1 Design Considerations for Wrist- Wearable Heart Rate Monitors Wrist-wearable fitness bands and smart watches are moving from basic accelerometer-based smart pedometers to include biometric sensing such as heart rate monitoring. This trend is being driven by manufacturers who are looking for differentiation in the rapidly growing wearables market and by educated consumers who want to maximize their performance and fitness with more effective training. Real-time continuous heart rate monitoring helps to motivate consumers to adapt their workouts based on heart rate zone feedback. Designers of these next-generation wearables face several challenges when implementing continuous heart rate monitoring in their products: Measurement accuracy of heart rate Tracking accuracy while exercising Validating performance across many unique individuals Reducing size and thickness of wearable designs Extending battery life Author Kevin Kilbane Senior Product Manager, Optical Sensors, Silicon Labs Wearable system designers have the choice of implementing non-invasive optical or electrical techniques. Electrical techniques typically use chest straps with two contacts on the skin to measure heart rate. This method is usually accurate but requires end users to wear a bulky band around their chest with a battery and wireless communication, which can be uncomfortable and inconvenient. Optical techniques are preferred as they eliminate the uncomfortable chest strap by using an optical emitter and sensor integrated with the wristband. Principle of Wrist Based Optical Heart Rate Monitoring As shown in Figure 1, infrared (IR) light from an LED shines on the skin of the user s wrist. Light entering the skin gets absorbed, scattered and reflected from tissue, bones, veins and arteries. An optical sensor is used to detect these weak reflections. Light reflected from tissue and bones is time-invariant and results in a direct current (DC)-only level. As the heart pumps, the light reflected from arterial blood changes and results in an alternating current (AC) signal. Advanced digital signal processing removes the DC signal and calculates the heart rate from the weak AC signal. This processing technique generates what is known as a photoplethysmography (PPG) signal.
2 Figure 1: IR Light Detection in a Heart-Rate Monitoring System Many factors contribute to accurate PPG signals in an optical design. These factors include the wavelength of the optical emitter(s), the sensitivity of the system at those wavelengths, spacing between the emitter and sensor, the amount of light generated by the LED, the number of emitters and noise in the system. The ideal system for heart rate monitoring for one user may not work well for another due to differences in skin pigmentation and other physiological characteristics of the unique wrist. Designers must choose carefully and use adaptive techniques where possible in order to work well with a broad consumer population. Failure to do so can result in dissatisfied customers and a high return rate of wearable products. Let s now consider a few of the critical aspects of heart rate monitor system design. Wavelength of Emitter Typical fingertip and ear-based heart rate monitoring systems use infrared (IR) emitters in the 850 nm to 940 nm band. IR emitters are not ideal for wrist-based solutions as the wrist does not have the same blood-rich capillaries close to the skin found in fingertip or ear locations. Green (525 nm) wavelengths have been found to provide superior performance in wrist locations for people with light complexions. Unfortunately, dark-pigmented skin can absorb green wavelengths. Yellow (590 nm) wavelengths have been found to work best with dark-pigmented skins. For best performance across the widest number of users, both green and yellow LEDs can be used at a small tradeoff to increased cost and power consumption. An adaptive technique is used to select the best signal to calculate heart rate for each unique individual. Opto-Coupling to Skin It is important to have good coupling of the optical signals between the wrist and the wearable device as air gaps reduce accuracy. Having a flexible wristband enables a close but comfortable fit. If the band is too tight, coupling is improved but blood flow can be restricted, resulting in worse accuracy. If the band is too loose, it will move freely. A common issue with single-led systems is that 2
3 the wristband may need to be adjusted farther up on the arm or rotated slightly for optimal performance. Using two LEDs spaced on opposite sides of the optical sensor minimizes issues due to placement of the band as well as with tilting. Tilting issues can occur during exercise when one side of the wristband has good skin coupling while the other side has an air gap. Three LEDs are recommended for wearable designs to ensure the highest accuracy across a wide cross-section of end users. For example, the high-performance Scosche Rhythm Plus fitness band uses two green LEDs and one yellow LED in a triangular setup as shown in Figure 2. Figure 2: Fitness Band Example with Multiple LEDs Emitter Light Energy The emitter light energy is primarily determined by the LED drive current, voltage, pulse on-time, half-angle and luminous intensity. Having an optical sensor system that enables control over several of these parameters allows the software to configure itself optimally for each unique individual. For example, green LEDs have a high forward voltage that may require a tradeoff with LED voltage and power output. A higher voltage for the LED is not always feasible, so a longer pulse-on time can be used to increase emitter light energy while staying within the LED s normal operating parameters. An auto-sensing function can adjust the LED driver current and/or the LED on-time to optimize the reflected signal for a unique individual. This automatic DC sensing helps minimize the dynamic range requirements of the system s analog-to-digital converter (ADC) and places the signal in the optimum range for detecting the weak AC heart rate signal. Tracking Accuracy When Exercising The biggest weakness of most commercially available wrist wearables is that they cannot accurately track heart rate while the user is exercising. It is very challenging to adequately compensate for motion and physiological artifacts when exercising. Accelerometers are typically used in wearables and can be effective when combined with advanced signal processing to reduce motion artifacts. These algorithms can use accelerometer data to reject heart rate samples corrupted by noise or to actively cancel the noise. Despite these algorithms, the heart rate signal can be temporarily lost. An adaptive algorithm that recognizes when the sensor data is not valid by using quality grading enables estimation techniques to provide consistent tracking accuracy when the 3
4 user is exercising. It is also important to validate the algorithms with a large sample of users representing different skin tones, ethnicities, ages and weights. Reducing End Product Size and Thickness Adding heart rate sensors requires more space in a wearable design. Many existing heart monitoring wearable designs use large discrete photodiodes combined with an analog front end (AFE) and MCU. The AFE includes the LED drivers, ADC, analog filtering and control. A smaller high-sensitivity photodiode integrated with the ADC along with analog filtering and LED drivers can have much lower noise floor, use fewer ADC bits, and smaller footprint. For example, Silicon Labs Si114x optical sensor integrates highsensitivity photodiodes, a 17-bit ADC, low-noise analog filtering, up to three dynamically configurable LED drivers and an I 2 C interface in a compact 2 mm x 2 mm clear QFN package. A typical AFE in 3 mm x 3 mm package and discrete photodiode in a 2 mm x 2 mm package has a 3x larger footprint. Increasing Battery Life The largest consumer of power for a heart rate monitor is typically the LED sampling power and the signal processing for motion artifact reduction. A key factor in power consumption is the sample rate used. Accurate heart rate monitoring can require faster sample rates when exercising with high beats per minute (BPMs). Using a dynamic algorithm that changes the sample rate based on BPM can maintain accuracy while minimizing power. Interpolation of samples can also enable lower power than raising the sample rate. Sensors that can dynamically change their LED drive currents can auto-sense DC level to reduce power and improve performance. Dynamically using one, two or three LEDs in the system design can also maintain high performance while minimizing power consumption. Conclusion Designing an optical heart rate monitoring solution in a wearable product presents many technical challenges. A high-performance integrated optical sensing solution such as Silicon Labs Si114x sensor family, combined with heart rate algorithms validated under clinical studies, enables developers to design robust wearable systems that maximize battery life and minimize the physical size of wrist-based heart rate monitoring devices. Silicon Labs has taken a major step toward making optical heart rate monitoring an integral feature for the fast-growing consumer wearable-device market. 4
5 5
Next Generation Biometric Sensing in Wearable Devices
Next Generation Biometric Sensing in Wearable Devices C O L I N T O M P K I N S D I R E C T O R O F A P P L I C AT I O N S E N G I N E E R I N G S I L I C O N L A B S C O L I N.T O M P K I N S @ S I L
More informationHealth and Fitness Analog solution. Wenbin Zhu Medical BDM June, 2015
Health and Fitness Analog solution Wenbin Zhu Medical BDM June, 2015 1 A Broad Market TI in Medical Devices Today TI HealthTech Engineering components for life. TI Solutions for Wearable Optical Bio-Sensing
More informationWRIST BAND PULSE OXIMETER
WRIST BAND PULSE OXIMETER Vinay Kadam 1, Shahrukh Shaikh 2 1,2- Department of Biomedical Engineering, D.Y. Patil School of Biotechnology and Bioinformatics, C.B.D Belapur, Navi Mumbai (India) ABSTRACT
More informationDesign of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals
Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals Mr. Vishwas Nagekar 1, Mrs Veena S Murthy 2 and Mr Vishweshwara Mundkur 3 1 Department of ECE, BNMIT, Bangalore 2 Assoc. Professor,
More informationSFH Photoplethysmography Sensor
SFH 7050 - Photoplethysmography Sensor Application Note draft version - subject to change without notice 1 Introduction This application note describes the use of the SFH 7050 (see Fig. 1) as the sensor
More informationRobust Wrist-Type Multiple Photo-Interrupter Pulse Sensor
Robust Wrist-Type Multiple Photo-Interrupter Pulse Sensor TOSHINORI KAGAWA, NOBUO NAKAJIMA Graduate School of Informatics and Engineering The University of Electro-Communications Chofugaoka 1-5-1, Chofu-shi,
More informationDESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR
DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR Srijan Banerjee 1, Subhajit Roy 2 1 Department of Electrical Engineering, Siliguri Institute of Technology, 2 Department of Electrical Engineering,
More informationdevices, to state of the art Infrared Data Association (IrDA) Transceivers and Sensor solutions.
1 Optical LITE-ON Technologies has evolved from just supplying traditional IR Discrete Emitter and Detecr devices, state of the art Infrared Data Association (IrDA) Transceivers and solutions. ISD (Infrared
More informationPHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION
PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION Tatyana Dimitrova Neycheva, Dobromir Petkov Dobrev Centre of Biomedical Engineering Ivan Daskalov Bulgarian Academy of Sciences, Bl. 105
More informationIntelligent Pillow for Heart Rate Monitor
International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 8, Issue 4 (August 2013), PP.47-52 Intelligent Pillow for Heart Rate Monitor Everlyn
More informationAPPLICATION OF HEART PHOTOPLETHYSMOGRAPHY
APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY 1 VICKY KUMAR SINGH, 2 SUMIT KUMAR THAKUR, 3 VINOD KUMAR 1,2,3 Department of Electronics Engineering, Bharati Vidyapeeth College of Engineering Pune E-mail: vickysingh229@gmail.com,
More informationHeart Rate Tracking using Wrist-Type Photoplethysmographic (PPG) Signals during Physical Exercise with Simultaneous Accelerometry
Heart Rate Tracking using Wrist-Type Photoplethysmographic (PPG) Signals during Physical Exercise with Simultaneous Accelerometry Mahdi Boloursaz, Ehsan Asadi, Mohsen Eskandari, Shahrzad Kiani, Student
More informationMasimo Corporation 40 Parker Irvine, California Tel Fax
Instruments and sensors containing Masimo SET technology are identified with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure accurate pulse oximetry
More informationA Twenty-Four Hour Tele-Nursing System Using a Ring Sensor
Proc. of 1998 Int. Conf. on Robotics and Automation Leuven, Belgium, May 16-20, 1998 A Twenty-Four Hour Tele-Nursing System Using a Ring Sensor Boo-Ho Yang, Sokwoo Rhee, and Haruhiko H. Asada d Arbeloff
More informationcommon type of cardiac diseases and may indicate an increased risk of stroke or sudden cardiac death. ECG is the most
ISSN: 0975-766X CODEN: IJPTFI Available Online through Research Article www.ijptonline.com DESIGNING OF ELECTRONIC CARDIAC EVENTS RECORDER *Dr. R. Jagannathan, K.Venkatraman, R. Vasuki and Sundaresan Department
More informationPhO 2. Smartphone based Blood Oxygen Level Measurement using Near-IR and RED Wave-guided Light
PhO 2 Smartphone based Blood Oxygen Level Measurement using Near-IR and RED Wave-guided Light Nam Bui, Anh Nguyen, Phuc Nguyen, Hoang Truong, Ashwin Ashok, Thang Dinh, Robin Deterding, Tam Vu 1/30 Chronic
More informationPreliminary Design Report with Diagram(s)
EEL 4914C Electrical Engineering Design (Senior Design) Preliminary Design Report with Diagram(s) 28 January 2008 Team Members: Name: Mark Oden Name: Carlos Manuel Torres Jr. Email: cerberus.rock@gmail.com
More informationAn Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient.
An Advanced Architecture & Instrumentation for Developing the System of Monitoring a Vital Sign (Oxygen Saturation) of a Patient. 1 Md.Mokarrom Hossain, 2 A.S.M.Mohsin*, 3 Md.Nasimul Islam Maruf, 4 Md.
More informationPHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION
PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION Dhiraj Sunehra 1, Thirupathi Samudrala 2, K. Satyanarayana 3, M. Malini 4 1 JNTUH College of Engineering,
More informationfnirs Sensor Data Sheet
FNIRS25102017 SPECIFICATIONS > Infrared emitter* >Peak emission: 860nm >Half intensity beam angle: ±13 deg >Spectral bandwitdth: 30nm >Radiant intensity: 750mW/sr > Red emitter* >Peak emission: 660nm >Half
More informationLaboratory Activities Handbook
Laboratory Activities Handbook Answer Key 0 P a g e Contents Introduction... 2 Optical Heart Rate Monitor Overview... 2 Bare Board Preparation... 3 Light Indicator... 5 Low Pass Filter... 7 Amplifier...
More informationCity, University of London Institutional Repository
City Research Online City, University of London Institutional Repository Citation: Rybynok, V., May, J.M., Budidha, K. and Kyriacou, P. A. (2013). Design and Development of a novel Multi-channel Photoplethysmographic
More informationD5.1 Report on the design of a fibre sensor based on NIRS
Optical Fibre Sensors Embedded into technical Textile for Healthcare Contract no.: FP6-027 869 Quality control Version : 2.0 Security: PU Nature: Prototype + Report (P, R) Workpackage: WP5 Start date of
More informationReal Time Heart Attack and Heart Rate Monitoring Android Application
Available Online at www.ijcsmc.com International Journal of Computer Science and Mobile Computing A Monthly Journal of Computer Science and Information Technology ISSN 2320 088X IMPACT FACTOR: 6.017 IJCSMC,
More informationDEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139
DEPARTMENT OF ELECTRICAL ENGINEERING AND COMPUTER SCIENCE MASSACHUSETTS INSTITUTE OF TECHNOLOGY CAMBRIDGE, MASSACHUSETTS 02139 Spring 2017 V2 6.101 Introductory Analog Electronics Laboratory Laboratory
More informationCompressive Wireless Pulse Sensing
Compressive Wireless Pulse Sensing CTS 205 Internet of Things Harvard University Kevin Chen Harnek Gulati HT Kung Surat Teerapittayanon Tracking reliable pulse waves for long term health diagnostics Motivation
More informationPulse Oximetry. Principles of oximetry
Pulse Oximetry The principal advantage of optical sensors for medical applications is their intrinsic safety since there is no electrical contact between the patient and the equipment. (An added bonus
More informationGet your daily health check in the car
Edition September 2017 Smart Health, Image sensors and vision systems, Sensor solutions for IoT, CSR Get your daily health check in the car Imec researches capacitive, optical and radar technology to integrate
More informationArterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects
Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Matti Huotari 1, Antti Vehkaoja 2, Kari Määttä 1, Juha
More informationDesign & Implementation of Pulseoxymeter to Measures the Oxygen Saturation in Blood
International Journal on Recent Innovation in Instrumentation & Control Engineering Vol. 2, Issue 1-2016 Design & Implementation of Pulseoxymeter to Measures the Oxygen Saturation in Blood INTRODUCTION
More informationELR 4202C Project: Finger Pulse Display Module
EEE 4202 Project: Finger Pulse Display Module Page 1 ELR 4202C Project: Finger Pulse Display Module Overview: The project will use an LED light source and a phototransistor light receiver to create an
More informationBiometric Data Collection Device for User Research
Biometric Data Collection Device for User Research Design Team Daniel Dewey, Dillon Roberts, Connie Sundjojo, Ian Theilacker, Alex Gilbert Design Advisor Prof. Mark Sivak Abstract Quantitative video game
More informationTowards inexpensive home Ambulatory BP Monitors [Work in Progress]
Towards inexpensive home Ambulatory BP Monitors [Work in Progress] 27 July 2009 Larry Beaty labeaty@ieee.org Phoenix Project, Twin Cities IEEE See http://www.phoenix.tc.ieee.org/ then sign up as a volunteer
More informationBy Pierre Olivier, Vice President, Engineering and Manufacturing, LeddarTech Inc.
Leddar optical time-of-flight sensing technology, originally discovered by the National Optics Institute (INO) in Quebec City and developed and commercialized by LeddarTech, is a unique LiDAR technology
More informationSpatially Resolved Backscatter Ceilometer
Spatially Resolved Backscatter Ceilometer Design Team Hiba Fareed, Nicholas Paradiso, Evan Perillo, Michael Tahan Design Advisor Prof. Gregory Kowalski Sponsor, Spectral Sciences Inc. Steve Richstmeier,
More informationBackground Suppression with Photoelectric Sensors Challenges and Solutions
Background Suppression with Photoelectric Sensors Challenges and Solutions Gary Frigyes, Product Manager Ed Myers, Product Manager Jeff Allison, Product Manager Pepperl+Fuchs Twinsburg, OH www.am.pepperl-fuchs.com
More informationPiezoresistive Carbon-based Hybrid Sensor for Body-Mounted Biomedical Applications
IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Piezoresistive Carbon-based Hybrid Sensor for Body-Mounted Biomedical Applications Recent citations - Sleep monitoring sensor
More informationChapter 2. Design and development of blood volume pulse sensor and heart rate meter. Abstract
Chapter 2 Design and development of blood volume pulse sensor and heart rate meter Abstract A low power, low cost sensor has been developed for sensing the blood volume pulse using transmission mode photoplethysmography
More informationA Digital Signal Processor for Musicians and Audiophiles Published on Monday, 09 February :54
A Digital Signal Processor for Musicians and Audiophiles Published on Monday, 09 February 2009 09:54 The main focus of hearing aid research and development has been on the use of hearing aids to improve
More informationSensors. CSE 666 Lecture Slides SUNY at Buffalo
Sensors CSE 666 Lecture Slides SUNY at Buffalo Overview Optical Fingerprint Imaging Ultrasound Fingerprint Imaging Multispectral Fingerprint Imaging Palm Vein Sensors References Fingerprint Sensors Various
More informationSMART LASER SENSORS SIMPLIFY TIRE AND RUBBER INSPECTION
PRESENTED AT ITEC 2004 SMART LASER SENSORS SIMPLIFY TIRE AND RUBBER INSPECTION Dr. Walt Pastorius LMI Technologies 2835 Kew Dr. Windsor, ON N8T 3B7 Tel (519) 945 6373 x 110 Cell (519) 981 0238 Fax (519)
More informationAn Intelligent Wearable e-belt for Continuous Monitoring of Sinus Rhythm
1 An Intelligent Wearable e-belt for Continuous Monitoring of Sinus Rhythm Arun.P 1, Amala M.C 2, Anjaly M 3, Githin T.S 4, Jomin J 5 1 Assistant Professor, 2,3,4,5 UG Scholar, Department of Electronics
More informationHow to Select the Right Positioning Sensor Solution A WHITE PAPER
How to Select the Right Positioning Sensor Solution A WHITE PAPER Published 10/1/2012 Today s machinery and equipment are continuously evolving, designed to enhance efficiency and built to withstand harsher
More informationLOW POWER WIRELESS PULSE OXIMETER TERMINAL
Technical Sciences and Applied Mathematics LOW POWER WIRELESS PULSE OXIMETER TERMINAL Dan LOZNEANU*, Paul BORZA*, Gheorghe PANĂ*, Horaţiu MOGA* * Transilvania University, Brasov, Romania Abstract: The
More informationPrinciple of Pulse Oximeter. SpO2 = HbO2/ (HbO2+ Hb)*100% (1)
Design of Pulse Oximeter Simulator Calibration Equipment Pu Zhang, Jing Chen, Yuandi Yang National Institute of Metrology, East of North Third Ring Road, Beijing, China,100013 Abstract -Saturation of peripheral
More informationECONOMICAL HEART RATE MEASUREMENT DEVICE WITH REMOTE MONITORING USING FINGERTIP
ECONOMICAL HEART RATE MEASUREMENT DEVICE WITH REMOTE MONITORING USING FINGERTIP PROJECT REFERENCE NO. : 37S1390 COLLEGE : SRI SIDDHARTHA INSTITUTE OF TECHNOLOGY, TUMKUR. BRANCH : TELECOMMUNICATION ENGINEERING
More informationMCT-MultiPlex Features Three Technologies
MCT-MultiPlex Features Three Technologies Near Infrared (NIR) based on MCT-360 NIR Transmitter; moisture, oil/fat, flavorings Visible (VIS) white light source color meter (200-800 nm); CIE L*, a*, b*;
More informationValidation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor
Phyllis K. Stein, PhD Associate Professor of Medicine, Director, Heart Rate Variability Laboratory Department of Medicine Cardiovascular Division Validation of the Happify Breather Biofeedback Exercise
More informationZKTECO COLLEGE- FUNDAMENTAL OF FINGER VEIN RECOGNITION
ZKTECO COLLEGE- FUNDAMENTAL OF FINGER VEIN RECOGNITION What are Finger Veins? Veins are blood vessels which present throughout the body as tubes that carry blood back to the heart. As its name implies,
More informationFrictioned Micromotion Input for Touch Sensitive Devices
Technical Disclosure Commons Defensive Publications Series May 18, 2015 Frictioned Micromotion Input for Touch Sensitive Devices Samuel Huang Follow this and additional works at: http://www.tdcommons.org/dpubs_series
More informationSignal Extraction Technology
Signal Extraction Technology Technical bulletin Introduction Masimo SET pulse oximetry is a new and fundamentally distinct method of acquiring, processing and reporting arterial oxygen saturation and pulse
More informationDesign of the Pulse Oximetry Measurement Circuit and Its Sensing System Based On CMOS
IOS Journal of Electrical and Electronics Engineering (IOS-JEEE) e-issn: 2278-1676,p-ISSN: 232-3331, Volume 12, Issue 1 Ver. IV (Jan. Feb. 217), PP 64-7 www.iosrjournals.org Design of the Pulse Oximetry
More informationRobust Wrist-Type Wireless Multiple Photo-Interrupter Pulse Sensor
Robust Wrist-Type Wireless Multiple Photo-Interrupter Pulse Sensor Toshinori Kagawa, Atsuko Kawamoto, and Nobuo Nakajima Abstract Long-term wearable vital sensors, monitoring parameters such as temperature,
More informationTechnical Explanation for Displacement Sensors and Measurement Sensors
Technical Explanation for Sensors and Measurement Sensors CSM_e_LineWidth_TG_E_2_1 Introduction What Is a Sensor? A Sensor is a device that measures the distance between the sensor and an object by detecting
More informationBio-Impedance Spectroscopy (BIS) Measurement System for Wearable Devices
Bio-Impedance Spectroscopy (BIS) Measurement System for Wearable Devices Bassem Ibrahim*, Drew A. Hall, Roozbeh Jafari* * Embedded Signal Processing (ESP) Lab, Texas A&M University, TX, USA BioSensors
More informationKeywords: Electronic Patch, Wireless Reflectance Pulse Oximetry, SpO2, Heart Rate, Body Temperature.
IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY Electronic Patch Wireless Reflectance Pulse Oximetry for Remote Health Monitoring S.Venkatesh Department of ECE, Anna University,Chennai,
More informationHeart Rate Monitoring using Adaptive Noise Cancellation
Heart Rate Monitoring using Adaptive Noise Cancellation 2015-2016 Q4 Bachelor Thesis by Bas Generowicz, 4029542 and Xenia Wesdijk, 4144074 Supervisors: R.C. Hendriks and S. Khademi at Delft University
More informationCity, University of London Institutional Repository
City Research Online City, University of London Institutional Repository Citation: Zaman, T., Kyriacou, P. A. & Pal, S. (2013). Free flap pulse oximetry utilizing reflectance photoplethysmography. 35th
More informationMULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN ARTERIES
IADIS International Journal on WWW/Internet Vol. 11, No. 3, pp. 116-125 ISSN: 1645-7641 MULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN Shusaku Nomura. Nagaoka University of Technology.
More informationChoosing and Using Photo Sensors
Part II Choosing and Using Photo Sensors Selection of the right photo sensor is the first step towards designing an optimal sensor-based system. The second step, and indeed a very important one, is the
More informationifeel Sensor USER GUIDE SUPPLEMENT
ifeel Sensor USER GUIDE SUPPLEMENT Choose Your ifeel Sensor There are two versions of the ifeel Sensor: USB and Bluetooth. Read the section of the user guide that matches your sensor. IFEEL BLUETOOTH SENSOR
More informationSelf-Contained, Passive, Non-Contact, Photoplethysmography: Real-Time Extraction Of Heart Rates From Live View Within A Canon PowerShot
Self-Contained, Passive, Non-Contact, Photoplethysmography: Real-Time Extraction Of Heart Rates From Live View Within A Canon PowerShot Henry Dietz, Chadwick Parrish, Kevin Donohue COIMG-146, 9:10AM, January
More informationTrans-African Hydro-Meteorological Observatory
Trans-African Hydro-Meteorological Observatory Sensor Design Competition A design by: 1. Kolyanga Emmanuel Email: emmakoly@gmail.com 2. Wogisha Benjamin Email: wogisha@gmail.com Executive Summary Proposal
More informationMoving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis
Moving from biomedical to industrial applications: OCT Enables Hi-Res ND Depth Analysis Patrick Merken a,c, Hervé Copin a, Gunay Yurtsever b, Bob Grietens a a Xenics NV, Leuven, Belgium b UGENT, Ghent,
More informationSensing. Autonomous systems. Properties. Classification. Key requirement of autonomous systems. An AS should be connected to the outside world.
Sensing Key requirement of autonomous systems. An AS should be connected to the outside world. Autonomous systems Convert a physical value to an electrical value. From temperature, humidity, light, to
More information6.111 Final Project Proposal HeartAware
6.111 Final Project Proposal HeartAware Michael Holachek and Nalini Singh Massachusetts Institute of Technology 1 Introduction Pulse oximetry is a popular non-invasive method for monitoring a person s
More informationPrelab for Laboratory Project 5: Operational amplifier
Worcester Polytechnic Institute Department of Electrical and Computer Engineering ECE3601 Intro to Electrical Engineering Laboratory Project 5: Operational amplifier I Prelab for Laboratory Project 5:
More informationSmall, Low Power, High Performance Magnetometers
Small, Low Power, High Performance Magnetometers M. Prouty ( 1 ), R. Johnson ( 1 ) ( 1 ) Geometrics, Inc Summary Recent work by Geometrics, along with partners at the U.S. National Institute of Standards
More informationVeraPath Optical Encoder Technology
TECHNICAL NOTES: OPTICAL TECHNOLOGIES VeraPath Optical Encoder Technology TN-1002 REV 160602 The Challenge MicroE s PurePrecision technology has enabled designers of precision motion control systems in
More informationThe OXY100C outputs four signals simultaneously, as shown in this graph: O 2 Saturation (beat-by-beat, CH 1) Pulse Waveform (beat-by-beat, CH 5)
Chapter 6 Specialty Modules NIBP100C OXY100C Pulse Oximeter Module The OXY100C Pulse Oximeter Module is primarily used to measure the blood oxygen saturation level in a non-invasive fashion. Via LEDs,
More informationWhat? nanowatt? Acquiring sensor data in wireless products with nanowatts of power consumption
11001101011010 10101010101010 10101010111010 01010011101101 010101 What? nanowatt? Acquiring sensor data in wireless products with nanowatts of power consumption Peggy Liska Texas Instruments Product Marketing
More informationBlood Group Detection and Mobile Monitoring System
International Conference on Innovative Trends in Electronics Communication and Applications 20 International Conference on Innovative Trends in Electronics Communication and Applications 2015 [ICIECA 2015]
More informationDRX Plus Detectors: Going from Good to Great
DRX Plus Detectors: Going from Good to Great Authors: Karin Töpfer, Tim Wojcik Introduction Carestream s introduction in 2009 of the world s first portable, wireless, cassette-sized detector the CARESTREAM
More informationAn acousto-electromagnetic sensor for locating land mines
An acousto-electromagnetic sensor for locating land mines Waymond R. Scott, Jr. a, Chistoph Schroeder a and James S. Martin b a School of Electrical and Computer Engineering b School of Mechanical Engineering
More informationRealizing Human-Centricity: Data-Driven Services
Realizing Human-Centricity: Data-Driven Services Ajay Chander R&D Lead, Data Driven Life Innovations Fujitsu Laboratories of America January 22, 2014 INTERNAL USE ONLY Copyright 2014 FUJITSU LIMITED Context:
More informationUltrasonic. Advantages
Ultrasonic Advantages Non-Contact: Nothing touches the target object Measures Distance: The distance to the target is measured, not just its presence Long and Short Range: Objects can be sensed from 2
More information(12) United States Patent (10) Patent No.: US 6,783,501 B2
USOO67835O1B2 (12) United States Patent (10) Patent No.: US 6,783,501 B2 Takahashi et al. (45) Date of Patent: Aug. 31, 2004 : (54) HEART RATE MONITOR AND HEART RATE 5,584,296 12/1996 Cui et al.... 600/479
More informationTABLE OF CONTENTS INTRODUCTION...04 PART I - HEALTH LEARNING...08 PART II - DEVICE LEARNING...12 PART III - BUILD...16 PART IV - DATA COLLECTION...
YOUTH GUIDE ENGINEER NOTES TABLE OF CONTENTS INTRODUCTION...04 PART I - HEALTH LEARNING...08 PART II - DEVICE LEARNING...12 PART III - BUILD...16 PART IV - DATA COLLECTION...18 PART V - COOL DOWN...22
More information(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( )
(19) (11) EP 1 424 934 B1 (12) EUROPEAN PATENT SPECIFICATION (4) Date of publication and mention of the grant of the patent: 06.08.08 Bulletin 08/32 (21) Application number: 01981641.2 (22) Date of filing:
More informationSimple Heartbeat Monitor for Analog Enthusiasts
Abigail C Rice, Jelimo B Maswan 6.101: Project Proposal Date: 18/4/2014 Introduction Simple Heartbeat Monitor for Analog Enthusiasts An electrocardiogram (ECG or EKG) is a simple, non-invasive way of measuring
More informationEE 230 Experiment 10 ECG Measurements Spring 2010
EE 230 Experiment 10 ECG Measurements Spring 2010 Note: If for any reason the students are uncomfortable with doing this experiment, please talk to the instructor for the course and an alternative experiment
More informationWEARABLE WIRELESS BODY AREA NETWORK
WEARABLE WIRELESS BODY AREA NETWORK Item Type text; Proceedings Authors Fajardo, Nicolas; Garrick, Kevin; Giroud, Xaviere; Kehn, Brian; Maggio, Andrew; Read, Cecilia Publisher International Foundation
More informationInput Stage Concerns. APPLICATION NOTE 656 Design Trade-Offs for Single-Supply Op Amps
Maxim/Dallas > App Notes > AMPLIFIER AND COMPARATOR CIRCUITS Keywords: single-supply, op amps, amplifiers, design, trade-offs, operational amplifiers Apr 03, 2000 APPLICATION NOTE 656 Design Trade-Offs
More informationNeuVision 500. Abundant and friendly display interface, multifold ECG display screen:
NeuVision 500 Features This monitoring system may be used to monitor patient s 6 physiological parameters: ECG, respiratory rate, body temperature, non-invasive blood pressure (NIBP), pulse oxygen saturation
More informationSoldier Tracking and Health Indication System Using ARM7 LPC-2148
Soldier Tracking and Health Indication System Using ARM7 LPC-2148 Shraddha Mahale, Ekta Bari, Kajal Jha Mechanism under Guidance of Prof. Elahi Shaikh (HOD) Electronics Engineering, Mumbai University Email:
More informationRevolutionizing 2D measurement. Maximizing longevity. Challenging expectations. R2100 Multi-Ray LED Scanner
Revolutionizing 2D measurement. Maximizing longevity. Challenging expectations. R2100 Multi-Ray LED Scanner A Distance Ahead A Distance Ahead: Your Crucial Edge in the Market The new generation of distancebased
More informationULS24 Frequently Asked Questions
List of Questions 1 1. What type of lens and filters are recommended for ULS24, where can we source these components?... 3 2. Are filters needed for fluorescence and chemiluminescence imaging, what types
More informationAdaptive Touch Sampling for Energy-Efficient Mobile Platforms
Adaptive Touch Sampling for Energy-Efficient Mobile Platforms Kyungtae Han Intel Labs, USA Alexander W. Min, Dongho Hong, Yong-joon Park Intel Corporation, USA April 16, 2015 Touch Interface in Today s
More informationE-health Project Examination: Introduction of an Applicable Pulse Oximeter
E-health Project Examination: Introduction of an Applicable Pulse Oximeter Mona asseri & Seyedeh Fatemeh Khatami Firoozabadi Electrical Department, Central Tehran Branch, Islamic Azad University, Tehran,
More informationPulse Sensor Individual Progress Report
Pulse Sensor Individual Progress Report TA: Kevin Chen ECE 445 March 31, 2015 Name: Ying Wang NETID: ywang360 I. Overview 1. Objective This project intends to realize a device that can read the human pulse
More informationNon-invasive Blood Oxygen Saturation Monitoring for Neonates Using Reflectance Pulse Oximeter
Non-invasive Blood Oxygen Saturation Monitoring for Neonates Using Reflectance Pulse Oximeter Wei Chen 1, Idowu Ayoola 1 1. Department of Industrial Design, Eindhoven University of Technology, Den Dolech
More informationVandal Proof Camera: v-cam 500 (D-WDR, 650 TVL, Sony Effio-E, 0.05 lx) Vandal Proof Camera: v-cam 500 (D-WDR, 650 TVL, Sony Effio-E, 0.
Vandal Proof Camera: v-cam 500 (D-WDR, 650 TVL, Sony Effio-E, 0.05 lx) Code: M10772 View of the camera View of the inside. Visible OSD keypad (on the left picture) and lens locking screws (on the right).
More informationPlaceLab. A House_n + TIAX Initiative
Massachusetts Institute of Technology A House_n + TIAX Initiative The MIT House_n Consortium and TIAX, LLC have developed the - an apartment-scale shared research facility where new technologies and design
More informationChapter 2 Mechatronics Disrupted
Chapter 2 Mechatronics Disrupted Maarten Steinbuch 2.1 How It Started The field of mechatronics started in the 1970s when mechanical systems needed more accurate controlled motions. This forced both industry
More informationWhitepaper EncoderBlue goes reflective
Whitepaper EncoderBlue goes reflective With the continuously increasing adoption of automated machinery in various applications, positioning devices are becoming an essential part of many systems. For
More informationMobileSOFT: U: A Deep Learning Framework to Monitor Heart Rate During Intensive Physical Exercise
MobileSOFT: U: A Deep Learning Framework to Monitor Heart Rate During Intensive Physical Exercise Vasu Jindal University of Texas, Dallas, TX vasu.jindal@utdallas.edu Abstract Wearable biosensors have
More informationMR Advance Techniques. Flow Phenomena. Class II
MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different
More informationA Radiation-Hardened, High-Resolution Optical Encoder for Use in Aerospace Applications
A Radiation-Hardened, High-Resolution Optical Encoder for Use in Aerospace Applications Pat Kreckie * Abstract Advances in aerospace applications have created a demand for the development of higher precision,
More informationMechatronics Engineering and Automation Faculty of Engineering, Ain Shams University MCT-151, Spring 2015 Lab-4: Electric Actuators
Mechatronics Engineering and Automation Faculty of Engineering, Ain Shams University MCT-151, Spring 2015 Lab-4: Electric Actuators Ahmed Okasha, Assistant Lecturer okasha1st@gmail.com Objective Have a
More informationMEMS Optical Scanner "ECO SCAN" Application Notes. Ver.0
MEMS Optical Scanner "ECO SCAN" Application Notes Ver.0 Micro Electro Mechanical Systems Promotion Dept., Visionary Business Center The Nippon Signal Co., Ltd. 1 Preface This document summarizes precautions
More information