Characterization of medical devices electromagnetic immunity to environmental RF fields.

Similar documents
BIODEX MULTI- JOINT SYSTEM

Measurement Procedure & Test Equipment Used

EMC Overview. What is EMC? Why is it Important? Case Studies. Examples of calculations used in EMC. EMC Overview 1

EMC Seminar Series All about EMC Testing and Measurement Seminar 1

Test Equipment. PHYS 401 Physics of Ham Radio

SERIES K: PROTECTION AGAINST INTERFERENCE

E-Field Uniformity Test Volume In Gtem Cell Based On Labview

Technician License Course Chapter 2. Lesson Plan Module 2 Radio Signals and Waves

Biological Safety. Electromagnetic Compatibility (EMC) Observe the following precautions related to biological safety.

Practical Considerations for Radiated Immunities Measurement using ETS-Lindgren EMC Probes

Harmonizing the ANSI-C12.1(2008) EMC Tests. Harmonizing the ANSI-C12.1(2008) EMC Tests

OtoRead - Technical Specifications Page 0. Technical Specifications. OtoRead D A 2017/06

General Safety/EMC and Electrical Information for i-limb ultra and i-limb digits

Power Sensors Ltd. PQube 3 AC Analyzer IEC Class 0,2 S Accuracy Compliance Report

Standardisation and Immunity Tests regarding IEMI

EMC Test Report. Report Number: M030826

Human Exposure Requirements for R&TTE and FCC Approval

An Introduction to EMC Testing (what can be done with scopes) Vincent Lascoste EMC Product Manager - RSF

EMC aspects associated to 5G networks

Amateur Wireless Station Operators License Exam

EMC TEST REPORT. Report No.: TS EME Model No.: 33XR-A Issued Date: Jan. 08, 2009

2620 Modular Measurement and Control System

8GHz RF EMF Strength Meter

How will the third edition of IEC affect your test facility?

TEST REPORT... 1 CONTENT...

IEEE Electromagnetic Compatibility Standards (Active & Archive) Collection: VuSpec

CONDUCTED RF EQUIPMENT POWER AMPLIFIERS. Practical RF Immunity System Design Considerations

10 Mb/s Single Twisted Pair Ethernet Noise Environment for PHY Proposal Evaluation Steffen Graber Pepperl+Fuchs

ETSI ETR 357 TECHNICAL January 1997 REPORT

EMC and Printed Circuit Board Design Design guidelines and concepts made simpler Part I Basics, Noise Sources, Tests

10 Mb/s Single Twisted Pair Ethernet Noise Environment for PHY Proposal Evaluation Steffen Graber Pepperl+Fuchs

EM-ISight Electromagnetic Scanning System. Since Page 1 of 6 Phone (613) Fax (613)

Emerging Standards for EMC Emissions & Immunity

GTEM cells. Emissions and immunity testing in a single, shielded environment

EMC and Variable Speed Drives

Guidance and Declaration - Electromagnetic Compatibility (EMC) for the Delfi PTS ii Portable Tourniquet System

Specification for Radiated susceptibility Test

FREQUENCY AGILE FM MODULATOR INSTRUCTION BOOK IB

DraftETSI EN V1.1.1 ( )

INTRODUCTION TO COMMUNICATION SYSTEMS AND TRANSMISSION MEDIA

EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS

Electromagnetic compatibility Guidance and manufacturer s declaration DIN EN :2007 (IEC :2007)

ISO INTERNATIONAL STANDARD

QM Radio Transmitter Module

CHAPTER 6 EMI EMC MEASUREMENTS AND STANDARDS FOR TRACKED VEHICLES (MIL APPLICATION)

Bulk Current Injection instead of Radiated immunity testing, in the range from 1 MHz upto 1 GHz: Measuring results

National 4. Waves and Radiation. Summary Notes. Name:

Future In Radiated Immunity Testing

D C 01/2019 3

Why/When I need a Spectrum Analyzer. Jan 12, 2017

LS200 TEST DATA IEC61000 SERIES

3.2 Measuring Frequency Response Of Low-Pass Filter :

Application Note #38B Automotive 600V/m Radar Pulse Test Solution

1. Electro-Static Discharge Test EN R Radiated Susceptibility Test EN R-2

Data Sheet, V1.0, Aug SMM310. Silicon MEMS Microphone. Small Signal Discretes

Bird Model 7022 Statistical Power Sensor Applications and Benefits

RF EMF Strength Meter

Overview of the ATLAS Electromagnetic Compatibility Policy

Certificate of Test AND KEEPS ALL REQUIREMENTS ACCORDING THE FOLLOWING REGULATIONS IEC :2001 IEC :2007

EMC standards. Presented by: Karim Loukil & Kaïs Siala

Double Probing Frequency Non-Linear Junction Detector «LORNET-0836» Technical Description & User Manual Certificate

PRINCIPLES OF COMMUNICATION SYSTEMS. Lecture 1- Introduction Elements, Modulation, Demodulation, Frequency Spectrum

Katran-Lux. Non-linear junction detector USER MANUAL

Physics of RFID. Pawel Waszczur McMaster RFID Applications Lab McMaster University

Test Plan for Hearing Aid Compatibility

An Introduction to Spectrum Analyzer. An Introduction to Spectrum Analyzer

Chapter 1: Introduction. EET-223: RF Communication Circuits Walter Lara

Accurate Harmonics Measurement by Sampler Part 2

A Comparison Between MIL-STD and Commercial EMC Requirements Part 2. By Vincent W. Greb President, EMC Integrity, Inc.

Announcements : Wireless Networks Lecture 3: Physical Layer. Bird s Eye View. Outline. Page 1

Application Note #60 Harmonic Measurement for IEC And other Radiated Immunity Standards

RF EMF Strength Meter

EN :2007+A1:2011 Electromagnetic compatibility Emission standard for residential, commercial and light-industrial environments

The influence of non-audible plural high frequency electrical noise on the playback sound of audio equipment (2 nd report)

National Voluntary Laboratory Accreditation Program

Antenna Measurements using Modulated Signals


Effectively Using the EM 6992 Near Field Probe Kit to Troubleshoot EMI Issues

Powered Traction Unit OPERATION MANUAL

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2011

A GTEM BEST PRACTICE GUIDE APPLYING IEC TO THE USE OF GTEM CELLS

SEMASPEC Provisional Test Method for Evaluating the Electromagnetic Susceptibility of Thermal Mass Flow Controllers

RESEARCH ON METHODS FOR ANALYZING AND PROCESSING SIGNALS USED BY INTERCEPTION SYSTEMS WITH SPECIAL APPLICATIONS

Where Safety Matters, Use The Latest Technology

ETSI EN V1.1.1 ( )

AMPLIFIER RESEARCH... APPLICATION NOTE: 23

Report for Excelsys EMC Measurements for 6Xgen Purchase Order: by Project Number EMT08J027

By order of ZHONGSHAN LIANGYI LIGHTING CO., LTD. at Zhongshan, China

EMC Testing Report. Dual-120CS. Yuan Hsun Electric Co., Ltd.

Mhow (MP) PIN c/o 56 APO RFI : PROCUREMENT OF FAST TRANSIENT RESPONSE ELECTROMAGNETIC PULSE (EMP) SIMULATOR

American National Standard for Methods of Measurement. Frequency allocations and radio treaty matters; general rules and regulations

Chapter 15: Radio-Wave Propagation

Device Detection and Monitoring of Unintentional Radiated Emissions

Improve Performance and Reliability with Flexible, Ultra Robust MEMS Oscillators

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

MEASUREMENTS OF COUPLING THROUGH BRAIDED SHIELD VIA NEW CONDUCTED IMMUNITY TECH- NIQUE

Laboratory Accreditation Programmes

Todd Hubing. Clemson Vehicular Electronics Laboratory Clemson University

Outline / Wireless Networks and Applications Lecture 3: Physical Layer Signals, Modulation, Multiplexing. Cartoon View 1 A Wave of Energy

EMC TEST REPORT. Report No. : EM/2004/10096 Page : 1 of 19

Transcription:

Characterization of medical devices electromagnetic immunity to environmental RF fields. INTRODUCTION The diffusion of personal communication devices and radio communication systems has strongly increased the electromagnetic field levels in environment. These levels can cause interference on sensitive electronic equipments. Personal use medical devices, like hearing aids, pace makers, infusion pumps are particularly interesting, because they may be used in non clinical environments, where EM field levels are not controlled. Electromagnetic interference produced by RF sources on hearing aids and infusion pumps was evaluated. In this work we present the immunity tests carried out on medical devices, in particular on hearing aids. We spend only some words about tests carried out on infusion pumps, since these tests put in evidence that the tested infusion pumps aren't susceptible to electromagnetic field (at least until to -30 V/m electromagnetic fields). All these tests were carried out in a GTEM cell (giga-hertz transverse electromagnetic mode), using signals of intensity and modulation comparable to those present in the environment. The purpose of this work is to characterise the interference, establishing immunity threshold for different frequencies and finding out which types of medical devices are more susceptible, and in which frequency range. METHODS Experimental setup Immunity tests were carried out in a GTEM cell. First working draft for revision of IEC 60118-13 (1999) [1] was utilized as reference for the experimental setup. Experimental setup is given in figure 1. Fig.1 Experimental setup

In order to quantify the immunity level of the hearing aids, IRIL (input related interference level) was used. IRIL is the sound pressure level that, in input at the hearing aid, produces an output pressure level with the same intensity to that of measured signal, when the hearing aid is exposed to an EM field. IRIL is found by subtracting the acoustic gain of the hearing aid at 1 khz from the interference level produced at the hearing aid output, at 1 khz, because of EM field exposure. The sound pressure level produced at the hearing aid output was measured by a phonometer kept out from the cell. An IEC711 2cc coupler and a 1 meter long plastic tubing of 2 mm inner diameter were used to connect the hearing aid to the microphone of phonometer. In particular the tubing was used to keep the coupler and the microphone out from the cell, because they are metallic and could distort the EM field generated inside the cell. The length of the tubing wasn't critical, because the hearing aid gain had to be measured in the test configuration. The EM field was generated by a RF generator and amplified by a power amplifier connected to the cell. The EM field exposure values were measured using a sensor that was placed in the same point where hearing aid was kept during the tests. The tests were conducted with hearing aids in microphone (M) mode of operation. Immunity tests were carried out in a GTEM cell. The experimental setup was the same used for tests on hearing aids, unless the phonometer was replaced by an oscilloscope or by a video camera (but the video camera was put inside the cell), according to the method used to evaluate the electromagnetic immunity of infusion pumps. In some case some parameters could be defined, like the alimentation tension of motor that pilots the cart that pushes the syringe or the cart advancement. The parameter choice depends on the infusion pump kind. When a parameter could not be defined, the pump was monitored with a video camera during the tests, to notice possible malfunctions (such as alarms on display). Exposure conditions 80% amplitude modulated with a 1 khz sine wave signals were used for testing hearing aids (as specified in IEC 61000-4-3 [2]). GSM and DCS signals were also used (although these tests are not provided for by normative in force). Although testing in the frequency range below 800 MHz is not considered necessary by the IEC 60118-13, the hearing aids were also tested on the radio and TV transmission range, because of the high EM field levels that are emitted by radio and TV signals sources and that may be found in the environment. Just AM signals were used, because FM signals don't cause interference on hearing aids (as it was verified). The exposure conditions are given in table 1. The maximum reported test level is subordinated to the performances of the used amplifiers.

Source Frequency range Frequency step size Test levels (V/m) AM Radio Long wave 150-285 khz 5% 1-2-3-5-10-15--25-30 Medium 525-1605 khz 5% 1-2-3-5-10-15--25-30 wave Short wave 2-26 MHz 10% 1-2-3-5-10-15--25-30 TV VHF 47-230 MHz 1% 1-2-3-5-10-15--25-30 UHF 470-862 MHz 1% 1-2-3-5-10-15--25-30 Frequencies in accordance to IEC 60118-13 Wireless digital phone 800-3000MHz 1% 1-2-3-5-10-15--25-30- 35--45-50-55 Table 1. Exposure conditions For each frequency and for each test level, the hearing aid was placed in the reference orientation (hearing aid microphone in front of the RF emitting source) and then rotated in steps of 90 in the horizontal plane around the vertical axis. For each orientation and for each test level, the carrier frequency was changed using the step size given in the table. The IRIL determination was carried out at the orientation where interference was maximum. The same exposure conditions were used, unless the maximum tested level was always 30 V/m and only some frequencies were tested for each frequency range (usually 3), because of the long times necessary to test a infusion pump. Requirements for immunity In order to be considered immune to an EM field, a hearing aid must have an IRIL below 55 db, when the hearing aid is exposed to the field. The EM field levels, which hearing aid should be immune to, are established in the draft version of IEC 60118-13. Those levels are established like E 55 values. E 55 is the unmodulated carrier field value where the hearing aid reaches an IRIL of 55 db. Increasing values of E 55 indicate increasing immunity. In order to be usable with a digital wireless device, a hearing aid has to be immune to 75 V/m EM fields in the 0.8 0.96 MHz frequency range and to 55 V/m EM fields in the 1.4 2 MHz frequency range [2]. CEI EN 60601-2-24 [7] was used as reference: infusion pumps must be immune to electromagnetic field of intensity at least 10V/m. Test procedures For each expositive condition (see table 1) the IRIL was measured and the EM field value where IRIL reached 55 db was determined.

Figure 2 shows the interference levels produced in an analogical hearing aid in M-mode in the 0.8 3 GHz frequency range for different EM field test levels. The E 55 value is 3V/m, since IRIL reaches 55 db in a 5V/m EM field. (e) 80 (d) IRIL (db) 60 (c) (b) (f) (a) 800 1000 10 10 1600 1800 00 20 2600 2800 3000 Frequency (MHz) Figure 2. Sample of hearing aid electromagnetic immunity valuation: interference levels produced in an analogical hearing aid by different values of electrical field; (a) 1 V/m, (b) 3V/m, (c) 5V/m, (d) 10V/m, (e) V/m, (f) limit. For each expositive condition, one or more parameters defined for the pump under test were measured and confronted with the same measured when the pump wasn't exposed to electromagnetic fields, to notice possible differences. Alternately the pump was monitored with a video camera (to notice possible alarms on display). RESULTS E CONCLUSIONS The immunity tests were conducted on 4 analogical and 3 digital hearing aids. The obtained results may be summarized as follows: are generally susceptible to AM and GSM signals, while they are practically immune to unmodulated and FM signals. In fact to difference of the FM signals or the unmodulated ones, that have constant amplitude in the time, amplitude modulation causes interference because the semiconductor junctions in the circuitry of hearing aids demodulate the input signal and produce a noise with the same frequency of the modulating signal. About GSM signal, the TDMA modulation results in the transmitter carrier being switched on and off at a rate of 217 Hz. This in effect causes an AM modulation of the carrier at this frequency. In the same way, therefore, this signal is demodulated by the circuitry of hearing aid and an audible 217 Hz buzz in the signal path is produced. In figure 3 E 55 values are shown for tested hearing aids. They were obtained in accordance with [2], using AM signals: some hearing aids are susceptible to AM signals already for field values of some V/m and just one hearing aid could be

considered usable with a digital wireless device (the hearing aid number 7 that is immune to the maximum tested EM field level). 80 (a) E55 (V/m) 60 (b) (c) 0 1(a) 2(a) 3(a) 4(a) 5(d) 6(d) 7(d) Analogical (a) and digital (d) hearing aids Figure 3. E 55 values and immunity requirements in M-mode; (a) requirements for immunity - 800-960 MHz, (b) maximum tested level, (c) requirements for immunity - 10-00 MHz; E 55 values - 800-960 MHz; E 55 values - 10-00 MHz E 55 medium values are shown in figure 4 for analogical and digital hearing aids for AM and GSM signals. The interference was evaluated at 1 khz, in accordance with [2]. Since AM signals modulating frequency is 1 khz, hearing aids are naturally more susceptible to AM signals than to GSM signals, at 1 khz. Really, while the interference to AM signals is essentially shown to 1 khz, ( and upper harmonics in the case of signal distortions ), the interference to GSM signals expands on all the frequency spectrum, as is shown in figures 5a) and 5b). 60 <E55> 0 800-960 10-00 Frequency range (MHz) Figure 4. E 55 medium values: AM signals and analogical hearing aids, GSM signals and analogical hearing aids; AM signals and digital hearing aids; GSM signals and digital hearing aids

Figure 5. Sound pressure level spectrum in output from the hearing aid: in absence of EM field in a V/m EM field (a: AM b:gsm) E 55 measured values are shown in figures 6 for the 7 tested hearing aids in the TV and radio transmission range: also EM fields in the frequency range below 0.8 GHz may cause interference on hearing aids. However hearing aids are more susceptible to electromagnetic fields in the 0.8-3 GHz range than in the TV and Radio transmission range (fig.3 and fig.7). This stronger interference could be due to the higher frequency. In fact when the frequency rises, the wave length decreases and the electrical connections in the circuitry are more electrically coupled and so more susceptible. 30 (a) E55 (V/m) 10 0 1(a) 2(a) 3(a) 4(a) 5(d) 6(d) 7(d) Figure 6. E 55 values in TV and AM radio transmission range in M-mode; (a) maximum tested level, AM radio, VHF, UHF Digital hearing aids are usually more immune to EM fields than analogical ones. In fact the analogical signal in output from the digital hearing aid microphone is converted to digital signal. So, after the conversion, the signal is naturally less susceptible to external signals.

Although just a hearing aid could be considered usable with a digital wireless device, the most of the tested hearing aids (5 at least) can be considered sufficiently immune to signals of intensity and modulation comparable to those usually present in the environment. 6 different kind infusion pumps were tested: no pumps resulted susceptible to electromagnetic fields. ACKNOWLEDGEMENT This work has been effected in the framework of the programme CNR/ENEA and MIUR Legge 95/95 REFERENCE 1. CEI EN 61000-4-3. Electromagnetic compatibility (EMC). Part 4: Testing and measurement techniques. Section 3: Radiated, radio-frequency, electromagnetic field immunity test (1997) 2. First Working draft for revision of IEC 60118-13:1997. Part 13: Electromagnetic compatibility (EMC) Product standard. TC 29/WG 13 (Ravn) 127 (1999) 3. CEI EN 61000-4-3/A1. Electromagnetic compatibility (EMC). Part 4: Testing and measurement techniques. Section 3: Radiated, radio-frequency, electromagnetic field immunity test (1999) 4. CEI EN 60118-0.. Part 0: Measurement of electro acoustical characteristics (1997) 5. CEI EN 60118-13.. Part 13 Electromagnetic compatibility (1998) 6. CEI EN 60601-1-2. Medical electrical equipment. Part 1: General requirements for safety. 2 Collateral standard: Electromagnetic compatibility Requirements and tests (1998) 7. CEI EN 60601-2-24. Medical electrical equipment. Part 2: Particular requirements for safety for infusion pumps and control devices. (1999)