Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14

Size: px
Start display at page:

Download "Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14"

Transcription

1 Electrical Safety Medical Instrumentation Application and Design, 4th Edition, Chapter 14 John G. Webster, Univ. of Wisconsin, Madison ISBN:

2 Taught Matter of Lectures Introduction Basic Theory of Measurements Beginnings of Basic Sensors Sensors [MEMS] Signals and Noise Amplifiers of Signals Connection and Protection of Signals Data Acquisition and Data Converters Electric Safety in Medical Systems

3 Electrical Safety Agenda Introduction Physiological Effects of Electricity Susceptibility Parameters Distribution of Electric Power Macroshock Hazards Microshock Hazards Electrical-Safety Codes and Standards Approaches to Protection Against Shock Power Distribution Equipment Design Testing the Electric System Tests of Electric Appliances Conclusion

4 Introduction ES as elemetary protection Medical technology has improved health care in ALL medical specialties, with rising complexity More than 10,000 device-related patient injuries Most patient injuries are attributable to improper use Medical personnel rarely read user manuals until a problem has occurred Result: Medical instrumentation safety Safe Design Safe Use... Safe EVERYTHING (one of the most regulated industrial market) ES is one of the basic protection mechanisms for patient, operater, and third persons and part of this chapter

5 Physiological effects Current can heal and harm For a physiological effect the body must become part of an electrical circuit Three phenomena occur when el. current flows 1. El. stimulation of excitable tissue (muscle, nerve) 2. Resistive heating of tissue 3. Electrochemical burns Further consideration are based on the following parameters Human body with contact to el. circuit at left and right hand Body weight: 70 kg Applied current time: 1 s to 3 s Current frequency: 60 Hz

6 Physiological effects Current can heal and harm Figure 14.1 Physiological effects of electricity Threshold or estimated mean values are given for each effect in a 70 kg human for a 1 to 3 s exposure to 60 Hz current applied via copper wires grasped by the hands.

7 Physiological effects Threshold of Perception (1) Current density is just large enough to excite nerve endings in the skin Subject feels tingling sensation Lower thresholds with moist/wet hands (decreases contact resistance) 0.5 ma at 60 Hz 2 ma to 10 ma DC The subject meight feel a slight warming

8 Physiological effects Let-go Current (2) The let-go current is defined as the maximal current at which the subject can withdraw voluntarily For higher current nerves and muscles are vigorously stimulated Involuntary contraction or reflex withdrawals may cause secondary physical injuries (falling off the ladder) The minimal threshold for the let-go current is 6 ma

9 Physiological effects Respiratory Paralysis, Pain, Fatigue (3) Higher current causes involuntary contraction of muscles and stimulation of nerves what can lead to pain and cause fatigue Example: stimulation of respiratory muscles lead to involuntary contraction with the result of asphyxiation if current is not interrupted Of course, today s ethics commission would never allow these experiments on human beings.

10 Physiological effects Ventricular Fibrillation (4) The heart is especially susceptible to electric current. Just 75 ma to 400 ma (AC) can rapidly disorganize the cardiac rhythm and death occurs within minutes Only a brief high-current pulse from a defibrillator can depolarize all the cells of the heart muscle simultaneously Within the U.S. occur approximately 1,000 death per year due to cordconnected appliances

11 Physiological effects Sustained Myocardial Contraction (5) When current is high enough to stimulate the entire heart muscle, it stops beating Usually the heart-beat ensues when the current is interrupted Minimal currents range from 1 A to 6 A (AC), like used in defibrillators

12 Physiological effects Burns and Physical Injury (6) Resistive heating cause burns Current can puncture the skin Brain and nerve tissue may lose all functional excitability Simultaneously stimulated muscles may contract strong enough to pull the attachment away from the bone or bread the bone

13 Susceptibility Parameters Introduction The current needed to produce each effect depends on these parameters Threshold of Perception and Let-Go Variability Frequency Duration Body Weight (and gender) Points of Entry Macroshock Microshock

14 Susceptibility Parameters Variability of threshold and Let-go current Figure 14.2 Distributions of perception thresholds and let-go currents These data depend on surface area of contact (moistened hand grasping AWG No. 8 copper wire). (Replotted from C. F. Dalziel, "Electric Shock," Advances in Biomedical Engineering, edited by J. H. U. Brown and J. F. Dickson III, , )

15 Susceptibility Parameters Frequency Let-go current versus frequency Minimal let-go currents occur for commercial power-line frequencies (50 Hz to 60 Hz) For frequencies below 10 Hz let-go current rises again (muscle can relax) Figure 14.3 Let-go current versus frequency Percentile values indicate variability of let-go current among individuals. Let-go currents for women are about two-thirds the values for men. (Reproduced, with permission, from C. F. Dalziel, "Electric Shock," Advances in Biomedical Engineering, edited by J. H. U. Brown and J. F. Dickson III, 1973, 3, )

16 Susceptibility Parameters Duration Geddes and Baker (1989) presented the excitation behavior of myocardial cells by a lumped parallel RC circuit that represents the resistance and capacitance of the cell membrane. This model determines the cell excitation thresholds that exceed about 20 mv for varying rectangular pulse duration d by assigning the rheobase currents I r and cell membrane time constant τ=rc. The strength-duration equation For a short duration: Stimulation current I d is inversely related to the pulse duration d (Figure 14.4)

17 The sub-threshold pulse increases the sodium activation m and decreases the sodium deactivation h whilst the potassium activation n increases enough to quench the firing of the neuron. The super-threshold pulse leads to the periodic oscillations of the states as shown above.

18 Susceptibility Parameters Duration Figure 14.4 Normalized analytical strength duration curve for current I, charge Q, and energy U. The x axis shows the normalized duration of d/τ (From Geddes, L. A., and L. E. Baker, Principles of Applied Biomedical Instrumentation, 3rd ed. New York: John Wiley & Sons, 1989).

19 Susceptibility Parameters Body weight Several studies (animals) show a clear dependency of the fibrillating current to the body weight (Figure 14.5) 50 ma rms for 6kg dogs to 130 ma rms for 24 kg dogs Figure 14.5 Fibrillation current versus shock duration. Thresholds for ventricular fibrillation in animals for 60 Hz ac current. Duration of current (0.2 to 5 s) and weight of animal body were varied. (From L. A. Geddes, IEEE Trans. Biomed. Eng., 1973, 20, Copyright 1973 by the Institute of Electrical and Electronics Engineers. Reproduced with permission.)

20 Susceptibility Parameters Points of Entry Macroshock: only a small fraction of the total current flows through the heart. Magnitude to harm the heart is far greater Microshock: all the current applied flows through the heart Figure 14.6 Effect of entry points on current distribution (a) Macroshock, externally applied current spreads throughout the body, (b) Microshock, all the current applied through an intracardiac catheter flows through the heart. (From F. J. Weibell, "Electrical Safety in the Hospital," Annals of Biomedical Engineering, 1974, 2, )

21 Susceptibility Parameters Points of Entry - Example 60 Hz for 5 s to a ventricular pacing catheter during implantable cardioverter-defibrillator implant testing in 40 patients Result Intermittent capture with a minimum current of 20 µa Continuous capture with a minimum current of 49 µa Resulting Regulation The widely accepted safety limit to prevent microshocks is 10 µa

22 Susceptibility Parameters Points of Entry - Example Figure Percentile plot of thresholds for continuous capture and VF (or sustained VT). Cumulative percent of patients is shown on abscissa and root-mean-square AC current (in µa) on ordinate. Squares denote unipolar data; circles, bipolar data. Solid symbols identify data from patients in whom the only clinical arrhythmia was atrial fibrillation (AF). Top, Thresholds for continuous capture. Current strength of 50 µa caused continuous capture in 5 patients (12%) with unipolar AC and in 9 (22%) with bipolar AC (P=0.49). Bottom, Thresholds for sustained VT/VF. These plots do not reach 100% because sustained-vt/vf thresholds exceeded maximum output of stimulator in 6 patients (15%) with bipolar AC and 8 (20%) with unipolar AC. From Swerdlow, C. D., W. H. Olson, M. E. O Connor, D. M. Gallik, R. A. Malkin, M. Laks, Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current Implications for electrical safety. Circulation., 1999, 99,

23 Distribution of Electric Power Introduction Electric Power is needed in health-care facilities not only for medical devices but also for any other electrical equipment like lightning, air condition, telephone, television etc. BUT Medical devices underlie special safety regulations as they might stay in special contact to and with patients, applicants and third persons 1. Overvoltage protection 2. Special ground Example A lightning causes an overvoltage at the public power supply. The overvoltage is transferred directly to the patients heart by applied ECG-Electrodes. => Over voltage protection

24 Distribution of Electric Power El. power-distribution from grid to receptacles Health-care facilities need an additional (green) ground path for all receptacles redundant to the metal (white) ground path Figure 14.8 Simplified electric-power distribution for 115 V circuits. Power frequency is 60 Hz.

25 Distribution of Electric Power Isolated-power systems Isolation transformer like this example protect systems from ground faults The Line-isolation monitor must be used to detect the occurrence of ground faults Figure 14.9 Power-isolation-transformer system with a line-isolation monitor to detect ground faults.

26 Macroshock Hazards Macroshock = current spreads through the body Two factors reduce danger in case of an electric shock 1. High skin resistance (15kOhm to 1 MOhm at 1 cm 2 ) 2. Spatial distribution Many medical devices Reduce the skin resistance with ionic gel (good electrode contact), or Bypass the natural protection by bypassing the skin (thermometer in the mouth, intravenous catheters, etc.) Many fluids conduct electricity (blood, urine, intravenous solution, etc.) Result Patients in medical-care facilities are much more susceptible to macroshocks

27 Macroshock Hazards Protection Ground fault with short circuit to a metal chassis a. not grounded chassis à macroshock b. grounded chassis à safe Figure Macroshock due to a ground fault from hot line to equipment cases for (a) ungrounded cases and (b) grounded chassis.

28 Microshock Hazards Microshock = all current flows through the heart Microshock accidents generally result from leakage-currents in line-operated equipment differences in voltage between grounded conductive surfaces due to large currents in the grounding system Microshock currents can flow either into or out of the electric connection to the heart Result Patient is only in danger of microshock if there is some electric connection to the heart

29 Microshock Hazards Protection Leakage-current flows a. through the ground wire no microshock occurs b. through the patient if he touches the chassis and has a grounded catheter etc. c. through the patient if he is touching ground and has a connected catheter etc. Figure Microshock leakage-current pathways. Assume 100 µa of leakage current from the power line to the instrument chassis, (a) Intact ground, and 99.8 µa flows through the ground, (b) Broken ground, and 100 µa flows through the heart, (c) Broken ground, and 100 µa flows through the heart in the opposite direction.

30 Microshock Hazards Conductive Path to the Heart Specific types of electric connections to the heart can be identified Epicardial or endocardial electrodes (i.e. temporary externalized pacemakers) < 1 Ohm Electrodes for intracardiac electrogramm (ECT) < 1 Ohm Liquid-filled catheters placed in the heart (i.e. measure blood preassure, withdraw blood samples, inject substances, etc.) usually 50 kohm to 1 MOhm Internal resistance of the body is about 300 Ohm

31 Microshock Hazards Conductive Path to the Heart Ventricular fibrillation and pump failure thresholds vs. electrode area Figure 14.11Microshock Assume 100 leakage-current pathways. µa of leakage current from the power line to the instrument chassis, (a) Intact ground, and 99.8 µa flows through the ground, (b) Broken ground, and 100 µa flows through the heart, (c) Broken ground, and 100 µa flows through the heart in the opposite direction.

32 Microshock Hazards Example of patient in the intensive-care unit Patient is connected to: ECG monitor that grounds the right-leg electrode to reduce 60 Hz interferce Blood-pressure monitor that monitors the left-ventricular blood-pressure Figure (a) Large ground-fault current raises the potential of one ground connection to the patient. The microshock current can then flow out through a catheter connected to a different ground, (b) Equivalent circuit. Only power-system grounds are shown.

33 Electrical-Safety Codes and Standards

34 Basic Approaches to Protection against Shock There are two fundamental methods of protecting patients against shock 1. Complete isolation and insulation from all grounded objects and all sources of electric current 2. Same potential of all conducting surfaces within reach of the patients Neither approach can be fully achieved in most practical environments, so some combination must usually suffice Protection must include patient, applicants and third party persons

35 Protection: Power Distribution Grounding System Low-resistance grounding system carry currents up to circuit-breaker ratings by keeping all conductive surfaces on the same potential (refer to Figure and 14.11) Patient-equipment grounding point Reference grounding point Connections for other patient-equipment Figure Grounding system All the receptacle grounds and conductive surfaces in the vicinity of the patient are connected to the patient-equipment grounding point. Each patientequipment grounding point is connected to the reference grounding point that makes a single connection to the building ground.

36 Protection: Power Distribution Ground-Fault Circuit Interrupter (GFCI) Ground-fault circuit interrupters disconnect the source power when a ground fault greater than about 6 ma occurs GFCI senses differences in the in- an outgoing current Most GFCI use differential transformer and solid-state circuitry Most GFCI are protectors against macroshocks as they are usually not as sensitive as 10 µa or the medical equipment has a fault current greater than that

37 Protection: Power Distribution Example of a GFCI Figure Ground-fault circuit interrupters (a) Schematic diagram of a solid-state GFCI (three wire, two pole, 6 ma). (b) Ground-fault current versus trip time for a GFCI. [Part (a) is from C. F. Dalziel, "Electric Shock," in Advances in Biomedical Engineering, edited by J. H. U. Brown and J. F. Dickson III, 1973, 3: ]

38 Protection: Equipment Design Introduction Most failures of equipment ground occur at the ground contact or in the plug and cable Molded plugs should be avoided because of invisible breaks Strain-relief devices are recommended No use of three-prong-to-two-prong adapters (cheater adapters) Reduction of leakage current Special use of low-leakage power cords Capacitance-minimized design (special layout-design and usage of insulation) Maximized impedance from patient leads to hot conductors and from patient leads to chassis ground Double-Insulated equipment Interconnection of all conducting surfaces Separate layer of insulation to prevent contact with conductive surfaces (e.g. non conductive chassis, switch levers, knobs, etc.) Operation at low voltages Electrical isolation

39 Protection: Equipment Design Introduction Operation at low voltages (< 10 V) Reduction of risk of macroshock with reduced operation voltage Risk of microshock still exists Electrical isolation with isolation amplifiers Isolation amplifiers break the ohmic continuity of electric signals Isolation amplifiers use different voltage sources and different grounds Isolation voltage ν iso is rated from 1 kv to 10 kv without breakdown and described by the isolation-mode rejection ration (IMMR) (lightning example form the beginning) Next slides describe isolation amplifiers

40 Protection: Equipment Design Electric Isolation Three fundamental design methods 1. Transformer isolation Frequency-modulated or pulse-width-modulated carrier signal with small signal bandwidths Possibility to transmit energy and/or information 2. Optical isolation Uses LED on source-side and photodiode on output-side Very fast signal transmission possible, but no energy 3. Capacitive isolation

41 Protection: Equipment Design Electric Isolation à Transformer Figure Electrical isolation of patient leads to biopotential amplifiers (a) General model for an isolation amplifier, (b) Transformer isolation amplifier (Courtesy of Analog Devices, Inc., AD202). (c) Simplified equivalent circuit for an optical isolator (Copyright 1989 Burr-Brown Corporation. Reprinted in whole or in part, with the permission of Burr- Brown Corporation. Burr Brown ISO100). (d) Capacitively coupled isolation amplifier (Horowitz and Hill, Art of Electronics, Cambridge Univ. Press, Burr Brown ISO106).

42 Protection: Equipment Design Electric Isolation à optical and capacitive methods

43 Protection: Equipment Design Good Practice for isolated Heart Connections The best way to minimize hazards of microshocks is to isolate or eliminate electric connections to the heart Fully insulated connectors for external cardiac pacemakers powered by batteries Blood-pressure sensors with triple insulation between the column of liquid, the sensor case, and the electric connections Catheters with conductive walls all the way inside the patient to distribute the shock throughout the body (enlarged surface)

44 Testing the Electric System Introduction Test equipment: electrical-safety analyzers Testing the electric system Receptacles Grounding system Isolated power system Testing the electric appliance Ground-pin-to-chassis resistance Chassis leakage current Leakage current in patient leads

45 Testing the Electric System Electrical-Safety Analyzers Wide product range of el.-safety analyzers is available e.g. Medical-facility power systems Medical appliances Medical devices Special use-cases Products range from simple conversion box to computerized automatic measurement systems

46 Testing the Electric System Testing Receptacles Figure Three-LED receptacle tester Ordinary silicon diodes prevent damaging reverse-led currents, and resistors limit current. The LEDs are ON for line voltages from about 20 V rms to greater than 240 V rms, so these devices should not be used to measure line voltage.

47 Testing the Electric System Ground-pin-to-Chassis Resistance Resistance between the ground pin of the plug and the equipment chassis and exposed metal objects should not exceed 0.15 Ohm during the life of the appliance Figure Ground-pin-to-chassis resistance test

48 Testing the Electric System Chassis Leakage Current Figure (a) Chassis leakagecurrent test, (b) Current-meter circuit to be used for measuring leakage current. It has an input impedance of 1 kω and a frequency characteristic that is flat to 1 khz, drops at the rate of 20 db/decade to 100 khz, and then remains flat to 1 MHz or higher. (Reprinted with permission from NFPA , "Health Care Facilities," Copyright 2005, National Fire Protection Association, Quincy, MA This reprinted material is not the complete and official position of the National Fire Protection Association, on the referenced subject, which is represented only by the standard in its entirety.)

49 Testing the Electric System Leakage Current in Patient Leads (1) Leakage current from patient leads to ground Figure Test for leakage current from patient leads to ground (Reprinted with permission from NFPA , "Health Care Facilities," Copyright 2005, National Fire Protection Association, Quincy, MA This reprinted material is not the complete and official position of the National Fire Protection Association, on the referenced subject, which is represented only by the standard in its entirety.)

50 Testing the Electric System Leakage Current in Patient Leads (2) Leakage current between patient leads Figure Test for leakage current between patient leads (Reprinted with permission from NFPA , "Health Care Facilities," Copyright 2005, National Fire Protection Association, Quincy, MA This reprinted material is not the complete and official position of the National Fire Protetion Association, on the referenced subject, which is represented only by the standard in its entirety.)

51 Testing the Electric System Leakage Current in Patient Leads (3) AC isolation current Figure Test for ac isolation current (Reprinted with permission from NFPA , "Health Care Facilities," Copyright 2005, National Fire Protection Association, Quincy, MA This reprinted material is not the complete and official position of the National Fire Protection Association, on the referenced subject, which is represented only by the standard in its entirety.)

52 Conclusion Adequate electrical safety in health-care facilities and systems can be achieved at moderate costs by combining: 1. Good power-distribution system 2. Well designed equipment 3. Periodic maintenance and testing of power systems and equipment 4. Modest training program for medical personnel

53 With special thanks to the author of the corresponding book: John G. Webster Further reading Medical Instrumentation Application and Design, 4th Edition, Chapter 14 John G. Webster, Univ. of Wisconsin, Madison ISBN:

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital

Electrical Safety. Principles and examples of safety. History of (electrical) safety. Rules and regulations. Safety in the Hospital Electrical Safety History of (electrical) safety Safety in the Hospital Role of Clinical Engineer Rules and regulations Food and Drug Administration (FDA), Underwriter s Laboratory (UL), Federal Communication

More information

The question pool for the current General Class license took effect on July 1, 2007.

The question pool for the current General Class license took effect on July 1, 2007. General Class License Manual Errata and Corrections 22 Dec 2010 The following text is intended to support or correct the 6 th edition of the General Class License Manual. Determine the version of the manual

More information

Frequently Asked Question on Isolated Power Supply(IPS)

Frequently Asked Question on Isolated Power Supply(IPS) Frequently Asked Question on Isolated Power Supply(IPS) Q: What is an IPS? A: IPS is called as Isolated Power Supply or System and is used in a Hospital to ensure safe power supply to OT Operation Theaters

More information

BME 701 Lecture 1. Measurement and Instrumentation

BME 701 Lecture 1. Measurement and Instrumentation BME 701 Lecture 1 Measurement and Instrumentation 1 Cochlear Implant 2 Advances in Vision (Retinal Stimulation) 3 Mini Gastric Imaging 4 5 Aspects of Measurement General Instrumentation Transducers (Electrodes)

More information

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition Dr. Qasem Qananwah BME 420 Department of Biomedical Systems and Informatics Engineering 1 Biopotential

More information

Electrocardiogram (ECG)

Electrocardiogram (ECG) Vectors and ECG s Vectors and ECG s 2 Electrocardiogram (ECG) Depolarization wave passes through the heart and the electrical currents pass into surrounding tissues. Small part of the extracellular current

More information

Introduction to Biomedical Engineering

Introduction to Biomedical Engineering Introduction to Biomedical Engineering Biomedical Instrumentation Kung-Bin Sung 5/8/007 Outline Chapter 8 and chapter 5 of st edition: Bioinstrumentation Bridge circuit Operational amplifiers, instrumentation

More information

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards Biomedical Instrumentation, Measurement and Design ELEC4623/ELEC9734 Electrical Safety and Performance Standards Contents Physiological Effects of Electrical Currents Safety Standards for Medical Instrumentation

More information

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013 Benha University June 2013 Benha Faculty of Engineering Electrical Department Hospital Instrumentations (E472) 4 Th year (control) Dr.Waleed Abdel Aziz Salem Time: 3 Hrs Answer the following questions.

More information

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان

Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم جامعة تكريت كلية طب االسنان Dr.Rajaa CH.11. أ.م.د. رجاء سهيل جنم طب االسنان Applications of Electricity and Magnetism In Medicine Electrical Shock: When an electrode is connected to each hand & 60 Hz currets of different levels are

More information

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output VERSION 4.0 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany

More information

I. Introduction to Animal Sensitivity and Response

I. Introduction to Animal Sensitivity and Response I. Introduction to Animal Sensitivity and Response The term stray voltage has been used to describe a special case of voltage developed on the grounded neutral system of a farm. If this voltage reaches

More information

Biopotential Electrodes

Biopotential Electrodes Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Biopotential Electrodes 1 2 Electrode electrolyte interface The current crosses

More information

ECE 528 Understanding Power Quality

ECE 528 Understanding Power Quality ECE 528 Understanding Power Quality http://www.ece.uidaho.edu/ee/power/ece528/ Paul Ortmann portmann@uidaho.edu 208-316-1520 (voice) 1 Today Wiring and grounding Why it s important References Terms and

More information

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence

Anaesthesia. Update in. Electricity and Magnetism. David Sanders, Bruce McCormick* *Correspondence Update in Anaesthesia Electricity and Magnetism David Sanders, Bruce McCormick* *Correspondence Email: Bruce.McCormick@rdeft.nhs.uk ELECTRICAL CHARGE Charge is a fundamental property of matter. Atoms,

More information

I. Introduction to Animal Sensitivity and Response

I. Introduction to Animal Sensitivity and Response Stray Voltage Field Guide Douglas J. Reinemann, Ph.D. Professor of Biological Systems Engineering University of Wisconsin Madison September 2007 Update I. Introduction to Animal Sensitivity and Response

More information

Electrical noise in the OR

Electrical noise in the OR Electrical noise in the OR Chris Thompson Senior Staff Specialist Royal Prince Alfred Hospital SYDNEY SOUTH WEST AREA HEALTH SERVICE NSW HEALTH Electrical noise in the OR Root causes Tiny little signals

More information

Application Note 175 Using the STMISOC Stimulus Isolator

Application Note 175 Using the STMISOC Stimulus Isolator APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 02.14.2018 Application Note 175 Using the The MP160/150 System stimulation features

More information

Electrical Severity Measurement Tool Revision 4

Electrical Severity Measurement Tool Revision 4 Electrical Severity Measurement Tool Revision 4 November 2017 Electrical Severity Measurement Tool 1.0 Purpose: This tool is intended to measure the severity of exposure to an electrical safety event based

More information

Grounding for Power Quality

Grounding for Power Quality Presents Grounding for Power Quality Grounding for Power Quality NEC 250.53 states that ground resistance should be less than 25 ohms. Is this true? Grounding for Power Quality No! NEC 250.53 states

More information

Webinar: An Effective Arc Flash Safety Program

Webinar: An Effective Arc Flash Safety Program Webinar: An Effective Arc Flash Safety Program Daleep Mohla September 10 th, 2015: 2pm ET Agenda Arc Flash Defined and Quantified NFPA 70E / CSA Z 462 - Recent Updates What is the ANSI Z10 Hierarchy of

More information

PRODUCT SHEET IMPORTANT SAFETY NOTES!

PRODUCT SHEET IMPORTANT SAFETY NOTES! STMISO STIMULUS ISOLATION ADAPTERS See also: Stimulator Setup notes in AcqKnowledge Software Guide BIOPAC offers three stimulus isolation adapters: STMISOC constant current or constant voltage (5X / 10X)

More information

APPLICATION NOTE. Overview

APPLICATION NOTE. Overview Application Note 111 APPLICATION NOTE Nerve Conduction Velocity 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 06.05.2018 This application note details

More information

Lab E5: Filters and Complex Impedance

Lab E5: Filters and Complex Impedance E5.1 Lab E5: Filters and Complex Impedance Note: It is strongly recommended that you complete lab E4: Capacitors and the RC Circuit before performing this experiment. Introduction Ohm s law, a well known

More information

User's Guide. AC Circuit Load Tester. Model CT70

User's Guide. AC Circuit Load Tester. Model CT70 User's Guide AC Circuit Load Tester Model CT70 Introduction Congratulations on your purchase of the CT70 AC Circuit Load Tester. This device can detect circuit and wiring problems such as: Poor ground

More information

Special-Purpose Operational Amplifier Circuits

Special-Purpose Operational Amplifier Circuits Special-Purpose Operational Amplifier Circuits Instrumentation Amplifier An instrumentation amplifier (IA) is a differential voltagegain device that amplifies the difference between the voltages existing

More information

Model 7000 Low Noise Differential Preamplifier

Model 7000 Low Noise Differential Preamplifier Model 7000 Low Noise Differential Preamplifier Operating Manual Service and Warranty Krohn-Hite Instruments are designed and manufactured in accordance with sound engineering practices and should give

More information

GROUNDED ELECTRICAL POWER DISTRIBUTION. Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER

GROUNDED ELECTRICAL POWER DISTRIBUTION. Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER GROUNDED ELECTRICAL POWER DISTRIBUTION Excerpt from Inverter Charger Series Manual BY: VIJAY SHARMA ENGINEER .0 Conductors for Electrical Power Distribution For single-phase transmission of AC power or

More information

4.2 SHORT QUESTIONS AND ANSWERS 1. What is meant by cell? The basic living unit of the body is cell. The function of organs and other structure of the body is understood by cell organization. 2. Give the

More information

EE 230 Experiment 10 ECG Measurements Spring 2010

EE 230 Experiment 10 ECG Measurements Spring 2010 EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment

More information

Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E.

Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E. Overview of Grounding for Industrial and Commercial Power Systems Presented By Robert Schuerger, P.E. HP Critical Facility Services delivered by EYP MCF What is VOLTAGE? Difference of Electric Potential

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 60479.1:2010 IEC/TS 60479-1, Ed. 4.0 (2005) AS/NZS 60479.1:2010 Australian/New Zealand Standard Effects of current on human beings and livestock Part 1: General aspects AS/NZS 60479.1:2010 This

More information

PRACTICAL PROBLEMS WITH SUBSTATION EARTHING

PRACTICAL PROBLEMS WITH SUBSTATION EARTHING 1 PRACTICAL PROBLEMS WITH SUBSTATION EARTHING Dr Hendri Geldenhuys Craig Clark Eskom Distribution Technology This paper considers the issues around substation sites where the soil resistivity is of particularly

More information

BIO-ELECTRIC MEASUREMENTS

BIO-ELECTRIC MEASUREMENTS BIO-ELECTRIC MEASUREMENTS OBJECTIVES: 1) Determine the amplitude of the electrical "noise" in the body. 2) Observe and measure the characteristics and amplitudes of muscle potentials due to the biceps.

More information

Biomedical Engineering Evoked Responses

Biomedical Engineering Evoked Responses Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer andreas.neubauer@medma.uni-heidelberg.de Tel.: 0621 383 5126 Stimulation of biological systems and data acquisition 1. How can biological

More information

In this equation, P is the power output, V is the voltage, and I is the current. This is an important equation and it will be used repeatedly.

In this equation, P is the power output, V is the voltage, and I is the current. This is an important equation and it will be used repeatedly. Solar Fundamentals 0 people liked this 0 discussions READING ASSIGNMENT Fundamental Electrical Concepts You need to explore some fundamental concepts from electrical theory to understand how photovoltaic

More information

LF442 Dual Low Power JFET Input Operational Amplifier

LF442 Dual Low Power JFET Input Operational Amplifier LF442 Dual Low Power JFET Input Operational Amplifier General Description The LF442 dual low power operational amplifiers provide many of the same AC characteristics as the industry standard LM1458 while

More information

1. The coulomb is a unit of. A. charge B. voltage C. energy D. capacitance E. current. 2. The following is not true about voltage:

1. The coulomb is a unit of. A. charge B. voltage C. energy D. capacitance E. current. 2. The following is not true about voltage: BioE 1310 - Review 1 - DC 1/16/2017 Instructions: On the Answer Sheet, enter your 2-digit ID number (with a leading 0 if needed) in the boxes of the ID section. Fill in the corresponding numbered circles.

More information

Fundamentals of Pacing Therapy

Fundamentals of Pacing Therapy 1 4 7 2 5 8 3 6 9 Fundamentals of Pacing Therapy This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations.

More information

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions

BENG 186B Principles of Bioinstrumentation. Week 7 Review. Solutions BENG 186B Principles of Bioinstrumentation Week 7 Review Solutions Selections from: 2015 Homework 5 2015 Homework 6 C d = 0.001 1 2 1.5 Normalized Voltage 1 0.5 0-0.5-1 -1.5-2 Time A B C b C b BENG 186B

More information

Technical data ESA614 Electrical Safety Analyzer

Technical data ESA614 Electrical Safety Analyzer Technical data ESA614 Electrical Safety Analyzer The ESA614 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology professionals

More information

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry Main problem: heart muscle fibers are continuously stimulated by adjacent muscles

More information

Suppliers' Information Note. BT Metallic Path Facility. Interface Description

Suppliers' Information Note. BT Metallic Path Facility. Interface Description SIN 349 Issue 2.5 August 2015 Suppliers' Information Note For The BT Network BT Metallic Path Facility Interface Description Each SIN is the copyright of British Telecommunications plc. Reproduction of

More information

TL082 Wide Bandwidth Dual JFET Input Operational Amplifier

TL082 Wide Bandwidth Dual JFET Input Operational Amplifier TL082 Wide Bandwidth Dual JFET Input Operational Amplifier General Description These devices are low cost, high speed, dual JFET input operational amplifiers with an internally trimmed input offset voltage

More information

Direct Current Circuits

Direct Current Circuits HAPTER20 C. Return to Table of Contents Direct Current Circuits Street Light by Giacomo Balla, 1909. (Museum of Modern Art, New York.) Fig. 20 1 Replica of the first incandescent lamp, which used sewing

More information

What tests should be done?

What tests should be done? Valleylab Force FX periodic testing Electrosurgery The passage of high frequency electrical current through tissue to create a desired clinical effect. 2/7/2009 1 What tests should be done? 1. Output tests

More information

High-Voltage, Internally Powered ISOLATION AMPLIFIER

High-Voltage, Internally Powered ISOLATION AMPLIFIER ISO17 High-Voltage, Internally Powered ISOLATION AMPLIFIER FEATURES SIGNAL AND POWER IN ONE TRIPLE-WIDE PACKAGE 8Vpk TEST VOLTAGE 5Vrms CONTINUOUS AC BARRIER RATING WIDE INPUT SIGNAL RANGE: 1V to 1V WIDE

More information

EXPERIMENT 5 Bioelectric Measurements

EXPERIMENT 5 Bioelectric Measurements Objectives EXPERIMENT 5 Bioelectric Measurements 1) Generate periodic signals with a Signal Generator and display on an Oscilloscope. 2) Investigate a Differential Amplifier to see small signals in a noisy

More information

NL800A - Stimulus Isolator

NL800A - Stimulus Isolator NL800A - Stimulus Isolator Introduction The NL800A STIMULUS ISOLATOR is a battery powered, opto-coupled isolator. It has a constant current output, with very high output impedance, making it suitable for

More information

A Superior Current Source with Improved Bandwidth and Output Impedance for Bioimpedance Spectroscopy

A Superior Current Source with Improved Bandwidth and Output Impedance for Bioimpedance Spectroscopy International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 12, Issue 12 (December 2016), PP.24-29 A Superior Current Source with Improved Bandwidth

More information

A Comprehensive Model for Power Line Interference in Biopotential Measurements

A Comprehensive Model for Power Line Interference in Biopotential Measurements IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 49, NO. 3, JUNE 2000 535 A Comprehensive Model for Power Line Interference in Biopotential Measurements Mireya Fernandez Chimeno, Member, IEEE,

More information

Concepts to be Reviewed

Concepts to be Reviewed Introductory Medical Device Prototyping Analog Circuits Part 3 Operational Amplifiers, http://saliterman.umn.edu/ Department of Biomedical Engineering, University of Minnesota Concepts to be Reviewed Operational

More information

Industrial and Commercial Power Systems Topic 7 EARTHING

Industrial and Commercial Power Systems Topic 7 EARTHING The University of New South Wales School of Electrical Engineering and Telecommunications Industrial and Commercial Power Systems Topic 7 EARTHING 1 INTRODUCTION Advantages of earthing (grounding): Limitation

More information

Device Interconnection

Device Interconnection Device Interconnection An important, if less than glamorous, aspect of audio signal handling is the connection of one device to another. Of course, a primary concern is the matching of signal levels and

More information

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing What is a signal? A signal is a varying quantity whose value can be measured and which conveys information. A signal can be simply defined as a function that conveys information. Signals are represented

More information

TL082 Wide Bandwidth Dual JFET Input Operational Amplifier

TL082 Wide Bandwidth Dual JFET Input Operational Amplifier TL082 Wide Bandwidth Dual JFET Input Operational Amplifier General Description These devices are low cost, high speed, dual JFET input operational amplifiers with an internally trimmed input offset voltage

More information

Grounding Complications

Grounding Complications Grounding Complications Sensitive Equipment Isolated grounding Supplemental grounds Sensitive Electronic Equipment NEC 647 [2002-2005] Originally intended for audio studios -- now Industrial/commercial

More information

Why? Why? "Fatal Four Bureau of Labor Statistics. Monitoring and Metering for Electrical Safety. Shock Fire Explosion Production

Why? Why? Fatal Four Bureau of Labor Statistics. Monitoring and Metering for Electrical Safety. Shock Fire Explosion Production Monitoring and Metering for Electrical Safety Shock Fire Explosion Production Why? Why? "Fatal Four Bureau of Labor Statistics 1. Falls 2. Electrocutions 3. Struck by Object 4. Caught in/between 1 Bureau

More information

STMISOLA LINEAR ISOLATED STIMULATOR

STMISOLA LINEAR ISOLATED STIMULATOR STMISOLA LINEAR ISOLATED STIMULATOR The Constant Current and Constant Voltage Isolated Linear Stimulator (STMISOLA) will connect to any analog output signal drive (±10 V input) and provides considerable

More information

Application Notes. Electrosurgical Analyzer Primer

Application Notes. Electrosurgical Analyzer Primer Electrosurgical Analyzer Primer Purpose Scope Overview The purpose is to outline the characteristics of electrosurgical analyzers, describe the challenges of measuring RF power and review the inspection

More information

Laboratory Activities Handbook

Laboratory Activities Handbook Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...

More information

Lecture 4 Biopotential Amplifiers

Lecture 4 Biopotential Amplifiers Bioinstrument Sahand University of Technology Lecture 4 Biopotential Amplifiers Dr. Shamekhi Summer 2016 OpAmp and Rules 1- A = (gain is infinity) 2- Vo = 0, when v1 = v2 (no offset voltage) 3- Rd = (input

More information

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title Basic system for Electrocardiography Customer/Clinical need A recent health care analysis have demonstrated

More information

FAN4146 Ground Fault Interrupter

FAN4146 Ground Fault Interrupter Features For Two-Wire ALCI and RCD Applications Precision Sense Amplifier and Bandgap Reference Built-in AC Rectifier Direct DC Coupled to Sense Coil Built-in Noise Filter Low-Voltage SCR Disable SCR Gate

More information

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V? Introduction In this lab we will examine the concepts of electric current and potential in a circuit. We first look at devices (like batteries) that are used to generate electrical energy that we can use

More information

Lab: Using filters to build an electrocardiograph (ECG or EKG)

Lab: Using filters to build an electrocardiograph (ECG or EKG) Page 1 /6 Lab: Using filters to build an electrocardiograph (ECG or EKG) Goal: Use filters and amplifiers to build a circuit that will sense and measure a heartbeat. You and your heartbeat Did you know

More information

*Notebook is excluded

*Notebook is excluded Biomedical Measurement Training System This equipment is designed for students to learn how to design specific measuring circuits and detect the basic physiological signals with practical operation. Moreover,

More information

Capacitive Voltage Substations Ferroresonance Prevention Using Power Electronic Devices

Capacitive Voltage Substations Ferroresonance Prevention Using Power Electronic Devices Capacitive Voltage Substations Ferroresonance Prevention Using Power Electronic Devices M. Sanaye-Pasand, R. Aghazadeh Applied Electromagnetics Research Excellence Center, Electrical & Computer Engineering

More information

EXPERIMENT 1 Safety, Instrumentation, and Measurement

EXPERIMENT 1 Safety, Instrumentation, and Measurement ELEC 2010 Laboratory Manual Experiment 1 - Prelab Page 1 of 8 EXPERIMENT 1 Safety, Instrumentation, and Measurement Introduction The objectives of this experiment are to: Learn and apply principles of

More information

Filter Considerations for the IBC

Filter Considerations for the IBC APPLICATION NOTE AN:202 Filter Considerations for the IBC Mike DeGaetano Application Engineering Contents Page Introduction 1 IBC Attributes 1 Input Filtering Considerations 2 Damping and Converter Bandwidth

More information

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement, KL-720 Biomedical Measurement System Supplied by: 011 683 4365 This equipment is intended for students to learn how to design specific measuring circuits and detect the basic physiological signals with

More information

Electric Hazards and the Human Body *

Electric Hazards and the Human Body * OpenStax-CNX module: m42350 1 Electric Hazards and the Human Body * OpenStax This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 Abstract Dene thermal

More information

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences EECS 145L: Electronic Transducer Laboratory FINAL EXAMINATION Fall 2013 You have three hours to

More information

LF353 Wide Bandwidth Dual JFET Input Operational Amplifier

LF353 Wide Bandwidth Dual JFET Input Operational Amplifier LF353 Wide Bandwidth Dual JFET Input Operational Amplifier General Description These devices are low cost, high speed, dual JFET input operational amplifiers with an internally trimmed input offset voltage

More information

Using Circuits, Signals and Instruments

Using Circuits, Signals and Instruments Using Circuits, Signals and Instruments To be ignorant of one s ignorance is the malady of the ignorant. A. B. Alcott (1799-1888) Some knowledge of electrical and electronic technology is essential for

More information

Random & Sinusoidal Jitter Injector. Main Unit Operation Manual

Random & Sinusoidal Jitter Injector. Main Unit Operation Manual Random & Sinusoidal Jitter Injector RJI12G Main Unit Operation Manual Rev 1.0 September 2012 Introduction... 2 Safety Instruction... 2 1. General... 4 1 1 Features... 4 1 2 Functions & Characteristics...

More information

LM2925 Low Dropout Regulator with Delayed Reset

LM2925 Low Dropout Regulator with Delayed Reset LM2925 Low Dropout Regulator with Delayed Reset General Description The LM2925 features a low dropout, high current regulator. Also included on-chip is a reset function with an externally set delay time.

More information

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer

INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer 110 Toledo Street Farmingdale, NY 11735 USA http://www.netech.org 510-USER-Manual Rev3 10/29/2007 Dear User, We appreciate your purchase of the LKG

More information

UNIT 3: FIELD EFFECT TRANSISTORS

UNIT 3: FIELD EFFECT TRANSISTORS FIELD EFFECT TRANSISTOR: UNIT 3: FIELD EFFECT TRANSISTORS The field effect transistor is a semiconductor device, which depends for its operation on the control of current by an electric field. There are

More information

SENSOR AND MEASUREMENT EXPERIMENTS

SENSOR AND MEASUREMENT EXPERIMENTS SENSOR AND MEASUREMENT EXPERIMENTS Page: 1 Contents 1. Capacitive sensors 2. Temperature measurements 3. Signal processing and data analysis using LabVIEW 4. Load measurements 5. Noise and noise reduction

More information

LCM100 USER GUIDE. Line Carrier Modem INDUSTRIAL DATA COMMUNICATIONS

LCM100 USER GUIDE. Line Carrier Modem INDUSTRIAL DATA COMMUNICATIONS USER GUIDE INDUSTRIAL DATA COMMUNICATIONS LCM100 Line Carrier Modem It is essential that all instructions contained in the User Guide are followed precisely to ensure proper operation of equipment. Product

More information

400 MHz Passive High-Voltage Probe R&S RT-ZH

400 MHz Passive High-Voltage Probe R&S RT-ZH Manual 400 MHz Passive High-Voltage Probe R&S RT-ZH11 1409.7737.02 Printed in Germany Test and Measurment Manufacturer ROHDE & SCHWARZ For comprehensive information about Rohde and Schwarz, please visit

More information

Super Low Noise Preamplifier

Super Low Noise Preamplifier PR-E 3 Super Low Noise Preamplifier - Datasheet - Features: Outstanding Low Noise (< 1nV/ Hz, 15fA/ Hz, 245 e - rms) Small Size Dual and Single Channel Use Room temperature and cooled operation down to

More information

Current transducer FHS 40-P/SP600

Current transducer FHS 40-P/SP600 Current transducer I PM = 0-100 A Minisens transducer The Minisens transducer is an ultra flat SMD open loop integrated circuit current transducer based on the Hall effect principle. It is suitable for

More information

EDL Group #3 Final Report - Surface Electromyograph System

EDL Group #3 Final Report - Surface Electromyograph System EDL Group #3 Final Report - Surface Electromyograph System Group Members: Aakash Patil (07D07021), Jay Parikh (07D07019) INTRODUCTION The EMG signal measures electrical currents generated in muscles during

More information

Lab #9: Compound Action Potentials in the Toad Sciatic Nerve

Lab #9: Compound Action Potentials in the Toad Sciatic Nerve Lab #9: Compound Action Potentials in the Toad Sciatic Nerve In this experiment, you will measure compound action potentials (CAPs) from an isolated toad sciatic nerve to illustrate the basic physiological

More information

M302RM OPERATING MANUAL

M302RM OPERATING MANUAL M302RM OPERATING MANUAL The Model 302RM is a Linear, high voltage, differential amplifier designed to drive a capacitive load such as Conoptics 350, 360, 370 series E.O. modulators. The amplifier is DC

More information

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual

WHALETEQ. ESU Neutral Electrodes Impedance Tester. Model: HFPA150. User Manual WHALETEQ ESU Neutral Electrodes Impedance Tester Model: HFPA150 User Manual Version2014-10-30 Hardware Version1.3.x 1. Introduction HFPA150 is a unique tester specifically designed for the testing of neutral

More information

IFB270 Advanced Electronic Circuits

IFB270 Advanced Electronic Circuits IFB270 Advanced Electronic Circuits Chapter 14: Special-purpose op-amp circuits Prof. Manar Mohaisen Department of EEC Engineering eview of the Precedent Lecture Introduce the level detection op-amp circuits

More information

PR-E 3 -SMA. Super Low Noise Preamplifier. - Datasheet -

PR-E 3 -SMA. Super Low Noise Preamplifier. - Datasheet - PR-E 3 -SMA Super Low Noise Preamplifier - Datasheet - Features: Low Voltage Noise (0.6nV/ Hz, @ 1MHz single channel mode) Low Current Noise (12fA/ Hz @ 10kHz) f = 0.5kHz to 4MHz, A = 250V/V (customizable)

More information

Application Note # 5438

Application Note # 5438 Application Note # 5438 Electrical Noise in Motion Control Circuits 1. Origins of Electrical Noise Electrical noise appears in an electrical circuit through one of four routes: a. Impedance (Ground Loop)

More information

AS SUPER 4 digital. Elektrischer Nadelstimulator / Electrical needle stimulator. Art.-Nr

AS SUPER 4 digital. Elektrischer Nadelstimulator / Electrical needle stimulator. Art.-Nr AS SUPER 4 digital Elektrischer Nadelstimulator Electrical needle stimulator Art.-Nr. 200510 Gebrauchsanweisung Instruction Manual - Art.-Nr. 101477 20 Contents Foreword...21 Purpose for use...21 Safety

More information

Healthcare Isolation Power Systems. Presented by: David Knecht

Healthcare Isolation Power Systems. Presented by: David Knecht Healthcare Isolation Power Systems Presented by: David Knecht E-Mail: David.Knecht@Bender-us.com Healthcare Isolation Power Systems Theory Offer Codes & Standards FAQs Isolated Power System Purpose To

More information

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1 Dr. Gari Clifford Hilary Term 2013 1. (Exemplar Finals Question) a) List the five vital signs which are most commonly recorded from patient monitors in high-risk

More information

BENG 186B Winter 2013 Final

BENG 186B Winter 2013 Final Name (Last, First): BENG 186B Winter 2013 Final This exam is closed book, closed note, calculators are OK. Circle and put your final answers in the space provided; show your work only on the pages provided.

More information

Potential Risks of MRI in Device Patients

Potential Risks of MRI in Device Patients Outline Potential Risks of MRI in Device Patients Redha Boubertakh r.boubertakh@qmul.ac.uk MRI and cardiac implantable electronic devices (CIED) Components of an MRI scanner MRI implant and device safety

More information

C3D04060A Silicon Carbide Schottky Diode Z-Rec Rectifier

C3D04060A Silicon Carbide Schottky Diode Z-Rec Rectifier C3D46A Silicon Carbide Schottky Diode Z-Rec Rectifier RM = 6 V I F(AVG) = 4 A Q c = 8.5 nc Features 6-Volt Schottky Rectifier Optimized for PFC Boost Diode Application Zero Reverse Recovery Current Zero

More information

Los A LA-UR Los Alamos National Laboratory Los Alamos, New Mexico 87545

Los A LA-UR Los Alamos National Laboratory Los Alamos, New Mexico 87545 LA-UR-98-1 Los Alamos NationalLaboratory is operated by the University of California for the United States Department of Energy under contract W-7405-ENG-36 TITLE: SUBMITTED TO: Electrical Potential Transfer

More information

Earthing of Electrical Devices and Safety

Earthing of Electrical Devices and Safety Earthing of Electrical Devices and Safety JOŽE PIHLER Faculty of Electrical Engineering and Computer Sciences University of Maribor Smetanova 17, 2000 Maribor SLOVENIA joze.pihler@um.si Abstract: - This

More information

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology Biomedical Sensor Systems Laboratory Institute for Neural Engineering Graz University of Technology 2017 Bioinstrumentation Measurement of physiological variables Invasive or non-invasive Minimize disturbance

More information