Measurement of SQUID noise levels for SuperCDMS SNOLAB detectors

Similar documents
Trigger Algorithms for the SuperCDMS Dark Matter Search

Pulse Tube Interference in Cryogenic Sensor Resonant Circuits

rf SQUID Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

11. AC-resistances of capacitor and inductors: Reactances.

Goals of the Lab: Photodetectors and Noise (Part 2) Department of Physics. Slide 1. PHYSICS6770 Laboratory 4

SIGNAL CONDITIONING FOR CRYOGENIC THERMOMETRY IN THE LHC

Superconducting quantum interference device (SQUID) and its application in science and engineering. A presentation Submitted by

Lab E2: B-field of a Solenoid. In the case that the B-field is uniform and perpendicular to the area, (1) reduces to

Evaluation of Laser Stabilization and Imaging Systems for LCLS-II

Shielding. Fig. 6.1: Using a Steel Paint Can

Detection Beyond 100µm Photon detectors no longer work ("shallow", i.e. low excitation energy, impurities only go out to equivalent of

First tests of prototype SCUBA-2 array

Thermal Johnson Noise Generated by a Resistor

APPLICATION NOTE. Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz. Abstract

Semiconductor Detector Systems

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

arxiv: v1 [physics.ins-det] 9 Apr 2016

Magnetic and Electromagnetic Microsystems. 4. Example: magnetic read/write head

Backgrounds in DMTPC. Thomas Caldwell. Massachusetts Institute of Technology DMTPC Collaboration

Simultaneous geomagnetic monitoring with multiple SQUIDs and fluxgate sensors across underground laboratories

A 200 h two-stage dc SQUID amplifier for resonant gravitational wave detectors

Super Low Noise Preamplifier

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Chemistry 985. Some constants: q e 1.602x10 19 Coul, ɛ x10 12 F/m h 6.626x10 34 J-s, c m/s, 1 atm = 760 Torr = 101,325 Pa

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

Teacher s notes Induction of a voltage in a coil: A set of simple investigations

ET1210: Module 5 Inductance and Resonance

Scintillators as an external trigger for cathode strip chambers

In the name of God, the most merciful Electromagnetic Radiation Measurement

Electric Circuit Fall 2017 Lab3 LABORATORY 3. Diode. Guide

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

Developing a Core Loss Model. Effect of Temperature on Core Loss Effect of Duty Cycle on Core Loss

Photon Count. for Brainies.

Exp 3 COLCULATE THE RESPONSE TIME FOR THE SILICON DETECTOR

Michael Hermansen. Low-Noise Piezoelectric Driver for External Cavity Diode Lasers. Physics 492R Capstone. 10 April Advisor: Dr.

PHY 123/253 Shot Noise

Figure Responsivity (A/W) Figure E E-09.

Common-Source Amplifiers

Chapter 5: Diodes. I. Theory. Chapter 5: Diodes

Physics 281 EXPERIMENT 7 I-V Curves of Non linear Device

A Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University

RC and RL Circuits Prelab

The Original SQUID. Arnold H. Silver. Josephson Symposium Applied Superconductivity Conference Portland, OR October 9, 2012

The preferred Exercise is shown in Exercises 5B or 5C.

6. Bipolar Diode. Owing to this one-direction conductance, current-voltage characteristic of p-n diode has a rectifying shape shown in Fig. 2.

8.2 Common Forms of Noise

SQUID Basics. Dietmar Drung Physikalisch-Technische Bundesanstalt (PTB) Berlin, Germany

Reference Sources. Prelab. Proakis chapter 7.4.1, equations to as attached

14.2 Photodiodes 411

ExperimentswithaunSQUIDbasedintegrated magnetometer.

Superconducting Gravity Gradiometers (SGGs)

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HSW S01

SENSOR AND MEASUREMENT EXPERIMENTS

Introduction to SQUIDs and their applications. ESAS Summer School Jari Penttilä Aivon Oy, Espoo, Finland

ECEN 4606, UNDERGRADUATE OPTICS LAB

Low Vibration, Low Thermal Fluctuation System for Pulse Tube and Gifford- McMahon Cryocoolers

Name Date: Course number: MAKE SURE TA & TI STAMPS EVERY PAGE BEFORE YOU START EXPERIMENT 10. Electronic Circuits

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

NOTICE. The above identified patent application is available for licensing. Requests for information should be addressed to:

- Datasheet - Features: Version 1.1. Cryogenic Low Pass Filter Unit Type KA-Fil 2a

Class #9: Experiment Diodes Part II: LEDs

Superconducting Transition-Edge Sensors and Superconducting Tunnel Junctions for Optical/UV Time-Energy Resolved Single-Photon Counters

United States Patent [19]

Common-source Amplifiers

Figure Figure E E-09. Dark Current (A) 1.

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Mercury Cadmium Telluride Detectors

Objective Type Questions 1. Why pure semiconductors are insulators at 0 o K? 2. What is effect of temperature on barrier voltage? 3.

Physics 1442 and 1444 Questions and problems Only

NEEP 427 PROPORTIONAL COUNTERS. Knoll, Chapters 6 & 14 Sect. I & II

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

AC Bias Characterization of Low Noise Bolometers for SAFARI Using an Open-Loop Frequency Domain SQUID-based Multiplexer Operating Between 1 and 5 MHz

Lab E5: Filters and Complex Impedance

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Photons and solid state detection

PMT Calibration in the XENON 1T Demonstrator. Abstract

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

Lecture - 19 Microwave Solid State Diode Oscillator and Amplifier

The Ins and Outs of Audio Transformers. How to Choose them and How to Use them

Final Project: FEDX X-ray Radiation Detector

d. Can you find intrinsic gain more easily by examining the equation for current? Explain.

CMOS Circuit for Low Photocurrent Measurements

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

EE320L Electronics I. Laboratory. Laboratory Exercise #6. Current-Voltage Characteristics of Electronic Devices. Angsuman Roy

ENEE 307 Laboratory#2 (n-mosfet, p-mosfet, and a single n-mosfet amplifier in the common source configuration)

PHY 122 Shot Noise. Complete Shot Noise Pre- Lab before starting this experiment

LIFETIME OF THE MUON

Gamma Spectrometer Initial Project Proposal

Glossary of VCO terms

CHAPTER 2 D-Q AXES FLUX MEASUREMENT IN SYNCHRONOUS MACHINES

Experiment 1: Instrument Familiarization (8/28/06)

(i) Determine the admittance parameters of the network of Fig 1 (f) and draw its - equivalent circuit.

Voltage Biased Superconducting Quantum Interference Device Bootstrap Circuit

EE-4022 Experiment 3 Frequency Modulation (FM)

UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE Department of Electrical and Computer Engineering

Effects of Incident Optical Power on the Effective Reverse Bias Voltage of Photodiodes This Lab Fact demonstrates how the effective reverse bias

EMC review for Belle II (Grounding & shielding plans) PXD DEPFET system

Workshop,, Nov , Hirschberg. DITANET-Workshop

AUTOMOTIVE CURRENT SENSOR DHAB S/25. Datasheet

Transcription:

Measurement of SQUID noise levels for SuperCDMS SNOLAB detectors Maxwell Lee SLAC National Accelerator Laboratory, Menlo Park, CA, 94025, MS29 SLAC-TN-15-051 Abstract SuperCDMS SNOLAB is a second generation direct dark matter search. In the SuperCDMS SNOLAB experiment, detectors are able to pick up from signals from dark matter nuclear recoil interactions which occur inside the bulk of the detectors. These interactions produce both phonon and charge signals. HEMTs read out charge signals whereas TES are used to detect phonon signals which are then read out by SQUID amplifiers. SQUID amplifiers must add negligible noise to the TES intrinsic noise which has been previously measured and is approximately 50pA/ Hz down to 100Hz for ease of signal distinguishability in dark matter nuclear interactions. The intrinsic noise level of the SQUID was tested in the SLAC 300mK fridge and determined to provide adequately low levels of noise with a floor of approximately 3pA/ Hz. Furthermore, a 10x amplifier was tested for addition of extraneous noise. This noise was investigated with and without this amplifier, and it was found that it did not add a significant amount of noise to the intrinsic SQUID noise. Introduction The Super Cryogenic Dark Matter Search (SuperCDMS) is a second generation direct dark matter detection experiment and successor to SuperCDMS Soudan. Germanium and silicon detectors are aimed at detecting dark matter interactions; this project is novel as it is aimed at detecting dark matter at masses <10GeV. The experiment is planned to be hosted at the SNOLAB underground lab in Sudbury, Canada, approximately 2km underground to shield the experiment from cosmic rays and background radiation. A crucial part of the experiment is the detectors and the readout electronics for the detectors. When a dark matter particle interacts with an atom in our detector, a nuclear recoil event takes place. Electron hole pairs are created along with prompt phonons. A voltage is then applied across the detector and the charges drift, producing subsequent Luke phonons. HEMTs are employed to read the charge signals. A TES exploits the very temperature dependent resistance of a superconducting phase transition in order to detect the phonon signal which are then read out and amplified by SQUID amplifiers. A SQUID is a superconducting loop that contains 2 Josephson junctions. These junctions have critical currents where under these currents, there is no measureable voltage or resistance. Flux is quantized in units of flux quanta in the superconducting loop which SLAC National Accelerator Laboratory, 2575 Sand Hill Road, Menlo Park, CA 94025 This material is based upon work supported by the U.S. Department of Energy, Office of Science, Office of Workforce Development for Teachers and Scientists (WDTS) under the Science Undergraduate Laboratory Internship (SULI) program, under Contract No. DE-AC02-76SF00515.

results in the current tracing a periodic function, such as the sin wave in figure 3. SQUID noise measurements are crucial for distinguishability of dark matter signals from noise in direct dark matter detection. In our experiment, a TES will detect phonon signals produced by a dark matter nuclear recoil interaction, or produced by other interactions. The TES then is read out and amplified by a SQUID magnetometer which is capable of detecting extremely subtle variations in the magnetic fields of the TES. In both the TES and the SQUID, Johnson noise resulting from thermal agitations in resistors will provide the main source of noise to the experiment. It is essential that this noise be minimized in order for the TES current noise to dominate the noise of the system and not the SQUID noise (1.3). SQUID amplifiers were manufactured and initially tested at UCDenver. They produced acceptable noise levels in the conditions at UCDenver. This experiment was the first time these SNOLAB SQUIDs (R4C5) were measured at SLAC in our 300mK fridge. A 10x amplifier box was also tested for the first time. By introducing the box, the signal along with the noise are amplified by a factor of 10. This favorably increases the signal to noise ratio. The difference between signal and noise is then more distinct and easier to analyze. Analysis of the amplified signal showed minimal extraneous noise compared to the original data. Materials and Methods Each detector has 12 channels (6 top, 6 bottom), each requiring a SQUID. A single TES sensor is coupled to each SQUID. SQUID arrays were provided by NIST based on pre-existing designs that were determined to be adequate for our uses. Preliminary tests at UCDenver have shown that they fit all project requirements, including noise levels. These SQUIDs are placed inside a copper detector tower prototype (Figure 2), which is then put in a cryogenic fridge. With the thermal load, the fridge is able to achieve temperatures of about 500mK for approximately 20 minutes. Since these SQUIDs are highly sensitive to magnetic fields, a room temperature high magnetic permeability shield is placed around the cryogenic tank. These experiments took place at SLAC National Accelerator Laboratory in the Building 33 (Light assembly building) clean room. The room is a 100k rated clean room but particle counts indicated that the room was under a 1k rating for the duration of the experiment. The experiment was placed in a Cryomech model PT405 two stage pulse tube cryocooler. Temperature sensors were disconnected for the duration of the measurements as they were determined to add excess noise. In order to measure the noise coming from an individual SQUID, a modulation curve must first be plotted to find the optimal region of operation. A signal generator and a spectrum analyzer at room temperature are hooked up to the SQUID. A triangle wave is sent through an inducting coil, and the resulting function is displayed on the spectrum analyzer (refer to figures 1 and 2). The voltage is then biased using the SQUID bias box to operate at the optimal point, just below the critical currents of the Josephson junctions. Once operating at this point, one can disconnect the signal generator and measure the signal output of the SQUID. Several bandwidths were measured between 0.25Hz-128Hz, and the results were taken. The same procedure was repeated with two different spectrum analyzers, and with and without the 10x amplifier box. 2

Figure 1 Figure 2 Figure 1 is a simple diagram of the circuit that was used for the SQUID noise measurements. The SQUID is pictured in the middle with Josephson junctions in green. A triangle wave is sent through the circuit on the right, which creates a magnetic field in the coil, inducing a voltage in the SQUID because of the change in magnetic flux. This voltage can then be measured and is the subject of this paper. Figure 2 is a picture of our detector tower prototype. The entire tower is made of copper and is a little over 2ft tall. Results Figure 3 Voltage vs. Time 0.1 0.08 Voltage (V) 0.06 0.04 0.02 0-0.02-0.04-0.06-0.08-6.00E-02-4.00E-02-2.00E-02 0.00E+00 Time (s) 3 2.00E-02 4.00E-02 6.00E-02

Figure 4 Modulation Curve 1 0.8 0.6 Volts 0.4 0.2 0-0.2-0.4-0.6-0.8-150 -100-50 0 50 100 150 Microamps This is the modulation curve that was measured from the initial run. The output from the spectrum analyzer was given in units of voltage vs. time (figure 3), but time was converted to current with a conversion rate of 100mA/s, which can be calculated from the frequency and voltage of the triangle wave sent through the initial resistor in figure 1. From figures 3 and 4, one is able to observe and calculate the following quantities. Peak to peak =, Modulation depth =, Gain =, Slope of inflection point of voltage time graph =, Turn ratio of coils =, Relation of time and current = Voltage before amplifier (1.1) Responsivity (1.2) Furthermore, we must look at the noise equation to understand the requirements for the SQUID. Noise (1.3) 4

Figure 5 Figure 6 0.000001 SQUID Noise Measurements 0.0000001 1E-08 A/ Hz 1E-09 1E-10 1E-11 1E-12 1 10 100 Hz 1000 10000 100000 Green: SR785 Red: Agilent 5

This graph is the culmination of all of the measurements with the SQUID. Bandwidths from 0.25Hz to 10KHz were used. This graph has 4 different lines. The red lines were measured with the Agilent spectrum analyzer. The green lines were measured with the SRS785 spectrum analyzer. The lighter lines represent signals that came from the 10x amplifier box and were subsequently divided by 10. These are hard to see; refer to the next graph for a better view. 1/f noise is very apparent at low frequencies from about 1-100Hz. At higher frequencies, the noise level bottoms out at ~3 pa/ Hz. Figure 7 1.00E-06 SQUID Noise Measurements 1.00E-07 1.00E-08 A/ Hz 1.00E-09 1.00E-10 1.00E-11 1.00E-12 1.00E+00 1.00E+01 1.00E+02 Hz Red: 10x 1.00E+03 1.00E+04 1.00E+05 This graph is the same as the previous figure, except the Agilent readings have been removed. The red line is the signal from the test with the amplifier box that has been divided by 10 and then overlain on the green original signal. A white noise level of ~3pA/ Hz was measured as well as a noise of ~20pA/ Hz at 20Hz for the amplified signal. The regular signal shows the same white noise level and ~19pA/ Hz of noise at 20Hz. Discussion The readouts for the SQUIDs from the spectrum analyzers were measured and the units were converted to the input current noise. The modulation curve resulted in a very nice sinusoidal curve. If there were problems with the SQUID, such as the possibility of having charge trapped, one would have seen a deviation in this curve which was not present. Furthermore, it can be seen that there seems to be some points in the curve that don t seem to be very precise. This is merely due to the maximum resolution from the oscilloscope, and can be ignored. A much sharper curve was observed on the oscilloscope (Figure 5). Looking at the calculations, one can see that the responsivity of the circuit is 550 ohms, which is as expected. The voltage of the 6

SQUID is set to approximately 0.2V, corresponding to the point of inflection on the graph and the region of highest responsivity. The noise graphs show many things. First of all, as seen in Figure 6, there is obviously more noise at lower frequencies. This can be attributed to the so called 1/f noise or pink noise as the power spectral density is inversely proportional to the frequency. This noise is expected. Secondly, the red lines on figure 2 show considerably more noise in almost all regions than the green lines. The only difference between the two measurements was that they were taken with two different spectrum analyzers, therefore it is easy to conclude that the Agilent machine produces more intrinsic noise than the SRS785. The region between 100-1000Hz shows the most noise, whereas at higher frequencies, both machines seem to catch the same amount of noise. Both machines also seem to output the same amount of noise at low frequencies, likely because of the 1/f noise. At very low frequencies there are few data points, so it is hard to make conclusions. The TES sensors have a current noise of ~50pA/ Hz at 100Hz from previous measurements. The SQUID must therefore have a noise level that is much lower in order for the TES sensors to dominate in (1.3). At 100 Hz, the SQUID produced approximately 11pA/ Hz whereas at 20Hz, it produced approximately 19 pa/ Hz. This is acceptable noise as even at low frequencies, the TES intrinsic noise will dominate and set the phonon energy resolution. At high frequencies, the SQUID produces approximately 3 pa/ Hz of white noise, which is well below the threshold of the TES and will not be a problem. This also indicates that the SQUIDs that we received are consistent with those measured at UCDenver. UCDenver SQUID measurements were done under ideal conditions in a liquid nitrogen tank with ideal shielding. UCDenver measured approximately 5pA/ Hz of noise at 200 Hz, almost mirroring the measurements taken here. In this experiment, there were many more factors including the vertical flex cable, the cryogenic fridge, the temperature sensors, and the pulse tube cooler that could have added extra noise to the measurements. One can clearly see that the noise of these SQUIDs is acceptably low and the TES term in the noise equation will dominate. Figure 7 shows the graph where only the SRS785 machine is used. One can see that the signal that has been amplified by 10x looks as if it is slightly above the original signal. At 100Hz, it produced approximately 12 pa/ Hz of noise and at 20Hz, it produced approximately 20 pa/ Hz, which is negligible compared to the TES intrinsic noise. No further noise peaks can be seen on the amplified noise measurements, suggesting that the box doesn t create sporadic noise. Since the amplified signal creates a more favorable signal to noise ratio, and therefore makes the distinction between signal and noise easier, it can be concluded that the amplifier should be used in all further experiments. Conclusion SQUID noise measurements are essential for making the distinction between noise and dark matter interaction signals in direct dark matter detection. The noise of these SQUIDs were measured for the first time in the conditions at SLAC. It produced a white noise level of roughly 3 pa/ Hz and at 20Hz, produced around 19 pa/ Hz. This is low enough to be negligible and let the TES noise dominate instead of the SQUID noise. An amplifier box was added, and was determined to not add significant sporadic noise. The amplifier box should be used in all future tests. Future work includes testing the floating SQUID amplifiers for the CDMS HV Detectors for their noise levels. 7

As this is an entirely novel project, expectations are hard to estimate, but the Acknowledgements hope is to achieve noise levels similar to that of this project. This material is based upon work supported by the U.S. Department of Energy, Office of Science, and the Office of Workforce Development for Teachers and Scientists. I would like to thank Richard Partridge, Tsuguo Aramaki, and Roger Romani who provided insight and opinions on the research, as well as teaching me the ins and outs of the project from the ground up. 8

References 1. The SuperCDMS Collaboration. SuperCDMS SNOLAB Conceptual Design Report. N.p., 27 Mar. 2015. 9