1 Standard(s) Customer Test Report Les Industries Spectralux Inc. Sphere Test Calibration Report CIE , IESNA LM-16-93, IESNA LM ABC Technologies, 1st Avenue, City, Province, Country, 1A1-1A1 General Information Lamp Details: CD89 Rated AC Values L C1 Tested 5 February 2007 Created 5 February 20 Sphere 25.0 C Temperature Humidity 7.5 % Lamp Type LED Stabilization Time: 30 Mninutes Tested By: Notes Description Manufacturer Catalog No. Traffic LED Module AAA GTFB Serial No Diameter Color 200 mm Green Approved Signatory: Signature: Type Voltage Current Apparent Power Power Factor N/A THD Direct AC V A 6.08 VA 1) Field performance may differ from laboratory measurements. Results are valid for tested material only. 2) The original electronic file or paper report cannot be edited in whole or in part without written consent of Spectra Lux Industries Inc. 3) This calibration report describes the performance of a single product and does not necessarily represent the average performance of a group of the same SSL product. N/A Test Report: L C1 Page 1 / 6

2 Lamp Calibration Method The calibration was performed in a 1.93 meter diameter Ulbricht sphere with a spectrophotometer covering the fullvisible spectrum range from 380 to 780 nanometers. During the test, supply current and voltage were monitored within ± 0.1 percent of the nominal electrical characteristics. ADC voltmeterand adc ammeter were connected between acalibrated DC power supply. The voltmeter is connected across the electrical power inputs of the SSL product. Good electrical contacts have been used to ensure the control of electrical parameters and an adequate stabilization period prior to collecting data. The self-absorbance was measured and ageometrical correction factor was applied to the luminous fluxvalue to take into account the sphere configuration. Equipment, reference standards and methods are traceable to NIST (National Institute of Standards and Technology). Test Report: L C1 Page 2 / 6

3 Equipment List Equipment Manufacturer Model Serial Number Auxiliary Power Supply Calibration Date Calibration Due Date American Reliance SPS150-7 B /01/ /01/04 Test Power Supply irdc CIF-3000A /01/ /01/08 Input Power Meter Yokogawa WT2 27E /08/ /08/02 Output Power Meter Yokogawa WT2 27E /08/ /08/02 Shunt Resistor Fluke Y /08/ /08/02 Current Multimeter HP HP3458A 25B /03/ /03/07 Voltage Multimeter HP HP3458A 25B /03/ /03/07 Spectrometer Ocean Optics USB2000N USB2E /07/ /08/24 Temperature Humidity Sensor Omega HH // //20 Test Report: L C1 Page 3 / 6

4 Electrical Measurements Input Frequency 60 Hz Active Power 5.92 W THDV [CSA] 0.27 % Voltage V(rms) Apparent Power 6.21 VA THDA [CSA] 3.82 % Current A(rms) Power Factor Max. Harmonic At 3rd order Percentage (%) Percentage (%) Input Voltage Harmonics Harmonic Input Current Harmonics Harmonic Test Report: L C1 Page 4 / 6

5 Spectral Radiant Flux (W/nm) Les Industries Spectralux Inc. Spectral Power Distribution Wavelength (nm) Peak Wavelength Luminous Flux Input Power 496 nm 62 lm 5.92 W Lumens/Watt.5 Full Width/Half Maximum Center Wavelength Centroid Wavelength Dominant Wavelength nm 899 nm 498 nm Excitation Purity Colorimeric Purity x y u v L* 0.11 a* b* Test Report: L C1 Page 5 / 6

6 y Color Region for Green Traffic Signal Lights x Test Report: L C1 Page 6 / 6

Sample Report. IES LM Test Report. LED Lamp 001. Test results reported for: Orb Optronix report: LEDC WG. Original issue date:

Sample Report. IES LM Test Report. LED Lamp 001. Test results reported for: Orb Optronix report: LEDC WG. Original issue date: IES LM 79 08 Test Report Test results reported for: report: Original issue date: LED Lamp 001 LEDC001 010 WG 1 January 2012 Prepared for: Testing performed by: LED ABC, Inc. Memorial Highway 1003 7th Ave.

More information

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P30LN-L540-C27-B25 tel: fax: lm x 0.

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P30LN-L540-C27-B25 tel: fax: lm x 0. Total Integrated Radiant Flux Total Integrated Luminous Flux Luminous Efficacy Peak Wavelength Dominant Wavelength Color Purity CCT CRI Ra Peak Candela Irradiance (W/m 2 *nm) Figure 3 Spectral Illuminance

More information

Report No.: HZ k/R1. (Lumens) Stabilization Time CRI (K) (Light & Power) Table 1: Executive Data Summary

Report No.: HZ k/R1. (Lumens) Stabilization Time CRI (K) (Light & Power) Table 1: Executive Data Summary Test Summary Sample Tested: ML4LS10SOLBK Luminous Efficacy Total Luminous Flux Power (Lumens /Watt) (Lumens) (Watts) Power Factor 55.2 552.2 10.01 0.9797 CCT Stabilization Time CRI (K) (Light & Power)

More information

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P30LN-L600-C30-B25 tel: fax: lm x 0.

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P30LN-L600-C30-B25 tel: fax: lm x 0. Total Integrated Radiant Flux Total Integrated Luminous Flux Luminous Efficacy Peak Wavelength Dominant Wavelength Color Purity CCT CRI Ra Peak Candela Irradiance (W/m 2 *nm) Figure 3 Spectral Illuminance

More information

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P20-L360-C27-B25 tel: fax: lm x 0.

Orb Optronix, Inc th Ave Kirkland, WA Device Ref #: P20-L360-C27-B25 tel: fax: lm x 0. Total Integrated Radiant Flux Total Integrated Luminous Flux Luminous Efficacy Peak Wavelength Dominant Wavelength Color Purity CCT CRI Ra Peak Candela Irradiance (W/m 2 *nm) Figure 3 Spectral Illuminance

More information

www.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-010 TEST OF PAR 38 LED LAMPS MODEL NOS. S8854 / S8861

More information

www.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-018 TEST OF PAR 30L LED LAMPS MODEL NOS. S8814 / S8821

More information

LM Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923

LM Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923 LM-79-08 Test Report for OSRAM SYLVANIA INC. 100 Endicott Street Danvers, MA 01923 Model(s): Spectrum Lighting SG12SQLED21W35KE1/12SQGW/3.5LA 8 May 2011 This test report presents the results of measurements

More information

www.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-039 TEST OF LED CANDLE LAMPS MODEL NOS. S8899 / S8903

More information

www.greenelectricalsupply.com REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G100125842 Date: August 17, 2010 REPORT NO. 100125842CRT-036 TEST OF LED CANDLE LAMPS MODEL NOS. S8891 / S8894

More information

IES LM MEASUREMENT AND TEST REPORT. Test Model: SRLED-RTUS24-4K. Jeanne Han/Safety Manager

IES LM MEASUREMENT AND TEST REPORT. Test Model: SRLED-RTUS24-4K. Jeanne Han/Safety Manager IES LM-79-08 MEASUREMENT AND TEST REPORT Test Model: SRLED-RTUS24-4K Report Type: Test Engineer: Electrical and Photometric tests including: Input Current, Power, Power Factor, Luminous Flux, Luminous

More information

Test Sample Description Catalog Number: Description:

Test Sample Description Catalog Number: Description: Project Number Date 16.01512 7/12/2016 Test Sample Description Catalog Number: Description: HS56-WH-30K9 LED Retro-fit Trim LAB 1 Sphere Photometry Test Method: Preburn Time: Time To Stabilize: Total Operating

More information

LM Test. for LED. Cabinet

LM Test. for LED. Cabinet LM-79-08 Test Report for ELEC-TECH INTERNATIONAL CO LTD No.1 Jinfeng Road, Tangjiawan Town, Xiangzhou District, ZhuhaiCity, Guangdong Province, P.R. China 5190855 LED Cabinet Light Model: 535021## #(##=11-30)

More information

IES LM MEASUREMENT AND TEST REPORT For

IES LM MEASUREMENT AND TEST REPORT For IES LM-79-08 MEASUREMENT AND TEST REPORT For WISDOM OPTOELECTRONICS TECHNOLOGY CO., LIMITED Floor 8,B2 Building,YangGuangn Industrial Park, Bao An District,ShenZhen,China Test Model: WSD-SB10W48-40K-D-T3-S

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB01842 PAGE: 1 OF 6 CATALOG NUMBER: SWISH2X2-29YN/D10 LUMINAIRE: WHITE PAINTED SHEET METAL HOUSING, 2 WHITE CIRCUIT BOARDS EACH WITH 60 LEDS, FROSTED POLYCARBONATE LENS IN THE CENTER,

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB01271 PAGE: 1 OF 5 LUMINAIRE: LUMINAIRE: FABRICATED WHITE PAINTED METAL HOUSING, 4 WHITE CIRCUIT BOARDS EACH WITH 120 LEDS, FROSTED HOLOGRAPHIC PLASTIC LENS. LENS FROSTED SIDE UP. LAMPS:

More information

LM-79 Test Report. Electrical and Photometric tests as required by IESNA LM-79 test standard

LM-79 Test Report. Electrical and Photometric tests as required by IESNA LM-79 test standard Lamp specifics: Luminaire Used: none none Rated Watts Voltage: 6.75 Watts Lamp Type: LPL-W18-120C-930 / 78875 / Rated Orientation CCT: VBD Test Requester E.Z. Cola Rated Voltage 12 V DC New Product: 1

More information

Report No.: HZ c

Report No.: HZ c Test Summary Model L13T8DF450-CG4 Luminous Efficacy (Lumens /Watt) 121.4 Total Luminous Flux (Lumens) 2038.0 Power (Watts)/2 16.79 Power Factor 0.9957 CCT (K) 4912 CRI 81.5 Stabilization Time (Light &

More information

Report No.: HZ a. 60 mins Note. 4000K, frosted lens Table 1: Executive Data Summary

Report No.: HZ a. 60 mins Note. 4000K, frosted lens Table 1: Executive Data Summary Test Summary Model MT8N28 Luminous Efficacy (Lumens /Watt) 101.5 Total Luminous Flux (Lumens) 1860.0 Power (Watts)/2 18.32 Power Factor 0.9972 CCT (K) 3948 CRI 82.9 Stabilization Time (Light & Power) 60

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB03176 PAGE: 1 OF 9 LUMINAIRE: CAST WHITE PAINTED FINNED METAL HEAT SINK,SHEETMETAL JUNCTION BOX AFFIXED TO TOP OF HEATSINK, 1 CIRCUIT BOARD WITH 10 LEDS. LAMPS: TEN WHITE MULTI-CHIP LIGHT

More information

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No Lighting Sciences, Inc. UL Verification Services 7826 E. Evans Rd. Scottsdale, Arizona 85260 USA P: 480-991-9260 F: 480-991-0375 lsi@lightingsciences.com www.lightingsciences.com Description: The sample(s)

More information

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No Lighting Sciences, Inc. UL Verification Services 7826 E. Evans Rd. Scottsdale, Arizona 85260 USA P: 480-991-9260 F: 480-991-0375 lsi@lightingsciences.com www.lightingsciences.com Description: The sample(s)

More information

LED Lighting Flux and Color Measurement System (LFC)

LED Lighting Flux and Color Measurement System (LFC) LED Lighting Flux and Color Measurement System (LFC) Fast, accurate and complete System for research and production LFC System is a high cost efficiency light measurement system. The system is designed

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB03414 PAGE: 1 OF 9 LUMINAIRE: EXTRUDED METAL HOUSING WITH HEAT SINK FINS, TWO WHITE CIRCUIT BOARD WITH THREE HUNDRED LEDS ON EACH BOARD, METAL REFLECTOR WITH SPECULAR FINISH, FLAT TRANSLUCENT

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB03145 PAGE: 1 OF 9 LUMINAIRE: EXTRUDED METAL HOUSING WITH HEAT SINK FINS, TWO WHITE CIRCUIT BOARD WITH THREE HUNDRED LEDS ON EACH BOARD, METAL REFLECTOR WITH SPECULAR FINISH, FLAT TRANSLUCENT

More information

L Test Report. LED TUBE 1odel: LED11.5WT8/48/850-PET-G7DR

L Test Report. LED TUBE 1odel: LED11.5WT8/48/850-PET-G7DR Quality Assured NVLAP LAB CODE 200960-0 L 1-79-08 Test Report for EIKO GLOBAL, LLC 23220 W 84th Street, Shawnee, KS 66227 LED TUBE 1odel: LED11.5WT8/48/850-PET-G7DR Laboratory: Leading Testing Laboratories

More information

LM Test Report. For. T8 LEDTUBE Model: LED9WT8 F/24/850K-G7D

LM Test Report. For. T8 LEDTUBE Model: LED9WT8 F/24/850K-G7D Quality Assured NVLAP LAB CODE 200960-0 LM-79-08 Test Report For EIKO GLOBAL, LLC 23220 W 84th Street, Shawnee, KS 66227 T8 LEDTUBE Model: LED9WT8 F/24/850K-G7D Laboratory: Leading Testing Laboratories

More information

Diogen Lighting,Inc 14 Inverness Drive East,Ste.E-128 Englewood, CO T8L W BEST Test Service Shenzhen Co., Ltd.

Diogen Lighting,Inc 14 Inverness Drive East,Ste.E-128 Englewood, CO T8L W BEST Test Service Shenzhen Co., Ltd. IESNA LM-79: 2008 Measurement and Test Report for 14 Inverness Drive East,Ste.E-128 Englewood, CO 80112 Jun 15, 2011 Product Name: T8 Tube Model No: T8L1200-18W-02-120-32 Test Engineer: David Zhang Report

More information

IESNA LM-79: Measurement and Test Report for Green Creative Ltd. Room , New Victory House, Wing Lok Street, Central, HONG KONG

IESNA LM-79: Measurement and Test Report for Green Creative Ltd. Room , New Victory House, Wing Lok Street, Central, HONG KONG IESNA LM-79: 2008 Measurement and Test Report for Green Creative Ltd. Room 1206-7, New Victory House, 93-103 Wing Lok Street, Central, HONG KONG Apr03,2013 Product Name: [MR16 Model No: [01-715-0/840-380

More information

Lighting Sciences

Lighting Sciences Lighting Sciences www.lightingsciences.com Lighting Sciences Inc. 7826 E. Evans Road Scottsdale, Arizona 85260 USA Tel: 480-991-9260 Fax: 480-991-0375 INTENSITY(CANDLEPOWER) SUMMARY ANGLE MEAN CP LUMENS

More information

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No Lighting Sciences, Inc. UL Verification Services 7826 E. Evans Rd. Scottsdale, Arizona 85260 USA P: 480-991-9260 F: 480-991-0375 lsi@ul.com www.lightingsciences.com Description: The sample(s) was(were)

More information

TEST REPORT. Job No SHA Date: February 19, 2017 REPORT NO SHA-003

TEST REPORT. Job No SHA Date: February 19, 2017 REPORT NO SHA-003 TEST REPORT Job No. 170101691SHA Date: February 19, 2017 REPORT NO. 170101691SHA-003 TEST OF Outdoor Full-Cutoff Wall-mounted Area Luminaires MODEL NO. WP-ADS30U-40B / WP-ADS30U-40B-PC RENDERED TO Maxlite,

More information

NVLAP LAB CODE:

NVLAP LAB CODE: REPORT NUMBER: RAB02177 PAGE: 1 OF 7 CATALOG NUMBER: SK21XL25RYY LUMINAIRE: STAMPED STEEL CEILING PAN WITH WHITE FINISH, 10 LED BOARDS EACH WITH 10 LEDS, ACRYLIC DROP LENS WITH SMOOTH FINISH. LAMPS: ONE-HUNDRED

More information

Lighting Sciences, Inc.

Lighting Sciences, Inc. Lighting Sciences, Inc. Color and Photometric Testing, According to IESNA LM 79 Standards and Procedures, Test Report Files for: Par 30, 9W, DWF, Dimmable, 120V, Glass (SKU #180064) Seesmart 4139 Guardian

More information

Report No.: HZ d

Report No.: HZ d Test Summary Model LLT-4-T8-15W-DBA-C-50K Luminous Efficacy (Lumens /Watt) 112.3 Total Luminous Flux (Lumens) 1953.0 Power (Watts)/2 17.39 Power Factor 0.9974 CCT (K) 5140 CRI 82.9 Stabilization Time (Light

More information

Test Sample Description Catalog Number: Description:

Test Sample Description Catalog Number: Description: Project Number Date 17.02759 10/12/2017 Test Sample Description Catalog Number: Description: P5641-3130K 6" 29W LED Down-Light Cylinder LAB 1 Sphere Photometry Test Method: Preburn Time: Time To Stabilize:

More information

Checked X.CAO

Checked X.CAO REPORT NUMBER: RAB02068 PAGE: 1 OF 7 CATALOG NUMBER: FXLED78SF - FXLED78T (SF=SLIPFITTER MOUNT & T=TRUNNION MOUNT) LUMINAIRE: CAST FINNED METAL HOUSING, MOLDED PLASTIC REFLECTOR WITH SPECULAR FINISH AND

More information

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No Lighting Sciences, Inc. UL Verification Services 7826 E. Evans Rd. Scottsdale, Arizona 85260 USA P: 480-991-9260 F: 480-991-0375 lsi@ul.com www.lightingsciences.com Description: The sample(s) was(were)

More information

INDEPENDENT TEST LABORATORY REPORT No

INDEPENDENT TEST LABORATORY REPORT No Lighting Sciences, Inc. UL Verification Services 7826 E. Evans Rd. Scottsdale, Arizona 85260 USA P: 480-991-9260 F: 480-991-0375 lsi@ul.com www.lightingsciences.com Description: The sample(s) was(were)

More information

LM-79 Test Report. GE Infusion M1000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard

LM-79 Test Report. GE Infusion M1000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard specifics: Luminaire Used: SRTGLEDGI SRTGLEDGI 15W 30K WD E1 GES66 15W 30KRated Voltage: Watts 15 watt Driver Type: Roal Strato GE RSLD035-7A Infusion M1000/83 Rated CCT: Orientation VBU GE Infusion M1000/830/W/N

More information

REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045

REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 REPORT 3933 US ROUTE 11 CORTLAND, NEW YORK 13045 Project No. G101183506 FOR THE SCOPE OF ACCREDITATION UNDER NVLAP LAB CODE 100402-0. Date: June 6, 2013 REPORT NO. 101183506CRT-001 TEST OF ONE LED FLOOD

More information

LM-79 Test Report. GE Infusion M2000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard

LM-79 Test Report. GE Infusion M2000/830/W/N. Electrical and Photometric tests as required by IESNA LM-79 test standard specifics: Luminaire Used: SGE4SQLEDGI SGE4SQLEDGI 30W 30K ND E1 AR0204LGI 30W 3SG Rated FGCC Voltage: Watts 30 watt Driver Type: Roal Strato GE RSLD035-07 Infusion M2000/83 Rated CCT: Orientation VBU

More information

LED14.5WPAR30S/FL/827-DIM

LED14.5WPAR30S/FL/827-DIM LED14.5WPAR30S/FL/827-DIM TABLE OF CONTENTS 1 - GENERAL INFORMATION... 3 1.1 PRODUCT DESCRIPTION FOR EQUIPMENT UNDER TEST (EUT)... 3 1.2 OBJECTIVE... 3 1.3 TEST FACILITY DESCRIPTION... 4 1.4 TEST EQUIPMENT

More information

Chameleon Spectroradiometer

Chameleon Spectroradiometer Chameleon Spectroradiometer SPEED / STABILITY / ACCURACY The Chameleon Series is designed to measure the absolute spectral of power, of the light under test. Adding optical probes and absolute calibration

More information

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION Transverse Distance in Units of Mounting Heights Street Side House Side 2 1 0 1 2 1.5.2 ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION Values based on 25 foot mounting height. 11 1/2 17 7/8 11 0 11 REPORT

More information

LM Test Report

LM Test Report LM-79-08 Test Report For ELEC-TECH INTERNATIONAL CO LTD No.1 Jinfeng Road, Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong Province, P.R. China 519085 Wall Pack Model name(s): 533041xx Representative

More information

Report No.: HZ f. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ f. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI Test Summary Sample Tested: L15T8SE450-G Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 125.1 1904.0 15.22 0.9679 CCT (K) CRI Stabilization Time (Light & Power)

More information

NVLAP LAB CODE LM Test Report. For. Maxlite INC. (Brand Name: Maxlite) 12 York Ave., West Caldwell, NJ 07006

NVLAP LAB CODE LM Test Report. For. Maxlite INC. (Brand Name: Maxlite) 12 York Ave., West Caldwell, NJ 07006 LM-79-08 Test Report For Maxlite INC (Brand Name: Maxlite) 12 York Ave., West Caldwell, NJ 07006 2x4 Luminaires for Ambient Lighting of Interior Commercial Spaces Model name(s): MLFP24EP50XX Remark: The

More information

Approved Method: Electrical and Photometric. Measurements of Solid-State Lighting Products

Approved Method: Electrical and Photometric. Measurements of Solid-State Lighting Products IESNA SUSTAINING MEMBER Date of issue Mar 17, 2015 Version 1.0 Total pages: 11 Test report of IES LM-79-08 Approved Method: Electrical and Photometric Measurements of Solid-State Lighting Products Rendered

More information

LM Test Report. for. Philips (China) Investment Co., Ltd. InstantFit LEDtube Model:

LM Test Report. for. Philips (China) Investment Co., Ltd. InstantFit LEDtube Model: LM-79-08 Test Report for Philips (China) Investment Co., Ltd. Building 9, Lane 888, Tianlin Road Shanghai, China InstantFit LEDtube Model: 9290002840 Laboratory: Leading Testing Laboratories NVLAP CODE:

More information

LM Test Report. for Elec-Tech International Co., Ltd. BR40 LAMP Model: XX(XX:01-10), WBR40L27S2A

LM Test Report. for Elec-Tech International Co., Ltd. BR40 LAMP Model: XX(XX:01-10), WBR40L27S2A LM-79-08 Test Report for Elec-Tech International Co., Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China BR40 LAMP Model: 525021XX(XX:01-10), WBR40L27S2A Laboratory:

More information

Approved Method: Electrical and Photometric Measurements of Solid-State Lighting Products

Approved Method: Electrical and Photometric Measurements of Solid-State Lighting Products Ref. No.: Version: 1.0 IESNA SUSTAINING MEMBER LCZP17060295 Date of issue: Jun. 29, 2017 Total pages: 12 Test report of IES LM-79-08 Approved Method: Electrical and Photometric Measurements of Solid-State

More information

IESNA LM-79: Measurement and Test Report for. Green Creative. Xinsen Science &Trade Park, Unit 327,839An'Yuan Road, Shanghai, China

IESNA LM-79: Measurement and Test Report for. Green Creative. Xinsen Science &Trade Park, Unit 327,839An'Yuan Road, Shanghai, China IESNA LM-79: 2008 Measurement and Test Report for Green Creative Xinsen Science &Trade Park, Unit 327,839An'Yuan Road, Shanghai, China I I l i \ II I: Feb 06, 2013 Product Name: LED Lamp Jl II Model No:

More information

REPORT 3933 US ROUTE 11, CORTLAND, NEW YORK 13045

REPORT 3933 US ROUTE 11, CORTLAND, NEW YORK 13045 FOR THE SCOPE OF. ACCREDITATION UNDER NVLAP LAB CODE 100402-0.. REPORT 3933 US ROUTE 11, CORTLAND, NEW YORK 13045 Project No. G101555841 Date: March 21, 2014 REPORT NO. 101555841CRT-001 TEST OF ONE LED

More information

Approved Method: Electrical and Photometric. Measurements of Solid-State Lighting Products

Approved Method: Electrical and Photometric. Measurements of Solid-State Lighting Products IESNA SUSTAINING MEMBER Ref. No.: LCGP15040508 Date of issue June 4, 2015 Version 1.0 Total pages: 11 Test report of IES LM-79-08 Approved Method: Electrical and Photometric Measurements of Solid-State

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Test #: L1212321 Date: 12/1/212 8165 E Kaiser Blvd. Anaheim, CA 9288 p. 714.282.227 f. 714.676.5558 NVLAP LAB CODE 2927 Test Report: L1212321 Model Number: LT24-4L-35K Report Prepared For: Raffino Inc.

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE MAXLITE 12 York Ave. West Caldwell, NJ.

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE MAXLITE 12 York Ave. West Caldwell, NJ. Test #: L08124408 Date:08/31/2012 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927 0 Test Report: Model Number: L08124408 RR61430W Report Prepared For: MAXLITE

More information

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION

ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION Transverse Distance in Units of Mounting Heights Street Side House Side 2 1 0 1.2.5 1 ISOFOOTCANDLE LINES OF HORIZONTAL ILLUMINATION Values based on 25 foot mounting height. 11 1/2 17 7/8 11 0.1.05.02.01.005

More information

Checked X.CAO

Checked X.CAO REPORT NUMBER: RAB01900 PAGE: 1 OF 7 CATALOG NUMBER: BULLET12Y LUMINAIRE: CAST FINNED METAL HEAT SINK, MOLDED TEXTURED PLASTIC REFLECTOR WITH SEMI-DIFFUSE FINISH, 1 CIRCUIT BOARD WITH 24 LEDS, CLEAR MICRO-PRISMATIC

More information

NVLAP LAB CODE LM Test Report. For. MaxLite Inc. (Brand Name: MaxLite) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For. MaxLite Inc. (Brand Name: MaxLite) 12 York Ave West Caldwell NJ 07006 LM-79-08 Test Report For MaxLite Inc. (Brand Name: MaxLite) 12 York Ave West Caldwell NJ 07006 Outdoor Wall-Mounted Area Luminaires Model name(s): BLP50AUT550BPCMA BLP50AUT530BPCMA Representative (Tested)

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Test #: L12125907R02 Date: 1/14/2013 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927 0 Test Report: Model Number: L12125907R02 LR24-5000L-35K Report Prepared

More information

TEST REPORT. Job No SHA Date: Apr 12,2017 REPORT NO SHA-006. TEST OF ONE LED Luminaire MODEL NO.

TEST REPORT. Job No SHA Date: Apr 12,2017 REPORT NO SHA-006. TEST OF ONE LED Luminaire MODEL NO. TEST REPORT Job No. 170400516SHA Date: Apr 12,2017 REPORT NO. 170400516SHA-006 TEST OF ONE LED Luminaire MODEL NO. RENDERED TO Overdrive Electronics Pvt. Ltd C-121, Hosiery Complex, Phase2 Extn. TEST:

More information

NVLAP LAB CODE LM Test Report. For. Maxlite Inc. (Brand Name: ) 12 York Avenue, West Caldwell, NJ

NVLAP LAB CODE LM Test Report. For. Maxlite Inc. (Brand Name: ) 12 York Avenue, West Caldwell, NJ LM-79-08 Test Report For Maxlite Inc (Brand Name: ) 12 York Avenue, West Caldwell, NJ 07006 LED Downlight Model name(s): RRECO84040W Test & Report By: Johnson Sun Engineer: Johnson Sun Date: Jun.10,2015

More information

Model No Stock No Calibration Due Date Chroma Programmable AC Source PS-AC02

Model No Stock No Calibration Due Date Chroma Programmable AC Source PS-AC02 Report No: L041608503 Date: 4/29/2016 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Report No: L041608503 Report Prepared For: GM LIGHTING 9830 W 190th St,

More information

Report No.: HZ j. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ j. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI Test Summary Sample Tested: T8FR17/835/DIR/LED Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts)/2 Power Factor 105.4 1761.0 16.70 0.9970 CCT (K) CRI Stabilization Time (Light

More information

IESNA LM-79: Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE. Dec 08, 2014

IESNA LM-79: Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE. Dec 08, 2014 IESNA LM-79: 2008 Measurement and Test Report for Halco Lighting Technologies ATL 2940-A PACIFIC DRIVE Dec 08, 2014 Product Name: LED Model No: 82031 Test Engineer: David Zhang Report No.: BTR66.181.14.0039.06

More information

Address: IES LM Luminaire Description: Indoor High Bay Light Source: LGIT 5630 Ballast/Driver: VPL Overdrive L320HBL46DIM/50K

Address: IES LM Luminaire Description: Indoor High Bay Light Source: LGIT 5630 Ballast/Driver: VPL Overdrive L320HBL46DIM/50K Customer Company & Address: Overdrive Electronics Pvt. Ltd C-121, Hosiery Complex, Phase2 Extn. Contact Person: Mohit K Mittal Phone Number: +91 9811204494 Email Address: mohit@overdriveasia.com Relevant

More information

ELATION LIGHTING 6122 S. EASTERN AVE, COMMERCE, CA Electrical and Photometric tests as required by the IESNA test standards.

ELATION LIGHTING 6122 S. EASTERN AVE, COMMERCE, CA Electrical and Photometric tests as required by the IESNA test standards. Test #: L09133804 Date: 10/9/2013 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927 0 Test Report: Model Number: L09133804 ELED Strip TW CW Report Prepared For:

More information

IESNA LM-79: Measurement and Test Report for Jiawei Technology (USA) Limited Lincoln Ave.Hayward, CA USA.

IESNA LM-79: Measurement and Test Report for Jiawei Technology (USA) Limited Lincoln Ave.Hayward, CA USA. IESNA LM-79: 2008 Measurement and Test Report for Jiawei Technology (USA) Limited 2305 Lincoln Ave.Hayward, CA 94545 USA Apr 03, 2015 Product Name: self-ballast LED Lamp Model No: DL-9PAR30-830-LN-25-D;ML-9PAR30-830-LN-25-D

More information

LM Test Report. for Elec-Tech International Co.,Ltd. LED PAR30 Spot light Model: XX(XX:41-50)

LM Test Report. for Elec-Tech International Co.,Ltd. LED PAR30 Spot light Model: XX(XX:41-50) LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED PAR30 Spot light Model: 524142XX(XX:41-50) Laboratory:

More information

LM Test Report. for Maxlite SK America Inc. 2*4 Retrofit Kits Model: RKT4514U5550DV

LM Test Report. for Maxlite SK America Inc. 2*4 Retrofit Kits Model: RKT4514U5550DV LM-79-08 Test Report for Maxlite SK America Inc. 12 York Ave West Caldwell NJ 07006 2*4 Retrofit Kits Model: RKT4514U5550DV Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806

More information

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10) LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540751XX(XX: 01-10) Laboratory:

More information

IES LM Report. Technical Requirements Table, V4.0, DLC Premium DesignLights Consortium Product Application Category

IES LM Report. Technical Requirements Table, V4.0, DLC Premium DesignLights Consortium Product Application Category IES LM-79-08 Report Partner: Partner Company Name Zhongshan Luxmode Industrial CO., LTD. Manufacturer/ OEM name (if different) Zhongshan Luxmode Industrial CO., LTD. Primary Contact Name Kevin Chao kevin.wlc@luxmodecorp.com

More information

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 61-70)

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 61-70) LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540751XX(XX: 61-70) Laboratory:

More information

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10)

LM Test Report. for Elec-Tech International Co.,Ltd. LED CEILING LIGHT Model: XX(XX: 01-10) LM-79-08 Test Report for Elec-Tech International Co.,Ltd No.1 Jinfeng Rd.,Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong province, China LED CEILING LIGHT Model: 540741XX(XX: 01-10) Laboratory:

More information

IESNA LM-79: Measurement and Test Report for EiKO Global, LLC. Sep 09, 2013 LED8WMR16/38/830-DIM. David Zhang BTR

IESNA LM-79: Measurement and Test Report for EiKO Global, LLC. Sep 09, 2013 LED8WMR16/38/830-DIM. David Zhang BTR IESNA LM-79: 2008 Measurement and Test Report for EiKO Global, LLC Sep 09, 2013 Product Name: LED Model No: LED8WMR16/38/830-DIM Test Engineer: David Zhang Report No.: BTR66.181.13.1294.01-1 Sample Received

More information

Photometric Indoor Test Report

Photometric Indoor Test Report Photometric Indoor Test Report Relevant Standards LM-79-28 Prepared For Sea Gull Lighting Andre Duljas 31 W. Washington Street Riverside, NJ 875 Catalog Number 96118S-32 LTL Test Number 2191 Test Date

More information

LM Test Report

LM Test Report LM-79-08 Test Report For ELEC-TECH INTERNATIONAL CO LTD No.1 Jinfeng Road, Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong Province, P.R. China 519085 LED High Bay Light Model name(s): 502288##

More information

Report No.: HZ h. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI

Report No.: HZ h. Total Luminous Flux (Lumens) Stabilization Time (Light & Power) Table 1: Executive Data Summary CRI Test Summary Sample Tested: 8PLE26LED35 Luminous Efficacy (Lumens /Watt) Total Luminous Flux (Lumens) Power (Watts) Power Factor 101.5 707.7 6.97 0.9815 CCT (K) CRI Stabilization Time (Light & Power) 3498

More information

IESNA LM-79: Measurement and Test Report for Green Creative Ltd. Room , New Victory House, Wing Lok Street, Central, HONG KONG

IESNA LM-79: Measurement and Test Report for Green Creative Ltd. Room , New Victory House, Wing Lok Street, Central, HONG KONG IESNA LM-79: 2008 Measurement and Test Report for Green Creative Ltd. Room 1206-7, New Victory House, 93-103 Wing Lok Street, Central, HONG KONG Sep 02,2013 Product Name: LED PAR30 Model No: 14PAR30G3DIM/827FL40

More information

Equipment Used Model No Stock No Calibration Due Date

Equipment Used Model No Stock No Calibration Due Date Report No: L091401801 Date: 9/16/2014 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Report No: L091401801 Report Prepared For: Olympia Lighting, inc. 50

More information

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MAXLED) 12 York Avenue West Caldwell, NJ

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MAXLED) 12 York Avenue West Caldwell, NJ LM-79-08 Test Report For MAXLITE INC. (Brand Name: MAXLED) 12 York Avenue West Caldwell, NJ 07006 LED Downlight Model name(s): RRECO61840W Representative (Tested) Model: RRECO61840W Test & Report By: Johnson

More information

Client submitted the sample. Catalog number is LRH240U. Received in working and undamaged condition. No modifications were necessary.

Client submitted the sample. Catalog number is LRH240U. Received in working and undamaged condition. No modifications were necessary. Report No: L031500702 Date: 3/13/2015 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Report No: L031500702 Report Prepared For: EiKO Global, LLC 23220 W.

More information

LM Test Report

LM Test Report LM-79-08 Test Report For ELEC-TECH INTERNATIONAL CO LTD No. 1 Jinfeng Road, Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong Province, P.R.China, 519085 LED Linear Ambient Light Model Name(s):

More information

LM Test Report. For FALKOR INDUSTRY INC. (Brand Name: Falkor Lighting) 5580 Bandini Blvd Bell,CA USA

LM Test Report. For FALKOR INDUSTRY INC. (Brand Name: Falkor Lighting) 5580 Bandini Blvd Bell,CA USA LM-79-08 Test Report For FALKOR INDUSTRY INC (Brand Name: Falkor Lighting) 5580 Bandini Blvd Bell,CA 90201 USA High-bay Luminaires for Commercial and Industrial Buildings Model name(s): FI-HB-2FT-110W

More information

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd. Rm 1902, Easey Comm Bldg Hennessy Rd Wanchai, HONG KONG

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd. Rm 1902, Easey Comm Bldg Hennessy Rd Wanchai, HONG KONG LM-79-08 Test Report For Hocan Group Co.,Ltd (Brand Name: SEPICN LED Lighting) Rm 1902, Easey Comm Bldg 253-261 Hennessy Rd Wanchai, HONG KONG LED Tube Light Model name(s): HC-T8-8FT-36W-ID(Clear) Representative

More information

NVLAP LAB CODE LM Test Report. For MAXLITE,INC. (Brand Name: N/A) 12 YORK AVENUE WEST CALDWELL,NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE,INC. (Brand Name: N/A) 12 YORK AVENUE WEST CALDWELL,NJ 07006 LM-79-08 Test Report For MAXLITE,INC (Brand Name: N/A) 12 YORK AVENUE WEST CALDWELL,NJ 07006 Direct Linear Ambient Luminaires Representative (Tested) Model: LSECO-4NU2050 Model Different: N/A Test & Report

More information

Calibration Due Date Chroma Programmable AC Source 61604

Calibration Due Date Chroma Programmable AC Source 61604 Test #: L04130402 Date: 5/2/2013 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Test Report: Model Number: L04130402 TVL4000 CW Report Prepared For: ELATION

More information

Test Report. Report Number: L14070 Date: Oct 9, Issued by: Dialight Optics Laboratory 1501 Route 34 South, Farmingdale, NJ 07727

Test Report. Report Number: L14070 Date: Oct 9, Issued by: Dialight Optics Laboratory 1501 Route 34 South, Farmingdale, NJ 07727 Test Report Report Number: L14070 Date: Oct 9, 2014 Issued by: 1501 Route 34 South, Farmingdale, NJ 07727 Test of one Vigilant 60K High Bay Unit manufacturer: Dialight Corporation Unit model number: H6UGNC44NXXXX

More information

IESNA LM (NOT INCLUDE SPATIAL NON-UNIFORMITY OF CHROMATICITY) ELECTRICAL AND PHOTOMETRIC MEASUREMENTS OF SOLID-STATE LIGHTING PRODUCTS

IESNA LM (NOT INCLUDE SPATIAL NON-UNIFORMITY OF CHROMATICITY) ELECTRICAL AND PHOTOMETRIC MEASUREMENTS OF SOLID-STATE LIGHTING PRODUCTS IESNA LM-79-08 (NOT INCLUDE SPATIAL NON-UNIFORMITY OF CHROMATICITY) ELECTRICAL AND PHOTOMETRIC MEASUREMENTS OF SOLID-STATE LIGHTING PRODUCTS MEASUREMENT AND TEST REPORT For ELEC-TECH INTERNATIONAL CO.,

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Test #: L10131601 Date: 10/10/2013 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Test Report: Model Number: L10131601 ELED Strip TW BOTH A Report Prepared

More information

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT)

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD. (Brand Name: THAILIGHT) LM-79-08 Test Report For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD (Brand Name: THAILIGHT) Sanhui Ind. Area, Cunwei, Hengli, Dongguan, China. Outdoor Pole/Arm-mounted Area and Roadway Luminaires

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Test #: L113411R1 Date: 1/28/213 8165 E Kaiser Blvd. Anaheim, CA 9288 p. 714.282.227 f. 714.676.5558 NVLAP LAB CODE 2927 Test Report: Model Number: L113411R1 LR24-4L-5K-D-S Report Prepared For: Raffino

More information

LM Test Report

LM Test Report LM-79-08 Test Report For ELEC-TECH INTERNATIONAL CO LTD No.1 Jinfeng Road, Tangjiawan Town, Xiangzhou District, Zhuhai City, Guangdong Province, P.R. China 519085 LED Ceiling Light Model name(s): 544661##

More information

LM Test Report. for Maxlite SK America Inc. WALLPACK Model: MLSWP30LED50DS

LM Test Report. for Maxlite SK America Inc. WALLPACK Model: MLSWP30LED50DS LM-79-08 Test Report for Maxlite SK America Inc. 12 York Ave West Caldwell NJ 07006 WALLPACK Model: MLSWP30LED50DS Laboratory: Leading Testing Laboratories NVLAP CODE: 200960-0 Tel: +86-571-56680806 www.ledtestlab.com

More information

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING)

NVLAP LAB CODE LM Test Report. For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD. (Brand Name: GI LED LIGHTING) LM-79-08 Test Report For GREEN INOVA LIGHTING TECHNOLOGY (SHENZHEN) LTD (Brand Name: GI LED LIGHTING) 4th floor, building 2, Zone 2, Hong Hua Ling Industrial Park, Liu xian Block,Nan Shan District, Shenzhen,

More information

LM Test Report. For. Revolution Lighting Technologies, Inc. (Brand Name: ) 2280 Ward Ave. Simi Valley, CA

LM Test Report. For. Revolution Lighting Technologies, Inc. (Brand Name: ) 2280 Ward Ave. Simi Valley, CA LM-79-08 Test Report For Revolution Lighting Technologies, Inc (Brand Name: ) 2280 Ward Ave. Simi Valley, CA. 93065 Outdoor Pole/Arm-Mounted Area and Roadway Luminaires Model name(s): 113XS8-3LT Remark:

More information

illumia Plus Temperature Controlled Light Measurement Systems

illumia Plus Temperature Controlled Light Measurement Systems illumia Plus Temperature Controlled Light Measurement Systems Allows testing over a broad range of temperatures Measure with Confidence When a solid-state light engine is used in a luminaire or fixture,

More information

Model No Stock No Calibration Due Date Chroma Programmable AC Source PS-AC02

Model No Stock No Calibration Due Date Chroma Programmable AC Source PS-AC02 Report No: L071600303 Date: 7/6/2016 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Report No: L071600303 Report Prepared For: Revolution Lighting Technologies

More information

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE

8165 E Kaiser Blvd. Anaheim, CA p f NVLAP LAB CODE Report No: L071600301 Date: 7/6/2016 8165 E Kaiser Blvd. Anaheim, CA 92808 p. 714.282.2270 f. 714.676.5558 NVLAP LAB CODE 200927-0 Report No: L071600301 Report Prepared For: Revolution Lighting Technologies

More information