Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions.

Size: px
Start display at page:

Download "Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions."

Transcription

1 Lesson 12: Doppler Principles This lesson contains 50 slides plus 26 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 59 through 80 in the textbook:

2 DOPPLER DOPPLER PRINCIPLES

3 DOPPLER TRANSDUCER SAME FREQUENCY During Doppler operation, the reflected sound has the same frequency as the transmitted sound if the blood is stationary.

4 DOPPLER TRANSDUCER LOWER FREQUENCY During Doppler operation, the reflected sound has a lower frequency if the blood is moving away from the sound source.

5 DOPPLER TRANSDUCER HIGHER FREQUENCY During Doppler operation, the reflected sound has a higher frequency if the blood is moving toward the sound source.

6 DOPPLER SHIFT f

7 DOPPLER SHIFT FORMULA

8 BLOOD VELOCITY FORMULA

9 DOPPLER RELATIONSHIPS TRANSMITTED FREQUENCY f o DOPPLER ANGLE q BLOOD VELOCITY V DOPPLER SHIFT f Increase Increase Decrease Decrease Increase Decrease Decrease Increase Increase Increase Decrease Decrease

10 DOPPLER CALCULATIONS Doppler shifts ( f) for V = 100 cm per second c = 1550 meters per second q cos q f o = 2.5 MHz f o = 5.0 MHz Hz 6452 Hz Hz 5588 Hz Hz 4562 Hz Hz 3226 Hz Hz 1670 Hz Hz 0 Hz

11 DOPPLER SHIFT EXAMPLE TRANSMITTED FREQUENCY RECEIVED FREQUENCY DOPPLER SHIFT f DIRECTION OF FLOW TO SOUND 5 MHz MHz 5 khz Away 5 MHz MHz 5 khz Toward

12 SPECTRAL DOPPLER FFT DISPLAY: POSITIVE DOPPLER SHIFT WITHOUT SPECTRAL BROADENING FREQUENCY SCALES FFT DISPLAY: POSITIVE DOPPLER SHIFT WITH SPECTRAL BROADENING

13 CW SPECTRAL DOPPLER NON-IMAGING CW B-MODE IMAGING AND CW

14 CW SPECTRAL DOPPLER NON-IMAGING CARDIAC CW DOPPLER TRANSDUCER

15 CW SPECTRAL DOPPLER CW DOPPLER DETECTING SHIFTS FROM MORE THAN ONE VESSEL

16 CW SPECTRAL DOPPLER B-MODE AND CW DOPPLER WITH SPECTRUM ANALYZER DISPLAY

17 PW SPECTRAL DOPPLER PW TRANSMITTING PW RECEIVING

18 PW SPECTRAL DOPPLER NON-IMAGING PW B-MODE IMAGING AND PW

19 PW SPECTRAL DOPPLER

20 PW SPECTRAL DOPPLER NORMAL SAMPLE VOLUME WITH NO SPECTRAL BROADENING LARGER THAN NORMAL SAMPLE VOLUME WITH SPECTRAL BROADENING

21 PW - ALIASING WRAP-AROUND INDICATES ALIASING

22 PW - ALIASING Example: The PRF is 4 khz, the Nyquist Limit is 2 khz. The Doppler shift exceeds the Nyquist Limit. POSITIVE DOPPLER SHIFT WITH THE NYQUIST LIMIT (N.L) EXCEEDED AS A RESULT OF HIGH VELOCITY FLOW

23 PW - ALIASING Example: The PRF is 8 khz, the Nyquist Limit is 4 khz. The Doppler shift does not exceed the Nyquist Limit. ALIASING: NYQUIST LIMIT EXCEEDED SAME DOPPLER SHIFT BUT ALIASING ELIMINATED SCALE ADJUSTED TO INCREASE THE PULSE DOPPLER PRF AND RAISE THE NYQUIST LIMIT

24 PW - ALIASING Example: The PRF is 8 khz, the Nyquist Limit is 4 khz. The Doppler shift does not exceed the Nyquist Limit. ALIASING: NYQUIST LIMIT EXCEEDED SAME DOPPLER SHIFT BUT ALIASING ELIMINATED BASE LINE LOWERED TO INCREASE THE PULSE DOPPLER PRF AND RAISE THE NYQUIST LIMIT

25 PW - ALIASING Example: The PRF is still 4 khz, the Nyquist Limit is still 2 khz. The cursor was adjusted to increase the Doppler angle, which reduced the Doppler shift to a value below the Nyquist Limit. ALIASING: NYQUIST LIMIT EXCEEDED ALIASING ELIMINATED DOPPLER ANGLE INCREASED SCALE, BASELINE, AND PRF UNCHANGED

26 PW - ALIASING Example: The PRF is still 4 khz, the Nyquist Limit is still 2 khz. The transducer frequency was changed from 10 MHz to 5 MHz, which reduced the Doppler shift to a value below the Nyquist Limit. ALIASING: NYQUIST LIMIT EXCEEDED ALIASING ELIMINATED TRANSDUCER FREQUENCY DECREASED SCALE, BASELINE, AND PRF UNCHANGED

27 PW - ALIASING Example: The PRF is 5 khz, the Nyquist Limit is 2.5 khz. The Doppler shift exceeds the Nyquist Limit. NO WRAP-AROUND BUT ALIASING STILL PRESENT THE BASE LINE WAS MOST LIKELY LOWERED TO ELIMINATE WRAP- AROUND BUT THE DOPPLER SHIFT STILL EXCEEDS THE NYQUIST LIMIT

28 NYQUIST LIMIT THE NYQUIST LIMIT IS EQUAL TO THE DOPPLER PRF DIVIDED BY 2

29 NYQUIST LIMIT NYQUIST LIMITS (PRF 2) PRF N.L. 1 khz (1000 Hz) 500 Hz 2 khz (2000 Hz) 1000 Hz 3 khz (3000 Hz) 1500 Hz 4 khz (4000 Hz) 2000 Hz 5 khz (5000 Hz) 2500 Hz 6 khz (6000 Hz) 3000 Hz

30 CW NO ALIASING CW - NO ALIASING:

31 ANGLE-CORRECT PW - VELOCITY CORRECT: PROPER ANGLE-CORRECT VELOCITY SCALES PW - VELOCITY INCORRECT: IMPROPER ANGLE-CORRECT

32 WALL FILTERS WALL FILTER SETTING NORMAL WALL FILTER SETTING TOO HIGH

33 DOPPLER GAIN DOPPLER GAIN SETTING NORMAL DOPPLER GAIN SETTING TOO HIGH

34 TRANSCRANIAL DOPPLER NON-IMAGING TCD STUDY WITH A SINGLE-ELEMENT PW TRANSDUCER

35 TRANSCRANIAL DOPPLER PW TRANSDUCERS USED FOR TRANSCRANIAL DOPPLER

36 ANALOG DOPPLER Non-spectral

37 ANALOG DOPPLER

38 PULSATILITY INDEX (PI) PI = (peak systolic velocity end diastolic velocity) mean velocity

39 RESISTIVITY INDEX (RI) RI = (peak systolic velocity end diastolic velocity) peak systolic velocity

40 COLOR-FLOW IMAGING

41 COLOR-FLOW IMAGING STANDARD COLOR MAP IN THIS SECTOR SCAN, SAMPLES ARE OBTAINED FROM MANY DIFFERENT ANGLES-TO-FLOW. THE RED HUES REPRESENT POSITIVE DOPPLER SHIFTS. THE BLUE HUES REPRESENT NEGATIVE DOPPLER SHIFTS. THE DIRECTION OF FLOW IS FROM THE LEFT (L) OF THE IMAGE TO THE RIGHT (R) OF THE IMAGE. THE BLACK AREA IN THE CENTER OF THE VESSEL IS THE RESULT OF A 90º DOPPLER ANGLE.

42 COLOR-FLOW IMAGING ENHANCED COLOR MAP IN THIS LINEAR SCAN, ALL SAMPLES ARE OBTAINED FROM THE SAME ANGLE-TO-FLOW. WHEN THE BOX-CURSOR FORMS A PARALLELOGRAM WITH THE TOP SKEWED TO THE RIGHT WITH THE BOTTOM SKEWED TO THE LEFT: THE BLUE HUES (UPPER COLOR SCALE) REPRESENT POSITIVE DOPPLER SHIFTS. THE REDDISH HUES (LOWER COLOR SCALE) REPRESENT NEGATIVE DOPPLER SHIFTS. BASED ON THE ENHANCED MAP USED FOR THIS COLOR BAR SCALE, THE YELLOWISH HUE (IN THE CENTER OF THE VESSEL) REPRESENTS INCREASED VELOCITY. THE REGION OF TURBULENCE REPRESENTS REVERSE FLOW.

43 COLOR-FLOW IMAGING STANDARD COLOR MAP IN THIS LINEAR SCAN, ALL SAMPLES ARE OBTAINED FROM THE SAME ANGLE-TO-FLOW. THE BLACK AREA IN THE CENTER OF THE VESSEL IS THE RESULT OF A 90º DOPPLER ANGLE.

44 COLOR VARIANCE MAP COLOR-FLOW IMAGING

45 COLOR-FLOW IMAGING CAROTID

46 COLOR-FLOW IMAGING STANDARD COLOR MAP COLOR DOPPLER SAMPLE ANGLES 2-D, SPECTRAL DOPPLER, & COLOR-FLOW DOPPLER ECHOCARDIOGRAPHIC STUDY (Triplex)

47 DOPPLER TISSUE IMAGING PARASTERNAL LONG AXIS VIEW, 2-D ECHOCARDIOGRAPHY & DTI

48 DOPPLER TISSUE IMAGING MITRAL VALVE PROLAPSE, COLOR M-MODE USING DTI

49 COLOR-FLOW & POWER DOPPLER IMAGING COLOR-FLOW DOPPLER POWER DOPPLER

50 POWER DOPPLER & SPECTRAL DOPPLER

51 POWER DOPPLER

52 Answers to the following TWENTY FOUR practice questions were derived from material in the textbook:

53 Question 1 What describes the Doppler shift from moving reflectors? low when the Doppler angle is close to zero the sum of the transmitted and received frequencies high if the Doppler angle is 90 degrees the difference between the transmitted and received frequencies Page 59

54 Question 1 What describes the Doppler shift from moving reflectors? low when the Doppler angle is close to zero the sum of the transmitted and received frequencies high if the Doppler angle is 90 degrees the difference between the transmitted and received frequencies Page 59

55 The drawings represent reflectors that are moving at the same speed. The arrows indicate the direction of movement relative to the respective transducer. Which reflector causes the LOWEST returning frequency? A B C D E Question 2 Pages 59 and 61

56 The drawings represent reflectors that are moving at the same speed. The arrows indicate the direction of movement relative to the respective transducer. Which reflector causes the LOWEST returning frequency? A B C D E Question 2 Pages 59 and 61

57 Question 3 What does a Doppler system measure? frequency shift while calculating blood velocity frequency shift while calculating sound velocity frequency shift while calculating attenuation blood velocity while calculating frequency shift Page 60

58 Question 3 What does a Doppler system measure? frequency shift while calculating blood velocity frequency shift while calculating sound velocity frequency shift while calculating attenuation blood velocity while calculating frequency shift Page 60

59 Question 4 Which one of the following factors does not affect the frequency of the Doppler shift? size of the Doppler probe angle at which the probe is pointed at the vessel velocity of blood in the vessel transmitted frequency Page 60

60 Question 4 Which one of the following factors does not affect the frequency of the Doppler shift? size of the Doppler probe angle at which the probe is pointed at the vessel velocity of blood in the vessel transmitted frequency Page 60

61 Question 5 What is an advantage of continuous wave Doppler over pulsed Doppler? a lower Nyquist limit spectral analysis is not required a wider range of shift frequencies without aliasing depth selectivity is possible Page 63

62 Question 5 What is an advantage of continuous wave Doppler over pulsed Doppler? a lower Nyquist limit spectral analysis is not required a wider range of shift frequencies without aliasing depth selectivity is possible Page 63

63 Question 6 When will aliasing begin to occur? when the Doppler shift is one-half the Nyquist limit when the Doppler shift exceeds one-fourth the PRF when the Doppler shift exceeds the Nyquist limit when the Doppler shift is 50% of the Nyquist limit Pages 66 through 68

64 Question 6 When will aliasing begin to occur? when the Doppler shift is one-half the Nyquist limit when the Doppler shift exceeds one-fourth the PRF when the Doppler shift exceeds the Nyquist limit when the Doppler shift is 50% of the Nyquist limit Pages 66 through 68

65 Question 7 Which one of the following uses a method other than just measuring frequency shift to detect movement? CW Doppler Power Doppler PW Doppler color-flow Doppler Page 80

66 Question 7 Which one of the following uses a method other than just measuring frequency shift to detect movement? CW Doppler Power Doppler PW Doppler color-flow Doppler Page 80

67 Question 8 What causes the problem in this PW spectral display? transducer frequency too low aliasing spectral broadening excessive output power Pages 66 through 68

68 Question 8 What causes the problem in this PW spectral display? transducer frequency too low aliasing spectral broadening excessive output power Pages 66 through 68

69 Question 9 What can be done to correct the problem in the spectral display? use a higher frequency transducer decrease the output power increase the Doppler angle decrease the PRF Pages 66 through 68

70 Question 9 What can be done to correct the problem in the spectral display? use a higher frequency transducer decrease the output power increase the Doppler angle decrease the PRF Pages 66 through 68

71 Question 10 What is the result of the gate settings in this image? aliasing and an accurate velocity calculation spectral broadening and an accurate velocity calculation a thin spectral envelope and an accurate velocity measurement spectral broadening and an inaccurate velocity calculation Pages 65 and 70

72 Question 10 What is the result of the gate settings in this image? aliasing and an accurate velocity calculation spectral broadening and an accurate velocity calculation a thin spectral envelope and an accurate velocity measurement spectral broadening and an inaccurate velocity calculation Pages 65 and 70

73 Question 11 What is an ultrasound scanner that uses a single probe to display a real time image along with spectral information? a duplex system an annular array a triplex system a CW Doppler system Page 65

74 Question 11 What is an ultrasound scanner that uses a single probe to display a real time image along with spectral information? a duplex system an annular array a triplex system a CW Doppler system Page 65

75 Question 12 What does the letter A represent in the spectral display? aliasing the peak Doppler shift the mean Doppler shift the negative Nyquist Limit Page 62

76 Question 12 What does the letter A represent in the spectral display? aliasing the peak Doppler shift the mean Doppler shift the negative Nyquist Limit Page 62

77 Question 14 What does the letter C represent in the spectral display? a large window the peak Doppler shift PW Doppler shifts from a very small sample volume spectral broadening Page 62

78 Question 13 What does the letter B represent in the spectral display? the zero baseline the positive velocity scale the peak Doppler shift the positive frequency shift scale Page 62

79 Question 13 What does the letter B represent in the spectral display? the zero baseline the positive velocity scale the peak Doppler shift the positive frequency shift scale Page 62

80 Question 14 What does the letter C represent in the spectral display? a large window the peak Doppler shift PW Doppler shifts from a very small sample volume spectral broadening Page 62

81 Question 15 What does the letter D represent in the spectral display? the spectral baseline the Nyquist Limit the peak Doppler shift the positive velocity scale Page 62

82 Question 15 What does the letter D represent in the spectral display? the spectral baseline the Nyquist Limit the peak Doppler shift the positive velocity scale Page 62

83 Question 16 What does the letter E represent in the spectral display? the negative PW frequency shift scale the negative CW frequency shift scale the peak Doppler shift the negative velocity scale Page 62

84 Question 16 What does the letter E represent in the spectral display? the negative PW frequency shift scale the negative CW frequency shift scale the peak Doppler shift the negative velocity scale Page 62

85 Question 17 Which of the following is a limitation of CW Doppler? frequent aliasing depth selectivity is not possible inability to detect peak velocities a low Nyquist Limit Page 63

86 Question 17 Which of the following is a limitation of CW Doppler? frequent aliasing depth selectivity is not possible inability to detect peak velocities a low Nyquist Limit Page 63

87 Question 18 What is the process used for color-flow Doppler to automatically assess the data from multiple sampling sites to produce a display representing mean Doppler shift frequencies? autocorrelation quadrature phase detection FFT analysis zero crossing detection Page 77

88 Question 18 What is the process used for color-flow Doppler to automatically assess the data from multiple sampling sites to produce a display representing mean Doppler shift frequencies? autocorrelation quadrature phase detection FFT analysis zero crossing detection Page 77

89 Question 19 When will aliasing begin to occur when using a 5 MHz Doppler system with a PW PRF of 15 khz? when the Doppler shift exceeds 5 khz when the Doppler shift exceeds 7.5 khz when the Doppler shift exceeds 15 khz when the Doppler shift exceeds 30 khz Pages 66 through 68

90 Question 19 When will aliasing begin to occur when using a 5 MHz Doppler system with a PW PRF of 15 khz? when the Doppler shift exceeds 5 khz when the Doppler shift exceeds 7.5 khz when the Doppler shift exceeds 15 khz when the Doppler shift exceeds 30 khz Pages 66 through 68

91 Question 20 Which Doppler waveform produces the highest pulsatility index? biphasic monophasic triphasic post-stenotic Page 73

92 Question 20 Which Doppler waveform produces the highest pulsatility index? biphasic monophasic triphasic post-stenotic Page 73

93 Question 21 Where is aliasing displayed in the color-flow Doppler image? A B C D Page 77

94 Question 21 Where is aliasing displayed in the color-flow Doppler image? A B C D Page 77

95 Question 22 Why is a black area present in the vessel on the colorflow Doppler image? Doppler angle of 90 degrees reversal of blood flow within the vessel aliasing plaque Page 75

96 Question 22 Why is a black area present in the vessel on the colorflow Doppler image? Doppler angle of 90 degrees reversal of blood flow within the vessel aliasing plaque Page 75

97 Question 23 The direction of flow in the color-flow Doppler image is red to blue right (R) to left (L) left (L) to right (R) red to black Page 75

98 Question 23 The direction of flow in the color-flow Doppler image is red to blue right (R) to left (L) left (L) to right (R) red to black Page 75

99 Question 24 Which letter in the color-flow Doppler image represents the greatest Doppler shift? A B C D E Pages 59, 75, and 80

100 Question 24 Which letter in the color-flow Doppler image represents the greatest Doppler shift? A B C D E Pages 59, 75, and 80

101 END OF LESSON 12 For information on the accompanying textbook, visit the Website:

SONOGRAPHIC PHYSICS, INSTRUMENTATION & DOPPLER REVIEW Part 3

SONOGRAPHIC PHYSICS, INSTRUMENTATION & DOPPLER REVIEW Part 3 SONOGRAPHIC PHYSICS, INSTRUMENTATION & DOPPLER REVIEW 2012 Part 3 1 Doppler Imaging 2 DOPPLER TRANSDUCER SAME FREQUENCY During Doppler operation, the reflected sound has the same frequency as the transmitted

More information

Doppler Ultrasound. Amanda Watson.

Doppler Ultrasound. Amanda Watson. Doppler Ultrasound Amanda Watson amanda.watson1@nhs.net Before we start Why does blood appear black on a B-mode image? B-mode echoes vs. Doppler echoes In B-Mode we are concerned with the position and

More information

Introduction to Ultrasound Physics

Introduction to Ultrasound Physics Introduction to Ultrasound Physics Vassilis Sboros Medical Physics and Cardiovascular Sciences University of Edinburgh Transverse waves Water remains in position Disturbance traverse producing more wave

More information

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher Chapter on Doppler ultrasound: principles and practice by Colin Deane INTRODUCTION Competent use of Doppler ultrasound techniques requires

More information

4 Working With Scan Modes

4 Working With Scan Modes 4 Working With Scan Modes Scan Modes Overview All of the information in this chapter pertains to live imaging. Many of the controls and functions change when you freeze the scan. For information on using

More information

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13

The Physics of Echo. The Physics of Echo. The Physics of Echo Is there pericardial calcification? 9/30/13 Basic Ultrasound Physics Kirk Spencer MD Speaker has no disclosures to make Sound Audible range 20Khz Medical ultrasound Megahertz range Advantages of imaging with ultrasound Directed as a beam Tomographic

More information

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล Diagnosis TTE TEE ICE 3D 4D Evaluation of Cardiac Anatomy Hemodynamic

More information

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures Optimisation of Image Acquisition Bordeaux 16th November 2016 J.S. McGhie W.B. Vletter R. Frowijn No disclosures Image optimisation: The Echo machine It looks difficult to drive an echo machine!! Some

More information

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D

Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D DINFO Dipartimento di Ingegneria dell Informazione Department of Information Engineering Nuove tecnologie per ecografia ad ultrasuoni: da 2D a 4D Piero Tortoli Microelectronics Systems Design Lab 1 Introduction

More information

S S S2 Operation Manual

S S S2 Operation Manual PREPARATION 1. How to create and input patient data? In the MAIN INTERFACE, press the key, to enter into the patient exam list interface. Then, click New patient item to create new patient files. Patient

More information

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes These lessons contain 26 slides plus 15 multiple-choice questions. These lesson were derived from pages 26 through 32 in the textbook: ULTRASOUND IMAGING

More information

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT Image Optimization: The Sonographer s Responsibility Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT Image Optimization: The Sonographer s Responsibility Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT Disclosure

More information

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions.

Lesson 02: Sound Wave Production. This lesson contains 24 slides plus 11 multiple-choice questions. Lesson 02: Sound Wave Production This lesson contains 24 slides plus 11 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 2 through 7 in the textbook: ULTRASOUND

More information

Ultrasound & Artifacts

Ultrasound & Artifacts ISSN 2005-7881 Journal of Neurosonology 3(Suppl. 2):1-17, 2011 Ultrasound & Artifacts Siryung Han The Catholic University of Korea Artifacts False image- echoes without anatomic correlate US image dose

More information

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust

The physics of ultrasound. Dr Graeme Taylor Guy s & St Thomas NHS Trust The physics of ultrasound Dr Graeme Taylor Guy s & St Thomas NHS Trust Physics & Instrumentation Modern ultrasound equipment is continually evolving This talk will cover the basics What will be covered?

More information

EMBEDDED DOPPLER ULTRASOUND SIGNAL PROCESSING USING FIELD PROGRAMMABLE GATE ARRAYS

EMBEDDED DOPPLER ULTRASOUND SIGNAL PROCESSING USING FIELD PROGRAMMABLE GATE ARRAYS EMBEDDED DOPPLER ULTRASOUND SIGNAL PROCESSING USING FIELD PROGRAMMABLE GATE ARRAYS Diaa ElRahman Mahmoud, Abou-Bakr M. Youssef and Yasser M. Kadah Biomedical Engineering Department, Cairo University, Giza,

More information

Medical Imaging (EL582/BE620/GA4426)

Medical Imaging (EL582/BE620/GA4426) Medical Imaging (EL582/BE620/GA4426) Jonathan Mamou, PhD Riverside Research Lizzi Center for Biomedical Engineering New York, NY jmamou@riversideresearch.org On behalf of Prof. Daniel Turnbull Outline

More information

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions. Lesson 06: Pulse-echo Imaging and Display Modes This lesson contains 22 slides plus 15 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 26 through 32 in

More information

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound Ultrasound Physics History: Ultrasound Ultrasound 1942: Dr. Karl Theodore Dussik transmission ultrasound investigation of the brain 1949-51: Holmes and Howry subject submerged in water tank to achieve

More information

12/26/2017. Alberto Ardon M.D.

12/26/2017. Alberto Ardon M.D. Alberto Ardon M.D. 1 Preparatory Work Ultrasound Physics http://www.nysora.com/mobile/regionalanesthesia/foundations-of-us-guided-nerve-blockstechniques/index.1.html Basic Ultrasound Handling https://www.youtube.com/watch?v=q2otukhrruc

More information

Quick Reference Guide

Quick Reference Guide siemens.com/nx3 Quick Reference Guide ACUSON NX3 Series Contents 2 System Overview 3 Getting Started 8 2D Mode and M-mode 12 Color and Spectral Doppler 24 Measurements and Calculations 38 Text, Arrows

More information

FREQUENTLY ASKED QUESTIONS

FREQUENTLY ASKED QUESTIONS FREQUENTLY ASKED S Q9: Expanding aperture is a technique in ultrasound beamforming where the objects very close to the aperture are imaged by a lower number of aperture elements. That is, as the depth

More information

Ultrasound Imaging Ultr Michael Dadd 2007

Ultrasound Imaging Ultr Michael Dadd 2007 Ultrasound Imaging Ultrasound Physics & Instrumentation - Recommended Reading 1. Diagnostic Ultrasound: Principles and Instruments (7th Ed) Frederick W Kremkau W B Saunders Company 2. Applied Physics &

More information

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time

Chapter 4. Pulse Echo Imaging. where: d = distance v = velocity t = time Chapter 4 Pulse Echo Imaging Ultrasound imaging systems are based on the principle of pulse echo imaging. These systems require the use of short pulses of ultrasound to create two-dimensional, sectional

More information

(12) United States Patent (10) Patent No.: US 6,390,983 B1. MO et al. (45) Date of Patent: May 21, 2002

(12) United States Patent (10) Patent No.: US 6,390,983 B1. MO et al. (45) Date of Patent: May 21, 2002 USOO6390983B1 (12) United States Patent (10) Patent No.: MO et al. (45) Date of Patent: May 21, 2002 (54) METHOD AND APPARATUS FOR (56) References Cited AUTOMATIC MUTING OF DOPPLER NOISE INDUCED BY ULTRASOUND

More information

MAKING TRANSIENT ANTENNA MEASUREMENTS

MAKING TRANSIENT ANTENNA MEASUREMENTS MAKING TRANSIENT ANTENNA MEASUREMENTS Roger Dygert, Steven R. Nichols MI Technologies, 1125 Satellite Boulevard, Suite 100 Suwanee, GA 30024-4629 ABSTRACT In addition to steady state performance, antennas

More information

BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM

BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM BEAM DISTORTION IN DOPPLER ULTRASOUND FLOW TEST RIGS: MEASUREMENT USING A STRING PHANTOM R. Steel, P. J. Fish School of Informatics, University of Wales, Bangor, UK Abstract-The tube in flow rigs used

More information

EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss

EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss EENG473 Mobile Communications Module 3 : Week # (12) Mobile Radio Propagation: Small-Scale Path Loss Introduction Small-scale fading is used to describe the rapid fluctuation of the amplitude of a radio

More information

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths. Q47. When performing a sonogram why the sonographer needs to use the TGC? TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow

More information

Physics of ultrasound

Physics of ultrasound 1 Physics of ultrasound Basic principles Nature of ultrasound Sound = longitudinal, mechanical wave particles move parallel to direction of travel Audible sound < 20 khz Ultrasound > 20 khz Sound cannot

More information

M5 Diagnostic Ultrasound System

M5 Diagnostic Ultrasound System V0807 M5 Diagnostic Ultrasound System Mindray s ultrasound family is now introducing a new member, M5 hand-carried color Doppler system. M5, coming in a laptop size with comprehensive ergonomic design,

More information

Detection of Targets in Noise and Pulse Compression Techniques

Detection of Targets in Noise and Pulse Compression Techniques Introduction to Radar Systems Detection of Targets in Noise and Pulse Compression Techniques Radar Course_1.ppt ODonnell 6-18-2 Disclaimer of Endorsement and Liability The video courseware and accompanying

More information

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University

Introduction to Medical Engineering (Medical Imaging) Ultrasound Imaging. Ho Kyung Kim Pusan National University Introduction to Medical Engineering (Medical Imaging) Suetens 6 Ultrasound Imaging Ho Kyung Kim Pusan National University Sound Sonic: 20 Hz 20 khz (audible frequency) Subsonic () Ultrasound

More information

Three-dimensional display of calculated velocity profiles for physiological flow waveforms

Three-dimensional display of calculated velocity profiles for physiological flow waveforms Three-dimensional display of calculated velocity profiles for physiological flow waveforms Ramez E. N. Shehada, MSc, Richard S. C. Cobbold, Phi), K. Wayne Johnston, MD, FRCS(C), and Rene Aarnink, MSc,

More information

SPATIAL PATTERNS and dynamics of blood flow are primary

SPATIAL PATTERNS and dynamics of blood flow are primary 524 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 53, NO. 2, APRIL 2004 Doppler Ultrasound Systems Designed for Tumor Blood Flow Imaging Christian Kargel, Member, IEEE, Gernot Plevnik, Birgit

More information

Diagnostic Ultrasound System. Operation Note

Diagnostic Ultrasound System. Operation Note M5 Diagnostic Ultrasound System Table of Contents System Introduction...3 Control Panel...4 Control Panel...5 Control Panel...6 Control Panel...7 Control Panel...8 Control Panel...9 Power ON / OFF the

More information

Keysight Technologies Pulsed Antenna Measurements Using PNA Network Analyzers

Keysight Technologies Pulsed Antenna Measurements Using PNA Network Analyzers Keysight Technologies Pulsed Antenna Measurements Using PNA Network Analyzers White Paper Abstract This paper presents advances in the instrumentation techniques that can be used for the measurement and

More information

Low wavenumber reflectors

Low wavenumber reflectors Low wavenumber reflectors Low wavenumber reflectors John C. Bancroft ABSTRACT A numerical modelling environment was created to accurately evaluate reflections from a D interface that has a smooth transition

More information

An Introduction to FFT EMI Receivers

An Introduction to FFT EMI Receivers An Introduction to FFT EMI Receivers Introduction An evolution in EMI receiver design is underway to take advantage of today s digital signal processing (DSP) technologies, using fast Fourier transform

More information

Pulse Doppler Flow-Dop

Pulse Doppler Flow-Dop EDUCATION Pulse Doppler Flow-Dop GAMPT-50100 User Manual Fon: +49 (0) 3461-278 691-0 Fax: +49 (0) 3461-278 691-101 email: info@gampt.de Gesellschaft für Angewandte Medizinische Physik und Technik mbh (GAMPT

More information

DC-6. Diagnostic Ultrasound System

DC-6. Diagnostic Ultrasound System DC-6 Diagnostic Ultrasound System DC-6 is a general purpose color Doppler ultrasound system aiming at most clinical areas both in exam and research with various transducers and multi software packages

More information

Advanced signal processing methods for planewave color Doppler ultrasound imaging

Advanced signal processing methods for planewave color Doppler ultrasound imaging Western University Scholarship@Western Electronic Thesis and Dissertation Repository May 2017 Advanced signal processing methods for planewave color Doppler ultrasound imaging Omar Mansour The University

More information

3. Ultrasound Imaging(2)

3. Ultrasound Imaging(2) 3. Ultrasound Imaging(2) Lecture 13, 14 Medical Imaging Systems Jae Gwan Kim, Ph.D. jaekim@gist.ac.kr, X 2220 Department of BioMedical Science and Engineering Gwangju Institute of Sciences and Technology

More information

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images Artifacts Artifacts Chapter 20 What are they? Simply put they are an error in imaging These artifacts include reflections that are: not real incorrect shape, size or position incorrect brightness displayed

More information

monitoring device ought to find wide clinical application. available, while section three describes the Doppler interface with the B-scan machine.

monitoring device ought to find wide clinical application. available, while section three describes the Doppler interface with the B-scan machine. THE YALE JOURNAL OF BIOLOGY AND MEDICINE 50 (1977), 367-373 Pulse-Doppler Ultrasound and Its Clinical Application PETER ATKINSON AND PETER N.T. WELLS Yale University School of Medicine, New Haven, Connecticut

More information

Photomultiplier Tube

Photomultiplier Tube Nuclear Medicine Uses a device known as a Gamma Camera. Also known as a Scintillation or Anger Camera. Detects the release of gamma rays from Radionuclide. The radionuclide can be injected, inhaled or

More information

KIZLON. Sentinel House,Harvest Crescent, Ancells Business Park, Fleet Hampshire GU51 2UZ, UK Website:

KIZLON. Sentinel House,Harvest Crescent, Ancells Business Park, Fleet Hampshire GU51 2UZ, UK   Website: KIZLON Sentinel House,Harvest Crescent, Ancells Business Park, Fleet Hampshire GU51 2UZ, UK Email: info@kizlon.com Website: www.kizlon.com Portable Ultrasound Scanners KUS-A100 & A101 Ultrasound Scanners

More information

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc. P a g e 1 ST985 TDR Cable Analyzer Instruction Manual Analog Arts Inc. www.analogarts.com P a g e 2 Contents Software Installation... 4 Specifications... 4 Handling Precautions... 4 Operation Instruction...

More information

Chapter 8. Ultrasound Medical Diagnostic Instrumentation PART-I

Chapter 8. Ultrasound Medical Diagnostic Instrumentation PART-I Chapter 8 Ultrasound Medical Diagnostic Instrumentation PART-I ULTRASONAGRAPHY Ultrasonography is a technique by which ultrasonic energy is used to detect internal body organs. Bursts of ultrasound energy

More information

A COMPACT SYSTEM WITH ADVANCED PERFORMANCE

A COMPACT SYSTEM WITH ADVANCED PERFORMANCE Samsung Medison is a global leading medical devices company. Founded in 1985, the company sells cutting-edge diagnostic ultrasound devices around the world in various medical fields. The company has attracted

More information

Recently there has been a dramatic increase in the. Performance Tests of Doppler Ultrasound Equipment With a Tissue and Blood Mimicking Phantom

Recently there has been a dramatic increase in the. Performance Tests of Doppler Ultrasound Equipment With a Tissue and Blood Mimicking Phantom Performance Tests of Doppler Ultrasound Equipment With a Tissue and Blood Mimicking Phantom Evan J. Boote, MS, James A. Zagzebski, PhD A tissue- and blood-mimicking phantom was assembled for assessing

More information

Radar-Verfahren und -Signalverarbeitung

Radar-Verfahren und -Signalverarbeitung Radar-Verfahren und -Signalverarbeitung - Lesson 2: RADAR FUNDAMENTALS I Hon.-Prof. Dr.-Ing. Joachim Ender Head of Fraunhoferinstitut für Hochfrequenzphysik and Radartechnik FHR Neuenahrer Str. 20, 53343

More information

Introduction to Radar Systems. Clutter Rejection. MTI and Pulse Doppler Processing. MIT Lincoln Laboratory. Radar Course_1.ppt ODonnell

Introduction to Radar Systems. Clutter Rejection. MTI and Pulse Doppler Processing. MIT Lincoln Laboratory. Radar Course_1.ppt ODonnell Introduction to Radar Systems Clutter Rejection MTI and Pulse Doppler Processing Radar Course_1.ppt ODonnell 10-26-01 Disclaimer of Endorsement and Liability The video courseware and accompanying viewgraphs

More information

Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization

Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization Doppler Ultrasound Blood Flow Measurement System for Assessing Coronary Revascularization J. Solano 1, M. Fuentes 1, A. Villar 2 J. Prohias 2, F. García-Nocetti 1 1 Universidad Nacional Autónoma de México,

More information

Explain what is meant by a photon and state one of its main properties [2]

Explain what is meant by a photon and state one of its main properties [2] 1 (a) A patient has an X-ray scan taken in hospital. The high-energy X-ray photons interact with the atoms inside the body of the patient. Explain what is meant by a photon and state one of its main properties....

More information

ACUSON X150 Knobology & User Guide

ACUSON X150 Knobology & User Guide ACUSON X150 Knobology & User Guide ACUSON X150 CONTROL PANEL Table of Contents The ACUSON X150 knobology and user guide is the clinicians quick reference of system terms, functions and capabilities. This

More information

Coded Ultrasound for Blood Flow Estimation Using Subband Processing

Coded Ultrasound for Blood Flow Estimation Using Subband Processing IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, vol. 55, no. 10, October 2008 2211 Coded Ultrasound for Blood Flow Estimation Using Subband Processing Fredrik Gran, Jesper Udesen,

More information

Beamforming in ultrasound

Beamforming in ultrasound Peter Pazmany Catholic University Faculty of Information Technology www.itk.ppke.hu Medical diagnostic systems (Orvosbiológiai képalkotó rendszerek) Beamforming in ultrasound ( Nyalábalkotás az ultrahangban)

More information

Pulsed VNA Measurements:

Pulsed VNA Measurements: Pulsed VNA Measurements: The Need to Null! January 21, 2004 presented by: Loren Betts Copyright 2004 Agilent Technologies, Inc. Agenda Pulsed RF Devices Pulsed Signal Domains VNA Spectral Nulling Measurement

More information

DOPPLER RADAR. Doppler Velocities - The Doppler shift. if φ 0 = 0, then φ = 4π. where

DOPPLER RADAR. Doppler Velocities - The Doppler shift. if φ 0 = 0, then φ = 4π. where Q: How does the radar get velocity information on the particles? DOPPLER RADAR Doppler Velocities - The Doppler shift Simple Example: Measures a Doppler shift - change in frequency of radiation due to

More information

Set No.1. Code No: R

Set No.1. Code No: R Set No.1 IV B.Tech. I Semester Regular Examinations, November -2008 RADAR SYSTEMS ( Common to Electronics & Communication Engineering and Electronics & Telematics) Time: 3 hours Max Marks: 80 Answer any

More information

Spectral Velocity Estimation using the Autocorrelation Function and Sparse Data Sequences

Spectral Velocity Estimation using the Autocorrelation Function and Sparse Data Sequences Spectral Velocity Estimation using the Autocorrelation Function and Sparse Data Sequences Jørgen Arendt Jensen Ørsted DTU, Build. 348, Technical University of Denmark, DK-8 Lyngby, Denmark Abstract Ultrasound

More information

Development and Application of 500MSPS Digitizer for High Resolution Ultrasonic Measurements

Development and Application of 500MSPS Digitizer for High Resolution Ultrasonic Measurements Indian Society for Non-Destructive Testing Hyderabad Chapter Proc. National Seminar on Non-Destructive Evaluation Dec. 7-9, 2006, Hyderabad Development and Application of 500MSPS Digitizer for High Resolution

More information

Blood Flow Analysis Using Ultrasound Doppler Technique

Blood Flow Analysis Using Ultrasound Doppler Technique IOSR Journal of Electronics and Communication Engineering (IOSR-JECE) e-issn: 2278-2834,p- ISSN: 2278-8735.Volume 10, Issue 5, Ver. I (Sep - Oct.2015), PP 31-35 www.iosrjournals.org Blood Flow Analysis

More information

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Paper presented at the 23rd Acoustical Imaging Symposium, Boston, Massachusetts, USA, April 13-16, 1997: COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES Jørgen Arendt Jensen and Peter

More information

Ultra Wide Band Communications

Ultra Wide Band Communications Lecture #3 Title - October 2, 2018 Ultra Wide Band Communications Dr. Giuseppe Caso Prof. Maria-Gabriella Di Benedetto Lecture 3 Spectral characteristics of UWB radio signals Outline The Power Spectral

More information

Multi-Path Fading Channel

Multi-Path Fading Channel Instructor: Prof. Dr. Noor M. Khan Department of Electronic Engineering, Muhammad Ali Jinnah University, Islamabad Campus, Islamabad, PAKISTAN Ph: +9 (51) 111-878787, Ext. 19 (Office), 186 (Lab) Fax: +9

More information

EE 3302 LAB 1 EQIUPMENT ORIENTATION

EE 3302 LAB 1 EQIUPMENT ORIENTATION EE 3302 LAB 1 EQIUPMENT ORIENTATION Pre Lab: Calculate the theoretical gain of the 4 th order Butterworth filter (using the formula provided. Record your answers in Table 1 before you come to class. Introduction:

More information

A Breakthrough in Sputtering Target Inspections: Ultra-High Speed Phased Array Scanning with Volume Focusing

A Breakthrough in Sputtering Target Inspections: Ultra-High Speed Phased Array Scanning with Volume Focusing 17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China A Breakthrough in Sputtering Target Inspections: Ultra-High Speed Phased Array Scanning with Volume Focusing Dominique Braconnier,

More information

Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar

Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar Lecture 21. Wind Lidar (3) Direct Detection Doppler Lidar Overview of Direct Detection Doppler Lidar (DDL) Resonance fluorescence DDL Fringe imaging DDL Scanning FPI DDL FPI edge-filter DDL Absorption

More information

Chapter-15. Communication systems -1 mark Questions

Chapter-15. Communication systems -1 mark Questions Chapter-15 Communication systems -1 mark Questions 1) What are the three main units of a Communication System? 2) What is meant by Bandwidth of transmission? 3) What is a transducer? Give an example. 4)

More information

Introduction to Lab Instruments

Introduction to Lab Instruments ECE316, Experiment 00, 2017 Communications Lab, University of Toronto Introduction to Lab Instruments Bruno Korst - bkf@comm.utoronto.ca Abstract This experiment will review the use of three lab instruments

More information

Radar Systems Engineering Lecture 12 Clutter Rejection

Radar Systems Engineering Lecture 12 Clutter Rejection Radar Systems Engineering Lecture 12 Clutter Rejection Part 1 - Basics and Moving Target Indication Dr. Robert M. O Donnell Guest Lecturer Radar Systems Course 1 Block Diagram of Radar System Transmitter

More information

Ultrasound Transducers for Q+ Systems

Ultrasound Transducers for Q+ Systems Analogic Ultrasound Product Specifications Ultrasound Transducers for Q+ Systems 00.050.231, Revision A, May 9, 2014 US Patents 6,216,029-6,325,759-6,558,326-6,911,008-7,274,325-8,088,070-8,556,815 - D654,178

More information

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives:

Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Objectives: Advanced Lab LAB 6: Signal Acquisition & Spectrum Analysis Using VirtualBench DSA Equipment: Pentium PC with National Instruments PCI-MIO-16E-4 data-acquisition board (12-bit resolution; software-controlled

More information

Vectrino Micro ADV Comparison

Vectrino Micro ADV Comparison Nortek Technical Note No.: TN-022 Title: Vectrino Micro ADV comparison Last edited: November 19, 2004 Authors: Atle Lohrmann, NortekAS, Chris Malzone, NortekUSA Number of pages: 12 Overview This brief

More information

Local Measurement of the Pulse Wave Velocity using Doppler Ultrasound. Minnan Xu

Local Measurement of the Pulse Wave Velocity using Doppler Ultrasound. Minnan Xu Local Measurement of the Pulse Wave Velocity using Doppler Ultrasound by Minnan Xu Submitted to the Department of Electrical Engineering and Computer Science in partial fulfillment of the requirements

More information

DC-6 Expert. Diagnostic Ultrasound System

DC-6 Expert. Diagnostic Ultrasound System DC-6 Expert Diagnostic Ultrasound System MINDRAY has newly released DC-6 Expert, a general purpose color Doppler ultrasound system with full ergonomic designs, supplying more accessible exams, higher imaging

More information

Blood velocities estimation using ultrasound. Elham Pirnia Supervised by: Andreas Jakobsson Examiner: Maria Sandsten

Blood velocities estimation using ultrasound. Elham Pirnia Supervised by: Andreas Jakobsson Examiner: Maria Sandsten Blood velocities estimation using ultrasound Elham Pirnia Supervised by: Andreas Jakobsson Examiner: Maria Sandsten Center for Mathematical Science, Mathematical Statistics Lund University, Sweden September

More information

Research Article Embedded System for Real-Time Digital Processing of Medical Ultrasound Doppler Signals

Research Article Embedded System for Real-Time Digital Processing of Medical Ultrasound Doppler Signals Hindawi Publishing Corporation EURASIP Journal on Advances in Signal Processing Volume 28, Article ID 418235, 7 pages doi:1.1155/28/418235 Research Article Embedded System for Real-Time Digital Processing

More information

Technician License Course Chapter 2. Lesson Plan Module 2 Radio Signals and Waves

Technician License Course Chapter 2. Lesson Plan Module 2 Radio Signals and Waves Technician License Course Chapter 2 Lesson Plan Module 2 Radio Signals and Waves The Basic Radio Station What Happens During Radio Communication? Transmitting (sending a signal): Information (voice, data,

More information

Chapter 5 Window Functions. periodic with a period of N (number of samples). This is observed in table (3.1).

Chapter 5 Window Functions. periodic with a period of N (number of samples). This is observed in table (3.1). Chapter 5 Window Functions 5.1 Introduction As discussed in section (3.7.5), the DTFS assumes that the input waveform is periodic with a period of N (number of samples). This is observed in table (3.1).

More information

Chapter Two. Fundamentals of Data and Signals. Data Communications and Computer Networks: A Business User's Approach Seventh Edition

Chapter Two. Fundamentals of Data and Signals. Data Communications and Computer Networks: A Business User's Approach Seventh Edition Chapter Two Fundamentals of Data and Signals Data Communications and Computer Networks: A Business User's Approach Seventh Edition After reading this chapter, you should be able to: Distinguish between

More information

Laboratory Experiment #1 Introduction to Spectral Analysis

Laboratory Experiment #1 Introduction to Spectral Analysis J.B.Francis College of Engineering Mechanical Engineering Department 22-403 Laboratory Experiment #1 Introduction to Spectral Analysis Introduction The quantification of electrical energy can be accomplished

More information

Muhammad Ali Jinnah University, Islamabad Campus, Pakistan. Fading Channel. Base Station

Muhammad Ali Jinnah University, Islamabad Campus, Pakistan. Fading Channel. Base Station Fading Lecturer: Assoc. Prof. Dr. Noor M Khan Department of Electronic Engineering, Muhammad Ali Jinnah University, Islamabad Campus, Islamabad, PAKISTAN Ph: +9 (51) 111-878787, Ext. 19 (Office), 186 (ARWiC

More information

Chapter 7. Introduction. Analog Signal and Discrete Time Series. Sampling, Digital Devices, and Data Acquisition

Chapter 7. Introduction. Analog Signal and Discrete Time Series. Sampling, Digital Devices, and Data Acquisition Chapter 7 Sampling, Digital Devices, and Data Acquisition Material from Theory and Design for Mechanical Measurements; Figliola, Third Edition Introduction Integrating analog electrical transducers with

More information

Product Description. System Architecture. Display Formats. Data Acquisition. Display Annotations. Data Processing. Display Screen

Product Description. System Architecture. Display Formats. Data Acquisition. Display Annotations. Data Processing. Display Screen Vivid i n Product Description The Vivid i n is a high-performance, battery-operated, ultra-portable diagnostic Ultrasound systems providing exceptional image quality. The Vivid i n is designed for cardiovascular

More information

Echo Artifacts: The Cause and Solution

Echo Artifacts: The Cause and Solution Echo Artifacts: The Cause and Solution David Adams, RCS, RDCS, FASE Duke University Medical Center Disclosures None My Happy / Sad ratio 20% Sad 1 st talk on Sunday (post party) Talk about Artifacts Artifacts

More information

Data and Computer Communications Chapter 3 Data Transmission

Data and Computer Communications Chapter 3 Data Transmission Data and Computer Communications Chapter 3 Data Transmission Eighth Edition by William Stallings Transmission Terminology data transmission occurs between a transmitter & receiver via some medium guided

More information

Velocity Estimation in muscular Tissue by Ultrasound

Velocity Estimation in muscular Tissue by Ultrasound Velocity Estimation in muscular Tissue by Ultrasound Trond-Olav Dahl Master of Science in Communication Technology Submission date: June 2007 Supervisor: Tor Audun Ramstad, IET Co-supervisor: Hans Torp,

More information

INTRODUCTION. Strong Performance: High resolution and penetration, achieving precise flaw detection

INTRODUCTION. Strong Performance: High resolution and penetration, achieving precise flaw detection Shantou Institute of Ultrasonic Instruments Co., Ltd. Add: 77 Jinsha Road, Shantou, Guangdong 515041, China Tel: 86-754-88250150 Fax: 86-754-88251499 Http://www.siui.com/ndt Product Data CTS-9009 Digital

More information

Where DSP meets Measurement Science: A Sound Example. By Andrew Hurrell PhD

Where DSP meets Measurement Science: A Sound Example. By Andrew Hurrell PhD Where DSP meets Measurement Science: A Sound Example By Andrew Hurrell PhD Measuring ultrasound why bother? 6 million ultrasound scans within NHS during 2004-2005 Ultrasound has potential for: Thermal

More information

Terminology (1) Chapter 3. Terminology (3) Terminology (2) Transmitter Receiver Medium. Data Transmission. Simplex. Direct link.

Terminology (1) Chapter 3. Terminology (3) Terminology (2) Transmitter Receiver Medium. Data Transmission. Simplex. Direct link. Chapter 3 Data Transmission Terminology (1) Transmitter Receiver Medium Guided medium e.g. twisted pair, optical fiber Unguided medium e.g. air, water, vacuum Corneliu Zaharia 2 Corneliu Zaharia Terminology

More information

: PT Philips Indonesia CERTIFICATE OF PRODUCTION SOLE AGENT/ DISTRIBUTOR : NA CERTIFICATE OF DISTRIBUTION (PAK)

: PT Philips Indonesia CERTIFICATE OF PRODUCTION SOLE AGENT/ DISTRIBUTOR : NA CERTIFICATE OF DISTRIBUTION (PAK) SOLE AGENT/ DISTRIBUTOR : PT Philips Indonesia CERTIFICATE OF PRODUCTION : NA CERTIFICATE OF DISTRIBUTION (PAK) : HK.07.Alkes/IV/625/AK.2.2012 ADDRESS : Jl. Buncit Raya Kav 99-100, Pasar Minggu, Jak-Sel

More information

Sonic Distance Sensors

Sonic Distance Sensors Sonic Distance Sensors Introduction - Sound is transmitted through the propagation of pressure in the air. - The speed of sound in the air is normally 331m/sec at 0 o C. - Two of the important characteristics

More information

Lecture 6 SIGNAL PROCESSING. Radar Signal Processing Dr. Aamer Iqbal Bhatti. Dr. Aamer Iqbal Bhatti

Lecture 6 SIGNAL PROCESSING. Radar Signal Processing Dr. Aamer Iqbal Bhatti. Dr. Aamer Iqbal Bhatti Lecture 6 SIGNAL PROCESSING Signal Reception Receiver Bandwidth Pulse Shape Power Relation Beam Width Pulse Repetition Frequency Antenna Gain Radar Cross Section of Target. Signal-to-noise ratio Receiver

More information

Comprehensive Ultrasound Research Platform

Comprehensive Ultrasound Research Platform Comprehensive Ultrasound Research Platform Functional Requirements List and Performance Specifications Emma Muir Sam Muir Jacob Sandlund David Smith Advisor: Dr. José Sánchez Date: November 9, 2010 Introduction

More information

UNIT-3. Electronic Measurements & Instrumentation

UNIT-3.   Electronic Measurements & Instrumentation UNIT-3 1. Draw the Block Schematic of AF Wave analyzer and explain its principle and Working? ANS: The wave analyzer consists of a very narrow pass-band filter section which can Be tuned to a particular

More information

Introduction. In the frequency domain, complex signals are separated into their frequency components, and the level at each frequency is displayed

Introduction. In the frequency domain, complex signals are separated into their frequency components, and the level at each frequency is displayed SPECTRUM ANALYZER Introduction A spectrum analyzer measures the amplitude of an input signal versus frequency within the full frequency range of the instrument The spectrum analyzer is to the frequency

More information