SUNTEST CPS+ / XLS+ Software documentation Version 1.4 SUNTEST ATLAS SUNTEST --> ENTER. Start. Stop ESCAPE

Size: px
Start display at page:

Download "SUNTEST CPS+ / XLS+ Software documentation Version 1.4 SUNTEST ATLAS SUNTEST --> ENTER. Start. Stop ESCAPE"

Transcription

1 SUNTEST CPS+ / XLS+ Software documentation Version 1.4 SUNTEST ATLAS SUNTEST --> Start Stop APE 0 APE

2 ATLAS SUNTEST --> PROGRAM PARAMETER PROGRAM PARAMETER Program Start Programming SYS Data XENON lamp Check Filter ok? Program Phases Selection Calibration NEW yes NEW No APE E Select Filters ABCDEFGH Calibration E Calibration BST Select Filters ABCDEFGH FW Check Print out? BST APE Phase Time Print: min Change /APE Switch off Start Program (1-6) Dose Time APE START SUNTEST Program is running Switch off criterion reached DOSE TIME Program diagram SUNTEST CPS+ / XLS+

3 Dear SUNTEST User! This software documentation is applicable for the following instruments: SUNTEST CPS+ SUNTEST XLS+ The SUNTEST CPS+ / XLS+ microprocessor control unit and the sensors for irradiance, black standard and test chamber temperature let you program your own test procedure in prestructured steps for different cycles. light and dark phasing, controlled irradiance levels, controlled black standard temperature, display of test chamber temperature, immersion control, cyclical or continuous and controlled water temperature, ON / OFF switching of a cooling unit. Page 3

4 Contents Page 1. Menu 5 2. Programming 5 3. Starting the program Calibration Calibrating the irradiance Calibrating the black standard temperature Changing the Xenon-lamp Trouble shooting Additional functions 13 Page 4

5 APE APE Software documentation SUNTEST CPS+ / XLS+ 1. Menu Display text Input/Key pad ATLAS SUNTEST --> ATLAS SUNTEST CPS+ Start Stop PROGRAM PARAMETER 0 SYSTEM DATA XENONLAMP PROGRAM START PROGRAMMING SELECTION CALIBRATION CALIBR. E (W/m²) CALIBR. BST ( C) 2. Programming PROGRAM PROGRAMMING Input of the program number: PROGRAM (1-6) n Comment: A program can be selected and changed when entering the program number. Input of number of phases: PHASES (1-6): _ n PROGRAM: _ PHASES: _ ok Comment: The program input can be interrupted by pressing STOP. Page 5

6 DISPLAY TEXT INPUT/KEY PAD Input of irradiance (E): E (0; W/m²) PH 1= PH 2= E (0; W/m²) PH 3= PH 4= E (0; W/m²) PH 5= PH 6= n n n Input of immersion function (0 = NO; 1 = YES) (FW): FW (0=OFF/1=ON) PH 1= PH 2= FW (0=OFF/1=ON) PH 3= PH 4= FW (0=OFF/1=ON) PH 5= PH 6= n n n Comment: When the cooling aggregate is connected and selected (refer to additional functions in Chapter 12, item 4), the following message is displayed instead of "Input of immersion function". By pressing "" the programming can be continued. COOLING UNIT = ON Input of the black standard temperature (BST): BST (0; 35 C C) PH 1= PH 2= BST (0; 35 C C) PH 3= PH 4= BST (0; 35 C C) PH 5= PH 6= n n n Comment: When choosing "0" the black standard temperature is not regulated, and the fan for the temperature control runs at constant speed. The lowest black standard temperature of 35 C is reached with min. irradiance and uncontrolled black standard temperature. When operating with a cooling aggregate, the black standard temperature can be dropped below 35 C even at max. irradiance. Page 6

7 DISPLAY TEXT INPUT/KEY PAD Input of the phase time: PHASE TIME [min] PH 1= PH 2= PHASE TIME [min] PH 3= PH 4= PHASE TIME [min] PH 5= PH 6= n n n Comment: Cycles between 1 and 999 minutes can be entered. Input off switch-off criterion: SWITCH OFF TIME DOSE SWITCH OFF TIME [h] n DOSE [kj/m²] n Comment: The maximum values are hours for the irradiation time, and kj/m² for the radiant exposure.. 3. Starting the program PROGRAM PROGRAM START CHECK FILTER! OK? /APE Comment: Before starting the program, the filter data can be checked and changed. By pressing the menu appears: FILTER:_ CHANGE? /APE APE Comment: By pressing you may access the menu for the inserted filters (see chapter 8). CHECK PRINTOUT! OK? /APE Comment: Before starting the program, the printout of the time parameters can be entered via the interface in minutes. By pressing the menu appears: Page 7

8 DISPLAY TEXT INPUT/KEY PAD PRINT: min CHANGE? /APE n Comment: The printing time can be modified via. 0 means no printout. By pressing APE the input can be skipped. CYCLE TIME PRINTER: Minutes ( ) n Input of program number: START PROGRAM (1-6): _ RUN PROGRAM: _ press START or APE n n START IGNITION! PLEASE WAIT Comment: The program will be started automatically, data can be retrieved from the indicated parameters. 4. Comment: The parameters of a running program can be scrolled through using the arrow keys. SUNTEST PROGRAM IS RUNNING PROGRAM NUMBER: PHASE NUMBER: PHASE TIME SET: PHASE TIME (min): TIME (h) SET: TIME (h) ACT: E (W/m²) SET VAL: E (W/m²) SET ACT: BST ( C) SET VAL: BST ( C) SET ACT: (switch-off criteria total time) Page 8

9 DISPLAY TEXT INPUT/KEY PAD CHT ( C): FW (ON/OFF): DOSE SET: DOSE (kj/m²): XENONLAMP (h): OPER: HOURS (h): (switch-off criterion: dose) Comment: Explanation of the abbreviations in the display text E= Irradiance in W/m², FW= Immersion function (ON/OFF), BST= Black standard temperature in C, CHT= Test chamber temperature in C, DOSE= Radiant exposure in kj/m². 5. SWITCH OFF CRITERIA REACHED Comment: Upon reaching the end of the program, the above message appears. Pressing the key displays the irradiance and the total time. Time values for total time and phase times are counted from lamp ignition. DOSE TIME : (kj/m²) h ok? 6. Comment: If during a running program a lamp has to be exchanged or the optical filter system has to cleaned (s. chapter 11, 'Trouble Shooting"), the program can be interrupted and continued afterwards. SUNTEST PROGRAM IS RUNNING STOP Page 9

10 DISPLAY TEXT INPUT/KEY PAD PROGRAM STOP CONTINUE APE START START/APE Comment: The fan to cool down the lamp is automatically switched off after 3 minutes in the SUNTEST XLS+ and after approximately 40 seconds in the SUNTEST CPS+. Then turn ON/OFF switch to OFF and pull mains plug. Based on the present operating manual the lamp can now be exchanged and the filter system can be cleaned. Plug in mains plug and reset ON/OFF switch to ON. PROGRAM INTERRUPTED CONTINUE --> START START/APE Comment: After pressing START the program will continue at the point of interruption, or press " to completely stop the program. 7. Comment: SUNTEST supports an RS232 to download measurement data to a computer or printer while a program is in progress. At the start of the program, the length in time of the interface record can be entered in a range of 0 to 999 minutes. With an entry of 0 there will be no printout. The data printout has the following format: OPER.HOUR : h XENON LAMP: h FILTER: _ SUNTEST SERIAL NUMBER: PROGRAM NUMBER: _ SWITCH OFF: (TIME/DOSE) PHASE NUMBER: _ PHASE TIME: (min) E: (W/m²) BST: ( C) FW: PN PT E H BST CHT T = PHASE NUMBER = PHASE TIME = IRRADIANCE = DOSE = BLACK STANDARD TEMPERATURE = CHAMBER TEMPERATURE = RUNNING TIME PN PT E H BST CHT T n min W/m² kj/m² C C min Page 10

11 DISPLAY TEXT INPUT/KEY PAD 8. Comment: Because the maximum irradiance value depends on the inserted filter, the relevant filter characteristics have to be entered. PARAMETER SYSTEM DATA SELECTION FILTER SELECTION A B C D E F G H Comment: The inserted filter configuration can be selected with the arrow keys and confirmed by pressing. The following letters correspond to the inserted filter configuration: A: coated quartz glass, B: coated quartz glass with UV special glass, C: coated quartz glass with special window glass, D: uncoated quartz glass, E: uncoated quartz glass with UV special glass, F: uncoated quartz glass with special window glass, G: coated quartz glass with 6 mm window glass "Solar ID 65" (filter acc. to ICH Guideline), H: coated quartz glass with 1 mm WG320 "Solar Standard" (filter acc. to COLIPA). 9. Calibration Comment: The calibration shall only be carried out when a coated quartz glass dish is installed. For this, a XenoCal resp. Radialux ( nm) (option) is placed in the test chamber and is connected to the reading unit. For more calibaration details please refer to the Operating Manual of the SUNTEST XLS/XLS+. PARAMETER SYSTEM DATA CALIBRATION 9.1 Calibrating the irradiance CALIBR. E (W/m²) CALIBRIBRATION: E= set E= sensor nnn Comment: With E= set, the irradiance value at which the calibration should be carried out is set and has to be confirmed with. Afterwards, the lamp ignites automatically and runs up to the set value. The irradiance value displayed by the XenoSoft software has to be entered at E= sensor. (For calibration with the Radialux, the irradiance value displayed on the reading unit has to be entered at E= set after multiplication with the filter factor of the filter systmes installed).the relevant calibration factor is calculated and indicated by the control unit. E (W/m²): BST ( C): ok? Page 11

12 DISPLAY TEXT INPUT/KEY PAD 9.2 Calibrating the black standard temperature Comment: To calibrate the black standard thermometer, a temperature sensor (option) has to be placed in the test chamber. CALIBR: BST( C) CALIBRIBRATION: E= set BST= set nnn (2x) Comment: The irradiance value has to be entered at E= set, and the value for the black standard temperature required for calibration has to be entered at BST= set. Afterwards the SUNTEST is started by the control unit and displays the setpoint and the actual black standard values. BST ( C) SET VAL: BST ( C) ACT VAL: STOP Comment: When the final temperature is reached the program is interrupted by pressing the STOP key, the temperature value is read out immediatele and after the automatic restart of the program (and after the automatic cooling period) the temperature value can be entered. BST SENSOR VALUE BST= SENSOR Comment: The calibration factor is automatically calculated by the control unit. E (W/m²): BST ( C): ok? WARNING! The sensor warms up during exposure in the SUNTEST. Before touching put on protective gloves. Burn hazard! 10. Changing the Xenon-lamp Comment: The lamp exchange is described in the enclosed operating manual. An interrupted program can be continued after the exchange of a lamp. PARAMETER XENONLAMP NEW XENONLAMP: YES NO Comment: When changing the lamp, the lamp s age can be set to zero by pressing. Page 12

13 11. Trouble shooting Comment: The following error messages can be displayed during operation of the SUNTEST. Trouble shooting and routine maintenance can be carried out by the operator according to the applicable safety regulations. Otherwise, please call the Xenotest service department. No. Displayed error messages Possible cause (C) and action (A) 1 DOOR OPEN or TEMPERATURE FUSE 2 E MAX. POWER REACHED CHANGE XENON LAMP? (A) Close test chamber door. (C) Temperature fuse is defective. (A) Replace temperature fuse by the Xenotest service department. (A) Replace XENON lamp (A) Clean the contaminated optical filter system. 3 FLOATING WATER FAILED (A) Switch on the immersion unit. (C) Water level in the water tank too low. (A) Refill water. 4 POWER FAIL OCCURED! CONTINUE START 5 ERROR 1: (h) XENON LAMP NO START! (C) Mains power failure. (A) Enter START (the program will be continued at the interrupted point). (C) Lamp is too hot. (A) Let lamp cool down and restart. (C) Lamp has not ignited due to too high lamp age (1500h). (A) Replace lamp. 6 COOLING UNIT failed (C) The immersion unit is connected. (C) The cooling aggregate has been selected. (A) Select the immersion unit under Additional Functions (Chapter 12), item 4 or connect the cooling aggregate. 12. Additional functions Comment: By pressing the number keys in the main menu (= 1st display after starting), you can get additional information on the SUNTEST and preset test parameters. No. Display text Function 2 SHOW SYSTEM CLOCK Unit age in hours, minutes and seconds 3 SUNTEST Version Software version 4 COOLING UNIT Floating water Selec the immersion unit or the cooling aggregate 5 INITIALISING NV-RAM Overwrite the program with the nominal values 6 AUTOMATIC START YES/NO Restarting after power failure can be switched on or off 7 SERIAL NUMBER Serial number of the SUNTEST CPS+ / XLS+ 8 E(W/m²): BST( C): ok 9 MAIN SERVICE MODE Start of the service program Shows the calibration for the irradiance and the black standard temperature Page 13

14 ATLAS Material Testing Technology GmbH Vogelsbergstr. 22 D Linsengericht / Altenhaßlau Phone: +49/6051/ Fax: +49/6051/ Ident No.:

Xenon Test instruments. Lightfastness Weatherfastness Photostability

Xenon Test instruments. Lightfastness Weatherfastness Photostability Xenon Test instruments Lightfastness Weatherfastness Photostability THE WORLD S MOST WIDELY USED FLATBED WEATHERING INSTRUMENTS Weathering is the adverse response of a material or product to climate, often

More information

SUNTEST XENON TEST INSTRUMENTS. Lightfastness Weatherfastness Photostability

SUNTEST XENON TEST INSTRUMENTS. Lightfastness Weatherfastness Photostability SUNTEST XENON TEST INSTRUMENTS Lightfastness Weatherfastness Photostability SUNTEST THE WORLD S MOST WIDELY USED FLATBED WEATHERING INSTRUMENTS Weathering is the adverse response of a material or product

More information

Conforming to the ICH Guideline for the Photostability Testing of New Drug Substances and Drug Products (ICH Q1B) Using the Atlas SUNTEST CPS+

Conforming to the ICH Guideline for the Photostability Testing of New Drug Substances and Drug Products (ICH Q1B) Using the Atlas SUNTEST CPS+ Conforming to the ICH Guideline for the Photostability Testing of New Drug Substances and Drug Products (ICH Q1B) Using the Atlas SUNTEST CPS+ This document summarizes the key requirements in the ICH Guideline

More information

SUNTEST XENON TEST INSTRUMENTS. Lightfastness Weatherfastness Photostability

SUNTEST XENON TEST INSTRUMENTS. Lightfastness Weatherfastness Photostability SUNTEST XENON TEST INSTRUMENTS Lightfastness Weatherfastness Photostability SUNTEST THE WORLD S MOST WIDELY USED FLATBED WEATHERING INSTRUMENTS Weathering is the adverse response of a material or product

More information

Consumer Goods Testing

Consumer Goods Testing Consumer Goods Testing LEAD E R S I N PH OTO S TA B I L I T Y T E S T I NG F OR T HE CONS UMER G OODS I ND US TRY Packaged Consumer Goods Environmental Stress Challenges of Product Photostability Three

More information

CONSUMER GOODS TESTING

CONSUMER GOODS TESTING CONSUMER GOODS TESTING Realistic Light Stability Testing of Ingredients, Products and Packaging Packaged Consumer Goods Challenges of Product Photostability The success of today s consumer goods like drinks,

More information

MANUAL. Textron Motors Diagnostic Tool. This manual is valid for the following Textron Motors Diagnostic Tool:

MANUAL. Textron Motors Diagnostic Tool. This manual is valid for the following Textron Motors Diagnostic Tool: MANUAL Textron Motors Diagnostic Tool This manual is valid for the following Textron Motors Diagnostic Tool: 0507 TD0507_HB Rev F 6..05 en_english Read the manual before performing the task on the engine.

More information

Network of Weathering

Network of Weathering Atlas Material Testing Technology LLC (p)+1.773.327.4520 (f) +1.773.327.5787 www.atlas mts.com Specifications and features of products, services and the standards they meet are subject to change without

More information

TC 405/30 THERMOCOMPUTER TC 405/30. temp rate. time. rate. temp time. fix prog. memo. pers. Operating Instructions. event kw/h.

TC 405/30 THERMOCOMPUTER TC 405/30. temp rate. time. rate. temp time. fix prog. memo. pers. Operating Instructions. event kw/h. TC 405/30 fix prog pers prog 0 memo THERMOCOMPUTER TC 405/30 7 8 9 4 5 6 1 rate temp rate o k temp time 2 event kw/h time 3 start stop Operating Instructions Brief Instructions To - start a fixed programme

More information

ADI-100 Interrupter. Operator s Manual. 526 S. Seminole Bartlesville, OK /

ADI-100 Interrupter. Operator s Manual. 526 S. Seminole Bartlesville, OK / ADI-100 Interrupter Operator s Manual 526 S. Seminole Bartlesville, OK 74003 918/336-1221 www.sescocp.com ADI - 100 Interrupter FEATURES Clock Accuracy 100% during GPS lock Clock Drift 30 µsec per degree

More information

Oriel Flood Exposure Sources

Oriel Flood Exposure Sources 218 Oriel Flood Exposure Sources High intensity outputs CALIBRATION SOURCES Highly uniform, large collimated beams Efficient out of band rejection Timed exposures DEUTERIUM SOURCES ARC SOURCES INCANDESCENT

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Conductive Conductivity ProMinent D1C2-Leit.-001-GB Conductivity 100 µs/cm DULCOMETER STOP START Conductivity 100

More information

NOTE: Refer to the User s Manual ( xxP) for more information on Lamp Operation Modes.

NOTE: Refer to the User s Manual ( xxP) for more information on Lamp Operation Modes. Introduction Applies to RPMS-D100U/UF (38-GFX204-01, 38-GFX205-01, 38-GFX206-01, 38-GFX207-01), CS70-D100U/UF (38-GFX052-xx, 38-GFX053-xx) and CS50-D100U (38-GFX054-xx). RPMS-D100U/UF, CS50-D100U and CS70-D100U/UF

More information

SL 200 instrument can process the energy intensity emitted by the solar radiation in a precise place of the earth.

SL 200 instrument can process the energy intensity emitted by the solar radiation in a precise place of the earth. Solarimeter SL 200 Table of contents 1 Introduction...4 2 General information...4 2.1 Measured units...4 2.2 Use...4 3 Operating principle...5 3.1 Keyboard presentation...5 3.2 Instrument offers 3 groups

More information

REFLOW OVEN, FORCED CONVECTION FT03 ( Item code V900320)

REFLOW OVEN, FORCED CONVECTION FT03 ( Item code V900320) www.cif.fr REFLOW OVEN, FORCED CONVECTION FT03 ( Item code V900320) INSTRUCTIONS MANUAL This document contains manufacturer s information which is protected by copyright. All rights reserved. This document

More information

Using DIAGNOSE.EXE with a ColorQuest 45/0

Using DIAGNOSE.EXE with a ColorQuest 45/0 Insight on Color November, 1999, Vol. 11, No. 11 (Rev. 5/02) Using DIAGNOSE.EXE with a ColorQuest 45/0 The DIAGNOSE.EXE program, which may be used to assess instrument performance and diagnose potential

More information

Pro Advantage Urine Analyzer. Operator s Manual. Pro Advantage by NDC. All rights reserved PS-733 Rev 00-3/10 1/23

Pro Advantage Urine Analyzer. Operator s Manual. Pro Advantage by NDC. All rights reserved PS-733 Rev 00-3/10 1/23 Pro Advantage Urine Analyzer Operator s Manual PS-733 Rev 00-3/10 1/23 The information in this manual was correct at the time of printing. However, the manufacturer continues to update and change this

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Ozone ProMinent D1C2-03-001-GB O 3 DULCOMETER STOP START O 3 DULCOMETER STOP START Type D Type W D1C A Please enter

More information

Contents I. APPLICATIONS II. SPECIFICATIONS III. PRINCIPLES IV OPERATION V. MAINTENANCE

Contents I. APPLICATIONS II. SPECIFICATIONS III. PRINCIPLES IV OPERATION V. MAINTENANCE Contents I. APPLICATIONS...- 1 - II. SPECIFICATIONS...- 2 - III. PRINCIPLES...- 3 - IV OPERATION...- 9 - V. MAINTENANCE...- 21 - VI SUPPLEMENT...- 23 - Warning: The responsible person must be clear that

More information

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT!

PD-3000UV. INSTRUCTION MANUAL UV-VIS Spectrophotometer OMA IMPORTANT! INSTRUCTION MANUAL UV-VIS Spectrophotometer PD-3000UV IMPORTANT! READ THIS OPERATION MANUAL! *Prior to operation, please read this manual carefully and retain it for future reference. Failure to read,

More information

Берг АБ Тел. (495) , факс (495) Initial operation and diagnosis

Берг АБ Тел. (495) , факс (495) Initial operation and diagnosis Axial/radial bearings with integral measuring system Initial operation and diagnosis Fitting and maintenance instructions MON 18 This technical publication has been produced with a great deal of care and

More information

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop 1. Log in usage using the SMIF web site 2. Turn power

More information

RQflex 2. Manual English

RQflex 2. Manual English RQflex 2 Manual English Contents 1 Introduction 3 2 Description of the instrument 5 display, keys 6 technical data 8 3 Setting up the instrument 9 insertion of batteries 10 setting the time and date 11

More information

XRC X-Ray Calibration System

XRC X-Ray Calibration System XRC X-Ray Calibration System Technical Description Contents 1. Equipment overview 2. X-ray Beam Specifications 3. The Control Console 4. Radiation Safety 5. Filters 6. X-ray Generator & Heat Exchange Equipment

More information

EdvoCycler & MegaCycler

EdvoCycler & MegaCycler Operation Manual: EdvoCycler & MegaCycler Cat. #541 and #542 Features: 25 x 0.2 ml Tube Block (EdvoCycler) 49 x 0.2 ml Tube Block (MegaCycler) Heated Oil-Free Lid with Magnetic Latch Vivid 7 Line LCD Display

More information

Operating Instructions

Operating Instructions TC 405 fix prog pers prog 0 memo t 0 C/h 7 8 9 4 5 6 1 2 event kw/h k THERMOCOMPUTER TC 405 tmp 1 t1 t 2 tmp 2 3 start stop k C/h C/h k 1050 C 1180 C 1250 C 150 C 800 C 900 C Operating Instructions Brief

More information

Product Specification SenseAir S8

Product Specification SenseAir S8 Product Specification SenseAir S8 Miniature CO 2 Sensor module Driesen + Kern GmbH SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-005

More information

KNX TH65-AP. Thermo-Hygrometer. Installation and Adjustment. Item number 70184

KNX TH65-AP. Thermo-Hygrometer. Installation and Adjustment. Item number 70184 EN KNX TH65-AP Thermo-Hygrometer Item number 70184 Installation and Adjustment 1 Content 1. Description... 3 1.1. Technical specifications... 3 2. Installation and commissioning... 4 2.1. Installation

More information

1000TR. Instructions

1000TR. Instructions 1000TR ph Instructions CONTENTS 1. INTRODUCTION... 2 1.1 COMMON INTRODUCTION... 2 1.2 PARTS & ACCESSORIES... 2 2. INSTALLATION... 3 2.1 CASING... 3 2.2 MOUNTING... 3 2.3 ELECTRICAL INSTALLATION... 3 2.3.1

More information

TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller. Mounting and Operating Instructions EB 5578 EN

TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller. Mounting and Operating Instructions EB 5578 EN TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller Mounting and Operating Instructions EB 5578 EN Firmware version 2.30 Edition November 2017 Definition of signal words DANGER!

More information

REFLOW OVEN, FORCED CONVECTION FT02 ( Item code F31114)

REFLOW OVEN, FORCED CONVECTION FT02 ( Item code F31114) www.cif.fr REFLOW OVEN, FORCED CONVECTION FT02 ( Item code F31114) INSTRUCTIONS MANUAL This document contains manufacturer s information which is protected by copyright. All rights reserved. This document

More information

Instruction Manual HAS3xE-IM-SW39 10/2005

Instruction Manual HAS3xE-IM-SW39 10/2005 Instruction Manual HAS3xE-IM-SW39 10/2005 Software Version 3.9.x NGA 2000 Software Manual for MLT or CAT 200 Analyzer and MLT or CAT 200 Analyzer Module (combined with NGA 2000 Platform, MLT, CAT 200 or

More information

USER MANUAL. EPP Intelligent Positioner Control Unit 1/22.

USER MANUAL. EPP Intelligent Positioner Control Unit 1/22. USER MANUAL - Intelligent Positioner Control Unit 1/22 Table of contents: 1 General... 3 1.1 Safety instructions... 3 2 Application... 4 3 Electrical specifications and terminals... 5 3.1 Control loop...

More information

fi-6130 / fi-6230 Cleaning and Maintenance

fi-6130 / fi-6230 Cleaning and Maintenance fi-6130 / fi-6230 Cleaning and Maintenance fi-6130 / fi-6230 Cleaning and Maintenance fi-6130 / fi-6230 Image Scanner Operator's Guide Chapter 4 DAILY CARE This chapter describes how to clean the scanner.

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D2C Part 2: Adjustment and Operation, Measured Variables ph/chlorine dioxide ProMinent D2C2-001-pH/CIO2-GB ph/clo 2 7.20 ph 0.45 ppm DULCOMETER STOP STAR T ph/clo 2 7.20

More information

User manual UV Flash Dry 15/700. HENSEL-VISIT GmbH & Co. KG Robert-Bunsen-Str. 3 D Würzburg GERMANY

User manual UV Flash Dry 15/700. HENSEL-VISIT GmbH & Co. KG Robert-Bunsen-Str. 3 D Würzburg GERMANY User manual UV Flash Dry 15/700 HENSEL-VISIT GmbH & Co. KG Robert-Bunsen-Str. 3 D-97076 Würzburg GERMANY Tel. +49 (0) 931 27881-0 Fax: +49 (0) 931 27881-50 Email: info@hensel.de Internet: http://www.hensel.de

More information

Reference Guide Brief explanations for routine operations

Reference Guide Brief explanations for routine operations Reference Guide Brief explanations for routine operations DCP-T30 DCP-T50W DCP-T70W Brother recommends keeping this guide next to your Brother machine for quick reference. Online User's Guide For more

More information

The Eyetracker. Bruno Bonet

The Eyetracker. Bruno Bonet The Eyetracker Bruno Bonet EYETRACKER The Eyetracker is at the end of the MRI room behind the machine. Do NOT move the Eyetracker EYETRACKER A bulb can light the pupil through mirrors and the reflected

More information

INSTALLATION & OPERATION MANUAL

INSTALLATION & OPERATION MANUAL INSTALLATION & OPERATION MANUAL PREFACE This installation & operation manual is intended as an instruction manual for trained person who is in charge of installation, maintenance, repair, etc. Before installation

More information

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS

TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) S. No. PARTICULARS BHEL SPECIFICATIONS TECHNICAL SPECIFICATIONS for 300KV CONSTANT POTENTIAL X RAY EQUIPMENT (for industrial applications) BIDDER s OFFER S. No. PARTICULARS BHEL SPECIFICATIONS 1.0 Area of Application The 300kV Constant potential

More information

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer Page 1 of 11 Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer The Aramis Raman system is a software selectable multi-wavelength Raman system with mapping capabilities with a 400mm monochromator and

More information

COMPACT-LINE. KOBO-RP Transmitter for ORP Model ARM-Z

COMPACT-LINE. KOBO-RP Transmitter for ORP Model ARM-Z KOBO-RP Transmitter for ORP Model ARM-Z measuring monitoring analysing COMPACT-LINE Measuring range -1999 to +1999 mv Switchable from ORP to ph ph or mv/orp (Oxidation Reduction Potential) and temperature

More information

SDMO. Changeover switch VERSO 200

SDMO. Changeover switch VERSO 200 User guide SDMO Changeover switch VERSO 200 V 1.1a 13/11/2013 33502029301_0_1 Summary 1 - Safety guidelines for installation and operation... 4 2 - Installation of the source changeover switch... 5 2.1

More information

Fan Coil Actuator FCA 2

Fan Coil Actuator FCA 2 Fan Coil Actuator FCA 2 FCA 2 4920210 Updated: Feb-16 (subject to changes) Page 1 of 89 Contents 1 Function description... 4 1.1 Operation and display... 5 1.2 Advantages of the FCA 2... 6 1.2.1 Special

More information

SAFE USE &OPERATIONS MANUAL

SAFE USE &OPERATIONS MANUAL www.uvprocess.com Product information Updates on UV Process Supply products ITEM PART NO. VersaProbe standard package M007-092 (Standard package includes UV-A+B probe.) Individual UV-A Remote Probe M007-093

More information

Model # PCYFC-10kW-250A PCYFC-20kW-250A PowerCycle Battery Conditioner Operation Manual

Model # PCYFC-10kW-250A PCYFC-20kW-250A PowerCycle Battery Conditioner Operation Manual Model # PCYFC-10kW-250A PCYFC-20kW-250A PowerCycle Battery Conditioner Operation Manual MAN-000014-00 REV. B TABLE OF CONTENTS SPECIFICATIONS... 1 CYCLER CONTROLS AND USER INTERFACE... 3 OPERATING PROCEDURE...

More information

TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller. Mounting and Operating Instructions EB 5578 EN

TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller. Mounting and Operating Instructions EB 5578 EN TROVIS 5500 Automation System TROVIS 5578 Heating and District Heating Controller Mounting and Operating Instructions EB 5578 EN Firmware version 2.2x Edition May 2015 Definition of signal words DANGER!

More information

Warning! ESD sensitive device!

Warning! ESD sensitive device! Datasheet & Manual SenseAir S8 Miniature CO2 sensor Model SE-0031 3.2% CO2 Warning! ESD sensitive device! Document PSP 108 Key technical specifications Item Target gas Operating Principle SenseAir S8 CO2

More information

COMPACT-LINE. KOBO-pH Transmitter for ph-value Model APM-Z

COMPACT-LINE. KOBO-pH Transmitter for ph-value Model APM-Z KOBO-pH Transmitter for ph-value Model APM-Z measuring monitoring analysing COMPACT-LINE Measuring range ph -1 to 14 Switchable from ph to ORP Display of ph value, mv/orp (Oxidation Reduction Potential)

More information

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5

Heiland electronic GmbH TD / TD1 / TD2. B&W-Densitometers. USERS MANUAL Version 5 Heiland electronic GmbH TD / TD1 / TD2 B&W-Densitometers USERS MANUAL Version 5 2 Table of Contents 1. GENERAL INFORMATION...4 2. SAFETY REGULATIONS...5 3. AREA OF APPLICATIONS...5 4. INSTRUMENT DESCRIPTION...6

More information

Driesen + Kern GmbH. Product Specification. SenseAir S8. Commercial. Miniature infrared CO 2 sensor module. Document Rev Page 1 (10)

Driesen + Kern GmbH. Product Specification. SenseAir S8. Commercial. Miniature infrared CO 2 sensor module. Document Rev Page 1 (10) Driesen + Kern GmbH Product Specification SenseAir S8 Commercial Miniature infrared CO 2 sensor module 1 (10) SenseAir S8 Commercial Miniature infrared CO 2 sensor module Warning! ESD sensitive device!

More information

Driesen + Kern GmbH. Product Specification. SenseAir S8. Miniature CO 2 sensor. Document PSP 113. Rev 3. Page 1 (12)

Driesen + Kern GmbH. Product Specification. SenseAir S8. Miniature CO 2 sensor. Document PSP 113. Rev 3. Page 1 (12) Driesen + Kern GmbH Product Specification SenseAir S8 Miniature CO 2 sensor 1 (12) SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0017

More information

NX Series Inverters. HVAC Pocket Programming Guide

NX Series Inverters. HVAC Pocket Programming Guide NX Series Inverters HVAC Pocket Programming Guide HVAC Pocket Programming Guide HVAC Pocket Programming Guide / Contents This guide provides a single reference document for the user of NXL HVAC (product

More information

Heavy Duty Station Ref. HD-B

Heavy Duty Station Ref. HD-B www.jbctools.com Heavy Duty Station Ref. HD-B Safety It is imperative to follow safety guidelines to prevent electric shock, injury, fire or explosion. - Do not use the units for any purpose other than

More information

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 4. Page 1 (10)

Product Specification. SenseAir S8 LP. Miniature infrared CO 2 sensor module. Document PSP 126. Rev 4. Page 1 (10) Product Specification SenseAir S8 LP Miniature infrared CO 2 sensor module Document PSP 126 Rev 4 Page 1 (10) SenseAir S8 LP Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure

More information

Please enter the identity code of your device here!

Please enter the identity code of your device here! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured variable chlorine dioxide ProMinent D1C2-Cl02-001-GB ClO 2 0.15 ppm DULCOMETER STOP START ClO 2 0.15 ppm DULCOMETER STOP

More information

Xenon Weathering. Making the Most Advanced Instruments Even Better. Relative Irradiance. Wavelength (nm) Simplified Control Navigation

Xenon Weathering. Making the Most Advanced Instruments Even Better. Relative Irradiance. Wavelength (nm) Simplified Control Navigation Xenon Weathering Making the Most Advanced Instruments Even Better For over 100 years, Atlas instruments have revolutionized the science of weather durability testing. The new Atlas Ci4400 Weather-Ometer

More information

Flowmetering System Controller KMA4000 / PLU4000 Operating instructions

Flowmetering System Controller KMA4000 / PLU4000 Operating instructions System Controller KMA4000 / PLU4000 Operating instructions AVL Pierburg Flow Technology GmbH D-41460 Neuss Leuschstr. 11 Phone:+49 2131 40256-0 Fax:+49 2131 40256-99 Notes Before connecting and operating

More information

EXPOSURE UNIT USER S MANUAL

EXPOSURE UNIT USER S MANUAL EXPOSURE UNIT USER S MANUAL Drix Semiconductor, optics by Ralon Labs Author: Guy VDB Date : 05/07/2013 Drix NV Lerenveld 26, 2547 Lint, Belgium Tel: +32 34466100 Fax: +32 34885618 www.drixsemi.com info@drix.be

More information

Installing the Stuller Portable Controller STULLER. Step 1: Position the portable Stuller controller

Installing the Stuller Portable Controller STULLER. Step 1: Position the portable Stuller controller STULLER SINCE 1970 Stuller 302 rue Louis XIV Lafayette, Louisiana 70508-7700 800-877-7777 info@stuller.com Installing the Stuller Portable Controller Important Guidelines 1 Do NOT cut or bend the thermocouple.

More information

Brief description of KR 15 controllers

Brief description of KR 15 controllers Brief description of KR 15 controllers E N E R G Y R E C O V E R Y Page 2 Page 4 Page 5 Page 6 Page 9 1. Function of the controller 2. Terminal connections 3. Technical data 4. Initial operation 5. Manual

More information

Gas and Air Sensors. Product Specification. SenseAir. Miniature CO2 sensor

Gas and Air Sensors. Product Specification. SenseAir. Miniature CO2 sensor Gas and Air Sensors Product Specification SenseAir S8 Miniature CO2 sensor SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0050

More information

Please completely read through operating instructions! Do not discard! The warranty shall be invalidated by damage caused by operating errors!

Please completely read through operating instructions! Do not discard! The warranty shall be invalidated by damage caused by operating errors! Operating Instructions DULCOMETER D1C Part 2: Adjustment and Operation, Measured Variable Conductivity ProMinent D1C2-Leit.-001-GB Conductivity 100 µs/cm DULCOMETER STOP START Please completely read through

More information

Convey-All BTS 550 Wireless Seed Tender Remote Users Manual

Convey-All BTS 550 Wireless Seed Tender Remote Users Manual Convey-All BTS 550 Wireless Seed Tender Remote Users Manual Intercomp Co. 3839 County Road 116 Medina, MN 55340, U.S.A. (763)-476-2531 1-800-328-3336 Fax: 763-476-2613 www.intercompcompany.com Manual#

More information

Operating Instructions

Operating Instructions Operating Instructions (For Copier) Digital Colour Imaging Systems Model No. DP-C406 / C306 / C266 Before operating this machine, please carefully read this manual and keep this documentation in a safe

More information

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER

USER MANUAL FOR VISIBLE SPECTROPHOTOMETER USER MANUAL FOR VISIBLE SPECTROPHOTOMETER 1 Table of Contents 1. MAIN USAGES...3 2. WORKING ENVIRONMENT...3 3. MAIN TECHNICAL DATA AND SPECIFICATIONS...4 4. WORKING PRINCIPLE...5 5. OPTICAL PRINCIPLE...6

More information

Instruction Notes for 108A L Sensor Input

Instruction Notes for 108A L Sensor Input Operation Manual Instruction Notes for 108A L14-1800 Digital Control Module ON/OFF Main Power Switch Alarm Limits/ Cycle Switch Load Outlet (x2) General Description Sensor Input This temperature control

More information

USER MANUAL ENGLISH 2985-SX BANKNOTE COUNTER

USER MANUAL ENGLISH 2985-SX BANKNOTE COUNTER USER MANUAL ENGLISH 2985-SX BANKNOTE COUNTER INTRODUCTION ENGLISH Thank you for purchasing Safescan 2985-SX banknote counter. For proper use and maintenance, we advise to read this user manual carefully.

More information

CROSS Chassis from 160 A to 450 A

CROSS Chassis from 160 A to 450 A CROSS Chassis from 160 A to 450 A STS CATALOGUE Important note! The technical data enclosed is for general information. Please note that the operating instructions and references indicated on the products

More information

Photoresist Absorbance and Bleaching Laboratory

Photoresist Absorbance and Bleaching Laboratory MCEE 505 Lithography Materials and Processes Page 1 of 5 Photoresist Absorbance and Bleaching Laboratory Microelectronic Engineering Rochester Institute of Technology 1. OBJECTIVE The objective of this

More information

Alemite FCS Fluid Control System

Alemite FCS Fluid Control System 343000 343000-1 Alemite FCS Fluid Control System Installation and Operation Guide 670996 TABLE OF CONTENTS DISCLAIMER... 3 BEFORE YOU BEGIN........................................................ 4 1.0

More information

REPORT OF CALIBRATION of One Standard of Spectral Irradiance ( nm)

REPORT OF CALIBRATION of One Standard of Spectral Irradiance ( nm) Newport Corp Stratford, CT 6/6/08 REPORT OF CALIBRATION of One Standard of Spectral Irradiance (250-2400 nm) Oriel Part# 63358 Lamp Serial Number: 7~1803 1. Material One 45 watt, quartz halogen, tungsten

More information

Operation Manual. Canon CXDI-1 System Digital Radiography

Operation Manual. Canon CXDI-1 System Digital Radiography Canon CXDI-1 System Digital Radiography Operation Manual Before using the instrument, be sure to read this manual thoroughly. Also, read the manuals of other instruments in this system. Keep the manual

More information

ecotrans Lf 01/02 Transmitter / Switching Device for Conductivity Type B Operating Instructions 02.07/

ecotrans Lf 01/02 Transmitter / Switching Device for Conductivity Type B Operating Instructions 02.07/ ecotrans Lf 01/02 Transmitter / Switching Device for Conductivity Type 202731 B 20.2731.0 Operating Instructions 02.07/00435548 1 Notes 4 2 Identifying the device version 5 2.1 Type designation 5 3 Mounting

More information

Ci3000+ Weather-Ometer & Fade-Ometer. Accelerating Your Expertise. Setting the Standard for Xenon Weathering

Ci3000+ Weather-Ometer & Fade-Ometer. Accelerating Your Expertise. Setting the Standard for Xenon Weathering Ci3000+ Weather-Ometer & Fade-Ometer Setting the Standard for Xenon Weathering Accelerating Your Expertise Focused On Your Goals The Atlas Vision Shaping the future of the materials testing world in partnership

More information

AH Auto-Darkening Welding Helmet User Manual. Made in China

AH Auto-Darkening Welding Helmet User Manual. Made in China AH7-220-0000 Auto-Darkening Welding Helmet User Manual Made in China 2016 TECHNICAL SPECIFICATIONS AH7-220-0000 Auto-Darkening Welding Helmet 2016 WARNING Please read all instructions before using the

More information

NLG513 SOFTWARE. Translation of the original German operating instructions

NLG513 SOFTWARE.   Translation of the original German operating instructions SOFTWARE NLG513 Translation of the original German operating instructions www.brusa.biz BRUSA Elektronik AG Neudorf 14 CH-9466 Sennwald +41 81 758 19 00 info@brusa.biz LEGAL NOTICE Publisher BRUSA Elektronik

More information

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION

Measurement Method of High Absorbance (Low Transmittance) Samples by UH4150 INTRODUCTION INTRODUCTION With UH4150, a detector can be selected depending on the analysis purpose. When analyzing a solid sample which doesn t contain any diffuse components, by selecting the direct light detector,

More information

APX Mobile and Portable Automated Test and Alignment

APX Mobile and Portable Automated Test and Alignment APX Mobile and Portable Automated Test and Alignment Software Updates First things first! Be sure to check that you are running the latest software versions for the 8800SX and its applications. Visit the

More information

Add the plate to the AvalonN engine At the AvalonN engine press Offline, press Set plate and OK. Click on the next available plate and OK.

Add the plate to the AvalonN engine At the AvalonN engine press Offline, press Set plate and OK. Click on the next available plate and OK. Configure AvalonN and Apogee Prepress for a new plate size Follow these steps to add and configure a new plate size for use with the AvalonN platesetter and Apogee Prepress. Audience: Customer, Technical

More information

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO2 sensor. Document PSP 107. Rev 8. Page 1 (10)

Gas and Air Sensors. Product Specification. SenseAir S8. Miniature CO2 sensor. Document PSP 107. Rev 8. Page 1 (10) Gas and Air Sensors Product Specification SenseAir S8 Miniature CO2 sensor 8 1 (10) SenseAir S8 Miniature infrared CO 2 sensor module Warning! ESD sensitive device! Figure 1. SenseAir S8 Article no. 004-0-0013

More information

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No

Portable Appliance Testers. OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver Code No Portable Appliance Testers OmegaPAT MI 2140 BetaPAT MI 2141 User Manual Ver. 1.2. Code No. 20 750 684 Distributor: Producer: METREL d.d. Ljubljanska 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si

More information

OPERATING MANUAL. ba77044e01 03/2014 FDO 1100 IDS OPTICAL D.O. SENSOR

OPERATING MANUAL. ba77044e01 03/2014 FDO 1100 IDS OPTICAL D.O. SENSOR OPERATING MANUAL ba77044e01 03/2014 FDO 1100 IDS OPTICAL D.O. SENSOR FDO 1100 IDS Copyright 2014, SI Analytics GmbH Reprinting - even as excerpts - is only allowed with the explicit written authorization

More information

Application Note 26. Optical Hazard Measurements with JETI specbos 1211UV

Application Note 26. Optical Hazard Measurements with JETI specbos 1211UV Optical Hazard Measurements with JETI specbos 1211UV Contents 1 Introduction 3 2 Regulations 3 3 Categories of Optical Hazard 3 4 Available Accessories for JETI specbos 1211UV 5 5 Schemes and Peculiarities

More information

Operating manual. ph/oxi 3400i. mg/l. ARng. ph/oxygen measuring instrument. ba75686e01 01/2007

Operating manual. ph/oxi 3400i. mg/l. ARng. ph/oxygen measuring instrument. ba75686e01 01/2007 Operating manual ph/oxi 3400i O 7 9 2 M 1 7 6 ng C TP mg/l STO CAL O ph/oxygen measuring instrument ba75686e01 01/2007 Accuracy when going to press The use of advanced technology and the high quality standard

More information

Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as

Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as Read and be familiar with this manual before operating or servicing this device. To ensure operator, technician, and patient safety, use only as specified in this manual. Product Manufacturer: Medical

More information

Easy-to-Use RF Device & User-Friendly Windows Software

Easy-to-Use RF Device & User-Friendly Windows Software December 4, 2018 Low Cost / Profile High resolution Frequency Stability Measurement Test Set Pico Second Resolution Instrument Easy-to-Use RF Device & User-Friendly Windows Software APPLICATIONS Calibration

More information

User Manual. High Efficiency Locator Beacon

User Manual. High Efficiency Locator Beacon User Manual RT-B77 HELB High Efficiency Locator Beacon Edited by: RHOTHETA Elektronik GmbH Kemmelpark Dr.-Ingeborg-Haeckel-Str. 2 82418 Murnau Germany Tel.: +49 8841 4879-0 Fax: +49 8841 4879-15 Internet:

More information

MINAC Systems Mobile Induction Heating Equipment USER MANUAL. MINAC 25/40, 50/80 MINAC 25 Twin. Version MCCC 2.2.1

MINAC Systems Mobile Induction Heating Equipment USER MANUAL. MINAC 25/40, 50/80 MINAC 25 Twin. Version MCCC 2.2.1 Mobile Induction Heating Equipment USER MANUAL MINAC 25/40, 50/80 MINAC 25 Twin Version MCCC 2.2.1 Manufacturer: EFD Induction a.s P.O. Box 363, N-3701 Skien, Norway Tel: +47 35506000 Fax: +47 35506010

More information

The New Standard in Lightfastness Testing

The New Standard in Lightfastness Testing The New Standard in Lightfastness Testing Xenon Lightfastness Tester Model B02 s The New Standard in Lightfastness Testing Designed specifically to meet ISO 105 B02, the new rotating rack Q-Sun Model

More information

OM-60-TH SERVICE LOGGER WITH TEMPERATURE/HUMIDITY/DEW POINT SENSOR GETTING STARTED

OM-60-TH SERVICE LOGGER WITH TEMPERATURE/HUMIDITY/DEW POINT SENSOR GETTING STARTED OM-60-TH SERVICE LOGGER WITH TEMPERATURE/HUMIDITY/DEW POINT SENSOR The OM-60-TH is a Service Logger with a temperature/humidity/dew point sensor. The OM-60-TH is specified for air use only. The sensor

More information

JOINT STOCK COMPANY. MULTICHANNEL RECEIVER RI-4010M (version RM ) User guide

JOINT STOCK COMPANY. MULTICHANNEL RECEIVER RI-4010M (version RM ) User guide JOINT STOCK COMPANY MULTICHANNEL RECEIVER RI-4010M (version RM1-60619) User guide Safety requirements Before using the multichannel receiver RI-4010M read this user guide and follows safety requirements!

More information

Learning to use your easypod. [somatropin (rdna origin) for injection]

Learning to use your easypod. [somatropin (rdna origin) for injection] Learning to use your easypod [somatropin (rdna origin) for injection] Get to know the device Before you use your easypod for the first time, take a moment to get to know the device. Cartridge Door Button

More information

Disclaimers. Important Notice

Disclaimers. Important Notice Disclaimers Disclaimers Important Notice Copyright SolarEdge Inc. All rights reserved. No part of this document may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means,

More information

TLC3-BCR-U Series Thermostat

TLC3-BCR-U Series Thermostat OVERVIEW TLC3-BCR-U Series Thermostat Features Low power energy consumption: < 1W per unit Relays switching for outputs each up to 300W Temperature control for 2 or 4-pipe heating or cooling systems. Optional

More information

Remote Maintenance Box, RMB

Remote Maintenance Box, RMB OPERATOR S MANUAL Remote Maintenance Box, RMB Warnings and legal information Operating procedure Voltage measurement Document no.: 4189340730B SW version AGC 3.40.0 or later and AGC 4.00.0 or later Table

More information

INSTRUCTION MANUAL DISSOLVED OXYGEN-METER MODEL CDO-01

INSTRUCTION MANUAL DISSOLVED OXYGEN-METER MODEL CDO-01 INSTRUCTION MANUAL DISSOLVED OXYGEN-METER MODEL CDO-01 EL-221, MIDC Electronic Zone, Mhape, Navi Mumbai-400710. Tel: 022- 61393000 CHAPTER 1 INTRODUCTION Contech CDO-01 measures Dissolved Oxygen and temperature

More information

QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS

QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS QUANTUM Qflash T2 / X2 OPERATING INSTRUCTIONS 1.0 DESIGNATIONS T2 AND X2 1. Removable Reflector, two positions Normal and Wide angle. 2. Flash-tube 2A. Modeling Lamp (for Model X2 only) 3. Bounce Head,

More information

PRODUCT SAFETY SOLUTIONS MODEL HWI-01. Model HWI-01 Serial Number:

PRODUCT SAFETY SOLUTIONS MODEL HWI-01. Model HWI-01 Serial Number: USER S MANUAL HOT WIRE IGNITION TESTER MODEL HWI-01 Model HWI-01 Serial Number: 1 HWI-01 HOT WIRE IGNITION TESTER IMPORTANT INSTRUCTIONS - - SAVE THESE INSTRUCTIONS Introduction Thank you for choosing

More information

Getting started with your NutriOpt On-site Adviser. Better feed choices today

Getting started with your NutriOpt On-site Adviser. Better feed choices today Getting started with your NutriOpt On-site Adviser Better feed choices today This manual explains how to get started with the NutriOpt On-site Adviser. To make the most of this innovative solution, please

More information