A Thesis. entitled. Automated Signal to Noise Ratio Analysis for Magnetic Resonance Imaging Using a. Noise Distribution Model. Abdullah M Aldokhail

Size: px
Start display at page:

Download "A Thesis. entitled. Automated Signal to Noise Ratio Analysis for Magnetic Resonance Imaging Using a. Noise Distribution Model. Abdullah M Aldokhail"

Transcription

1 A Thesis entitled Automated Signal to Noise Ratio Analysis for Magnetic Resonance Imaging Using a Noise Distribution Model by Abdullah M Aldokhail Submitted to the Graduate Faculty as partial fulfillment of the requirements for the Master of Science Degree in Biomedical Sciences - Medical Physics E. Ishmael Parsai, Ph.D., Committee Chair Kerry Krugh, Ph.D., Committee Member Diana Shvydka, Ph.D., Committee Member Dr. Amanda Bryant-Friedrich, Dean College of Graduate Studies The University of Toledo August 2016

2 Copyright 2016, Abdullah M Aldokhail This document is copyrighted material. Under copyright law, no parts of this document may be reproduced without the expressed permission of the author.

3 An Abstract of Automated Signal to Noise Ratio Analysis for Magnetic Resonance Imaging Using a Noise Distribution Model by Abdullah M Aldokhail Submitted to the Graduate Faculty as partial fulfillment of the requirements for the Master of Science in Degree in Biomedical Science - Medical Physics The University of Toledo August 2016 Signal to noise ratio (SNR) is a fundamental index of image quality in Magnetic Resonance Imaging (MRI) and it has traditionally been considered as a sensitive test to monitor the performance of MRI systems. As part of quality assurance program, SNR measurement should have a high level of reproducibility and accuracy. Currently accepted methods to evaluate SNR rely upon manual analysis and are observer dependent. The goal of this project is to develop a robust automated method for SNR analysis to eliminate individual s variability when performing such a test. This method uses a macro file that has been written using imagej software to automatically segment the phantom from the background region, measure the mean signal from an area covering 80% of the phantom and determine the noise by fitting a noise distribution model to the background histogram. Fifty-four phantom scans from a variety of RF coils were used to compare the automated SNR analysis method with the manual SNR methods described in the ACR QC Manual. The automated SNR method proved to be accurate, highly reproducible, and relatively insensitive to the existence of minor artifacts in the noise measurement region. iii

4 Acknowledgements I would like to express my sincere gratitude to my professor and advisor Dr. Kerry Krugh for introducing the topic to me and for the continuous guidance and support during my study and my research. This thesis would not have been possible without his help and support. I would like to thank the MRI staff for giving me the time to spend on the machines and for being helpful and kind. Finally, I would like to express my deep gratitude to my parents and my wife for the continuous support and encouragement throughout my years of study. This achievement would not have been possible without them. iv

5 Table of Contents Abstract Acknowledgements Table of Contents List of Tables List of Figures iii iv v vii viii 1 Introduction Basic Principles of MRI Magnetic Proprieties of Nucleus Interaction of Protons with an External Magnetic Feld MR Signal Detection and Image Formation Primary Components of MRI Scanner Quality Control for MRI System Signal to Nose Ratio Noise Distribution in Magnitude MR Image Signal to Noise Ratio Measurements Automated SNR Measurement Objective of the Study 18 2 Materials and Methods Materials MRI Scanners RF Coils and Phantoms Image Analysis Tools 21 v

6 2.2 Methods Scanning Procedure and Parameters Automated Assessment of Signal to Noise Ratio Validation of the Automated SNR Method Background Artifacts and Noise Estimation 41 3 Results 43 4 Discussion 52 5 Conclusion 57 References 58 Appendix A : Phantom Set Up 61 Appendix B : ImageJ Macro File 62 vi

7 List of Tables 1.1 Properties of selected MR active nuclei Correction factors for multi-channels coils as a function of channel numbers ImageJ built in functions utilized in the automated SNR method Scanning protocols for each RF coil Mean SNR and standard deviation for the automated SNR analysis method (SNRauto) Mean SNR and standard deviation for the single-image SNR analysis method (SNRsingle) Mean SNR and standard deviation for the two-image SNR analysis method (SNRdual) A comparison between the three methods of SNR measurement over all phantom scans of different radiofrequency coils (and number of channels) Mean (R 2 ), standard deviation (SD), and %CV for different RF coils Comparison between noise estimation obtained from the background ROI standard deviation and the noise determined by the noise model fit SNR based on SNR single and SNR auto in presence and absence of background artifacts.51 vii

8 List of Figures 1.1: Protons distribution in presence and absence of external magnetic field : precession of a proton in a magnetic field (Bo) : signal localization using read out gradient. The read out gradient induces a change in the proton precessional frequency (ω) depending on the spatial location within the sampling volume : Flowchart showing the main steps in the automated SNR measurement : A schematic illustration of a typical bimodal distribution : A schematic illustration of a histogram acquired from MRI phantom scan with a threshold value (T) corresponds to the signal intensity at the local minimum between the background and phantom peaks : Phantom selection in the phase encode direction Background selection for noise measurement : Phantom selection in the phase encode direction : background selection for noise measurement : Signal truncation along the edges of FOV : Schematic illustration of estimating the number of pixels with zero count : background signal histogram including signal truncation (left) and after eliminating the truncation artifact (right) : Background histogram before and after signal smoothing : SNR report with acquisition parameters : Single image method for SNR measurement : Two image method for SNR measurement...40 viii

9 2.15: Spine phantom scan shows background ROI placement to include structural artifacts (left) and away from visible artifacts (right) : Knee phantom scan shows background ROI placement in the phase encode direction to intentionally include structural artifacts (left) and the proper placement of ROI (right).42 Figure 3.1: SNR measurements of six RF coils from three different SNR methods.46 Figure 3.2: Background histogram for knee coil with artifacts (top) and without artifacts (bottom).49 Figure 3.3: Background histogram for spine coil with artifacts (top) and without artifacts (bottom).50 Figure 4.1: background noise histogram for shoulder coil before smooth (top) and after smooth (bottom) 53 Figure 4.2: difference image artifacts due to temporal instability in MRI data acquisition for breast (left) and spine (right) phantom scan 54 Figure 4.2: difference image artifacts due to temporal instability in MRI data acquisition for breast (left) and spine (right) phantom scan 56 A-1: Photographs show reference position for phantom set up ix

10 x

11 Chapter One Introduction Magnetic Resonance Imaging (MRI) is a non-invasive medical imaging modality that is used to create cross-sectional images of the internal body structures to aid physicians in the study and diagnosis of disease. MRI uses a very strong magnetic field (most commonly used field strength are 1.5T and 3T) and non-ionizing radiation (radio frequency waves) to generate the MR signal. MRI is based on principles of Nuclear Magnetic Resonance (NMR) that is when MR active nuclei are placed in a strong magnetic field, the protons will align with the magnetic field in one of two distinct energy states; either parallel (low energy state) or antiparallel (high energy state) to the direction of the applied magnetic field. A few excess protons will align in the low energy level resulting in a net magnetization pointing in the direction of the magnetic field (Mz). Additionally, the protons will start to precess around the main magnetic field at a specified frequency proportional to the strength of the applied magnetic field (precession frequency). If an external energy (radiofrequency pulse at the precessional frequency) is applied to the sampling volume, some of the protons in the low energy state will absorb the energy and will be flipped to the higher energy state. In addition, the RF pulse will cause protons to synchronize and precess in-phase with each other resulting in a displacement of the net magnetization to the transverse plane (Mx,y). The rotation of the net magnetization on the transverse plane will induce a voltage in an RF receiver coil that could be detected and subsequently used as an MR signal. This phenomenon was independently described by Felix Bloch and Edward Purcell in 1946 for which they shared the Nobel Prize in Physics in In 1973, Paul Lauterbur and Peter Mansfield, independently proposed a method to spatially localize NMR signals by using magnetic field 1

12 gradients for which they shared the Nobel Prize for Medicine in The first MR image of a human body was achieved by Raymond Damadian in MRI found wide applications in neurological and musculoskeletal imaging due to the excellent contrast resolution for soft tissue this modality provides. 1.1 Basic Principles of MRI Atoms consist of a central nucleus surrounded by negatively charged electrons. The nucleus contains positively charged particles called protons and electrically neutral particles called neutrons. There are three types of motion within an atom. In addition to the electron motion around the nucleus, both electrons and the nucleus spin around their own axis Magnetic properties of the Nucleus Magnetic properties of the nucleus are determined by the number of protons and neutrons. Nuclei with an odd mass number (i.e. the number of protons plus the number of neutrons is odd) possess a spin or an angular momentum and magnetic moment. Such nuclei, tend to interact with an applied magnetic field (MR active nuclide). On the other hand, nuclei with even mass number and even atomic number (i.e. the number of protons and the number of neutrons are both even), have no spin and no magnetic moment and will not interact with the external magnetic field (MR inactive nuclei )(Mark, Richard, & Thomas, 2004). Table 1.1 lists examples of relevant MR active nuclei. 2

13 Table 1.1: Properties of selected MR active nuclei (Bushberg, 2012). Nucleus Spin Number % Isotopes Magnetic Gyromagnetic Relative Abundance Moment Ratio (MHz/T) sensitivity H-1 1/ F-19 1/ x 10 6 Na-23 3/ x10 4 P-31 1/ x10 5 Of primary interest in MRI is the hydrogen atom (proton). Hydrogen is the most abundant nucleus in the human body. In addition, hydrogen which consists of a single proton, has a relatively large magnetic moment. For these reasons, MRI uses H-1 protons as a source for MR signals (Bushberg, 2012) Interaction of Protons with an External Magnetic Feld Each proton can be considered as a small magnet with a magnetic moment represented by magnitude and direction. In absence of an external magnetic field, the magnetic moment of a sample of protons will be randomly oriented with a net magnet moment of zero (Figure 1.1 A). When a strong magnetic field is applied to a sample containing a large number of protons, two important phenomena can be observed. The magnetic moment of protons will align with the magnetic field in two distinct energy levels; low energy level (parallel to the direction of magnetic field) and high energy level (antiparallel to the direction of the magnetic field). Alignment of protons depends on the thermal energy of the nuclide; high thermal energy nuclei will align antiparallel to the magnetic field, while low thermal energy nuclei align with the 3

14 magnetic field. At equilibrium, a few more protons will align parallel to the magnetic field ( low energy level) resulting in small, but measurable net magnetic moment in the direction of the applied magnetic field (Figure 1.1B) (Westbrook, Roth, & Talbot, 2011). Figure 1.1: Protons distribution in presence and absence of external magnetic field. Figure taken from Bushberg, J. T. (2012): The essential physics of medical imaging. The energy separation between the two energy levels is proportional to the strength of the applied magnetic field (Bushberg, 2012). As the energy separation increases, fewer protons will have thermal energy high enough to oppose the magnetic field. Therefore, a stronger magnetic field increases the net magnetization resulting in higher MR signal. In addition to the protons alignment in the magnetic field, the protons will start to precess around the magnetic field at specific angular precessional frequency (ω) as shown in Figure1.2. 4

15 Figure 1.2: precession of a proton in a magnetic field (Bo). Figure taken from Bushberg, J. T. (2012): The essential physics of medical imaging. The processional frequency (ω) is governed by Larmor equation: ω = γ x B0 (1-1) Where: ω: is the angular frequency ( radian/s) γ: is Gyromagnetic ratio (radian/s-t) ; B0: is the magnetic field strength (T) 5

16 In terms of linear frequency (f), the Larmor equation becomes: f = γ B0 (1-2) 2π Where: f : precessional frequency in MHz. γ/2π: is Gyromagnetic Ratio in (MHz/T). B0: is magnetic field strength in (T). The Larmor equation states that the precessional frequency depends on the magnetic field strength and the gyromagnetic ratio (a constant value for each MR active nuclei). Gyromagnetic ratio for selected MR active nuclei are listed in Table MR Signal Detection and Image Formation Recall that when protons are placed in a strong magnetic field, they tend to align with the magnetic field with slight majority aligned with direction of the magnetic field ( low energy state) resulting in a net magnetization vector pointing in the direction of the magnetic field (Mz). When an external radiofrequency pulse (precisely matching the Larmor frequency of protons) is applied to the sampling volume, the net magnetization will be tipped away from the z direction at an angle determined by the magnitude and the duration of the RF pulse. If the RF pulse is left on long enough to flip the net magnetization vector entirely into the x-y plane, a 90 O flip angle is 6

17 achieved. Once the 90 O RF pulse is switched off, the transverse net magnetization vector will rotate on the x-y plane and after some time delay will realign with the main magnetic field in the z direction due to a process known as relaxation. During relaxation, the hydrogen atoms release the absorbed RF energy in two simultaneous, but independent process named T1 and T2 relaxation. Relaxation results in regrowth of the net magnetization in the z direction (T1 recovery) and decay of transverse magnetization (T2 decay).the rotation of the transverse magnetization in the x-y plane will induce a voltage in a receiver RF coil as stated by Faraday s law of electromagnetic induction. The voltage induced in the RF coil is subsequently used as the MR signal (Westbrook et al., 2011). The collected data from MR signals are stored in a frequency domain k-space matrix. The horizontal axis of the k-space represents the frequency information and the vertical axis represents the phase information. After the k-space is filled, an inverse Fourier transform is applied to the data to produce a spatial domain representation of the data (Bushberg, 2012). 1.2 Primary Components of MRI Scanner There are four main components of MRI unit: A strong magnet, gradient coils, radiofrequency system, and computer system. The magnet is the main and the most expensive part of the MRI system. The strength of magnetic field is measured in tesla (T) (1.0T = 10,000 Gauss). Typical magnetic field strength for clinical applications ranges from 0.3T to 3T with 1.5T and 3T being the most commonly used field strength. Another important component of the MRI scanner is the gradient coils. To obtain an MR image with spatial information, the MR signal from different physical positions must be 7

18 localized. Gradient coils are used to produces a slight change in the magnetic field strength across the imaging volume which results in slight change in the precessional frequency of protons across the sampling volume as stated by the Larmor equation. Location of MR signals in a two dimensional image is determined by the frequency and phase of the precessional protons. Figure 1.3 shows an example of signal localization using the read out gradient. Figure 1.3: signal localization using read out gradient. The read out gradient induces a change in the proton precessional frequency (ω) depending on the spatial location within the sampling volume. Taken from Mark, B., Richard, S., & Thomas, K. N. (2004). MRI: Basic Principles and Applications, 3rd edition. The radiofrequency system is composed of a radiofrequency transmitter (to generate radiofrequency pulses tuned at the Larmor frequency) and a sensitive radiofrequency receiver coil to detect MR signals. MRI utilizes different types of RF coils depending on the anatomical area of interest. Volume coils surround the entire imaging tissue and produce a uniform signal intensity and SNR over the entire imaging volume. Volume coils can provide large FOV and are suitable for brain, knee, and total body imaging, however, this type of coils produces images with relatively low SNR compared with other types of coils such as surface coils. Surface coils, on the other hand, are placed adjacent to the imaging tissue and produce high SNR when imaging tissues near the surface of the coil. This type of coils is used to image structures near the surface of the patient such as the spine. Surface coils are usually small and have limited FOV. 8

19 Another type of MRI coil are phased array coils. This type of coil consists of multiple independent sub-coils that offer the high SNR of surface coils while achieving the large FOV of the volume coils. Phased array coils can be configured as a surface coil as well as volume coil. Finally, a powerful computer system is needed to control and coordinate the functionality of MRI unit components and process the MRI data. 1.3 Quality Control for MRI System Image quality, for medical purpose, can be defined as how well the image conveys anatomical or functional information to the interpreting physician (Bushberg, 2012). Like any other medical imaging modality, MRI system needs to be checked periodically to ensure the scanner works properly and to detect any degeneration of image quality that can adversely affect the clinical decision. Recommendations and protocols for image quality assurance have been established by several organization such as the American College of Radiology (ACR: MRI Quality Control Manual, 2015), American Association of Physics in Medicine (AAPM report No.100, 2010), and the National Electrical Manufacturers Association (NEMAMS ). A comprehensive MRI quality program has been established by The American College of Radiology (ACR) providing a set of standardized quality control tests and performance criteria to ensure a high level of performance. The image quality portion of this program includes various daily, weekly, and annual assessments of the MRI scanner. The recommended tests include, but are not limited to, slice thickness and slice position accuracy, low and high contrast resolution, magnetic field homogeneity, image uniformity, and signal to noise ratio (ACR: MRI Quality Control Manual, 2015). 9

20 1.3.1 Signal to Nose Ratio Signal to noise ratio (SNR) is an important factor in describing the image quality of MRI scanners. It is often used to evaluate the performance of MRI systems, compare and evaluate pulse sequences and imaging protocols, and to monitor the system performance for ongoing quality control program purposes. (Dietrich, Raya, Reeder, Reiser, & Schoenberg, 2007; Murphy et al., 1993; Redpath, 1998). Since the contrast sensitivity of the MR image is limited by SNR (Bushberg, 2012), degeneration of SNR impairs the detectability of low contrast objects. In such case, lesions with signal intensities close to the surrounding tissue can be difficult to be identified by interpreting physician and can result in inaccurate diagnosis. Thus, it is important to have a reliable and accurate method to evaluate SNR in MR imaging to detect a small degeneration in SNR before it adversely impact the clinical decisions. As the name implies, SNR is the ratio of the signal relative to the noise. MR signal is proportional to the amplitude of the voltage induced in the receiver coil while the noise can be random and/ or nonrandom. Random noise is present in every image and is caused by the electrical fluctuation of the MR system and the presence of the patient in magnet. Nonrandom or structural noise arises from system malfunction or some type of motion artifacts such as ghosting artifact. Several factors affect SNR in MR image which include, the main magnetic field strength, pulse sequence parameters (TR, TE, and flip angle), scanning parameters such as voxel size, the number of signal averages (NEX), receiver bandwidth (BW), image acquisition and reconstruction algorithms, and RF coil type and location relative to the imaging volume (Bushberg, 2012; Westbrook et al., 2011). For instance, SNR can be maximized by using a large voxel volume and increasing the number of signal averages (NEX), however, increasing the 10

21 NEX increases the scan time while using large voxel volume reduces image spatial resolution. Therefore, scanning parameters are adjusted to produce optimal image quality with reasonable scan time Noise Distribution in Magnitude MR Image In MRI, the acquired complex data in the k-space are assumed to be superimposed by a Gaussian distributed noise. Since Fourier transform- is a linear operation, it will preserve the Gaussian distribution of the noise. In clinical practice, however, magnitude reconstruction algorithm is applied to the complex data to produce the image. Magnitude operation is non-linear and it will alter the noise distribution in the magnitude images to a Rician distribution (Sijbers, den Dekker, Van Audekerke, Verhoye, & Van Dyck, 1998). Noise properties in magnitude images have been studied by Henkelman which shows that image processing used to produce the magnitude image changes the noise distribution in the image and leads to under estimation of the true noise value (σ). In areas with no MR signal(background regions), the true noise value (σ) was found to be equal to the background standard deviation (SD air) divided by (Henkelman, 1985). The correction factor (0.655) arises because the noise distribution in the background region is not Gaussian. Further analysis of the noise distribution shows that the noise distribution in a magnitude MR image does not follow Gaussian distribution; in fact, it follows Rician distribution. This is more prominent for areas with low signal intensities, while for areas of high signal intensities (SNR > 3) the noise distribution approaches Gaussian distribution. It was also shown that in regions with no MR signal (background region), the noise distribution follows a special case of 11

22 Rician distribution known as Rayleigh distribution with probability distribution function given in equation (1-3) (Gudbjartsson & Patz, 1995; Sijbers et al., 1998). Where: S: is the noise signal in background area, σ : is the noise. P(S) = S S2 σ2 e ( 2σ 2 ) (1-3) Gaussian standard deviation or the true noise in the image (σ) can be estimated from the Rayleigh standard deviation (SD air ) using equation (1-4) (Edelstein, Bottomley, & Pfeifer, 1984) which yields the same result presented by Henkelman. SD air = (2 π ) σ (1-4. A) 2 σ = SD(air) / (1-4. B) Constantinides C.D et al showed that the number of channels of the receiver coil influences the noise distribution in the magnitude image and proposed equations to evaluate the noise distribution not only for a single channel coils, but also to include multi-channel phased array coils (Constantinides, Atalar, & McVeigh, 1997). Phased array coils consist of multiple independent closely positioned sub-coils (channels). Each sub-coil receives MR signal from a small area of the imaged structure and the output signal from each sub-coil can be displayed as a composite or combined image. The output signal of the composite image in most MRI systems is computed as the root of the sum-of-sqaures of the signal from each sub-coil. According to Constantinides C.D et al, the composite noise distribution in area with no MR signal 12

23 (background) was shown to be governed by a central chi statistics with probability distribution function given by: 2 P(S) = (σ 2) 2n (n 1)! S2n 1 e ( 2σ 2 ) (1-5) Where S: is the signal in background areas (noise signal), n: is the number coil elements, σ: is the noise. S 2 For a single channel coil (n=1), central chi distribution (1-5) reduces to the Rayleigh distribution function (1-3). To summarize, in magnitude MR image, the noise distribution in regions with noise signal only (background noise) was found to be governed by Rayleigh distribution for a single channel coils while for multi-channel coils it follows the central chi distribution Signal to Noise Ratio Measurements While measuring the MRI signal is somewhat straight forward, estimating the noise is more challenging. A review of literature has shown many different methods to perform SNR measurement; some are based on a single image acquisition and computing the mean signal from the signal producing phantom and estimating the noise from the standard deviation of the background region (i.e. region with no MR signal) (Henkelman, 1985; Kaufman, Kramer, Crooks, & Ortendahl, 1989). Another method is based on acquisition of two identical images and computing the noise from the subtracted image (NEMA MS 1, 2008 & AAPM report No. 100, 13

24 2010). Still another method is based on acquiring an image with no signal producing phantom (pure noise image) and estimating the noise from the pure noise image (NEMA MS 1, 2008). The most commonly used methods for QA purposes found during literature review are discussed below Single Image Method SNR measurement based on single image acquisition is performed by measuring the mean signal (S) from the phantom ROI and estimating the noise (σ) from the standard deviation from an artifact-free background region in the phase encode direction (SDair). Since the reconstruction algorithm used in MRI to produce a magnitude image changes the noise distribution in the reconstructed image, true noise in the image for a single channel coil (σ) = SD(air) The factor (0.655) arises because the noise in the image is not Gaussian distributed (Henkelman, 1985; Kaufman et al., 1989). Since the number of channels influences the noise distribution, different correction factors need to be applied for multi-channel coils. Table 1.3 shows appropriate correction factors for multi-channel coils as suggested by NEMA Standards Publication MS Table 1.3: correction factors for multi-channels coils as a function of channel numbers. Number of channels Correction factor (CF)

25 Signal to noise ratio in this case is given by: SNR = CF S SD(air) (1-6) Although this method is simple to perform, it is sensitive to the presence of structural artifacts in the background region such as ghosting artifacts. Because the single image method assumed the signal in the background region is essentially zero, subtle signal intensities projected in the background region due to image artifacts can lead to artificially elevated SD. As a consequence, the SNR is underestimated (Firbank, Coulthard, Harrison, & Williams, 1999; Kaufman et al., 1989). An additional factor that complicates the SNR measurement with the single image method is the spatial inhomogeneity of the noise in the background region due to noise correlation (Aja-Fernandez & Tristan-Vega, 2012). In addition, for ongoing QC measurements, it is crucial to reproduce the exact same ROIs in terms of size and location for signal and noise measurement. This is especially true for measuring the mean signal for surface coils where the signal falls significantly depending on the distance from the coil surface (AAPM report No. 100, 2010) Dual Image Method SNR measurement based on two image acquisition was proposed by the National Electrical Manufacturers Association (NEMA MS ). This method required an acquisition of two identical phantom scans with minimum time interval between the two scans (less than five minutes). The signal intensity was suggested to be computed as the mean signal (S) from a 15

26 regular phantom ROI covering at least 75% of the phantom. These images are then subtracted from each other and the noise is computed as the standard deviation from the same ROI in the difference image (SD diff). The noise (σ) in the subtracted image is given by SD(diff). Since the noise is evaluated in a region within the phantom which has high signal intensity, this method assumed that the noise has a Gaussian distribution and, therefore, no need for a correction factor. The square root of two, however, arise because the noise is computed from the difference image. 2 SNR in this case is given by: SNR= 2 S SD(diff) (1-7) Since the noise is measured from an area within the phantom, this method is not sensitive to the structural artifacts projected in the background region. However, one of the identified problems of this method is that it is sensitive to temporal system instability which leads to artificially elevated standard deviation in the subtracted image (SD diff) and,hence, underestimation of SNR. In this case, the method is not valid for noise estimation (Sijbers et al., 1998). Additionally, unlike the single image method, the dual image method requires two MRI phantom scans (doubling the scanning time) and images need to be manipulated for SNR analysis which makes it time consuming. Finally, like the single image method, it is imperative for ongoing QA tests to reproduce the same size and location of ROIs from previous test which can be challenging. One can choose either the single or the dual acquisition method to evaluate SNR for the scanner. However, the most important factor when measuring SNR for quality control purposes 16

27 is reproducibility. In addition to scan parameters and phantom position, the method of noise analysis and measurements must be identical from one test to another (ACR MRI QC Manual, 2015). For some coils where the signal drops sharply relative to the distance from the coil, it is imperative to have a reference for SNR measurements such as a photograph showing the location of the ROIs to accurately reproduce them each time (AAPM report No. 100, 2010). To avoid such a burden from individuals performing SNR measurements and to eliminate human variability, an automated method to measure SNR with high accuracy and reproducibility is proposed in this project. 1.4 Automated SNR Measurement. Review of literatures reveals many proposed methods to evaluate the performance of MRI scanners using automated approach to keep the human interaction at minimum and to expedite the process of the measurements. Keeping the human interaction to minimum will mitigate the observer dependence and will avoid the subjectivity of individuals performing quality control measurements. One of the early works toward automating the process of evaluating MRI performance using a custom designed parallel rod test object was proposed to allow a semi-automated assessment of scanners performance in term of image distortion, spatial resolution, slice thickness and separation, and SNR with minimum human interaction (Covell et al., 1986; Hyde, Ellis, Gardner, Zhang, & Carson, 1994). Bourel et al have developed an automated QC software and a designated cylindrical test object aiming to detect degrading MRI system performance in term of SNR, spatial resolution, slice thickness, image uniformity, and geometric distortion (Bourel, Gibon, Coste, Daanen, & Rousseau, 1999). Recent published work for an automated 17

28 MRI quality assurance program to compare and evaluate the MRI systems performance within multicenter studies. In this work, the ACR phantom was utilized to evaluate several scanner performance parameter such as SNR, image uniformity, spatial resolution and linearity (Davids et al., 2014). Although these approaches were investigated to provide an automated assessment of the main parameters related to the performance of MRI scanner including SNR measurement, those methods utilize software that strictly applied to a particular test object which might not be readily available for clinical medical physicists. In addition, the design and size of the test objects seems to be impractical for evaluating SNR for surface coils and special purpose coils such as breast and knee coils. Finally, the noise measurement is based on the standard deviation measured from the background region or from the difference image. In contrast, the automated SNR method proposed in this project is based on a noise distribution model and it can be used to evaluate SNR for a variety of coil configurations using the readily available quality control phantoms provided by the scanner vendor. 1.5 Objective of the study SNR is a fundamental index of image quality in MRI and it provides a sensitive measure to monitor the performance of MRI systems. As part of a quality assurance program, the phantom set up, scan parameters and the method for analyzing SNR should have a high level of reproducibility. The main goal of this project is to develop an automated SNR method using a model-based noise determination to improve the reproducibility of the measurement by eliminating individual s variability when performing such measurements. 18

29 This study has the following specific objectives: 1) Propose a reliable automated method for SNR analysis by developing a robust software to efficiently segment the phantom from the background region and determine the noise by fitting a noise distribution model to the background histogram. 2) Evaluate the robustness of the automated SNR method on a variety of RF coils types and channel numbers. 3) Assess the accuracy and precision of the proposed automated method by comparison with the recommended and commonly used methods for evaluating SNR (single and dual image method). 4) Investigate the impact of the presence of minor background artifacts on the noise determined by the fit. 19

30 Chapter Two Materials and Methods A total of 54 phantom scans were performed on two different MRI systems using six different radiofrequency coils. The acquired images were used to compare three different methods of measuring SNR, namely: single image method, two-image method and the proposed method in this study (automated method). 2.1 Materials MRI Scanners MRI scanners used for this project are a 3.0T GE (Signa Excite HDx, General Electric) and a 1.5T Siemens (Magnetom Espree, Siemens) both are located at The University of Toledo Medical Center. Toledo, OH RF Coils and phantoms A random sampling of three RF coils were chosen for each scanner to represent a variety of coil configurations ( single-channel volume coil, phased-array surface coil, and phased-array volume coil) over a range of channel numbers. For the 3.0T scanner, phantom scans were performed with a single-channel body coil, 8-channel brain coil, and 8-channel breast coil. While for the 1.5T scanner, phantom scans were performed with a spine coil operated with 3 channels, a4-channel shoulder coil, and a 8-channel knee coil. Scans were performed using the quality assurance fluid-filled phantoms provided by the manufacturer. 20

31 2.1.3 Image analysis tools Images were analyzed using the freely available ImageJ software from the National Institute of Health (NIH) which was also utilized to develop a script for the automated SNR measurement (Schneider, Rasband, & Eliceiri, 2012). Useful built-in function utilized in this project are listed in Table 2.1 Table 2.1: ImageJ built in functions utilized in the automated SNR method. Function getstatistic( area, mean) gethistogram ( values, counts, nbins) createselection setthreshold (lower, upper) getselectionbounds (x, y, width, height ) getinfo (DICOM_TAG) makerectangle ( x, y, width, height) makeinverse Fit.doFit ( equation, xpoints, ypoints) Description Returns the area and the mean pixel value of the image or selection. Returns the histogram ( pixel values and number of counts) of the image or selection with a bin width of max pixel value / nbin. Creates a composite selection from a thresholded image Sets the lower and upper threshold values Returns the smallest rectangle that circumscribe the active selection with origin coordinate (x,y) Returns the value of a specified DICOM TAG) Creates a rectangular selection with origin coordinate of (x,y) Invert the active selection (creates an inverse selection). Fits a defined equation to a specified x and y data. 21

32 2.2 Methods Scanning Procedure and Parameters: For the each RF coil used in this project, three identical, consecutive MRI scans were performed in compliance with the ACR MRI QC Manual (ACR:MRI Quality Control Manual, 2015). These scans were repeated three times in three different scanning sessions over a period of two months. Within each scanning session, phantom scans were performed without repositioning the phantom so that the only variable in SNR measurement should be related to the image analysis method such as the size and/ or the location of ROIs. Although care was taken to ensure the same scanning set up for each session, performing the scans on different sessions is important to take into account any possibility of set up changes which may cause some variation in the SNR measurement. It should be noted that after positioning the phantom of interest inside the magnet bore, a period of three minutes was allowed to pass before starting the scan to allow the fluid in the phantom to settle. As recommended by the ACR QC Manual, all phantom scans were performed with a T1 weighted image using a single spin-echo (SE) sequence with repetition time (TR) of 500 ms, echo time (TE) of 20 ms, number of signal averages (NEX) of one, and matrix size of Appropriate field of view (FOV) and slice thickness were chosen according to the RF coil of interest. FOV was chosen to provide a background region, in the frequency encode direction, large enough to obtain adequate sampling of the noise (ACR: MRI Quality Control Manual, 2015). Table (2.2) shows the scanning parameter for each RF coil. 22

33 Table 2.2: Scanning protocols for each RF coil. RF Coil Pulse Matrix Slice TR(ms) TE(ms) NEX FOV(cm) Sequence Size Thickness(mm) GE Body SE GE Brain SE GE Breast SE Siem. Spine SE Siem. Shoulder SE Siem. Knee SE Automated Assessment of Signal to Noise Ratio The proposed method utilizes a script that has been written using ImageJ software to automatically measure the mean signal intensity (S) from an area covering 80% of the phantom and noise measurement by fitting the noise model discussed in section to the signal distribution in the background region (excluding the phase encode direction). The script was developed to perform the following main tasks: 1) automatically segment the phantom from the background region using a histogram-based thresholding technique 2) define signal ROI as an area covering 80% of the phantom and measure the mean signal (S) 3) define the background ROI in the frequency encode direction and extract the signal distribution 4) background signal processing to correct for signal truncation and to smooth the histogram 5) fit the signal distribution to the noise model in equation (1-5) to obtain the noise fit parameter (σ) 6) calculate SNR as (S/σ) and print the result along with some optional parameters. Figure 2.1 shows a flowchart of the automated SNR measurement. 23

34 Number of Counts MRI Phantom Image Segmented Phantom Signal ROI Thresholding Define 80% of the phantom & find mean signal (S) Define background area in frequency encode direction Background ROI SNR Report Fit background signal to noise model & determine the noise (σ) from the fit Calculate SNR as (S/σ) and print the result Bkg. Distribution Noise Model Fit Signal Intensity Figure 2.1: Flowchart showing the main steps in the automated SNR measurement. 24

35 Phantom Segmentation and Defining the Signal ROI The first task in the scrip is to segment the phantom from the background region using a histogram-based threshold technique with threshold value (T) corresponding to a signal intensity at the valley (local minimum) between the phantom and background peaks. Segmentation is a process that subdivides the image into regions with similar properties. A simple method to segment an image can be obtained by using a thresholding technique by which the image is divided into two categories depending on the pixel intensity. Pixels with a certain range of intensity will be assigned to one category, while the rest of the pixels will be assigned to the other category (Russ, 1995). A histogram-based thresholding technique, was found to work well when the ROI has signal intensity different than that of the background region (Sharma & Aggarwal, 2010). Since phantom scans show two distinctive histogram peaks (high phantom signals and low background signals), the resultant histogram will be bimodal (Fig 2.2). Is this case, mode method can be used to segment the image with threshold value (T) corresponding to the pixel value at the local minimum between the two peaks (Glasbey, 1993; Sahoo, Soltani, & Wong, 1988). 25

36 Frequency (arbitrary) Bimodal Distribution Intensity ( arbitrary) Figure 2.2: A schematic illustration of a typical bimodal distribution. In some cases, however, the phantom scan shows a histogram with three peaks separated by two minima due to truncation artifact (Figure.2.3). To circumvent this issue, the valleys in the histogram are sorted and the threshold value (T) is chosen as the pixel value of the second valley (i.e. the valley between the background peak and phantom peak). Additionally, a pop-up window will appear showing the automated threshold value and the user can change the threshold value if needed. The automated threshold value determined by the algorithm was found to be appropriate for all phantom scans performed on this project. 26

37 Frequency (arbitrary) MRI Phantom Histogram Truncation peak Background peak Phantom peak T Intensity ( arbitrary) Figure. 2.3: A schematic illustration of a histogram acquired from MRI phantom scan with a threshold value (T) corresponds to the signal intensity at the local minimum between the background and phantom peaks. Phantom segmentation is accomplished by the following steps: i- Obtaining the intensity histogram of the image using gethistogram function. ii- iii- iv- Finding the peaks in the histogram using findmaxima functions. Finding and sorting the valleys in the histogram. Defining the threshold value (T) as the pixel value at the valley between the background and the phantom peaks. v- Thresholding the image using setthreshold function with threshold value of (T). After segmenting the phantom from the background region, the boundaries of the phantom are defined using the create selection function and the mean signal from an area covering 80% of the selection (phantom) is acquired using the getstatistics function. The mean signal (S) is saved as a variable to be used later in SNR calculation. 27

38 In short, the phantom is segmented from the background using a histogram-based threshold technique with a threshold value corresponding to a pixel value in the valley between the phantom and the background peaks. Then, the mean signal from an area covering 80% of the phantom (S) is obtained and saved to be used in SNR calculation Defining the Background ROI and Extracting the Noise Signal: The next step in the script is to define the background region excluding the phase encode direction. The phase encode must be excluded from noise measurement because it is susceptible to major artifact (ghosting artifact) (ACR MRI QC Manual, 2015 & Henkelman, 1985). In order to accomplish this, the script will obtain the phase encode direction from the image DICOM header using the getinfo function and the background ROI will be defined accordingly as follows: i- If the phase encode direction is column (vertical axis of the image), the script will create a mask covering the phantom (with a buffer to exclude potential artifacts that may occur at the boundaries of the phantom) in the phase encode direction using the makerectangle function (Figure 2.4). 28

39 Phase Encode Direction Phase Encode Direction Figure 2.4: phantom selection in the phase encode direction. The selected rectangle in Figure 2.4 represents the area that needs to be excluded from the noise measurement. To exclude this region, the makeinverse function is used to select the entire image excluding the previous selection as shown in Figure 2.5. Figure 2.5: Background selection for noise measurement ii- If the phase encode direction is row (along the horizontal axis of the image), a horizontal rectangle covering the phantom on the phase encode direction will be 29

40 created using the function makerectangle (Figure 2.6). The selection then will be inverted and the resulting selection is shown in Figure Figure 2.6: Phantom selection in the phase encode direction. Frequency Encode Direction Frequency Encode Direction Figure2.7: background selection for noise measurement. 30

41 After defining the background region, signal distribution (noise signal) from the selected area is obtained from the histogram using the gethistogram function Background Signal Processing After extraction the background signal, the signal is processed to exclude the counts from the truncation area in the image from being included in the noise estimation and to smooth the noise histogram. It is not uncommon to observe signal truncation artifact (dark bands along the outer edges of the FOV) where the signal has artificially been zeroed by the system (Figure 2.8). Figure 2.8: Signal truncation along the edges of FOV. Multiple approaches were investigated to eliminate the signal truncation from the histogram used for noise determination. One approach was to eliminate all the pixels that have zero signal from the background histogram (i.e. assuming the background region has signal intensity greater than zero). Another approach was to approximate the number of pixels that have no signal to an arbitrary number. Yet another approach was to extrapolate the number of pixels that have zero 31

42 Frequency ( Y) signal intensity from the first three points in the noise histogram (x1, x2, and x3). This approximation is done by using linear regression to determine the y-intercept (Figure 2.9). The number of counts from pixels with no signal in the background region is assumed to be equal to the y-intercept value. Background histograms with signal truncation and after extrapolating the number of pixels with signal intensity of zero are shown in Figure x1 x2 x3 Intensity (X) Figure 2.9: Schematic illustration of estimating the number of pixels with zero count. 32

43 The slope of the three points was found using the line function: y= mx + b (2-1) where y: the number of counts (frequency) m: the slope of the line x: is the pixel value (intensity) b : is the y-intercept Since the truncation signal represent one point in the histogram, manipulating this point should have minimum effect on the curve fitting process. Linear regression approach was implemented to estimate the number of zero counts in the background region since it was shown to maintain the trend of the noise histogram Figure 2.10: background signal histogram including signal truncation (left) and after eliminating the truncation artifact (right). 33

44 Frequency Histogram signal smoothing is implemented in the software because it was shown to improve the consistency of the automated method. Smoothing of the background signal is done using a five-point moving average technique by which each point in the signal histogram is replaced with an average of five adjacent points prior to applying the noise model fit. Figure 2.11 shows the background histogram before and after signal smoothing Signal Intensity Original Histogram Smoothed Histogram Figure 2.11: Background histogram before and after signal smoothing The noise signal is smoothed using the following equations: For i=2 to i= n-2: Ss(i) = S(i 2)+S(i 1)+S(i)+S(i+1)+S(i+2) 5 (2-2) Where: n : is the total number of points in the histogram Ss(i) is the i th smoothed point S(i) is the original i th point in the histogram 34

45 For the second point( i=1), there is not enough points to perform the five -point average. Therefore, a three-point smoothing is done as: Ss(1) = S(0)+S(1)+S(2) 3 (2-3) Similarly, for the first point in the histogram (i=0) the smoothing is done as: Ss(0)= S(0)+S(1) 2 (2-4) Since the last two points are located at the tail of the curve, which is almost flat, these points should have no effect on the noise estimation and, therefore, they can be assumed to be zero (O'Haver, 1997) Noise Model Fit and Computing the Image Noise: Recall from equation (1-5) that the probability distribution of the noise in the background region is a function of the signal, the number of channels in the receiver coil, and the noise in the image. The background signal is already obtained in a previous step, but the number of channels is still undefined. It is possible to incorporate a manual input where the user can enter the number of channels. However, a study has shown that the presence of noise correlation in the background region results in a reduction of the effective number of channels of the coil (Aja- Fernandez & Tristan-Vega, 2012). Therefore, to optimize the accuracy of noise estimation and to make the process fully automated, the algorithm will execute an iteration process to find the number of channels (n) that produce the best fit of the noise model to the background noise distribution. In other words, the program will fit the noise model to the background signal histogram using the Fit.doFit function over a range of channel numbers (from n=1 to j ) where j is the number of iterations. In this project, j was chosen to be eight because the maximum 35

46 Frequency number of coil elements utilized in this study was eight channels. After the iteration process, the goodness of the fit is determined by the coefficient of determination (R 2 ) and the number of channels that produces the best fit will be used as the variable (n) for the noise estimation. The curve fitter function uses an iterative procedure by which 1) the parameters of the fitting model are initially estimated, 2) the fitting model is compared with the data points, 3) the initial parameters values are then adjusted to improve the fit, 4) the function will return to step 2. This iterative process is repeated until the best possible fit is achieved (Schneider et al., 2012). Figure 2.11 shows the background histogram and the model fit based on the background signal. Bkg. Distribution Noise Model Fit Signal Intensity Figure 2.11: Background ROI distribution with noise model fit based on background signal. The final step in the program is to obtain the noise parameter (σ) from the optimal fit and calculate SNR calculation as: SNR = S σ (2-5) 36

47 Where S : is the mean signal measured from signal ROI σ: is the noise parameter obtained from the curve fit The result will be printed on the screen with some optional acquisition parameters as shown in Figure Figure 2.12: SNR report with acquisition parameters 37

48 2.2.3 Validation of the Automated SNR Measurement method: To investigate the reliability of the automated SNR method, accuracy and precision were assessed. Precision was assessed by obtaining three consecutive phantom scans under identical scanning conditions from six different radiofrequency coils (section 2.1.2). These scans were repeated three times in three different scanning sessions in reproducible manner. For each radiofrequency coil, the SNR was measured using the automated method and the variation in the measurements from each scan is determined by the coefficient of variation (CV) as: CV = SD μ Where: SD : is the standard deviation of the three SNR measurements μ : is the mean value of the three SNR measurements The CV in the measurements from each scanning session (i.e from consecutive scans without repositioning the phantom) will be referred as intra-session CV and it should reflect only the variation in the image analysis method such as the location and size of ROIs. The CV in the measurements from the three scanning session will be referred as inter-session CV. In addition to the variation in the image analysis method, inter-session CV takes into account a potential variation in the phantom set up and possible slight changes in the scanner performance. Accuracy of the automated methods was tested on 52 phantom scans from six different radiofrequency coils (section 2.1.2). For each image, the automated SNR measurement value is compared with that obtained from the manual SNR methods described by the ACR QC Manual, 38

49 namely, single-image SNR method and two- image SNR method (ACR: MRI Quality Control Manual, 2015). The accuracy of the automated SNR method is reported as the percentage deviation (% Dev) from the manual method as: Deviation % Dev = ( manual SNR value ) 100 As recommended by the ACR manual, SNR measurement using the single-image method is performed by measuring the mean signal from a manually drawn signal ROI covering 80% of the phantom and measuring the standard deviation from an ROI placed in the background region avoiding the phase encode direction and visible artifacts (Figure 2.13). Effort was made to perform the analysis in a reproducible manner. Signal ROI of a regular shaped area was sized to cover 80 ± 2% of the area of the phantom and placed at the center of the phantom. Background measurement is obtained from two identically sized rectangular ROIs covering as large area as possible in the frequency encode direction avoiding signal truncation and visible artifacts. Mean Signal ROI Background ROIs SD 1 SD 2 S Figure 2.13: Single image method for SNR measurement. This method will be referred as (SNR single) and it is computed as: SNR single = CF S SD(air) 39

50 Where S : is the mean signal in the phantom SD(air): is the combined background standard deviation = (SD SD 2 2 )/2 CF: is Gaussian equivalent correction factor For the two-image method, two consecutive identical phantom scans are acquired and the mean signal is measured from two centered ROIs of a regular shape covering 80 ±2% of the area the phantoms. These images are, then, subtracted and the noise is measured as the standard deviation from the same signal ROI in the subtracted image (Figure 2.14). In most cases, the boundaries of the phantom in the difference image are difficult to identify. Therefore, the mean signal ROI is copied and placed in the difference image. Mean Signal ROI1 Mean Signal ROI2 Difference Image ROI S 1 S 2 SD diff Image 1 Image 2 Image1-Image2 Figure 2.14: Two image method for SNR measurement. This method will be referred as (SNR dual) and is computed as: SNR= 2 S SD(diff) 40

51 Where S: is the average of the mean signal in ROI1 and ROI2 SD(diff): is the standard deviation in the difference image 2 : To account for the noise in the difference image Background Artifacts and Noise Estimation To investigate the effect of minor background artifacts in the noise estimation determined by the automated method, structural artifact was intentionally included and excluded in two separate background ROIs. The noise was then estimated using the model-based fit and by the ROI standard deviation. This experiment was performed on spine phantom scan (Figure 2.15) and knee phantom scan (Figure 2.16). For the knee scan, the background ROI was intentionally placed in the phase encode direction to include ghosting artifact. Figure 2.15: Spine phantom scan shows background ROI placement to include structural artifacts (left) and away from visible artifacts (right). 41

52 Figure 2.16: Knee phantom scan shows background ROI placement in the phase encode direction to intentionally include structural artifacts (left) and the proper placement of ROI (right). 42

53 Chapter Three Results Using the methods described in the previous section, 52 phantom scans from two MRI scanners using different radiofrequency coils were evaluated using the three different SNR methods. 3.1 Reproducibility For each radiofrequency coil, the mean SNR value (and standard deviation) and the percentage coefficient of variation (%CV) using the automated SNR method (SNR auto), single image SNR method (SNR single), and two image SNR method (SNR dual) are shown in Tables 3.1, 3.2 and 3.3, respectively. Looking at each scanning session, the deviation in the measurements ranges from 0.07 to 0.86 for SNRauto, 0.31 to 3.80 for SNRsingle, and 0.20 to 6.85 for SNRdual. The average intra-session %CV ranges from 0.23 to 0.34 for SNR auto, 0.63 to 1.28 for SNR single, and 1.98 to 2.18 for SNR dual. Across all scanning sessions, the deviation in the measurements range from 1.27 to 1.83 for SNR auto, 0.85 to 3.10 for SNR single, and 1.13 to 8.58 for SNR dual with an inter-session %CV that on average are 1.48, 1.53, and 4.34 for SNR auto, SNR single, and SNR dual, respectively. 43

54 Table 3.1: Mean SNR and standard deviation for the automated SNR analysis method (SNRauto) Session 1 Session 2 Session 3 All Phantom Scans RF Coil SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV Brain (8CH) (0.41) (0.21) (0.16) (1.83) 1.38 Body (1CH) 81.75(0.15) (0.30) (0.07) (1.27) 1.56 Breast (8CH) (0.17) (0.39) (0.38) (1.43) 0.88 Knee (8CH) (0.31) (0.25) (0.36) (1.68) 1.62 Shoulder (4CH) 83.57(0.44) (0.12) (0.15) (1.42) 1.71 Spine (3CH) 84.15(0.22) (0.20) (0.86) (1.43) 1.74 Average Table 3.2: Mean SNR and standard deviation for the single-image SNR analysis method (SNRsingle) Session 1 Session 2 Session 3 All Phantom Scans RF Coil SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV Brain (8CH) (1.22) (0.76) (0.70) (1.48) 0.98 Body (1CH) 81.93(1.03) (1.06) (0.35) (1.26) 1.56 Breast (8CH) (2.46) (0.78) (3.80) (3.10) 1.91 Knee (8CH) (0.94) (0.47) (1.68) (2.24) 2.12 Shoulder (4CH) 88.06(0.90) (0.57) (1.24) (1.31) 1.50 Spine (3CH) 76.56(0.91) (0.31) (1.01) (0.85) 1.11 Average

55 Table 3.3: Mean SNR and standard deviation for the two-image SNR analysis method (SNRdual) Session 1 Session 2 Session 3 All Phantom Scans RF Coil SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) % CV Brain (8CH) (6.85) (1.47) (3.43) (5.12) 3.88 Body (1CH) 81.59(0.38) (1.05) (0.55) (1.13) 1.38 Breast (8CH) (1.44) (3.26) (5.01) (8.58) 6.28 Knee (8CH) (1.57) (3.99) (0.20) (3.25) 3.34 Shoulder (4CH) 75.93(3.19) (1.84) (1.04) (2.47) 3.18 Spine (3CH) 74.65(0.47) (0.49) (2.73) (5.63) 7.98 Average Accuracy A comparison between the mean SNR measurements based on the automated SNR method, single image method, and dual image method over all phantom scans are shown in Table 3.4 and Figure 3.1. The automated SNR measurements show a percentage deviation range from % to 8.16% and -1.55% to 16.44% relative to the single and the dual image methods, respectively. On average, the automated SNR value is ± 6.98% of the single image method and 7.88 ± 8.45% of the dual image method. 45

56 SNR Table 3.4: A comparison between the three methods of SNR measurement over all phantom scans of different radiofrequency coils (and number of channels). The average percentage deviation is reported ± one standard deviation *. SNR singe SNR dual SNR auto % % Dev. of % Dev. of RF Coil SNR (±SD) % CV SNR (±SD) % CV SNR (±SD) CV SNR single SNR dual Brain (8CH) (1.48) (5.12) (1.83) Body (1CH) 80.62(1.26) (1.13) (1.27) Breast (8CH) (3.10) (8.58) (1.43) Knee (8CH) (2.24) (3.25) (1.68) Shoulder (4CH) 87.62(1.31) (2.47) (1.42) Spine (3CH) 76.14(0.85) (5.63) (1.43) Average ±6.98* 7.88±8.45* SNR Measurements of Different Methods 0 8-Ch Brain 1ch Body 8-Ch Breast 8-Ch Knee 4-Ch Shoulder 3-Ch Spine SNR single SNR dual SNR auto Figure 3.1: SNR measurements of six RF coils from three different SNR methods. 46

57 The goodness of the noise model fit in terms of the mean coefficient of determination (R 2 ) for each RF coils is shown in Table 3.5. In all phantom measurements, the noise model fit yields R 2 value that on average is 0.98 with an average %CV of 0.34%. Table3.5: Mean R 2, standard deviation (SD), and %CV for different RF coils. Averaged Over All Phantom Scans RF Coil R 2 (±SD) %CV Brain (8CH) Body (1CH) Breast (8CH) Knee (8CH) Shoulder (4CH) Spine (3CH) Average

58 3.3 Background Artifacts and Noise Measurement The impact of minor artifacts in the background ROI on the noise estimation obtained by computing the background standard deviation from a manually drawn ROI compared to the model-based noise is presented in Table 3.6. The noise estimation based on the standard deviation of the background region is significantly increased in the presence of artifacts 17.61% and 18.91% for knee coil and spine coil, respectively. The noise determined by the fit in the presence of artifacts increased by 2.71% and 1.59% for the knee coil and the spine coil, respectively. Background histograms and noise model fit with and without artifacts for the knee and spine coils are shown in Figure 3.2 and 3.3, respectively. Table 3.6: Comparison between noise estimation obtained from the background ROI standard deviation and the noise determined by the noise model fit. Knee coil Spine coil Noise estimation from background SD Without With % Difference artifacts artifacts Noise determined by the fit Without With % Difference artifacts artifacts

59 Frequency Frequency Artifact signal Signal Intensity Signal Intensity Figure 3.2: Background histogram for knee coil with artifacts (top) and without artifacts (bottom). The open circles represent the noise signal and solid line shows the model fit. Parameter b represents the noise estimate from the fit. 49

60 Frequency Frequency Signal Intensity Artifact signal Signal Intensity Figure 3.3: Background histogram for spine coil with artifacts (top) and without artifacts (bottom). The open circles represent the noise signal and solid line shows the model fit. Parameter b represents the noise estimate from the fit. 50

61 Table 3.7 shows the effects of the presence of artifacts in the background ROI on the measured SNR using the single image method and the automated SNR method. The reduction in the SNR value based on the single image method in the presence of artifacts is 14.96% and 15.95% for the knee coil and spine coil, respectively. Based on the automated method, the reduction in the measured SNR due to the presence of artifacts is 2.61% and 1.49% for the knee coil and the spine coil, respectively. Table 3.7: SNR based on SNR single and SNR auto in presence and absence of background artifacts. Knee coil Spine coil SNR based on single image method Without With % Difference artifacts artifacts SNR based on automated method Without With % Difference artifacts artifacts

62 Chapter Four Discussion The proposed method for measuring SNR is fully automated and simple to operate; the user only needs to open the image and the macro file (script) on ImageJ software, execute Run Macro from the tool bar and confirm the selected threshold value. The time a medical physicist spent on analyzing an image and trying to reproduce the exact size and locations of the ROIs from previous tests can be significant. In contrast, the automated SNR measurement is obtained in a few seconds. Background signal truncation was eliminated from the histogram used for noise determination by using linear regression to determine the y-intercept and equating the number pixels with zero count to the value of the y-intercept. Background signal smoothing was incorporated in the automated method since it was found to improve the consistency of the measurements. This is especially true for the shoulder coil. During the early work in developing the program, the SNR measurements for this particular coils were inconsistent from one scan to another. Applying the five-point smooth was shown to improve the consistency of the automated method with insignificant effect on the SNR values. Therefore, smoothing was implemented in the software. Figure 4.1 shows background histogram for shoulder coil scan before and after signal smooth. 52

63 Frequency Frequency Signal Intensity Signal Intensity Figure 4.1: background noise histogram for shoulder coil before smooth (top) and after smooth (bottom). Noise determined by the fit for non-smoothed and smoothed histogram are and 17.54, respectively. 53

64 Accuracy of the automated method is demonstrated by comparing the SNR values obtained from the automated method with that obtained by the single and dual image methods described in the ACR manual. Averaged over all phantom measurements, the automated measurements shows a good agreement with the single image method. The deviations in the measurements from the single method were found to be 1.67±6.91%. In term of the dual image method, the deviations in the measurements were 7.88±8.45%. The dual image method was found to be vulnerable to artifacts in the difference image which increase the standard deviation and results in underestimation of SNR (Figure 4.2). These artifacts arise due to temporal instability of the machine which result in miss-registration of the raw data from one scan relative to the other and/or non-cancellation of systemic artifacts (De Wilde, Lunt, & Straughan, 1997; Sijbers et al., 1998). Figure 4.2: difference image artifacts due to temporal instability in MRI data acquisition for breast (left) and spine (right) phantom scan. 54

65 Consistency of the automated SNR measurement was assessed with repeated phantom scans with and without repositioning of the phantom. The variation in the SNR measurements without repositioning of the phantom attributed only to the methodology of the image analysis is reported as intra-session %CV. The variation in the measurement with repositioning of the phantom is reported as inter-session %CV which takes into account the potential variation of phantom set up and possible change in the scanner performance. Considered together, the results show that the automated method has a higher level of reproducibility compared to the manual methods ( SNR single and SNR dual). The impact of minor background artifacts on the noise estimation obtained by the automated method was investigated by intentionally including structural artifacts in the background ROI and compared to the effect that these artifacts have on the SNR obtained with the single-image method. The results show that the automated SNR method is less sensitive to the presence of artifacts compared to the single image method. Unlike the single image method which determines noise from the background ROI standard deviation, the noise in the automated method is determined by fitting the background signal to a noise model making it relatively insensitive to minor artifacts. For a single channel coil, the histogram of the background noise shows asymmetrical intensity distribution as shown in Figure 4.3. The noise distribution in the background region shows a good fit to Rayleigh distribution function. For multi-channel coils, background noise distribution is more symmetrical than that for single channel coil (Figure 4.4). The background noise distribution for multi-channel coils fits well with central chi distribution function. These 55

66 observations show that the background noise distributions in magnitude MR images are in agreement with the noise models discussed in section Background Noise Histogram CH Body Coil Figure 4.3: Background noise distribution for single channel coil. The solid line shows Rayleigh distribution based on the noise signal Background Noise Histogram 3CH Spine Coil Figure 4.4: Background noise distribution for three-channel coil. The solid line shows central chi distribution based on the noise signal 56

67 Chapter 5 Conclusions An automated method for SNR analysis using model-based noise determination was developed utilizing the freely available ImageJ software. The SNRauto demonstrated a higher level of reproducibility than the traditional single- and dual-image methods described in the ACR Manual The SNRauto results have good agreement with those obtained by the traditional single- and dual-image methods The noise model fits well with the background noise histogram for a variety of coil types and channel numbers. The SNRauto was proven to be relatively insensitive to the presence of minor artifacts in background region The developed program is fully automated, user friendly, time-efficient and can be a valuable tool to medical physicists who routinely perform MRI annual QC testing. 57

68 References American Association of Physicists in Medicine. AAPM Report No. 100: Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities. College Park, MD: AAPM;2010. American College of Radiology (ACR). MRI Quality Control Manual Aja-Fernandez, S., & Tristan-Vega, A. (2012). Influence of noise correlation in multiple-coil statistical models with sum of squares reconstruction. Magn Reson Med, 67(2), doi: /mrm Bourel, P., Gibon, D., Coste, E., Daanen, V., & Rousseau, J. (1999). Automatic quality assessment protocol for MRI equipment. Med Phys, 26(12), doi: / Bushberg, J. T. (2012). The essential physics of medical imaging: Philadelphia : Wolters Kluwer Health/Lippincott Williams & Wilkins, c rd ed. Constantinides, C. D., Atalar, E., & McVeigh, E. R. (1997). Signal-to-noise measurements in magnitude images from NMR phased arrays. Magnetic Resonance in Medicine, 38(5), doi: /mrm Covell, M. M., Hearshen, D. O., Carson, P. L., Chenevert, T. P., Shreve, P., Aisen, A. M.,... Martel, W. (1986). Automated analysis of multiple performance characteristics in magnetic resonance imaging systems. Med Phys, 13(6), doi: / Davids, M., Zollner, F. G., Ruttorf, M., Nees, F., Flor, H., Schumann, G., & Schad, L. R. (2014). Fullyautomated quality assurance in multi-center studies using MRI phantom measurements. Magn Reson Imaging, 32(6), doi: /j.mri De Wilde, J. P., Lunt, J. A., & Straughan, K. (1997). Information in magnetic resonance images: evaluation of signal, noise and contrast. Med Biol Eng Comput, 35(3),

69 Dietrich, O., Raya, J. G., Reeder, S. B., Reiser, M. F., & Schoenberg, S. O. (2007). Measurement of signalto-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging, 26(2), doi: /jmri Edelstein, W. A., Bottomley, P. A., & Pfeifer, L. M. (1984). A signal to noise calibration procedure for NMR imaging systems. Medical Physics, 11(2), doi:doi: Firbank, M. J., Coulthard, A., Harrison, R. M., & Williams, E. D. (1999). A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol, 44(12), N Glasbey, C. A. (1993). An Analysis of Histogram-Based Thresholding Algorithms. CVGIP: Graphical Models and Image Processing, 55(6), doi: Gudbjartsson, H., & Patz, S. (1995). THE RICIAN DISTRIBUTION OF NOISY MRI DATA. Magnetic Resonance in Medicine, 34(6), doi: /mrm Henkelman, R. M. (1985). MEASUREMENT OF SIGNAL INTENSITIES IN THE PRESENCE OF NOISE IN MR IMAGES. Medical Physics, 12(2), doi: / Hyde, R. J., Ellis, J. H., Gardner, E. A., Zhang, Y., & Carson, P. L. (1994). MRI scanner variability studies using a semi-automated analysis system. Magn Reson Imaging, 12(7), Kaufman, L., Kramer, D. M., Crooks, L. E., & Ortendahl, D. A. (1989). MEASURING SIGNAL-TO-NOISE RATIOS IN MR IMAGING. Radiology, 173(1), Mark, B., Richard, S., & Thomas, K. N. (2004). MRI: Basic Principles and Applications, 3rd edition. Medical Physics, 31(1), 170. Murphy, B. W., Carson, P. L., Ellis, J. H., Zhang, Y. T., Hyde, R. J., & Chenevert, T. L. (1993). Signal-to-noise measures for magnetic resonance imagers. Magn Reson Imaging, 11(3), O'Haver, T. (1997). A Pragmatic Introduction to Signal Processing. 59

70 Redpath, T. W. (1998). Signal-to-noise ratio in MRI. Br J Radiol, 71(847), doi: /bjr Russ, J. C. (1995). The image processing handbook (2nd ed.). Boca Raton: CRC Press. Sahoo, P. K., Soltani, S., & Wong, A. K. C. (1988). A survey of thresholding techniques. Computer Vision, Graphics, and Image Processing, 41(2), doi: Schneider, C. A., Rasband, W. S., & Eliceiri, K. W. (2012). NIH Image to ImageJ: 25 years of image analysis. Nat Methods, 9(7), Sharma, N., & Aggarwal, L. M. (2010). Automated medical image segmentation techniques. Journal of Medical Physics / Association of Medical Physicists of India, 35(1), doi: / Sijbers, J., den Dekker, A. J., Van Audekerke, J., Verhoye, M., & Van Dyck, D. (1998). Estimation of the noise in magnitude MR images. Magn Reson Imaging, 16(1), Westbrook, C., Roth, C. K., & Talbot, J. (2011). MRI in practice: Chichester, West Sussex ; Malden, Mass. : Wiley-Blackwell, c th ed. 60

71 Appendix A Phantom Set Up Body phantom Brain phantom Spine phantom Knee phantom Figure A-1: Photographs show reference position for phantom set up 61

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia M R I Physics Course Magnetic Resonance Imaging Spatial

More information

Background (~EE369B)

Background (~EE369B) Background (~EE369B) Magnetic Resonance Imaging D. Nishimura Overview of NMR Hardware Image formation and k-space Excitation k-space Signals and contrast Signal-to-Noise Ratio (SNR) Pulse Sequences 13

More information

MAGNETIC RESONANCE IMAGING

MAGNETIC RESONANCE IMAGING CSEE 4620 Homework 3 Fall 2018 MAGNETIC RESONANCE IMAGING 1. THE PRIMARY MAGNET Magnetic resonance imaging requires a very strong static magnetic field to align the nuclei. Modern MRI scanners require

More information

Image Quality/Artifacts Frequency (MHz)

Image Quality/Artifacts Frequency (MHz) The Larmor Relation 84 Image Quality/Artifacts (MHz) 42 ω = γ X B = 2πf 84 0.0 1.0 2.0 Magnetic Field (Tesla) 1 A 1D Image Magnetic Field Gradients Magnet Field Strength Field Strength / Gradient Coil

More information

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005

MARP. MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 ACR MRI accreditation program MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR keener@marpinc.com MARP Medical & Radiation Physics,

More information

H 2 O and fat imaging

H 2 O and fat imaging H 2 O and fat imaging Xu Feng Outline Introduction benefit from the separation of water and fat imaging Chemical Shift definition of chemical shift origin of chemical shift equations of chemical shift

More information

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging

NEMA Standards Publication MS (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging NEMA Standards Publication MS 1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging Published by: National Electrical Manufacturers Association 1300 North

More information

1 Introduction. 2 The basic principles of NMR

1 Introduction. 2 The basic principles of NMR 1 Introduction Since 1977 when the first clinical MRI scanner was patented nuclear magnetic resonance imaging is increasingly being used for medical diagnosis and in scientific research and application

More information

Pulse Sequence Design and Image Procedures

Pulse Sequence Design and Image Procedures Pulse Sequence Design and Image Procedures 1 Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant 2 A pulse sequence is a timing diagram designed with a series of RF pulses, gradients switching, and signal readout

More information

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 - COMPUTERIZED IMAGING Section I: Chapter 2 RADT 3463 Computerized Imaging 1 SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS RADT 3463 COMPUTERIZED IMAGING Section I: Chapter 2 RADT

More information

HETERONUCLEAR IMAGING. Topics to be Discussed:

HETERONUCLEAR IMAGING. Topics to be Discussed: HETERONUCLEAR IMAGING BioE-594 Advanced MRI By:- Rajitha Mullapudi 04/06/2006 Topics to be Discussed: What is heteronuclear imaging. Comparing the hardware of MRI and heteronuclear imaging. Clinical applications

More information

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai 1. Introduction NMR or nuclear magnetic resonance occurs when nuclei are placed in a magnetic field. It is a physical

More information

Pulse Sequence Design Made Easier

Pulse Sequence Design Made Easier Pulse Sequence Design Made Easier Gregory L. Wheeler, BSRT(R)(MR) MRI Consultant gurumri@gmail.com 1 2 Pulse Sequences generally have the following characteristics: An RF line characterizing RF Pulse applications

More information

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010 (N)MR Imaging Lab Course Script FMP PhD Autumn School Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder Date: November 3rd, 2010 1 Purpose: Understanding the basic principles of MR imaging

More information

RF and Electronic Design Perspective on Ultra-High Field MRI systems

RF and Electronic Design Perspective on Ultra-High Field MRI systems RF and Electronic Design Perspective on Ultra-High Field MRI systems A DISSERTATION SUBMITTED TO THE FACULTY OF THE GRADUATE SCHOOL OF THE UNIVERSITY OF MINNESOTA BY SUNG-MIN SOHN IN PARTIAL FULFILLMENT

More information

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images 10. Phase Cycling and Pulsed Field Gradients 10.1 Introduction to Phase Cycling - Quadrature images The selection of coherence transfer pathways (CTP) by phase cycling or PFGs is the tool that allows the

More information

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging JOURNAL OF MAGNETIC RESONANCE IMAGING 20:1046 1051 (2004) Technical Note The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging James W. Goldfarb, PhD* Purpose: To describe a known (but undocumented)

More information

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel.

BACKGROUND. ** 78% of all MRI scanners have Image Quality problems. *** *** 25% of all Multi-Channel RF coils have at least one bad channel. Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period Moriel NessAiver, Ph.D. - Simply Physics - Baltimore, MD moriel@simplyphysics.com

More information

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements MR Data for Treatment Planning: Issues, Protocol Optimization, and s (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements TG-117 Use of MRI Data in Treatment Planning

More information

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo Gradient Spoiling Average balanced SSFP magnetization Reduce sensitivity to off-resonance FFE, FISP, GRASS, GRE, FAST, Field Echo 1 Gradient-Spoiled Sequence (GRE, FFE, FISP, GRASS) RF TR G z G y G x Signal

More information

Chapter 2. The Physics of Magnetic Resonance Imaging

Chapter 2. The Physics of Magnetic Resonance Imaging Chapter 2. The Physics of Magnetic Resonance Imaging 2.1. Introduction The origins of the Nuclear Magnetic Resonance (NMR) signal and how it is manipulated to form images are the subjects of this chapter.

More information

MR Advance Techniques. Flow Phenomena. Class II

MR Advance Techniques. Flow Phenomena. Class II MR Advance Techniques Flow Phenomena Class II Flow Phenomena In this class we will explore different phenomenona produced from nuclei that move during the acquisition of data. Flowing nuclei exhibit different

More information

A Conceptual Tour of Pulsed NMR*

A Conceptual Tour of Pulsed NMR* A Conceptual Tour of Pulsed NMR* Many nuclei, but not all, possess both a magnetic moment, µ, and an angular momentum, L. Such particles are said to have spin. When the angular momentum and magnetic moment

More information

2014 M.S. Cohen all rights reserved

2014 M.S. Cohen all rights reserved 2014 M.S. Cohen all rights reserved mscohen@g.ucla.edu IMAGE QUALITY / ARTIFACTS SYRINGOMYELIA Source http://gait.aidi.udel.edu/res695/homepage/pd_ortho/educate/clincase/syrsco.htm Surgery is usually recommended

More information

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging Poster No.: C-1125 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. I. Peltonen,

More information

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review MRES 7005 - Fast Imaging Techniques Module 2 Artefacts and Imaging Optimisation for single shot methods Content: Introduction Phase error Phase bandwidth Chemical shift review Chemical shift in pixels

More information

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Dr John Ridgway Leeds Teaching Hospitals NHS Trust, UK Cardiac MR Physics for clinicians: Part I Journal of Cardiovascular Magnetic Resonance 2010, 12:71 http://jcmr-online.com/content/12/1/71

More information

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves Experiment 1 Goal: Examine the effect caused by changing flip angle on image contrast in a simple gradient echo sequence and derive T1-curves. Image

More information

8.2 IMAGE PROCESSING VERSUS IMAGE ANALYSIS Image processing: The collection of routines and

8.2 IMAGE PROCESSING VERSUS IMAGE ANALYSIS Image processing: The collection of routines and 8.1 INTRODUCTION In this chapter, we will study and discuss some fundamental techniques for image processing and image analysis, with a few examples of routines developed for certain purposes. 8.2 IMAGE

More information

MRI SYSTEM COMPONENTS Module One

MRI SYSTEM COMPONENTS Module One MRI SYSTEM COMPONENTS Module One 1 MAIN COMPONENTS Magnet Gradient Coils RF Coils Host Computer / Electronic Support System Operator Console and Display Systems 2 3 4 5 Magnet Components 6 The magnet The

More information

MR Basics: Module 6 Pulse Sequences

MR Basics: Module 6 Pulse Sequences Module 6 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol PHY3902 PHY3904 Nuclear magnetic resonance Laboratory Protocol GETTING STARTED You might be tempted now to put a sample in the probe and try

More information

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging Medical Imaging X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging From: Physics for the IB Diploma Coursebook 6th Edition by Tsokos, Hoeben and Headlee And Higher Level Physics 2 nd Edition

More information

RAD 229: MRI Signals and Sequences

RAD 229: MRI Signals and Sequences RAD 229: MRI Signals and Sequences Brian Hargreaves All notes are on the course website web.stanford.edu/class/rad229 Course Goals Develop Intuition Understand MRI signals Exposure to numerous MRI sequences

More information

2015 Spin echoes and projection imaging

2015 Spin echoes and projection imaging 1. Spin Echoes 1.1 Find f0, transmit amplitudes, and shim settings In order to acquire spin echoes, we first need to find the appropriate scanner settings using the FID GUI. This was all done last week,

More information

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters.

The ACR magnetic resonance accreditation phantom (ACR MRAP) has been designed to examine a broad range of instrument parameters. OVERVIEW OF THE ACR MRI ACCREDITATION PHANTOM Geoffrey D. Clarke, Ph.D. University of Texas Southwestern Medical Center at Dallas email: GEOFFREY.CLARKE@MAIL.SWMED.EDU INTRODUCTION The ACR magnetic resonance

More information

Magnetic Resonance Imaging Principles, Methods, and Techniques

Magnetic Resonance Imaging Principles, Methods, and Techniques Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls Jr., Emory University Publisher: Medical Physics Publishing Corporation Publication Place: Madison, Wisconsin Publication Date:

More information

MR Basics: Module 8 Image Quality

MR Basics: Module 8 Image Quality Module 8 Transcript For educational and institutional use. This transcript is licensed for noncommercial, educational inhouse or online educational course use only in educational and corporate institutions.

More information

RF pulse design and the Small Tip Angle Approximation

RF pulse design and the Small Tip Angle Approximation RF pulse design and the Small Tip Angle Approximation Dr Shaihan J Malik Lecturer in Imaging Sciences Division of Imaging Sciences & Biomedical Engineering King s College London shaihan.malik@kcl.ac.uk

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging Principles, Methods, and Techniques Perry Sprawls, Ph.D., FACR, FAAPM, FIOMP Distinguished Emeritus Professor Department of Radiology Emory University Atlanta, Georgia Medical

More information

Novel Concepts for RF Surface Coils with Integrated Receivers

Novel Concepts for RF Surface Coils with Integrated Receivers Novel Concepts for RF Surface Coils with Integrated Receivers by Sonam Tobgay A Thesis Submitted to the Faculty of the WORCESTER POLYTECHNIC INSTITUTE in partial fulfillment of the requirements for the

More information

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP

MR in Tx Planning. Acknowledgements. Outline. Overview MR in RTP MR Data for Treatment Planning and Stereotactic Procedures: Sources of Distortion, Protocol Optimization, and Assessment (Preview of TG117 Report) Debra H. Brinkmann Mayo Clinic, Rochester MN Acknowledgements

More information

EFFECTS OF PHASE AND AMPLITUDE ERRORS ON QAM SYSTEMS WITH ERROR- CONTROL CODING AND SOFT DECISION DECODING

EFFECTS OF PHASE AND AMPLITUDE ERRORS ON QAM SYSTEMS WITH ERROR- CONTROL CODING AND SOFT DECISION DECODING Clemson University TigerPrints All Theses Theses 8-2009 EFFECTS OF PHASE AND AMPLITUDE ERRORS ON QAM SYSTEMS WITH ERROR- CONTROL CODING AND SOFT DECISION DECODING Jason Ellis Clemson University, jellis@clemson.edu

More information

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Half-Pulse Excitation Pulse Design and the Artifact Evaluation Half-Pulse Excitation Pulse Design and the Artifact Evaluation Phillip Cho. INRODUCION A conventional excitation scheme consists of a slice-selective RF excitation followed by a gradient-refocusing interval

More information

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain Available online at www.derpharmachemica.com ISSN 0975-413X CODEN (USA): PCHHAX Der Pharma Chemica, 2016, 8(17):15-20 (http://derpharmachemica.com/archive.html) The role of in increasing resolution in

More information

MRI Metal Artifact Reduction

MRI Metal Artifact Reduction MRI Metal Artifact Reduction PD Dr. med. Reto Sutter University Hospital Balgrist Zurich University of Zurich OUTLINE Is this Patient suitable for MR Imaging? Metal artifact reduction Is this Patient suitable

More information

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 )

EE225E/BIOE265 Spring 2011 Principles of MRI. Assignment 6 Solutions. (y 0 + vt) dt. 2 y 0T + 3 ) EE225E/BIOE265 Spring 211 Principles of MRI Miki Lustig Handout Assignment 6 Solutions 1. Nishimura 6.7 (Thanks Galen!) a) After the 9 y pulse, the spin is in the ˆx direction (using left-handed rotations).

More information

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

TimTX TrueShape.  The parallel transmit architecture of the future. Answers for life. www.siemens.com/trueshape TimTX TrueShape The parallel transmit architecture of the future. The product/feature (mentioned herein) is not commercially available. Due to regulatory reasons its future availability

More information

Acceptance Testing of a Digital Breast Tomosynthesis Unit

Acceptance Testing of a Digital Breast Tomosynthesis Unit Acceptance Testing of a Digital Breast Tomosynthesis Unit 2012 AAPM Spring Clinical Meeting Jessica Clements, M.S., DABR Objectives Review of technology and clinical advantages Acceptance Testing Procedures

More information

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents

Philips Site Yearly Performance Evaluation Philips Openview 16-Jan-08. Table of Contents Philips Site Yearly Performance Evaluation Philips Openview 6-Jan-8 Table of Contents Summary and Signature Page 2 Specific Comments 3 Site Information 4 Equipment Information 4 Table Position Accuracy

More information

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of

3D-Printed Microstrip Resonators for 4.7T MRI. Saeed Javidmehr. A thesis submitted in partial fulfillment of the requirements for the degree of 3D-Printed Microstrip Resonators for 4.7T MRI by Saeed Javidmehr A thesis submitted in partial fulfillment of the requirements for the degree of Master of Science in Electromagnetics and Microwaves Department

More information

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging

Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Alae Tracker: Tracking of the Nasal Walls in MR-Imaging Katharina Breininger 1, Andreas K. Maier 1, Christoph Forman 1, Wilhelm Flatz 2, Catalina Meßmer 3, Maria Schuster 3 1 Pattern Recognition Lab, Friedrich-Alexander-Universität

More information

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance

Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance FDA Workshop April 13, 2010 Standards for Imaging Endpoints in Clinical Trials: Standardization and Optimization of Image Acquisitions: Magnetic Resonance Edward F. Jackson, PhD Professor and Chief, Section

More information

MRI at a Glance. Catherine Westbrook. Blackwell Science

MRI at a Glance. Catherine Westbrook. Blackwell Science MRI at a Glance Catherine Westbrook Blackwell Science MRI at a Glance MRI at a Glance CATHERINE WESTBROOK MSC DCRR CTC Director of Training and Education Lodestone Patient Care Ltd Blackwell Science 2002

More information

Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging

Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging Lecture 7: Basics of magnetic resonance imaging (MRI): one dimensional Fourier imaging Lecture aims to explain: 1. Basic aims of magnetic resonance imaging 2. Signal demodulation in magnetic resonance

More information

SIEMENS MAGNETOM Skyra syngo MR D13

SIEMENS MAGNETOM Skyra syngo MR D13 Page 1 of 12 SIEMENS MAGNETOM Skyra syngo MR D13 \\USER\CIND\StudyProtocols\PTSA\*ep2d_M0Map_p2_TE15 TA:7.9 s PAT:2 Voxel size:2.5 2.5 3.0 mm Rel. SNR:1.00 :epfid Properties Routine Contrast Prio Recon

More information

Correction of the local intensity nonuniformity artifact in high field MRI

Correction of the local intensity nonuniformity artifact in high field MRI Correction of the local intensity nonuniformity artifact in high field MRI Poster No.: C-0346 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Paper S. Kai, S. Kumazawa, H. Yabuuchi, F. Toyofuku;

More information

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC? Slide 1 Multi-Slice CT Artifacts and Quality Control Dianna Cody, Ph.D. Chief, Radiologic Physics UT MD Anderson Cancer Center Houston, TX Slide 2 What are the rules or recommendations for CT QC? AAPM

More information

PULSED/CW NUCLEAR MAGNETIC RESONANCE

PULSED/CW NUCLEAR MAGNETIC RESONANCE PULSED/CW NUCLEAR MAGNETIC RESONANCE The Second Generation of TeachSpin s Classic Explore NMR for both Hydrogen (at 21 MHz) and Fluorine Nuclei Magnetic Field Stabilized to 1 part in 2 million Homogenize

More information

Pulse Sequences: Rapid Gradient Echo

Pulse Sequences: Rapid Gradient Echo Pulse Sequences: Rapid Gradient Echo M229 Advanced Topics in MRI Holden H. Wu, Ph.D. 2018.04.17 Department of Radiological Sciences David Geffen School of Medicine at UCLA Class Business Office hours -

More information

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon,

More information

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Advanced MSK MRI Protocols at 3.0T Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University Outline Why High Field for MSK? SNR and Relaxation Times Technical Issues Example

More information

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical 3T Unlimited ipat on MAGNETOM Allegra The Importance of ipat at 3T s medical ipat on MAGNETOM Allegra The Importance of ipat at 3T The rise of 3T MR imaging Ultra High Field MR (3T) has flourished during

More information

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI Val M. Runge, MD Institute for Diagnostic and Interventional Radiology Clinics for Neuroradiology and Nuclear Medicine University Hospital Zurich

More information

High Field MRI: Technology, Applications, Safety, and Limitations

High Field MRI: Technology, Applications, Safety, and Limitations High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. The University of Texas M. D. Anderson Cancer Center, Houston, TX Introduction The amount of available signal

More information

NMR Basics. Lecture 2

NMR Basics. Lecture 2 NMR Basics Lecture 2 Continuous wave (CW) vs. FT NMR There are two ways of tuning a piano: - key by key and recording each sound (or frequency). - or, kind of brutal, is to hit with a sledgehammer and

More information

A Novel Approach for MRI Image De-noising and Resolution Enhancement

A Novel Approach for MRI Image De-noising and Resolution Enhancement A Novel Approach for MRI Image De-noising and Resolution Enhancement 1 Pravin P. Shetti, 2 Prof. A. P. Patil 1 PG Student, 2 Assistant Professor Department of Electronics Engineering, Dr. J. J. Magdum

More information

Bayesian Estimation of Tumours in Breasts Using Microwave Imaging

Bayesian Estimation of Tumours in Breasts Using Microwave Imaging Bayesian Estimation of Tumours in Breasts Using Microwave Imaging Aleksandar Jeremic 1, Elham Khosrowshahli 2 1 Department of Electrical & Computer Engineering McMaster University, Hamilton, ON, Canada

More information

A Comparative Analysis of Noise Reduction Filters in MRI Images

A Comparative Analysis of Noise Reduction Filters in MRI Images A Comparative Analysis of Noise Reduction Filters in MRI Images Mandeep Kaur 1, Ravneet Kaur 2 1M.tech Student, Dept. of CSE, CT Institute of Technology & Research, Jalandhar, India 2Assistant Professor,

More information

EE469B: Assignment 1 Solutions

EE469B: Assignment 1 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment Solutions Due Thursday Oct 6 Introduction This assignment concerns typical Fourier transform designs of excitation pulses. This includes designing

More information

INSTRUMENTATION FOR PARALLEL MAGNETIC RESONANCE IMAGING. A Dissertation DAVID GERALD BROWN

INSTRUMENTATION FOR PARALLEL MAGNETIC RESONANCE IMAGING. A Dissertation DAVID GERALD BROWN INSTRUMENTATION FOR PARALLEL MAGNETIC RESONANCE IMAGING A Dissertation by DAVID GERALD BROWN Submitted to the Office of Graduate Studies of Texas A&M University in partial fulfillment of the requirements

More information

Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing

Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing Refined Slanted-Edge Measurement for Practical Camera and Scanner Testing Peter D. Burns and Don Williams Eastman Kodak Company Rochester, NY USA Abstract It has been almost five years since the ISO adopted

More information

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging Poster No.: C-0709 Congress: ECR 2014 Type: Scientific

More information

Intermediate and Advanced Labs PHY3802L/PHY4822L

Intermediate and Advanced Labs PHY3802L/PHY4822L Intermediate and Advanced Labs PHY3802L/PHY4822L Torsional Oscillator and Torque Magnetometry Lab manual and related literature The torsional oscillator and torque magnetometry 1. Purpose Study the torsional

More information

Image Processing for feature extraction

Image Processing for feature extraction Image Processing for feature extraction 1 Outline Rationale for image pre-processing Gray-scale transformations Geometric transformations Local preprocessing Reading: Sonka et al 5.1, 5.2, 5.3 2 Image

More information

Image Enhancement in Spatial Domain

Image Enhancement in Spatial Domain Image Enhancement in Spatial Domain 2 Image enhancement is a process, rather a preprocessing step, through which an original image is made suitable for a specific application. The application scenarios

More information

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare

Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare GE Healthcare Designing an MR compatible Time of Flight PET Detector Floris Jansen, PhD, Chief Engineer GE Healthcare There is excitement across the industry regarding the clinical potential of a hybrid

More information

EE469B: Assignment 2 Solutions

EE469B: Assignment 2 Solutions EE469B Fall 26-7 RF Pulse Design for MRI EE469B: Assignment 2 s Due Thursday Oct 3 Introduction This assignment concerns the design of small-tip-angle 2D excitation pulses based on spiral k-space trajectories.

More information

Digital Image Processing. Lecture # 3 Image Enhancement

Digital Image Processing. Lecture # 3 Image Enhancement Digital Image Processing Lecture # 3 Image Enhancement 1 Image Enhancement Image Enhancement 3 Image Enhancement 4 Image Enhancement Process an image so that the result is more suitable than the original

More information

Digital Imaging CT & MR

Digital Imaging CT & MR Digital Imaging CT & MR January 22, 2008 Digital Radiography, CT and MRI generate images in a digital format What is a Digital Image? A digital image is made up of picture elements, pixels row by column

More information

Supplementary Material

Supplementary Material Supplementary Material Orthogonal representation of sound dimensions in the primate midbrain Simon Baumann, Timothy D. Griffiths, Li Sun, Christopher I. Petkov, Alex Thiele & Adrian Rees Methods: Animals

More information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information Ordering Information Please contact us if you have any questions or if you would like a quote or delivery schedule regarding the Catphan phantom. phone 800-525-1190, or 518-692-1190 fax 518-692-3329 mail

More information

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID)

6.S02 MRI Lab Acquire MR signals. 2.1 Free Induction decay (FID) 6.S02 MRI Lab 1 2. Acquire MR signals Connecting to the scanner Connect to VMware on the Lab Macs. Download and extract the following zip file in the MRI Lab dropbox folder: https://www.dropbox.com/s/ga8ga4a0sxwe62e/mit_download.zip

More information

Encoding of inductively measured k-space trajectories in MR raw data

Encoding of inductively measured k-space trajectories in MR raw data Downloaded from orbit.dtu.dk on: Apr 10, 2018 Encoding of inductively measured k-space trajectories in MR raw data Pedersen, Jan Ole; Hanson, Christian G.; Xue, Rong; Hanson, Lars G. Publication date:

More information

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

PULSED NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 (revised, 2/12/07) PULSED NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706 Abstract A pulsed nuclear magnetic resonance technique (spin-echo)

More information

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs

MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs MRI imaging in neuroscience Dr. Thom Oostendorp Lab class: 2 hrs 1 Introduction In tomographic imaging techniques, such as MRI, a certain tissue property within a slice is imaged. For each voxel (volume

More information

DIGITAL IMAGE PROCESSING Quiz exercises preparation for the midterm exam

DIGITAL IMAGE PROCESSING Quiz exercises preparation for the midterm exam DIGITAL IMAGE PROCESSING Quiz exercises preparation for the midterm exam In the following set of questions, there are, possibly, multiple correct answers (1, 2, 3 or 4). Mark the answers you consider correct.

More information

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations, Water suppression n a typical biological sample the concentration of the solute is 1 mm or less. n many situations, the signals of interest are those of amide protons that exchange with the solvent water.

More information

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. Works-in-Progress package Version 1.2 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD

More information

1.Discuss the frequency domain techniques of image enhancement in detail.

1.Discuss the frequency domain techniques of image enhancement in detail. 1.Discuss the frequency domain techniques of image enhancement in detail. Enhancement In Frequency Domain: The frequency domain methods of image enhancement are based on convolution theorem. This is represented

More information

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE Rev. 1.0 DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE DoseLab users may reference the following instructions to perform Tomotherapy Quality Assurance tests as recommended

More information

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003 Works-in-Progress package Version 1.0 For the Installation and User s Guide NUMARIS/4VA21B January 22, 2003 Section of Medical Physics, University Hospital Freiburg, Germany Contact: Klaus Scheffler PhD,

More information

New developments in near-field acoustic holography

New developments in near-field acoustic holography Please leave this heading unchanged! New developments in near-field acoustic holography N.B. Roozen*, A.C. Geerlings, B.T. Verhaar, T. Vliegenthart. Philips Applied Technologies, High Tech Campus 7, 5656

More information

New Features of IEEE Std Digitizing Waveform Recorders

New Features of IEEE Std Digitizing Waveform Recorders New Features of IEEE Std 1057-2007 Digitizing Waveform Recorders William B. Boyer 1, Thomas E. Linnenbrink 2, Jerome Blair 3, 1 Chair, Subcommittee on Digital Waveform Recorders Sandia National Laboratories

More information

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging Award: Magna Cum Laude Poster No.: C-1988 Congress: ECR 2014 Type: Scientific Exhibit Authors: T. Hamaguchi,

More information

An Efficient Color Image Segmentation using Edge Detection and Thresholding Methods

An Efficient Color Image Segmentation using Edge Detection and Thresholding Methods 19 An Efficient Color Image Segmentation using Edge Detection and Thresholding Methods T.Arunachalam* Post Graduate Student, P.G. Dept. of Computer Science, Govt Arts College, Melur - 625 106 Email-Arunac682@gmail.com

More information

The development of the RF-pulse for the low level SAR used by the MRI.

The development of the RF-pulse for the low level SAR used by the MRI. The development of the RF-pulse for the low level SAR used by the MRI. Kojiro Yamaguchi a*, Eizo Umezawa a, Sachiko Ueoku b, Kazuhiro Katada c a Faculty of radiological technology, School of Health Science,

More information

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter

A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter A Study On Preprocessing A Mammogram Image Using Adaptive Median Filter Dr.K.Meenakshi Sundaram 1, D.Sasikala 2, P.Aarthi Rani 3 Associate Professor, Department of Computer Science, Erode Arts and Science

More information

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb.

H Micro-Imaging. Tuning and Matching. i. Open any 1H data set and type wobb. - 1-1 H Micro-Imaging The NMR-specific properties of the objects are visualized as multidimensional images. Translational motion can be observed and spectroscopic information can be spatially resolved.

More information

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging Feng Liu, Joe Li, Ian Gregg, Nick Shuley and Stuart Crozier School of Information Technology and Electrical Engineering,

More information