Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Similar documents
The impact of increasing SID on patient dose in pa abdomen imaging

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Improvement of CT image quality with iterative reconstruction idose4

Lumbar disc height assessment: a comparative study between EOS and digital radiography

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Correction of the local intensity nonuniformity artifact in high field MRI

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation.

Aim. Images for this section: Page 2 of 13

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

Development of new dosimeter for measuring dose distribution in CT

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization

Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing

Suppression of metal artifacts using image-based monoenergetic DECT imaging

Practical guidelines for color calibration and quality assurance of medical displays

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PATIENT EFFECTIVE DOSES IN DIAGNOSTIC RADIOLOGY, NA

The importance of radiation quality for optimisation in radiology

Dose Reduction and Image Preservation After the Introduction of a 0.1 mm Cu Filter into the LODOX Statscan unit above 110 kvp

The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

Exposure in Dental Radiology: A Comparison Between Intra-oral, Panoramic and Tomographic Examinations

Ask EuroSafe Imaging Tips & Tricks. Paediatric Imaging Working Group. Dose Management in Digital Radiography

Teaching Digital Radiography and Fluoroscopic Radiation Protection

Iranian Journal of Medical Physics

Exposure Indices and Target Values in Radiography: What Are They and How Can You Use Them?

Artefacts found in computed radiography

While digital techniques have the potential to reduce patient doses, they also have the potential to significantly increase them.

Predicted image quality of a CMOS APS X-ray detector across a range of mammographic beam qualities

biij Optimisation in general radiography CJ Martin, PhD, FIPEM, FioP Biomedical Imaging and Intervention Journal REVIEW PAPER

R/F. Comparison of Long View Radiography Systems. 1. Introduction. 2. Methods of Long View Radiography

Comparison of new digital mammography systems: physical characterisation and image quality evaluation

2 COMPENSATING FILTERS PETER J. BARGER

The effect of compensating filter on image quality in lateral projection of thoraco lumbar radiography

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners

Voice recognition versus transcriptionist

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

2217 US Highway 70 East Garner, NC Main: Fax:

Beam-Restricting Devices

Appropriate Inspection Distance of Digital X-Ray Imaging Equipment for Diagnosis

Practical Medical Physics Session: TG-151 Dose Monitoring. August 5, 2013 Katie Hulme, M.S.

Half value layer and AEC receptor dose compliance survey in Estonia

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Effects of Plastic Wedges on Whole Foot Radiograph in Anteroposterior and Oblique Positions

Reflex TM Surgical Technique. Anterior Cervical Plate

A comparative study of several digital flat panel X-ray units: patients doses and image quality in chest radiography

ddr Compact Series Setting a new benchmark in digital radiography.

Comparison of computed radiography and filmõscreen combination using a contrast-detail phantom

I. Introduction.

Imaging Technique Optimization of Tungsten Anode FFDM System

Effect of patient support pads on image quality and dose in fluoroscopy a)

Evaluation of a quality control phantom for digital chest radiography

COMPUTED RADIOGRAPHY CHAPTER 4 EFFECTIVE USE OF CR

Clinical Experience Using the Open Bore Multislice CT System Supria (16 slice CT) MEDIX VOL. 61 P.8 P.11

Preliminary studies of a new monitor ionization chamber

The Evaluation of Collimator Alignment of Diagnostic X-ray Tube Using Computed Radiography System

Digital radiography. bucky table and wall stand as a dual detector or wireless system. Amadeo DR Systems

Objective Evaluation of Radiographic Contrast- Enhancement Masks

X-RAYS - NO UNAUTHORISED ENTRY

Estimation of signal transfer property for wireless digital detector in different measurement schemes

X o- System. Chiro- Systems. Chiropractic X-Ray Systems and Digital Solutions. Digital Radiography Computerized Radiography Chiropractic Systems

Image quality evaluation of turbo-spin echo diffusion weighted image (TSE-DWI) : A phantom study

Introduction of Computed Radiography in Two Mammography Services: Image Quality and Dose Analysis

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Report on the study and optimisation of chest and lumbar spine X-ray imaging

Diagnostic Imaging Equipment & Consumables. Patient Positioning

Reducing Radiation Exposure from Survey CT Scans

Authors: Cabral, Ricardo 1 ; Carvoeiras, Pedro 2 ; Fatana, João 2, ; Alves, Rita 1. 1 Centro Hospitalar Lisboa Norte - Hospital de Santa Maria; 2

Optima ZS Spinal Fixation System

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Open. the Digitized world. Fuji Computed Radiography

efficiency ddrcompact Series

Investigation of the line-pair pattern method for evaluating mammographic focal spot performance

Calibration of KAP meters

European Journal of Radiology

Luminos RF Classic. Where value meets performance.

Improved Imaging Plate Dosimetry for X-rays in Interventional Radiology

Clinical Experiences with a Patient Skin Dose Monitoring and Tracking Program

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

SPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by

X-ray Tube and Generator Basic principles and construction

QC in Diagnostic Radiology. Main steps for a QC survey in Diagnostic Radiology

Introduction. Sam R. Kottamasu Lawrence R. Kuhns

Derivation of total filtration thickness for diagnostic x-ray source assembly

Ring-Down versus Comet Tail: Two artifacts uncovered

Optimization of Digital Mammography Resolution Using Magnification Technique in Computed Radiography 1

Transcription:

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis Poster No.: C-0585 Congress: ECR 2015 Type: Scientific Exhibit Authors: K. Minehiro, K. Ichikawa, S. demura, K. Noto, T. Takata, T. Gabata; Kanazawa/JP Keywords: DOI: Radiation safety, Digital radiography, Radioprotection / Radiation dose, Bones, Congenital 10.1594/ecr2015/C-0585 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 12

Aims and objectives Adolescent idiopathic scoliosis (AIS) is present in 2%-4% of children aged between 10 and 16 years(fig. 1 on page 2). It is defined as a lateral (LAT) curvature of the spine greater than 10, accompanied by vertebral rotation. Although the LAT curvature is the main component, it can also be associated with rotation of the spine and curvatures in different planes (1). AIS is usually first identified by a family member, during school screening or by a paediatric or family physician. Because AIS is usually painless, a fullness or prominence of the back is particularly noted while bending forward. However, as the vertebrae (spinal bones) rotate, the rib cage is affected, which in turn can compromise the heart and lungs (i.e. shortness of breath). For evaluation of spinal curvature progression, AIS patients have to undergo radiographic full-spine examination at least once every few months. However, a previous report has suggested that exposure to X-rays during childhood and adolescence may increase the risk of breast cancer (2). Therefore, the dose of the full-spine examination should be reduced to prevent from the risk increase. Though some K-edge type filters, which offer selective spectra, have been proposed to reduce the dose of general x-ray imaging, it is noteworthy that the conventional filters with aluminium (Al) and copper (Cu) reduced the dose as efficiently as did the K-edge type filters with negligible additional tube loading (3). The purpose of the present study was to investigate the effectiveness of a dose reduction method using Cu-filters in full-spine examination. Images for this section: Page 2 of 12

Fig. 1: Patients with adolescent idiopathic scoliosis (AIS) Surgery is usually indicated for 50 or 40 curves in skeletally immature patients, curves that progress despite bracing, and cause 'unacceptable' (to the patient) deformity. Page 3 of 12

Methods and materials Dose measurement The entrance surface dose (ESD) was measured using a dosimeter (Model 9015, Radcal Co., USA) with an ion chamber (10 5-6, Radcal Co.).Fig. 2 on page 5 shows the measurement geometry we used. The exposure field size was 10 cm 10 cm. We measured ESDs for non-filter and 0.1-mm, 0.2-mm and 0.3-mm Cu filters (0.1-mm Cu, 0.2-mm Cu, and 0.3-mm Cu), using average exposure conditions of anterior-posterior (AP) (81 kv#20 mas) and lateral (LAT) (90 kv#50 mas) images for AIS patients. The formula for calculating the ESD is shown in Fig. 3 on page 5. Phantom image evaluation Image acquisition The full-spine A-P and LAT view images of an anthropomorphic body phantom, Rando Phantom (The Phantom Laboratory, Salem, NY) were obtained by using a computed radiography system (FCR XL-2; Fuji medical systems, Tokyo, Japan) including a imaging plate (Fuji IP Cassette LC 14B2X100002000078; Fuji medical systems, Tokyo, Japan), for non-filter, 0.1-mm Cu, 0.2-mm Cu, and 0.3-mm Cu. The exposure conditions of A-P and LAT images were 81kV 20 mas and 90 kv 90 mas, respectively. The mas value for LAT image was increased from the standard value of 50 mas, because the lateral width of Rando phantom was larger than that of the standard AIS patient. Other parameters are shown in Fig. 4 on page 6 Subjective study Three orthopaedists independently assessed the image qualities of the full-spine A-P and LAT images. Five locations (cervical spine, C4 and C7; thoracic spine, T8 and T12; lumber spine and L3) for A-P view and six locations (C4, C7, T8, T12, L3 and pelvic bone) for LAT view were graded using a three-point scale: 3 = superior, 2 = suboptimal, 1 = unacceptable ( Fig. 5 on page 6 ). This scale was classified on the basis of the visually observed edge of the bone. Thus 'superior', 'suboptimal' and 'unacceptable' meant 'completely observable', 'partially observable' and 'unobservable', respectively. For each location, the score of each Cu filter was compared with that of non-filter. In addition, the average scores of all locations were compared between no filter, 0.1-mm Cu, 0.2-mm Cu, and 0.3-mm Cu images. Page 4 of 12

Images for this section: Fig. 2: Exposure doses were measured using an ion chamber. The chamber was placed at 60 cm above the floor to avoid influence from backscatter fractions. All measurements were made at the centre of the X-ray beam with a fixed field size 10 cm 10 cm. Page 5 of 12

Fig. 3: We measured the entrance surface doses (ESD) for no filter, 0.1-, 0.2- and 0.3- mm Cu filters using average exposure conditions of anterior-posterior (AP) and lateral (LAT) views of AIS patients. Fig. 4: Materials and radiographs settings Page 6 of 12

Fig. 5: Full-spine images of RANDO taken using no filter: Cu-filters were subjectively assessed by three orthopedists using a 3-point rating scale: superior, suboptimal and unacceptable. Page 7 of 12

Results ESD ESD of A-P views for non-filter was 0.31mGy, whereas ESDs for 0.1-mm Cu, 0.2-mm Cu and 0.3-mm Cu were 0.16, 0.11 and 0.08 mgy, respectively. ESDs of LAT views for non-filter, 0.1-mm Cu, 0.2-mm Cu and 0.3-mm Cu were 1.67, 0.94, 0.65 and 0.49 mgy, respectively ( Fig. 6 on page 8 ). The dose was remarkably reduced with increase of the thickness of Cu-filter. Subjective study For the A-P view, there were small differences in the subjective score between non filter 0.1-mm Cu, 0.2-mm Cu and 0.3-mm Cu. The score for 0.3-mm Cu tended to be lower in the location of the low thoracic and lumbar spine (Fig. 7 on page 9, Fig. 8 on page 9 ). The LAT view image quality ( Fig. 9 on page 10, Fig. 10 on page 10 ) showed small differences between non filter, 0.1-mm Cu, 0.2-mm Cu, and the 0.3-mm Cu image presented significantly degraded image qualities as shown in Fig. 10 on page 10. The average score of the 0.3-mm Cu-filter image, particularly of the LAT view, was lower than those of the others( Fig. 11 on page 11 ). The scores of non-filter, 0.1- mm Cu, and 0.2-mm Cu were not statistically significant. As a result, it was indicated that the 0.2-mm Cu-filter could be used with approximately 60% dose reduction, maintaining the image quality of the full spine image for AIS patients. Images for this section: Page 8 of 12

Fig. 6: Dose reduction using Cu-filters for full-spine radiography. ESDs associated with no filter, 0.1-, 0.2- and 0.3-mm Cu filters for AP and LAT views were 0.31, 0.16, 0.11 and 0.08 mgy; 1.67, 0.94, 0.65 and 0.49 mgy, respectively. Fig. 7: A-P images and evaluated scores for non-filter and 0.1-mm Cu. Page 9 of 12

Fig. 8: A-P images and evaluated scores for 0.2- and 0.3-mm Cu. Fig. 9: LAT images and evaluated scores for non-filter and 0.1-mm Cu. Page 10 of 12

Fig. 10: LAT images and evaluated scores for 0.2- and 0.3-mm Cu. Fig. 11: Average subjective image quality ratings of orthopedists. Small differences shown in the scores between no filter, 0.1-mm and 0.2-mm Cu-filters (AP: 2.80, 2.80 and 2.73; LAT: 2.78, 2.67 and 2.72, respectively). The 0.3-mm Cu-filter presented significantly degraded image quality (AP: 2.53 and LAT: 2.28). The 0.2-mm Cu-filter could yield satisfactory images with an approximately 60% dose reduction. Page 11 of 12

Conclusion The 0.1-mm, 0.2-mm, and 0.3-mm Cu filters offered approximately 50%, 60%, 75% dose reductions, respectively, as compared with non-filter. The visually evaluated image quality scores of 0.1-mm and 0.2-mm Cu filters was comparable with non-filter. Our findings indicated that the 0.2-mm Cu filter would be useful for dose reduction in the full spine x-ray examination for AIS. Personal information References 1. Weinstein SL, Dolan LA, Wright JG, Dobbs MB. Effects of bracing in adolescents with idiopathic scoliosis. N Engl J Med. 2013;369(16)1512-21 2. D M Morin, J E Lonstein, M Stovall, D G Hacker, N Luckyanov, C E Land. Breast cancer mortality after diagnostic radiography: findings from the US Scoliosis Cohort Study. Spine 2000;25:2052-63 3. Koedooder K, Venema HW. Filter materials for dose reduction in screen-film radiography. Phys Med Biol. 1986;31(6):585-600 4. Faulkner K, Jones AP, Walker A (1995) Safety in diagnostic radiology. National radiological protection board report No. 72, The institute of physical sciences in medicine, York: UK. Page 12 of 12