INSTRUCTIONS MODELS AVX-CA-SEMA TEST JIG FOR USE WITH AVTECH AVR-EB2A-B REVERSE RECOVERY TEST SYSTEMS SERIAL NUMBER:

Similar documents
INSTRUCTIONS MODELS AVX-TRR-SQMELF TEST JIG FOR USE WITH AVTECH AVR-EB4-B REVERSE RECOVERY TEST SYSTEMS SERIAL NUMBER:

INSTRUCTIONS MODELS AVX-TFR-MELF TEST JIG FOR USE WITH AVTECH AVR-EBF6-B FORWARD RECOVERY TEST SYSTEMS SERIAL NUMBER:

INSTRUCTIONS MODEL AVX-TFR-STUD TEST JIG FOR USE WITH AVTECH AVR-EBF6-B FORWARD RECOVERY TEST SYSTEMS SERIAL NUMBER:

INSTRUCTIONS MODEL AVC MONOCYCLE GENERATOR MODULE SERIAL NUMBER:

INSTRUCTIONS MODEL AVX-S1-P2-EX1 PLUG-IN SOCKET OUTPUT MODULE SERIAL NUMBER:

INSTRUCTIONS AVX-FILT SERIES OF RISE TIME FILTERS FOR USE WITH AVR-EBF6-B TFR TEST SYSTEMS SERIAL NUMBER:

INSTRUCTIONS MODEL AVX-S1-P0D-MS1A PLUG-IN SOCKET OUTPUT MODULE SERIAL NUMBER:

AVTECH TECHNICAL BRIEF 15 (TB15) A COMPARISON OF REVERSE RECOVERY MEASUREMENT SYSTEMS

INSTRUCTIONS MODEL AV-1000-C ±10 VOLT / TTL / ECL GENERAL PURPOSE PULSE GENERATOR SERIAL NUMBER:

TETRIS 1000 High Impedance Active Probe. Instruction Manual

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

SCXI MHz Quad 8x1 50 Ω Multiplexer

TETRIS User's Guide. High Impedance Active Probe DO177-1

INSTRUCTIONS MODEL AVM-1-C. 5 VOLT, 25 MHz HIGH SPEED PULSE GENERATOR. WITH 100 ps RISE TIME SERIAL NUMBER:

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

PHV RO High Voltage Passive Probe. Instruction Manual

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

NI PXI/PXIe-2540 Specifications

PHV 1000-RO. High impedance passive probe. Features: CeramCore TM hybrid probe. Modular construction. Coaxial design

NI PXI-2520 Specifications 80-Channel SPST Relay Module

NI PXI-2521 Specifications 40-Channel DPST Relay Module

PML 791-RO. High impedance passive probe. Features: 2.5 mm Diameter Tip. Coaxial Design. Interchangeable Spring Contact Tip

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

CURRENT MEARUREMENT SHUNT CS-10/500 USER MANUAL

PHV RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

User s Guide. RP7000 Series Active Probe. Dec RIGOL Technologies, Inc.

NI PXIe-5601 Specifications

PKT 512A-RO High Impedance Passive Cable Divider

BROADBAND LINEAR AMPLIFIER Model F1020 (models F10A and F20A combined)

BROADBAND LINEAR AMPLIFIER Model P150

NI PXI/PXIe-2532 Specifications

Appendix A: Specifications

NI PXI/PXIe-2527 Specifications

NI PXI-2530 Specifications

Operator s Manual. PP016 Passive Probe

Instruction Manual Model Upconverter

NI PXI/PXIe-2529 Specifications

NI PXI-2557 Specifications

NI PXI-2555 Specifications

BROADBAND'LINEAR'AMPLIFIER WITH'A±B'INPUTS'AND'DC5OFFSET Model&A400X

PCX-7401 Pulsed Current Source Operation Manual

Instruction Manual Model Upconverter

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

Operator s Manual. PP022 Passive Probe

Baro Transmitter PTB x1.073

Supplied in carry case with additional accessories The PHV 1000-RO is a 400 MHz, standard sized, 100:1 passive probe designed for instruments

PMT 211A-RO. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

Charge Recovery Measurements with TLP

Programmable K-Factor Scaler B and Programming Software Kit B

NI SCXI Specifications

BROADBAND'LINEAR'AMPLIFIER Model&F10AD

As all PMK probes the PML 751-RO features CeramCore TM technology. The entire probe

PML 711A-RO High impedance passive probe Features:

400 MHz Passive High-Voltage Probe R&S RT-ZH

Kentech Instruments Ltd. CPS3/S pulser Serial No. J03*****

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

Operator s Manual. PP017 and PP018 Passive Probes

TK 100 Probe Calibrator. Instruction Manual

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95

K-Factor Scaler F5140 and Programming Kit F5141 Installation & Operating Instructions

PMT 221A. High impedance passive probe. Features: CeramCore TM Hybrid Probe. Modular Construction. Coaxial Design. Interchangeable Spring Contact Tip

DUAL%CHANNEL%LINEAR%AMPLIFIER %WITH%PHASE%INVERTER Model&A400DI

RTD), LM335 Voltage, or AD592 current. c. Constant TE Current (8350 only)

CIRCUIT-TEST ELECTRONICS

Microwave. Accessories for Microwave Scalar and System Analyzers

Model 9305 Fast Preamplifier Operating and Service Manual

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 40 A Fault Insertion Unit

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 10 A Fault Insertion Unit

Model 113 Scintillation Preamplifier Operating and Service Manual

AMP-13 OPERATOR S MANUAL

AMERITRON RCS-12 AUTOMATIC ANTENNA SWITCH

Glass Electrode Meter

Part No. Data Rate Media type Distance

High Efficiency AC Input 8A 19V Laser Driver

High Efficiency AC Input 12A 12V Laser Driver

Operation and Service Manual. 350 MHz Preamplifier SIM914. Stanford Research Systems

GT-1050A 2 GHz to 50 GHz Microwave Power Amplifier

PXI Contents. PXI-2585 Specifications. Topology DEVICE SPECIFICATIONS. 10-Channel Multiplexer

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

INSTRUCTIONS MODEL AVOZ-B3-B. 100 AMP, 100 V, 35 ns RISE TIME LASER DIODE DRIVER WITH IEEE AND RS-232 CONTROL SERIAL NUMBER:

MHz FUNCTION GENERATOR INSTRUCTION MANUAL

PXIe Contents. Required Software CALIBRATION PROCEDURE

NI PXI-2532 Specifications

PXIe Contents SPECIFICATIONS. RF Preselector Module

NI PXIe-5630 Specifications

ETSI EN V1.2.1 ( )

HO118 Binning Interface User Manual * * Version 02. User Manual. Test & Measurement

PXIe Contents SPECIFICATIONS. RF Amplifier

Model BPG-1P Biphasic Pulse Generator

z47524 PXI Quad SPDT RF Changeover Switch Module

Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

PI-10 Broadband Power Indicator

SNR-SFP100-T. SNR-SFP100-T 10/100M Copper SFP Series

PULSE DISTRIBUTION AMPLIFIER OPERATING MANUAL

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

SIMATIC RTLS. Localization systems SIMATIC RTLS PCB OEM AX (6GT2700-8DF00-0AX0) Device manual 10/2018 C79000-G8976-C518-01

SCC-FV01 Frequency Input Module

NI PXIe-5663 Specifications

Transcription:

AV T E C H ELECTROSYSTEMS N A N O S E C O N D P.O. BOX 265 OGDENSBURG, NY U.S.A. 13669-0265 W AVE F O R M E L E CT RO NI C S SI N C E 1 9 7 5 TEL: 888-670-8729 (USA & Canada) or +1-613-686-6675 (Intl) FAX: 800-561-1970 (USA & Canada) or +1-613-686-6679 (Intl) info@avtechpulse.com - LTD. http://www.avtechpulse.com/ INSTRUCTIONS MODELS AVX-CA-SEMA TEST JIG FOR USE WITH AVTECH AVR-EB2A-B REVERSE RECOVERY TEST SYSTEMS SERIAL NUMBER: X BOX 5120, LCD MERIVALE OTTAWA, ONTARIO CANADA K2C 3H5

2 WARRANTY Avtech Electrosystems Ltd. warrants products of its manufacture to be free from defects in material and workmanship under conditions of normal use. If, within one year after delivery to the original owner, and after prepaid return by the original owner, this Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings. This warranty is the extent of the obligation assumed by Avtech with respect to this product and no other warranty or guarantee is either expressed or implied. TECHNICAL SUPPORT Phone: 888-670-8729 (USA & Canada) or +1-613-686-6675 (International) Fax: 800-561-1970 (USA & Canada) or +1-613-686-6679 (International) E-mail: info@avtechpulse.com World Wide Web: http://www.avtechpulse.com

3 TABLE OF CONTENTS WARRANTY...2 TECHNICAL SUPPORT...2 TABLE OF CONTENTS...3 INTRODUCTION...4 REGULATORY NOTES...5 FCC PART 18... 5 EC DECLARATION OF CONFORMITY...5 DIRECTIVE 2002/95/EC (RoHS)...6 DIRECTIVE 2002/96/EC (WEEE)...6 BASIC AMPLITUDE CONTROL...7 SETTING THE AMPLITUDE LEVELS...8 AMPLITUDE ACCURACY... 8 CURRENT MONITOR... 9 INCORRECT ORIENTATION...9 TEST JIG MECHANICAL ASPECTS...10 SAFETY INTERLOCK...12 OVERRIDE... 12 WIRING DIAGRAM...13 PERFORMANCE CHECKSHEET...14 Manual Reference: /fileserver2/officefiles/instructword/avx-ca/avx-ca-sema,ed1.odt. Last modified February 12, 2016. Copyright 2016 Avtech Electrosystems Ltd, All Rights Reserved.

4 INTRODUCTION The AVX-CA-SEMA test jig is designed for use with the AVR-EB2A-B series of reverse recovery time test systems. Two customer-specified Loranger sockets are installed on the jig. The standard Loranger 03186 021 6215 socket will accept DO-41 and similar axial packages. The special-order Loranger 03453 041 X215 socket, provided by the customer, will accept certain SQMELF packages. This test jig can be used as a replacement or an alternative for the test jigs originally supplied with the Avtech AVR-EB2A-B units.

5 REGULATORY NOTES FCC PART 18 This device complies with part 18 of the FCC rules for non-consumer industrial, scientific and medical (ISM) equipment. This instrument is enclosed in a rugged metal chassis and uses a filtered power entry module. The main output signal is provided on a shielded connector that is intended to be used with shielded coaxial cabling and a shielded load. Under these conditions, the interference potential of this instrument is low. If interference is observed, check that appropriate well-shielded cabling is used on the output connectors. Contact Avtech (info@avtechpulse.com) for advice if you are unsure of the most appropriate cabling. Also, check that your load is adequately shielded. It may be necessary to enclose the load in a metal enclosure. If any of the connectors on the instrument are unused, they should be covered with shielded metal dust caps to reduce the interference potential. This instrument does not normally require regular maintenance to minimize interference potential. However, if loose hardware or connectors are noted, they should be tightened. Contact Avtech (info@avtechpulse.com) if you require assistance. EC DECLARATION OF CONFORMITY We Avtech Electrosystems Ltd. P.O. Box 5120, LCD Merivale Ottawa, Ontario Canada K2C 3H5 declare that this pulse generator meets the intent of Directive 2004/108/EG for Electromagnetic Compatibility. Compliance pertains to the following specifications as listed in the official Journal of the European Communities: EN 50081-1 Emission EN 50082-1 Immunity

6 and that this pulse generator meets the intent of the Low Voltage Directive 72/23/EEC as amended by 93/68/EEC. Compliance pertains to the following specifications as listed in the official Journal of the European Communities: EN 61010-1:2001 Safety requirements for electrical equipment for measurement, control, and laboratory use DIRECTIVE 2002/95/EC (RoHS) This instrument is exempt from Directive 2002/95/EC of the European Parliament and of the Council of 27 January 2003 on the Restriction of the use of certain Hazardous Substances (RoHS) in electrical and electronic equipment. Specifically, Avtech instruments are considered "Monitoring and control instruments" (Category 9) as defined in Annex 1A of Directive 2002/96/EC. The Directive 2002/95/EC only applies to Directive 2002/96/EC categories 1-7 and 10, as stated in the "Article 2 - Scope" section of Directive 2002/95/EC. DIRECTIVE 2002/96/EC (WEEE) European customers who have purchased this equipment directly from Avtech will have completed a WEEE Responsibility Agreement form, accepting responsibility for WEEE compliance (as mandated in Directive 2002/96/EC of the European Union and local laws) on behalf of the customer, as provided for under Article 9 of Directive 2002/96/EC. Customers who have purchased Avtech equipment through local representatives should consult with the representative to determine who has responsibility for WEEE compliance. Normally, such responsibilities with lie with the representative, unless other arrangements (under Article 9) have been made. Requirements for WEEE compliance may include registration of products with local governments, reporting of recycling activities to local governments, and financing of recycling activities.

7 BASIC AMPLITUDE CONTROL The AVX-CA-SEMA is intended for use with an Avtech AVR-EB2A-B reverse recovery test system, ordered separately. Please refer to the manual supplied with the AVREB2A-B instrument for detailed usage instructions. The basic test arrangement for the AVR-EB2A-B is shown in the figure below. The OUT output on the instrument mainframe is connected to the IN input on the test jig using the coaxial cable supplied with the AVR-EB2A-B, and the control cable is connected using the supplied DB-9 cable. CURRENT THROUGH DUT (IDUT) VOLTAGE PULSE 0V 60 cm SMA-to-SMA COAXIAL CABLE IF, +100 ma MAX. VPULSE, -24V MAXIMUM 200 ns AVTECH AVR-EB2A-B MAINFRAME VOUT = IDUT 50Ω RPULSE = 50Ω DUT IN Other signals... CONTROL OSCILLOSCOPE (BW > 1 GHz) CH A 60 cm SMA-to-SMA COAXIAL CABLE RSCOPE = 50Ω CONTROL AVX-CA-SEMA TEST JIG RDC = 950Ω Up to VDC +115V IR IF + VPULSE / 100 Ω OUT IDUT Other signals... IR, -100 ma MAX. IF (VDC VDIODE) / 1 kω VSCOPE = IDUT 50Ω / 10 OUT VPULSE, up to -24V 0 ma REVERSE RECOVERY TIME (trr) 2 meter DB-9 CABLE EXT TRIG 20 db ATTENUATOR (USE SUPPLIED SMA-BNC ADAPTER, IF REQUIRED) SYNC The OUT output on the test jig must be connected (using the 60 cm SMA-to-SMA cable and 20 db attenuator supplied with the AVR-EB2A-B) to the 50 Ohm input of a highbandwidth (> 1 GHz) user-supplied oscilloscope. The voltage on the output is proportional to the current through the DUT. By observing the current waveform through

8 the DUT, the reverse recovery time may be determined. When connected as shown above, the oscilloscope voltage waveform is related to the device current by: VSCOPE = IDUT 50Ω / 10 This is the key equation for relating the observed voltage waveform to the DUT current. SETTING THE AMPLITUDE LEVELS The test jig contains a bias tee circuit to combine a DC forward bias with a reverse bias pulse. An adjustable DC voltage of up to +115V is generated in the AVR-EB2A-B mainframe. This DC voltage is controlled by the front-panel OS menu, or by the source:voltage:low command. The DC voltage source has R DC = 950 Ω in series with it, inside the mainframe. The DC forward bias current through the DUT is then given by: IF = (VDC VDIODE) / (RDC + RSCOPE) IF = (VDC VDIODE) / 1000 Ω where VDIODE is the diode forward voltage drop (typically 0.7V). The reverse bias pulse is capacitively coupled on top of this DC bias current. The pulse generator has an adjustable amplitude of up to -24V, controlled by the front-panel AMP menu, or by the source:voltage command. The pulse generator has 50 Ω in series with it, inside the mainframe, as well as the 50 Ω of the oscilloscope, so the size of the current change will be VPULSE / (50 Ω + 50 Ω). Thus, the reverse current will given by: IR IF + VPULSE / 100 Ω In this equation, IF is positive, VPULSE is negative, and VPULSE must be set sufficiently high such that IR is negative. This equation is only approximate, because the diode itself may introduce some poorly-defined parasitic resistance during the switching transient. This basic arrangement of DC and pulse sources is specified by MIL-STD-750E Method 4031.4 Test Condition A. AMPLITUDE ACCURACY The amplitude settings should not be relied upon for any degree of accuracy, because the diode characteristics (forward voltage drop, reverse parasitic resistance) are not known accurately in advance. Instead the voltage at the OUT terminal on the test jig should be measured with a calibrated oscilloscope.

9 CURRENT MONITOR The AVR-EB2A-B monitors the actual forward bias current through the DUT. If the DUT has been incorrectly seated, and the forward bias current can not flow properly, the instrument will detect this problem, display an error message, and turn off the output. If this occurs, check the seating of the DUT and close both lids again. The monitored forward current level is displayed on the front panel LCD, for the user's convenience. The accuracy is ± 2% ± 1 ma. INCORRECT ORIENTATION The instrument and the DUT will not be damaged if the diode is installed with the incorrect orientation (i.e., with the anode and cathode reversed). However, incorrect waveforms will be generated.

10 TEST JIG MECHANICAL ASPECTS The AVX-CA-SEMA test jig will accept a variety of axial and SQMELF packages. To install a DUT, open the main (black) lid like this: Install a diode in one of the two sockets. An example of an axial DO-41 diode installed is shown below: An example of a SQMELF package installed is shown below:

11 The two sockets are wired in parallel. Only one DUT may be installed at a time, or the resulting waveforms will be incorrect. Once the DUT is installed in the appropriate socket, close the main lid. It must be closed for proper operation. The AVR-EB2A-B monitors the actual forward bias current through the DUT. If the DUT has been incorrectly seated, and the forward bias current can not flow properly, the instrument will detect this problem, display an error message, and turn off the output. If this occurs, check the seating of the DUT and close both lids again. The instrument and the DUT will not be damaged if the diode is installed with the incorrect orientation (i.e., with the anode and cathode reversed). However, incorrect waveforms will be generated. The IN, OUT, and CONTROL connectors are on the rear of the jig, below the hinges:

12 SAFETY INTERLOCK The AVX-CA-SEMA contains safety interlock devices, for use with the AVR-EB2A-B mainframe. The DB-9 female CONTROL connector should be connected to the corresponding connector on the AVR-EB2A-B mainframe using the straight-through DB-9 cable supplied with the AVR-EB2A-B. The DB9 CONTROL connector pinout is as follows: Pin 1 - To test jig switch 1. Pin 2 - To test jig switch 2. Pin 5 - Ground. Pin 6 - To test jig switch 1. Pin 7 - To test jig switch 2. Pin 9 - Safety sensor power supply (+15V through 680 Ohms). When the test jig lid is safely closed, Pin 1 is shorted to Pin 6, and Pin 2 is shorted to Pin 7. Switch 1 is a passive mechanical switch. Switch 2 is an active photosensor, which requires power from Pin 9. OVERRIDE If the user has implemented other safety steps that make the interlock circuit unnecessary, it is possible to disable the interlock by installing a jumper across the two pins of header J2 on the jig. Caution: Voltages as high as +115V may briefly (< 100 us) be present on the jig when the output is on, if no DUT is installed. Protective circuits will disable the high voltage output if this condition occurs, but for maximum safety, the interlock circuit should not be overridden.

WIRING DIAGRAM

PERFORMANCE CHECKSHEET