Gas proportional scintillation counter pulse-signature analysis using digital techniques

Similar documents
nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

A digital method for separation and reconstruction of pile-up events in germanium detectors. Abstract

Digital coincidence acquisition applied to portable β liquid scintillation counting device

nanodpp datasheet I. FEATURES

Digital trigger system for the RED-100 detector based on the unit in VME standard

nanomca-sp datasheet I. FEATURES

Electronic Instrumentation for Radiation Detection Systems

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

nanomca datasheet I. FEATURES

A PC104 Multiprocessor DSP System for Radiation Spectroscopy Applications

A high energy gamma camera using a multiple hole collimator

Digital Signal Processing for HPGe Detectors

X-ray polarimetry with a conventional gas proportional counter through rise-time analysis

CAEN Tools for Discovery

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS

Method for digital particle spectrometry Khryachkov Vitaly

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

Chapter 6 Pulse Processing

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

A user-friendly fully digital TDPAC-spectrometer

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS*

PROGRESS in TOF PET timing resolution continues to

nanomca-ii-sp datasheet

Development of New Peak Detection method for Nuclear Spectroscopy

The Absolute Detection Efficiency of Vacuum Electron Multipliers, to kev Protons and Ar + Ions.

Simulation of Algorithms for Pulse Timing in FPGAs

Amptek Silicon Drift Diode (SDD) at High Count Rates

arxiv:hep-ex/ v1 8 Jul 1999

Prototype of a Compact Imaging System for GEM Detectors Tomohisa Uchida, Member, IEEE, Yowichi Fujita, Manobu Tanaka, Member, IEEE, and Shoji Uno

Performance Assessment of Pixelated LaBr 3 Detector Modules for TOF PET

A novel type epithermal neutron radiography detecting and imaging system

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

arxiv: v2 [physics.ins-det] 17 Oct 2015

PMT Calibration in the XENON 1T Demonstrator. Abstract

Solid State Photomultiplier: Noise Parameters of Photodetectors with Internal Discrete Amplification

The domino sampling chip: a 1.2 GHz waveform sampling CMOS chip

R&D on CsI(Tl) crystals + LAAPD at USC Activities report

PX4 Frequently Asked Questions (FAQ)

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Positron Emission Tomography

Amptek Inc. Page 1 of 7

MAROC: Multi-Anode ReadOut Chip for MaPMTs

SILICON DRIFT DETECTORS (SDDs) [1] with integrated. Preliminary Results on Compton Electrons in Silicon Drift Detector

High collection efficiency MCPs for photon counting detectors

K 223 Angular Correlation

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s

Week 11: Chap. 16b Pulse Shaping

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

MWPC Gas Gain with Argon-CO 2 80:20 Gas Mixture

XRF Instrumentation. Introduction to spectrometer

EMC Pulse Measurements

50 MHz Voltage-to-Frequency Converter

Thick GEM versus thin GEM in two-phase argon avalanche detectors

An ASIC dedicated to the RPCs front-end. of the dimuon arm trigger in the ALICE experiment.

Traditional analog QDC chain and Digital Pulse Processing [1]

A CMOS INTEGRATED CIRCUIT FOR PULSE-SHAPE DISCRIMINATION*

Charge Sharing Effect on 600 µm Pitch Pixelated CZT Detector for Imaging Applications *

Performance measurements of a depth-encoding PET detector module based on positionsensitive

Testing the Electronics for the MicroBooNE Light Collection System

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

FPGA-BASED CONTROL SYSTEM OF AN ULTRASONIC PHASED ARRAY

Investigation of low noise, low cost readout electronics for high sensitivity PET systems based on Avalanche Photodiode arrays

ORTEC. High-Count-Rate Spectroscopy with Ge Detectors: Quantitative Evaluation of the Performance of High-Rate Systems 1. I.

Final Results from the APV25 Production Wafer Testing

PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

Atomic and Nuclear Physics

The Liquid Xenon Detector for PET: Recent Results

Analysis and performance Evaluation of trapezoidal filter in pulse shaping

Analysis of 1=f Noise in CMOS Preamplifier With CDS Circuit

Design of the Front-End Readout Electronics for ATLAS Tile Calorimeter at the slhc

AN-DPP-003 Rev A2: Using the DP5 with HPGe USING THE DP5 WITH GERMANIUM DETECTORS

Average energy lost per unit distance traveled by a fast moving charged particle is given by the Bethe-Bloch function

X-ray Detectors: What are the Needs?

Evaluating the Performance of a Commercial Silicon Drift Detector for X-ray Microanalysis

Goal of the project. TPC operation. Raw data. Calibration

THE USE OF CdTe DETECTORS FOR DENTAL X-RAY SPECTROMETRY

PCS-150 / PCI-200 High Speed Boxcar Modules

Front-End Electronics and Feature-Extraction Algorithm for the PANDA Electromagnetic Calorimeter

Pixel hybrid photon detectors

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

M. K. Schultz, R. M. Keyser, R. C. Trammell, and D. L. Upp

Initial Results from a Cryogenic Proton Irradiation of a p-channel CCD

Measurement of X-ray Photon Energy and Arrival Time Using a Silicon Drift Detector

LIFETIME OF THE MUON

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

PoS(ICRC2017)449. First results from the AugerPrime engineering array

An aging study ofa MICROMEGAS with GEM preamplification

This is a repository copy of A two-phase argon avalanche detector operated in a single electron counting mode.

The LUX Experiment Trigger and Data Acquisition Systems. Eryk Druszkiewicz April 15 th 2013

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

DSA-LX. Digital Signal Analyzer. Radiation Safety. Amplified.

Transcription:

Nuclear Instruments and Methods in Physics Research A 422 (1999) 341 346 Gas proportional scintillation counter pulse-signature analysis using digital techniques P.C.P.S. Simo es *, J.M.F. dos Santos, C.A.N. Conde Departamento de Fı&sica, F.C.T.U.C., Universidade de Coimbra, P-3000 Coimbra, Portugal Departamento de Biofı&sica e Biomatema& tica, Faculdade de Medicina, Universidade de Coimbra, P-3000 Coimbra, Portugal Abstract The use of short shaping time-constants (50 ns) followed by digital sampling of gas proportional scintillation counter pulses is investigated. Such short shaping time causes each pulse to closely resemble the scintillation light burst, allowing for an effective pulse-signature analysis. Pulse amplitude is obtained by numerical integration of the digital pulse samples. The shorter total pulse durations, lead to pulse-height distributions that are much less dependent on the total interaction rate occurring in the detector. Total interaction rates up to 12 k s can be tolerated without significant degradation of the obtained pulse-height distribution. It was also verified that the maximum throughput achieved by the digital pulse-height analyser improves from 1 to 1.7 k s when using the short shaping time-constants. 1999 Elsevier Science B.V. All rights reserved. 1. Introduction * Correspondence address: Departamento de Física, F.C.T.U.C., Universidade de Coimbra, P-3000 Coimbra, Portugal. Gas proportional scintillation counters (GPSC) have been widely applied in important areas such as X-ray astronomy and energy dispersive X-ray fluorescence analysis [1 3]. They combine room temperature operation with large detection area and high counting rates without space charge effects [4 6]. The interest in GPSCs has led to a continued research effort in understanding its working principles and in improving its performance. Digital signal processing developments have been mostly oriented to nuclear spectrometry for solid-state detector applications. However, the application of these techniques to the analysis of GPSC pulses proved to be a simple but effective method to improve the detector performance in X-ray spectrometry applications, through background reduction and peak enhancement. Recent studies investigated the applications of digital pulse processing techniques to X-ray spectrometry with GPSCs [7 9] using a digital pulse-height analyser (DPHA). In GPSCs applications long pulse shape timeconstants are routinely used (typically 5 μs for both integration and differentiation time-constants) for best performance. Pulse integration was performed directly by the main amplifier [7,8] or else digitally by the DPHA [9]. X-ray pulse-height distributions were improved on the basis of digital pulse risetime discrimination [7 9]. 0168-9002/99/$ see front matter 1999 Elsevier Science B.V. All rights reserved. PII: S 0 1 6 8-9 0 0 2 ( 9 8 ) 0 0 9 7 8-4 IV. INFORMATION PROCESSING

342 P.C.P.S. Simo es et al./nucl. Instr. and Meth. in Phys. Res. A 422 (1999) 341 346 In the present work we investigate the advantages of using very short time-constants (50 ns) in the main amplifier. Because of the short time-constants the pulse shape will closely resemble the scintillation light burst. This allows pulse-signature identification, an advantage in applications where pulse shape discrimination of multiple events is needed [10,11]. On the other hand, since the total pulse duration is shorter, the obtained pulse-height distributions will be much less dependent on the interaction rates. This will present some advantages in GPSC applications where high interaction rates are involved, such as in X-ray solar flares, synchrotron radiation and X-ray structure analysis [3,6]. 2. Results and discussion 2.1. Experimental set-up The schematic diagram of the uniform field GPSC used in this work is presented in Fig. 1. The detector has a 4 cm deep absorption region and a 1 cm deep scintillation region and it is described in detail in Ref. [12]. The GPSC signals are fed to the DPHA through a HP5554A preamplifier and a HP5582A main amplifier using 50 ns integration and differentiation time-constants. The short differentiation time constant causes the pulse shape to follow the scintillation light burst, while the short integration timeconstant eliminates high-frequency noise. Each pulse is digitised at a 20 MHz rate with 8-bit resolution and processed by the DPHA [7]. In a typical run, each pulse is processed by a series of algorithms as follows. The digitised pulse samples are smoothed by a 3-point median filter to reduce high-frequency oscillation [13]. The pulse amplitude is obtained by numerical integration of the pulse samples. Pulse duration is taken as the time that the pulse spends above a threshold value set just above noise level. A pulse is rejected whenever a sampled value overshoots the ADC scale. Pulse discrimination is now performed on the basis of pulse duration. As previously [7], partial pulse-height distributions discriminated according Fig. 1. Schematic diagram of the GPSC used in this work. to different pulse duration can be simultaneously acquired. 2.2. Pulse-signature analysis A non-collimated Cd radioactive source was placed in the active volume of the GPSC filled with pure xenon at 1330 mbar. Each Cd decay may involve the ejection of a conversion electron (CE) (e "62.5 kev or e "84.6 kev) accompanied by the simultaneous emission of an Ag fluorescence X-ray photon (K α "22.1 kev, K β "25.0 kev, α"3.0 kev, β"3.3 kev) or Auger electrons. In Fig. 2 are presented a set of digitised pulses. Only pulses with an area corresponding to 22.1 kev (Ag K α X-rays or background events) are shown. As seen, different pulse signatures can be distinguished. Pulses a result from events occurring in the drift region near the detector axis and leading to a single primary electron cloud. Pulses b result from events where the electron cloud due to the Xe photoelectron emission and the electron cloud due to Xe L-fluorescence secondary emission reach the scintillation region at different instants. Pulses c result from events where the primary electron cloud is strongly distorted, e.g. events occurring near detector walls. Note that the number of events a, b and c presented in Fig. 2 are not normalised to their occurring probability but the number of events a is largely dominant.

P.C.P.S. Simo es et al./nucl. Instr. and Meth. in Phys. Res. A 422 (1999) 341 346 343 Fig. 2. Digitised pulses, from a Cd X-ray source, with an area correspondent to 22.1 kev. Pulses a result from events occurring in the drift region and leading to a single scintillation light burst. Pulses b result from events where the scintillation light burst from Xe L-fluorescence secondary emission is identifiable. Pulses c result from events where the primary electron cloud is strongly distorted. Fig. 2 shows that the method in study is an effective and versatile technique to discriminate events that produce a characteristic pulse signature, e.g. events leading to double interaction points like pulses b in Fig. 2. In fact this is the working principle of a gated GPSC [10,14]. In Fig. 3, curve a, represents the pulse-height distribution obtained from the non-collimated Cd radioactive source. The real count rate is about 2 k s. The Ag K-fluorescence lines and the e and e CE peaks can be distinguished. The low-energy tail associated to each peak results from solid angle effects and primary electron loss to the detector walls, since the radioactive source is not collimated. Using pulse-duration discrimination the background under the 22.1 kev X-ray line is efficiently reduced as in previous work [7 9]: curve b of Fig. 3 is the pulse-height distribution obtained when only the pulses with 3.6 4.0 μs pulse duration time are counted. The e "62.5 kev CE emission occurs simultaneously either with a 22.1 kev fluorescence emission (83% of e events) or with Auger electrons (17% of e events). Thus, this allows an unique identification of the 62.5 kev CE, since a unique pulse signature is produced whenever the electron clouds of the 22.1 kev fluorescence emission and of the 62.5 kev CE reach the scintillation region at Fig. 3. Pulse-height distribution obtained from a non-collimated Cd radioactive source placed inside the detector active volume. Curve a shows the distribution obtained without pulse-duration discrimination. Curve b is obtained after pulseduration discrimination in the range from 3.6 to 4 μs. To the pulse-height distribution shown in curve c contribute only the pulse that appears within a 20 μs time interval after a triggerpulse with area correspondent to 22.1 kev with time-duration in the range 3.6 4 μs. Curve c is represented with a vertical scale multiplied by a factor of 100. different instants. When the electron clouds of both events reach the scintillation region simultaneously a single pulse results. In this case, the pulse cannot be distinguished from the 84.6 kev CE pulses. Curve c in Fig. 3 represents the pulse-height distribution obtained when only the first pulse following a trigger pulse, within a 20 μs time-window interval, is considered. The pulses selected for triggering are those with an area corresponding to 22.1 kev and a 3.6 to 4.0 μs time-duration. As it is shown the 62.5 kev CE is the only prominent feature on the pulse-height distribution. Its energy resolution and peak-to-background ratio improves relatively to the non-triggered pulse-height distribution from 11.6% to 9.7% and from 2 to 6, respectively, thus improving the detector sensitivity to the 62.5 kev CE. The detection efficiency is reduced by about two orders of magnitude, though, due mainly to the Ag K-fluorescence yield, to the Ag K α /K β relative emission probability, to the detector absorption efficiency of the 22.1 kev X-rays and to the probability of the electron clouds of CE and K α events lead to a single pulse. IV. INFORMATION PROCESSING

344 P.C.P.S. Simo es et al./nucl. Instr. and Meth. in Phys. Res. A 422 (1999) 341 346 2.3. Performance at high count rates Another advantage of using very short time-constants in the main amplifier is the less dependence of the pulse-height distribution on the total interaction rate, since pulse duration is shorter. Pulse-height distributions were obtained from 2 mm collimated Fe, Cd and Am X-ray sources, directly irradiating the GPSC, and for total rates from 10 to 3 10 interactions per second. In Fig. 4 curves a, b and c depict the pulse-height distributions obtained from a Cd X-ray source irradiating the detector at 0.1, 12 and 18 k total counts per second, respectively. The pulse-height distributions in Fig. 4.1 are obtained with a standard Wilkinson-type pulse-height analyser (PHA) (Nucleus-PCAII), while the distributions in Fig. 4.2 are obtained with the DPHA. The shaping time-constants used for the analog PHA acquisitions are selected to obtain the best performance for each count rate: 5, 2 and 2 μs, respectively. In Fig. 5 we present the energy resolution as a function of the real count rate for 5.9, 22.1 and 59.6 kev X-rays, as measured on the pulse-height distributions obtained with the Wilkinson-type PHA and with the DPHA. A pulse-duration discrimination was selected in the DPHA so that a counting-peak efficiency of more than 90% is maintained. As it can be seen, the energy resolution obtained using the DPHA is less dependent on the count rate than the resolution obtained using the Wilkinson-type PHA. Since at high count rates the best shaping time-constant used with the Wilkinson-type PHA decreases with decreasing X-ray energy (e.g. 0.5 μs shaping time-constants for the Fe at 30 k counts per second), the difference in the performance shown for both PHAs is smaller for low X-ray energy, as shown in Fig. 5. One of the reasons for the versatility of the DPHA technique is the ability to perform pulse analysis using always the same short shaping time-constant while for the Wilkinson-type PHA one must select the best shaping time-constants according to the total count rate and the X-ray energy. The performance obtained with the DPHA operating in the earlier conditions [7 9] presents the same dependence with the count rate as that obtained with the Wilkinson-type PHA. In Fig. 6 we depict the pulse-height distributions obtained from lead (&100 ppm) in a water sample excited with a Cd X-ray source. The total count rate is about 18 k s. An energy resolution and a peak-to-background ratio for the Pb L β line (12.6 kev) of 9.3% and 0.29 are obtained when using the Wilkinson-type PHA and shaping timeconstants of 2 μs, curve a, while the values 7.9% and 1.7 are achieved when using the DPHA, curve b. As before, a pulse-duration discrimination was selected in the DPHA so that a counting peak Fig. 4. Pulse-height distributions obtained from a Cd X-ray source irradiating the detector at 0.1, 12 and 18 k total counts per second, curves a, b and c, respectively: (4.1) using a standard Wilkinson-type PHA; (4.2) using the DPHA. The area under each curve is normalised. Fig. 5. Energy resolution as a function of the real count rate for 5.9, 22.1 and 59.6 kev X-rays, as measured for the pulse-height distributions obtained with the Wilkinson-type PHA and with the DPHA.

P.C.P.S. Simo es et al./nucl. Instr. and Meth. in Phys. Res. A 422 (1999) 341 346 345 Fig. 6. Pulse-height distributions obtained from lead (&100 ppm) in a water sample excited with a Cd source for a total count rate of about 18 k s, obtained with a standard Wilkinson-type PHA, curve a, and obtained with the DPHA, curve b. efficiency of more than 90% is maintained. For a total count rate of about 150 s, the energy resolution is 7.8% and 7.1% and the peak-to-background ratio is 0.78 and 1.7 for the Wilkinson-type PHA and the DPHA, respectively. 3. Conclusions We have demonstrated that the use of short shaping time-constants allows an effective pulsesignature analysis. Since the total pulse duration is shorter, the obtained pulse-height distribution will also be much less dependent on the total interaction rate occurring in the detector. Total interaction rates up to 12 k s can be tolerated without significant degradation of the obtained pulseheight distribution. Pulse duration and its undershoot are the limiting factors in high count-rate applications. The use of pole-zero cancellation and base-line restorer techniques reduces the effect caused by the pulse undershoot. However, this effect can be simply and likely more efficiently corrected by digital subtraction of the pulse base line. We have also found that pulse discrimination based on total pulse duration-time is as effective as the previous pulse rise-time discrimination technique [7 9,15]. The peak-to-background ratio is always better when using the DPHA than using the Wilkinsontype PHA [7 9,15], but this difference becomes smaller as the count rate increases. A slight degradation of the performance obtained using this method compared to other techniques was observed at low count rates, in cases where the noise is not efficiently eliminated by the electronics (e.g. when HV-supply ripple is significant). The maximum throughput achieved by the DPHA operating with short shaping time-constants improves from 1 to 1.7 k counts per second when compared to the throughput achieved by the DPHA operated in previous configurations. This improvement results from the fact that the number of digitised samples, per pulse, used by the processing algorithms is smaller. The maximum throughput achieved by the DPHA can be significantly increased using an upgraded digital signal processor and faster data transfer (e.g. using direct memory access). Acknowledgements This work was carried out in the Atomic and Nuclear Instrumentation Group of the Instrumentation Centre (Unit 217/94) of the Physics Department, University of Coimbra. This work was supported by CERN/P/FIS/1166/97. Travel support is acknowledged from Fundac7 a o Calouste Gulbenkian, Lisboa. The authors wish to thank T.H.V.T. Dias for manuscript reviewing. References [1] A. Peacock, R.D. Andresen, E.A. Leimann, A.E. Long, G. Manzo, B.G. Taylor, Nucl. Instr. and Meth. 169 (1980) 613. [2] A. Smith, A. Peacock, T.Z. Kowalski, IEEE Trans. Nucl. Sci. NS-34 (1987) 57. [3] D.A. Goganov, A.A. Schultz, Nucl. Instr. and Meth. 394 (1997) 151. [4] A.J.P.L. Policarpo, Space Sci. Instr. 3 (1977) 77. [5] V.P. Varvaritsa, I.V. Vikulov, V.V. Ivashov, M.A. Panov, V.I. Filatov, K.I. Shchekin, Instr. Exp. Tech. 35 (1992) 745. [6] A. Smith, M. Bavdaz, Rev. Sci. Instr. 63 (1992) 689. IV. INFORMATION PROCESSING

346 P.C.P.S. Simo es et al./nucl. Instr. and Meth. in Phys. Res. A 422 (1999) 341 346 [7] P.C.P.S. Simo es, J.F.C.A. Veloso, J.M.F. dos Santos, C.A.N. Conde, IEEE Trans. Nucl. Sci. NS-44 (1997) 521. [8] P.C.P.S. Simo es, J.M.F. dos Santos, C.A.N. Conde, X-Ray Spectrom. 26 (1997) 182. [9] P.C.P.S. Simo es, J.M.F. dos Santos, C.A.N. Conde, IEEE Trans. Nucl. Sci. NS-45 (1998), accepted for publication. [10] G. Manzo, J. Davelaar, A. Peacock, R.D. Andresen, B.G. Taylor, Nucl. Instr. and Meth. 177 (1980) 595. [11] V. Dangendorf, H. Schmidt-Bocking, Proc. SPIE-EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy V, 6 11 August 1989, San Diego, California, vol. 1159, 1989, p. 192. [12] J.M.F. dos Santos, J.F.C.A. Veloso, R.E. Morgado, C.A.N. Conde, Nucl. Instr. and Meth. 353 (1994) 195. [13] E. Cosulich, F. Gatti, Nucl. Instr. and Meth. 321 (1992) 211. [14] M.R. Sims, G. Manzo, A. Peacock, B.G. Taylor, Nucl. Instr. and Meth. 211 (1983) 499. [15] P.C.P.S. Simo es, J.C. Martins, C.M.B.A. Correia, IEEE Trans. Nucl. Sci. NS-43 (1996) 1804.