Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

Similar documents
CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

PoS(PhotoDet 2012)022

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

Silicon Photomultiplier

High granularity scintillating fiber trackers based on Silicon Photomultiplier

K 223 Angular Correlation

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

Electronic Instrumentation for Radiation Detection Systems

Solid-State Photomultiplier in CMOS Technology for Gamma-Ray Detection and Imaging Applications

ORTEC Experiment 3. Gamma-Ray Spectroscopy Using NaI(Tl) Equipment Required. Purpose. Gamma Emission

HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

Arrays of digital Silicon Photomultipliers Intrinsic performance and Application to Scintillator Readout

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

Testing the Electronics for the MicroBooNE Light Collection System

PET Performance Evaluation of MADPET4: A Small Animal PET Insert for a 7-T MRI Scanner

PoS(PhotoDet 2012)058

Study of Silicon Photomultipliers for Positron Emission Tomography (PET) Application

SPM Series Quick Start Experiment Guide Rev.1.0, May 2011

Traditional analog QDC chain and Digital Pulse Processing [1]

Characterisation of SiPM Index :

B-Series Fast, Blue-Sensitive Silicon Photomultiplier Sensors

J-Series High PDE and Timing Resolution, TSV Package

COMPTON SCATTERING. Purpose. Introduction. Fundamentals of Experiment

SensL B-Series Silicon Photomultipliers for TOF- PET. NDIP2014 Kevin O Neill 4 th July, 2014

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

Tutors Dominik Dannheim, Thibault Frisson (CERN, Geneva, Switzerland)

Characterization of SiPMs for Large Scale Applications

High collection efficiency MCPs for photon counting detectors

A Novel Design of a High-Resolution Hodoscope for the Hall D Tagger Based on Scintillating Fibers

Application Notes: Discrete Amplification Photon Detector 5x5 Array Including Pre- Amplifiers Board

PMT tests at UMD. Vlasios Vasileiou Version st May 2006

PML Channel Detector Head for Time-Correlated Single Photon Counting

Instruction manual and data sheet ipca h

Measure the roll-off frequency of an acousto-optic modulator

nanomca datasheet I. FEATURES

Thomas Frach, Member, IEEE, Walter Ruetten, Member, IEEE, Klaus Fiedler, Gunnar Maehlum, Member, IEEE, Torsten Solf, and Andreas Thon

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

SiPM Module PRELIMINARY

arxiv: v2 [physics.ins-det] 10 Jan 2014

A user-friendly fully digital TDPAC-spectrometer

APPLICATION NOTE. Steffen Pahlke KETEK VITUS Silicon Drift Detector. KETEK VITUS Silicon Drift Detector (SDD) with new ultra low capacitance FET

A test of silicon photomultipliers as readout for PET

A combined NIR filter and trigger sensor for use with a supercontinuum laser

Silicon Carbide Solid-State Photomultiplier for UV Light Detection

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

Non-amplified High Speed Photodetectors

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

PCS-150 / PCI-200 High Speed Boxcar Modules

Amplified High Speed Photodetectors

Highlights of Poster Session I: SiPMs

arxiv:hep-ex/ v1 19 Apr 2002

Time-of-flight PET with SiPM sensors on monolithic scintillation crystals Vinke, Ruud

nanomca-sp datasheet I. FEATURES

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Gain and Breakdown Voltage Measurements

The digital Silicon Photomultiplier A novel Sensor for the Detection of Scintillation Light

Dual 500ns ADC User Manual

Direct Measurement of Optical Cross-talk in Silicon Photomultipliers Using Light Emission Microscopy

Characterizing a single photon detector

mesytec GmbH & Co. KG Wernher-von-Braun-Str Putzbrunn Germany Tel.: Fax: REPRESENTED BY:

Review of Solidstate Photomultiplier. Developments by CPTA & Photonique SA

Non-amplified Photodetectors

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

PH2510 Nuclear Physics Laboratory Use of Scintillation Counters (NP5)

Scintillation Counters

Positron Emission Tomography

An electrical photon source

RECENTLY, the Silicon Photomultiplier (SiPM) gained

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Timing Resolution Performance Comparison for Fast and Standard Outputs of SensL SiPM

A Measurement of the Photon Detection Efficiency of Silicon Photomultipliers

nanodpp datasheet I. FEATURES

IR Antibunching Measurements with id201 InGaAs Gated SPAD Detectors

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

On the initiation of lightning in thunderclouds (Instrumentation, Supplementary information)

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

Results on the LED Pulser System for the Hall A DVCS Experiment

Balanced Photoreceivers Models 1607-AC & 1617-AC

MSCF-16-LN (Data sheet V5.0_01)

Software Module MDPP-16-QDC V0003

CHAPTER 8 PHOTOMULTIPLIER TUBE MODULES

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

PERFORMANCE OF PHOTODIGM S DBR SEMICONDUCTOR LASERS FOR PICOSECOND AND NANOSECOND PULSING APPLICATIONS

The Influence of Crystal Configuration and PMT on PET Time-of-Flight Resolution

Performance of 8-stage Multianode Photomultipliers

Variable-Gain High Speed Current Amplifier

InGaAs SPAD freerunning

Activities in Electronics Lab Associates are: Mrs. Arti Gupta, K.S.Golda, S.Muralithar & Dr.R.K.Bhowmik

Contens: 1. Important Notes 1.1 Technical Recommendations 1.2 Mechanical Recommendations 2. Operating the CPM 2.1 Selecting Operating Mode 2.2 Calcula

INGAAS FAST PIN (RF) AMPLIFIED PHOTODETECTORS

e t Development of Low Cost γ - Ray Energy Spectrometer

Variable-Gain High Speed Current Amplifier

Gamma Spectrometer Initial Project Proposal

Investigation of Solid-State Photomultipliers for Positron Emission Tomography Scanners

CAEN Tools for Discovery

Nuclear Instruments and Methods in Physics Research A

Transcription:

KETEK GmbH Hofer Str. 3 81737 Munich Germany www.ketek.net info@ketek.net phone +49 89 673 467 70 fax +49 89 673 467 77 Silicon Photomultiplier Evaluation Kit Quick Start Guide Eval Kit

Table of Contents 1 Introduction...3 2 Setup Example...3 3 Single Photon Measurements...4 3.1 Dark Count Spectrum... 4 3.2 Single Photon Spectrum... 4 3.3 Using the Full Dynamic Range... 4 4 Measurements with Scintillator...5 4.1 Energy Spectrum... 5 4.2 Coincidence Time Resolution... 6 5 Appendix...7 5.1 Variant for Optical Bench Mount... 7 5.2 Bias Source... 7 This quick start guide covers the KETEK Evaluation Kit only. For other products please refer to www.ketek.net. Revision history: Rev. 1.0, August 2016 - Initial release

1 Introduction The KETEK Evaluation Kit allows an easy operation and evaluation of any KETEK. It can be used for a wide range of applications which require e.g. single photon counting or measurements with scintillators. Two of the Evaluation Kits can be mounted face to face for coincidence measurements. The Evaluation Kit is equipped with a preamplifier and set up modularly so that the PCBs with the presoldered s can be swapped to a different model. For the operation of the Evaluation Kit only a +12 V DC power supply, a bias source and an oscilloscope are required. For full technical details of the s and the Evaluation Kit, please refer to the KETEK Datasheet and the KETEK Evaluation Kit Datasheet. Additionally, KETEK provides a Bias Source which can be directly connected to the Evaluation Kit. Please refer to the KETEK Bias Source Datasheet for further information. 2 Setup Example The signal and bias connectors are SMA type, the preamplifier is equipped with solder pins. An overview is shown in figure 1. Signal lines need to be be terminated with 50 Ω. Preamplifier power: +12V DC, 150 ma Bias: positive with max. + 40 V, typical current limit 2 ma Signal: connected to the preamplifier Note: Signal corresponds to 90% of the signal Monitor: connected to an oscilloscope (50 Ω DC) Note: Monitor corresponds to 10% of the signal Amplified Signal: connected to an oscilloscope (50 Ω DC) Since the is a highly sensitive photodetector, it must be operated under dark conditions. After biasing the with e.g. 4 V above the breakdown voltage, dark counts should be visible at the Amplified Signal path, using e.g. a timebase of 100 ns/div and a vertical resolution of 20 mv/div. The Monitor is used for higher photon fluxes for which the preamplifier saturates (e.g. when coupling a bright scintillator). Note: The procedure may vary depending on the used readout electronics. For the examples shown here, a digital oscilloscope is used. Fig. 1 Evaluation Kit Connection Scheme +12 V DC GND Monitor Amplified Signal Signal Mounting Holes for Coincidence Setup Bias Preamplifier PCB with 08/2016 Rev. 1.0 Page 3

3 Single Photon Measurements 3.1 Dark Count Spectrum At the oscilloscope, set the trigger to the amplified signal at 0.5 pe amplitude (1 pe is the smallest occurring pulse height, corresponding to a photoelectron pe). Note: This measurement has to be done under dark conditions without illuminating the. Measure the area respectively the charge of the signal at the position of the trigger point in e.g. a 25 ns wide gate. Best results are obtained by integrating the whole positive pulse area. Histogramming of the integrated charge is the resulting dark count spectrum. Note: E.g. crosstalk probability can be extracted from the dark count spectrum. 3.2 Single Photon Spectrum For this measurement, a pulsed light source is needed. This can be e.g. a pulsed LED or a pulsed laser. Connect the electrical trigger output of the pulser to the oscilloscope and set the trigger to it. Note: In case a synchronized electrical trigger output is not provided by the pulser, a second Evaluation Kit can be used to generate the trigger signal. This second should be fully illuminated by the same optical pulse e.g. with an optical beam splitter. The other path for the single photon measurement can be attenuate e.g. with neutral density filters. Measure the area respectively the charge of the signal in e.g. a 25 ns wide gate. The histogram of the measured charge is the single photon spectrum (cf. fig. 2). Note: E.g. the relative gain, breakdown voltage and photo detection efficiency can be extracted. 3.3 Using the Full Dynamic Range The amplified signal output and the Monitor output can be connected and used simultaneously. Typically for low light levels down to single photons, the measurement is done with the amplified output. For higher light levels, the preamplifier will saturate at 0.5 V signal amplitude. After this point, the Monitor output is measured. This allows to exploit the the full dynamic range of the. Fig. 2 Example of a Single Photon Spectrum Measured with PM3350 at 4 V overvoltage Signal 20 mv/div Laser Trigger Signal 5 ns/div Histogram Counts [a.u.] pedestal 1 pe 2 pe 8 pe Start Integration Gate End Integration Gate 9 pe 10 pe 11 pe 3 pe 4 pe 5 pe 6 pe 7 pe 12 pe 13 pe 14 pe 15 pe 16 pe 17 pe Integrated Charge [nvs] Evaluation Kit Quick Start Guide

4 Measurements with Scintillator With bright scintillators typically the charge released by the exceeds the dynamic range of the preamplifier. In this case, the Monitor output is used. The charge released by the is a measure for the detected number of scintillating photons and thus a measure for the deposited energy in the scintillator. Note: Alternatively the preamplifier can be disconnected to connect the Signal output directly to the oscilloscope (50 Ω DC). This is an option in case timing is not measured. For simultaneous measurements of timing and energy, the Amplified Signal is used for timing and the Monitor output for the energy measurement. Note: Even though the preamplifier may be in saturation, timing can still be measured at the steep rising edge of the Amplified Signal close to the baseline. Usually the monitor output is used to measure the charge released by the corresponding to the deposited energy in the scintillator. Evaluation Kits can be ordered either with unmounted or premounted LYSO crystals on the s: For PM11 (1.2 x 1.2 mm² active area) LYSO with 1.2 x 1.2 x 4.0 mm³ For PM33 (3.0 x 3.0 mm² active area) LYSO with 3.0 x 3.0 x 5.0 mm³ For PM66 (6.0 x 6.0 mm² active area) LYSO with 5.0 x 5.0 x 5.0 mm³ 4.1 Energy Spectrum In analogy to section 3.1, at the oscilloscope the trigger has to be set to the signal above the baseline and the pulse area as a measure for the energy is histogrammed. Figure 3 shows an example energy spectrum of a ²²Na source measured with the Monitor output. The used scintillator is LYSO with 3 x 3 x 5 mm³. Fig. 3 Example of an Energy Spectrum Measured with PM3350 at 4 V overvoltage, ²²Na source and LYSO 3.0 x 3.0 x 5.0 mm³ 350 300 Counts [a.u.] 250 200 150 Energy Resolution 11.9% FWHM at 511 kev (corrected for saturation) 100 50 0 0 250 500 750 1000 1250 Energy [kev] 08/2016 Rev. 1.0 Page 5

4.2 Coincidence Time Resolution Two Evaluation Kits are mounted facing each other Note: Two M6 mounting bolts are supplied for alignment when ordering two Evaluation Kits, they are screwed to one of the Evaluation Kits, the second Evaluation Kit is slid on the mounting bolts This example uses two PM3350, each equipped with 3.0 x 3.0 x 5.0 mm³ LYSO with a ²²Na source placed exactly in between the s Note: ²²Na has two decay branches. It emits two 511 kev annihilation γ rays in opposing directions, originating from a β+ decay. The second decay emits a single 1.27 MeV γ ray. Of interest here are the two 511 kev γ rays for the coincidence measurement. Monitor outputs are connected to a coincidence logic, e.g. realized by the oscilloscope Lower threshold of the coincidence logic defines the lower energy cut-off. The thresholds for both Monitor outputs are set to the minimum between Compton edge and 511 kev photopeak to filter for the energy respectively coincident events. The Coincidence Timing can then be measured by the time difference of both Amplified Signals. The output of the coincidence logic triggers the measurement of the timestamps of both Amplified Signals. Note: Best timinig is achieved with a timing measurement close to the baseline. Typically with LYSO the optimum is around 2-3 pe. The time stamps of both Amplified Signals can be measured e.g. with a leading edge discriminator, a constant fraction discriminator or using an oscilloscope. The time difference between both time stamps is then histogrammed as shown in figure 4. The Coincidence Time Resolution is the FWHM of the Gaussian distribution. Fig. 4 Example of a CTR Histogram Measured with two PM3350 at 4 V overvoltage, ²²Na source and LYSO 3.0 x 3.0 x 5.0 mm³ 2500 Counts [a.u.] 2000 1500 1000 CTR 180 ps FWHM 500 0-1 -0.5 0 0.5 1 Time Difference [ns] Evaluation Kit Quick Start Guide

5 Appendix 5.1 Variant for Optical Bench Mount The Evaluation Kit as also available in a different form factor for optical bench mount including the preamplifier (cf. fig. 5). The PCB with the is premounted to a Thorlabs CP02/M Frame (SM1-Threaded 30 mm Cage Plate, 0.35 Thick, 2 Retaining Rings, M4 Tap). The is centered on the optical axis. The cables are MCX to SMA and are supplied together with the Evaluation Kit. 5.2 Bias Source As an addon to both Evaluation Kits, a Bias Source is available (cf. fig. 5). It offers an adjustable bias voltage from 20 to 40 V with either positive or negative polarity. Current limit can be set to either 2 or 20 ma. For full technical specifications please refer to the KETEK Bias Source Datasheet. Fig. 5 Evaluation Kit for Optical Bench Mount with Bias Source Bias Source premounted in optomechanic frame Preamplifier 08/2016 Rev. 1.0 Page 7