DETECTING THE RATIO OF I AC

Similar documents
Detecting the Ratio of I ac. /I ave. photoelastic modulators

STOKES POLARIMETRY STOKES POLARIMETRY BY DR. THEODORE C. OAKBERG APPLICATION NOTE PHOTOELASTIC MODULATORS

225 Lock-in Amplifier

Optical Power Meter Basics

Application Note (A12)

THE BENEFITS OF DSP LOCK-IN AMPLIFIERS

ELEC3242 Communications Engineering Laboratory Frequency Shift Keying (FSK)

UNIT-3. Electronic Measurements & Instrumentation

Notes on Noise Reduction

Chapter 2 Signal Conditioning, Propagation, and Conversion

Available online at ScienceDirect. Energy Procedia 55 (2014 )

Physics 262. Lab #1: Lock-In Amplifier. John Yamrick

Part 2: Second order systems: cantilever response

Continuous Wave Magneto Optic Kerr

1. Explain how Doppler direction is identified with FMCW radar. Fig Block diagram of FM-CW radar. f b (up) = f r - f d. f b (down) = f r + f d

Introduction. sig. ref. sig

Measuring of small AC signals using lock-in amplifiers. Narrow band selective amplifiers + amplitude detector. Lock-in amplifiers

Electronic Instrumentation & Automation. ET-7th semester. By : Rahul Sharma ET & TC Deptt. RCET, Bhilai

PCS-150 / PCI-200 High Speed Boxcar Modules

Measuring of small AC signals using lock-in amplifiers. Narrow band selective amplifiers + amplitude detector. Lock-in amplifiers

Using Circuits, Signals and Instruments

Experiment 7: Frequency Modulation and Phase Locked Loops

Measuring of small AC signals using lock-in amplifiers. Narrow band selective amplifiers + amplitude detector. Lock-in amplifiers

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

Linear Time-Invariant Systems

SHF Communication Technologies AG. Wilhelm-von-Siemens-Str. 23D Berlin Germany. Phone Fax

First and second order systems. Part 1: First order systems: RC low pass filter and Thermopile. Goals: Department of Physics

About Lock-In Amplifiers Application Note #3

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

SHRI ANGALAMMAN COLLEGE OF ENGINEERING & TECHNOLOGY (An ISO 9001:2008 Certified Institution) SIRUGANOOR,TRICHY

Experiment 19:Small Signal Measurement

8.5 Modulation of Signals

14.2 Photodiodes 411

PLL Synchronizer User s Manual / Version 1.0.6

AC Resistance Thermometry Bridges and their Advantages By Peter Andrews

Calibration of photoeiastic modulators in the vacuum UV

1 Lock-in Amplifier Introduction

HARDWARE IMPLEMENTATION OF LOCK-IN AMPLIFIER FOR NOISY SIGNALS

Chapter 2 Analog-to-Digital Conversion...

Exercise 6 AC voltage measurements average responding voltmeters

Signal Conditioning Fundamentals for PC-Based Data Acquisition Systems

Receiver Signal to Noise Ratios for IPDA Lidars Using Sine-wave and Pulsed Laser Modulation and Direct Detections

ELEC3242 Communications Engineering Laboratory Amplitude Modulation (AM)

A Prototype Wire Position Monitoring System

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences

Photometer System Mar 8, 2009

Stability of a Fiber-Fed Heterodyne Interferometer

Stabilizing an Interferometric Delay with PI Control

LFR: flexible, clip-around current probe for use in power measurements

Meeting Measurement Challenges For Low-Power, Pulsed, Or Modulated Light Sources

Polarization Experiments Using Jones Calculus

Chapter 1. Overview. 1.1 Introduction

Model Series 400X User s Manual. DC-100 MHz Electro-Optic Phase Modulators

CA330 RTD Calibrator: High-speed Response and High-resolution Resistance Simulator

Chlorophyll a/b-chlorophyll a sensor for the Biophysical Oceanographic Sensor Array

First Time User Manual

Definitions. Spectrum Analyzer

EE-4022 Experiment 3 Frequency Modulation (FM)

Chapter 2 TELEMETRY SYETEMS

PHY 123/253 Shot Noise

Table of Contents...2. About the Tutorial...6. Audience...6. Prerequisites...6. Copyright & Disclaimer EMI INTRODUCTION Voltmeter...

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

WELCOME TO PHYC 493L Contemporary Physics Lab

Laser Diode Characterization and Its Challenges

1.5µm PbSe Power Detector

SIGNAL RECOVERY. Model 7265 DSP Lock-in Amplifier

Keysight Technologies Pulsed Antenna Measurements Using PNA Network Analyzers

rf SQUID Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

SEQUENTIAL NULL WAVE Robert E. Green Patent Pending

BLACKBODY RADIATION PHYSICS 359E

Variable Gain Photoreceiver - Fast Optical Power Meter

As Published on EN-Genius.net

PHASE TO AMPLITUDE MODULATION CONVERSION USING BRILLOUIN SELECTIVE SIDEBAND AMPLIFICATION. Steve Yao

Modulation is the process of impressing a low-frequency information signal (baseband signal) onto a higher frequency carrier signal

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Quantum frequency standard Priority: Filing: Grant: Publication: Description

Name Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

Initial Results from the C-Mod Prototype Polarimeter/Interferometer

Electro-optic components and systems Toll Free:

Chapter 1. Electronics and Semiconductors

Electron Spin Resonance v2.0

LLS - Introduction to Equipment

Optical design of shining light through wall experiments

Goals of the Lab: Photodetectors and Noise (Part 2) Department of Physics. Slide 1. PHYSICS6770 Laboratory 4

Exercise 1-3. Radar Antennas EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION OF FUNDAMENTALS. Antenna types

Introduction. In the frequency domain, complex signals are separated into their frequency components, and the level at each frequency is displayed

Sampling and Reconstruction

Lab 12 Microwave Optics.

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

MAHARASHTRA STATE BOARD OF TECHNICAL EDUCATION (Autonomous) (ISO/IEC Certified) SUMMER 14 EXAMINATION Model Answer

Directly Chirped Laser Source for Chirped Pulse Amplification

9. Microwaves. 9.1 Introduction. Safety consideration

MAE334 - Introduction to Instrumentation and Computers. Final Exam. December 11, 2006

Timing Noise Measurement of High-Repetition-Rate Optical Pulses

Spectral phase shaping for high resolution CARS spectroscopy around 3000 cm 1

note application Measurement of Frequency Stability and Phase Noise by David Owen

Putting it all Together

H represents the value of the transfer function (frequency response) at

Transcription:

T E C H N O L O G Y F O R P O L A R I Z A T I O N M E A S U R E M E N T DETECTING THE RATIO OF I AC MEASUREMENT OF THE RAGE INTENSITY OF A MODULATED LIGHT BEAM In any experiment using photoelastic modulators (PEMs) it is necessary to compare the time average intensity of the light at the detector with the amplitude of a single frequency component of the light intensity. 1,2 For example, in experiments to measure circular dichroism the effect being measured is proportional to the ratio I f ave where I f is the amplitude of the Fourier series component of the light intensity function at the modulator frequency f and I ave is the average of the light intensity function, integrated over a time interval much longer than the period of oscillation of the PEM. 3 The case for linear dichroism is the same, except that the frequency component needed is twice the modulator frequency, or 2f. In many other experiments these ratios may be used to normalize the optical signal so that the results obtained are independent of fluctuations in the intensity of the transmitted light beam, for example changes in the intensity of the light source or changes in absorption, scattering, etc. in the optical components. The measurement of this average light intensity is not trivial, and it presents some important challenges for the experimenter. Of course, what is measured directly is the electrical output of a transducer which converts the light intensity information into an electrical signal, either a current or a voltage. The problem reduces to determining the ratio of electrical signals, rather than optical quantities. The determination of these ratios. and especially the measurement of the average intensity, current or voltage is the subject of this application note. Detection of the AC Signal Component The analysis done on the light intensity function entering the detector (or its electrical signal counterpart) is commonly based on a Fourier series analysis of this intensity function. amplifiers are used to measure amplitudes of the AC components of this signal, for example at the modulator frequency (f), or twice the modulator frequency (2f). The lock-in amplifier requires an input reference signal (provided by the PEM controller) which is precisely synchronized with the optical signal. The result is an electronic voltmeter with a very narrow frequency bandwidth. Typical lock-in amplifiers may be adjusted to detect f, 2f, or in some cases even higher harmonics of the PEM frequency. The basic phase sensitive detection scheme of the lock-in amplifier is sensitive to odd harmonics of the detected signal frequency. Many typical PEM setups produce rather strong harmonic components, especially the third harmonic of the detected signal frequency. Using electronic filtering (typically provided by the lock-in) to minimize the effects of these higher harmonics is recommended. amplifiers provide several output modes: 1) visual, by a meter or digital display, 2) analog, a voltage output proportional to the amplitude of the detected signal, and 3) a digital output to an RS-232 or GPIB computer interface. Many lock-ins can also digitize other signals (such as the average signal component) or even compute the ratio of the lock-in output to another signal. For modern PEM setups the digital output to a computer is the probably the most important. Page 1

Detection of the Average Intensity Component The average or zero frequency Fourier component of the light intensity function (or the detector electrical signal) may be determined by either of two methods: DC electronic techniques or AC electronic techniques which involve chopping or periodically interrupting the light beam. Each technique has its own advantages and disadvantages, with the choice frequently being dictated by the type of detector that is selected. These techniques are described in succession below. DETECTION OF RAGE SIGNAL COMPONENTS USING DC TECHNIQUES Electronic circuits with electrical bandwidths down to DC may be used for measurement of the average component of the electronic signal. Such circuits, if they can be used, have an advantage of simplicity and low cost. They have the disadvantage that electrical offsets (DC components in the electrical signal not related to the average light-intensity) must be dealt with very carefully. Offsets These electrical offsets arise from three main sources: 1. The dark current or intrinsic electrical signal from the detector under conditions of no light exposure. 2. Ambient light, not a part of the experiment light beam, failing on the detector. 3. Electrical offsets which arise from the electronic circuitry associated with the detector. These electrical offsets must either be eliminated or reduced to acceptably low levels, or characterized and corrected for in the computation of the average signal, or V ave, The three sources of electrical offsets are discussed in detail below. 1. Dark current Most detectors have a DC dark current component in their output which is an intrinsic characteristic of the device. The dark current is dependent on temperature, and in the case of detectors such as PMTs and avalanche photodiodes, strongly depends on the operating voltage applied. In many cases, it may be possible to reduce the dark current to negligible levels by cooling the detector. Silicon photovoltaic photodiodes have low dark currents at room temperature. 2. Ambient light The traditional method of eliminating the effect of ambient light is to operate the experimental setup in a dark environment. Alternatively, if a monochromatic light source is being used, placing a narrow band spectral filter over the detector aperture may reduce the ambient light falling on the detector to an acceptable level. This would result in much better access to the equipment during the experiment. 3. Electrical offsets Active electrical components such as operational amplifiers have DC offsets associated with their operation. It is usually possible to include in their circuits a means of reducing the offset to zero. In some cases such nulling circuits might also be used to compensate for dark current or ambient light effects. Caution must be used, since the null adjustment may drift with time and with changes of the operating temperature of the electronic equipment. If DC detection techniques are being considered, a careful analysis of the offsets and their effects must be made. Questions which should be asked include: What accuracy is required in the polarization modulation measurements being made? What is the magnitude of the polarization modulation signal produced by the effect being observed? What is the magnitude and stability of the offsets in the system? Are the offsets small enough to be ignored, or can they by corrected by using electronic methods? If not, can the offsets be characterized and corrections be made during computation of the V AC /V ave ratio? Page 2

Input + FIGURE 1. OPERATIONAL AMPLIFIER CIRCUIT FOR MEASURING THE TIME RAGE OF A COMPLEX PERIODIC SIGNAL Average signal detection An electrical circuit for detecting an average electrical signal composed of complex AC and DC components is shown below. This is an active low-pass filter circuit using an operational amplifier. The time constant τ (characteristic response time) of the circuit is given by τ = 1/RC and the output is proportional to the average voltage or current of the input signal. Of course, other circuit elements are required for practical use of this circuit. Hinds Instruments uses this circuit in its Model SCU-90 signal conditioning unit. The instrument derives, from an input detector signal, a wide-band AC signal and a low-pass or DC signal. The signal conditioner also provides adjustable gain, primarily for the DC output. The output is an analog signal proportional to l ave which is suitable for input to a digitizing circuit such as is frequently provided by a lock-in amplifier. Figure 2 shows a block diagram for typical DC-coupled signal detection. AC Signal Path C Amplifier Low-pass Filter R Operational Amplifier Adjustable Gain Amplifier Output A-D Converter Computer RS232 or GPIB FIGURE 2. BLOCK DIAGRAM FOR DC-COUPLED RAGE SIGNAL DETECTION Another approach to average signal detection would be to use a digital voltmeter (DVM) connected to the output of the detector in parallel with the lock-in amplifier. Many such instruments are equipped with RS-232 or GPIB interfaces. In deciding whether such an instrument could be used for average signal measurement, the magnitude of the expected signal, the sensitivity of the DVM, and the digital resolution of the A-D converter need to be considered. Also, the experimenter must determine that the DC measurement of a complex electrical signal is in fact proportional to the time average of the signal. Detection of Average Signal Components using AC Techniques (Chopping) The major advantage of using chopping (periodic interruption of the light beam) as a means of average signal measurement is that DC offsets are eliminated. The primary disadvantage is an additional instrument will need to added to the system. A high-spped optical chopper is manufactured by Hinds Instruments. This chopper uses the resonance modulation of the PEM combined with a polarizer and a waveplate to produce an on/off effect at high speeds. Chopping frequencies are typically 50 khz. Detection of both the polarization modulated signal and the chopping signal is done using lock-in amplifiers. The most direct method is shown in Figure 3. Page 3

Chopper Reference FIGURE 3. BLOCK DIAGRAM FOR DETECTOR SIGNAL ANALYSIS USING CHOPPING FOR RAGE SIGNAL DETECTION amplifier I must have frequency response capable of accepting the polarization modulated signal at the modulator frequency f and twice this value, 2f. A lower frequency model lockin (probably less expensive) may be used for lock-in 2. Many experimental setups require the measurement of both f and 2f components of the detector signal. If this is the case, two lock-in amplifiers of type I would be required for real time measurements. If the light source and signals are stable and if measurement speed requirements are not severe, a single lock-in amplifier could be switched under computer control to measure these two quantities sequentially. Many lock-in amplifiers have provision for calculating the ratio of the signal detected by the lock-in amplifier to another analog signal. For example, the analog output of lock-in 2 could be sent to lockin 1, the ratio calculated and transmitted to the computer. Alternatively, the analog output of the ratio circuit could be displayed with an external digital voltmeter. Chopper Reference AC Signal Path Amplifier 1 PEM Reference Amplifier 2 Average Signal Path Amplifier 1 PEM Reference Amplifier 2 Amplifier 3 RS232/ GPIB Computer Computer FIGURE 4. ALTERNATIVE BLOCK DIAGRAM FOR DETECTOR SIGNAL ANALYSIS USING CHOPPING FOR RAGE SIGNAL DETECTION I in Figure 3 must detect an interrupted polarization modulation signal. While the phase relationship between the polarization modulation signal and the modulator reference is preserved accurately, this situation is not ideal. Better accuracy and rejection of noise can be obtained with the setup shown in Figure 4. For this setup, lock-in amplifier 1 must have an output time constant much less than the chopper period. For a chopper frequency of a few hundred khz, a time constant of 1 msec or less would be required. Usually this dictates the use of a high quality analog lock-in amplifier.4 amplifier 3 has frequency requirements similar to lockin 2. Calibrating the relationship between V AC and V ave Whether AC or DC electronic techniques are used for detecting the average signal V ave an arbitrary experimental constant is likely to appear in the output signal and thus in the ratio of V AC /V ave. In the case of a Hinds signal conditioner, it is likely that some amplification has been used in the average signal circuit. In the case of the lock-in amplifier, the use of filters or chopping may alter the response of the instrument. The lock-in amplifier also measures r.m.s. quantities, whereas polarization modulation theory gives peak or amplitude quantities. Page 4

There is therefore a need to produce an optical signal for which the ratio of I AC ave is known with precision. Use of such a signal will enable the experimenter to determine the experimental constant K 1 which relates the ratio of the experimental outputs to the ratio of optical intensities, as given in equation 1. I AC V = AC K1 I ave V ave (1) Fortunately this is easily done with the PEM and other equipment common to polarization modulation experiments. Figure 5 shows the optical setup for determining the experimental constant K 1. Monochromatic light source Polarizer 45 A PEM 0 Polarizer 45 PEM Controller Oscilloscope FIGURE 5. OPTICAL BENCH SET-UP FOR V AC /V CALIBRATION In general, this experimental setup should resemble as closely as possible the optical bench setup for the experiment in question. Some additional precautions need to be mentioned. First, if the light source is a laser, the effects of modulated interference must be minimized or eliminated. 5 Second, the polarizers should be of high quality (such as Glan-Thompson calcite polarizers) with very good extinction characteristics. Third, all DC offsets must be minimized or eliminated, or properly compensated for in the calculation of V ave. Jones calculus analysis of the above optical setup leads to the following expression for the timedependent intensity of the light at point B. 6 The intensity of the linearly polarized light at point A is I 0 I = I 0 K 2 1 - cos(a 0 cos( t)) 2 [ ] B f reference (trigger) to detector signal analysis where K 2 is an experimental constant which corrects for the transmission losses in the PEM and second polarizer, A 0 is the peak retardation of the PEM expressed in units of phase angle (e.g. radians) and the angular frequency Ω= 2πf. Expanding equation 2 in a Fourier series yields equation 3, which displays the terms (I ave and I 2f ) involved in the intensity function. The functions J n (A 0 ) are Bessel functions of integer order. Input I = I 0 K 2 1 - J 0 (A 0 ) + 2 J 2 (A 0 )cos(2 t) 2 J 4 (A 0 )cos(4 t) +... 2 [ { I ave { I 2f ] (2) (3) Page 5

The ratio of the AC intensity amplitude (I 2f ) to the average intensity (I ave ) is therefore given by: I 2f = 2J 2 (A 0 ) I (4) ave 1 J 0 (A 0 ) The condition of half-wave peak retardation is particularly useful for this calibration. The oscilloscope waveform is very distinctive, as shown in Figure 6. FIGURE 6. OSCILLOSCOPE WFORMS HALF-W RETARDATION The characteristic flat top (or flat bottom) of this waveform can be used to establish that the PEM has been accurately adjusted to half-wave retardation. (See the Calibration Application Note.) 7 For A 0 = π, J 0 (π) = -0.30424 and J 2 (π) = 0.48543, giving I 2f I ave = 0.744 (5) This with the experimentally determined ratio V 2f /V ave is sufficient for determining the constant K, in equation 1. References 1. K.W. Hipps and G.A. Crosby, Applications of the photoelastic modulator to polarization spectroscopy, J. Phys. Chem. 83, 555-562 (1979). 2. A.F. Drake, Polarization modulation--the measurement of linear and circular dichroism, J. Phys. E. 19, 170-181 (1986). 3. Note: In many articles on this subject the notation I DC or V DC is used rather than I ave and V ave. The latter notation will be used in this application note to avoid confusion when electronic techniques are discussed later. 4. Mr. Ted Schurter, EG&G Instruments, Princeton Applied Research. Private communication 5. T.C. Oakberg, Modulated interference effects: use of photoelastic modulators with lasers, Opt. Engr. 34, 1545-1550 (1995). 6. J.C. Kemp, Polarized Light and its Interactions with Modulating Devices, Hinds Instruments, Inc., Hillsboro, OR (1987). 7. T.C. Oakberg, Calibration Application Note, Hinds Instruments, Inc, Hillsboro, OR (1991). Hinds Instruments, Inc / 7245 NW Evergreen Pkwy / Hillsboro, OR 97124 / USA T: 503.690.2000 / F: 503.690.3000 / sales@hindsinstruments.com / www.hindsinstruments.com PEMlabs is a Trademark of Hinds Instruments, Inc. Manufactured in USA 2005, 2009, 2010 Hinds Instruments, Inc. All rights reserved. Printed in USA Page 6