Response characteristics of a tandem ionization chamber in standard X-ray beams

Similar documents
Preliminary studies of a new monitor ionization chamber

DETERMINATION OF INHERENT AND ADDITIONAL FILTRATION IN ORDER TO ESTABLISH RADIATION QUALITIES ACCORDING TO IEC 61267

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

Cylindrical Ion Chambers Victoreen Model 550 Series

Determination of the tube voltage from clinic mammographic system using two types of detectors

Cylindrical Ion Chambers

Calibration of KAP meters

MEASURING CONDITIONS AND UNCERTAINTIES FOR THE COMPARISON AND CALIBRATION OF NATIONAL DOSIMETRIC STANDARDS AT THE BIPM *

Electronic Brachytherapy Sources. Thomas W. Rusch

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

IONIZING RADIATION 2018 / 2019 DETECTORS. Including Codes of Practice

DETECTORS. Including Codes of Practice IONIZING RADIATION

Technical Protocol. SIM Comparison of Calibration Coefficients at Radiotherapy Level for Orthovoltage X-ray Beams. Pilot laboratory: NIST

When small things matter. Small Field Dosimetry Application Guide

熊本大学学術リポジトリ. Kumamoto University Repositor

Derivation of total filtration thickness for diagnostic x-ray source assembly

RADIATION THERAPY DETECTORS. For Small Field Dosimetry

A new approach to measure dwell position inaccuracy in HDR ring applicators quantification and corrective QA

Estimation of signal transfer property for wireless digital detector in different measurement schemes

Introduction of a Single Chip TLD System for Patient Dosimetry

Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010

Effective energy measurement using radiochromic film: application of a mobile scanner

I. Introduction.

Title: A COMPARISON OF Cs-137 AND X-RAY SOURCES AS CALIBRATION REFERENCES FOR THERMOLUMINESCENT DOSIMETER CHIPS

Development of the Mini-Ionization Chamber for High-Dose Real-Time Monitoring Inside a Gamma Irradiation Facility

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

Half value layer and AEC receptor dose compliance survey in Estonia

Effect of pressure, temperature and humidity in air on photon fluence and air kerma values at low photon energies

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

IV. 4. An Optical Common-mode Rejection for Improving the Sensitivity Limit of a Radiochromic Imaging Film

Key words: fluoroscopy, dose-area-product, kerma-area-product, calibration of KAP meters, patient exposure

Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

Y11-DR Digital Radiography (DR) Image Quality

Evaluation of X-Ray Beam Quality Based on Measurements and Estimations Using SpekCalc and Ipem78 Models

Nuclear Associates

Post-irradiation colouration of Gafchromic EBT radiochromic film

Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array

ABSORBED DOSE DISTRIBUTIONS USING THE ISODENSITOMETRIC METHOD FOR EXPOSURES WITH FILTER EMPLOYED FOR MAMMOGRAPHIES

A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering

NOT FOR DISTRIBUTION JINST_083P_0914 v1

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

Characterization of an in vivo diode dosimetry system for clinical use

Update on NIST X-Ray Air-Kerma Standards and Calibrations

Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector

SUBCHAPTER 14. THERAPEUTIC INSTALLATIONS

Disclosures. Outline 7/31/2017. Current Implementation Status of IEC Standard : Exposure Index (EI) for Digital Radiography

New Detectors for X-Ray Metal Thickness Measuring

Abstract. This investigation was supported by PHS grant CA10953 awarded by the NCI, DHHS. Introduction

Progress Report from the HDR Brachytherapy Working Party. Margaret Bidmead

A Multichannel Dosimeter Based on Scintillating Fibers for Medical Applications

Received 3 December 2016; revised 17 May 2017; editorial decision 21 June 2017; accepted 4 July 2017

Monte Carlo study on a new concept of a scanning photon beam system for IMRT

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners

Test Equipment for Radiology and CT Quality Control Contents

DOSE EVALUATION IN UROGRAPHY RADIOLOGICAL PROCEDURES

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing of fusionwelded

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response

RULES OF TENNESSEE DEPARTMENT OF ENVIRONMENT AND CONSERVATION DIVISION OF RADIOLOGICAL HEALTH CHAPTER USE OF X-RAY APPARATUS

Assessment of Beam Alignment, Collimation and Half Value Layer of Some Selected X-Ray Machines in Plateau State, Nigeria

SCINTILLATING FIBER DOSIMETER ARRAY

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

Technical data CAMARGUE CS-VH50/300. VARIABLE Height Bucky Table With Ceiling Suspension

SRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

QUALITY CONTROL TESTS IN SOME DIAGNOSTIC X-RAY UNITS IN BANGLADESH

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Technical aspects on DAP calibration and CT calibration

RADIATION SAFETY DIRECTORATE

A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3

Monitoring the eye lens

Development of new dosimeter for measuring dose distribution in CT

Nuclear Associates , , CT Head and Body Dose Phantom

Development of Personal Dosimeter Using Electronic Dose Conversion Method

MASINZA, ALWYN STANSLAUS. BSc (Hons.), PGDSc

RaySafe Xi. Specifications

HIGH RESOLUTION COMPUTERIZED TOMOGRAPHY SYSTEM USING AN IMAGING PLATE

Imaging Technique Optimization of Tungsten Anode FFDM System

NEMA XR X-ray Equipment for Interventional Procedures User Quality Control Mode

X-ray tube with needle-like anode

Energy Monitoring Device for Electron Beam Facilities

ISO INTERNATIONAL STANDARD. Non-destructive testing of welds Radiographic testing Part 1: X- and gamma-ray techniques with film

3D Diode Array Commissioning: Building Confidence in 3D QA Technology

Barcelona, January 2011 Workshop on optimization of Radiation Protection of Medical Staff

ADVANCED THERAPY DOSIMETER MODEL 35040

QA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion

COMPUTED RADIOGRAPHY SIMULATION USING THE MONTE CARLO CODE MCNPX

GE AMX 4+ Portable X-Ray

Physics of small megavoltage photon beam dosimetry

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Rotational total skin electron irradiation with a linear accelerator

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

State of the Art Film Dosimetry

The Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

A Scintillating Fiber Dosimeter for Radiology and Brachytherapy. with photodiode readout

Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

Atomic and nuclear physics

Transcription:

Applied Radiation and Isotopes 58 (2003) 495 500 Response characteristics of a tandem ionization chamber in standard X-ray beams Alessandro M. Costa*, Linda V.E. Caldas Instituto de Pesquisas Energ!eticas e Nucleares, Comiss*ao Nacional de Energia Nuclear, P.O. Box 11049, S *ao Paulo SP 05422-970, Brazil Received 14 November 2002; accepted 9 December 2002 Abstract A tandem ionization chamber was developed for quality control programs of the kilovoltage X-ray equipment used in radiotherapy. This chamber was tested and calibrated in low-energy X-ray beams in accordance with international recommendations, and it showed a satisfactory level of performance, mainly with regard to its main use, namely formation of a tandem system for confirmation of half-value layers in X-radiation beams previously determined by the conventional method. In this developed system no absorbers or special setups are necessary. r 2003 Elsevier Science Ltd. All rights reserved. Keywords: Ionization chambers; X-radiation; Quality control; Radiotherapy; Tandem system 1. Introduction Quality control in radiotherapy is needed for providing accurate dose delivery. Quality control in radiotherapy comprises specific recommendations for each type of treatment unit (linear accelerators, 60 Co units, kilovoltage X-ray equipment, simulators, brachytherapy equipment, etc). The tests for kilovoltage X-ray equipment include the constancy check of the beam quality in terms of the half-value layer (HVL). The conventional method for determining the HVL by adding absorbers of known thicknesses and materials takes a long time; consequently, the beam dosimetry becomes a delayed process and is never carried out at clinics at the recommended frequency. The tandem method has been in use for several decades (Kenney and Cameron, 1963; Gorbics and Attix, 1968; Spurny et al., 1973; da Rosa and Nette, *Corresponding author. Tel.:+55-11-3816-9211; fax: +55-11-3816-9209. E-mail address: amcosta@net.ipen.br (A.M. Costa). 1988). It consists of two individual dosimeters with different energy dependences and makes it possible to determine the effective energy in unknown radiation fields. At the Calibration Laboratory of Instituto de Pesquisas Energ!eticas e Nucleares, tandem systems composed of different ionization chambers were formed (Albuquerque and Caldas, 1989; Caldas, 1991). They consist of two individual chambers with different energy dependences, which enable one to confirm the values of the HVLs or effective energies in X-radiation beams previously determined by the conventional method. Plane parallel ionization chambers are recommended for beam dosimetry in radiation therapy with low-and high-energy X-rays; and also with high-energy electrons (IAEA, 1997; Almond et al., 1999; IAEA, 2000; Ma et al., 2001; Solimanian et al., 2001). In this work, a special double-faced plane parallel ionization chamber was developed with collecting electrodes of different materials in a tandem system. It was tested and calibrated in low-energy X-ray beams in accordance with international recommendations (IAEA, 1994; IEC, 1997). 0969-8043/03/$ - see front matter r 2003 Elsevier Science Ltd. All rights reserved. doi:10.1016/s0969-8043(03)00026-5

496 A.M. Costa, L.V.E. Caldas / Applied Radiation and Isotopes 58 (2003) 495 500 2. Materials and methods The special plane parallel ionization chamber with a double face, a tandem chamber, is disc-shaped, and the only difference between the two faces is in the inner (collecting) and guard electrode materials: one has aluminum electrodes (face A), and the other has graphite electrodes (face G). The chamber body is made of polymethylmethacrylate (Lucite). The inner electrode is 5 mm thick and 20 mm in diameter. The guard rings are 10 mm thick and 3 mm wide. The outer electrodes (entrance windows) are made of an aluminized polyester film (1.7 mg cm 2 ). The distance between the inner and the outer electrodes is 2 mm. The measuring volumes are approximately 0.6 cm 3, and the chamber is unsealed. Fig. 1 shows a schematic diagram of the tandem chamber. This chamber was attached to an electrometer PTW (Physikalisch-Technische Werkst-.atten), model MULTIDOS. The irradiations were carried out using a Rigaku- Denki X-ray system, model Geigerflex, with a Philips tube, model PW 2184/00, a beryllium window of 1 mm and a tungsten target, and maximum generating potential of 60 kv. Table 1 presents the characteristics of this X-ray system. The reference standard system consists of a plane parallel ionization chamber model M23344 (0.2 cm 3 ) and an electrometer model UNIDOS 10001, both from PTW. For the repeatability test of the combined response of the tandem chamber and electrometer, a stability check device was utilized. This device consists of a radioactive source of 90 Sr (nominal activity of 33.3 MBq, 1988) from PTW, type 8921, and a special holder was developed in order to provide a reproducible geometry for the measurements. As the chambers used in this work are unsealed, the measurements were corrected to the reference conditions of temperature and pressure, i.e, 20 C and 101.3 kpa. The reported uncertainties of all measurements are expanded uncertainties based on standard uncertainties multiplied by a coverage factor 2, corresponding to a confidence level of approximately 95%. 3. Results and discussion The tandem ionization chamber was studied in relation to the characteristics of leakage current without irradiation, linearity of response, short-term stability, dependence on radiation quality, dependence on field size and chamber orientation. 3.1. Leakage current without irradiation The leakage current without irradiation was determined from charge measurements for 20 min after the Fig. 1. Schematic diagram of the tandem ionization chamber: (a) entrance windows, (b) collecting electrodes, (c) guard rings, (d) entrance window fixation ring, and (e) cables. Table 1 Characteristics of the Rigaku-Denki X-ray sxystem; inherent filtration: 1 mm Be Generating potential Tube current Added filtration Air kerma rate (kv) (ma) (mm Al) (mm Al) (mgy min 1 ) 25 30 0.44 0.25 400 30 30 0.54 0.36 421 40 30 0.68 0.53 592 45 25 0.73 0.59 562 50 25 1.02 0.89 464

A.M. Costa, L.V.E. Caldas / Applied Radiation and Isotopes 58 (2003) 495 500 497 application of the polarizing voltage of +400 V sequentially to both faces of the tandem chamber. The leakage current, expressed as percentage of the ionization current produced by the minimum air kerma rate to which the chamber was exposed, was less than 0.02% in all cases for both the faces of the tandem chamber. This chamber meets the requirement of the International Electrotechnical Commission (IEC, 1997), which prohibits values exceeding 70.5%. 3.2. Linearity of response The linear relationship between ionization current and air kerma rate was determined by irradiating sequentially both faces of the tandem chamber at 30 kv X-ray tube voltage (HVL of 0.36 mm Al) and varying the tube current. The chamber was placed in the radiation field at a distance of 50 cm from the source taking its front faces (entrance windows) as the reference points. The air kerma rates were determined by utilizing the reference standard. The data are presented in Figs. 2 and 3. The straight lines are the results of a linear fit to these data. The uncertainty obtained for the angular coefficient, i.e., the uncertainty obtained in linearity of response was 70.45% for face A and 70.37% for face G. 3.3. Short-term stability Ten successive measurements corrected to the reference conditions of temperature and pressure were taken for both faces of the tandem chamber exposed to the 90 Sr source. The experimental standard deviation was 0.06% of the mean value obtained for face A and 0.03% for face G. The tandem chamber meets the IEC (IEC, 1997) requirement, namely, that the standard deviation Fig. 2. Ionization current as a function of air kerma rate for the tandem chamber, face A, 30 kv X-ray tube voltage and HVL 0.36 mm Al. The straight line is the result of a linear fit to the data. Fig. 3. Ionization current as a function of air kerma rate for the tandem chamber, face G, 30 kv X-ray tube voltage and HVL of 0.36 mm Al. The straight line is the result of a linear fit to the data. Table 2 Calibration coefficients of the tandem chamber (mmal) Calibration coefficient (mgy nc 1 ) Face A of a single measurement with the stability check device as determined from repeated measurements shall not exceed 0.3% of the mean output value. 3.4. Dependence on radiation quality Face G 0.25 23.670.5 36.670.8 0.36 22.170.5 37.070.8 0.53 21.070.5 37.670.9 0.59 20.670.5 37.870.9 0.89 19.670.5 37.970.9 The energy variation in response of the tandem chamber over the range of low-energy X-rays was studied using the radiation qualities listed in Table 1. The irradiation conditions were as follows: focus-tochamber distance was 50 cm, and the diameter of the circular field was 6.66 cm. The response of the tandem chamber in terms of the calibration coefficients (Meghzifene and Shortt, 2002) was determined in relation to the conventional true value of air kerma rate by utilizing a reference standard. The calibration coefficients of both faces of the tandem chamber are given in Table 2, and the energy dependence of the correction factors is shown in Fig. 4. The correction factors were normalized to the HVL of 0.36 mm Al (IEC, 1997). As can be seen, the maximum variation in response of face A (20%) is greater than in response of face G (3.4%). Taking into account the uncertainties in the measurements (72.3%), one can assume that face G presents a flat response over the

498 A.M. Costa, L.V.E. Caldas / Applied Radiation and Isotopes 58 (2003) 495 500 energy range concerned. In this case, face G of the tandem chamber can be utilized for the measurement of air kerma (and air kerma rate) in low-energy X-radiation fields as recommended for dosimetry in radiation therapy (IEC, 1997; IAEA, 2000). The response variation range specified by the IEC (IEC, 1997) is72.0%; the IAEA recommendation is that a variation in response would be less than 5% over the energy range used (IAEA, 2000). The ionization chamber used for the determination of HVLs by the conventional method should be selected to have a minimum dependence on radiation quality over the range in question. Therefore, face G of the tandem chamber can also be used for attenuation measurements in low-energy X-radiation beams. The different energy responses of the two faces of the tandem chamber can be used for the constancy check of the beam quality in the quality control program by utilizing the tandem system. The tandem curve, obtained by the ratio between the responses of face G and the responses of face A as a function of the HVLs is shown in Fig. 5. It shows a very useful curvature, and the HVLs can be determined with an uncertainty of approximately 73%. 3.5. Dependence on field size According to the requirements of IEC (IEC, 1997), the limit of variation in the response due to the field size shall not exceed 72.0%. For this test, circular fields with different diameters were utilized. Both faces of the tandem chamber were irradiated at the focus to chamber distance of 50 cm in the radiation qualities given in Table 1, and varying the beam limiting collimators. The compliance with this performance requirement was checked by comparing the tandem chamber response with the response of a chamber model 23342 from PTW, which is independent of field size (IEC, 1997). Tables 3 and 4 present the relative response as a function of the field size for the tandem chamber. Table 5 shows that the relative response of the PTW 23342 chamber is independent of the field size. The Fig. 4. Correction factors of the tandem chamber as a function of the radiation quality normalized to the HVL of 0.36 mm Al. Fig. 5. Tandem curve of the tandem chamber for low-energy X-radiation beams. Table 3 Relative response as a function of the field size for the tandem chamber, Face A Field diameter (cm) 0.25 mm Al 0.36 mm Al 0.53 mm Al 0.59 mm Al 0.89 mm Al 3.04 0.982 0.976 0.969 0.966 0.960 3.52 0.991 0.986 0.981 0.978 0.973 4.01 0.991 0.987 0.984 0.982 0.979 5.35 0.994 0.993 0.992 0.991 0.990 6.66 1.000 1.000 1.000 1.000 1.000 9.95 0.997 0.996 0.997 0.997 0.998 10.0 1.000 0.999 0.999 1.000 1.000 The relative uncertainty is less than 0.7%. The values are normalized to field diameter of 6.66 cm.

A.M. Costa, L.V.E. Caldas / Applied Radiation and Isotopes 58 (2003) 495 500 499 responses were normalized to the field size normally used at the calibration procedure of such chambers, i.e., diameter of 6.66 cm. The tandem chamber has an outer diameter of 6.0 cm. As can be seen from Tables 3 and 4, the maximum variation of the response occurs at the field sizes in which the tandem chamber is not uniformly irradiated, i.e., between diameters of 3.04 and 5.35 cm. Consequently, the developed tandem chamber must be used in field sizes greater than 6.0 cm in diameter. 3.6. Chamber orientation The variation in response of the tandem chamber was studied in function of the plane of the entrance window tilted by 75 from its reference position, perpendicular to the axis of the radiation beam. The tests were performed measuring the tandem chamber response, faces A and G, at tilt angles of 1 apart using the minimum and the maximum radiation qualities given in Table 4 Relative response as a function of the field size for the tandem chamber, Face G Field diameter (cm) 0.25 mm Al 0.36 mm Al 0.53 mm Al 0.59 mm Al 0.89 mm Al 3.04 0.967 0.957 0.941 0.936 0.923 3.52 0.978 0.973 0.962 0.958 0.949 4.01 0.979 0.978 0.969 0.968 0.961 5.35 0.987 0.989 0.986 0.986 0.984 6.66 1.000 1.000 1.000 1.000 1.000 9.95 0.993 0.998 0.999 1.002 1.003 10.0 0.991 0.996 0.996 0.999 0.998 The relative uncertainty is less than 0.8%. The values are normalized to the field diameter of 6.66 cm. Table 5 Relative response as a function of the field size for the PTW 23342 chamber Field diameter (cm) 0.25 mm Al 0.36 mm Al 0.53 mm Al 0.59 mm Al 0.89 mm Al 3.04 0.996 0.995 0.994 0.986 0.983 3.52 0.999 0.999 0.997 0.995 0.993 4.01 0.996 0.995 0.996 0.990 0.988 5.35 0.998 0.997 0.998 0.988 0.986 6.66 1.000 1.000 1.000 1.000 1.000 9.95 1.002 1.002 1.003 1.007 1.009 10.0 1.002 1.002 1.003 1.002 1.003 The relative uncertainty is less than 1%. The values are normalized to the field diameter of 6.66 cm. Table 6 Relative response of the tandem chamber as a function of the angle of tilt Angle of tilt (deg.) Face A Face G 0.25 mm Al 0.89 mm Al 0.25 mm Al 0.89 mm Al 5 1.00770.011 0.99370.008 1.00470.012 1.00270.009 4 1.00270.009 0.99170.008 1.00570.010 1.00270.008 3 1.00270.008 0.99170.008 1.00470.010 1.00470.009 2 1.00170.009 0.98770.008 1.00670.010 1.00370.009 1 0.99870.008 0.98870.007 1.00570.010 1.00170.009 0 1.00070.008 1.00070.008 1.00070.012 1.00070.009 +1 0.99670.009 0.98970.007 1.00370.010 0.99970.009 +2 0.99970.008 0.99470.008 1.00570.010 1.00470.009 +3 0.99970.008 0.99370.008 1.00870.013 1.00370.009 +4 1.00070.008 0.99470.008 1.00570.010 1.00470.008 +5 1.00070.008 0.99570.007 1.00170.013 1.00270.009

500 A.M. Costa, L.V.E. Caldas / Applied Radiation and Isotopes 58 (2003) 495 500 Table 1, focus to chamber distance of 100 cm and field size diameter of 6.0 cm. The results are given in Table 6. The responses were normalized to the reference position (0 ). As can be seen, both faces of the tandem chamber meet the IEC requirement, i.e., the limits of variation of response are within 71.0% (IEC, 1997). 4. Conclusions A double-faced chamber was developed and tested in accordance to international recommendations (IEC, 1997) at the Calibration Laboratory of Instituto de Pesquisas Energ!eticas e Nucleares for use in therapeutic beams. It showed a satisfactory level of performance. The different energy response of the two faces of the tandem chamber allowed the formation of a tandem system for the constancy check of the beam quality of low-energy X-radiation fields. The main advantage of this type of chamber is that it allows simple and quick measurements, which enables quality control tests to be performed at a much higher frequency than usually in practice. No absorbers or any special setups are necessary for dosimetry procedures with the tandem chamber. Acknowledgements The authors would like to thank the partial financial support from Funda@*ao de Amparo "a Pesquisa do Estado de S*ao Paulo (FAPESP) and Conselho Nacional de Desenvolvimento Cient!ıfico e Tecnol!ogico, Brazil. References Albuquerque, M.P.P., Caldas, L.V.E., 1989. New ionization chambers for beta and X-radiation. Nucl. Instrum. Methods A 280, 310 313. Almond, P.R., Biggs, P.J., Coursey, B.M., Hanson, W.F., Saiful Huq, M., Nath, R., Rogers, D.W.O., 1999. AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med. Phys. 26 (9), 1847 1869. Caldas, L.V.E., 1991. A sequential tandem system of ionisation chambers for effective energy determination of radiation fields. Radiat. Prot. Dosimetry 36, 47 50. da Rosa, L.A.R., Nette, H.P., 1988. Thermoluminescent dosimeters for exposure assessment in gamma or X radiation fields with unknown spectral distribution. Appl. Radiat. Isot. 39 (3), 191 197. Gorbics, S.G., Attix, F.H., 1968. LiF and CaF 2 :Mn thermoluminescent dosimeters in tandem. Int. J. Appl. Radiat. Isot. 19 (1), 81 89. IAEA (International Atomic Energy Agency), 1994. Calibration of dosimeters used in radiotherapy. Technical Report Series no. 374, IAEA, Vienna. IAEA (International Atomic Energy Agency), 1997. The use of plane parallel ionization chambers in high energy electron and photon beams: an international code of practice for dosimetry. Technical Report Series no. 381, IAEA, Vienna. IAEA (International Atomic Energy Agency), 2000. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absobed dose to water. Technical Report Series no. 398, IAEA, Vienna. IEC (International Electrotechnical Commission), 1997. Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy, Standard IEC 60731, IEC, Geneva. Kenney, G.N., Cameron, J.R., 1963. X-ray beam quality measurements utilizing TL dosemeters. Progress Report, AEC Contract AT 11-1-1105, TID-19112. Ma, C.-M., Coffey, C.W., DeWerd, L.A., Liu, C., Nath, R., Seltzer, S.M., Seuntjens, J.P., 2001. AAPM protocol for 40 300 kv X-ray beam dosimetry in radiotherapy and radiobiology. Med. Phys. 28 (6), 868 892. Meghzifene, A., Shortt, K.R., 2002. Calibration factor or calibration coefficient? SSDL Newsletter No. 46, p. 33. Solimanian, A., Ensaf, M.R., Ghafoori, M., 2001. Design, construction and calibration of plane-parallel ionization chambers at the SSDL of Iran. SSDL Newsletter No. 45, pp. 15 24. Spurny, Z., Milu, C., Racoveanu, N., 1973. Comparison of X-ray beams using thermoluminescent dosemeters. Phys. Med. Biol. 18 (2), 276 278.