Basic Communication Laboratory Manual. Shimshon Levy&Harael Mualem

Similar documents
Elizabethtown College Department of Physics and Engineering PHY104. Lab # 9- Oscilloscope and RC Circuit

Combinational logic: Breadboard adders

ME 365 EXPERIMENT 1 FAMILIARIZATION WITH COMMONLY USED INSTRUMENTATION

What the LSA1000 Does and How

EE 201 Function / Arbitrary Waveform Generator and Oscilloscope Tutorial

Experiment 1: Instrument Familiarization (8/28/06)

Experiment 1: Instrument Familiarization

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

UCE-DSO212 DIGITAL OSCILLOSCOPE USER MANUAL. UCORE ELECTRONICS

Sampling and Reconstruction

2 Oscilloscope Familiarization

EET 223 RF COMMUNICATIONS LABORATORY EXPERIMENTS

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

Introduction to Basic Laboratory Instruments

New Features of IEEE Std Digitizing Waveform Recorders

The University of Jordan Mechatronics Engineering Department Electronics Lab.( ) Experiment 1: Lab Equipment Familiarization

EXPERIMENT NUMBER 2 BASIC OSCILLOSCOPE OPERATIONS

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

The Oscilloscope. Vision is the art of seeing things invisible. J. Swift ( ) OBJECTIVE To learn to operate a digital oscilloscope.

ENGR 210 Lab 6 Use of the Function Generator & Oscilloscope

The oscilloscope and RC filters

Agilent 81133A/81134A

Exercise 2: FM Detection With a PLL

Laboratory Exercise 6 THE OSCILLOSCOPE

Sonoma State University Department of Engineering Science Spring 2017

CHAPTER 6. Motor Driver

AP034-OM-E Rev D ISSUED: January 2000 ²

Superheterodyne Spectrum Analyzer and Spectrum Analysis. Shimshon Levy&Harel Mualem

Name Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits

Multiple Instrument Station Module

U1604A Handheld Oscilloscopes, 40 MHz

B. Equipment. Advanced Lab

Frequency and Time Domain Representation of Sinusoidal Signals

Model Hz to 10MHz Precision Phasemeter. Operating Manual

332:223 Principles of Electrical Engineering I Laboratory Experiment #2 Title: Function Generators and Oscilloscopes Suggested Equipment:

EE-4022 Experiment 2 Amplitude Modulation (AM)

OPERATING AND MAINTENANCE MANUAL

ArbStudio Arbitrary Waveform Generators

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1

Laboratory 3 (drawn from lab text by Alciatore)

U1571A Ni-MH Battery Pack for U1600A Handheld Oscilloscopes

SIGNAL RECOVERY. Model 7265 DSP Lock-in Amplifier

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

4. Digital Measurement of Electrical Quantities

EE EXPERIMENT 1 (2 DAYS) BASIC OSCILLOSCOPE OPERATIONS INTRODUCTION DAY 1

Equipment: You will use the bench power supply, function generator and oscilloscope.

Appendix A: Specifications

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Introduction to basic laboratory instruments

Introduction to Lab Instruments

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

LABORATORY 4. Palomar College ENGR210 Spring 2017 ASSIGNED: 3/21/17

MODELS WW5061/2. 50MS/s Single/Dual Channel Arbitrary Waveform Generators

LAB I. INTRODUCTION TO LAB EQUIPMENT

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

Experiment #2: Introduction to Lab Equipment: Function Generator, Oscilloscope, and Multisim

DS 6000 Specifications

Characterizing High-Speed Oscilloscope Distortion A comparison of Agilent and Tektronix high-speed, real-time oscilloscopes

Model 745 Series. Berkeley Nucleonics Test, Measurement and Nuclear Instrumentation since Model 845-HP Datasheet BNC

Chapter 2 Analog-to-Digital Conversion...

Lab 0: Orientation. 1 Introduction: Oscilloscope. Refer to Appendix E for photos of the apparatus

AME140 Lab #2 INTRODUCTION TO ELECTRONIC TEST EQUIPMENT AND BASIC ELECTRONICS MEASUREMENTS

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term...

User Manual Series. Digital Storage Oscilloscope 6810, 6806, March Copyright Protek Test & Measurement 2005 All Rights Reserved

Lab 1: Basic Lab Equipment and Measurements

LAB I. INTRODUCTION TO LAB EQUIPMENT

ECE 231 Laboratory Exercise 3 Oscilloscope/Function-Generator Operation ECE 231 Laboratory Exercise 3 Oscilloscope/Function Generator Operation

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

Transient Data Acquisition System, TAS 4-40 Potential-free measurement of fast rise pulses:

Precalculations Individual Portion Introductory Lab: Basic Operation of Common Laboratory Instruments

EE ELECTRICAL ENGINEERING AND INSTRUMENTATION

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

Keysight Technologies E8257D PSG Microwave Analog Signal Generator. Data Sheet

N acquisitions, all channels simultaneously, N is selectable from 4, 16, 64, and 128 Inputs

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives:

781/ /

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

Oscilloscope Measurements

Notes on Experiment #1

Cleverscope Model CS320A - CS328A Data Sheet

Measurement Procedure & Test Equipment Used

SIGNAL GENERATORS. MG3633A 10 khz to 2700 MHz SYNTHESIZED SIGNAL GENERATOR GPIB

Table of Contents...2. About the Tutorial...6. Audience...6. Prerequisites...6. Copyright & Disclaimer EMI INTRODUCTION Voltmeter...

Fourier Theory & Practice, Part II: Practice Operating the Agilent Series Scope with Measurement/Storage Module

Analog Arts SL987 SL957 SL937 SL917 Product Specifications [1]

HAMEG Programmable Measuring Instruments Series 8100

How to Setup a Real-time Oscilloscope to Measure Jitter

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

ECE 440L. Experiment 1: Signals and Noise (1 week)

LeCroy 9304A, 9304AM Digital Oscilloscopes 200 MHz Bandwidth, 100 MS/s. Main Features

Department of Electronic Engineering NED University of Engineering & Technology. LABORATORY WORKBOOK For the Course SIGNALS & SYSTEMS (TC-202)

Moku:Lab. Specifications INSTRUMENTS. Moku:Lab, rev

INTRODUCTION TO ENGINEERING AND LABORATORY EXPERIENCE Spring, 2015

Introduction to Oscilloscopes Instructor s Guide

EECS 318 Electronics Lab Laboratory #2 Electronic Test Equipment

Arbitrary/Function Waveform Generators 4075B Series

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope

Department of Electrical and Computer Engineering. Laboratory Experiment 1. Function Generator and Oscilloscope

Transcription:

Basic Communication Laboratory Manual Shimshon Levy&Harael Mualem September 2006

CONTENTS 1 The oscilloscope 2 1.1 Objectives... 2 1.2 Prelab... 2 1.3 Background Theory- Analog Oscilloscope...... 3 1.4 Digital Oscilloscopes- Coupling and input section... 3 1.5 SamplingMethods... 5 1.6 Real-timeSamplingwithInterpolation... 5 1.7 Equivalent-timeSampling... 6 1.8 XYDisplayMode... 6 1.9 AdvancedMeasurement-TimingJitter... 8 1.10 DefiningShort-Term:Jittervs.Wander... 8 1.11THD-TotalHarmonicDistortionandDistortionAnalyzer... 8 2 Experiment Procedure 11 2.1 RequiredEquipment... 11 2.2 Bandwidth... 11 2.3 InputImpedance... 12 2.4 BandwidthandImpedanceLimitation... 13 2.5 Input coupling... 13 2.6 InstantaneousandaveragePower... 14 2.7 DistortionMeasurement... 14 2.8 Jitter... 15 2.9 FinalReport... 15 2.10Appendix-1... 16 2.10.1 Tophaselocktwofunctiongenerator.... 16 2.10.2 Setting a Zero phase reference at the end of the cable...... 17

Chapter 1 THE OSCILLOSCOPE 1.1 Objectives This experiment deals with basic and advanced time domain measurement, and. The purpose of the laboratory is to develop the skill needed to use basic laboratory equipment, to characterize basic electronic circuits, which demonstrate the real capability and limitation of time domain measurement. Upon completion of the experiment the student will: Understand the function of each main block of the oscilloscope. Making basic and advanced measurements with a oscilloscope. Know- the limitations of the oscilloscope. Possess the necessary tools to evaluate signals in time domain. 1.2 Prelab 1. Find the dc value (average value) and rms value of sawtooth signal as indicated in figure -1 2. Using Matlab, draw a graph of trianglewave, frequency 100k Hz, amplitude 1V p,in time domain and frequency domain (magnitude only), using the data calculate the THD and SINAD(dB) of the trianglewave. (see figure-2) amplitude(5v) 1 2 3 4 time(ms) Figure 1 Problem-1 sawtooth signal

Background Theory- Analog Oscilloscope 3 amplitude(5v) time Figure 2 Problem-2 trianglewave 1.3 Background Theory- Analog Oscilloscope In order to use an analog oscilloscope, you need to adjust three basic settings to accommodate an incoming signal: * The time base- This sets the amount of time per division (usually written as time/div or sec/div) represented horizontally across the screen. The horizontal scale control on your oscilloscope lets you adjust the time base. * The attenuation or amplification of the signal- The vertical scale control lets you adjust the amplitude of the signal before it is applied to the vertical deflection system. * The triggering of the oscilloscope- This control lets you stabilize a repeating signal, as well as triggering on a single event. 1.4 Digital Oscilloscopes- Coupling and input section Digital Oscilloscopes uses the same system as those in analog oscilloscopes; however, digital oscilloscopes contain additional data processing capability systems. (See Figure 2.) Digital oscilloscope collects data for the entire waveform and then displays the incoming signal. When you attach a probe to a circuit, the vertical system coupled to the signal and adjusts the amplitude of the signal using attenuator or amplifier as necessary. Coupling means the way used to connect an electrical signal from one circuit to the another. in our case the coupling can be set to DC, AC or ground. DC coupling connect the signal directly to the vertical system and shows all of an input signal components (see Fig.-4). AC coupling connect the input signal through capacitor, therefore capacitor blocks the DC component of a signal (see Fig.-3), so that you see the waveform centered at zero volts.next, the analog-to-digital converter (ADC) in the acquisition system samples the signal at discrete points in time and converts the signal s voltage at these points to digital values called sample points, and store the data at channel memory. The horizontal system s time base clock determines the rate of the A/D converter sample. The rate at which the clock ticks is called the sample rate, and is measured in samples per second. Sampling methods used usually fall into two types:

Digital Oscilloscopes- Coupling and input section 4 Acquisition System Processing Attenuator Fast A/D Converter Channel Memory Vertical Amplifier Input Trigger System Time Base Clock Horizontal System CRT Display Figure 3 Major blocks of digital oscilloscope 1 0 10 0 Amp(v) Amp(v) 10v DC+ 1V pp AC AC coupling Time 10v DC+ 1V pp AC DC coupling Time Figure 4 AC and Dc coupling of composite signal

Sampling Methods 5 * Real-time sampling. * Equivalent time sampling. ThesamplepointsfromtheA/Dconverterarestoredinmemoryaswaveform points. More than one sample point may make up one waveform point. Together, all the stored waveform points make up one waveform record. The number of waveform points used to make a waveform record is called the record length. The trigger system determines the start and stop points of the record. The display receives these record points after being stored in memory. Depending on the capabilities of your oscilloscope, additional processing of the sample points may take place, enhancing the display. Pre-trigger may be available, allowing you to see events before the trigger point. With a digital oscilloscope, you need to adjust these main control settings to take a measurement. * The time base. This sets the amount of time represented by the width of the screen. The horizontal scale control (Time/Div) on the oscilloscope lets you adjust thetimebase. * The attenuation or amplification of the input signal. The vertical scale control (Volt/Div) lets you adjust the amplitude of the signal before sending it to the Analog to Digital Converter. * The triggering. This control (Trigger) mark the system, the process of the acquisition signal, stabilize a repeating signal, capture a single-shot signal, * The acquisition mode and sampling method. These settings depend on the frequency and complexity of the signal you are trying to measure. Front panel controls let you adjust these settings. 1.5 Sampling Methods The sampling method tells the oscilloscope how to collect sample points. At low frequencies a digital oscilloscope easily collects more than enough sample points to construct an accurate picture of the original signal. However, at higher frequencies (how high depends on the oscilloscope s maximum sample rate) the oscilloscope cannot collect enough samples. The digital oscilloscope can do two things: * It can collect and build a picture of the waveform, over several period over time, as long as the signal is repetitive. (equivalent - time sampling mode). * It can collect a few sample points of the signal in a single pass ( about 10 points in real-time sampling mode) and then use interpolation. Interpolation is a processing algorithm to estimate the shape of the waveform looks like based on a sampled points. 1.6 Real-time Sampling with Interpolation In real-time sampling mode, the oscilloscope collects as many samples as it determine to reconstruct the signal, depending

Equivalent-time Sampling 6 Time Figure 5 Real time sampling on the duration shape and the amount of memory available for the channel (see Figure 5).For a single-shot or transient signals user must choose real time sampling mode. Interpolation is the way that the oscilloscope use to connects the dots. Linear interpolation connects sample points with straight lines, and sine interpolation connects sample points with curves (see Figure 6).Digital oscilloscope use linear interpolation for pulses and sine interpolation for sinusoidal waves as a default. 1.7 Equivalent-time Sampling Digital oscilloscope use equivalent-time sampling for a very fast repeating signal (when the time base scale is about 500 ns or faster). Equivalent-time sampling extends the useful frequency range of an oscilloscope. Equivalent-time sampling lets you construct a picture of a repetitive signal by capturing a little bit of information from each cycle (see Figure 7). 1.8 XY Display Mode The normal display of the oscilloscope is volts versus time. The XY mode convert the display to a volt versus volts display. You can use various application such as volts versus current or another two physical relating phenomena using transducers. If you have two sinusoidal wave with the same frequency, but with arbitrary phase, one method of measuring phase shift is to use the XY mode. This involves in applying one signal into the horizontal system and another signal to vertical system. The waveform resulting from this arrangement, is called Lissajous pattern. From the shape of the Lissajous pattern, you can tell the phase difference between the two signals (see figure -8).

XY Display Mode 7 Linear interpolation Amplitude Sinewave interpolation Time Figure 6 Linear and sinewave interpolation First acuisition Second acuisition Third acuisition Fourth acuisition Time Figure 7 Equivalent time sampling

Advanced Measurement-Timing Jitter 8 A B sinθ = A B 0 90 180 270 Figure 8 XY display of two sinewave, with equal frequency, and different phase difference 1.9 Advanced Measurement-Timing Jitter What Is Jitter? one definition of jitter is "The slight movement of a transmission signal in time or phase that can introduce errors and loss of synchronization. More jitter will be encountered with longer cables, cables with higher attenuation, and signals at higher data rates. Jitter lead to, timing distortion, or intersymbol interference. other definition of timing jitter is Jitter is defined as the short-term variations (instability) of a digital signal s significant instants from their ideal positions in time. This captures the essence of jitter, but some of the individual terms (shortterm, significant instants, ideal positions) need to be clarified before this definition can be unambiguously used. In all real applications, jitter has a random component, so it must be specified using statistical terms. Metricssuchasmeanvalue, standard deviation, and other qualifiers such as confidence interval, must be used to establish meaningful, repeatable measurements. 1.10 Defining Short-Term: Jitter vs. Wander By convention, timing variations are split into two categories, called jitter and wander, based on a Fourier analysis of the variations vs. time. Timing variations that occur slowly (typically minutes or hours) are called wander. Jitter describes timing variations that occur more rapidly (scale of seconds). The threshold between wander and jitter is definedtobe10hzaccording to the ITU2 but other definitions may be encountered. In many cases wander is of little or no consequence on serial communications links, where a clock recovery circuit effectively eliminates it. 1.11 THD- Total Harmonic Distortion and Distortion Analyzer Distortion analyzers measure the distortion produced in the Device Under Test (DUT). Basically, the distortion analyzer applies a very pure sine wave (with very low har-

THD- Total Harmonic Distortion and Distortion Analyzer 9 monics) to the input of a circuit and measures the resulting output. If the circuit is distortion free, the output will also be a pure sine wave. However, in real circuits, some distortion may be produced (typically harmonics). The distortion Analyzer determines the amount of distortion in the output by filtering out the original sine wave and measuring the RMS voltage of all residual signals, including harmonics, spurious responses and noise at defined bandwidth. Figure 9 shows the simplified block diagram of a distortion analyzer. The sine wave output is connected to the circuit under test. This sine wave must be very pure, its distortion will limit the accuracy of the instrument. The output of the circuit under test is connected to the input of the distortion analyzer. The analyzer uses a narrow sharp notch filter to remove the original sine wave from the input signal. Anything left over is assumed to be distortion in the signal caused by the DUT. The distortion is then compared to the waveform, including the fundamental sine wave, and the distortion is read out on the analyzer as a percentage of the original signal. It is important to note that noise and spurious responses present in the circuit will also contribute to the distortion reading, since they will not be removed by the notch filter. Most distortion analyzers include selectable low-pass and high-pass filtering to help eliminate these unwanted signals. Advanced distortion analyzers provide the capability of selecting the test frequency. These instruments have a tunable notch filter over the measured bandwidth which automatically tracks the frequency of the internal oscillator. The purity of the measured signal may be expressed in several different ways. For audio system, distortion (either in percent or db relative to the fundamental) is defined as distortion (%) = 100 V RMS of all signal without fundametal V RMS of all signal alternative expression for distortion is distortion (%) = 100 V RMS of all signal without fundametal V RMS of fundamental signal Another alternative is expressing the imperfections in terms of either SINAD V RMS of (signal + noise + distortion) SINAD (db) =20log 10 V RMS of (noise + distortion) SINAD measurements are often used to measure the sensitivity of a radio receiver. A signal generator with a 1-kHz modulating tone is connected to the antenna connection of the receiver. The signal generator s frequency is set to match the receiver frequency. A distortion meter is connected to the audio output of the receiver where it measures the SINAD of the recovered audio. The amplitude of the signal generator is reduced until a particular SINAD value is reached, typically 12dB. The signal generator amplitude at this point defines the sensitivity of the receiver.

THD- Total Harmonic Distortion and Distortion Analyzer 10 Pure sinewave oscillator Device under Test Notch filter RMS voltmeter Figure 9 Major blocks of Distortion Analyzer

Chapter 2 EXPERIMENT PROCEDURE 2.1 Required Equipment 1. Oscilloscope -Agilent 54621A 2. Function generator- Agilent 33120A 3. Signal generator -Agilent 8647B or equivalent. 4. 50 Ohm feedthrough. 5. 100kHz notch filter. 2.2 Bandwidth In this part of the experiment you will understand the meaning of the oscilloscope Bandwidth, (60 MHz or 100 MHz depend on your oscilloscope) by measuring the oscilloscope bandwidth using signal generator which have constant amplitude across all the frequency range, and 50Ω feedthrough termination to achieve correct termination. 1. Connect the signal generator using 50Ω feedthrough termination to the oscilloscope, as indicated in Figure-1. 2. Set the frequency of the signal generator to 250 khz, amplitude 800mV. Signal generator HP-8647A 55.000, MHz 50 ohm Feedthrough Figure 1 Bandwidth measuring system

Input Impedance 12 15.000,000 MHz 100κΩ BNC connectors 100κΩ Voltage divider Figure 2 Input impedance affects on measurement 3. Press Autoscale bottom on the oscilloscope, adjust the output of the signal generator to exceed exactly 8 vertical division (record the value O. db). 4. Observe the display of the oscilloscope, while changing the frequency of the signal generator, and the time base by steps to 60 MHz (100 MHz), and 5 ns/div respectively. 5. Verify that the amplitude of the signal is within ±3dB at 60 MHz (100 MHz). (record the value). 6. Continue to change the frequency according to the table below, up to the one that you got a noisy waveform (record the value), explain those three records. Frequency 250kHz 20MHz 40MHz 80MHz 110MHz 140MHz V m db 0-3 2.3 Input Impedance The input impedance of the oscilloscope is 1MΩ, parallel with 13pf. In some cases such impedance could load the DUT and cause a measurement error. 1. Connect the function generator directly to the oscilloscope. 2. Set the function generator to sinewave 1 khz, amplitude 1V pp. 3. Compare the amplitude reading of the oscilloscope and function generator? what is the ratio and why? 4. Change the signal to squarewave and triangle wave and measure V pp and V RMS of each signal, compare the results to prelab exercise. Store the data on magnetic media. 5. Connect the divider to the function generator, as indicated in Figure-2 6. Set the function generator to sinewave 1kHz, amplitude 1V pp. 7. Verify that the Voltage across each resistor is about 1V pp (why the voltage remain 1V pp after the division).

Bandwidth and Impedance Limitation 13 8. Change the frequency of the function generator to 1MHz, measure the voltage on each resistor, and the source, why the sum of the two voltage is less then the displayed voltage source, storethevoltageoftheresistorandthesource on magnetic media. 9. Change the frequency to 3MHz and explain why the signal change to triangle wave, what happen to the amplitude of the signal? Storethevoltageof one resistor on magnetic media. 2.4 Bandwidth and Impedance Limitation Sometime you asked to measure the rise time of a single or repetitive pulse. Such measurement could be meaningless if you don t pay attention to the oscilloscope bandwidth and the rise time of the oscilloscope. 1. Connect the function generator using 50Ω feedthrough to the oscilloscope, asindicatedinfigure-1 2. Set the function generator to squarewave frequency 15 MHz, amplitude 1V pp. 3. Watch the signal at the oscilloscope, try to explain why the signal looks like a sinewave, reduce the frequency by step to 1MHz and verify your answer. 4. change the signal to squarewave 1MHz and measure the rise time of the square wave signal. According to the specification the rise time is 20 ns. Store the data on magnetic media. 5. Disconnect the 50Ω feedthrough termination, measure again the rise time, whathappentotherisetimeofthesignal,explain. Store the data on magnetic media. 2.5 Input coupling Sometimes you have to measure small AC signal in a presence of a large DC signal e.g. power supply ripple and noise measurement. In such a case using AC coupling enable the measurement. In our case we have to measure 20 mv pp sinewave in the presence of a 5 VDC. 1. Connect the function generators to the oscilloscope, as indicated in Figure-3 2. Set one of the function generator to 5VDC. 3. Set the second function generator to sinewave 1 khz, amplitude 40mV. 4. Press Autoscale (DC coupling) explain why the AC signal disappears. 5. Measure V RMS of the signal, record your measurement, replace the oscilloscope with digital multimeter, set the multimeter to VAC, why the two results are so different? if you set the multimeter to VDC the results will be approximately the same explain?

Instantaneous and average Power. 14 15.000,000 MHz 15.000,000 MHz Figure 3 6. Reconnect the oscilloscope and change the coupling of the oscilloscope to AC and measure the amplitude (V pp ) of the sinewave signal. 2.6 Instantaneous and average Power. In this section you multiply two sinusoid signal e.g. voltage and current, and the multiply signal is the power of the system. 1. Connect two function generators to the oscilloscope, as indicated in Fig-3 2. Set the two function generators to sinewave 1V pp, frequency 10kHz. 3. Lock the two function generators to thesamephaseaccordingtotheprocedure of Appendix-1. 4. Multiply the two channels, by pressing Math button. Operate the average function if necessary to stabilize the multiplication signal (instantaneous power). 5. Turn off the Math button and measure the amplitude of each signal.(two function generators). 6. Turn on the Math button and measure the amplitude peak to peak, and the frequency of the third signal (use cursors)compare the result to the theory. 7. Set the Y cursor at the minimum point of the signal (assume that is the point of the ground). Change the phase of one signal continuously, what happen to the signal? explain. 2.7 Distortion Measurement In this part of the experiment, we measure the Total Harmonic Distortion (THD) of 100kHz triangle wave signal. (assume that the triangle wave is the distorted output of a system, where the input is a pure sinewave) 1. Connect function generators to the oscilloscope, as indicated in Fig-4 2. Set the function generators to trianglewave 1V pp, frequency 100kHz. 3. Measure the RMS value of the harmonics.

Jitter 15 15.000,000 MHz 100kHz Notch filter Figure 4 4. Disconnect the notch filter and measure the RMS value of the signal+noise +distortion.. 5. Calculate the THD and SINAD(dB) of the trianglewave signal, compare your result to the prelab results. 2.8 Jitter In this part of the experiment, we measure the jitter (time domain instability) of 1kHz squarewave signal. 1. connect the function generator directly to the oscilloscope. 2. Set the function generator to squarewave, frequency 1kHz, amplitude 100mv. 3. Set the oscilloscope to infinite persistence by pressing Display Persist,and set the trigger to Slope Falling, by pressing the proper keys of the trigger. 4. Set the rising edge of the signal at the center of the oscilloscope, and change the time base to 50 ns/div, the jitter of the squarewave will be displayed. 5. Measure peak to peak jitter using proper cursor (see Fig.-5) Save the data on magnetic media. 2.9 Final Report 1. Explain why you need to add 50Ω feedthrough termination, when you measured the bandwidth of the oscilloscope, what will happen if you connect the signal without the feedthrough termination. 2. Using Matlab or other mathematics software: a. Expand 10 MHz squarewave to Fourier series. b. Choose the first four odd harmonics, add them and draw a graph (voltage versus time), and compare the graph to the graph of your result (time domain and rise time).

Appendix-1 16 Figure 5 A jitter 1kHz squarewave signal, time base set to 50ns/div 3. Explain why the rise time measurement change, when you disconnect the feedthrough termination, prove your answer using graphs. 4. Explain why Kirchof voltage law fail when you use the voltage divider? Prove your answer using the stored graphs. 2.10 Appendix-1 2.10.1 To phase lock two function generator. 1. Connect rear-pane Ref Out 10MHz output terminal of the master Arbitrary Waveform Generator HP-33120A to Ext Ref in on the rear panel of the slave HP-33120A asindicatedinfigure-6 2. Connect the two AWG s to the oscilloscope as shown in Figure-6 3. set the oscilloscope to standard display-arbitrary phase between the two channel, by pressing AUTOSCALE. 4. Turn on the menu of the AWG by pressing shift menu on/off the display then looks like A:MOD MENU. 5. Move across to G:PHASE MENU by pressing the < button. 6. Move down one level to the ADJUST command, by pressing, the display looks like 1:ADJUST 7. press one level and set the phase offset, change the phase continuously between the two AWG s until one trace cover completely the other trace (zero phase). You see then a display like 120.000DEG.

Appendix-1 17 Ref Out Ext Ref in Ref Out Ext Ref in Rear panel connection. 15.000,000 MHz HP-33120A 15.000,000 MHz Oscilloscope 54600A Oscilloscope display Channel-1 Channel-2 Figure 6 8. Turn off the menu by pressing ENTER. You have the exited the menu. 2.10.2 Setting a Zero phase reference at the end of the cable. 1. Turn on the menu by pressing shift menu On/off the display looks like A:MOD MENU. 2. Move across to the PHASE MENU choice on this level, by pressing <, the display looks like G:PHASE MENU. 3. Move down one level and then across to the SET ZERO by pressing and > bottoms, the display shows the message 2:SET ZERO. 4. Move down a level to set the zero phase reference, by pressing the displayed message indicates PHASE=0. 5. Press ENTER, save the phase reference and turn off the menu. IMPORTANT : At this point; the function generator HP-33120A is phase locked to another HP-33120A or external clock signal with the specified phase relationship. The two signals remain locked unless you change the output frequency.