Digital Detector Array Image Quality for Various GOS Scintillators

Similar documents
Amorphous Selenium Direct Radiography for Industrial Imaging

Performance evaluation of a photon counting detector for high energy NDT applications

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Investigation of Effective DQE (edqe) parameters for a flat panel detector

10/3/2012. Study Harder

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

10/26/2015. Study Harder

Digital Radiography : Flat Panel

Conversion to Digital Radiography from Film Radiography

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Features and Weaknesses of Phantoms for CR/DR System Testing

Digital Radiography for the Inspection of Small Defects

Film Replacement in Radiographic Weld Inspection The New ISO Standard

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

A COMPARATIVE STUDY ON THE PERFORMANCE OF DIGITAL DETECTOR SYSTEMS FOR HIGH ENERGY APPLICATIONS

MTF and NPS of single-shot dual-energy sandwich detectors

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

ON THE WAY TO DIGITAL RADIOGRAPHY

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Computed Radiography

Migrating from traditional to Digital Radiography in Aerospace

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Performance of Image Intensifiers in Radiographic Systems

Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations

ISO INTERNATIONAL STANDARD

Any question concerning this ballot item should be addressed to Cliff Bueno at or

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Modeling detector performance in digital mammography using the linear cascaded systems approach

VOLUMETRIC computed tomography (CT) based on flatpanel

DRX Plus Detectors: Going from Good to Great

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

APPLICATION OF THE DIGITAL RADIOGRAPHY IN WELD INSPECTION OF GAS AND OIL PIPELINES

FOCUS. Digital Radiography in Industry: Digital Detector Arrays in Radiographic Testing. Introduction. Definitions. by Brad Kraai

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Some Essential Physics of Flat-Panel Detectors

BASICS OF FLUOROSCOPY

Shad-o-Box HS Product Family

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

DELWORKS DR MEDICAL. take the next step

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Current technology in digital image production (CR/DR and other modalities) Jaroonroj Wongnil 25 Mar 2016

Determination of the detective quantum efficiency of a prototype, megavoltage indirect detection, active matrix flat-panel imager

Advances in High Energy X-ray Digital Detector Arrays

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

7/24/2014. Image Quality for the Radiation Oncology Physicist: Review of the Fundamentals and Implementation. Disclosures. Outline

Advancements In Digital Radiography: CR, DR, and DICONDE

Research Article Image Quality Assessment of a CMOS/Gd 2 O 2 S:Pr,Ce,F X-Ray Sensor

Setting up digital imaging department!

Digital Radiography Systems Techniques and Performance Evaluation for Space Applications

» There are many possible techniques for making good radiographs. » Following a defined procedure will simplify technique

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Do you have any other questions? Please call us at (Toll Free) or , or

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

SYLLABUS. TITLE: Equipment Operation I. DEPARTMENT: Radiologic Technology

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Evaluation and Comparison of Image Quality for Indirect Flat Panel Systems with CsI and GOS Scintillators

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

AeroDR HD. Wireless Digital Radiography System

Alma Mater Studiorum Università di Bologna

Basis of Computed Radiography & PACS

Recommended Training Curriculum For Digital Radiography Personnel (Level II)

HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

COPYRIGHT 2002 by Srinivasan Vedantham ALL RIGHTS RESERVED. Use or inclusion of any portion of this document in another work intended for

Fluoroscopy - Chapter 9

Hardware for High Energy Applications 30 October 2009

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

X-ray light valve (XLV): a novel detectors technology for digital mammography*

Real Time Linear Array Imaging. Brian Caccamise

A Comprehensive Review of Image Production

Digital Mammography: From Physical Performance Evaluation To Image Quality Analysis

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 2: X- and gamma-ray techniques with digital detectors

Effect of Ink Spread and Opitcal Dot Gain on the MTF of Ink Jet Image C. Koopipat, N. Tsumura, M. Fujino*, and Y. Miyake

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

An aging study of the signal and noise characteristics in large-area CMOS detectors

O 2. Image Quality Assessment of a CMOS/Gd 2. S:Pr,Ce,F X-ray Sensor. Journal of Physics: Conference Series PAPER OPEN ACCESS

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing

radiography detector

Rad-icon Large-Area Industrial X-Ray Detectors

Seminar 8. Radiology S8 1

A Study of Slanted-Edge MTF Stability and Repeatability

Digital Images & Image Quality

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

X-ray Tube and Generator Basic principles and construction

Dürr NDT GmbH & Co. KG Höpfigheimer Straße 22 D Bietigheim-Bissingen Germany. Contract No. BAM ZBA Dürr

Fast MTF measurement of CMOS imagers using ISO slantededge methodology

Characteristics of a new cone beam computed tomography

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

Rad-icon Imaging Corp A Division of DALSA Corporation

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Digital Imaging Considerations Computed Radiography

Modelling Computed Radiography Detectors with a Cascaded Linear System Model

Transcription:

Digital Detector Array Image Quality for Various GOS Scintillators More info about this article: http://www.ndt.net/?id=22768 Brian S. White 1, Mark E. Shafer 2, William H. Russel 3, Eric Fallet 4, Jacques Roussilhe 5, and Karin Toepfer 6 1 Carestream Health, Inc., United States, brian.white@carestream.com 2 Carestream Health, Inc., United States, mark.shafer@carestream.com 3 Carestream Health, Inc., United States, william.russel@carestream.com 4 Carestream Health, Inc., France, eric.fallet@carestream.com 5 Carestream Health, Inc., France, jacques.roussilhe@carestream.com 6 Carestream Health, Inc., United States, karin.toepfer@carestream.com Abstract Digital detector arrays (DDAs) are becoming more prevalent for industrial radiographic inspection. The type of scintillator used with a DDA determines the overall achievable image quality. DRZ High, DRZ Plus, DRZ Standard, and DRZ Fine gadolinium oxysulfide terbium (GOS: Tb) activated scintillators were investigated in combination with the Carestream DRX PLUS digital detector array panel. Three beam conditions were utilized for exposure: RQA-5 (70 kvp, 21 mm aluminum), RQA-9 (120 kvp, 1 mm copper and 4 mm aluminum), and NDT (220 kvp, 8 mm copper). The detective quantum efficiency (DQE), modulation transfer function (MTF), sensitivity, and interpolated basic spatial resolution (isrb) were determined for each scintillator and beam condition. The results of this investigation are presented in this paper. 1. Introduction DDA technology is gaining acceptance within the NDT industry due to workflow advantages 1,2,3,4. Nearly all medical radiological applications have converted to indirect DDAs. DDAs are electronic imaging devices that convert X-rays or gamma rays to light, which is stored as a voltage and is subsequently sampled to form the digital image. There are two common types of scintillators for radiology and radiography: cesium iodide thallium (CsI:Tl) and GOS. CsI-type scintillators are not commonly used for industrial radiography because of ghosting issues above 150 kv; therefore, GOStype scintillators are almost always chosen for NDT applications. GOS scintillators are optically coupled to glass that contains a patterned array of pixels. Each pixel contains a photodiode, which senses the light from the scintillator, and a thin film transistor (TFT), which stores the charge created by the photodiode. The thickness and phosphor size of the scintillator largely influence the overall quality of the digital image, which is a combination of brightness, sharpness, and noise. Creative Commons CC-BY-NC licence https://creativecommons.org/licenses/by-nc/4.0/

DQE is a measure of image quality as a function of spatial frequency 5, and it is defined as: DQE f 2 2 q g T f NNPS f where DQE is the detective quantum efficiency, f denotes that the term is a function of spatial frequency, q is the density of incident quanta per unit area at the detector (flux), g is the system gain, T is the MTF, and NNPS is the normalized noise power spectrum. The DQE provides a quantitative way to measure the detection capability of an imaging system for given exposure parameters as a function of spatial frequency. MTF is a way to measure the achievable detail that a system can obtain. The MTF describes the contrast of an image as a function of spatial frequency. MTF is calculated by measuring the edge response of a tungsten phantom with a very sharp angle. A line profile is drawn in the radiograph across the sharp angle, resulting in an edge spread function (ESF). Taking the derivative of the ESF results in a rate of change across the angled edge, which is known as the line spread function (LSF). An oversampled LSF has a Fourier transform applied to it to yield the contrast modulation as a function of spatial frequency, as the Fourier transform decomposes the LSF into the frequency domain. Basic spatial resolution (SRb) is a measure of the amount of detail that can be seen in an image with a duplex wire gauge placed directly on the detector 6. The duplex wire gauge consists of several elements. Each element has two wires with a specific diameter and spacing between them. In an image, a line profile is drawn perpendicular to the elements. The element is said to be resolved if the intensity difference is greater than 20% of the wires against their background., 2. Experimental Four different GOS-type scintillators were investigated in combination with the Carestream DRX PLUS DDA for three different X-ray beam conditions. Table 1 describes the GOS scintillator types that were investigated. Scintillator Type Support Table 1: DRZ scintillator types. Phosphor Overcoat Total DRZ Fine 250 110 6 366 DRZ Standard 250 140 6 396 DRZ Plus 250 210 6 466 DRZ High 180 310 9 499 2

The DRZ Plus screen is always chosen for DDAs that utilize GOS for medical radiology. The DRZ Standard screen is normally utilized for NDT applications; however, the DRZ Fine screen has been introduced for DDAs with smaller pixel pitch. We wanted to determine, through scientific analysis, the best scintillator for our DRX PLUS DDA panel, which has a pixel pitch of 139 µm. The four GOS-type screens were placed in pressure contact with a TFT photodiode array. Exposures were performed with standardized beam conditions, RQA-5 and RQA- 9, which are used in medical radiology, and NDT, which is commonly used for industrial radiography. Individual images were acquired for the dark calibration, gain calibration, flat-field noise power spectrum, slanted-edge MTF target, and duplex wire gauge. MATLAB was utilized for the analysis of the DQE and MTF. For the DQE comparisons, the exposure was adjusted to match the DRZ Plus signal level. 3. Discussion As the thickness and phosphor size of the GOS scintillator is changed, the brightness, sharpness, and noise of the image changes dramatically. Most radiographers do not realize that the scintillator choice determines the image quality of the DDA, and it is often overlooked and taken for granted. Thinner GOS scintillators with smaller phosphors will result in images that are sharper, with reduced brightness and improved noise uniformity. Likewise, thicker GOS scintillators with larger phosphors will result in images that have reduced sharpness, with increased brightness and degraded noise uniformity. The balance of the sharpness and the signal-to-noise ratio will determine the overall image quality. Therefore, the proper choice of scintillator determines whether or not the DDA can meet inspection requirements. The brightness of the scintillator helps to determine the amplification or gain of the system. Brighter scintillators can result in improved image quality if all else is equal. Relative to the DRZ Plus, the DRZ Fine was 2.0 times less sensitive, the DRZ Standard was 1.27 times less sensitive, and the DRZ High was 1.40 times more sensitive for the RQA-5 beam condition. Table 2 provides a comparison for the two different beam conditions. Table 2: DRZ scintillator sensitivity. ADu per mr RQA5 RQA9 DRZ Fine 2224 2607 DRZ Standard 3503 4132 DRZ Plus 4445 5477 DRZ High 6229 7871 3

The overall image quality was measured by DQE analysis. Figures 1, 2, and 3 present the DQE results for the four GOS screens at the three beam conditions. The DRZ Standard scintillator had the best DQE above 2 cycles/mm. Below 2 cycles/mm, the DRZ Plus scintillator was the best choice as the DRZ High lacked sufficient sharpness for consideration. The DRZ Fine screen did not have suitable image quality except at very high spatial frequencies. As the kv increased, the overall achievable image quality decreased. Figure 1: DQE Comparison for RQA5, 70 kv, 21 mm Al DRZ scintillators. Figure 2: DQE Comparison for RQA9, 120 kv, 1 mm Cu and 4 mm Al DRZ scintillators. 4

Figure 3: DQE Comparison for 220 kv, 8 mm Cu DRZ scintillators. Figures 4, 5, and 6 present the MTF results for four different screens at the three beam conditions. The sharpness of the DRZ Fine scintillator was clearly the best across all spatial frequencies, followed by the DRZ Standard, DRZ Plus, and the DRZ High. The sharpness decreased at higher kv for all GOS screens. Figure 4: MTF, RQA5, 70 kv, 21 mm Al. 5

Figure 5: MTF, RQA9, 120 kv, 1 mm Cu and 4 mm Al. Figure 6: MTF, 220 kv, 8 mm Cu. Table 2 presents the interpolated basic spatial resolution results for the duplex wire gauge method. The four scintillator screens were tested using the three beam conditions. Resolution below the pixel pitch of the detector was achieved with the DRZ Fine screen. The DRZ Standard screen gave a resolution near or below the pixel pitch, whereas the DRZ Plus screen resulted in a resolution above the pixel pitch. The DRZ High screen had a resolution much higher than the pixel pitch of the detector. In general, the sharpness became worse as the kv was increased. 6

Table 2: Duplex wire gauge resolution values. Resolution (µm) RQA5 RQA9 220 kv, 8 mm Cu DRZ Fine 95 100 105 DRZ Standard 120 125 130 DRZ Plus 160 160 160 DRZ High 225 225 225 3. Summary The DRZ Standard GOS scintillator screen was the best choice for NDT applications utilizing the Carestream DRX PLUS digital detector array panel. The choice of the DRZ Standard screen resulted in the best overall image quality above 2 cycles/mm, and it resulted in images that had sharpness that was at or below the pixel pitch of the detector. Acknowledgements The author wishes to acknowledge Sharon Ryan who graciously reviewed this paper prior to publication. References 1. ASTM 2736, Standard Guide for Digital Detector Array Radiography, ASTM International. 2. ASTM E2698, Standard Practice for Radiological Examination Using Digital Detector Arrays, ASTM International. 3. ASTM 2737, Standard Practice for Digital Detector Array Performance Evaluation and Long-Term Stability, ASTM International. 4. ASTM 2597, Standard Practice for Manufacturing Characterization of Digital Detector Arrays, ASTM International. 5. Dainty, J. C. and Shaw, R., Image Science: Principles, Analysis and Evaluation of Photographic Type Imaging Processes, Academic Press, New York, NY, 1974. 6. ASTM E2002, Standard Practice for Determining Total Image Unsharpness and Basic Spatial Resolution in Radiography, ASTM International. 7