NATIONAL UNIVERSITY of SINGAPORE

Similar documents
TEST EQUIPMENT PLUS. Signal Hound USB-SA44B / USB-TG44A. Application Note 1: The Smith Chart. Rev. 0

Network Analysis Basics

FieldFox Handheld Education Series Part 3: Calibration Techniques for Precise Field Measurements

TTR500 Series Vector Network Analyzer Printable Help

Agilent AN Applying Error Correction to Network Analyzer Measurements

Platform Migration 8510 to PNA. Graham Payne Application Engineer Agilent Technologies

AV3672 Series Vector Network Analyzer

On Wafer Load Pull and Noise Measurements using Computer Controlled Microwave Tuners

Waveguide Calibration with Copper Mountain Technologies VNA

Circuit Characterization with the Agilent 8714 VNA

MWA REVB LNA Measurements

PLANAR 814/1. Vector Network Analyzer

Compact Series: S5065 & S5085 Vector Network Analyzers KEY FEATURES

ECE 4265/6265 Laboratory Project 7 Network Analyzer Calibration

ME1000 RF Circuit Design. Lab 4. Filter Characterization using Vector Network Analyzer (VNA)

Coaxial TRL Calibration Kits for Network Analyzers up to 40 GHz

Project Description and Guidelines

RF Characterization Report

Vector Network Analyzers (VERY) Basics. Tom Powers USPAS SRF Testing Course 19 Jan. 2014

Challenges and Solutions for Removing Fixture Effects in Multi-port Measurements

Calibration and De-Embedding Techniques in the Frequency Domain

RF and Microwave Test and Design Roadshow 5 Locations across Australia and New Zealand

application In-Fixture Measurements Using Vector Network Analyzers Network Analysis Solutions Application Note

New Ultra-Fast Noise Parameter System... Opening A New Realm of Possibilities in Noise Characterization

Vector Network Analyzer

Keysight Technologies In-Fixture Measurements Using Vector Network Analyzers. Application Note

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

Impedance 50 (75 connectors via adapters)

Wafer-Level Calibration & Verification up to 750 GHz. Choon Beng Sia, Ph.D. Mobile:

Measurements with Scattering Parameter By Joseph L. Cahak Copyright 2013 Sunshine Design Engineering Services

EE290C - Spring 2004 Advanced Topics in Circuit Design

IVCAD VNA Base Load Pull with Active/Hybrid Tuning. Getting Started v3.5

PART III LABORATORY MANUAL. Electromagnetic Waves and Transmission Lines By Dr. Jayanti Venkataraman

Vector Network Analyzer Application note

There is a twenty db improvement in the reflection measurements when the port match errors are removed.

PXIe Contents CALIBRATION PROCEDURE

Agilent PNA Microwave Network Analyzers

Overcoming Mixer Measurement Challenges

CHAPTER - 3 PIN DIODE RF ATTENUATORS

E/O and O/E Measurements with the 37300C Series VNA

(a) The insertion loss is the average value of the transmission coefficient, S12 (db), in the passband (Figure 1 Label A)

Network Analysis Seminar. Cables measurement

Focus Microwaves Inc. 277 Lakeshore Road Pointe-Claire Quebec, H9S-4L2, Canada Tel Fax Application Note 26

FABRICATING AND USING A PCB-BASED TRL PATTERN WITH A CMT VNA

Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Configuration of PNA-X, NVNA and X parameters

Designing a 2 GHz to 10 GHz Vector Reflectometer. Jonathan Klein, University of Alaska, Fairbanks

Application Note: Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements. Product Note A

On-Wafer Noise Parameter Measurements using Cold-Noise Source and Automatic Receiver Calibration

Preamplifier Options for Reducing Cable-Braid Loop Error

PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE

MEASUREMENT OF LARGE SIGNAL DEVICE INPUT IMPEDANCE DURING LOAD PULL

Vector Network Analyzers T - Series

Agilent Accurate Measurement of Packaged RF Devices. White Paper

PLANAR S5048 and TR5048

Keysight Technologies In-ixture Microstrip Device Measurements Using TRL * Calibration. Application Note

Technologies Vector Reflectometers

Keysight MOI for USB Type-C Connectors & Cable Assemblies Compliance Tests (Type-C to Legacy Cable Assemblies)

Custom Interconnects Fuzz Button with Hardhat Test Socket/Interposer 1.00 mm pitch

Keysight Technologies PNA Microwave Network Analyzers

Keysight MOI for USB Type-C Connectors & Cable Assemblies Compliance Tests (Type-C to Legacy Cable Assemblies)

Conversion Gain Measurements on Mixers with Different Input and Output Impedances

Radio ETI031 Laboratory Experiments 2: VECTOR NETWORK ANALYSER, ANTENNA and RECEIVER MEASUREMENTS

Managing Complex Impedance, Isolation & Calibration for KGD RF Test Abstract

The 2-Port Shunt-Through Measurement and the Inherent Ground Loop

Department of Electrical and Computer Engineering ECE332. Lab 3: High Frequency Measurements

A Comparison of Harmonic Tuning Methods for Load Pull Systems

Experiment 12 - Measuring X-Parameters Using Nonlinear Vector Netowrk Analyzer

Keysight Technologies Ampliier Linear and Gain Compression Measurements with the PNA Microwave Network Analyzers. Application Note

Product Overview: Main Features: AV3672A/B/C -S Vector Network Analyzer. (10MHz~13.5 GHz/26.5 GHz/43.5 GHz)

Demo / Application Guide for DSA815(-TG) / DSA1000 Series

PLANAR R54. Vector Reflectometer KEY FEATURES

EXPERIMENT EM3 INTRODUCTION TO THE NETWORK ANALYZER

Introduction to VNA Basics PRIMER

Vector network analysis Calibration and advanced measurements

Migrating 4195A to E5061B LF-RF Network Analyzer. April 2010 Agilent Technologies

SMT Hybrid Couplers, RF Parameters and Applications

CALIBRATION TYPES & CONSIDERATIONS

MACRO FILE AND DESIGN WINDOW COMPRESSION LOAD PULL MEASUREMENTS

Hot S 22 and Hot K-factor Measurements

Optoelectronic Components Testing with a VNA(Vector Network Analyzer) VNA Roadshow Budapest 17/05/2016

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

Agilent MOI for MIPI D-PHY Conformance Tests Revision 1.00 Dec-1, 2011

VVM measurement with E5061B for replacing 8508A vector voltmeter. May 2013 Agilent Technologies

Signal Integrity Testing with a Vector Network Analyzer. Neil Jarvis Applications Engineer

Description RF Explorer RFEAH-25 1 is a 25mm diameter, high performance near field H-Loop antenna.

Compact Series: S5048 & TR5048 Vector Network Analyzers KEY FEATURES

Reflectometer Series:

UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN )

CHAPTER 4. Practical Design

Cable and Antenna Analyzer

Measuring Photonic, Optoelectronic and Electro optic S parameters using an advanced photonic module

Description RF Explorer RFEAH-25 1 is a 25mm diameter, high performance near field H-Loop antenna.

Network Analyzers. R3765G/3767G Series. R3765G/67G Series Network Analyzers. 300 khz to 3.8 GHz/300 khz to 8 GHz

External Source Control

CAL U100B CAL U100B CDN M016 CAL U100B CDN M016 CAL U100B. Used as M2 CDN. Used as M3 CDN

S-Parameter Measurements with the Bode 100

LXI -Certified 2.4mm & 1.85mm Automated Tuners

High Performance, Broadband Network Analysis Solutions

Transcription:

NATIONAL UNIVERSITY of SINGAPORE Faculty of Engineering Electrical & Computer Engineering Department EE3104 Introduction to RF and Microwave Systems & Circuits Experiment 1 Familiarization on VNA Calibration & S-parameter Measurement 1. OBJECT The object of this experiment is to introduce the students to familiarize complex S- parameter measurement using the vector network analyzer (VNA). 2. INTRODUCTION 2.1 Vector Network Analyzer At microwave frequencies, it is difficult to perform direct measurement of voltage and current at the ports of a network. Furthermore, open and short circuit terminations required for Z and Y parameter measurement may cause active microwave devices to oscillate. On the other hand, measurement of waves at microwave frequencies is easily accomplished with the aid of directional couplers and a vector analyzer. Therefore in practice, S-parameters are measured and then converted to either Z or Y parameters as required. Because S-parameter measurements are performed when the device under test (DUT) ports are terminated with a resistive load (typically 50 ohms), active DUTs are usually prevented from oscillating. Figure 1 shows the schematic diagram of a VNA used to measure the complex S- parameters of a two-port network. The interconnections between each component of this schematic are transmission lines. It can be seen from Figure 1 that both Ports have directional couplers, and the RF stimulus is continuously switched from port 1 to port 2, usually with a solid state PIN switch. The frequency converter usually has one channel for each raw parameter. The all four parameters can be measured early simultaneously, and the calibration applied to the display data.

2.1 VNA SOLT calibration method Figure 1: Simple block diagram of a VNA A VNA measures the S-parameters of a device-under-test (DUT) over a wide frequency range. In order to make accurate measurements it is necessary to be fully familiar with calibration techniques. Calibration refers to the procedure whereby the unknown reflections and loss from cables, connectors, couplers, bias tees, etc., are determined via measurements on a series of known calibration standard. There are many methods these days for calibrating a VNA. The simplest to understand is the method using a short circuit, open circuit, 50Ω load and a through line. These must be precision calibrated standards with known characteristics. In the ideal case: SHORT S11=-1 OPEN S11=+1 LOAD S11=0 THRU S11=0 S21=1 angle 0 By measuring these known standards, the errors in the measurements can be extracted. Once the errors have been calculated (as vectors) they can be eliminated from measurements of DUT. This calibration method is referred to as the SOLT method. Here use one-port measurement to simply explain this SOLT method. In order to analyze how error correction is achieved, the sources of error must be identified and represented as a mathematical model. The sources of error for one-port measurement are shown below in figure 3 and they are:

(1) LEAKAGE in the non-ideal directional bridge/coupler. (2) REFLECTIONS in the test components before the DUT Cables, adaptors etc. (3) NON-50Ω impedance of the system itself. Source/Load MATCH (4) Losses in the whole setup Transmission or tracking error Figure 3: Errors in the one-port measurement system All these various unwanted signal components are lumped together into what is known as an error model. The one-port error model is shown as Figure 4; each signal is a branch in a signal flow chart. The 3 error terms are: E DF is the DIRECTIVITY term 1 E RF is the REFLECTION TRACKING term S 11 meas E DF E SF S 22 S 11 DUT E SF is the SOURCE MATCH term Figure 4: One-port error model flow chart Referring to the one-port error model, the three error vectors can be found by measuring the three calibration standards [S11=+1 (open), S11=-1 (short) and S11=0 (load), and solving simultaneous equations as follows: E RF S 11meas S E 11DUT RF = EDF + 1 E SF S 11DUT (1) For the LOAD measurement: S11meas For the OPEN measurement: S11 meas = E (2) DF ERF = EDF + (3) 1 E SF For the SHORT measurement: S11 meas ERF = EDF (4) 1 + E SF

Once the values (magnitude and phase) of the error terms are known at each frequency, the computer within a modern VNA can correct them out and display the true value of S 11DUT. Figure 2 is an example of S21 (0.5 db per division) and S11 (10 db per division) for a thru line, BEFORE/AFTER calibration (the reference line is 0 db). Figure 2: S 21 and S 11 for a thru line, BEFORE/AFTER calibration (the reference line is 0 db). 3. EQUIPMENTS HP 8753D Network Analyzer (30 KHz ~ 6 GHz) SOLT calibration Kit Device under test: band-pass filter, 10dB attenuator Cables and adaptors 4. EXPERIMENT 4.1 General Preparation and Information The instrumentation will be configured for you. Please note that the VNA, cables and adaptors are extremely expensive. Further, the connectors are easily damaged by carelessness and negligence. The onus is on you to handle with care. Ask if in doubt. In the instructions that follow for the VNA HP 8753D: BOLD ITALIC ARIAL font means a front panel button; whereas ITALIC ARIAL means a screen menu selection.

Switch on VNA HP 8753D, press Start, set the start frequency as 1 GHz, and then press Stop, set the start frequency as 6 GHz. 4.2 SOLT Calibration (1) Press CAL, choose CALIBRATION MENU. There are several options in the screen as S11 1-PORT, S22 1-PORT, FULL 2-PORT, TRL 2-PORT etc. Here we choose FULL 2-PORT, and then there are REFLECTION, TRANSMISSION and ISOLATION in the side of the VNA screen. (2). Press REFLECTION first, you will see 6 options in the screen. They are: FORWORD: OPEN, SHORT, LOAD; REVERSE: OPEN, SHORT, LOAD. a. Connect calibration kit OPEN to port 1 using top wrench, and then click the FORWORD: OPEN, wait a while until the label underlined; b. Replace the calibration kit OPEN from port 1 with calibration kit SHORT, and then click the FORWORD: SHORT, wait a while until the label underlined; c. Replace the calibration kit SHORT from port 1 with calibration kit LOAD, and then click the FORWORD: LOAD, wait a while until the label underlined; d. Connect calibration kit OPEN to port 2 using top wrench, and then click the REVERSE: OPEN, wait a while until the label underlined; e. Replace the calibration kit OPEN from port 2 with calibration kit SHORT, and then click the REVERSE: SHORT, wait a while until the label underlined; f. Replace the calibration kit SHORT from port 2 with calibration kit LOAD, and then click the REVERSE: LOAD, wait a while until the label underlined; g. Press STANDARDS DONE. I, and wait for a while until the VNA internal computing process done. (3) Press TRANSMISSON button beside the screen, and then connect the 2-port calibration kit THRU to port 1 and port 2. You can press DO BOTH FWD + REV, and the VNA will run the transmission calibration automatically. (4) Remove the design kit from port 1 and port 2, and then press ISOLATION button beside the screen. a. Press OMIT ISOLATION, wait a while until the label underlined; b. Press ISOLATION DONE. (5) Press DONE 2-PORT CAL to finish the whole full 2-port calibration procedure.

(6) Press SAVE/RECALL to save your cal data. Save it to a blank record and rename it as EE3104 so you can recall it easily in the future. 4.3 Device Measurement a) Band-pass Filter Connect the 2-port band-pass filter to port 1&2 of VNA, press MEAS in the VNA control panel, and then choose S11, S22, S21, S22 beside the VNA screen to observe its S-parameter performance. b) 3-dB Attenuator Connect the 2-port 3-dB attenuator to port 1&2 of VNA, press MEAS in the VNA control panel, and then choose S11, S22, S21, S22 beside the VNA screen to observe its S-parameter performance. (Note: You can Press FORMAT in the VNA control panel, and then choose LOG/MAG, PHASE, SMITH CHART, POLAR beside the VNA screen to decide the display format of your observation.) 5. QUESTIONS 5.1 Observe/sketch all the S-parameter for THRU connection after SOLT calibration. 5.2 Observe/sketch all the S-parameter of Band-pass Filter. (Freq range: 1~6GHz) 5.3 Observe/sketch all the S-parameter of the 10-dB attenuator. (Freq range: 1~6GHz) 5.4 During the calibration, what are the values of S12 and S21 for the OPEN, SHORT and LOAD circuit calibration measurements and S11 and S22 for the THRU calibration measurements? 6. REFERRENCES [1] D. M. Pozar, Microwave Engineering, 2 nd Edition, John Wiley, 1998, Chapt. 4, Section 4.3. [2] G.Gonzalez, Microwave Transistor Amplifiers Analysis and Design, 2 nd Edition, Prentice-Hall, 1997, Chapt. 1, Section 1.9.