Commissioning. Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests
|
|
- Brianna Foster
- 5 years ago
- Views:
Transcription
1 Acknowledgements David Shepard, Ph.D. Daliang Cao, Ph.D. Muhammad K. N. Afghan, Ph.D. Jinsong Ye, M.S. Tony P. Wong, Ph.D. Fan Chen, Ph.D. Min Rao, Ph.D. Vivek Mehta, M.D. Igor Gomola, Ph.D. David Housley Gerry Vantellingen
2 Commissioning Basic machine performance MLC Dose rate control Gantry speed control End-to-end tests
3 Picket fence test
4 Picket fence test with simulated error Gap 1.5 mm instead of 1 mm mm offset
5 Dose rate and gantry speed control Ratio of RapidArc to open field (%) x (mm)
6 Leaf speed control Ratio of RapidArc to open field (%) x (mm)
7 End-to-end test: Prostate - coronal y (mm) x (mm) 4.9% of pixels have γ > 1 (3%/3 mm)
8 End-to-end test: Prostate - coronal x-profile at y = 0.0 mm y-profile at x = 0.0 mm Film TPS Dose (cgy) Dose (cgy) x (mm) y (mm)
9 Interrupted delivery y (mm) Interrupted Normal x (mm)
10 Treatment planning system Treatment planning system may have additional commissioning requirements
11 The picket fence test during gantry rotation can detect gap and offset errors as small as 20% 20% 20% 20% 20% 0.1 mm 0.5 mm 1 mm 5 mm 1 cm 10
12 The picket fence test during gantry rotation can detect gap and offset errors as small as (2) 0.5 mm C. C. Ling, P. Zhang, Y. Archambault, J. Bocanek, G. Tang, and T. LoSasso, Commissioning and quality assurance of RapidArc delivery system, Int. J. Radiat. Oncol. Biol. Phys. 72, (2008)
13 IMAT QA Most IMRT QA devices measure the fluence from each beam angle for fixed-field IMRT plans. The dynamic nature of VMAT delivery, however, make the fluence measurement undesirable. A composite dose measurement represent a better fit for the purposes of VMAT plan QA.
14 VMAT QA Techniques 1. Film and ion chamber. 2. 2D diode array Mapcheck in MapPhan phantom 3. 2D ion chamber array MatriXX in MULTICube phantom 4. Other 2D/3D diode system Delta4 & ArcCheck 5. Dose reconstruction DosimetryCheck and IBA Compass
15 Film QA Absolute dose measurement Film Coronal Sagittal
16 2D Diode Array A MapCheck device inserted into a MapPhan solid water phantom
17 2D ion Chamber Array.07 cc air-vented ionization chambers Parallel read-out w/o dead time Sampling time: 20 msec 1020 detectors; 24.4 cm x 24.4 cm
18 MapCheck with dose rate correction relative dose response Mapcheck diode MatriXX ion chamber dose rate (MU/minute) The variation of dose response with the dose rate can reach up to 2.5% for MapCheck diode.
19 MapCheck with dose rate correction Dose calibration was performed at 224 MU/Min Gamma passing rate increased to 98.8%
20 Angular dose response of MapCheck diode detector array & phantom Couch Percent Error (%) phantom setup Gantry Angle A variation of the dose response of >20% can be found when the incidence beam angle is parallel to the detector plane. Each diode in MapCheck may have different angular dose response curve.
21 Angular variation of Matrixx dose response detector array & phantom Couch percent error (%) phantom setup Gantry Angle Up to 8% dose variation can be found when the photon beam is parallel to the detector plane
22 Angular response correction for MatriXX Before correction: passing rate 72.7% After correction: passing rate 98.1%
23 Gantry Angle Sensor Sensor is Fixed on the Gantry Inclinometer Detects Gantry Angle OmniPro IMRT Tags Measurement with Angle Correction is Applied
24 Arc Check Sun Nuclear 1386 diode detectors arranged in cylindrical geometry Measures entrance and exit dose Result is a composite dose for entire delivery
25 ArcCheck QA result (H&N) Gamma analysis passing rate is 95.2% for this case.
26 Scandidos Delta 4 A cylinder-shaped plastic phantom with 2 imbedded orthogonal crossing detector planes diode detectors Dose is recorded in 2 planes and a 3D dose is reconstructed for comparison with the QA plan.
27 Delta 4 3D Analysis From the 2D measurments, a 3D dose is reconstructed. User can scroll through slice-byslice. DVH analysis can also be performed. Dose can also be evaluated on a per beam basis and even a per control point basis. Courtesy of Scandidos
28 Gamma Comparison Display (TPS vs. COMPASS)
29 Dosimetry Check Dosimetry Check is a software package available through CIVCO and developed by Dean Renner. Fluences are recorded using the EPID during IMRT/IMAT beam delivery. The dose in patient is then determined as the measured fluence times a Dose Spread Function summed over the irradiated volume. Allows overlay of predicted and measured doses.
30 Reconstructed dose: green from plan, magenta from Dosimetry Check plan courtesy Art Pinkerton, Mississippi. Courtesy W.D. Renner
31 Summary There are a number of viable QA systems for VMAT patient specific QA. It is critical to take the time to characterize the system and understand its limitations. The implementation of a VMAT program may necessitate the purchase of additional QA tools.
32 The variation in angular response of a 2D ionization chamber array is typically 20% 1. Negligible 20% 20% 20% 20% 2. < 2% 3. 2% to 4% 4. 2% to 8% 5. > 8% 10
33 The variation in angular response of a 2D ionization chamber array is typically (4) 2% to 8% A. Van Esch, P. Zhang, C. Clermont C, M. Devillers, M. Iori, and D. P. Huyskens, On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom, Med. Phys. 34, (2007) J. Bocanek, I. Gomola, T. Depuydt, and F. Van den Heuvel, The feasibility of a 2D ion chamber array and the MultiCube phantom for quality assurance of RapidArc, Radiother. Oncol. 88 (Suppl. 2) S446 (2008)
34 Starting an IMAT program Form an implementation team Develop an implementation plan Establish a timeline Develop written procedures Evaluate training needs Determine case selection guidelines
35 Case selection Simple targets < 12 cm diameter Spherical or cylindrical no bifurcations Start with this... not this
36 Initial cohort standard plans For initial cohort of cases, develop standard plans Use as reference for comparison with IMAT plans Available as contingency treatment plan
3D Diode Array Commissioning: Building Confidence in 3D QA Technology
3D Diode Array Commissioning: Building Confidence in 3D QA Technology Caroline Yount, MS CANCER CENTER 3D QA The complex three-dimensional (3D) shapes of intensity modulated radiation therapy (IMRT) dose
More informationArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung
ArcCHECK, ein neuartiger QS-Ansatz bei der Rotationsbestrahlung Treffen des Arbeitskreises IMRT der DGMP Würzburg, 26 + 27.03.2009 Salih Arican Sun Nuclear Corporation QA Challenge for Rotational Beams
More informationThe Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry.
The Ultimate 4D QA Solution A 4D isotropic cylindrical detector array for arc delivery QA and Dosimetry. U.S.Patent No. 8,044,359; 6,125,335 Your Most Valuable QA and Dosimetry Tools 2 Y o u r M o s t
More informationVolumetric Modulated Arc Therapy. David Shepard Swedish Cancer Institute Seattle, WA
Volumetric Modulated Arc Therapy David Shepard Swedish Cancer Institute Seattle, WA Disclaimer Our VMAT work has been sponsored in part by Elekta. Outline David Shepard VMAT Basics and VMAT Plan Quality
More informationArcCHECK. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools
ArcCHECK The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359; 6,125,335 Compatible with: FFF Beams VMAT RapidArc TomoTherapy
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film As clinics strive to treat with the smallest field possible, the demand for SRS/SBRT QA grows as well.
More informationGantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 Gantry angle determination during arc IMRT: evaluation of a simple EPID-based technique and two commercial inclinometers Pejman Rowshanfarzad,
More informationSRS MapCHECK. SRS Patient QA, No Film. Your Most Valuable QA and Dosimetry Tools
SRS MapCHECK SRS Patient QA, No Film Your Most Valuable QA and Dosimetry Tools SRS Patient QA, No Film With improvements in targeting and localization, stereotactic treatments have become prevalent. To
More informationArcCHECKTM. The Ultimate 4D QA Solution. Your Most Valuable QA and Dosimetry Tools. VMAT RapidArc TomoTherapy Pinnacle 3 SmartArc Conventional IMRT
TM The Ultimate 4D QA Solution A 4D isotropical cylindrical detector array for arc delivery QA and Dosimetry U.S.Patent No. 8,044,359 What is? The world s first true 4D detector array The world s first
More informationCharacterization, Commissioning and Evaluation of Delta 4 IMRT QA System. Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX.
Characterization, Commissioning and Evaluation of Delta 4 IMRT QA System Ram Sadagopan 1 UTMD Anderson Cancer Center Houston, TX. 1 Acknowledgements Collaborators: Jose Bencomo, Rafael. M. Landrove, Peter
More informationMonica Kishore. Medical Physics Graduate Program Duke University. Approved: Jennifer O Daniel, Co-Supervisor. Fang-Fang Yin, Co-Supervisor
Accuracy of Planar Dosimetry for Volumetric Modulated Arc Therapy Quality Assurance by Monica Kishore Medical Physics Graduate Program Duke University Date: Approved: Jennifer O Daniel, Co-Supervisor Fang-Fang
More informationSUN NUCLEAR. EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm. corporation. Your Most Valuable QA and Dosimetry Tools
EPIDose : An Overview of EPIDose and the EPIDose Process and Algorithm SUN NUCLEAR corporation Your Most Valuable QA and Dosimetry Tools introduction Pre-treatment dose QA is an important process required
More informationA Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images
A Generalized Strategy for 3D Dose Verification of IMRT/VMAT Using EPID-measured Transit Images Aiping Ding, Bin Han, Lei Wang, Lei Xing Department of Radiation Oncology, Stanford University School of
More informationInstallation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter
Installation und Kommissionierung des Viewray MRIdian Linac Hamburg, 28. Mai 2018 Sebastian Klüter MR-guided RT in Heidelberg Funded by the German Research Foundation (DFG) Heidelberg consortium received
More informationDevelopment of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA
Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA Gena M.A.H 1, Ahmed L.El-Attar 2, Elbadry M. Zahran 3, Hany El-Gamal
More informationThe Current State of EPID-Based Linear Accelerator Quality Assurance. Disclosures. Purpose of this First Talk 8/3/2017
The Current State of EPID-Based Linear Accelerator Quality Assurance Timothy Ritter, PhD, DABR, FAAPM 1 Disclosures Employed by the Veterans Health Administration Faculty appointment with the University
More informationEmerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA
Emerging Technology: Real-Time Monitoring of Treatment Delivery EPID Exit Dose QA Arthur Olch, PhD, FAAPM AAPM Spring Clinical Meeting, March 21, 2017 Or.. What Dose are the Patients Really Getting???
More informationIMRT Delivery System QA. IMRT Dose Delivery. Acceptance testing. Why: specific tests for IMRT? Accuracy of leaf positioning (gaps) MLC Alignment
1 IMRT Delivery System Q Thomas LoSasso, PhD Memorial Sloan Kettering Cancer Center IMRT Dose Delivery cceptance testing Commissioning Quality assurance Verification Q Why: specific tests for IMRT? 2.
More informationConflict Disclosure. Rotational IMRT. Arc therapy. Dynamic Arc therapy. Intensity Modulated Arc Therapy Principles and Perspectives
Intensity Modulated Arc Therapy Principles and Perspectives Cedric Yu University of Maryland Conflict Disclosure Advisory Council on Advanced Treatment Delivery, Varian Medical Systems, Inc. Patent License:
More informationEvaluation of a diode array for QA measurements on a helical tomotherapy unit
Evaluation of a diode array for QA measurements on a helical tomotherapy unit K. M. Langen, a S. L. Meeks, D. O. Poole, T. H. Wagner, T. R. Willoughby, O. A. Zeidan, and P. A. Kupelian Department of Radiation
More informationIntensity-modulated radiation therapy dose verification using fluence and portal imaging device
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Intensity-modulated radiation therapy dose verification using fluence and portal imaging device Iori Sumida, 1a Hajime Yamaguchi,
More information7/23/2014. Acknowledgements. Implementing a new digital medical accelerator. New Generation of Medical Accelerators
Implementing a new digital medical accelerator John Wong Johns Hopkins University AAPM, Austin, 2014 Acknowledgements Yin Zhang, Ken Wang, Kai Ding (Commissioning - JHU) Esteban Velarde, Joe Moore (QA
More information8/3/2016. The EPID Strikes Back. Novel Applications for Current EPID Technology. Joerg Rottmann, PhD. Disclosures and acknowledgements
The EPID Strikes Back Joerg Rottmann Brigham and Women s Hospital / Dana-Farber Cancer Institute Harvard Medical School Disclosures and acknowledgements Disclosures Varian MRA grant Acknowledgements Boston
More informationPHYSICS QUESTIONNAIRE FORM
PHYSICS QUESTIONNAIRE FORM Institution Name: Date: Contact Information (name, address, phone, fax, email): Physicist: Radiation Oncologist: Dosimetrist (if applicable): Study Coordinator (if applicable):
More informationPixel response-based EPID dosimetry for patient specific QA
Received: 16 May 2016 Accepted: 26 September 2016 DOI: 10.1002/acm2.12007 RADIATION ONCOLOGY PHYSICS Pixel response-based EPID dosimetry for patient specific QA Bin Han 1 Aiping Ding 1 Minghui Lu 2 Lei
More informationThe evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response
Iran. J. Radiat. Res., 2005; 3 (1): 3-10 The evaluation of minimum detectable phantom thickness change using a scanning liquid filled ion chamber EPID dose response M. Mohammadi 1,2,3* and E. Bezak 1,2
More informationPhysical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy
Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Thomas LoSasso, a) Chen-Shou Chui, and C. Clifton Ling Department
More information8/3/2017. Use of EPIDs for Non-Routine Linac QA. Disclosure. Learning Objectives. Parts of this project received support from Varian Medical System.
Use of EPIDs for Non-Routine Linac QA Bin Cai PhD Disclosure Parts of this project received support from Varian Medical System. Learning Objectives Learn the recent development of EPID based Non-routine
More informationAperture Based Inverse Planning AAPM Summer School 2003
Aperture Based Inverse Planning AAPM Summer School 003 D.M. Shepard, M.A. Earl, Y. Xiao, C.X. Yu Acknowledgements Ziping Jiang Allen Li Shahid Naqvi James Galvin Di Yan Prowess, Inc. University of Maryland
More informationData Acquisition Systems design for Quality Assurance in advanced radiation dosimetry
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2016 Data Acquisition Systems design for Quality Assurance in advanced
More informationCommissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010
Commissioning and Calibrating a Linear Accelerator State-of-the-Art in 2010 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation
More informationSensitivity study of an automated system for daily patient QA using EPID exit dose images
Received: 27 June 2017 Revised: 8 December 2017 Accepted: 27 January 2018 DOI: 10.1002/acm2.12303 RADIATION ONCOLOGY PHYSICS Sensitivity study of an automated system for daily patient QA using EPID exit
More informationUniversity of Wollongong. Research Online
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2013 Validation of a general model for Intensity Modulated Radiation
More informationCharacterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy
Received: 5 January 1 Accepted: 1 September 1 DOI: 1.1/acm.11 RADIATION ONCOLOGY PHYSICS Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external
More informationStability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Stability of the Helical TomoTherapy Hi Art II detector for treatment beam irradiations Karin Schombourg, François Bochud, Raphaël
More informationIsocenter and Field of View Accuracy Measurement Software for Linear Accelerator
Isocenter and Field of View Accuracy Measurement Software for Linear Accelerator Aleksei E. Zhdanov 1 and Leonid G. Dorosinskiy 1 Ural Federal University named after the first President of Russia B. N.
More informationEvaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array
Evaluation of dosimetry parameters of photons and electron beams using a linear ionization chamber array José A. Bencomo, * Geoffrey Ibbott, Seungsoo Lee, and Joao A. Borges Department of Radiation Physics.
More informationA SIMPLE METHOD TO BACK-PROJECT ISOCENTER DOSE OF RADIOTHERAPY TREATMENTS USING EPID TRANSIT DOSIMETRY
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 05-03 (2017) 01-18 A SIMPLE METHOD TO BACK-PROJECT ISOCENTER DOSE OF RADIOTHERAPY TREATMENTS USING EPID TRANSIT DOSIMETRY T. B. Silveira 1,2 ; B. Q. Cerbaro
More informationMobius3D. Software based IMRT QA
Mobius3D Software based IMRT QA What is Mobius Medical Systems? Clinical Expertise Software Expertise Nathan Childress, Ph.D., Founder Eli Stevens, Chief Technical Officer Support Expertise Physicists
More informationA 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery
A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery Paul A. Jursinic a) Medical College of Wisconsin, Radiation Oncology Department, Milwaukee, Wisconsin
More informationA diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A diagnostic tool for basic daily quality assurance of a tomotherapy Hi Art machine Iwein Van de Vondel, 1 Koen Tournel, 1 Dirk
More informationDepartment of Physics, State University of New York at Buffalo, Buffalo NY, USA
124 research article A fully electronic intensity-modulated radiation therapy quality assurance (IMRT QA) process implemented in a network comprised of independent treatment planning, record and verify,
More informationClinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification
Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification L. N. McDermott, M. Wendling, B. van Asselen, J. Stroom, J.-J. Sonke, M. van Herk, and B. J. Mijnheer a Department
More informationMonte Carlo study on a new concept of a scanning photon beam system for IMRT
NUKLEONIKA 2011;56(4):291 297 ORIGINAL PAPER Monte Carlo study on a new concept of a scanning photon beam system for IMRT Anna M. Wysocka-Rabin, Günter H. Hartmann Abstract. Intensity-modulated radiation
More informationInitial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA. Katja Langen Mariana Guerrero Shifeng Chen Shh..
Initial Experience with a Commercial System for Volumetric Analysis of Patient Specific QA Katja Langen Mariana Guerrero Shifeng Chen Shh..Mobius3D ACKOWLEDGEMENTS Dr. Katja Langen Dr. Shifeng Chen Dr.
More informationCharacterization of an in vivo diode dosimetry system for clinical use
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 2, SPRING 2003 Characterization of an in vivo diode dosimetry system for clinical use Kai Huang, 1, * William S. Bice, Jr., 2, and Oscar Hidalgo-Salvatierra
More informationA positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 10, NUMBER 4, FALL 2009 A positioning QA procedure for 2D/2D (kv/mv) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup Huaiqun Guan,
More informationThe physical characteristics of a SLIC-EPID for transmitted dosimetry
Iran. J. Radiat. Res., 2005; 2 (4): 175-183 The physical characteristics of a SLIC-EPID for transmitted dosimetry M. Mohammadi 1,2,3* and E. Bezak 1,2 1 School of Chemistry and Physics, The University
More informationUsing the frame averaging of as500 EPID for IMRT verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Using the frame averaging of as500 EPID for IMRT verification J. Chang* and C. C. Ling Medical Physics Department, Memorial Sloan
More informationRemy Manigold University of Nevada, Las Vegas, UNLV Theses, Dissertations, Professional Papers, and Capstones
UNLV Theses, Dissertations, Professional Papers, and Capstones 5-1-2016 Creating A Dynamic, Multi-Purpose Correction for Multiple Geometries and Field Sizes to Account for Off-Axis and Asymmetric Backscatter
More informationQA Considerations. QA for LGK Perfexion : : Follow NRC licensing guidelines (10( CFR ) Leksell Gamma Knife Perfexion
Leksell Gamma Knife Perfexion QA Considerations Paula L. Petti, Ph.D. Taylor McAdam Bell Neuroscience Institute Washington Hospital Healthcare System Fremont, CA 1 QA for LGK Perfexion : : Follow NRC licensing
More informationDosimetric IMRT verification with a flat-panel EPID
Dosimetric IMRT verification with a flat-panel EPID B. Warkentin Department of Medical Physics, Cross Cancer Institute and Department of Physics, University of Alberta, 11 University Avenue, Edmonton,
More informationDose-response characteristics of an amorphous silicon EPID
Dose-response characteristics of an amorphous silicon EPID Peter Winkler a Division of Medical Radiation Physics, Department of Radiotherapy and Radiobiology, Medical University of Vienna, Waehringer Guertel
More informationFour-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method. JiHyung Yoon.
Four-dimensional in vivo dosimetry by dose reconstruction using continuous EPID images and phase sorting method Director of Dissertation: Dr. Jae Won Jung Major Department: Physics by JiHyung Yoon July,
More informationComparison of peripheral dose measurements using Ionization chamber and MOSFET detector
ORIGINAL ARTICLES Comparison of peripheral dose measurements using Ionization chamber and MOSFET detector Gopiraj ANNAMALAI 1, Ramasubramanian VELAYUDHAM 2 ABSTRACT Received: 7.07.2009 Accepted: 2.11.2009
More informationAccurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Markus Wendling, Robert J. W. Louwe, a Leah N. McDermott, Jan-Jakob Sonke, Marcel van Herk, and Ben J. Mijnheer
More informationISPFILMQATM STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE. Supports all major radiotherapy technologies! FilmQA TM
FILMQA STATE-OF-THE-ART RADIOTHERAPY VERIFICATION SOFTWARE Supports all major radiotherapy technologies! FilmQA is optimized for use with Gafchromic film products, including EBT2 and RTQA2. FILMQA helps
More informationGafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications
GafChromic QuiCk Phantom with EBT3P/3+P Film and FilmQA Pro for Radiation Therapy Dosimetry Applications I. SCOPE The protocol applies to GafChromic EBT3P and EBT3+P films exposed in GafChromic QuiCk Phantom
More informationState of the Art Film Dosimetry
State of the Art Film Dosimetry Micke A., Lewis D. Advanced Materials Ashland proprietary technology, patents pending Film Dosimetry Radiochromic Film EBT2/EBT3 One-Scan Protocol Multi-channel Film Dosimetry
More informationThe on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO)
The on-line detectors of the beam delivery system for the Centro Nazionale di Adroterapia Oncologica(CNAO) A. Ansarinejad1,2, A. Attili1, F. Bourhaleb2,R. Cirio1,2,M. Donetti1,3, M. A. Garella1, S. Giordanengo1,
More informationSCINTILLATING FIBER DOSIMETER ARRAY
SCINTILLATING FIBER DOSIMETER ARRAY FIELD OF THE INVENTION [0001] This invention relates generally to the field of dosimetry and, more particularly, to rapid, high-resolution dosimeters for advanced treatment
More informationEffect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging
The University of Toledo The University of Toledo Digital Repository Theses and Dissertations 2008 Effect of slit scan imaging techniques on image quality on radiotherapy electronic portal imaging Dean
More informationA feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate cancer *
IOP PUBLISHING Phys. Med. Biol. 52 (7) 2147 2156 PHYSICS IN MEDICINE AND BIOLOGY doi:1.188/31-9155/52/8/7 A feasibility study of using conventional jaws to deliver IMRT plans in the treatment of prostate
More informationClinical helical tomotherapy commissioning dosimetry
Clinical helical tomotherapy commissioning dosimetry John Balog and Gustavo Olivera TomoTherapy Incorporated, Madison, Wisconsin 53717 and Department of Medical Physics, University of Wisconsin at Madison,
More informationQUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING
1 QUALITY CONTROL PHANTOMS FOR RADIOTHERAPY AND MEDICAL IMAGING QualiFormeD Phantoms A selection of test objects facilitating regulatory quality controls in radiation therapy and medical imaging Practical,
More informationComparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Comparison of measured Varian Clinac 21EX and TrueBeam accelerator electron field characteristics Samantha A.M. Lloyd, 1a Sergei Zavgorodni,
More informationA simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems
Lin et al. Radiation Oncology (2018) 13:222 https://doi.org/10.1186/s13014-018-1164-1 RESEARCH Open Access A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end
More informationCyberKnife Iris Beam QA using Fluence Divergence
CyberKnife Iris Beam QA using Fluence Divergence Ronald Berg, Ph.D., Jesse McKay, M.S. and Brett Nelson, M.S. Erlanger Medical Center and Logos Systems, Scotts Valley, CA Introduction The CyberKnife radiosurgery
More informationQuality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT
Quality control of Gamma Camera By Dr/ Ibrahim Elsayed Saad 242 NMT WHAT IS QUALITY? The quality of a practice is to fulfill the expectations and demands from: Patient Clinicain Your self Quality assurance
More informationDOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING
DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended
More informationCHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY
14 CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY 2.1 INTRODUCTION kv-cbct integrated with linear accelerators as a tool for IGRT, was developed to
More informationPerformance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 4, 2016 Performance evaluation of the RITG148 + set of TomoTherapy quality assurance tools using RTQA 2 radiochromic film Eric C. Lobb Department
More informationWhen small things matter. Small Field Dosimetry Application Guide
R A D I AT I O N T H E R A P Y When small things matter. Small Field Dosimetry Application Guide Contents 1 Introduction 1 Introduction 2 2 The Physics of Small Fields 3 3 Detector Types 10 4 Detector
More informationBeam Production, Characteristics and Shaping
Beam Production, Characteristics and Shaping Dr. Manfred Sassowsky Outline X-ray production 60 Co units Linear Accelerators Beam characteristics Beam shaping Beam Production, Characteristics and Shaping
More informationIMRT verification with a camera-based electronic portal imaging system
Home Search Collections Journals About Contact us My IOPscience IMRT verification with a camera-based electronic portal imaging system This article has been downloaded from IOPscience. Please scroll down
More informationElectronic Brachytherapy Sources. Thomas W. Rusch
Electronic Brachytherapy Sources Thomas W. Rusch Educational Objectives Understand key elements of ebx source construction & operation Understand the rationale and methods for air kerma strength calibration
More informationImpact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT doses for flattening filter free (FFF) beam of TomoTherapy Hi-Art TM machines
JBUON 2014; 19(4): 1105-1110 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Impact of energy variation on Cone Ratio, PDD10, TMR20 10 and IMRT
More informationMONTE CARLO MODELLING OF A-SI EPID RESPONSE: THE
5 10 15 MONTE CARLO MODELLING OF A-SI EPID RESPONSE: THE EFFECT OF SPECTRAL VARIATIONS WITH FIELD SIZE AND POSITION Laure Parent, Joao Seco, Phil M Evans Joint Department of Physics, The Institute of Cancer
More informationThe most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 BUSINESS UNIT OF ISP
The most Comprehensive, Reliable, Economical and Easy to use GAFCHROMIC film based RT QA system Updated Feb 08 GAFCHROMIC EBT dosimetry film Designed and optimized for ALL RT procedures Can be cut into
More informationTotal body irradiation dose optimization based on radiological depth
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 3, 2012 Total body irradiation dose optimization based on radiological depth Amjad Hussain, 1,3a Peter Dunscombe, 1,2,3 J. Eduardo Villarreal-
More informationPhantoms in Medical Physics (RT) U. Oelfke. Division of Radiotherapy & Imaging
in partnership with Phantoms in Medical Physics (RT) U. Oelfke Division of Radiotherapy & Imaging uwe.oelfke@icr.ac.uk Making the discoveries that defeat cancer 1. Introduction What is a phantom? Wiki:
More informationLow-cost flexible thin-film detector for medical dosimetry applications
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 2, 2014 Low-cost flexible thin-film detector for medical dosimetry applications P. Zygmanski, 1a C.Abkai, 2 Z. Han, 1 Y. Shulevich, 2 D. Menichelli,
More informationAn Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.
An Introduction to TG-142 Imaging QA Using Standard Imaging Products Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc. Goals Understand the nature and intent of TG 142 imaging
More informationInitial setup and subsequent temporal position monitoring using implanted RF transponders
Initial setup and subsequent temporal position monitoring using implanted RF transponders James Balter, Ph.D. University of Michigan Has financial interest in Calypso Medical Technologies Acknowledgements
More informationCOMPREHENSIVE TG-142 IMAGING AND MACHINE QA
QA SOFTWARE COMPREHENSIVE TG-142 IMAGING AND MACHINE QA Automate the analysis of over thirty TG-142 recommended QA tasks The rapid progress of Radiation Therapy has created the need for Quality Assurance
More informationInnovative silicon detectors for dosimetry in external beam radiotherapy
University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2015 Innovative silicon detectors for dosimetry in external beam
More informationAnalysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose
J. Radiat. Res., 53, 301 305 (2012) Analysis of Post-exposure Density Growth in Radiochromic Film with Respect to the Radiation Dose Katsumi SHIMA 1,2, Kunihiko TATEOKA 1 *, Yuichi SAITOH 1,2, Junji SUZUKI
More informationComparative performance evaluation of a new a-si EPID that exceeds quad high-definition resolution
JBUON 2018; 23(2): 507-513 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Comparative performance evaluation of a new a-si EPID that exceeds quad
More informationA proposed method for linear accelerator photon beam steering using EPID
Received: 13 January 2018 Revised: 11 May 2018 Accepted: 29 June 2018 DOI: 10.1002/acm2.12419 RADIATION ONCOLOGY PHYSICS A proposed method for linear accelerator photon beam steering using EPID Michael
More informationRotational total skin electron irradiation with a linear accelerator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Rotational total skin electron irradiation with a linear accelerator Eric P. Reynard, 1,a Michael D.C. Evans, 1 Slobodan Devic,
More informationSPECIFICATION. Kilovoltage X-ray calibration system for protection and diagnostic level dosimetry. Prepared by
SPECIFICATION Kilovoltage X-ray Prepared by Igor Gomola, Technical Officer, Project ECU6023, Date 2015-Oct-06 Revision Date Status Comments 0.1 2015-Oct-06 Draft Igor Gomola Page 1 of 12 1. Scope This
More informationBeam Commissioning and Annual QA Phantoms, Detectors & Accessories. iba-dosimetry.com
Beam Commissioning and Annual QA Phantoms, Detectors & Accessories iba-dosimetry.com Benefit from efficient and smartly designed beam commissioning & annual QA tools that make your life EASIEST. For MORE
More informationImprovements in dose calculation accuracy for small off-axis targets in high dose per fraction tomotherapy
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2012 Improvements in dose calculation accuracy
More informationPhysics of small megavoltage photon beam dosimetry
Physics of small megavoltage photon beam dosimetry P Andreo, Professor of Medical Radiation Physics Karolinska University Hospital, Stockholm, Sweden ICARO2 IAEA, Vienna, 2017 IAEA-AAPM working group IAEA
More informationCommissioning an Elekta Versa HD linear accelerator
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Commissioning an Elekta Versa HD linear accelerator Ganesh Narayanasamy, 1,2 Daniel Saenz, 1 Wilbert Cruz, 1,3 Chul S. Ha, 1 Niko
More informationPeace of Mind. Automated.
1 Peace of Mind. Automated. Automated and guided beam commissioning Why SMARTSCAN? Get thousands of beam scans done effortlessly. Save your valuable time and manual operations at the water phantom and
More informationDetectors for small field dosimetry
Detectors or small ield dosimetry Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK 1 080915 Overview Ideal detector Water calorimeter Ionization chamber
More informationNote: Gating Characteristics of an Elekta Radiotherapy Treatment Unit Measured with Three Types of Detector
Note: Gating Characteristics of an Elekta Radiotherapy Treatment Unit Measured with Three Types of Detector Short title: Gating Characteristics of an Elekta Radiotherapy Treatment Unit Philip M Evans,
More information) was derived from 50% of the central axis dose created by nominal light field using geometry and mathematical methods. Leaf position (X mlc.
JOURNAL O APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 6, 2012 A light field-based method to adjust rounded leaf end MLC position for split shape dose calculation correction in a radiation therapy
More informationA Guide to Radiochromic Film Dosimetry with EBT2 and EBT3
A Guide to Radiochromic Film Dosimetry with EBT2 and EBT3 David F. Lewis Advanced Materials Group Ashland Specialty Ingredients Spain, April 2014 What is Radiochromic Film? A film that instantly changes
More information