2 + R3 + R4 + R1 R2 R3 R4 ANT Semi-rigid cable T4 Figure 1. Block diagram of the E5091A multiport test set (Option 009 left, Option 016 right) The E5091A utilizes an architecture based on multiple SPDT (Single-Pole-Double-Throw; one input and two outputs) es. By turning on and off these internal SPDT es with the softkey of the ENA, the RF signal from the ENA can be routed to each port on multiport devices under test (DUT). Since the solid state es are installed inside the test set, ing time is minimized for fast measurements compared to es with different technologies. Also the timing of ing in the test set is highly optimized by considering the wait time of frequency sweep of the ENA. As a result, the additional time in the total measurement sequence is dramatically reduced. This is defi nitely valuable for highvolume manufacturers with automated test equipment (ATE) systems, because measurement speed is the most signifi cant parameter and high throughput in measurements has a huge impact on the total cost of test. Though the measurement speed is highly optimized with the E5091A, there are trade-offs for some applications with the test set. For example, the solid state ing solution cannot handle high-power measurements because of its specifi cation on the damage level. The internal solid state es limit the maximum damage level of the test set at approximately +20 dbm, which is not suffi cient for characterization of components in GSM handsets. Amplifi ers for this application are usually operated in the non-linear region for effi ciency and high-power measurements with more than +30 dbm (= 1 Watt) are required in some cases. 3

To avoid the electrical stress from outside conditions and hardware failure on internal components in the test set, engineers today can use external EM es to extend the capability of the E5091A. The application example is discussed in detail in the following section, highlighting the key benefi ts of this fl exible confi guration, as well as providing details of selection criteria. EM es can be used to expand the functionality of the E5091A because these es have superior distortion characteristics than solid-state es. Agilent offers high-performance coaxial es for a wide range of required applications, providing the long life cycle, guaranteed 0.03 db insertion loss repeatability, and reliability required for achieving higher performance in the majority of ing applications. Measurement system performance can be optimized by using various es for specifi c applications. In the following measurement application example, we are recommending a combination of SPDT and single-pole-four-throw (SP4T) multiport EM es. Agilent provides a broad selection of es which provides confi guration fl exibility for various application and needs. The Agilent N1810TL is recommended for the SPDT, which routes signals from one input to two output paths. It provides superior insertion loss of 0.03 db, as well as an exceptional isolation of >60 db. As for a SP4T, which allows a single input and four output paths, users may opt for the Agilent 87104A or L7104A. For high-volume manufacturing, the 87104A is preferred due to its guaranteed 0.03 db repeatability, which ensures accuracy and reduces calibration cycles for the entire 5 million cycle operating life (10 million cycles typical). For economic considerations, Agilent offers the L7104A, which is a high-performance SP4T at an affordable price. It has guaranteed 0.03 db repeatability, which ensures accuracy and reduces calibration cycles for the entire 2 million cycles operating life (5 million cycles typical). In the following section, a procedure of setting up the measurement will be discussed. The details of each component in the setup will then be explained to enhance your understanding of how to expand the potential of the Agilent E5071C ENA RF Network Analyzer. 4

Driving The Electro-Mechanical Switches Agilent 11713 series attenuator/ driver In order to drive the electromechanical es, a driver is needed to supply the necessary DC voltage. The ing set up should be completed in advance of a measurement sweep by the network analyzer so that accurate and stable measurement results can be achieved. The Agilent 11713C attenuator/ driver is capable of driving as many as 20 SPDT es. The supported es can range from SPDT to single-pole-six-throw (SP6T) multiport es. The 11713C comes with a tri-voltage selection of 5V, 15V and 24V and also allows for a user-defi ned voltage supply. The confi guration of the ENA with the 11713B/C is described in Figure 2. The USB/GPIB interface (i.e. Agilent 82357B) is connected between the two instruments for remote control operation from the ENA. The interface is connected to the USB connector on the ENA and the GPIB connector on the 11713B/C. With this setup, the ENA can be utilized as a system controller, and is able to control external es via the 11713B/C. The 11713B/C provides the ease of point-to-point connection from the driver to the es by providing nine cable options, enabling easy connection between the instrument and the es. For the Agilent N1810TL SPDT, which is recommended in this solution, cable option 201 (Viking connector to 12-pin bare wire conductor cables) and option 501 (Viking connector to 4 9-pin DSUB connectors) can be used. On the other hand, the 87104A or L7104A can be connected to the driver by using cable option 201 and option 601 (Viking connector to 16-pin DIP connector). More information can be obtained from the 11713B/C Confi guration Guide, Agilent literature number 5989-7277EN. USB Interface USB/GPIB Interface (82357B) Cable for SPDT: 11713B-201 (bare wire) or 11713B-501 (DSUB) Cable for SP4T: 11713B-201 (bare wire) or 11713B-501 (16-pin DIP) Figure 2. Connection of the ENA and the 11713B/C 5

Note that the E5091A multiport test set has its own internal driving circuit which can supply the DC voltage through the D-sub 15-pin connector on the front panel (Figure 3). This circuit was originally designed for enabling control bits of DUTs such as antenna modules in cellular handsets, but it is possible to drive other devices within the hardware specifi cation of the test set. The output DC voltage can range from 0 to +5 V for positive signals and 5 V to 0 V for negative signals. The maximum supply current is limited below 50 ma. Figure 3. D-sub 15-pin connector for control voltage on the E5091A Software Connectivity For multiport measurements, there are many measurement parameters that need to be input into the test instruments. It can be very time consuming and sometimes the preparation of testing, as well as the calibration procedure, takes longer than the actual measurement time. Thus, the setup software becomes critical for reducing the time on multiport measurements. The ENA has a Measurement Wizard Assistant (MWA) software option (E5071C-790) that is defi nitely benefi cial to all operators of multiport measurements. Figure 4 indicates the overall concept of the MWA. The software consists of two main applications -- front-end and back-end. The front-end application running on Microsoft Excel generates a setup fi le that includes all parameters for network analysis such as power level, frequency range, IFBW or number of points. Once the necessary parameters of the front-end application have been input with the Excel-based user interface, they no longer have to input by softkey on the ENA. That minimizes fatigue in operation time, and testing productivity is greatly improved. The generated setup fi le (.mwa) is recalled by the back-end application of the Microsoft VBA software on the ENA, and all the included parameters are set up automatically within the ENA. The back-end application also has the calibration wizard function that offers the calibration procedure when using the multiport test set. This wizard minimizes the number of connections required for the calibration procedure, and eliminates the possibility of connecting wrong ports during calibration. 6

Front-End Application Back-End Application Operation environment Software based on the Microsoft Excel Can be operated on any PC with Microsoft Office 2000 installed Operation environment Software based on the Microsoft VBA Macro working in the ENA Option E5071C-790 for the ENA is required for full capability operation Setup file (.mwa) Features and benefits Easy and fast setup of multiport measurements for the ENA Generates one setup file (.mwa) including all the measurement parameters Step-by-step setup wizard with Excel-based user interface Features and benefits Recall the.mwa setup file and set all the parameters on the ENA automatically Calibration wizard for minimum operation time Automatic test procedure with go/no go limit test Figure 4. MWA concept Figure 5 shows an example of the GPIB sheet in the front-end application, which can be utilized for a remote control between the ENA and external instruments in multiport measurements. Figure 5. GPIB sheet of the MWA for controlling es via driver As previously mentioned, when the ENA is connected to peripherals via the USB/GPIB interface, the ENA can be operated as a system controller in the test system and it is possible to transmit GPIB commands to peripherals with a selected GPIB address. By sending commands to turn on or off the es, a desired port is selected via the 11713C driver within the test procedure that is defi ned by the front-end application of the MWA. 7

Measurement Example In this section, an example of a multiport measurement is introduced. The DUT is an RF multiport typically used in a cellular handset, and it is required to perform both S-parameter measurements on all paths and a harmonic distortion measurement on the transmission path of the DUT. To perform both of these measurements with a single connection, it is benefi cial to use external EM es. The basic confi guration examples of the measurements are shown in Figures 6 and 7. Because both confi gurations are focused on 2-port measurements out of multiple ports on the device, additional EM es may be required for a complete measurement setup. Figure 6 is for the S-parameter measurement with a low-power signal applied to the DUT, while Figure 7 is for the harmonic distortion measurement with a high-power signal to the DUT. Both measurements can be selected by changing the EM ports. USB/GPIB Interface (82357B) SPDT SP4T HPF Booster amplifier Isolator LPF DUT ATT Calibration plane Figure 6. Configuration example of low-power measurement (S-parameter measurement) USB/GPIB Interface (82357B) SPDT SP4T HPF Booster amplifier Isolator LPF DUT ATT Calibration plane Figure 7. Configuration example of high-power measurement (harmonics measurement) 8

Because the DUT is normally connected to power amplifi ers for transmission operation within a cellular handset, distortion measurements of the device are normally performed under the condition of high input power level. An external booster amplifi er is necessary to obtain the high power level that is well beyond the maximum output power level of the network analyzer. An isolator is inserted after the booster amplifi er to improve mismatch due to poor S22 characteristics of the amplifi er. For harmonic performance measurements, a low pass fi lter (LPF) needs to be placed to remove unwanted harmonic products generated by the amplifi er or the signal source of the network analyzer. Also a high pass fi lter (HPF) is necessary for suppression of the fundamental frequency on the receiver path in order to maximize the Signal-to-Noise (S/N) ratio of absolute measurements using the receivers of the network analyzer. Agilent EM es are suitable for this application as they provide guaranteed 0.03 db insertion loss repeatability, which ensures accuracy and reduces calibration cycles. The exceptional isolation performance of > 60 db is also a key contributor that maximizes measurement accuracy and improves system dynamic range. Harmonic Measurement Procedure In this section, the harmonic distortion measurement procedure with the ENA will be described in detail. The signal from port 1 of the ENA is applied to the 2-port DUT, and the output of the DUT is routed to port 2 of the ENA for an absolute measurement at the fundamental frequency, and to port 3 for the measurement at harmonic frequencies. The E5091A multiport test set can be useful when ports on the DUT are increased. The necessary steps consist of the following: (1) Parameters setup (2) Receiver calibration (3) Source power calibration (4) Measurement (1) Parameters setup (1-1) Measurement parameters setup All the necessary measurement parameters such as frequency, number of points, and IFBW should be selected appropriately. For this measurement example, measurement frequency ranges from 800 MHz to 1 GHz for a fundamental signal from the source port of the ENA. The absolute measurement at the fundamental frequency is performed on channel 1 of the ENA, while measurement of the 2nd harmonic stimulus is done on channel 2. Set the start and stop frequency [Start] > [Start] (800 MHz) [Stop] > [Stop] (1 GHz) Allocate two channels in the display for absolute measurements [Display] > [Allocate Channels] > [x2] [Meas] > [Absolute] > [B (1)] (for channel 1) [Meas] > [Absolute] > [C (1)] (for channel 2) (1-2) Turn on Frequency Offset Mode (FOM) By turning on the FOM, the source and receiver of the ENA can be tuned at different frequencies; harmonic stimulus generated in the DUT can be obtained at the receivers of the ENA. Option E5071C-008 for the ENA or upgrade product E5003A is required for enabling this functionality. Turn on the FOM for channels 1 and 2 [Sweep Setup] > [Frequency Offset] > [Frequency Offset] (On) Tune at the harmonic frequency on the source (port 1) and receiver (port 3) for channel 2 [Sweep Setup] > [Frequency Offset] > [Port 1] > [Multiplier] (2) > [Port 3] > [Multiplier] (2) 9

The frequency range of channel 1 is set at 800 MHz to 1 GHz for the source (port 1) and receiver port (port 2), and the range for channel 2 is from 1.6 GHz to 2 GHz for the both ports (port 1 and port 3). An integer of the multiplier should be selected depending on the harmonic stimulus of interest. (2) Receiver calibration The absolute power level at the receiver ports on the ENA can be calibrated with the receiver calibration. The effect of insertion loss on the receiver path is eliminated with the calibration, and values of absolute power level at the calibration plane are displayed on the screen of the ENA. Receiver calibration is done at the frequency range of a selected source port. Thus, for the harmonic measurement, the frequency range of the source port should be the same range as the receiver port. It is recommended to perform the power calibration in advance of the receiver calibration in order to achieve a more accurate power level on the input signal. (2-1) Power calibration Set the power level of the ENA on channels 1 and 2. [Sweep Setup] > [Power] > [Power] (dbm) Note: The input power level for the receiver calibration is limited under the maximum output power level of the ENA. Perform the power calibration for channels 1 and 2. The ENA and power meter are connected via USB/GPIB interfaces such as the Agilent 82357B. (Figure 8) [Cal] > [Power Calibration] > [Take Cal Sweep] USB/GPIB Interface (82357B) Calibration plane SP4T HPF ATT Power sensor Power meter Figure 8. Power calibration (for receiver calibration) 10

(2-2) Receiver calibration Perform receiver calibration after connecting the source and receiver paths. (Figure 9) For channel 1: [Cal] > [Receiver Calibration] > [Select Port] (2) > [Source Port] (1) > [Take Cal Sweep] For channel 2: [Cal] > [Receiver Calibration] > [Select Port] (3) > [Source Port] (1) > [Take Cal Sweep] SP4T HPF ATT Calibration plane Figure 9. Receiver calibration (3) Source power calibration (3-1) Lower the power level of the ENA For high-power measurements with a booster amplifi er, the output power level from the source port on the ENA must be set to a proper value so that the input signal to the receiver ports will not cause a hardware failure in the ENA. Note that maximum damage level of the test ports on the ENA is typically +26 dbm, and +20 dbm for the ports on the E5091A. For channels 1 and 2: [Sweep Setup] > [Power] > [Power] (dbm) (3-2) Set the frequency of the source port Set the source port (port 1) of channel 2 at the fundamental frequency range. For channel 2: [Sweep Setup] > [Frequency Offset] > [Port 1] > [Multiplier] (1) Now the frequency of the source port (port 1) in both channels ranges from 800 MHz to 1 GHz for fundamental signals. The frequency at port 2 is tuned for the fundamental measurement (800 MHz to 1 GHz), while the frequency at port 3 is tuned for the harmonic measurement (1.6 GHz to 2 GHz). 11

(3-3) Power calibration at high-power level The confi guration for high-power calibration with an external amplifi er is illustrated in Figure 10. The power level of the input of the DUT is adjusted to a desired value by the power meter and sensor connected to the ENA. USB/GPIB Interface (82357B) SPDT SP4T HPF Booster amplifier Isolator LPF ATT Calibration plane Power sensor Power meter Figure 10. Power calibration with high-power signal Insertion loss or gain on the source path can be compensated by inputting appropriate values in the loss compensation table (Figure 11). At high-power measurements, for example, when the output power level of the ENA is set at 0 dbm and a negative value such as 32 db is added in the compensation table, the power level at the calibration plane of the power sensor is adjusted to +32 dbm for the entire frequency range. Figure 11. Loss compensation table for power calibration For channels 1 and 2: [Cal] > [Power Calibration] > [Compen] > [Compensation] (On) > [Take Cal Sweep] ( Take Cal Sweep key is located under Power Calibration ) 12

(4) Measurement The measurement result is shown in Figure 12. The frequency dependency of the fundamental signal of the output of the DUT is displayed on channel 1, and the 2nd harmonic stimulus of the DUT is displayed on the channel 2. Figure 12. Results of harmonic distortion measurement Summary This application note explains how the potential of a multiport solution for network analysis can be expanded using the E5071C ENA network analyzer with external electromechanical es for handling high-power and low-power signal measurements. Appropriate es should be selected depending on the system individual performance needs, given the wide portfolio of ing solutions from Agilent es with the 11713B/C driver. In today s manufacturing environment, reducing cost of test and optimzing manufacturing yields are of primary importance. Agilent has the combined portfolio of test instruments and premium grade RF and microwave test accessories to address the critical demands of today s test and measurement world. 13

Recommended Instruments and Accessories E5071C ENA RF Network Analyzer, 9 khz to 8.5 GHz E5091A Multiport Test Set 87104A/B Multiport Coaxial Switch, DC to 4/20 GHz, SP4T L7104A/B Multiport Coaxial Switch, DC to 4/20 GHz, SP4T N1810TL Coaxial Switch, DC to 2/4/20/26.5 GHz, SPDT References ENA series web page: www.agilent.com/find/ena Multiport Test Sets: www.agilent.com/find/multiport Switch Drivers: www.agilent.com/find/drivers RF & Microwave es: www.agilent.com/find/es Related Literature E5071C ENA RF Network Analyzers, Data Sheet (5989-5479EN) Agilent RF & Microwave Switches, Performance you can count on, Brochure (5989-6947EN) Agilent 11713B/C Attenuator/Switch Drivers, Confi guration Guide (5989-7277EN) Agilent L Series Multiport Electromechanical Coaxial Switches, Technical Overview (5989-6030EN) 87104/87106A,B,C Multiport Coaxial Switches, Product Overview (5091-3366E) N1810/1/ 2 Coaxial Switch, Product Overview (5968-9653E) 14

Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence. Agilent Open www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development. www.lxistandard.org LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium. Remove all doubt Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance, onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to: www.agilent.com/find/removealldoubt www.agilent.com For more information on Agilent Technologies products, applications or services, please contact your local Agilent offi ce. The complete list is available at: www.agilent.com/find/contactus Americas Canada (877) 894-4414 Latin America 305 269 7500 United States (800) 829-4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Thailand 1 800 226 008 Europe & Middle East Austria 0820 87 44 11 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 /minute Germany 01805 24 6333** **0.14 /minute Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland 0800 80 53 53 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: Feb 15, 2008 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2008 Printed in USA, March 10, 2008 5989-7916EN