000W Single Output with PFC Function HRPG-000 series Bauart gepruft Sicherheit egelma ge od o s be wac g www. tuv.com ID 000000000 File Name:HRPG-000-SPEC 08-05-04
000W Single Output with PFC Function HRPG-000 series SPECIFICATION MODEL HRPG-000- HRPG-000-5 HRPG-000-4 HRPG-000-48 DC VOLTAGE RATED CURRENT CURRENT RANGE RATED POWER V 80A 0 ~ 80A 60W (max. 000W for 3 sec.) 5V 64A 0 ~ 64A 60W (max. 000W for 3 sec.) 4V 4A 0 ~ 4A 008W 48V A 0 ~ A 008W RIPPLE & NOISE (max.) Note. 50mVp-p 50mVp-p 00mVp-p 50mVp-p OUTPUT VOLTAGE ADJ. RANGE VOLTAGE TOLERANCE Note.3 LINE REGULATION LOAD REGULATION ~ 4V ±.0% ±.0% 4 ~ 7V ±.5% ±.5% ~ 8V ±.0% 46 ~ 56V ±.0% SETUP, RISE TIME 000ms, 50ms/30VAC 000ms, 50ms/5VAC at full load HOLD UP TIME (Typ.) 6ms/30VAC 6ms/5VAC at full load VOLTAGE RANGE Note.4 0 ~ 64VAC(300VAC for 5 sec.) 7 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz INPUT POWER FACTOR (Typ.) EFFICIENCY (Typ.) PF>0.5/30VAC.5% PF>0./5VAC at full load % 3% 4% AC CURRENT (Typ.) 8.5A/5VAC 5A/30VAC INRUSH CURRENT (Typ.) LEAKAGE CURRENT 5A/5VAC <.ma / 40VAC 40A/30VAC OVERLOAD 05 ~ 35% rated output power Protection type : Constant current limiting, recovers automatically after fault condition is removed PROTECTION 4.5 ~ 6.5V 8. ~ 0.6V ~ 33V 58 ~ 65V OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover OVER TEMPERATURE Shut down o/p voltage, recovers automatically after temperature goes down CURRENT SHARING Up to 4000W or (3+) units. Please refer to the Function Manual. REMOTE ON-OFF CONTROL Power ON : short; Power OFF : open. Please refer to the Function Manual. FUNCTION REMOTE SENSE Compensate voltage drop on the load wiring up to 0.5V. Please refer to the Function Manual. DC-OK SIGNAL The TTL signal out, PSU turn on = 3.3 ~ 5.6V ; PSU turn off = 0 ~ V. Please refer to the Function Manual. 5V STANDBY 5VSB : 5V@0.3A ; tolerance± 5%, ripple : 50mVp-p(max.) FAN CONTROL Fan on/off by NTC(RT50) or 30% load min. WORKING TEMP. -40 ~ +70 (Refer to "Derating Curve") WORKING HUMIDITY 0 ~ 0% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 0 ~ 5% RH non-condensing TEMP. COEFFICIENT ± 0.03%/ (0 ~ 50 ) VIBRATION 0 ~ 500Hz, 5G 0min./cycle, 60min. each along X, Y, Z axes SAFETY & EMC (Note 7) OTHERS NOTE SAFETY STANDARDS WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC EMISSION EMC IMMUNITY MTBF DIMENSION UL6368-, TUV EN6368-, EAC TP TC 004 approved I/P-O/P:3KVAC I/P-FG:KVAC O/P-FG:0.5KVAC I/P-O/P, I/P-FG, O/P-FG:00M Ohms / 500VDC / 5 / 70% RH Parameter Standard Conducted Radiated Harmonic Current Voltage Flicker EN5504, EN6000-6- Parameter ESD Radiated EFT / Burst Surge Conducted Magnetic Field Voltage Dips and Interruptions EN5503 (CISPR3) / EN550 (CISPR) Class B EN5503 (CISPR3) / EN550 (CISPR) Class A EN6000-3- Class A EN6000-3-3 ----- Standard EN6000-4- EN6000-4-3 EN6000-4-4 EN6000-4-5 EN6000-4-6 EN6000-4-8 EN6000-4- 86.6K hrs min. Telcordia SR-33 (Bellcore) ; 05.8K hrs min. MIL-HDBK-7F (5 ) 8*05*63.5mm (L*W*H) Test Level / Note Test Level / Note Level 3, 8KV air ; Level, 4KV contact Level 3 Level 3 Level 4, KV/Line-Line 4KV/Line-Earth Level 3 Level 4 >5% dip 0.5 periods, 30% dip 5 periods, >5% interruptions 50 periods PACKING.53Kg;8pcs/3.3Kg/.34CUFT File Name:HRPG-000-SPEC 08-05-04
000W Single Output with PFC Function HRPG-000 series Block Diagram PFC : 65KHz PWM : 0KHz I/P FG EMI FILTER ACTIVE INRUSH CURRENT LIMITING RECTIFIERS & PFC O.T.P. PFC CONTROL AUX POWER POWER SWITCHING O.L.P. PWM CONTROL RECTIFIERS & FILTER RECTIFIERS & FILTER DETECTION CIRCUIT O.T.P. O.V.P. REMOTE CONTROL CURRENT SHARING AUX AUXG +S +V -V -S DC OK CS RC+ RC- Derating Curve Output Derating VS Input Voltage 00 80 4V,48V 00 0 80 LOAD (%) 60 40 0 V,5V LOAD(%) 70 60 50 40-40 0 0 0 30 40 50 60 70 (HORIZONTAL) AMBIENT TEMPERATURE ( ) 0 INPUT VOLTAGE (V) 60Hz 00 64 File Name:HRPG-000-SPEC 08-05-04
000W Single Output with PFC Function HRPG-000 series Function Description of Pin No. Function Description AUXG Auxiliary voltage output ground. AUX Auxiliary voltage output, 4.75~5.5V, referenced to pin (AUXG). The maximum load current is 0.3A. This output is not controlled by the "remote ON/OFF control". 3 4 RC+ RC- Turns the output on and off by electrical or dry contact between pin 4 (RC-), Short: Power ON, Open: Power OFF. Remote control ground. 5 CS Current sharing signal. When units are connected in parallel, the CS pins of the units should be connected to allow current balance between units. 6,8 GND This pin connects to the negative terminal(-v). Return for DC-OK signal output. 7 DC-OK DC-OK signal is a TTL level signal, referenced to pin8(dc-ok GND). High when PSU turns on. +S Positive sensing. The +S signal should be connected to the positive terminal of the load. The +S and -S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. 0 -S Negative sensing. The -S signal should be connected to the negative terminal of the load. The -S and +S leads should be twisted in pair to minimize noise pick-up effect. The maximum line drop compensation is 0.5V. Function Manual.Remote Sense The remote sensing compensates voltage drop on the load wiring up to 0.5V. 0 +V LOAD -V Sense lines should be twisted in pairs Fig..DC-OK Signal DC-OK signal is a TTL level signal. High when PSU turns on. Between DC-OK(pin7) and GND(pin6,8) 3.3 ~ 5.6V 0 ~ V Output Status ON OFF 0 Fig. File Name:HRPG-000-SPEC 08-05-04
000W Single Output with PFC Function HRPG-000 series 3.Remote ON-OFF Control The PSU can be turned ON/OFF by using the "Remote Control" function. Between RC+(pin3) and RC-(pin4) Output Status SW ON (Short) SW OFF (Open) ON OFF SW 0 Fig 3. 4.Current Sharing HRPG-000 has the built-in active current sharing function and can be connected in parallel, up to 4 units, to provide higher output power as exhibited below : The power supplies should be paralleled using short and large diameter wiring and then connected to the load. Difference of output voltages among parallel units should be less than 0.V. The total output current must not exceed the value determined by the following equation: Maximum output current at parallel operation=(rated current per unit) (Number of unit) 0. When the total output current is less than 5% of the total rated current, or say (5% of Rated current per unit) (Number of unit) the current shared among units may not be fully balanced. PSU PSU PSU3 PSU4 +V +V +V +V -V -V -V -V PSU PSU PSU3 PSU4 AUXG RC+ CS DC-OK +S AUXG RC+ CS DC-OK +S AUXG RC+ CS DC-OK +S AUXG RC+ CS DC-OK +S AUX RC- GND GND -S 0 AUX RC- GND GND -S 0 AUX RC- GND GND -S 0 AUX RC- GND GND -S 0 -V +V LOAD Fig 4. File Name:HRPG-000-SPEC 08-05-04
000W Single Output with PFC Function HRPG-000 series Mechanical Specification Case No. 77 Unit:mm 5.4 3.8 Air flow direction 8. 0 I/P 3 63.5 0.75 4-M4 L=6 3 4 5 6 O/P. 0 05 8 5.4 3.8 LED (Vo ADJ.)SVR max. 8-M4 L=4 (Both Sides) 8max. 38.3 63.5.5 AC Input Terminal Pin No. Assignment Pin No. Assignment DC Output Terminal Pin No. Assignment Pin No. Assignment AC/L ~3 +V AC/N 4~6 -V 3 FG Connector Pin No. Assignment() : HRS DF-0DP-DS or equivalent Pin No. Assignment Pin No. Assignment Mating Housing Terminal AUXG 6,8 GND AUX 7 DC-OK HRS DF-0DS HRS DF-**SC 3 RC+ +S or equivalent or equivalent 4 RC- 0 -S 5 CS Installation Manual Please refer to : http://www.meanwell.com/manual.html File Name:HRPG-000-SPEC 08-05-04